To proceed directly to "Quick Start", go to Section
2.4.
To proceed directly to "Setup Mode", go to Section
2.5.9.
The Victoreen NERO™ mAx Model 8000, Non-invasive Evaluator of Radiation Output, uses an
innovative system of menus and softkeys to provide an intuitive, user friendly operating environment. All
measurement modes and options are displayed on the NERO mAx’s LCD and all functions are controlled
by the 5 softkeys beneath the display and the 3 keys to the right of the display.
NOTE
The NERO mAx consists of the NERO mAx control console, detector, detector cable, filter slides, AC
adapter, HVL plates, manual, Microsoft® Excel Add-in and carrying case.
The NERO mAx control console is compact and easy to use. The sophisticated electronics necessary to
provide highly accurate, reproducible measurements while maintaining an intuitive, user friendly operating
system are in the NERO mAx control console. The NERO mAx’s rechargeable battery is also housed in
the control console. The front panel of the control console contains a backlit 240 x 64 pixel, dot matrix
LCD display and eight push buttons. Connectors for power input, RS-232, printer, scope output and the
NERO mAx detector are located on the control console’s rear panel.
1-1
Victoreen 8000
Operators Manual
Figure 1-1.Control Console Front Panel
Figure 1-2. Control Console Rear Panel
The NERO mAx detector contains sensors for simultaneously measuring kV, exposure or rate and
invasive mA or mAs. Solid-state detectors are used to measure kV. An ion chamber, located in the top
of the detector, is used for exposure/rate measurements. In addition, connectors for an external ion
chamber are provided on the rear panel of the 8000 detector. The NERO mAx detector’s interface
connector is also located on the detector’s rear panel. The front panel has a keyed opening for the model
8000-filter slides and a connector for mAs leads.
1-2
General Information
Product Description
The NERO mAx filter cards contain the various filters needed to accurately measure kilovoltage. Each
filter card is coded so that the NERO mAx “knows” which filter is in use and its position. The NERO mAx
also verifies that the filter card is valid for the selected measurement mode. In addition, the filter cards
are keyed so that they may only be inserted one way. The W/Al filter card is labeled with the kVp ranges
for which it is calibrated. The Mammo filter card is labeled for the x-ray tube targets for which it is
calibrated.
The serial numbers of the NERO mAx control console unit, detector and filter cards must be matched in
order to obtain accurate results. Since the control console unit, detector and filter cards are calibrated
together, they
must be used together for accurate measurements.
1
1.2 New Features of Firmware Release 2.3
This release adds several new features to the Radiographic mode that give the NERO mAx greater
flexibility when making kV and exposure time measurements on all types of radiographic and dental x-ray
machines.
1. The %kV setting provides more time measurement options:
• Time measurements from 90%, 80%, and 75% of the peak kV.
• Pulse counting and zero crossing settings for single-phase generator time measurements.
2. User settable measurement delay allows:
• kV overshoot portion of waveform to be excluded from kV analysis.
• Exclusion of x-ray generator preheat pulses from kV and time analysis.
More information on these new features may be found in “Using a Measurement Delay” and “Using %kV
and Exposure Time Measurements” at the end of Section 2.5.1--Radio Mode.
Accuracy: ± 2%
Reproducibility: ± 1% or 0.2 mAs
Range: 1 - 1000 mA
HVL
Accuracy: ± 5%
Range: .1 - 99.9 mmAl
Physical
Display: 240 x 60 pixel, super twist LCD w/ccfl backlight
Power: 115 or 230 VAC External Supply. Rechargeable internal batteries supply more than
4 hours of continuous service with overnight charge.
1-4
Size:
Console: 9.00" x 9.12" x 3.25" (22.86 mm x 23.17 mm x 8.26 mm)
Detector: 6.56" x 3.70" x 2.58" (16.66 mm x 9.4 mm x 6.55 mm)
Slides Only: 2.4" x 6.25" x 0.31" (6.1 mm x 15.88 x 0.8 mm)
Operating Conditions:
10° C to 40°C (50° F to 104° F)
Maximum 90% relative humidity (non-condensing)
Weight:
Console: 4 lbs. 9.0 oz. (2.067 kg)
Detector: 1 lb. 10.4 oz. (with slide) (.747 kg)
Slides Only: 2.9 oz. & 3.2 oz. (.090 kg & .094 kg)
HVL Set
2.30 mm, 1.0 mm, 0.3 mm
Calibration
W/Al calibrated with 4.5 millimeters of Aluminum total filtration
Mo/Mo calibrated with 30 microns of Molybdenum filtration
Mo/Rh calibrated with 25 microns of Rhodium filtration
Mo/Al calibrated with 1 millimeter of Aluminum filtration
Rh/Rh calibrated with 25 microns of Rhodium filtration
Rh/Al calibrated with 1 millimeter of Aluminum filtration
General Information
Specifications
1
1.4 Battery Operation and Charging
The NERO mAx has an internal rechargeable battery which provides up to four hours of continuous
operation depending upon usage. The NERO mAx draws twice as much power from its from its battery
when it is actively making measurements than it does when in an idle state with its backlight off. To
conserve and extend battery life, the NERO mAx incorporates several power saving features. The
display backlight automatically turns off after one minute of inactivity. The backlight turns on when any
key is pressed or an exposure is made while in any measurement mode. In addition, the NERO mAx
exits from any measurement mode after five minutes of no activity. Pressing the ENTER key restores the
NERO mAx to its previous measurement mode.
The NERO mAx utilizes two levels of protection to assure reliable operation when the battery charge
becomes low. The first level of protection is a warning that is displayed when there is approximately 20
minutes of battery life remaining. During this time, the AC adapter may be plugged into the NERO mAx to
continue operation without interruption. The second level occurs when the battery charge is insufficient to
guarantee proper operation. When this occurs, the low battery indicator in the lower left corner of the
front panel illuminates and the instrument shuts down, becoming inoperable. When this happens, the AC
adapter can be plugged into the NERO mAx to restore operation. The NERO mAx “remembers” what
mode it was in before it shut down and returns to that mode upon power up. Pressing the ENTER key
returns the NERO mAx to its measurement mode.
The battery is charged whenever the NERO mAx is connected to its AC adapter and the adapter is
plugged into a suitable power source. When the power switch is on, the battery is charged at a low rate
that is enough to sustain the battery’s charge. When the power switch is off, the battery is charged at a
high rate.
To fully charge the battery, make sure that the NERO mAx is turned OFF, plug the AC adapter into the
rear of the NERO mAx console and plug the adapter into a suitable power source. The green battery
1-5
Victoreen 8000
Operators Manual
charge indicator on the front panel of the NERO mAx console illuminates when the battery is charging.
When the battery charge indicator is off, the battery is charged. It may take up to 16 hours to fully
recharge a severely discharged battery.
1.5 Printing
All of the exposure results displayed by the NERO mAx may be printed automatically if desired. The
data that is sent to the printer includes the NERO mAx’s mode of operation and selected options, the
current time and date, and the measured data. The NERO mAx uses a standard IBM compatible PC
printer cable.
With the NERO mAx turned off, plug the computer end of a standard IBM compatible PC printer cable into
the printer port at the rear of the NERO mAx then plug the printer end of the cable into the printer and turn
the printer on. Plug the AC adapter into the 8000 if needed (see Section 1.4--Battery Operation and
Charging) and turn the instrument on. From the readout menu, select setup screen and turn the
automatic printing on as follows:
SETUP CLOCK − > OFF
CAL PRINT − > ON
HVL UNITS − > R
EXP AIR −−− > 20.5 C 734 mmHg
CT EXP DATE −−> Oct. 10, 1996
AMSE TIME −−> 10 30 45
MODE SELECT ON/OFF
From the SETUP screen, use the SELECT softkey (under column 1) to select PRINT. When PRINT is
selected, the print selection blinks and a highlight (reverse video) extends across the other display field.
Press the ON/OFF softkey (under column 2) to turn automatic printing to ON. Now, whenever a
measurement is made, all of the measured results that are displayed on the NERO mAx’s screen will also
be sent to the printer.
If the printer is off line, out of paper or is otherwise non functional, the NERO mAx will display a printer
error message and printing will be disabled. If the printer becomes functional and returns to an on line
status with no errors, the NERO mAx will resume printing with the next exposure.
To turn automatic printing off, follow the procedure outlined above and toggle PRINT to OFF.
1.6 Scope Output
The NERO mAx scope output provides a real time output of the radiation waveform from the NERO mAx
detector. This output is from the less filtered detector; “channel A”. This signal can have a maximum
amplitude of approximately 5 volts. This signal is always available at the scope output BNC connector.
The NERO mAx does not need to be in a kVp measurement mode to provide a real time scope output,
but the NERO mAx detector must be in the beam.
To use the real time scope output, connect the NERO mAx’s scope output to an oscilloscope input using
a suitable BNC cable. Set the oscilloscope horizontal deflection controls to the desired sweep period and
adjust the scope to trigger on a positive slope. Some experimentation will be necessary to get the trigger
level and the vertical deflection adjusted properly. Generally, exposures made at the top of the selected
kV range will have signals above one volt and exposures made near the bottom of the selected kV range
will have signals in the tens of millivolts.
1-6
General Information
Procedures, Warnings, and Cautions
1
1.7 Procedures, Warnings, and Cautions
The equipment described in this manual is intended to be used for the detection and measurement of
ionizing radiation. It should be used only by persons who have been trained in the proper interpretation of
its readings and the appropriate safety procedures to be followed in the presence of radiation.
Although the equipment described in this manual is designed and manufactured in compliance with all
applicable safety standards, certain hazards are inherent in the use of electronic and radiometric
equipment.
WARNINGS and CAUTIONS are presented throughout this document to alert the user to potentially
hazardous situations. A WARNING is a precautionary message preceding an operation that has the
potential to cause personal injury or death. A CAUTION is a precautionary message preceding an
operation that has the potential to cause permanent damage to the equipment and/or loss of data.
Failure to comply with WARNINGS and CAUTIONS is at the user’s own risk and is sufficient cause to
terminate the warranty agreement between Fluke Biomedical and the customer.
Adequate warnings are included in this manual and on the product itself to cover hazards that may be
encountered in normal use and servicing of this equipment. No other procedures are warranted by Fluke
Biomedical. It shall be the owner’s or user’s responsibility to see to it that the procedures described here
are meticulously followed, and especially that WARNINGS and CAUTIONS are heeded. Failure on the
part of the owner or user in any way to follow the prescribed procedures shall absolve Fluke Biomedical
and its agents from any resulting liability.
Indicated battery and other operational tests must be performed prior to each use to assure that the
instrument is functioning properly. If applicable, failure to conduct periodic performance tests in
accordance with ANSI N323-1978 (R1983) Radiation Protection Instrumentation Test and Calibration, paragraphs 4.6 and 5.4, and to keep records thereof in accordance with paragraph 4.5 of the
same standard, could result in erroneous readings or potential danger. ANSI N323-1978 becomes, by
this reference, a part of this operating procedure.
Warning Summary
The following WARNINGS are provided for your reference and may appear throughout the NERO mAx
manual:
Extreme caution should be used when making
connections to the mAs terminals of the X-ray
generator or detector. Improper connections may
result in injury, damage to the NERO mAx, and/or
damage to the x-ray generator. Tube current (mA
and mAs) measurements should only be made by
persons familiar with the calibration and repair of xray machines.
An electric shock hazard exists between the ion
chamber bias connector and ground.
WARNING
WARNING
1-7
Victoreen 8000
Operators Manual
Caution Summary
The following CAUTIONS are provided for your reference and may appear throughout the NERO mAx
manual:
Use extreme caution when connecting to the mAs
terminal of the detector.
If line voltage surges beyond 15% of normal, a
power line conditioner must be used, otherwise
damage to the charging circuit will occur.
Note Summary
In the event of a transient induced lockup of the
Model 8000 NERO mAx, it is necessary to reset the
unit by cycling its power (turning it off then on).
After reset, the unit will power up in its normal
operating mode.
1.8 Receiving Inspection
CAUTION
CAUTION
NOTE
Upon receipt of the package:
1. Inspect the carton(s) and contents for damage. If damage is evident, file a claim with the carrier
and notify Fluke Biomedical at 440.248.9300.
2. Remove the contents from the packing material.
3. Verify that all items listed on the packing list have been received and are in good order.
1.9 Storage
If the unit is to be stored prior to use, pack it in the original container, if possible, and store in an
environment free of corrosive materials, fluctuations in temperature and humidity, and vibration and
shock.
Prior to use, check the condition and functionality of the device. Also check that the calibration is still
valid. Periodic recalibrations are usually required by individual radiation safety and/or quality assurance
programs. Please consult your local radiation safety or quality assurance office if you have any
questions.
1-8
Operation
Description
2
Section 2
Operation
2.1 Description
The NERO mAx consists of the NERO mAx control console, detector, detector cable, filter slides, AC
adapter, HVL plates, manual, Microsoft Excel Add-in and carrying case.
The NERO mAx control console is compact and easy to use. The sophisticated electronics necessary to
provide highly accurate, reproducible measurements while maintaining an intuitive, user-friendly operating
system are contained in the NERO mAx control console. The NERO mAx’s rechargeable battery is also
housed in the control console. The front panel of the control console contains a backlit 240 x 64 pixel, dot
matrix LCD display and eight push buttons. Connectors for power input, RS-232, printer, scope output
and the NERO mAx detector are located on the control console’s rear panel. The ON/OFF switch is
located on the right side of the control console.
The NERO mAx detector contains sensors for simultaneously measuring kV, exposure or rate and
invasive mA or mAs. Solid-state detectors are used to measure kV. An ion chamber, located in the top
of the detector, is used for exposure/rate measurements. In addition, connectors for an external ion
chamber are provided on the rear panel of the 8000 detector. The NERO mAx detector’s interface
connector is also located on the detector’s rear panel. The front panel has a keyed opening for the model
8000-filter slides and a connector for mAs leads.
The NERO mAx filter cards contain the various filters needed to accurately measure kilovoltage. Each
filter card is coded so that the NERO mAx “knows” which filter is in use and its position. The NERO mAx
also verifies that the filter card is valid for the selected measurement mode. In addition, the filter cards
are keyed so that they may only be inserted one way. The W/Al filter card is labeled with the kVp ranges
that it is calibrated for. The Mammo filter card is labeled for the x-ray tube targets that it is calibrated for.
The serial numbers of the NERO mAx control console unit, detector and filter slides must be matched in
order to obtain accurate results. The control console unit, detector and filter slides are calibrated
together, and
must be used together for accurate measurements.
2.2 General
Positioning the Control Console
Position the NERO mAx control console on a stable, flat surface within 25 feet of the detector. If a printer
is to be used with the NERO mAx it should also be placed on a stable, flat surface.
Positioning the Detector
Correct and reproducible positioning of the NERO mAx detector in the x-ray beam is very important in
obtaining accurate and reproducible results from the NERO mAx. Fluke Biomedical has printed several
alignment marks on top of the detector to assure correct, reproducible positioning of the detector.
The black circle is the minimum collimated beam size required for accurate exposure, rate and kVp
measurements. This circle defines the diameter of the NERO mAx’s internal ion chamber.
2-1
Victoreen 8000
Operators Manual
NOTE
The dashed gray rectangle is the minimum
collimated beam size required for accurate kVp
measurements only. This rectangle outlines the
NERO mAx’s kV detectors. Exposure and rate
results from an x-ray beam collimated to this area
will be incorrect because the beam is not
illuminating all of the internal ion chamber.
The black square and the dashed black crosshairs are alignment marks to aid in positioning the NERO
mAx detector in the x-ray beam.
The red arrows on the detector indicate the axis of the detector that should be aligned with the x-ray tube
axis for the most accurate measurements. This minimizes any heel effect.
Radiographic
Install the W/Al filter slide in the detector and set the filter slide to the desired kVp range. Position the
detector under the x-ray tube with the top of the detector facing up. Align the detector along the axis of
the x-ray tube to minimize heel effect. Set x-ray tube SDD (normally 26 inches) and collimate the beam
size to the round or square alignment marks on the top of the detector. Align the x-ray beam by making
the light field crosshairs coincident with the crosshairs on the top of the detector.
Dental
Set the detector on a flat, stable surface and position the x-ray tube so that the cone is just above the
detector's top surface.
2-2
Operation
General
NOTE
2
Do not rest the cone on the detector. This may
depress the top of the ion chamber and cause
incorrect exposure measurements.
Make sure that the detector is aligned along the axis of the x-ray tube and that the tube is perpendicular
to the detector's top surface and is centered over the detector crosshairs.
Mammography
Install the Mammo filter slide in the detector and set the filter slide for the x-ray machine’s target material.
Position the detector under the x-ray tube with the top of the detector facing up. Align the detector along
the axis of the x-ray tube to minimize heel effect. Set x-ray tube SSD (normally 26 inches) and collimate
the beam size to the square alignment marks on the top of the detector. Line up the front edge of the
square alignment mark on the top of the detector with the front edge (toward the chest wall) of the
collimated x-ray beam.
Fluoroscopy
Install the W/Al filter slide in the detector and set the filter slide to the desired kVp range. The kVp range
is the range closest to the detector front panel.
Position the detector
ray tube) and align the detector along the axis of the x-ray tube to minimize heel effect. Using the
centering marks provided on the table (or other centering methods), center the detector over the x-ray
beam. Don protective clothing and energize the fluoroscope to view the detector on the fluoro screen.
Move the fluoroscope so that the lead shield containing the NERO mAx’s detector diodes (opaque
rectangle) is centered on the screen.
upside down on the table (the top of the detector must be positioned toward the x-
Collimate to this area
X-Ray Tube Axis
Fluoro Detector Positioning
2-3
Victoreen 8000
Operators Manual
For automatic brightness control machines, place appropriate shielding over the detector to drive the
output to its maximum. A folded lead apron, gloves or a lead sheet may be used. For manual systems,
set the machine for its maximum output and use appropriate shielding to protect the image intensifier.
Other Applications
The basic principles are the same for other applications. In general, the detector should be located 18 to
40 inches from the x-ray source. The detector should be aligned with the x-ray tube axis with the top of
the detector facing the x-ray source. The detector should be in the center of the x-ray beam and the
beam should be collimated to the round or square alignment marks on the top of the detector.
For chest x-ray machines, the detector may be strapped or taped to the table in front of the film cassette.
For panoramic dental machines, the detector may be strapped or taped to the film cassette holder.
Inserting the Filter Slide
The NERO mAx filter slides are inserted into the front of the detector. The filter slides are keyed so that
they may only be inserted one way. To insert the filter slide, place the filter slide in the opening in the
front of the detector and gently push. The slide will click into place at each of its positions. Move the filter
slide until the desired kVp range (W/Al) or target material (Mammo) is closest to the detector front panel
(below “kVp RANGE”).
Connections to the NERO mAx
The NERO mAx detector and control console are connected to each other via a 25-foot cable. The
NERO mAx power should be turned
an external ion chamber to the detector or connecting the mAs leads.
off when connecting the detector to the control console, connecting
mAs Connections
Extreme caution should be used when making
connections to the mAs terminals of the X-ray
generator. Improper connections may result in
injury, damage to the NERO mAx, and/or damage
to the x-ray generator. Tube current (mA and
mAs) measurements should only be made by
persons familiar with the calibration and repair of xray machines.
Two mAs cables are provided with the NERO mAx; 12-ft. long mAs leads and 12-ft. long mAs extension
leads. The mAs leads are black, with a miniature phone plug at one end and alligator clips at the other
end, the mAs extension leads have clear insulation with two banana plugs at one end and alligator clips
at the other end. The miniature phone jack of the black mAs leads plugs into the mAs jack on the front
panel of the NERO mAx detector. The alligator clips may then be connected to the generator’s mAs
terminals. If the black mAs leads have insufficient length, the mAs extension leads may be used to
provide additional length. The mAs extension leads may be used in two ways; the extension lead banana
jacks may be plugged into the generator’s mAs terminals (if the generator has banana jacks) or the
extension leads may be reversed and the alligator clips can be clipped to the generator’s mAs terminals.
WARNING
The opposite end of the mAs extension leads are to be connected to the alligator clips at the end of the
black mAs leads. When using the mAs extension leads, care should be used to make sure that the leads
are not shorted together when they are connected.
2-4
Operation
General
The mAs leads should always be connected in the ground return of the high voltage transformer.
Damage to the generator or the NERO mAx or inaccurate measurements may result if the mAs leads are
connected to any point other then the ground return of the x-ray tube current. Tube current
measurements should only be made on generators providing open and short circuit protection of the
metering circuit and where the circuit operates near ground potential. The polarity of the mAs input signal
is not important because a full wave bridge is used in the NERO mAx mAs input circuit.
2
Printer Connections
When using the NERO mAx with a printer, both the printer and the NERO mAx should be turned off prior
to connection. The NERO mAx uses a standard IBM PC printer cable. For more information on printing
with the NERO mAx, see Section 1.5--Printing.
AC Adapter
The AC adapter may be plugged into the NERO mAx at any time. For more information on using the AC
adapter with the NERO mAx, see Section 1.4--Battery Operation and Charging.
Scope Output
The NERO mAx’s scope output may be connected to a suitable oscilloscope at any time. For more
information regarding the real time scope output, see Section 1.6--Scope Output.
2.3 Using The NERO mAx
The NERO mAx uses an innovative system of menus and softkeys to provide an intuitive, user-friendly
operating environment. All measurement modes and options are displayed on the NERO mAx’s LCD and
all functions are controlled by the 5 softkeys beneath the display and the 3 keys to the right of the display.
2-5
Victoreen 8000
Operators Manual
The NERO mAx display is divided into 5 columns, each column corresponds to the push button (softkey)
directly beneath it. The left most column is the “MODE” column that is used to select the NERO mAx’s
operating mode. The remaining four columns (1 through 4) display various options for each mode.
A highlight (reverse video block) denotes the selected menu element in each column. A blinking highlight
(reverse video block) locates the “active” menu column and highlights the selected menu element in that
column.
Legends, which describe the function of each softkey, appear along the bottom of the display. These
legends are separated from the rest of the display by a horizontal line. The time and date may be
displayed in the upper right corner of the display.
There are eight push buttons on the front panel of the NERO mAx. The five buttons directly beneath the
display are “softkeys”, their functions change according to the NERO mAx’s mode of operation. These
softkeys are used to select the NERO mAx’s mode of operation and various options within each mode.
These softkeys are also used to increment numerical values such as time or date. The “MODE” softkey
has two functions. First, it may be used to select the NERO mAx’s operating mode. Second, it may be
used to exit from any active measurement mode.
The “UP” and “DOWN” keys to the right of the display are also used to select the NERO mAx’s mode of
operation and various options within each mode. The “UP” and “DOWN” buttons move the highlight up
and down within each selected column to select various options. When the “UP” button is pressed, the
highlight moves up one menu element and wraps around when it reaches the top of the column and
returns to the bottom. When the “DOWN” button is pressed, the highlight moves down one menu element
and wraps around when it reaches the bottom. If only two options are available, the highlight toggles
between the two selections. In the calibration and setup modes, the “UP” and “DOWN” buttons are used
to toggle between options or to increment and decrement numeric values.
The “ENTER” button to the right of the display is used to initiate data acquisition and measurement with
options that have been selected. It also is used to accept numeric data in situations requiring data entry.
Each of the NERO mAx’s operating modes has a menu. All available menu options for each mode are
displayed in columns 1 through 4 above each softkey. All menus are left justified. This means that
changes made in columns on the left (such as the MODE column) may affect the columns of options to
the right. Softkey legends appear along the bottom of the display and as the menu options change, the
softkey legends change. The left most menu column is the ‘mode’ column, which displays the available
operating modes. The legend above the “MODE” softkey never changes, as this key always selects the
mode of operation of the NERO mAx.
Pressing the softkey under any column (MODE, 1 - 4) moves the blinking highlight into that menu column,
thereby selecting that column, and moves the highlight up through the available options. The highlight
wraps around when it reaches the top of the column and returns to the bottom. If only two options are
available, the highlight toggles between the two selections. In the calibration and setup modes, some of
the softkeys are used to toggle between options or to increment numeric values.
To make a measurement with the NERO mAx, simply highlight the desired measurement mode and
options and press the enter button. For more information on a specific measurement mode, see the
manual section that applies to that mode or x-ray generator type.
All of the NERO mAx’s operating modes are described in "Modes of Operation", following "Quick Start".
2.3.1 LCD Backlight Control
As a power saving feature, the NERO mAx LCD backlight turns off after one minute of no activity. The
backlight turns back on when any button is pressed. When the NERO mAx backlight turns off while
exposure results are being displayed, the backlight may be restored without affecting the displayed
exposure results by pressing any key except the MODE key. Pressing the mode key exits from the active
measurement mode and returns to the menu screen, erasing all exposure data.
2-6
Operation
Using the NERO mAx
2
2.3.2 Measured Quantities
kV
The NERO mAx calculates kVp from the ratio between two differentially filtered detector channels (A&B).
When an x-ray exposure is made, the NERO mAx samples the two detector channels simultaneously at a
rate of 100,000 samples per second. The detector waveforms are digitized by a pair of 100 kHz 16 bit
A/D converters and stored in memory.
The NERO mAx has sufficient memory to store up to 480 milliseconds of waveform data. For
radiographic exposures that exceed 480 milliseconds in length, the NERO mAx stores the first 320
milliseconds of the waveform and the last 160 milliseconds. This method allows storage of both the rising
and falling edges of the x-ray waveform. In the fluoro and AMSE modes, the NERO mAx stores a 480millisecond sample of the x-ray waveform after the SAVE key is pressed.
A delay of up to 999 milliseconds may be used to delay the start of kV data acquisition in the Radio and
Mammo modes. This delay may be used to skip events that occur at the beginning of an exposure, such
as an overshoot or undershoot. When a delay is used, only the kV data acquisition is effected.
Exposure, mAs and time measurements are not delayed and are measured over the entire exposure.
After the exposure is complete, the NERO mAx searches the stored channel A and B waveforms for ratio
peaks, when a peak is found it is stored. These peaks are then averaged and the average kVp is
calculated. While the NERO mAx searches for ratio peaks, it also looks for the highest peak, when the
highest peak is found it is stored and peak kV is calculated. Effective kV is calculated from the ratio of the
integrated A and B waveforms which is analogous to the density ratio in the kVp film cassette.
When calculating kVp average in the radiographic mode, the NERO mAx only includes peaks that are
above the selected %kV in its kVp average calculation. When ZERO or 1ØPULSE are selected, the
NERO mAx includes all detected peaks in the kVp average calculation.
The NERO mAx calculates kV as a function (F) of the ratio between the two detector channels, A and B
(r). The calculations used to calculate kV peak, kV effective and kVp average are summarized below:
kVp Average = F (r), where r is the average peak ratio B/A
kV Peak = F (r), where r is the peak ratio B/A
kV Effective = F (r), where r is the ratio
The calibration information that the NERO mAx uses to calculate kV is stored in nonvolatile memory in the
NERO mAx control console.
∑
B/A
∑
Time
The NERO mAx measures exposure time by determining the time between the first and last passage
through a preset percentage of kVp average. To accomplish this, the NERO mAx calculates the detector
ratio that corresponds to the preset percentage of the exposure's kVp, then measures the time between
those points on the rising and falling edges of the ratio waveform. In the Radiographic mode, the
percentage of kVp average over which the NERO mAx measures radiographic exposure time may be
selected using the %kV key. In the Mammographic mode, the percentage is fixed at 90% of kVp
average.
In the radiographic mode, when 75%, 80%, or 90% kV is selected, the NERO mAx measures exposure
time between the 75%, 80%, or 90% points on the kV waveform. When zero crossing (ZERO) is
selected, the NERO mAx measures radiographic exposure time from the moment x-rays are detected
until they are no longer detected. When 1ØPULSE is selected, the NERO mAx counts the number of xray pulses in a pulsed or single-phase radiographic exposure. This is primarily for use with single phase
full and half wave rectified generators.
In pulsed and single-phase applications when time is measured in pulses (1ØPULSE), use of a
measurement delay may introduce errors in the pulse count because the NERO mAx does not count
pulses during the delay time. In pulsed and single-phase applications when time is measured at a
2-7
Victoreen 8000
Operators Manual
percentage of kV, use of a measurement delay can also introduce errors. If the delay terminates between
pulses, when no x-rays are present, the NERO mAx waits until the beginning of the next pulse to begin
timing. If the delay terminates during a pulse, when x-rays are present, the NERO mAx begins timing
immediately.
When a delay is used in making measurements in the Radio and Mammo Modes, the NERO mAx
calculates exposure time differently. If a positive measurement delay is used in making a measurement,
the NERO mAx includes the delay time in its calculated exposure time. In addition, the exposure time
may be slightly longer than the exposure time between the selected %kV points on the kV waveform.
This is because the Model 8000 does not store the leading edge of the x-ray output waveform; it waits for
the specified delay time before acquiring kVp data. Because of this, the %kV point on the rising edge of
the kV waveform is not stored and the rise time of the waveform appears to be instantaneous. The
difference between the actual and measured exposure times is the rise time between zero and the
specified %kV. When a negative measurement delay is specified, the NERO mAx does not include the
delay time in the measured exposure time. The measured exposure time is from the beginning of data
acquisition or the %kV point on the first rising edge after the delay until passage through the %kV point on
the falling edge of the kV waveform.
More information on using the %kV function or using a measurement delay may be found at the end of
the Radio Mode section under "Using %kV and Exposure Time Measurements" or "Using a Measurement
Delay".
Exposure and Exposure Rate
The NERO mAx measures exposure by integrating the signal from an ion chamber (either internal or
external) and applying the proper conversion factor(s) to calculate exposure (R or Gy). Exposure rate is
calculated by sampling the integrated charge from the ion chamber at one-second intervals and dividing
the integrated charge by the sample period and applying the proper conversion factor(s). In the Pulsed
Fluoro and AMSE modes, the NERO mAx calculates exposure per pulse or frame by sampling the
integrated charge from the ion chamber at one second intervals and dividing by the number of pulses or
frames that it counts in a one second interval and applying the proper conversion factor(s). The NERO
mAx's ion chamber is internally vented and all exposure and rate measurements are corrected for air
density based upon user entered temperature and pressure.
The NERO mAx’s internal ion chamber is factory calibrated to provide accurate exposure and rate
measurements over the entire kVp range of the NERO mAx. These exposure and rate measurements
are adjusted by applying energy dependent correction factors that are determined by comparison to
applicable N.I.S.T. techniques.
In addition to the factory calibration, a user entered multiplier is available for exposure and rate
measurements made with the NERO mAx’s internal ion chamber.
When using external ion chambers, exposure and rate are calculated using the user entered calibration
factor (R/nC or Gy/nC) for the selected chamber. When making measurements in the CT Exposure mode
with a CT chamber, user entered beam width is also used in calculating exposure and rate.
mAs and mA
The NERO mAx measures mAs by integrating the signal from the mAs input during an x-ray exposure.
To calculate mA, the integrated signal from the mAs input is sampled at one-second intervals and divided
by the sample period. In the Pulsed Fluoro and AMSE modes, the NERO mAx calculates mAs per pulse
or frame by sampling the integrated signal from the mAs input at one second intervals and dividing by the
number of pulses or frames that it counts in a one second interval. The mA(s) circuitry is factory
calibrated using a calibrated current source and is not user adjustable.
HVL
In the HVL mode, the NERO mAx calculates half value layer based upon a series of exposure or rate
measurements made with varying thicknesses of aluminum absorbers placed in the x-ray beam. The
exposure or rate measurements may be made using the NERO mAx’s internal ion chamber or an external
2-8
Operation
Using the NERO mAx
ion chamber. HVL is calculated using a linear regression of the natural log of the normalized exposure
versus absorber thickness in millimeters of aluminum. Below is the formula used by the NERO mAx to
calculate half value layer.
()
BxAHVL+∗=ln
Where: x = Normalized dose at Half Value Layer (0.5)
A = Slope of ln(x) vs. mmAl
B = Intercept of ln(x) vs. mmAl (Usually very close to zero)
2.4 Quick Start
Locate a suitable radiographic x-ray generator. Set the generator for approximately 80 kV, 100 mA, .1
seconds or 10 mAs.
2
CAUTION
Make sure the NERO mAx is turned off.
Plug the detector cable into the back of the NERO mAx at the connector marked DETECTOR.
Plug the other end of the detector cable into the NERO mAx detector.
Insert the W/Al filter slide into the front filter slot of the NERO mAx detector at the 50 - 100 kVp position.
The filter slide will click into place.
Place the detector on the x-ray table at approximately 26” SDD. Make sure that the top of the detector is
facing the x-ray tube and the detector is aligned with the x-ray tube axis.
Plug the AC adapter into the NERO mAx and plug the adapter into a suitable AC power outlet.
Turn on the NERO mAx.
After the NERO mAx performs its power up diagnostics, the main menu screen is displayed.
a. Press the MODE button, the reverse video highlight in the mode column of the display (far left
column above the mode key) will start blinking.
b. Press the down arrow key until RADIO is highlighted.
c. Press the "1" key until 75% is highlighted.
d. Press the "2" key to toggle sensitivity between high and low, set the sensitivity to LOW.
e. Press the "3" key to change the measurement delay. Using the up and down arrow keys, set the
delay to +10 milliseconds.
f. Press the ENTER button.
Please wait a moment while the NERO mAx prepares to measure an exposure.
When the NERO mAx beeps and displays MAKE EXPOSURE, make an exposure.
2-9
Victoreen 8000
Operators Manual
Wait a moment while the 8000 analyzes the exposure then displays the measured kVp, exposure, time
and mAs (mAs will be zero because the mAs input was not used).
80.0
79.2
81.1
RADIO 75% LOW + 10 ms Please
MODE %kV SENS DELAY Wait…
The 8000 will immediately display Please Wait… in the lower right corner of its screen while it prepares to
take another exposure.
After the 8000 beeps and displays Make Exposure in the lower right corner of the display, another
exposure may be made.
To stop making exposures and return to the main menu screen, press the MODE button.
kVp Avg
kV Eff
kV Peak
100
392
0.0
msec
mR
mAs
2.5 Modes of Operation
The NERO mAx Nero has ten modes of operation as listed below (in selection order):
1. RADIO
2. MAMMO
3. FLUORO
4. AMSE
5. CT EXP
6. EXP
7. HVL
8. CAL
9. SETUP
10. UNIT ID
The NERO mAx's mode of operation is controlled from the main screen. When the menu cursor (blinking
reverse video) is moved to a mode selection, by pressing the mode key or using the up and down arrow
keys, all of the options available for that mode are displayed. Pressing a softkey (1 thru 4) moves the
menu cursor into the selected menu column and starts it blinking. Pressing the softkey again will move
the blinking menu cursor up through the column of available options. Pressing the up and down arrow
keys moves the blinking menu cursor up or down through the selected column of available options. After
options have been highlighted, the enter key is pressed to start the selected mode with the selected
options. Pressing the mode key at any time returns to the mode select screen.
UNIT
SETUP
CAL
HVL
EXP
CT EXP
AMSE ZERO
FLUORO 75%
MAMMO 80%
RADIO 90%
MODE %kV
1Ø PULSE
LOW
HIGH 10 mS
SENS DELAY
NERO mAx Operating Modes
Radio Mode
Radio mode is used to make measurements on tungsten target, aluminum filtered radiographic x-ray
generators. Radio mode simultaneously measures kVp, exposure, exposure time and mAs from a single
radiographic exposure. Radiographic mode may also be used to verify the NERO mAx’s kV calibration.
2-10
Operation
Modes of Operation
2
Mammo Mode
Mammo mode is used to make measurements on mammographic x-ray generators. Mammo mode
simultaneously measures kVp, exposure, exposure time and mAs from a single mammographic
exposure.
Fluoro Mode
Fluoro mode is used to make measurements on fluoroscopic x-ray generators. Fluoro mode supports
both continuous fluoro and pulsed fluoro measurements. In the continuous fluoro mode, the NERO mAx
measures kVp, exposure rate (R/min) and mA. In the pulsed fluoro mode, the NERO mAx measures
kVp, exposure rate (R/min and mR/pulse) and mAs/pulse.
AMSE Mode
AMSE mode is used for Automated Measurement of Sequential Exposures. This mode is used to
measure the output of CINE X-ray generators. In AMSE mode, the NERO mAx measures kVp, exposure
rate (mR/frame), mAs/frame and time/frame (mS/frame).
CT Exposure Mode
CT Exposure mode is used to make CT exposure measurements using the Victoreen 6000-100 or 6000200 CT ion chamber. A CT probe must be connected to the NERO mAx detector’s external ion chamber
input in this mode. The exposure is calculated using the user entered beam width (in mm) and the CT
probe’s calibration factor (Rcm/nC). This mode functions in the same manner as the Exposure Mode with
the addition of beam width entry.
Exposure Mode
Exposure mode is used to make exposure and rate measurements using the NERO mAx’s internal ion
chamber or an external ion chamber. The exposure is calculated using the selected ion chamber’s stored
calibration factor.
HVL Mode
In the HVL mode, the NERO mAx calculates half value layer based upon a series of exposure or rate
measurements made with varying thicknesses of aluminum absorbers placed in the x-ray beam. A
minimum of two exposures are required and up to ten exposures may be used. The exposure or rate
measurements may be made using the NERO mAx’s internal ion chamber or an external ion chamber.
Exposure and rate are calculated using the selected ion chamber’s calibration factor and when using a
CT chamber, beam width.
Calibrate Mode
The Calibrate mode is used to enter and store calibration factors for ion chambers used with the NERO
mAx. Calibration factors are available for the NERO mAx’s internal chamber and external chambers,
including the Victoreen CT chamber.
Setup Mode
The Setup Mode is used to setup various features of the NERO mAx. From the setup screen the user
can set the instruments internal real time clock, set the temperature and pressure used in air density
correction of exposure measurements, select exposure units of either Roentgens or Grays, turn automatic
printing on or off, turn the clock display on or off and select normal or reverse video on screen clock
display.
Unit ID
Displays the NERO mAx’s serial number, firmware part number and revision.
2-11
Victoreen 8000
Operators Manual
2.5.1 Radio Mode
Radio mode is primarily used to make measurements on tungsten target, aluminum filtered radiographic
x-ray machines. Radio mode simultaneously measures kVp, exposure, exposure time and mAs from a
single radiographic exposure. A measurement delay of up to 999 milliseconds may be used to delay the
start of kV data acquisition in order to skip over waveform anomalies at the beginning of an exposure.
See "Using a Measurement Delay" (at the end of this section) for more information on using a
measurement delay. The percentage of peak kV over which exposure time is measured may be selected
from 90%, 80%, or 75% of the kV peak. In addition, exposure time may be measured between zero
crossings or x-ray pulses may be counted (for single-phase generators). See "Using %kV and Exposure
Time Measurements" (at the end of this section) for more information about using %kV.
Generally, to measure radiographic exposures, choose a filter card (kV) setting such that the measured
(or expected) kV is in the upper end of the filter range. For instance, use the 50 - 100 kVp range instead
of the 80 - 160 kVp range to make measurements at 80 kVp. Staying in the upper end of the filter's kVp
range improves the signal to noise performance of the NERO mAx and allows the NERO mAx to "receive"
more of the radiation output that improves its measurements accuracy. Also, start out in low sensitivity, if
a channel A or B overrange occurs, switch to the next kVp filter range. If the NERO mAx does not
respond to an exposure and displays "MAKE EXPOSURE", switch to high sensitivity.
Radio mode may also be used to make measurements on dental x-ray generators. When using the
NERO mAx with dental x-ray machines, follow the instructions in section "Positioning the Detector" to
properly locate the detector. Generally, low sensitivity should be used, however if the NERO mAx does
not respond to an exposure, switch to high sensitivity. Select the proper %kV for the type of generator
being tested, generally ZERO or 1ØPULSE modes are needed for self-rectified generators, 75%, 80%, or
90% may be needed for three phase and medium to high frequency generators. If necessary, use a
measurement delay in order to disregard any filament preheat effects at the beginning of an exposure.
See "Using a Measurement Delay" and "Using %kV and Exposure Time Measurements" at the end of this
section for more information about using the %kV and measurement delay functions.
To make a radiographic measurement
Make sure that the Model 8000 is turned off. Plug one end of the Model 8000's detector cable into the
Model 8000 detector. Plug the other end of the detector cable into the Model 8000's detector connector.
Insert the W/Al filter card into the detector and place the filter card in the correct position for the kV range
to be measured. Place the Model 8000 detector on the x-ray table with the top of the detector facing the
x-ray tube. Position the detector so that the detector is centered in the beam and is aligned with the x-ray
tube axis. If mAs measurements are to be made, plug the mAs cable into the 8000 detector's mAs input
and connect the mAs leads to the generators mAs terminals. For more information on positioning the
Model 8000 detector, see section "Positioning the Detector".
WARNING
Extreme caution should be used when making
connections to the mAs terminals of the X-ray
generator or detector. Improper connections may
result in injury, damage to the NERO mAx, and/or
damage to the x-ray generator.
If results are to be printed, plug the printer cable into the printer port at the rear of the Model 8000 then
plug the other end of the printer cable into the printer and turn the printer on (see Section 1.5--Printing).
Plug the AC adapter into the 8000 if needed (see Section 1.4--Battery Operation and Charging) and turn
the instrument on.
Select the RADIO mode, %kV, high or low sensitivity and measurement delay, then press the ENTER
key. To use a measurement delay, press the DELAY key to select and increment the delay. The delay
may also be incremented or decremented by pressing the up or down keys when the delay is selected.
Select the interval, as a percentage of kV peak, over which exposure time is measured, using the %kV
key. ZERO selects measurement between the zero crossings on the rising and falling edges of the kV
waveform. 1ØPULSE selects the single-phase pulse counting mode for exposure time measurement.
The measurement delay and %kV settings are retained from one exposure to the next.
Please wait… FILTER = 50 - 100
RADIO 75% LOW 10 ms
MODE %kV SENS DELAY
2
Please wait while the Model 8000 prepares to take an exposure. Note that the selected filter kV range is
displayed. The kV filter range may be changed at any time without exiting from the measurement mode.
If the Model 8000 detects a fault condition such as an invalid filter card or position, an error message is
displayed and the user must correct the error to continue (see Section 4.3--Error Messages--for more
information). When the Model 8000 is ready for an exposure, it will beep and prompt for an exposure.
MAKE EXPOSURE FILTER = 50 - 100
RADIO 75% LOW 10 ms
MODE %kV SENS DELAY
Make an exposure. If nothing happens, there is insufficient x-ray intensity to make a measurement. To
remedy this situation, switch to high sensitivity, switch to the next lower kVp filter range (if possible),
increase mA or decrease the distance between the x-ray tube and Model 8000 detector.
ANALYZING DATA FILTER = 50 - 100
RADIO 75% LOW 10 ms
MODE %kV SENS DELAY
Please wait while the NERO mAx analyzes the exposure data.
80.0
kVp Avg
100
msec
2-13
Victoreen 8000
Operators Manual
79.2
81.1
RADIO 75% LOW 10 ms Please
MODE %kV SENS DELAY Wait…
After data analysis is complete, kV, exposure, time and mAs are displayed. If an overrange is detected
an error message is displayed. If the calculated kV is above or below the selected filter kV range, "High"
or "Low" is displayed instead of kV and measured time may be "----". If the NERO mAx cannot find the
selected %kV on the kV waveform, a %kV TOO LOW message is displayed (see Section 4.3--Error
Messages--for more information). Please wait while the Model 8000 prepares for the next exposure. If
the Model 8000 detects a fault condition such as an invalid filter card or position, an error message is
displayed and the user must correct the error to continue (see Section 4.3--Error Messages--for more
information).
80.0
79.2
81.1
RADIO 75% LOW 10 ms MAKE
MODE %kV SENS DELAY EXPOSURE
The Model 8000 is now ready to take another exposure. Pressing the mode key at any time exits from
this measurement mode and returns to the mode selection screen.
kV Eff
kV Peak
kVp Avg
kV Eff
kV Peak
392
0.0
100
392
0.0
mR
mAs
msec
mR
mAs
As a power saving feature, the Model 8000's display backlight turns off after one minute of no activity.
The backlight turns back on when any button is pressed or an exposure is made. In addition, after five
minutes of no activity, the 8000 exits from any measurement mode and returns to the mode selection
screen.
Using A Measurement Delay
A measurement delay may be used to postpone the start of data acquisition in order to skip over
waveform anomalies (such as overshoots or preheat effects) that may occur at the beginning of an
exposure.
When a delay is used, the NERO mAx waits for the specified delay time after its radiation threshold is
exceeded before starting data acquisition. Data acquisition starts immediately after the delay time has
elapsed if radiation is above the NERO mAx radiation threshold. If no radiation is detected after the delay
time has elapsed, data acquisition is delayed for up to one second after the delay time has elapsed. If no
radiation is detected for one second after the delay time has elapsed, the NERO mAx assumes that no
exposure has occurred and displays the “DELAY TOO LONG” message.
The delay range is from -999 to +999 milliseconds, and the polarity of the measurement delay only affects
how exposure time measurements are performed.
2-14
Operation
Modes of Operation
2
When a negative measurement delay is specified, the NERO mAx does not include the delay time in the
measured exposure time. A negative measurement delay should be used in cases when x-ray generator
filament preheat pulses or other waveform anomalies should be excluded from kV and exposure time
measurements. In the example shown below, a 10-pulse exposure is preceded by 14 filament preheat
pulses. Using a -230 millisecond delay to skip over the filament preheat pulses results in a measured
Negative Measurement Delay
Relative Radiation
120%
100%
80%
60%
40%
20%
0%
050100150200250300350400
A negative delay is used to skip over
filament preheat pulses
msec
Exposure time measured
with a negative delay
exposure time of 10 pulses.
An example showing this use of a measurement delay can also be found in troubleshooting Section 4.8-Waveforms - Dental With Filament Preheat.
When the measurement delay is positive, the NERO mAx includes the delay time in its measured
exposure time. A positive measurement delay should be used in cases when the kV waveform contains
leading edge overshoot or other waveform anomalies that should be excluded from kV measurements but
included in the exposure time measurement. In the example shown below, an overshoot occurs in the
first few milliseconds of a 50-millisecond exposure. Using a +10 millisecond delay to skip the overshoot
results in a measured kVp of 72 versus 82 and an exposure time of approximately 50 milliseconds.
2-15
Victoreen 8000
Operators Manual
Positive Measurement Delay
120
A positive delay
100
is used to skip
overshoots
80
60
kV
40
Exposure time measured with positive delay
20
10 mill isecond
delay
0
0 102030405060
msec
An example showing this use of a measurement delay can be found in troubleshooting Section 4.5-Waveforms - Overshoot.
When a positive measurement delay is used in making a measurement in the Radio or Mammo mode,
the measured exposure time may be slightly longer than the exposure time between the selected %kV
points on the kV waveform. This is because the Model 8000 does not store the leading edge of the x-ray
output waveform when a measurement delay is used; it waits for the specified delay time before acquiring
waveform data. As a result of this, the %kV point on the rising edge of the kV waveform is not stored and
the time from exposure start to this point cannot be subtracted from the total time.
When exposure time is measured in pulses (1ØPULSE), use of a measurement delay may cause an error
in the pulse count because no pulses are counted during the delay time. The polarity of the delay has no
effect on pulse counting; for example, the number of pulses counted with a -10 millisecond delay is the
same as the number counted with a +10 millisecond delay.
When using a measurement delay in pulsed and single-phase applications, care must be used in
determining the correct delay period. If x-rays are detected at the end of the delay period, such as during
a pulse, the NERO mAx begins timing immediately. If no x-rays are detected at the end of the delay
period, such as between pulses, the NERO mAx waits up to one second for the beginning of the next
pulse to begin timing. As a result of this, the delay period should terminate prior to the first x-ray pulse to
be included in data analysis.
The NERO mAx Excel Add-In should be used to view radiation and kV waveforms to determine the
optimum measurement delay to use. Information on using the NERO mAx Excel Add-In to view
waveforms may be found in the NERO mAx Toolkit for Excel Instruction Manual. In addition, a digital
storage oscilloscope may be connected to the scope output on the rear panel of the NERO mAx readout
to view radiation output waveforms. Information on using the NERO mAx scope output may be found in
Section 1.6 of the NERO mAx instruction manual.
2-16
Operation
Modes of Operation
2
Using %kV and Exposure Time Measurements
The %kV setting determines how the NERO mAx measures exposure time. When 75%, 80%, or 90% kV
is selected, the NERO mAx measures exposure time between the 75%, 80%, or 90% points on the kV
waveform. For best results when selecting 75%, 80%, or 90% kV, make sure that the percentage of the
kV waveforms peak kV is within the selected filter range. If the %kV or the measured kV is too low, a
"%kV TOO LOW" error message will be displayed.
When zero crossing (ZERO) is selected, the NERO mAx measures radiographic exposure time from the
moment x-rays are detected by the NERO mAx until they are no longer detected. This corresponds to the
time between first and last passage through the NERO mAx internal radiation detection threshold of the
channel A radiation signal. When using zero crossing, measured exposure times may be lengthened by
cable charging or by output filter capacitors used in some single phase generators to smooth the
generator’s output waveform.
When 1ØPULSE is selected, the NERO mAx counts the number of x-ray pulses in a pulsed or singlephase radiographic exposure. Pulses are detected by counting each passage through the NERO mAx
internal radiation detection threshold of the rising and falling edges of the pulses on channel A radiation
signal. This is primarily for use with single-phase full and half-wave rectified generators. Pulse counting
may not function properly on single-phase generators employing output filter capacitors to smooth the
generator’s output because the generator’s output may not drop to zero between pulses. Additionally,
use of a measurement delay may cause an error in the pulse count. For more information, see "Using a
Measurement Delay".
When calculating kVp average, the NERO mAx only includes peaks that are above the selected %kV in
the kVp average calculation. When ZERO or 1ØPULSE are selected, the NERO mAx includes all
detected peaks in the kVp average calculation.
2.5.2 Mammo Mode
Mammo mode is used to make measurements on mammographic generators. Mammo mode
simultaneously measures kVp, exposure time and mAs from a single mammographic exposure.
Exposure time in the Mammo mode is measured between the first and last passage through the 90%
points on the kV waveform. A measurement delay of up to 999 milliseconds may be used to delay the
start of kV data acquisition in order to skip over waveform anomalies at the beginning of an exposure.
See "Using a Measurement Delay" (at the end of this section) for more information on using a
measurement delay.
To make a mammographic measurement
Make sure that the NERO mAx is turned off. Plug one end of the NERO mAx’s detector cable into the
NERO mAx detector. Plug the other end of the detector cable into the NERO mAx’s detector connector.
Insert the MAMMO filter card into the detector and place the filter card in the correct position for the x-ray
tube target material. Place the NERO mAx detector on the x-ray table with the top of the detector facing
the x-ray tube. Position the detector so that the detector is centered in the beam and is aligned with the
x-ray tube axis. If mAs measurements are to be made, plug the mAs cable into the NERO mAx detector’s
mAs input and connect the mAs leads to the generators mAs terminals. For more information on
positioning the NERO mAx detector, see section "Positioning the Detector".
Extreme caution should be used when making
connections to the mAs terminals of the X-ray
generator or detector. Improper connections may
result in injury, damage to the NERO mAx, and/or
damage to the x-ray generator.
WARNING
2-17
Victoreen 8000
Operators Manual
If results are to be printed, plug the printer cable into the printer port at the rear of the NERO mAx then
plug the other end of the printer cable into the printer and turn the printer on (see Section 1.5--Printing).
Plug the AC adapter into the NERO mAx if needed (see Section 1.4--Battery Operation and Charging)
and turn the instrument on.
EXP
CT EXP
AMSE
FLUORO Mo 30μ
MAMMO LOW RHODIUM Rh 25μ
RADIO HIGH MOLY Al 1MM 0 ms
MODE SENS TARGET FILTER DELAY
Select the MAMMO mode, high or low sensitivity, target and filter and shot delay, then press the ENTER
key. To use a measurement delay, press the DELAY key to select and increment the shot delay. The
measurement delay may also be incremented or decremented by pressing the up or down keys when the
delay is selected. The measurement delay is retained from one exposure to the next and can only be
reset to zero by the user.
Please wait… FILTER = 22 - 35
MAMMO HIGH MOLY Mo 30μ 0 ms
MODE SENS TARGET FILTER DELAY
Please wait while the NERO mAx prepares to take an exposure. Note that the selected filter kV range is
displayed. If the NERO mAx detects a fault condition such as an invalid filter card or position, an error
message is displayed and the user must correct the error to continue (see Section 4.3--Error Messages-for more information). When the NERO mAx is ready to take an exposure it will beep and prompt for an
exposure.
MAKE EXPOSURE FILTER = 22 - 35
MAMMO HIGH MOLY Mo 30μ 0 ms
MODE SENS TARGET FILTER DELAY
Make an exposure. If nothing happens, there is insufficient x-ray intensity to make a measurement. To
remedy this situation, switch to high sensitivity, increase mA or decrease the distance between the x-ray
tube and NERO mAx detector.
ANALYZING DATA FILTER = 22 - 35
MAMMO HIGH MOLY Mo 30μ 0 ms
MODE SENS TARGET FILTER DELAY
Please wait while the NERO mAx analyzes the exposure data.
2-18
Loading...
+ 68 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.