Calibrate temperature,
pressure, and low dc voltage
process tools in a single
benchtop instrument
Page 2
The 7526A Precision Process Calibrator
calibrates a wide range of pressure gauges,
temperature transmitters, digital process
simulators, multimeters and more.
Page 3
Versatility, precision and value
—
all combined into a single
benchtop process calibration tool
The Fluke Calibration 7526A Precision Process Calibrator offers the best balance of economy
and accuracy for benchtop calibration of pressure and temperature process instrumentation.
Incorporating an isolated measurement channel, the 7526A lets you simultaneously source
and measure voltage, current or resistance, making it easy to calibrate temperature and pressure
transmitters, RTD and thermocouple readouts, pressure gauges, digital process simulators, data
loggers, multimeters and more.
• Sources and measures dc voltage, current, resistance, RTDs and thermocouples
• Measures pressure using Fluke 700 Pressure Modules
• Measures 4-20 mA loop current
• Sources 24 V dc transmitter loop power supply
• Tests pressure and thermal switches with an automated switch-test function
• Measures thermistors up to 4 kΩ
• Stores up to nine programmable setpoints for each input/output parameter
• Accepts ITS-90 coefficients for accurate SPRT measurements
• Compatible with MET/CAL
A “best fit” for your process calibration requirements
“Doing more with less” is a requirement every process manufacturer faces today. To maintain
product quality, reduce waste, improve efficiency and conform to regulatory standards, instruments
that measure process variables such as temperature and pressure must be calibrated at regular
intervals. Selecting the right calibrator for the job, however, can be tricky—especially when
attempting to balance precision and versatility with cost. Less expensive than high-end
multi-product calibrators, yet more precise
and versatile than handheld field
calibrators, the 7526A is a “best
fit” for instrument shops that
value precision, versatility
and economy.
®
Plus Calibration Management Software
Everything you
need in one box
The 7526A packs a lot of capability
into one box, allowing you to calibrate a
wide-ranging and varied workload. One
calibrator performs all of these functions:
• Simulates and measures nine RTD and
thirteen thermocouple types
• Accurately measures pressure up to 10 000
PSI (69 MPa) when combined with Fluke 700
Pressure Modules
• Sources and measures dc voltage to within
0.004 % of reading
• Sources and measures dc resistance up to 4 kΩ
• Sources dc current from 0 mA to 100 mA
• Accurately measures dc current from 0 mA to 50 mA
• Sources 24 V dc loop power
7526A Precision Process Calibrator 3
Page 4
Intuitive user interface
The user interface includes cursor controls, function keys and a ten-key keypad, making it easy
to navigate through intuitive menus; store and recall up to nine setpoints for each input/output
parameter; enter RTD or SPRT coefficients; and easily change display units with a key press.
With two LCD displays, you can easily view both source and measurement results simultaneously.
DC voltage output terminals. 0 mV to 100 V.
Accuracy: 30 ppm (+3 µV), 1 year*. Five-way berylliumcopper binding posts reduce thermal EMFs and accept
common cable terminations (standard single/dual pin
banana plug, spade lug, pin connector, bare wire).
DC current output terminals. 0 mA to 100 mA.
Accuracy: 50 ppm, 1 year
Thermocouple input/output terminal. Accuracy: ± 0.1 °C *.
User selectable internal or external cold junction compensation
for improved thermocouple measurement accuracy. Accepts
thermocouple types: B,C,E,J,K,L,N,R,S,T,U,XK,BP
Isolated input terminals for dc voltage/
current measurement, unique switch-test
input for testing pressure and thermal
switches, and a 24 V dc loop power supply
for powering 4-20 mA transmitters.
Isolated pressure module input. Connect a Fluke
700 Pressure Module to the front panel LEMO
connector and the calibrator auto-detects the type
and value of the module.
Isolated input controls. A HART key enables
a 250 Ω loop resistor allowing HART devices
to read a digital HART signal superimposed
on the 4-20 mA loop current.
* See extended specifications
for more details.
Page 5
Don’t forget the sensor
Calibrating the electronics portion of a temperature transmitter
isonly a part of a complete calibration. You also need to
calibrate the temperature sensor itself, whether an RTD or
thermocouple. Ignoring the sensor can be a mistake, because
temperature sensors are responsible for more than 75 % of the
output errors in temperature transmitters. You can calibrate the
temperature sensor individually, or calibrate both the sensor and
the transmitter as a system using a dry-block calibrator such
as the Fluke Calibration 914X Series Field Metrology Wells. The
Field Metrology Wells were designed specifically with process
calibration in mind—optimizing speed to temperature, portability,
stability and accuracy. The 7526A and a 914X dry-block calibrator
make a perfect combination for calibrating just about any
temperature transmitter.
Source as well as measure pressure
The P5500 Series Comparison Test Pumps can be used with the 7526A and Fluke 700 Series
Pressure Modules to generate or control test pressures. The unique test port configuration
allows for easy, tool-free connections of a wide variety of connection sizes and types, including NPT, BSP, and metric threads. The P5510 includes a built-in hand pump to generate vacuum
or positive pressure up to 300 psi (2 MPa). The P5513 allows for precise control of pneumatic
pressures up to 3000 psi (20 MPa). An external pressure supply is required. The P5514 allows
for generation of hydraulic pressures up to 10 000 psi (70 MPa). The P5515 includes a built-in
priming pump and screw pump that allows for pressure
generation up to 20 000 psi (140 MPa). Both the P5514
and P5515 can be used with oil or water. Special
versions are available with ethylene
propylene seals
for use with more
aggressive fluids.
Calibration and repair services
Fluke Calibration offers extensive calibration support
and service to ensure your long-term satisfaction
and return on investment. Our worldwide network
of calibration centers offers accredited calibrations
traceable to national standards. We also offer fast,
quality repair and calibration services including a
module exchange program and full support in
setting up your lab.
7526A Precision Process Calibrator 5
Page 6
Specifications
Summary specifications
DC voltage, output
[1]
Range
Absolute uncertainty,
± (ppm of output + µV),
1 year
0 mV to 100 mV3031 µV
0 V to 1 V301010 µV
0 V to 10 V30100100 µV
0 V to 100 V301 mV1 mV
TC output and input
–10 mV to 75 mV30210 Ω
[1] All outputs are positive only unless otherwise noted
DC voltage, isolated input
RangeAbsolute uncertainty,
± (ppm of reading + mV),
1 year
0 V to 10 V500.2100 µV
10 V to 100 V502.01 mV
DC current, output
[1]
Range
0 mA to 100 mA5011 µA
[1]
For line voltages less than 95 V, ±100 ppm of reading
Absolute uncertainty,
± (ppm of reading + µA),
1 year
DC current, isolated input
RangeAbsolute uncertainty,
0 mA to 50 mA10010.1 µA
0 mA to 24 mA
(Loop Power)
[1]
Loop Power: 24 V ±10 %
[2]
HART Resistor: 250 Ω ± 3 %
[1][2]
± (ppm of reading + µA),
10010.1 µA
1 year
Resistance, output
RangeAbsolute
uncertainty,
tcal ± 5 °C,
± ohms, 1 year
ResolutionNominal
5 Ω to 400 Ω 0.0150.001 Ω1 mA to 3 mA
5 Ω to 4 kΩ0.30.01 Ω100 µA to 1 mA
Resistance, input
RangeAbsolute uncertainty,
± (ppm of reading + Ω),
1 year
0 Ω to 400 Ω 200.0040.001 Ω
0 Ω to 4 kΩ 200.040.01 Ω
Sample thermocouple accuracy, input/output
(does not include all available TC types)
Total uncertainty, one year for temperature range O °C to +5O °C. Total uncertainty, 1.0 % of full span for temperature range -10 °C to O °C. For POO module only,
compensated temperature range is 15 °C to 35 °C. 2 “Dry” indicates dry air or non-corrosive gas as compatible media.
“316SS” indicates media compatible with Type 316 Stainless Steel. “C276” indicates media compatible with Hastelloy C276. Use of pressure zero is required prior
to measurement or source. Max. overpressure specification includes common mode pressure. Modules are CE rated. Metric adapter(s}: 1/4 inch NPr female-to-male
BSP/ISO 1/4-19, tapered thread, included with all modules except P29, P3O, and P31, all modules include a NIST traceable certificate and test data.
1
2
side media
2
Low
side
mediaFitting material
0.025 %316 SSN/A316 SS
0.025 %316 SSN/A316 SS
0.040 %316 SSN/A316 SS
7526A Precision Process Calibrator 7
Page 8
Ordering Information
7526A Precision Process Calibrator
ModelDescription
7526A Precision Process Calibrator
Includes traceable calibration report, user manual CD,
getting started guide, power cord, thermocouple
shorting jumper and USB-to-serial adapter cable
Recommended Accessories
ModelDescription
Y7526ARack Mount Kit
7526A-CASECarrying Case
5520A-525A/LEADSThermocouple and Test Leads Set
FLUKE-700PTP-1 Pneumatic Test Pump
FLUKE-700LTP-1 Low-pressure Test Pump
FLUKE-700PRV-1 Pressure Relief Valve Kit
Comparison Test Pumps
ModelDescription
P5510-2MPnuematic Test Pump, vacuum to 300 psi (2 MPa)
P5513-20MPneumatic Test Pump, vacuum to 3000 psi (20 MPa)
P5514-70MHydraulic Test Pump, 0 psi to 10 000 psi (70 MPa)
P5515-140MHydraulic Test Pump, 0 psi to 20 000 psi (140 MPa)
Hydraulic Test Pump
ModelDescription
FLUKE-700HTH-1 Hydraulic Test Hose
FLUKE-700HTP-2Hydraulic Test Pump, 10 000 PSI (690 bar)
Fluke Calibration
PO Box 9090,
Everett, WA 98206 U.S.A.
For more information call:
In the U.S.A. (877) 355-3225 or Fax (425) 446-5116
In Europe/M-East/Africa +31 (0) 40 2675 200 or Fax +31 (0) 40 2675 222
In Canada (800)-36-FLUKE or Fax (905) 890-6866
From other countries +1 (425) 446-5500 or Fax +1 (425) 446-5116
Web access: http://www.flukecal.com