Fluke 789 Operating Manual

Page 1
Manual Supplement
Manual Title: 789 Users Supplement Issue: 1 Print Date: August 2002 Issue Date: 8/04 Revision/Date: 1, 2/03 Page Count: 4
This supplement contains information necessary to ensure the accuracy of the above manual. This manual is distributed as an electronic manual on the following CD-ROM:
CD Title: 789 ProcessMeter CD Rev. & Date: 1, 2/2003 CD PN: 1636493
© 2004 Fluke Corporation. All rights reserved. Printed in the U.S.A. x
Page 2
789 Users Manual Supplement
Change #1
On page 2, under Safety Information, replace the first paragraph with the following:
The Meter complies with EN61010-1:2001, ANSI/ISA S82.01-2004, CAN/CSA C22.2 No. 1010-1:2004, UL61010-1, Measurement Category III, 1000V, Pollution Degree 2 and Measurement Category IV, 600V, Pollution Degree 2.
On page 4, Table 1, replace the CAT III and add the CAT IV:
CAT III IEC Overvoltage Category III
Category III equipment is designed to protect against transients in equipment in fixed equipment installations, such as distribution panels, feeders and short branch circuits, and lighting systems in large building.
8/04 1
CAT IV IEC Overvoltage Category IV
CAT IV equipment is designed to protect against transients from the primary supply level, such as an electricity meter or an overhead or underground utility service.
Page 3
Manual Supplement 789 Users
On page 5, replace Figure1 with the following:
'LVSOD\
352&(660(7(5

3XVKEXWWRQV
5$1*( +2/'
0,10$;
5RWDU\ 6ZLWFK

6SDQ&KHFN

9
2))
$
&2$56(
67(3
P9
9
P$
),1(
5(/
+]
P$
$
P$
P$
 +$57
P$
/22332:(5
9
,QSXW2XWSXW
&20
-DFNV
anw014f.eps
Figure1. Fluke 789 ProcessMeter
2 8/04
Page 4
789 Users Manual Supplement
On page 6, replace Figure 2, with the following:
1
A
2
Figure 2. Input/Output Jacks
mA
COM
V
4
3
anw001f.eps
On page 31, replace the Safety specifications with:
Safety:
1:2004, UL61010-1,Measurement Category III, 1000V, Pollution Degree 2 and Measurement Category IV, 600V, Pollution Degree 2.
8/04 3
The Meter complies with, EN61010-1:2001, ANSI/ISA S82.01-2004, CAN/CSA C22.2 No. 1010-
Page 5
Manual Supplement 789 Users
Change #2
On page 44, under Loop Power Supply,
Replace: Loop Power Supply: Minimum 24 V @ 24 mA in 1200 e load With: Loop Power Supply: Voltage: 24 V
Short Circuit protected
Change #3
On page 4 and 31, replace the UL, with the following:
! To 61010-1 2
4 8/04
nd
Edition
Loading...