7526A Precision
Process Calibrator
2
Technical Data
1
5
7526A at a glance
1. DC voltage output terminals
0 mV to 100 V
Accuracy: 30 ppm (+3 µV*), 1-year
2. DC current output terminals
0 mA to 100mA
Accuracy: 50 ppm, 1-year
3. RTD/Ω output terminals (two-wire)
5 Ω to 4 kΩ
Accuracy: ± 0.05 °C, –200 °C to 630 °C,
(Pt 385, 100 Ω), 1-year
4. Thermocouple input/output terminal
Accuracy: ± 0.1 °C,
–100 °C to 800 °C (type K), 1-year
5. Four-wire RTD/Ω input terminals
Accuracy: ± 0.02 °C, –80 °C to 100 °C
(Pt 385, 100 Ω), 1-year
6. Isolated pressure module input.
Accuracy determined by the pressure
modules
7. Primary input/output controls
8. Cursor controls
9. Isolated input terminals for dc
voltage/current measurement,
switch-test input and 24 V dc
loop power supply
10. Isolated input controls
3
10
4
6
7
8
9
Versatility, precision and value, combined into
a single benchtop process calibration tool
The Fluke Calibration 7526A Precision Process Calibrator
offers the best balance of economy and accuracy for benchtop
calibration of temperature and pressure process instrumentation. Easily calibrate RTD and thermocouple readouts, pressure
gauges, temperature transmitters, digital process simulators,
data loggers, multimeters and more.
• Sources and measures dc voltage, current, resistance,
RTDs and thermocouples
• Precision pressure measurement using Fluke 700 or
525A-P series pressure modules
• Includes 24 V dc transmitter loop power supply
• Measures 4-20 mA loop current
• Includes automated switch-test function
• Accepts ITS-90 coefficients for accurate SPRT
measurements
• Compatible with MET/CAL
®
Calibration Software
A “best fit” for your process calibration
requirements
In today’s competitive global markets with the need to
continuously cut manufacturing costs, precise temperature
and pressure process control is required to maintain product
quality, reduce waste and ensure compliance to regulatory
standards. Regular calibration of a wide variety of process
measurement instrumentation is required to meet these
* Floor specication increases with range.
See Specications for more information.
standards. The 7526A Precision Process Calibrator puts all the
necessary tools for process instrumentation calibration into
one box. The 7526A simulates and measures nine RTD and
thirteen thermocouple types, accurately measures pressure
to within 0.008 % of reading when combined with Fluke
525A-P Series Pressure Modules, sources and measures dc
voltage from 0 to 100 V to within 0.004 % of reading, sources
dc current from 0 mA to 100 mA, accurately measures dc
current to within 0.01 % from 0 mA to 50 mA and supplies
24 V dc loop power. Combined with MET/CAL
®
Calibration
Software, the 7526A is an efficient, versatile and affordable
calibrator that truly is a best fit for your process calibration
requirements.

Consolidated specications
DC voltage, output
[1]
Range
Absolute Uncertainty, ±
(ppm of output + µV), 1-yr
Resolution
0 to 100 mV 30 3 1 µV
0 to 1 V 30 10 10 µV
0 to 10 V 30 100 100 µV
0 to 100 V 30 1 mV 1 mV
TC Output and Input
–10 mV to 75 mV 30 2 10 Ω
[1] All outputs are positive only unless otherwise noted
DC voltage, isolated input
Range Absolute Uncertainty, ±
(ppm of reading + mV),
1-yr
Resolution
0 to 10 V 50 0.2 100 µV
10 V to 100 V 50 2.0 1 mV
DC Current, output
Range[1] Absolute Uncertainty,
± (ppm of reading + µA),
1-yr
Resolution
0 to 100 mA 50 1 1 µA
[1] For line voltages less than 95 V, ±100 ppm of reading
DC current, isolated input
Range Absolute Uncertainty, ±
(ppm of reading + µA),
1-yr
Resolution
0 mA to 50 mA 100 1 0.1 µA
0 mA to 24 mA [1][2]
(Loop Power)
100 1 0.1 µA
[1] Loop Power: 24 V ±10% [2] HART Resistor: 250 Ω ± 3%
Resistance, output
Range Absolute
Uncertainty,
tcal ± 5 °C, ±
Ohms, 1-yr
Resolution Nominal
current
5 Ω to 400 Ω 0.015 0.001 Ω 1 to 3 mA
5 Ω to 4 kΩ 0.3 0.01 Ω 100 µA to 1 mA
Resistance, input
Range Absolute Uncertainty, ±
(ppm of reading + Ω), 1-yr
Resolution
0 Ω to 400 Ω 20 0.004 0.001 Ω
0 Ω to 4 kΩ 20 0.04 0.01 Ω
Sample RTD and thermistor, intput
(does not include all available RTD types)
RTD Type Temperature Range (°C) Absolute
[1]
Min Max
Uncertainty,
tcal ±5 °C,
± (°C), 1-yr
Pt 385, 100 Ω –80 100 0.020
100 300 0.024
YSI 400 15 50 0.007
[1] See extended specifications for all RTD types: Pt-100 (385, 3926, 3916), Pt200, Pt-500, Pt-1000, Ni-120, Cu-427, SPRT
General specications
Standard interface RS-232, IEEE-488 (GPIB)
Temperature
performance
Electromagnetic
compatibility
Temperature coefficient Temperature coefficient for temperatures
Relative humidity Operating: <80% to 30 °C
Altitude Operating: 3,000 m (9,800 ft) max
Safety EN/IEC 61010-1:2010 3rd Edition,
Analog low isolation 20 V
Line power 120 V~: 100 V to 120 V
Line frequency 47 Hz to 63 Hz
Line voltage variation ± 10 % about setting
Power consumption 15 VA maximum
Dimensions Height: 14.6 cm (5.75 in)
Weight
(without options)
Operating: 0 °C to 50 °C
Calibration (tcal): 18 °C to 28 °C
Storage: –20 °C to 70 °C
CE: Conforms to EN61326;
operation in controlled EM environments
outside tcal 5 °C is 10% of the 90-day
specification (or 1 year if applicable) per °C
<70% to 40 °C
<40% to 50 °C
Non-operating: 12,200 m (40,000 ft) max
UL 61010-1:2012, CAN/CSA 22.2 No. 61010-1-12
240 V~: 220 V to 240 V
Width: 44.5 cm (17.5 in)
Depth: 29.8 cm (11.75 in)
4.24 kg (9.35 lb)
Sample thermocouple accuracy, input/output
(does not include all available TC types)
TC type Temperature Range (°C) Absolute
[1]
Min Max
Uncertainty,
tcal ±5 °C,
± (°C), 1-yr
J –210 1200 0.09
K –250 1372 0.1
S –50 1767 0.29
T –250 400 0.11
[1] See extended specifications for all TC types (B,C,E,J,K,L,N,R,S,T,U,XK,BP)
[2] Best accuracy within specified TC temperature range
Sample RTD and thermistor, output
(does not include all available RTD types)
RTD Type Temperature Range (°C) Absolute
[1]
Min Max
Uncertainty,
tcal ±5 °C,
± (°C), 1-yr
Pt 385, 100 Ω –200 630 0.05
YSI 400 15 50 0.007
[1] See extended specifications for all RTD types: Pt-100 (385, 3926, 3916), Pt200, Pt-500, Pt-1000, Ni-120, Cu-427, SPRT
2 Fluke Calibration 7526A Precision Process Calibrator
[2]
Fluke Calibration. Precision, performance, confidence.
Fluke Calibration
PO Box 9090,
Everett, WA 98206 U.S.A.
Fluke Europe B.V.
PO Box 1186, 5602 BD
Eindhoven, The Netherlands
For more information call:
In the U.S.A. (877) 355-3225 or Fax (425) 446-5116
In Europe/M-East/Africa +31 (0) 40 2675 200 or Fax +31 (0) 40 2675 222
In Canada (800)-36-FLUKE or Fax (905) 890-6866
From other countries +1 (425) 446-5500 or Fax +1 (425) 446-5116
Web access: http://www.flukecal.com
©2012 Fluke Calibration. This information is preliminary and subject to change.
Printed in U.S.A. 7/2012 4253244B_EN
Pub_ID: 11959-eng
Modification of this document is not permitted without written permission
from Fluke Calibration.
™