Fluke 787 User Manual

April 1997 Rev. 4, 3/13
© 1997-2013 Fluke Corporation. All rights reserved. Specifications are subject to change without notice. All product names are trademarks of their respective companies.
787
ProcessMeter™

Users Manual

LIMITED WARRANTY AND LIMITATION OF LIABILITY
THIS WARRANTY IS YOUR ONLY REMEDY. NO OTHER WARRANTIES, SUCH AS FITNESS FOR A PARTICULAR PURPOSE, ARE EXPRESSED OR IMPLIED. FLUKE IS NOT LIABLE FOR ANY SPECIAL, INDIRECT, INCIDENTAL OR CONSEQUENTIAL DAMAGES OR LOSSES, ARISING FROM ANY CAUSE OR THEORY. Since some states or countries do not allow the exclusion or limitation of an implied warranty or of incidental or consequential damages, this limitation of liability may not apply to you.
Fluke Corporation P.O. Box 9090 Everett, WA 98206-9090 U.S.A.
11/99
Fluke Europe B.V. P.O. Box 1186 5602 BD Eindhoven The Netherlands

Table of Contents

Title Page
Introduction .................................................................................................................... 1
Contacting Fluke ............................................................................................................ 1
Safety Information .......................................................................................................... 2
How to Get Started ......................................................................................................... 6
Getting Acquainted with the Meter ................................................................................. 7
Measuring Electrical Parameters .................................................................................... 18
Input Impedance ........................................................................................................ 18
Ranges ...................................................................................................................... 18
Measuring a Composite Signal .................................................................................. 18
Testing Diodes ........................................................................................................... 19
Displaying Minimum, Maximum, and Average ........................................................... 19
Using TouchHold ....................................................................................................... 20
Compensating for Test Lead Resistance ................................................................... 20
Using the Current Output Functions ............................................................................... 21
Source Mode ............................................................................................................. 21
Simulate Mode ........................................................................................................... 23
Changing the Current Span ....................................................................................... 23
Producing a Steady mA Output ................................................................................. 25
i
787
Users Manual
Manually Stepping the mA Output ............................................................................ 26
Auto Ramping the mA Output ................................................................................... 27
Power-Up Options ......................................................................................................... 28
Battery Life ..................................................................................................................... 29
Using the Holster ........................................................................................................... 29
Maintenance .................................................................................................................. 30
General Maintenance ................................................................................................ 30
Calibration ................................................................................................................. 30
Replacing the Battery ................................................................................................ 30
Replacing a Fuse ...................................................................................................... 32
If the Meter does not Work ........................................................................................ 33
Replacement Parts and Accessories ............................................................................. 34
Specifications ................................................................................................................. 37
ii

List of Tables

Table Title Page
1. Symbols ................................................................................................................................. 5
2. Input/Output Jacks................................................................................................................. 8
3. Rotary Switch Positions for Measurements ........................................................................... 10
4. Rotary Knob Positions for mA Output .................................................................................... 12
5. Pushbuttons .......................................................................................................................... 13
6. Display ................................................................................................................................... 16
7. Range Requirements for Measuring a Composite Signal ...................................................... 18
8. Steady mA Output Adjustment .............................................................................................. 25
9. Steady mA Output Adjustment .............................................................................................. 26
10. mA Step Values ..................................................................................................................... 27
11. Power-Up Options ................................................................................................................. 28
12. Typical Alkaline Battery Life .................................................................................................. 29
13. Replacement Parts ................................................................................................................ 35
iii
787
Users Manual
iv

List of Figures

Figure Title Page
1. Fluke 787 ProcessMeter ........................................................................................................ 6
2. Input/Output Jacks................................................................................................................. 7
3. Rotary Switch Positions for Measurements ........................................................................... 9
4. Rotary Switch Positions for mA Output .................................................................................. 11
5. Pushbuttons .......................................................................................................................... 13
6. Elements of the Display ......................................................................................................... 15
7. Sourcing Current ................................................................................................................... 22
8. Simulating a Transmitter ........................................................................................................ 24
9. Using the Holster ................................................................................................................... 30
10. Replacing the Battery ............................................................................................................ 31
11. Replacing a Fuse .................................................................................................................. 33
12. Replacement Parts ................................................................................................................ 36
v
787
Users Manual
vi

Introduction

Warning
Read “Safety Information” before you use the Meter.
Your Fluke 787 ProcessMeter (referred to as “the Product or Meter”) is a handheld, battery-operated tool for measuring electrical parameters and supplying steady or ramping current to test process instruments. It has all the features of a digital multimeter, plus current output capability.
If the Meter is damaged or something is missing, contact the place of purchase immediately.
Contact your Fluke distributor for information about DMM accessories. To order replacement parts or spares, see Table 13 near the end of this manual.

Contacting Fluke

To contact Fluke, call one of the following telephone numbers:
Technical Support USA: 1-800-44-FLUKE
(1-800-443-5853)
Calibration/Repair USA: 1-888-99-FLUKE
(1-888-993-5853)
Canada: 1-800-36-FLUKE (1-800-363-5853)
Europe: +31 402-675-200
Japan: +81-3-6714-3114
1
787
Users Manual
Singapore: +65-6799-5566
Anywhere in the world: +1-425-446-5500
Or, visit Fluke's website at www.fluke.com.
To register your product, visit http://register.fluke.com.
To view, print, or download the latest manual supplement, visit http://us.fluke.com/usen/support/manuals.

Safety Information

The symbols used on the Product and in this manual are shown in Table 1.
A Warning identifies conditions and procedures that are dangerous to the user. A Caution identifies conditions and procedures that can cause damage to the Product or the equipment under test.
 Warning
To prevent possible electrical shock, fire, or personal injury:
Read “Safety Information” before using
the Product.
Do not use the Meter if it is damaged.
Before you use the Meter, inspect the case. Look for cracks or missing plastic. Pay particular attention to the insulation surrounding the connectors.
Make sure the battery door is closed and
latched before operating the Meter.
Remove test leads from the Meter before
opening the battery door.
Inspect the test leads for damaged
insulation or exposed metal. Check test lead continuity. Replace damaged test leads before you use the Meter.
Measure a known voltage first to make
sure that the Meter operates correctly. If you are unsure, have the Meter examined.
Do not operate the Meter around
explosive gas, vapor, or dust. Do not use in a damp or wet environment.
Use only a single 9 V battery, properly
installed in the meter case, to power the Meter.
When servicing the Meter, use only
specified replacement parts.
2
ProcessMeter™ Safety Information
Before a current measurement, do the
fuse test.
Use the correct terminals, function, and
range for measurements.
Do not work alone.
For current measurements, connect the
Meter to the circuit after you remove circuit power. Always put the Meter in series with the circuit.
Comply with local and national safety
requirements when in hazardous locations.
Only use test leads that have the same
voltage, category, and amperage ratings as the Meter and that are approved by a safety agency. When used with optional accessories, the lowest category rating applies.
Do not use TouchHold to determine if
dangerous voltage is present. TouchHold will not capture unstable or noisy readings.
To avoid false readings, which could lead
to possible electric shock or personal injury, replace the battery as soon as the battery indicator () appears.
Remove test leads from the Meter before
you open the battery door.
Close and latch the battery door before
using the Meter.
To avoid personal injury or damage to the
Meter, use only the specified replacement fuse, 440 mA 1000 V fast-blow, Fluke PN
943121.
Do not exceed the Measurement Category
(CAT) rating of the lowest rated individual component of a Product, probe, or accessory.
Do not use the TL175 or TP175 test
probes in CAT III or CAT IV environments without the probe tip fully extended and correct category rating visible in the window.
When the TL175 is used with instruments
or other accessories, the lowest category rating of the combination applies. One exception is when the probe is used with the AC172 or AC175.
3
787
Users Manual
Caution
To prevent possible damage to Product or to equipment under test:
Disconnect the power and discharge all
high-voltage capacitors before testing resistance or continuity.
Use the proper jacks, function, and range
for your measurement or sourcing application.
For best mechanical performance, the
Product must remain in the holster at all times.
To protect yourself, adhere to the following guidelines:
Use caution when working above 30V ac rms, 42V ac
pk, or 60V dc. Such voltages pose a shock hazard.
When using the probes, keep your fingers behind the
finger guards on the probes.
Connect the common test lead before you connect the
live test lead. When you disconnect test leads, disconnect the live test lead first.
4
ProcessMeter™
Safety Information
Table 1. Symbols
Symbol Meaning Symbol Meaning
AC (Alternating Current)
DC (Direct Current)
CAT II
CAT III
CAT IV
Risk of Danger. Important information. See Manual.
Alternating or direct current Battery Inspected and licensed by TÜV Product
Services Meets Underwriters’ Laboratories safety
requirements Fuse Double Insulation Conforms to CAN/CSA-C22.2 No 61010-1,
second edition, including Amendment 1. Measurement Category II is applicable to test and measuring circuits connected directly to utilization
points (socket outlets and similar points) of the low-voltage MAINS installation. Measurement Category III is applicable to test and measuring circuits connected to the distribution part of
the building’s low-voltage MAINS installation. Measurement Category IV is applicable to test and measuring circuits connected at the source of the
building’s low-voltage MAINS installation. This product complies with the WEEE Directive (2002/96/EC) marking requirements. The affixed label
indicates that you must not discard this electrical/electronic product in domestic household waste. Product Category: With reference to the equipment types in the WEEE Directive Annex I, this product is classed as category 9 "Monitoring and Control Instrumentation" product. Do not dispose of this product as unsorted municipal waste. Go to Fluke’s website for recycling information.
Earth ground
Conforms to European Union directives
Hazardous voltage.
Conforms to relevant North American Safety Standards.
Conforms to relevant Australian EMC requirements
Conforms to relevant South Korean EMC Standards.
5
787
Users Manual

How to Get Started

If you are familiar with the Fluke 80 Series DMM, read “Using the Current Output Functions,” review the tables and figures in “Getting Acquainted with the Meter,” and begin using your Meter.
If you are unfamiliar with Fluke 80 Series DMMs, or DMMs in general, read “Measuring Electrical Parameters” in addition to the sections referenced in the previous paragraph.
The sections following “Using the Current Output Functions” contain information about the power-up options, and battery and fuse replacement instructions.
Display
Rotary Switch
PROCESSMETER
787
Pushbuttons
REL Hz
mA
A
COM
CAT
1000V
OUTPUT
mA
mA
V
H
Input/Output Jacks
MIN MAX RANGE HOLD
% STEP COARSE FINE
mV
V
V
OFF
OUTPUT 0-24mA
SOURCE SIMULATE
+ +
A
mA
0.44A
(1A/ 30 sec)
FUSED
30mA
FUSED
Figure 1. Fluke 787 ProcessMeter
ee014f.eps
6
ProcessMeter™ Getting Acquainted with the Meter

Getting Acquainted with the Meter

To become familiar with the features and functions of the Meter, study the following figures and tables.
Figure and Table 2 describe the input/output jacks.
Figure and Table 3 describe the input functions you
get with the first five rotary switch positions.
OUTPUT 0-24mA
1
2
SOURCE SIMULATE
+ +
A
0.44A
(1A /30 sec)
FUSED
Figure 2. Input/Output Jacks
mA
Figure and Table 4 describe the output functions you
get with the last two rotary switch positions.
Figure and Table 5 describe the functions of the
pushbuttons.
Figure and Table 6 explain what all the elements of
the display indicate.
3
30mA
FUSED
COM
CAT
1000V
V
4
ee001f.eps
7
787
Users Manual
Table 2. Input/Output Jacks
Item Jack Measurement Functions Source Current Function
c A
d mA
V
COM
Input for current to 440 mA continuous. (1A for up to 30 seconds.) Fused with a 440 mA fuse.
Input for current to 30 mA. Fused with a 440 mA fuse.
Input for voltage to 1000V, Ω, continuity, and diode test.
Common for all measurements.
Output for dc current to 24 mA.
Common for dc current output to 24 mA.
Common for transmitter
Simulate Transmitter
Function
Output for transmitter simulation to 24 mA. (Use in series with an external loop supply.)
simulation to 24 mA. (Use in series with an external loop supply.)
8
ProcessMeter™ Getting Acquainted with the Meter
4
6
3
5
mV
V
V
mA
A
OUTPUT
mA
mA
2
OFF
1
ee002f.eps
Figure 3. Rotary Switch Positions for Measurements
9
787
Users Manual
Table 3. Rotary Switch Positions for Measurements
No. Position Function(s) Pushbutton Actions
OFF
V A
d V
d mV
O
mA A L
10
Meter off
Default: measure ac V
F Frequency counter
Measure dc V Same as above
Measure dc mV Same as above
Default: measure Ω
T for continuity
BLUE D test
High test lead in cA: measure A dc
BLUE selects ac
High test lead in
dmA: measure mA dc
M Selects a MIN, MAX, or AVG action (see pg. 18) KSelects a fixed range (hold 1 second for auto range) I Toggles TouchHold C Toggles relative reading (sets a relative zero point)
Same as above, except diode test has only one range
Same as above, except there is only one range for each input jack position, 30 mA or 1A
ProcessMeter™ Getting Acquainted with the Meter
mV
V
V
OFF
mA
A
OUTPUT
mA
mA
1
2
ee008.eps
Figure 4. Rotary Switch Positions for mA Output
11
787
Users Manual
Table 4. Rotary Switch Positions for mA Output
No. Position Default Function Pushbutton Actions
OUTPUT
[ mA
OUTPUT
mA J
12
Test leads in SOURCE:
Source 0 % mA
Test leads in SIMULATE:
Sink 0 % mA
Test leads in SOURCE:
Source repeating 0 % -100 %-0 % slow ramp (E)
Test leads in SIMULATE:
Sink repeating 0 % -100 %-0 % slow ramp (E)
% STEP X or W: Adjusts output up or down to the next 25 % step
COARSE Z or Y: Adjusts output up or down 0.1 mA
FINE X or W: Adjusts output up or down 0.001 mA
BLUE cycles through:
Fast repeating 0 % -100 % - 0 % ramp (P on display)
Repeating 0 % -100 % - 0 % ramp in 25 % steps (N on display)
Slow repeating 0 % -100 % - 0 % ramp (E on display)
ProcessMeter™ Getting Acquainted with the Meter
2
3
MIN MAX RANGE HOLD
% STEP COARSE FINE
REL Hz
4
H
1
78
Figure 5. Pushbuttons
Table 5. Pushbuttons
No. Pushbutton Function(s)
b Toggles the backlight
U
(BLUE)
Rotary switch in mA A Lposition and test lead plugged into c A jack: Toggles between ac and dc ampere measure
Rotary switch in O position: Selects diode test function (D) Rotary switch in OUTPUT mA Jposition: Cycles through
Slow repeating 0 % -100 % - 0 % ramp (Eon display)
Fast repeating 0 % -100 % - 0 % ramp (P on display)
Repeating 0 % -100 % - 0 % ramp in 25 % steps (N on display)
5
6
ee003f.eps
13
787
Users Manual
Table 5. Pushbuttons (cont.)
No. Pushbutton Function(s)
X
M
% STEP
Z
K
COARSE
X
I
FINE FINE
F
W
COARSE
C
Y
% STEP
T
W
Measuring: Selects a MIN, MAX, or AVG action (see pg. 18) mA Output: Adjusts mA output up to the next higher 25% step
Measuring: Selects a fixed range (hold for 1 second for auto range) mA Output: Adjusts output up 0.1 mA
Measuring: Toggles TouchHold, or in MIN MAX recording, suspends recording mA Output: Adjusts output up 0.001 mA
Measuring: Toggles between frequency counter and ac voltage measurement functions mA Output: Adjusts output down 0.001 mA
Measuring: Toggles relative reading (sets a relative zero point) mA Output: Adjusts output down 0.1 mA
Measuring: Toggles between Ω measure and continuity functions mA Output: Adjusts mA output down to the next lower 25 % step
14
ProcessMeter™ Getting Acquainted with the Meter
6 7
8
9
5
4
10
3
2
1
11
12
ee004f.eps
Figure 6. Elements of the Display
15
787
Users Manual
Table 6. Display
No. Element Meaning
Percentage display
OUTPUT
D
S
Q
B
Numerals
H
MINMAXAVGR
16
Shows the mA measured value or output level in %, in a 0-20 mA or 4-20 mA scale
(change scales with power-up option)
Lights when mA output (source or simulate) is active
Lights in diode test function
Lights in continuity function
Lights when relative reading is on
Lights when the battery is low
Show the input or output value
Lights when TouchHold is on
MIN MAX recording status indicators:
MIN means the display is showing the minimum recorded value. MAX means the display is showing the maximum recorded value. AVG means the display is showing the average value since starting recording (up to
about 35 hours continuous recording time). Rmeans MIN MAX recording is on.
ProcessMeter™ Getting Acquainted with the Meter
Table 6. Display (cont.)
No. Element
mA, DC, mV, AC, M
or kΩ, kHz
Auto
400100030
EP N
Meaning
Show the input or output units and multipliers associated with the numerals
Range status indicators:
Auto means autoranging is on. The number plus the unit and multiplier indicate the active range.
One of these lights in mA ramping or step output (rotary switch position mA J):
E means slow continuous 0% - 100% - 0% ramping.
P means fast continuous 0% - 100% - 0% ramping.
N means ramping in 25% steps.
17
787
Users Manual

Measuring Electrical Parameters

The proper sequence for taking measurements is as follows:
1. Plug the test leads into the appropriate jacks.
2. Set the rotary knob.
3. Touch the probes to the test points.

Input Impedance

For the voltage measurement functions, input impedance is 10 MΩ. See the specifications for more information.

Ranges

A measurement range determines the highest value the Meter can measure. Most meter measurement functions have more than one range (see the Specifications).
Being in the right range is important:
If the range is too low, the display shows OL
(overload).
If the range is too high, the Meter will not be
displaying its most accurate measurement.
The Meter normally automatically selects the lowest range that will measure the applied input signal (Auto showing on the display). Press Kif you want to lock the
range. Each time you press K, the Meter selects the next higher range.
If you have locked the range, the Meter resumes auto ranging when you change to another measurement function or you press K and hold it for 1 second.

Measuring a Composite Signal

Because the input is dc-coupled, to measure an ac voltage or frequency with a dc bias, you must manually select the range specified in Table 7. For example, to measure 100 mV ac with 20 V dc superimposed, select the 4 V range.
Table 7. Range Requirements for Measuring a
Composite Signal
Range (ac) Max. Allowable AC + DC
400.0 mV 3 V
4.000 V 30 V
40.00 V 300 V
400.0 V 400 V
1000 V 1000 V
18
ProcessMeter™ Measuring Electrical Parameters

Testing Diodes

To test a single diode:
1. Insert the red test lead into the Vjack and black test lead into the COM jack.
2. Set the rotary switch to O.
3. Press the BLUE pushbutton so that the D symbol is on the display.
4. Touch the red probe to the anode and the black probe to the cathode (side with band or bands). The Meter should indicate the appropriate diode voltage drop.
5. Reverse the probes. The Meter should display OL, indicating high impedance.
6. The diode is good if it passes the tests in steps 4 and 5.

Displaying Minimum, Maximum, and Average

MIN MAX recording stores the lowest and highest measurements, and maintains the average of all measurements.
Press M to turn on MIN MAX recording. Readings are stored until you turn the Meter off, switch to another measurement or source function, or turn MIN MAX off. The beeper sounds when a new maximum or minimum is recorded. Auto power-off is disabled and auto ranging is turned off during MIN MAX recording.
Press M again to cycle through the MAX, MIN, and AVG displays. Press and hold M for 1 second to erase stored measurements and exit.
If MIN MAX recording is on continuously for over 40 hours, minimum and maximum readings are still recorded, but the displayed average no longer changes.
In MIN MAX recording, press I to suspend recording; press I again to resume recording.
19
787
Users Manual

Using TouchHold

Note
You must have MIN MAX recording off to use TouchHold.
! Warning
To avoid possible electric shock, do not use TouchHold to determine if dangerous voltage is present. TouchHold will not capture unstable or noisy readings.
Activate TouchHold display on each new stable reading (except in the frequency counter function). Press I to activate TouchHold. This feature allows you to take measurements in situations in which it is difficult to look at the display. The Meter beeps and updates the display with each new stable reading.
®
if you want the Meter to freeze the

Compensating for Test Lead Resistance

Use the relative reading feature (Q on the display) to set the present measurement as a relative zero. A common use for this is to compensate for test lead resistance when measuring Ω.
Select the Ω measure function, touch the test leads together, then press C. Until you press C again, or switch to another measurement or source function, the readings on the display will subtract the lead resistance.
20
ProcessMeter™ Using the Current Output Functions

Using the Current Output Functions

The Meter provides steady, stepped, and ramped current output for testing 0-20 mA and 4-20 mA current loops. You can choose source mode, in which the Meter supplies the current, or simulate mode, in which the Meter regulates current in an externally-powered current loop.

Source Mode

Source mode is selected automatically by inserting the test leads into the SOURCE + and jacks as shown in Figure 7. Use source mode whenever you need to supply
current into a passive circuit such as a current loop with no loop supply. Source mode depletes the battery faster than simulate mode, so use simulate mode whenever possible.
The display looks the same in source and simulate modes. The way to tell which mode is in use is to see which pair of output jacks is in use.
21
787
Users Manual
787
PROCESSMETER
22
MIN MAX RANGE HOLD
% STEP COARSE FINE
mV
V
V
OFF
OUTPUT 0-24mA
SOURCE SIMULATE
+ +
A
mA
0.44A
(1A/30 sec)
FUSED
30mA
FUSED
REL Hz
mA
A
COM
CAT
1000V
OUTPUT
mA
mA
V
H
ee010f.eps
Figure 7. Sourcing Current
ProcessMeter™ Using the Current Output Functions

Simulate Mode

Simulate mode is so named because the Meter simulates a current loop transmitter. Use simulate mode when an external dc voltage of 24 to 30V is in series with the current loop under test.
Caution
Set the rotary switch to one of the mA output settings BEFORE you connect the test leads to a current loop. Otherwise, a low impedance from the other rotary switch positions could be presented to the loop, causing up to 50 mA to flow in the loop.
Simulate mode is selected automatically by inserting the test leads into the SIMULATE + and jacks as shown in Figure 8. Simulate mode conserves battery life, so use it instead of source mode whenever possible.
The display looks the same in source and simulate modes. The way to tell which mode is in use is to see which pair of output jacks is in use.

Changing the Current Span

The Meter’s current output span has two settings (with overrange to 24 mA):
4 mA = 0%, 20 mA = 100% (factory default)
0 mA = 0%, 20 mA = 100%
To find out which span is selected, short the OUTPUT SOURCE + and jacks, turn the rotary switch to OUTPUT [ mA, and observe the 0% output level.
To toggle and save the current output span in nonvolatile memory (retained when the power is turned off):
1. Turn off the Meter.
2. Hold down the Kpushbutton while you turn the rotary switch to OUTPUT [ mA.
3. Wait at least 2 seconds, then release K.
23
787
Users Manual
dc V
Power Supply
COM +24V
787
PROCESSMETER
24
REL Hz
mA
A
COM
CAT
1000V
OUTPUT
mA
V
H
mA
MIN MAX RANGE HOLD
% STEP COARSE FINE
mV
V
V
OFF
OUTPUT 0-24mA
SOURCE SIMULATE
+ +
A
mA
0.44A
(1A/30 sec)
FUSED
30mA
FUSED
Figure 8. Simulating a Transmitter
ee011f.eps
ProcessMeter™ Using the Current Output Functions

Producing a Steady mA Output

When the rotary switch is in the OUTPUT [ mA position, and the OUTPUT jacks are connected to an appropriate load, the Meter produces a steady mA dc output. The Meter begins sourcing or simulating 0%. Use the pushbuttons to adjust the current as shown in Table 8.
Select either sourcing or simulating by choosing the SOURCE or SIMULATE output jacks.
If the Meter cannot deliver the programmed current because the load resistance is too high or the loop supply voltage is too low, dashes (-----) appear on the numeric display. When the impedance between the SOURCE jacks is low enough, the Meter will resume sourcing.
Note
The STEP pushbuttons described on the next page are available when the Meter is producing a steady mA output. The STEP pushbuttons go to the next multiple of 25 %.
Pushbutton Adjustment
Table 8. mA Output Adjust Pushbuttons
Z
Adjusts up 0.1 mA
K
COARSE
X
Adjusts up 0.001 mA
I
FINE FINE
Adjusts down 0.001 mA
F
W
COARSE
Adjusts down 0.1 mA
C
Y
25
787
Users Manual

Manually Stepping the mA Output

When the rotary switch is in the OUTPUT [ mA position, and the OUTPUT jacks are connected to an appropriate load, the Meter produces a steady mA dc output. The Meter begins sourcing or simulating 0 %. Use the pushbuttons to step the current up and down in 25 % steps as shown in Table 9. See Table 10 for mA values at each 25 % step.
Select either sourcing or simulating by choosing the SOURCE or SIMULATE output jacks.
If the Meter cannot deliver the programmed current because the load resistance is too high or the loop supply voltage is too low, dashes (-----) appear on the numeric display. When the impedance between the SOURCE jacks is low enough, the Meter will resume sourcing.
Note
The COARSE and FINE adjustment pushbuttons described on the previous page are available when you are manually stepping the mA output.
Pushbutton Adjustment
Table 9. mA Stepping Pushbuttons
X
M
Adjusts up to the next higher
25 % step % STEP % STEP
T
Adjusts down to the next lower
25 % step
W
26
ProcessMeter™ Using the Current Output Functions
Table 10. mA Step Values
Step
0 % 4.000 mA 0.000 mA
25 % 8.000 mA 5.000 mA
50 % 12.000 mA 10.000 mA
75 % 16.000 mA 15.000 mA
100 % 20.000 mA 20.000 mA
125 % 24.000 mA
120 % 24.000 mA
Value (for each span setting)
4 to 20 mA 0 to 20 mA
When the rotary switch is in the OUTPUT mA J position, the Meter produces a continuously repeating 0 %
- 100 % - 0 % ramp in your choice of three ramp waveforms:
E 0 % - 100 % - 0 % 40-second smooth ramp,
P 0 % - 100 % - 0 % 15-second smooth ramp N 0 % - 100 % - 0 % Stair-step ramp in 25 % steps,
The ramp times are not adjustable. Press the BLUE pushbutton to cycle through the three waveforms.

Auto Ramping the mA Output

Auto ramping gives you the ability to continuously apply a varying current stimulus from the Meter to a transmitter, while your hands remain free to test the response of the transmitter. Select either sourcing or simulating by choosing the SOURCE or SIMULATE jacks.
(default)
pausing 5 seconds at each step. Steps are listed in Table 10.
Note
At any time during auto ramping, you can freeze the ramp simply by moving the rotary switch to the [ mA position. Then you can use the COARSE, FINE, and % STEP adjust pushbuttons to make adjustments.
27
787
Users Manual

Power-Up Options

To select a power-up option, hold down the pushbutton shown in Table 11 while turning the rotary switch from OFF to any on position. Wait 2 seconds before you release the pushbutton after powering up the Meter. The Meter beeps to acknowledge the power-up option.
Table 11. Power-Up Options
Option Pushbutton Default Action Taken
Change current span 0 % setting K Remembers
Disable beeper T Enabled Disables beeper
Disable auto power-off BLUE Enabled Disables the feature that turns off the Meter
Only the setting for current span is retained when the power is turned off. The others have to be repeated for each operating session.
You may activate more than one power-up option by holding down more than one pushbutton.
Toggles between 0 and 4 mA
last setting
power after 30 minutes of inactivity. Auto power off is disabled regardless of this option if MIN MAX recording is on.
28
ProcessMeter™ Battery Life

Battery Life

Warning
To avoid false readings, which could lead to possible electric shock or personal injury, replace the battery as soon as the battery indicator (B) appears.
Table 12 shows typical alkaline battery life. To preserve battery life:
Use current simulation instead of sourcing when
possible.
Avoid using the backlight.
Do not disable the automatic power-off feature.
Turn the Meter off when you are not using it.
Table 12. Typical Alkaline Battery Life
Meter Operation Hours
Measuring any parameter or simulating current
Sourcing 12 mA into 500Ω 12
80

Using the Holster

The Meter is supplied with a snap-on holster that absorbs shocks and protects the Meter from rough handling. You can turn the Meter over in the holster to protect the face of the meter from scratches when carrying the Meter.
Caution
For best mechanical performance, the Product must remain in the holster at all times.
29
787
Users Manual
Meter in holster face down for protection
Figure 9. Using the Holster
Holster with probe clip
ee009f.eps

Maintenance

This section provides some basic maintenance procedures. Repair, calibration, servicing not covered in this manual must be performed by qualified personnel. For maintenance procedures not described in this manual, contact a Fluke Service Center.

General Maintenance

Periodically wipe the case with a damp cloth and detergent; do not use abrasives or solvents.

Calibration

Calibrate your Meter once a year to ensure that it performs according to its specifications. Contact a Fluke Service Center for instructions.

Replacing the Battery

! Warning
To prevent possible electrical shock, fire, or personal injury:
Remove test leads from the Meter before
opening the battery door.
Make sure the battery door is closed and
latched before operating the Meter.
30
ProcessMeter™ Maintenance
Remove test leads from the Meter before you open the battery door.
Replace the battery as follows. Refer to Figure 10. Use an alkaline 9V battery, type ANSI/NEDA 1604A or IEC 6LR61.
1. Remove the test leads and set the rotary switch to OFF.
2. With a standard blade hand screwdriver, turn each battery door screw counterclockwise so that the slot is parallel with the screw picture molded into the case.
3. Lift off the battery door.
ee007f.eps
Figure 10. Replacing the Battery
31
787
Users Manual

Replacing a Fuse

! Warning
To avoid personal injury or damage to the Meter, use only the specified replacement fuse, 440 mA 1000V fast-blow, Fluke PN 943121.
Both current input jacks are fused with a separate 440 mA fuse. To determine if a fuse is blown:
1. Turn the rotary switch to mA A L.
2. Plug the black test lead into COM, and the red test lead into cA.
3. Using an ohmmeter, check the resistance between the meter test leads. If the resistance is about 1Ω, the fuse is good. An open means the fuse is blown.
4. Move red test lead to dmA.
5. Using an ohmmeter, check the resistance between the meter test leads. If the resistance is about 14Ω, the fuse is good. An open means the fuse is blown.
If a fuse is blown, replace it as follows. Refer to Figure 11 as necessary:
1. Remove the test leads from the Meter and turn the rotary switch to OFF.
2. Remove the battery door.
3. Remove the three Hexalobular (star) screws from the case bottom and turn the case over.
4. Gently lift the bottom of the front of the case (nearest the input/output jacks) until the top unsnaps from the rear half of the case.
5. Replace the blown fuse with the exact type specified: 440 mA 1000V fast-blow fuse, Fluke PN 943121. Both fuses are the same type.
6. Make sure the rotary switch is in the OFF position.
7. Fit the top of case together, engaging the two snaps (item ). Make sure that the gasket is properly seated.
8. Close the case and reinstall the three screws.
9. Replace the battery door.
32
ProcessMeter™ Maintenance

If the Meter does not Work

Examine the case for physical damage. If there is
damage, make no further attempt to use the Meter, and contact a Fluke Service Center.
Check the battery, fuses, and test leads.
Review this manual to make sure you are using the
correct jacks and rotary switch position.
If the Meter still does not work, contact a Fluke Service Center. If the Meter is under warranty, it will be repaired or replaced (at Fluke’s option) and returned at no charge. See the Warranty on the back of the title page for terms. If the warranty has lapsed, the Meter will be repaired and returned for a fixed fee. Contact a Fluke Service Center for information and price.
F1
F2
1
ee012f.eps
Figure 11. Replacing a Fuse
33
787
Users Manual

Replacement Parts and Accessories

! Warning
To avoid personal injury or damage to the Meter, use only the specified replacement fuse, 440 mA 1000V fast-blow, Fluke PN 943121.
Note
When servicing the Meter, use only the replacement parts specified here.
Replacement parts and some accessories are shown in Figure 12 and listed in Table 13. Many more DMM accessories are available from Fluke. For a catalog, contact your nearest Fluke distributor.
To find out how to order parts or accessories use the telephone numbers or addresses shown on page 1 of this manual.
34
ProcessMeter™ Replacement Parts and Accessories
Table 13. Replacement Parts
Item Description Fluke PN or Model no. Quantity
BT1 9V battery, ANSI/NEDA 1604A or IEC 6LR61 614487 1 MP103 Holster, Yellow 2074033 1 ! F1, 2 Fuse, 440 mA, 1000V fast-blow 943121 2 MP85 Case top 619962 1 MP86 Case bottom 619939 1 H2, 3, 4 Case screw 832246 3 MP89, 90 Non-skid foot 824466 2 MP8 O-ring for input/output receptacle 831933 1 MP92 Battery door 619947 1 H5, 6 Battery door fasteners 948609 2 S1 Keypad 646932 1 Not shown Test Leads variable Not shown Alligator Clips variable Not shown Industrial test lead set variable
[1]
1 (set of 2)
[1]
1 (set of 2)
[1]
Option Not shown Quick Reference Guide 4276679 1 Not shown Users Manual (CD-ROM) 1586721 1 Not shown Calibration Manual 641891 Option [1] See www.fluke.com for more information about the test leads and alligator clips available for your region.
35
787
Users Manual
MP85
Holster MP103
S1
36
MP8
MP86
H2, 3, 4
F2
F1
BT1
H5, 6
MP89, 90
MP92
Figure 12. Replacement Parts
Tilt Stand
CD-ROM (Users Manual)
ee015c.eps
ProcessMeter™ Specifications

Specifications

All specifications apply from +18°C to +28°C unless stated otherwise.
All specifications assume a 5 minute warmup period.
The standard specification interval is 1 year.
Note
“Counts” means number of increments or decrements of the least significant digit.
DC Volts Measurement
Range (V dc) Resolution Accuracy, ±(% of Reading + Counts)
4.000 0.001 V 0.1 % + 1
40.00 0.01 V 0.1 % + 1
400.0 0.1 V 0.1 % + 1
1000 1 V 0.1 % + 1
Input impedance: 10 MΩ (nominal), <100 pF Normal mode rejection ratio: >60 dB at 50 Hz or 60 Hz Common mode rejection ratio: >120 dB at dc, 50 Hz, or 60 Hz Overvoltage protection: 1000V
37
787
Users Manual
DC Millivolts Measurement
Range (mV dc) Resolution Accuracy (% of Reading + Counts)
400.0 0.1 mV 0.1 % + 1
AC Volts Measurement
Range (ac) Resolution
400.0 mV 0.1 mV 0.7 % + 4 1.2 % + 4 7.0 % + 4
4.000 V 0.001 V 0.7 % + 2 1.2 % + 4 7.0 % + 4
40.00 V 0.01 V 0.7 % + 2 1.2 % + 4 7.0 % + 4
400.0 V 0.1 V 0.7 % + 2 1.2 % + 4 7.0 % + 4
1000 V 1 V 0.7 % + 2 1.2 % + 4 7.0 % + 4
Specifications are valid from 5 % to 100 % of amplitude range. AC conversion: true rms Maximum crest factor: 3
±
For non-sinusoidal waveforms, add
Ω
Input impedance: 10 M Common mode rejection ratio: >60 dB at dc, 50 Hz, or 60 Hz
(nominal), < 100 pF, ac-coupled
(2 % reading + 2 % f.s.) typical
50 Hz to 60 Hz 45 Hz to 200 Hz 200 Hz to 500 Hz
Accuracy, ±(% of Reading + Counts)
38
ProcessMeter™ Specifications
AC Current Measurement
Range
45 Hz to 2 kHz
1.000 A (Note) 0.001 A 1 % + 2 1.5 V/A
Note: 440 mA continuous, 1 A 30 seconds maximum on, 5 minutes off
Specifications are valid from 5 % to 100 % of amplitude range. AC conversion: true rms Maximum crest factor: 3 For non-sinusoidal waveforms, add Overload protection 440 mA, 1000V fast-blow fuse
Resolution Accuracy, ±(% of Reading + Counts)
±
(2 % reading + 2 % f.s.) typical
Typical Burden
Voltage
DC Current Measurement
Range Resolution Accuracy, ±(% of Reading + Counts)
30.000 mA 0.001 mA 0.05 % + 2 14 mV/mA
1.000 A (Note) 0.001 A 0.2 % + 2 1.5 V/A
Note: 440 mA continuous, 1A 30 seconds maximum on, 5 minutes off
Overload protection: 440 mA, 1000V fast-blow fuse
Typical Burden
Voltage
39
787
Users Manual
Ohms Measurement
Range Resolution Measurement Current Accuracy, ±(% of Reading + Counts)
400.0 Ω 0.1 Ω 220 μA 0.2 % + 2
4.000 kΩ 0.001 kΩ 59 μA 0.2 % + 1
40.00 kΩ 0.01 kΩ 5.9 μA 0.2 % + 1
400.0 kΩ 0.1 kΩ 590 nA 0.2 % + 1
4.000 MΩ 0.001 MΩ 220 nA 0.35 % + 3
40.00 MΩ 0.01 MΩ 22 nA 2.5 % + 3
Overload protection: 1000 V
Open circuit voltage: <3.9 V
40
ProcessMeter™ Specifications
Frequency Counter Accuracy
Range Resolution Accuracy, ±(% of Reading + Counts)
199.99 Hz 0.01 Hz 0.005 % + 1
1999.9 Hz 0.1 Hz 0.005 % + 1
19.999 kHz 0.001 kHz 0.005 % + 1
Display updates 3 times/second at >10 Hz
Frequency Counter Sensitivity
Input Range
1 V 0.1 V
4 V 1 V
40 V 3 V
400 V 30 V
1000 V 300 V
Minimum Sensitivity (rms Sinewave)
5 Hz to 5 kHz*
* Usable 0.5 Hz to 20 kHz with reduced sensitivity.
41
787
Users Manual
Diode Test and Continuity Test
Diode test indication: display voltage drop: 0.2 mA nominal test current at 0.6V: 2.4V full scale, accuracy ±(2% + 1 count)
Continuity test indication: continuous audible tone for test resistance <100 Ω
Open circuit voltage: <3.9 V Short circuit current: 1.2 mA typical Overload protection: 1000 V rms
DC Current Output
Source mode:
Span: 0 mA or 4 mA to 20 mA, with overrange to 24 mA Accuracy: 0.05 % of span Compliance voltage: 12 V with battery voltage >8.5 V
Simulate Mode:
Span: 0 mA or 4 mA to 20 mA, with overrange to 24 mA Accuracy: 0.05 % of span Loop voltage: 24 V nominal, 30 V maximum, 15 V
minimum
Compliance voltage: 21 V for 24 V supply
Burden voltage: <3 V
42
ProcessMeter™ Specifications
General Specifications
Maximum voltage applied between
any jack and earth ground .................................... 1000 V
Storage temperature ............................................. -40 °C to 60 °C
Operating temperature .......................................... -20 °C to 55 °C
Operating altitude .................................................. 2000 meters maximum
Temperature coefficient ........................................ 0.05 x specified accuracy per °C for temperatures <18 °C or >28 °C
Relative humidity ................................................... 95 % up to 30 °C, 75 % up to 40 °C, 45 % up to 50 °C, and 35 % up to 55 °C
Vibration ................................................................. Random Mil-prf-28800f, 10 Hz to 500 Hz
Shock ...................................................................... 1 meter drop test
Power requirements .............................................. Single 9V battery (ANSI/NEDA 1604A or IEC 6LR61)
Size ......................................................................... 32 mm H x 87 mm W x 187 mm L (1.25 in H x 3.41 in W x 7.35 in L)
With holster ..................................................... 52 mm H x 98 mm W x 201 mm L (2.06 in H x 3.86 in W x 7.93 in L)
Weight .................................................................... 369 g (13 oz)
With holster ..................................................... 638 g (22.5 oz)
Safety ...................................................................... IEC 61010-1, 1000 V CAT III, Pollution Degree 2
Electromagnetic Environment .............................. IEC 61326-1, Portable
Electromagnetic Compatibility ............................. Accuracy for all ProcessMeter functions is not specified in RF field >3V/m
Applies to use in Korea only ............................ Class A Equipment (Industrial Broadcasting & Communication Equipment)
[1] This product meets requirements for industrial (Class A) electromagnetic wave equipment and the seller or user should take notice of it. This
equipment is intended for use in business environments and not to be used in homes.
[1]
43
787
Users Manual
44
Loading...