Fluke 8060, 8062a Specification

Fluke 8060/8062A Technical Specifications
VDC
Range and Resolution: 200.00 mV, 2.0000V, 20.000V,
200.00V, 1000.0V 8060A: 200.00 mV, 2.0000V, 20.000V, 200.00V,
1000.0V
8062A: ±(0.05%+2)
VAC (True-RMS, AC Coupled)
dBm, 600Ohms reference
ADC
AAC (True-RMS) Range and Resolution: 200.00 µA, 2.0000 mA, 20.000
Resistance
Frequency
Range and Resolution: 200.00 mV, 2.0000V, 20.000V,
200.00V, 750.0V
8060A: ±(0.2%+10) 8062A: ±(0.5%+10)
Range and Resolution: -50.00 - +60.00 dBm 8060A: ±(0.15 dB) 8062A: NA
Range and Resolution: 200.00 µA, 2.0000 mA, 20.000
mA, 200.00 mA, 2000.0 mA
8060A: ±(0.2%+2) 8062A: ±(0.3%+2)
mA, 200.00 mA, 2000.0 mA
8060A: ±(0.75%+10) 8062A: ±(0.75%+10)
Range and Resolution: 200.00Ohms, 2.0000 kOhms,
20.000 kOhms, 200.00 kOhms, 300.0 MOhms
8060A: ±(0.07%+2) 8062A: ±(0.1%+2)
Range and Resolution: 200.00 Hz, 2.0000 kHz, 20.000
kHz, 200.00 kHz
8060A: ±(0.05%+1) 8062A: NA
Conductance
Range and Resolution: 2000.0 nS (equivalent to 500
kOhms)
8060A: ±(0.5%+20) 8062A: NA
General Specifications
Battery Life
Weight
Size
One Year Warranty
170 hours typical (9V alkaline)
0.41 kg (0.90 lb)
180 mm L x 86 mm W x 45 mm D (7.1" L x 3.4" W x 1.8" D)
One year guaranteed calibration
Loading...