Fluke 787 User Manual

787
ProcessMeter
Users Manual
®
April 1997, Rev.3, 12/01
© 1997, 1898, 2000, 2001 Fluke Corporation, All rights reserved. Printed in U.S.A. All product names are trademarks of their respective companies.

LIMITED WARRANTY & LIMITATION OF LIABILITY

or handling. Resellers are not authorized to extend any other warranty on Fluke’s behalf. To obtain service during the war­ranty period, send your defective meter to the nearest Fluke Authorized Service Center with a description of the problem.
THIS WARRANTY IS YOUR ONLY REMEDY. NO OTHER WARRANTIES, SUCH AS FITNESS FOR A PARTICULAR PURPOSE, ARE EXPRESSED OR IMPLIED. FLUKE IS NOT LIABLE FOR ANY SPECIAL, INDIRECT, INCIDENTAL OR CONSEQUENTIAL DAMAGES OR LOSSES, ARISING FROM ANY CAUSE OR THEORY.
Since some states or countries do not allow the exclusion or limitation of an implied warranty or of incidental or consequen­tial damages, this limitation of liability may not apply to you.
Fluke Corporation Fluke Europe B.V. P.O. Box 9090 P.O. Box 1186 Everett, WA 5602 B.D. Eindhoven 98206-9090 USA The Netherlands

Table of Contents

Title Page
Introduction....................................................................................................................1
Contacting Fluke ........................................................................................................... 1
Safety Information......................................................................................................... 2
How to Get Started........................................................................................................ 5
Getting Acquainted with the Meter................................................................................ 6
Measuring Electrical Parameters................................................................................... 17
Input Impedance....................................................................................................... 17
Ranges..................................................................................................................... 17
Measuring a Composite Signal................................................................................. 17
Testing Diodes.......................................................................................................... 18
Displaying Minimum, Maximum, and Average.......................................................... 18
Using TouchHold...................................................................................................... 19
Compensating for Test Lead Resistance.................................................................. 19
Using the Current Output Functions .............................................................................. 20
Source Mode............................................................................................................ 20
i
787
Users Manual
Simulate Mode.......................................................................................................... 22
Changing the Current Span ...................................................................................... 22
Producing a Steady mA Output................................................................................. 24
Manually Stepping the mA Output ............................................................................ 25
Auto Ramping the mA Output................................................................................... 26
Power-Up Options.......................................................................................................... 27
Battery Life..................................................................................................................... 28
Using the Holster and Flex-Stand.................................................................................. 28
Maintenance.................................................................................................................. 28
General Maintenance................................................................................................ 28
Calibration................................................................................................................. 28
Replacing the Battery................................................................................................ 30
Replacing a Fuse...................................................................................................... 31
If the Meter does not Work........................................................................................ 32
Replacement Parts and Accessories............................................................................. 33
Specifications................................................................................................................. 36
Index
ii
ProcessMeter

Introduction

WWarning
Read “Safety Information” before you use the meter.
Your Fluke 787 ProcessMeter (referred to as “the meter”) is a handheld, battery-operated tool for measuring electrical parameters and supplying steady or ramping current to test process instruments. It has all the features of a digital multimeter, plus current output capability.
Your meter is shipped with a Flex-Stand holster, one set of TL75 test leads, one set of AC70A Alligator Clips, this manual, and a laminated Quick Reference Card that fits inside the holster.
If the meter is damaged or something is missing, contact the place of purchase immediately.
Contact your Fluke distributor for information about DMM accessories. To order replacement parts or spares, see Table 13 near the end of this manual.

Contacting Fluke

To order accessories, receive operating assistance, or get the location of the nearest Fluke distributor or Service Center, call:
USA : 1-888-99-FLUKE (1-888-993-5853) Canada: 1-800-36-FLUKE (1-800-363-5853) Europe: +31 402-678-200 Japan: +81-3-3434-0181 Singapore: +65-738-5655 Anywhere in the world: +1-425-446-5500
1
787
Users Manual
Address correspondence to: Fluke Corporation Fluke Europe B.V.
P.O. Box 9090, P.O. Box 1186, Everett, WA 98206-9090 5602 BD Eindhoven USA The Netherlands
Or visit us on the World Wide Web: www.fluke.com

Safety Information

The meter complies with IEC1010-1, ANSI/ISA S82.01­1994 and CAN/CSA C22.2 No. 1010.1-92 Overvoltage Category III. Use the meter only as specified in this manual, otherwise the protection provided by the meter may be impaired.
A Warning identifies conditions and actions that pose hazard(s) to the user; a Caution identifies conditions and actions that may damage the meter or the equipment under test.
International symbols used on the meter and in this manual are explained in Table 1.
WWarning
To avoid possible electric shock or personal injury:
Do not use the meter if it is damaged.
Before you use the meter, inspect the case. Look for cracks or missing plastic. Pay particular attention to the insulation surrounding the connectors.
2
ProcessMeter
Safety Information
Make sure the battery door is closed and
latched before you operate the meter.
Remove test leads from the meter before
you open the battery door.
Inspect the test leads for damaged
insulation or exposed metal. Check test leads continuity. Replace damaged test leads before you use the meter.
Do not use the meter if it operates
abnormally. Protection may be impaired. When in doubt, have the meter serviced.
Do not operate the meter around
explosive gas, vapor, or dust.
Use only a single 9V battery, properly
installed in the meter case, to power the meter.
When servicing the meter, use only
specified replacement parts.
Caution
To avoid possible damage to meter or to equipment under test:
Disconnect the power and discharge all high­voltage capacitors before testing resistance or continuity.
Use the proper jacks, function, and range for your measurement or sourcing application.
To protect yourself, adhere to the following guidelines:
Use caution when working above 30V ac rms, 42V ac pk, or 60V dc. Such voltages pose a shock hazard.
When using the probes, keep your fingers behind the finger guards on the probes.
Connect the common test lead before you connect the live test lead. When you disconnect test leads, disconnect the live test lead first.
3
787
Users Manual
Table 1. International Symbols
Symbol Meaning Symbol Meaning
Alternating current Earth ground
Direct current Fuse
CAT III
4
c
Alternating or direct current Conforms to European Union directives Refer to the manual for information about
this feature. Battery
Meets Underwriters’ Laboratories safety requirements
Overvoltage (Installation) Category III, Pollution Degree 2 per IEC1010-1 refers to the level of Impulse Withstand Voltage protection provided. Typical locations include; Mains, wall outlets, main distribution levels connected closer to the supply system but less than the primary supply system (CAT IV).
T
Conforms to relevant Canadian Standards Association directives
Double insulated
Inspected and licensed by TÜV Product Services

How to Get Started

ProcessMeter
How to Get Started
If you are familiar with the Fluke 80 Series DMM, read
“Using the Current Output Functions,” review the tables and figures in “Getting Acquainted with the Meter,” and begin using your meter.
If you are unfamiliar with Fluke 80 Series DMMs, or DMMs in general, read “Measuring Electrical Parameters” in addition to the sections referenced in the previous paragraph.
The sections following “Using the Current Output Functions” contain information about the power-up options, and battery and fuse replacement instructions.
Later, use the Quick Reference Card to refresh your memory about the various functions and features that you can use.
Display
Rotary Switch
PROCESSMETER
787
REL Hz
mA
A
COM
CAT
1000V
OUTPUT
mA
mA
V
H
MIN MAX RANGE HOLD
% STEP COARSE FINE
mV
V
V
OFF
OUTPUT 0-24mA
SOURCE SIMULATE
+ +
A
mA
0.44A
(1A/ 30 sec)
FUSED
30mA
FUSED
Figure 1. Fluke 787 ProcessMeter
Pushbuttons
Input/Output Jacks
ee014f.eps
5
787
Users Manual

Getting Acquainted with the Meter

To become familiar with the features and functions of the meter, study the following figures and tables.
Figure and Table 2 describe the input/output jacks.
Figure and Table 3 describe the input functions you
get with the first five rotary switch positions.
OUTPUT 0-24mA
1
2
SOURCE SIMULATE
+ +
A
0.44A
(1A/30 sec)
FUSED
Figure 2. Input/Output Jacks
mA
Figure and Table 4 describe the output functions you get with the last two rotary switch positions.
Figure and Table 5 describe the functions of the pushbuttons.
Figure and Table 6 explain what all the elements of the display indicate.
3
30mA
FUSED
COM
CAT
1000V
V
4
ee001f.eps
6
Table 2. Input/Output Jacks
V
ProcessMeter
Getting Acquainted with the Meter
Item Jack Measurement Functions Source Current
Function
A c A
B d mA
C
D COM
Input for current to 440 mA continuous. (1A for up to 30 seconds.) Fused with a 440 mA fuse.
Input for current to 30 mA. Fused with a 440 mA fuse.
Input for voltage to 1000V, Ω, continuity, and diode test.
Common for all measurements.
Output for dc current to 24 mA.
Common for dc current output to 24 mA.
Simulate Transmitter
Function
Output for transmitter simulation to 24 mA. (Use in series with an external loop supply.)
Common for transmitter simulation to 24 mA. (Use in series with an external loop supply.)
7
787
Users Manual
4
6
3
5
mV
V
V
mA
A
OUTPUT
mA
mA
2
OFF
1
Figure 3. Rotary Switch Positions for Measurements
ee002f.eps
8
Getting Acquainted with the Meter
Table 3. Rotary Switch Positions for Measurements
No. Position Function(s) Pushbutton Actions
ProcessMeter
A OFF B V A
C d V D d mV
E O
F mA A L
Meter off Default:
measure ac V
F
Frequency counter Measure dc V Same as above Measure dc mV Same as above Default: measure
T for continuity BLUE D test
High test lead in
cA: measure A dc BLUE selects ac
M Selects a MIN, MAX, or AVG action (see pg. 18) KSelects a fixed range (hold 1 second for auto range) I Toggles TouchHold C Toggles relative reading (sets a relative zero point)
Same as above, except diode test has only one range
Same as above, except there is only one range for each input jack position, 30 mA or 1A
High test lead in
dmA: measure mA dc
9
787
Users Manual
10
mV
V
V
mA
A
OUTPUT
mA
mA
OFF
Figure 4. Rotary Switch Positions for mA Output
1
2
ee008.eps
Getting Acquainted with the Meter
Table 4. Rotary Switch Positions for mA Output
No. Position Default Function Pushbutton Actions
ProcessMeter
A OUTPUT
[ mA
B OUTPUT
mA
J
Test leads in
SOURCE:
Source 0% mA
Test leads in
SIMULATE:
Sink 0% mA
Test leads in
SOURCE:
Source repeating 0% -100%-0% slow ramp (E)
Test leads in
SIMULATE:
Sink repeating 0% -100%-0% slow ramp (E)
% STEP X or W: Adjusts output up or down to the next 25% step COARSE Z or Y: Adjusts output up or down 0.1 mA FINE X or W: Adjusts output up or down 0.001 mA
BLUE cycles through:
Fast repeating 0% -100% - 0% ramp (P on display)
Repeating 0% -100% - 0% ramp in 25% steps (N on display)
Slow repeating 0% -100% - 0% ramp (E on display)
11
787
Users Manual
2
3
MIN MAX RANGE HOLD
% STEP COARSE FINE
REL Hz
4
H
1
78
Figure 5. Pushbuttons
Table 5. Pushbuttons
No. Pushbutton Function(s)
A
B
b Toggles the backlight
U
(BLUE)
Rotary switch in mA A Lposition and test lead plugged into c A jack: Toggles between ac and dc ampere measure
Rotary switch in O position: Selects diode test function (D) Rotary switch in OUTPUT mA Jposition: Cycles through
Slow repeating 0% -100% - 0% ramp (Eon display)
Fast repeating 0% -100% - 0% ramp (P on display)
Repeating 0% -100% - 0% ramp in 25% steps (N on display)
5
6
ee003f.eps
12
Table 5. Pushbuttons (cont.)
No. Pushbutton Function(s)
ProcessMeter
Getting Acquainted with the Meter
C
D
E
F
G
H
X
M
% STEP
Z
K
COARSE
X
I
FINE FINE
F
W
COARSE
C
Y
% STEP
T
W
Measuring: mA Output:
Measuring: mA Output:
Measuring: mA Output:
Measuring: mA Output:
Measuring: mA Output:
Measuring: mA Output:
Selects a MIN, MAX, or AVG action (see pg. 18)
Adjusts mA output up to the next higher 25% step
Selects a fixed range (hold for 1 second for auto range)
Adjusts output up 0.1 mA
Toggles TouchHold, or in MIN MAX recording, suspends recording
Adjusts output up 0.001 mA
Toggles between frequency counter and ac voltage measurement functions
Adjusts output down 0.001 mA
Toggles relative reading (sets a relative zero point)
Adjusts output down 0.1 mA
Toggles between measure and continuity functions
Adjusts mA output down to the next lower 25% step
13
787
Users Manual
6 7
8
9
5
4
10
3
2
1
11
12
Figure 6. Elements of the Display
ee004f.eps
14
Table 6. Display
No. Element Meaning
ProcessMeter
Getting Acquainted with the Meter
A Percentage display
B OUTPUT
C D
D S
E Q
F B G Numerals
H H I MINMAXAVGR
Shows the mA measured value or output level in %, in a 0-20 mA or 4-20 mA scale (change scales with power-up option)
Lights when mA output (source or simulate) is active Lights in diode test function Lights in continuity function Lights when relative reading is on Lights when the battery is low Show the input or output value Lights when TouchHold is on MIN MAX recording status indicators:
MIN means the display is showing the minimum recorded value. MAX means the display is showing the maximum recorded value. AVG means the display is showing the average value since starting recording (up to
about 35 hours continuous recording time). Rmeans MIN MAX recording is on.
15
787
Users Manual
Table 6. Display (cont.)
No. Element
J mA, DC, mV, AC,
M or kΩ, kHz
K Auto
400100030
L EP N
Meaning
Show the input or output units and multipliers associated with the numerals
Range status indicators:
Auto means autoranging is on. The number plus the unit and multiplier indicate the active range.
One of these lights in mA ramping or step output (rotary switch position mA J):
E means slow continuous 0% - 100% - 0% ramping. P means fast continuous 0% - 100% - 0% ramping. N means ramping in 25% steps.
16
ProcessMeter

Measuring Electrical Parameters

Measuring Electrical Parameters
The proper sequence for taking measurements is as follows:
1. Plug the test leads into the appropriate jacks.
2. Set the rotary knob.
3. Touch the probes to the test points.

Input Impedance

For the voltage measurement functions, input impedance is 10 M. See the specifications for more information.

Ranges

A measurement range determines the highest value the meter can measure. Most meter measurement functions have more than one range (see the Specifications).
Being in the right range is important:
If the range is too low, the display shows OL (overload).
If the range is too high, the meter will not be displaying its most accurate measurement .
The meter normally automatically selects the lowest range that will measure the applied input signal (Auto showing on the display). Press Kif you want to lock the range. Each time you press K, the meter selects the next higher range.
If you have locked the range, the meter resumes auto ranging when you change to another measurement function or you press K and hold it for 1 second.

Measuring a Composite Signal

Because the input is dc-coupled, to meaure an ac voltage or frequency with a dc bias, you must manually select the range specified in Table 7. For example, to measure 100 mV ac with 20 V dc superimposed, select the 4 V range.
Table 7. Range Requirements for Measuring a
Composite Signal
Range (ac) Max. Allowable AC + DC
400.0 mV 3 V
4.000 V 30 V
40.00 V 300 V
400.0 V 400 V 1000 V 1000 V
17
787
Users Manual

Testing Diodes

To test a single diode:
1. Insert the red test lead into the Vjack a nd black test lead into the COM jack.
2. Set the rotary switch to O.
3. Press the BLUE pushbutton so that the D symbol is on the display.
4. Touch the red probe to the anode and the black probe to the cathode (side with band or bands). The meter should indicate the appropriate diode voltage drop.
5. Reverse the probes. The meter should display OL, indicating a high impedance.
6. The diode is good if it passes the tests in steps 4 and 5.

Displaying Minimum, Maximum, and Average

MIN MAX recording stores the lowest and highest measurements, and maintains the average of all measurements.
Press M to turn on MIN MAX recording. Readings are stored until you turn the meter off, switch to another measurement or source function, or turn MIN MAX off. The beeper sounds when a new maximum or minimum is recorded. Auto power-off is disabled and auto ranging is turned off during MIN MAX recording.
Press M again to cycle through the MAX, MIN, and AVG displays. Press and hold M for 1 second to erase stored measurements and exit .
If MIN MAX recording is on continuously for over 40 hours, minimum and maximum readings are still recorded, but the displayed average no longer changes.
In MIN MAX recording, press I to suspend recording; press I again to resume recording.
18
ProcessMeter
Measuring Electrical Parameters

Using TouchHold

Note
You must have MIN MAX recording off to use TouchHold.
! Warning
To avoid possible electric shock, do not use TouchHold to determine if dangerous voltage is present. TouchHold will not capture unstable or noisy readings.
Activate TouchHold display on each new stable reading (except in the frequency counter function). Press I to activate TouchHold. This feature allows you to take measurements in situations in which it is difficult to look at the display. The meter beeps and updates the display with each new stable reading.
if you want the meter to freeze the

Compensating for Test Lead Resistance

Use the relative reading feature (Q on the display) to set the present measurement as a relative zero. A common use for this is to compensate for test lead resistance when measuring Ω.
Select the measure function, touch the test leads together, then press C. Until you press C again, or switch to another measurement or source function, the readings on the display will subtract the lead resistance.
19
787
Users Manual

Using the Current Output Functions

The meter provides steady, stepped, and ramped current output for testing 0-20 mA and 4-20 mA current loops. You can choose source mode, in which the meter supplies the current, or simulate mode, in which the meter regulates current in an externally-powered current loop.

Source Mode

Source mode is selected automatically by inserting the test leads into the SOURCE + and jacks as shown in Figure 7. Use source mode whenever you need to supply
current into a passive circuit such as a current loop with no loop supply. Source mode depletes the battery faster than simulate mode, so use simulate mode whenever possible.
The display looks the same in source and simulate modes. The way to tell which mode is in use is to see which pair of output jacks i s in use.
20
787
PROCESSMETER
ProcessMeter
Using the Current Output Functions
40
60
MIN MAX RANGE HOLD
% STEP COARSE FINE
mV
V
V
OFF
OUTPUT 0-24mA
SOURCE SIMULATE
+ +
A
mA
0.44A
(1A/30 sec)
FUSED
30mA
FUSED
REL Hz
mA
A
COM
CAT
1000V
OUTPUT
mA
mA
V
20
H
0
80
100
ee010f.eps
Figure 7. Sourcing Current
21
787
Users Manual

Simulate Mode

Simulate mode is so named because the meter simulates a current loop transmitter. Use simulate mode when an external dc voltage of 24 to 30V is in series with the current loop under test.
Caution
Set the rotary switch to one of the mA output settings BEFORE you connect the test leads to a current loop. Otherwise, a low impedance from the other rotary switch positions could be presented to the loop, causing up to 50 mA to flow in the loop.
Simulate mode is selected automatically by inserting the test leads into the SIMULATE + and jacks as shown in Figure 8. Simulate mode conserves battery life, so use it instead of source mode whenever possible.
The display looks the same in source and simulate modes. The way to tell which mode is in use is to see which pair of output jacks i s in use.

Changing the Current Span

The meter’s current output span has two settings (with overrange to 24 mA):
4 mA = 0%, 20 mA = 100% (factory default)
0 mA = 0%, 20 mA = 100%
To find out which span is selected, short the OUTPUT SOURCE + and jacks, turn the rotary switch to OUTPUT [ mA, and observe the 0% output level.
To toggle and save the current output span in nonvolatile memory (retained when the power is turned off):
1. Turn off the meter.
2. Hold down the Kpushbutton while you turn the rotary switch to OUTPUT [ mA.
3. Wait at least 2 seconds, then release K.
22
dc V
Power Supply
COM +24V
787
PROCESSMETER
ProcessMeter
Using the Current Output Functions
40
60
20
80
REL Hz
mA
A
COM
CAT
1000V
OUTPUT
mA
V
H
mA
MIN MAX RANGE HOLD
% STEP COARSE FINE
mV
V
V
OFF
OUTPUT 0-24mA
SOURCE SIMULATE
+ +
A
mA
0.44A
(1A/30 sec)
FUSED
FUSED
30mA
Figure 8. Simulating a Transmitter
0
100
ee011f.eps
23
787
Users Manual

Producing a Steady mA Output

When the rotary switch is in the OUTPUT [ mA position, and the OUTPUT jacks are connected to an appropriate load, the meter produces a steady mA dc output. The meter begins sourcing or simulating 0%. Use the pushbuttons to adjust the current as shown in Table 8.
Select either sourcing or simulating by choosing the SOURCE or SIMULATE output jacks.
If the meter cannot deliver the programmed current because the load resistance is too high or the loop supply voltage is too low, dashes (-----) appear on the numeric display. When the impedance between the SOURCE jacks is low enough, the meter will resume sourcing.
Note
The STEP pushbuttons described on the next page are available when the meter is producing a steady mA output. The STEP pushbuttons go to the next multiple of 25%.
Table 8. mA Output Adjust Pushbuttons
Pushbutton Adjustment
Z
K
COARSE
X
I
FINE FINE
F
W
COARSE
C
Y
Adjusts up 0.1 mA
Adjusts up 0.001 mA
Adjusts down 0.001 mA
Adjusts down 0.1 mA
24
ProcessMeter
Using the Current Output Functions

Manually Stepping the mA Output

When the rotary switch is in the OUTPUT [ mA position, and the OUTPUT jacks are connected to an appropriate load, the meter produces a steady mA dc output. The meter begins sourcing or simulating 0%. Use the pushbuttons to step the current up and down in 25% steps as shown in Table 9. See Table 10 for mA values at each 25% step.
Select either sourcing or simulating by choosing the SOURCE or SIMULATE output jacks.
If the meter cannot deliver the programmed current because the load resistance is too high or the loop supply voltage is too low, dashes (-----) appear on the numeric display. When the impedance between the SOURCE jacks is low enough, the meter will resume sourcing.
Note
The COARSE and FINE adjustment pushbuttons described on the previous page are available when you are manually stepping the mA output.
Table 9. mA Stepping Pushbuttons
Pushbutton Adjustment
X
M
% STEP % STEP
T
W
Adjusts up to the next higher 25% step
Adjusts down to the next lower 25% step
25
787
Users Manual
Table 10. mA Step Values
Step Value (for each span setting)
4 to 20 mA 0 to 20 mA
0% 4.000 mA 0.000 mA 25% 8.000 mA 5.000 mA 50% 12.000 mA 10.000 mA 75% 16.000 mA 15.000 mA 100% 20.000 mA 20.000 mA 125% 24.000 mA 120% 24.000 mA

Auto Ramping the mA Output

Auto ramping gives you the ability to continuously apply a varying current stimulus from the meter to a transmitter, while your hands remain free to test the response of the transmitter. Select either sourcing or simulating by choosing the SOURCE or SIMULATE jacks.
When the rotary switch is in the OUTPUT mA Jposition, the meter produces a continuously repeating 0% - 100% - 0% ramp in your choice of three ramp waveforms:
E 0% - 100% - 0% 40-second smooth ramp,
(default)
P 0% - 100% - 0% 15-second smooth ramp N 0% - 100% - 0% Stair-step ramp in 25% steps,
pausing 5 seconds at each step. Steps are listed in Table 10.
The ramp times are not adjustable. Press the BLUE pushbutton to cycle through the three waveforms.
Note
At any time during auto ramping, you can freeze the ramp simply by moving the rotary switch to
[ mA
the
COARSE, FINE
pushbuttons to make adjustments.
position. Then you can use the
, and
% STEP
adjust
26
ProcessMeter

Power-Up Options

Power-Up Options
To select a power-up option, hold down the pushbutton shown in Table 11 while turning the rotary switch from OFF to any on position. Wait 2 seconds before you release the pushbutton after powering up the meter. The meter beeps to acknowledge the power-up option.
Table 11. Power-Up Options
Option Pushbutton Default Action Taken
Change current span 0% setting K Remembers
Disable beeper T Enabled Disables beeper Disable auto power-off BLUE Enabled Disables the feature that turns off the meter
Only the setting for current span is retained when the power is turned off. The others have to be repeated for each operating session.
You may activate more than one power-up option by holding down more than one pushbutton.
Toggles between 0 and 4 mA
last setting
power after 30 minutes of inactivity. Auto power off is disabled regardless of this option if MIN MAX recording is on.
27
787
Users Manual

Battery Life

WWarning
To avoid false readings, which could lead to possible electric shock or personal injury, replace the battery as soon as the battery indicator (B) appears.
Table 12 shows typical alkaline battery life. To preserve battery life:
Use current simulation instead of sourcing when possible.
Avoid using the backlight.
Do not disable the automatic power-off feature.
Turn the meter off when you are not using it.
Table 12. Typical Alkaline Battery Life
Meter Operation Hours
Measuring any parameter or simulating current
Sourcing 12 mA into 500 12
80

Using the Holster and Flex-Stand

The meter is supplied with a snap-on holster that absorbs shocks and protects the meter from rough handling. You can turn the meter over in the holster to protect the face of the meter from scratches when carrying the meter.
The holster is equipped with a Flex-Stand bail. Some uses of the holster with Flex-Stand are shown in Figure 9.

Maintenance

This section provides some basic maintenance procedures. Repair, calibration, servicing not covered in this manual must be performed by qualified personnel. For maintenance procedures not described in this manual, contact a Fluke Service Center.

General Maintenance

Periodically wipe the case with a damp cloth and detergent; do not use abrasives or solvents.

Calibration

Calibrate your meter once a year to ensure that it performs according to its specifications. Contact a Fluke Service Center for instructions.
28
ProcessMeter
Maintenance
Holster with Flex-Stand bent
Meter in holster face down
for protection
(Store Quick Reference Card under meter)
Figure 9. Using the Holster and Flex-Stand
Holster with
Flex-Stand extended
Holster with Flex-Stand
looped over wall
Holster with probe in clip
Holster with Flex-Stand
looped over belt
ee009f.eps
29
787
Users Manual

Replacing the Battery

! Warning
To avoid electrical shock, remove test leads from the meter before you open the battery door.
Close and latch the battery door before you use the meter.
Remove test leads from the meter before you open the battery door.
Replace the battery as follows. Refer to Figure 10. Use an alkaline 9V battery, type ANSI/NEDA 1604A or IEC 6LR61.
1. Remove the test leads and set the rotary switch to OFF.
2. With a standard blade hand screwdriver, turn each battery door screw counterclockwise so that the slot is parallel with the screw picture molded into the case.
3. Lift off the battery door.
30
Figure 10. Replacing the Battery
ee007f.eps
ProcessMeter
Maintenance

Replacing a Fuse

! Warning
To avoid personal injury or damage to the meter, use only the specified replacement fuse, 440 mA 1000V fast-blow, Fluke PN
943121.
Both current input jacks are fused with a separate 440 mA fuse. To determine if a fuse is blown:
1. Turn the rotary switch to mA A L.
2. Plug the black test lead into COM, and the red test lead into cA.
3. Using an ohmmeter, check the resistance between the meter test leads. If the resistance is about 1, the fuse is good. An open means the fuse is blown.
4. Move red test lead to dmA.
5. Using an ohmmeter, check the resistance between the meter test leads. If the resistance is about 14Ω, the fuse is good. An open means the fuse is blown.
If a fuse is blown, replace it as follows. Refer to Figure 11 as necessary:
1. Remove the test leads from the meter and turn the rotary switch to OFF.
2. Remove the battery door.
3. Remove the three Phillips-head screws from the case bottom and turn the case over.
4. Gently lift the bottom of the front of the case (nearest the inpu t/output jacks) until the top unsnaps from the rear half of the case.
5. Replace the blown fuse with the exact type specified: 440 mA 1000V fast-blow fuse, Fluke PN 943121. Both fuses are the same type.
6. Make sure the rotary switch is in the OFF position.
7. Fit the top of case together, engaging the two snaps (item A). Make sure that the gasket is properly seated.
8. Close the case and reinstall the three screws.
9. Replace the battery door.
31
787
Users Manual

If the Meter does not Work

Examine the case for physical damage. If there is damage, make no further attempt to use the meter, and contact a Fluke Service Center.
Check the battery, fuses, and test leads.
Review this manual to make sure you are using the
correct jacks and rotary switch position.
If the meter still does not work, contact a Fluke Service Center. If the meter is under warranty, it will be repaired or
replaced (at Fluke’s option) and returned at no charge. See the Warranty on the back of the title page for terms. If the warranty has lapsed, the meter will be repaired and returned for a fixed fee. Contact a Fluke Service Center for information and price.
F1
F2
1
32
Figure 11. Replacing a Fuse
ee012f.eps
ProcessMeter

Replacement Parts and Accessories

Replacement Parts and Accessories
! Warning
To avoid personal injury or damage to the meter, use only the specified replacement fuse, 440 mA 1000V fast-blow, Fluke PN
943121.
Note
When servicing the meter, use only the replacement parts specified here.
Replacement parts and some accessories are shown in Figure 12 and listed in Table 13. Many more DMM accessories are available from Fluke. For a catalog, contact your nearest Fluke distributor.
To find out how to order parts or accessories use the telephone numbers or addresses shown on page 1 of this manual.
33
787
Users Manual
Table 13. Replacement Parts
Item Description Fluke PN or Model no. Quantity
BT1 9V battery, ANSI/NEDA 1604A or IEC 6LR61 614487 1 CG81Y Holster, Yellow CG81G 1 ! F1, 2 Fuse, 440 mA, 1000V fast-blow 943121 2 MP85 Case top 619962 1 MP86 Case bottom 619939 1 H2, 3, 4 Case screw 832246 3 MP89, 90 Non-skid foot 824466 2 MP8 O-ring for input/output receptacle 831933 1 MP92 Battery door 619947 1 H5, 6 Battery door fasteners 948609 2 S1 Keypad 646932 1 TL75 Standard test lead set TL75 1 AC70A Alligator clips for use with TL75 test lead set AC70A 1 TL20 Industrial test lead set TL20 Option TM1 Product Overview Manual 1586717 1 TM2 Users Manual (CD-ROM) 1586721 1 TM3 Calibration Manual (not shown) 641891 Option
34
ProcessMeter
Replacement Parts and Accessories
MP85
MP8
MP86
H2, 3, 4
F2
F1
BT1
H5, 6
S1
MP92
TL20 (Option) Industrial
Test Lead Set
AC70A
Alligator Clips
MP89, 90
Figure 12. Replacement Parts
Product Overview Manual
CD-ROM (Users Manual)
TL75
Test Lead Set
Holster
CG81Y
ee015c.eps
35
787
Users Manual

Specifications

All specifications apply from +18°C to +28°C unless stated otherwise.
All specifications assume a 5 minute warmup period.
The standard specification interval is 1 year.
Note
“Counts” means number of increments or decrements of the least significant digit.
DC Volts Measurement
Range (V dc) Resolution Accuracy, ±(% of Reading + Counts)
4.000 0.001V 0.1% + 1
40.00 0.01V 0.1% + 1
400.0 0.1V 0.1% + 1 1000 1V 0.1% + 1
Input impedance: 10 MΩ (nominal), < 100 pF Normal mode rejection ratio: >60 dB at 50 Hz or 60 Hz Common mode rejection ratio: >120 dB at dc, 50 Hz, or 60 Hz Overvoltage protection: 1000V
36
DC Millivolts Measurement
Range (mV dc) Resolution Accuracy (% of Reading + Counts)
400.0 0. 1 mV 0.1% + 1
AC Volts Measurement
Range (ac) Resolution Accuracy, ±(% of Reading + Counts)
50 Hz to 60 Hz 45 Hz to 200 Hz 200 Hz to 500 Hz
400.0 mV 0.1 mV 0.7% + 4 1.2% + 4 7.0% + 4
4.000V 0.001V 0.7% + 2 1.2% + 4 7.0% + 4
40.00V 0.01V 0.7% + 2 1.2% + 4 7.0% + 4
400.0V 0.1V 0.7% + 2 1.2% + 4 7.0% + 4 1000V 1V 0.7% + 2 1.2% + 4 7.0% + 4
Specifications are valid from 5% to 100% of amplitude range. AC conversion: true rms Maximum crest factor: 3 For non-sinusoidal waveforms, add Input impedance: 10 M Common mode rejection ratio: >60 dB at dc, 50 Hz, or 60 Hz
(nominal), < 100 pF, ac-coupled
±
(2% reading + 2% f.s.) typical
ProcessMeter
Specifications
37
787
Users Manual
AC Current Measurement
Range
45 Hz to 2 kHz
1.000A (Note) 0.001A 1% + 2 1.5V/A
Note: 440 mA continuous, 1A 30 seconds maximum Specifications are valid from 5% to 100% of amplitude range.
AC conversion: true rms Maximum crest factor: 3 For non-sinusoidal waveforms, add Overload protection 440 mA, 1000V fast-blow fuse
Resolution Accuracy, ±(% of Reading + Counts) Typical Burden
Voltage
±
(2% reading + 2% f.s.) typical
DC Current Measurement
Range Resolution Accuracy, ±(% of Reading + Counts) Typical Burden
Voltage
30.000 mA 0.001 mA 0.05% + 2 14 mV/mA
1.000A (Note) 0.001A 0.2% + 2 1.5V/A
Note: 440 mA continuous, 1A 30 seconds maximum Overload protection: 440 mA, 1000V fast-blow fuse
38
Ohms Measurement
Range Resolution Measurement Current Accuracy, ±(% of Reading + Counts)
400.0 0. 1 220 µA 0.2% + 2
4.000 k 0.001 k 59 µA 0.2% + 1
40.00 k 0.01 k 5.9 µA 0.2% + 1
400.0 k 0.1 k 590 nA 0.2% + 1
4.000 M 0.001 M 220 nA 0.35% + 3
40.00 M 0.01 M 22 nA 2.5% + 3
Overload protection: 1000V Open circuit voltage: <3.9V
ProcessMeter
Specifications
39
787
Users Manual
Frequency Counter Accuracy
Range Resolution Accuracy, ±(% of Reading + Counts)
199.99 Hz 0.01 Hz 0.005% + 1
1999.9 Hz 0.1 Hz 0.005% + 1
19.999 kHz 0.001 kHz 0.005% + 1
Display updates 3 times/second at >10 Hz
Frequency Counter Sensitivity
Input Range Minimum Sensitivity (rms Sinewave)
5 Hz to 5 kHz*
1 V 0.1 V 4 V 1 V
40 V 3 V
400 V 30 V
1000 V 300 V
* Usable 0.5 Hz to 20 kHz with reduced sensitivity.
40
ProcessMeter
Specifications
Diode Test and Continuity Test
Diode test indication: display voltage drop: 0.2 mA
nominal test current at 0.6V: 2.4V full scale, accuracy ±(2% + 1 count)
Continuity test indication: continuous audible tone for test resistance <100
Open circuit voltage: <3.9V Short circuit current: 1.2 mA typical Overload protection: 1000V rms
DC Current Output
Source mode:
Span: 0 mA or 4 mA to 20 mA, with overrange to 24 mA Accuracy: 0.05% of span Compliance voltage: 12V with battery voltage >8.5V
Simulate Mode:
Span: 0 mA or 4 mA to 20 mA, with overrange to 24 mA Accuracy: 0.05% of span Loop voltage: 24V nominal, 30V maximum, 15V minimum
Compliance voltage: 21V for 24V supply Burden voltage: <3V
General Specifications
Maximum voltage applied between any jack and earth ground: 1000V
Storage temperature: -40°C to 60°C Operating temperature: -20°C to 55°C Operating altitude: 2000 meters maximum Temperature coefficient: 0.05 x specified accuracy per
°C for temperatures <18°C or >28°C
Accuracy adders for use in RF Fields: In an RF field of 3V/m, change the accuracy specifications as follows:
For DC Millivolts Measrement, add 0.03% of range For AC Volts Measurement, add 0.37% of range For DC Current Measurement, 30.000 mA range, add
0.14% or range
For DC Current Output, add 0.02% of span Accuracy for all meter functions is not specified in RF
fields > 3V/m.
Relative humidity: 95% up to 30°C, 75% up to 40°C, 45% up to 50°C, and 35% up to 55°C
41
787
Users Manual
Vibration: Random 2g, 5 to 500 Hz Shock: 1 meter drop test Water and dust protection: Complies with IEC529 IP52
(normal operating vacuum used for dust test) Safety: Complies with IEC1010-1, ANSI/ISA S82.01-
1994 and CAN/CSA C22.2 No. 1010.1-92 Overvoltage Category III.
Certifications: CSA, UL, TÜV Power requirements: Single 9V battery (ANSI/NEDA
1604A or IEC 6LR61) Size: 32 mm H x 87 mm W x 187 mm L (1.25 in H x 3.41
in W x 7.35 in L); With holster and Flex-Stand: 52 mm H x 98 mm W x 201
mm L (2.06 in H x 3.86 in W x 7.93 in L) Weight: 369 g (13 oz); With holster and Flex-Stand: 638 g (22.5 oz)
42
—A—
Auto
Ramping, 17, 26 Ranging, 17 Stepping, 17
—B—
Battery
Replacement, 30
Buttons, 12

Index

Current output
Compliance, 24 Load impedance, 24 Ramping, auto, 26 Simulating a transmitter, 22 Sourcing, 20 Span (4-20 mA or 0-20 mA), 22 Steady, 24 Stepping, auto, 26 Stepping, manually, 25 With external loop supply, 22
—D—
—C—
Calibrating the meter, 28 Compensating for test lead resistance, 19 Composite Signals, 17
Diodes, testing, 18 Display, 15
43
787
Users Manual
—F—
Flex-Stand, 28 Freezing a reading (TouchHold), 19 Fuse, checking and replacing, 31
—H—
Holster, 28
—I—
Input/Output jacks, 7
—J—
Jacks, 7
—K—
Knob positions, 9, 11
—L—
Loop supply, external, 22
M—
mA output. Mailing address for Fluke, 2 Maintenance, 28 Malfunction, 32 Measuring, 17 MIN MAX recording, 18
See
Current output
—O—
Offset, programming an, 19 Options, power-up, 27 Output.
See
Current output
—P—
Parts, replacement, 34 Power-up options, 27 Pushbuttons, 12
44
Index
(continued)
—R—
Range
Auto, 17
Locking, 17 Ranges, 17 Relative reading, 19 Rotary switch positions, 9, 11
—S—
Safety information, 2 Screen, 15 Service and sales, 1 Settings, 27
Simulating.
See
Current output
Sourcing. Specifications, 36 Switch positions, 9, 11 Symbols, international, 4
See
Current output
—T—
Telephone numbers, 1 TouchHold, 19
—W—
Warranty.
Service, 32
WWW address for Fluke, 2
See
back of Title Page
45
787
Users Manual
46
Loading...