This Fluke product will be free from defects in material and workmanship for three years from the date of purchase. This
warranty does not cover fuses, disposable batteries, or damage from accident, neglect, misuse, alteration, contamination, or
abnormal conditions of operation or handling. Resellers are not authorized to extend any other warranty on Fluke’s behalf.
To obtain service during the warranty period, contact your nearest Fluke authorized service center to obtain return
authorization information, then send the product to that Service Center with a description of the problem.
THIS WARRANTY IS YOUR ONLY REMEDY. NO OTHER WARRANTIES, SUCH AS FITNESS FOR A PARTICULAR
PURPOSE, ARE EXPRESSED OR IMPLIED. FLUKE IS NOT LIABLE FOR ANY SPECIAL, INDIRECT, INCIDENTAL OR
CONSEQUENTIAL DAMAGES OR LOSSES, ARISING FROM ANY CAUSE OR THEORY. Since some states or countries
do not allow the exclusion or limitation of an implied warranty or of incidental or consequential damages, this limitation of
liability may not apply to you.
Fluke Corporation
P.O. Box 9090
Everett, WA 98206-9090
U.S.A.
11/99
Fluke Europe B.V.
P.O. Box 1186
5602 BD Eindhoven
The Netherlands
The Fluke 729 and 729 FC Automatic Pressure
Calibrators (the Product)
calibration tools for lab or field use. This battery-operated
product performs automatic pressure calibration of
transmitters, pressure switches, and gauges. The
automatic functions are done by the Product or by
downloaded tasks from calibration-management software
(CMS). The Product also supplies basic HART
communicator functions when used with HART-capable
transmitters. “FC” designates Fluke Connect
this manual.
Product key features and benefits include:
•Automatic pressure calibration of transmitters,
pressure switches and pressure gauges.
•Source and measure pressure to test and calibrate
pressure sensors, transmitters, and other pressure
instruments.
are portable field pressure
®
®
throughout
•Source and simulate milliamp signals while
measuring pressure for testing current to pressure
converters (I/P).
•HART (Highway Addressable Remote Transducer)
communication to configure and perform trim on
HART smart-pressure transmitters.
•Provides Loop Power and simultaneously measures
mA output from a connected device.
•Measures pressure with any of 50 Fluke-750P Series
pressure modules.
Power button. Push to turn on or turn off
the Product.
BACK button. Push to move to the
previous user interface (UI) screen. When
navigating through the screens, BACK can
move out of most screens and with
repeated pushes, can take you all the way
to the startup screen.
F1-F4 softkeys. Performs the specified
function of the label above each softkey on
the display. These labels change
throughout the UI.
Arrow buttons. Push to move the display
cursor left, right, up, or down. Cursor
highlights in yellow over the selection.
SELECT button. Push to make a selection
on the UI. This button cannot confirm
number or character choices (push
ENTER).
Display brightness button. Push to change the
display brightness from dim to bright and back again.
Numeric keypad. Used when a numeric entry is
necessary.
ENTER button. Push to make a selection on the UI.
Only ENTER confirms numbers and characters.
Confirm all editable items with ENTER.
CLEAR (ZERO) button. Push to clear highlighted
entry. When PI/PV/IP/SWT starts, push to zero the
pressure reading.
mA button. Push to select the mA measure function.
SOURCE PRESSURE button. Use this function
to output (source) a target pressure from the
Product. Use the arrows to select the Setpoint
field and enter the values with the number
buttons.
VDC button. Push to select the measure dc
voltage function.
SETUP button. Push to enter the Setup menu
to change operating parameters. See Setup Menu.
TASKS button. Push to enter the Tasks menu.
See TASKS Menu.
VENT button. Push to release pressure at the
internal pressure port. VENTING shows on the
display as the Product vents. VENTED shows
after venting is complete.
MEASURE PRESSURE button. Use this function
for pressure measurement mode.
HART button. Push to enter HART
communication mode.
HOLD button. Push to freeze the reading on the
display. Push it a second time to release the hold
on the display. Push and hold to capture the
screen and save into memory.
External display shows selected device that is
connected. The device can be a pressure module,
Setpoint indicator
temperature probe, process variable (PV) of
connected HART device, or Fluke Connect
secondary device reading.
Fluke Connect on indicator (729 FC only) Softkey indicator
HART on indicator Current, voltage, or pressure switch display
24 volt loop power active indicator Main display with measured or sourced value
Battery charge level Time and date display
Pressure port status
When the third function (RTD, External Pressure
Module, HART Transmitter, or FC secondary device) is
connected, the Product changes to triple-function display
automatically. In the subsequent triple-function display,
the original bottom half of the display is divided into two
parts:
• The left display continues to show the electrical
measurement display.
• The right display shows the model name, reading,
and units from the selected third function. See
Figure 5. The figure shows a transmitter device
(3051C).
When more than one third function is available, use the
arrow keys to highlight the device (in yellow) in the right
display. Then use the softkeys to select the device to
show.
The display can show RTD temperature measurements in
the selected unit (C/F) as the third-function. When the
temperature unit field highlights, push the softkey to toggle
the temperature setting. The ohms reading shows.
Note
The Product supports PT100-385 RTD only.
Pressure Module Connection
The Product automatically detects external pressure module
installation or removal. Move the cursor to the unit type and
push E to configure the pressure-unit type (it can be
different from the pressure unit of internal pressure). Change
the unit type of the internal pressure to automatically change
the unit type of the pressure module.
Figure 6 shows a pressure module as the third function.
•When you plug in an external pressure module, the
pressure-unit type defaults to the internal pressure unit
type.
•When you remove or install an external pressure
module, the zero offset resets.
Figure 6. Triple-Function Display with External Pressure
Module Connected
Battery charger/universal power supply port. Use the battery charger for bench-top applications that have ac
line power available. Use the battery charger to charge the battery while installed in the Product.
USB port – Used for communication between the Product and a PC.
Push to drain condensation (see Maintenance Menu).
Drain port fitting connector.
External temperature port for RTD Temperature Probe
External pressure module port
Pressure connection
COM Port
Measure, source, and simulate mA; measure V dc and supply loop power.
Use the multi-page Setup menu to configure the Product
features.
Note
For some of the submenu functions, the Product
password is required. The default password is
1234.
To access the Setup menu, push S.
The Setup menu includes submenus for Product
configuration. The submenus are:
1. 729 Information
2. 729 Setup
3. Manage Users
4. Manage Test Results
5. Manage Screen Shots
6. Manage Custom Tasks
7. Maintenance
Push the arrow buttons, R, or E to access these
submenus. Alternatively, push the corresponding number
button for access. Push F to go to the previous menu.
To clear all stored files, push D (Clear All Files).
Enter the Product password to use this function. This
action clears all stored test results, screen shots, and
custom tasks.
The subsequent sections explain the Setup menu and
submenus.
Manage FC Devices (729 FC Only)
1. Enter the Setup menu.
2. Push B (Manage FC Device) to set 729 FC as
the primary device and discover other units that have
Fluke Connect. The connected product shows on the
display.
Push C (Enable Fluke Connect) to set 729 FC as
the secondary device so it can be discovered by the
Fluke Connect App on smart phone.
Locator
When the Product shows several FC devices on the
Fluke Connect screen, highlight a product and push
B (Locator). This causes the Fluke Connect button
on the listed product to flash repeatedly. The FC device
exits locator mode automatically or when B
(Locator) is pushed again.
The 729 Information submenu is an information-only
screen that shows information about the Product. The
screen shows:
• The Product model (729 or 729 FC)
• Serial Number
• Firmware Revision
• The last calibration date of the Product pressure
components.
•The last calibration date of the Product electrical
components.
There are also softkeys at the bottom of this screen.
These are:
•A - Calibrate Pressure – Follow the prompts
on the screen and the Product takes you through a
procedure that calibrates the pressure components
of the Product. Enter the Product password to use
this function. D (Options) lets you save the
calibration, go back a step, skip a step, or move to
the next section. See Calibration Manual for details.
•B - Calibrate Electrical – Follow the prompts
on the screen and the Product takes you through a
procedure that calibrates the electrical components
of the Product. Enter the Product password to use
this function. D (Options) lets you save the
calibration, go back a step, skip a step, or move to
the next section. See Calibration Manual for details.
•C - Battery Information – An information-only
screen that shows if the battery is present, the
remaining battery charge level, battery temperature,
and battery voltage.
The 729 Setup submenu shows the information in
Table 5.
Table 5. 729 Setup Menu
Menu Parameter Function
Push E and use the arrows to
select the UI language: English,
Language
Date and Time
Simplified Chinese, Traditional
Chinese, French, German, Italian,
Portuguese, Spanish, Russian,
Japanese, or Korean.
Push E and use the arrows to
access the Date and Time menu.
Once inside the menu, use the
arrows to select a parameter and
use the number keys to make
changes to these settings. For am or
pm, push R. Once you make
changes, push D (Done) to
store the new parameters. The
timestamp saves with any stored
results.
Table 5. 729 Setup Menu (cont.)
Menu Parameter Function
Set the format of the date. Use
Date Format
Time Format
Number Format
Temperature Unit
the softkeys to select
yyyy-mm-dd, mm/dd/yyyy, or
dd/mm/yyyy.
Push the necessary softkey to
select 12-hour or 24-hour time
format. The format changes in
the Date Format menu. E or
R also changes the format.
Push the necessary softkey to
select a decimal point or comma
for the numbering format. E or
R also changes the format.
Push the necessary softkey to
[1]
select between °C or °F. E or
R also changes the units.
The Product has a programmable
Auto Battery Timeout feature to
save the battery charge. Push the
arrow keys to highlight this line.
Push E or R to reach the
Auto Battery
Timeout
PCM Resolution
HART
Communication
Fluke Connect Enable/Disable
timeout menu. The choices are:
• Never
• 5 minutes
• 10 minutes
• 15 minutes
• 30 minutes
• 60 minutes
Choices are High and Low. High
resolution is the normal state, and
Low reduces the resolution one
digit.
Enable/Disable
Manage Users
The Manage Users submenu controls access to add or
delete Product user names from this menu:
• Add a new user: push A (New):
Use the onscreen keyboard to add a new user. Push
the softkeys for Capslock ON (or OFF), Backspace,
and to complete the entry (Done).
Use this screen to check the Product buttons. Push any
button and verify the onscreen response.
For the power button, push and hold for 2 seconds.
Exhaust
The Exhaust function releases all pressure inside the
Product including the pressure reservoir. This reservoir
keeps the pressure stable and provides pressure if only
small pressure steps are necessary. Push A for
Exhaust.
Tasks Menu
Tasks let you set up calibration and measurement tasks
to save and recall for later use.
To go to the Tasks menu, push T. The Tasks menu
has these submenus:
• Pressure Transmitter (P/I)
• Pressure Switch (SW)
• Current to Pressure Test (I/P)
• Pressure Leak Test
• Pressure Transmitter (P/V)
• Pneumatic Pressure Transmitter (P/P)
Note
P/P tasks require a pressure module connected
to the Product.
Use this feature to calibrate a pressure to voltage
transmitter with voltage measurement. For connections, see
Figure 11.
1. From page 1 of the Tasks menu, push the arrows to
highlight Pressure Transmitter (P/V).
2. Push E or R.
3. From this screen, use the arrows, numeric keypad, and
E to change the values of the source pressure and
measure voltage.
Push D (Continue) to edit these other
parameters:
• Tolerance of Range
• Test Strategy
• Auto Settling Time
• Transfer Function A (Linear or Square
Root).
Push B (Save as Customized) to save these
settings as a customized task. Recall these customized
tasks from the Tasks menu at any time.
Pressure Transmitter (P/P)
To calibrate a pressure-to-pressure transmitter, use this
feature. This feature requires a pressure module to
measure pressure output from a connected transmitter.
To use the function, see Figure 15:
1. From page 1 of the Tasks menu, push the arrows
to highlight Pressure Transmitter (P/P).
2. Push E or R.
3. From this screen, use the arrows, numeric keypad,
and E to change the values of the source
pressure and measure pressure.
4. Push D (Continue) and these parameters
can be edited:
• Tolerance of Range
• Test Strategy
• Auto Settling Time
5. Push B (Save as Customized) to save these
settings as a customized task. Recall this
customized task from the Tasks menu at any time.
Push G and the Product searches (polls) the bus to find
any connected HART devices. From this screen, there are
these softkeys:
A - Communicator Only
B - Loop Power +24V ON
C - Enable HART 250Ω
D - Restart BUS Polling
Once Polling finds the transmitter(s), move the cursor to the
desired tag number and select D (Continue) to select.
Loop Power +24V ON
The Product supplies loop power at 24 V dc to a current
transmitter disconnected from the system. To supply 24 V
loop power:
1. From the HART menu, push B (Loop Power
+24V ON). The Product polls the bus to see if a
transmitter is connected.
2. With the transmitter disconnected from normal loop
power, connect the Product test leads in series with the
instrument current loop as shown in Figure 16.
The Product has a selectable 250 Ω HART resistor to
facilitate use with HART devices. Enable the HART resistor
when you connect the Product to a HART device. The
HART resistor defaults to OFF.
To turn on or off the HART resistor, push C (Enable HART 250Ω) or (Disable HART 250Ω).
HART Menu
The HART menu has these functions that configure and
calibrate a HART device:
• HART Data
• HART Service
• Calibrate (Ad Hoc)
• Find Task by Tag
• Bus Polling
From the HART menu, push A (Disconnect) to
disconnect from HART communication, or push D
(Continue) to continue to the highlighted entry.
HART Data
The HART data screen shows information about the
connected HART device. Push B (Page Up) or C
(Page Down) where necessary.
Highlight the HART Service menu selection and push
D (Continue). When you enter some of the
submenus, a warning tells you to remove the loop from
AUTOMATIC control. Push D (Continue).
PV Zero Trim
This function sets the digital pressure value of the
transmitter to zero.
1. Push C (Trim Zero) to trim the PV output value to
zero.
2. Push D (Done) when completed.
mA Output Trim
In the mA output trim screen, you can adjust the transmitter
output.
1. Start with the 4 mA range. Push A (Fetch) to get
the current measured mA value.
2. Push B (Trim 4 mA) to send the measured value
for correction. Push C (Next Point). The Product
moves to the 20 mA range.
3. Repeat steps 1 and 2.
4. When complete, push C to return to the previous
point or D (Done).
Trim to Applied Values
1. Enter the LRV pressure value necessary in the Setpoint
entry and then push E and let the pressure settle.
2. Push B (Trim LRV) to trim the lower range of
pressure applied.
3. Enter the necessary URV pressure value in the
Setpoint entry and push E and let the pressure
settle.
4. Push C (Trim URV) to trim the upper range of
pressure applied.
Enter the necessary mA value to output from the connected
transmitter.
• Push B (Step Up 4mA) to increase mA output in
4 mA steps.
• Push C (Step Down 4mA) to decrease mA output
in 4 mA steps.
Re-range Transmitter
Change the Upper Range Value (URV) and the Lower
Range Value (LRV) of the connected transmitter from this
screen.
1. Enter the Upper Range Value (PV URV) necessary.
2. Push E to select.
3. Push C (Send) to change the URV.
4. Enter the Lower Range Value (PV LRV) necessary.
5. Push E to select.
6. Push C (Send) to change the LRV.
Device Diagnostics
Select Device Diagnostics to send the self-diagnostics
command to the connected transmitter and view the
transmitter Self Test result.
Calibrate (Ad hoc)
Calibrate (Ad Hoc) calibrates a connected HART device
without a downloaded task.
The opening screen shows the task type and
source/measure configuration for the device. The
transmitter HART configuration is copied into this
configuration screen. Correct as needed for your test.
1. Push B (PV Calibrate) to change the measure
from mA to the PV source value of the transmitter.
The next screen starts the As Found Calibration. These
softkeys are active:
A – Abort
B – Manual Test
D – Auto Test
Test results show after all strategy points have recorded the
measurements. Measurements in specification show as
black. Measurements out of specification show as red.
1. Push D (Done) or A (Abort).
2. Enter the Tag ID, serial number of the device, and the
User ID.
3. Push D (Done)
Adjust
Push B (HART Adjust). These choices are shown:
• PV Zero Trim
• mA Output Trim
• Trim to Applied Values
PV Zero Trim
To set the digital pressure value of the Product to zero:
1. Push C (Trim Zero).
2. Push D (Done).
mA Output Trim
From the mA output trim screen, you can fetch the current
PV value and send this value to the transmitter to adjust the
PV to the correct mA output.
1. Starting with the 4 mA range, push A (Fetch) to
get the current measured mA value.
2. Push B (Trim 4mA) to send the measured value
for correction.
3. Push C (Next Point). The Product moves to the
20 mA range.
4. Repeat steps 1, 2, and 3.
5. When complete, push Cto return to the previous
point or D (Done).
1. Enter the LRV pressure value necessary in the Setpoint
entry and let the pressure settle.
2. Push B (Trim LRV) to trim to the lower range of
pressure applied.
3. Enter the URV pressure value necessary in the
Setpoint entry and let the pressure settle.
4. Push C (Trim URV) to trim to the upper range of
pressure applied.
As Left
Push D (As Left). Repeat the procedure from As
Found.
Calibration results save to the Test Results menu.
Find Task by Tag
Downloaded tasks from DPCTrack2 or other supported
software are shown. Use the up and down arrows to scroll
through the tasks. Push B (Page Up) or C
(Page Down) where necessary. Push D (Recall) to
recall the highlighted task.
Bus Polling
Bus Polling goes back into the bus polling screen and scans
for connected HART devices.
If no HART device is found after Polling has completed,
push D (Restart Bus Polling) to restart bus polling to
search of the Product BUS for HART devices. The Product
locates the device.
Measure Mode
The measure mode has these functions:
• mA dc measure
• mA dc source
• mA dc simulate
• V dc measure
• Switch (continuity)
The lower left of the display shows the operation mode.
The Product supports the 700P and 750P series pressure
modules. See Accessories. Before you use a pressure
module, read its instruction sheet. The modules are different
in how they are used, zeroed, what types of process
pressure media are allowed, and accuracy specifications.
Figure 18 shows gage and differential modules. Differential
modules also operate in gage mode when you leave the low
fitting open to atmosphere.
To measure pressure, attach the applicable pressure
module for the process pressure you will test as described
in the module’s instruction sheet.
To measure pressure:
W Warning
To prevent personal injury, shut off the valve
and slowly bleed off the pressure before
attaching the pressure module to the pressure
line to avoid a violent release of pressure in a
pressurized system.
To prevent possible damage to the Product or
to equipment under test:
•Never apply more than 10 lb-ft. of torque
between the pressure module fittings, or
between the fittings and the body of the
module.
•Always apply correct torque between the
pressure module fitting and connecting
fittings or adapters.
•Never apply pressure above the rated
maximum printed on the pressure module.
•Use the pressure module only with
specified materials. See the printing on the
pressure module or the pressure module
instruction sheet for the acceptable
material compatibility.
Connect a pressure module to the Product as shown in
Figure 19. The pressure module shows on the display after
several seconds once installed.
The Product automatically senses the pressure module
attached and sets its range accordingly.
Note
Before doing a task that sources or measures
pressure, zero the Product.
1. With the pressure module highlighted in the third
screen, push D (Zero External) to zero the
pressure. When zeroing absolute pressure modules,
enter the current barometric pressure when the
displayed value turns to yellow highlight
2. Zero the pressure. See the module’s Instruction Sheet.
Modules can have different zeroing procedures that
depend on module type.
When ramped, the source sweeps up or down in value. Use
the Auto Ramp feature to verify a switch or alarm, or when a
smooth increase or decrease of the output function is
necessary. Set the Product to ramp up Pressure or Current.
1. From the main screen, push D (Auto Ramp).
2. Set the Product for the necessary source function:
• B (Pressure)
• C (Current)
3. Use the arrows, numeric keypad, and E to change
the Start Values, End Values, and Slew Rate (Ramp
Time).
4. Select the Repeat Mode:
• B (One Time)
• C (Repetitive)
5. Select the Step Style:
• B (Sawtooth)
• C (Triangle)
6. Push D (Start).
While the signal ramps, the output adjusts to the value. The
selection of endpoints and ramp time determines the size of
steps. For example, if you set the Product to ramp from
1 mV to 1 V over 10 seconds, the output adjusts in
approximately 25 mV steps. The Ramp function continues
until you get the selected limit.
To measure temperature with an RTD probe accessory,
connect the probe into the RTD port on the right of the
Product. Make sure the correct probe type is used. Use
Fluke-720RTD, PN 4366669 or the Fluke-720URTDA,
PN 4382695.
The standard probe has a 10-inch insertion depth with a
¼-inch diameter stainless steel sheath. See Figure 20.
Note
The factory default type is PT100-385 so if the
Product is used with the Fluke 720 RTD Probe
(PN 4366669), it is not necessary to set the probe
type. Connect the probe to the Product and
configure the display to read temperature.
The display shows “OL” when the measured
temperature is outside the nominal measurement
range of the RTD function (below -40 °C or above
150 °C).
The operation mode (for example, measure or source)
shows on the display. If the Product is not in mA source
mode, push M.
To source mA:
1. Connect the test leads as necessary.
2. Key in the necessary output value (for example, to
source 5.5 mA, push 5.5E).
3. Push E.
4. To change the output value, record a new value and
then push E.
4. To set the output value in the present source function,
push P then enter the desired value and push E.
5. To turn off sourcing completely, select another function.
Note
Use the source current function to drive a current
input device. This is different from the loop power
function in which the Product is powering a
process instrument.
4 to 20 mA Simulation
To configure the Product as a load on a current loop:
1. Move the cursor to the analog function indicator and
push D (Simulate Current).
2. Enter the necessary value.
3. Connect an external loop supply to the mA port as
shown in Figure 21.
Users can record a series of pressure or mA
measurements for later upload to a host computer. The
Product records a maximum of 8000 readings, depending
on the reading rate, duration, and how much memory is in
use for other functions such as tasks or stored results.
Record the reading rate and duration in minutes.
To log data:
1. From the main screen, push B (Log). Adjust
the Reading Rate and Duration with the arrow
buttons and then push E to select parameters
from the Log screen.
2. Choose the Log Source from any of the available
measured values with A, B, or C.
The log source can be internal pressure, external
pressure, mA, V dc, HART PV, or temperature.
3. For Reading Rate: push E to adjust the reading
rate from 1 /m to 60 /m.
4. For Duration: use the numeric keypad to record the
duration in minutes, followed by E. The maximum
depends on the reading rate and how much memory
is available to log data.
5. Push D (Start) to begin. The Product logs the
measured values of the selected source for the preselected duration. Push A (Abort) to stop
logging.
Results show after the Product logs the data. Push
A (Abort) or D (Done). If you push D
(Done), enter Tag, S/N, and a User ID. The Product
saves the results and are recalled under the
Setup>Manage Test Results menu. Upload the logged
Product data to DPCTrack2 for review in a graphical
report.
Communication with a PC
Download procedures from a PC to the Product and
upload test results to a PC from the Product. A PC,
Microsoft Windows, USB cable (supplied), and Fluke
DPCTrack2 application software, or a qualified Fluke
partner’s software are required. See the DPCTrack2
Users Manual for further instructions. See Figure 22 for
The Product features a rechargeable battery. Charge the
battery while it is inside or outside of the Product. This
allows you to have more than one fully charged battery on
hand.
Charge the Battery
Before the Product is used, charge the battery. To charge
the battery while in the Product, connect the battery
charger to the Product.
The battery fully charges in 8 hours.
To charge the battery outside of the Product, see
Figure 24 for battery access:
1. Place the Product face down.
2. Lift the Product stand to expose all screws.
3. Remove the six screws with a Phillips screwdriver.
4. Pull off the back.
5. Remove the battery.
6. Connect the battery charge to the input on the
battery.
The battery charge indicator (top-right of display) shows
while the battery is outside of the Product. Solid green
bars show the level of charge on the battery. When all
bars are illuminated and solid, the battery is fully charged.
The bars progressively illuminate to show that the battery
is currently charging.
Battery Life
The battery charge indicator shows on the upper right of
the display.
Table 6 shows the typical operation time for a new, fully
charged battery. Product performance meets its
specification until the battery charge indicator reads
empty.
Table 6. Typical Battery Life
Operation Modes Battery Life
Measure, continuous 20 Hours
Measure and source,
with loop power on,
continuous
To prevent possible electrical shock, fire, or
personal injury:
•Have an approved technician repair the
Product.
•Do not operate the Product with covers
removed or the case open. Hazardous
voltage exposure is possible.
•Remove the input signals before you
clean the Product.
• Use only specified replacement parts.
• Run “Exhaust” before you open the battery
door.
Clean the Product
Clean the Product and pressure modules with a soft cloth
dampened with water or water and mild detergent.
WCaution
To prevent possible damage to the Product,
do not use solvents or abrasive cleansers.
Clean the Pump Valve
1. Remove the battery door (see Battery Replacement)
and locate the pump valve caps, see Figure 23 ().
2. Use a small screwdriver to remove the two valve
retention caps located in the oval shaped opening on
the underside of the Product.
3. Gently remove the spring and O-ring assembly.
4. Set aside the valve assemblies and clean out the
valve body with a cotton swab soaked in isopropyl
alcohol (IPA).
5. Repeat this process several times with a new cotton
swab until there is no sign of residue.
6. Run the pump for a few seconds.
7. Clean the O-ring assembly and O-ring on the
retention caps with IPA and inspect the O-rings
closely for any cuts, nicks, or wear. Replace if
necessary.
8. Inspect the springs for wear or loss of tension. They
should be approximately 3.8 mm (0.15 in) long in the
relaxed state. If they are shorter, the O-ring will not
seat properly. Replace if necessary.
9. Clean and inspect all parts and then reinstall the
O-ring and spring assemblies into the valve body.
10. Reinstall the retention caps and gently tighten the
cap.
11. Seal the output of the Product and pump up the unit
to at least 50 % of its rated pressure.
12. Release the pressure and repeat several times to
ensure that the O-rings seat properly.
To prevent possible electrical shock, fire, or
personal injury, do not use the Product if it
operates abnormally. Protection may be
impaired. When in doubt, have the Product
serviced.
If the Product does not turn on, check the battery charge.
Disconnect the battery charger. If the Product receives
power, the power button is illuminated. If the button is
illuminated, but the Product does not turn on, have the
Product serviced. See Contact Fluke.
Replace the battery when it no longer holds a charge for
the rated interval. The battery normally lasts for up to 300
charge/discharge cycles. To order a replacement battery,
see Contact Fluke and User-Replaceable Parts.
Note
Take spent batteries to a qualified recycler or
hazardous materials handler for disposal.
Contact an authorized Fluke Service Center for
recycling information.
To replace the battery, see Figure 24:
1. Push S and select Maintenance.
2. Push A(Exhaust) to release Product internal
pressure.
3. Turn the Product Off.
4. Remove test leads.
5. Make sure the Product is unplugged from its charger.
6. Turn the Product over.
7. Lift up the bail and remove the six screws with a
Phillips screwdriver.
2. Connect the USB cable (provided) to a PC (see
Figure 23).
4. Click on “Find your software”.
5.
Search for “729”.
6.
On the results page, select the Software Downloads tab.
7. Click on the necessary software link.
8. Read the instructions on this page.
9. Download the Firmware file.
10. Click on the Firmware exe file.
Calibration Data
The date of the last calibration and verification shows on
the sticker on the calibration certificate and on the
Instrument Information screen in screen in the Setup
menu. The CAL. STATUS number on the sticker should
always match the Calibration Status number in the
calibration screen. Only qualified personnel should
calibrate the Product.
Service Center Calibration or Repair
Only qualified service personnel must do calibration,
repairs, or service not included in this manual. If the
Product fails, examine the battery pack first, and replace
it if necessary.
Make sure to operate the Product in accordance with the
instructions in this manual. If the Product is faulty, send a
description of the failure with the Product. Pressure
modules do not need to accompany the Product unless
the module is faulty also. Be sure to pack the Product
securely, in the original shipping container if it is
available. See Contact Fluke and the Warranty
Statement.
The Product uses calibration management software to
communicate with documenting software. The Product
documents As Found, As Left, and single-reading results.
Download tasks by pre-defined tag numbers and save the
results to Product memory. Review or delete these results
when necessary.
Strap
Attach the Product carrying strap as shown in Figure 25.
Adjust the straps as necessary to hang the Product on
any sturdy support.
International ................................................... IEC 61326-1: Basic Electromagnetic Environment;
CISPR 11: Group 1, Class A
Group 1: Equipment has intentionally generated and/or uses conductively-
coupled radio frequency energy that is necessary for the internal function of
the equipment itself.
Class A: Equipment is suitable for use in all establishments other than
domestic and those directly connected to a low-voltage power supply network
that supplies buildings used for domestic purposes. There may be potential
difficulties in ensuring electromagnetic compatibility in other environments
due to conducted and radiated disturbances.
Caution: This equipment is not intended for use in residential environments
and may not provide adequate protection to radio reception in such
environments.
Korea (KCC) .................................................. Class A Equipment (Industrial Broadcasting & Communication Equipment)
Class A: Equipment meets requirements for industrial electromagnetic wave
equipment and the seller or user should take notice of it. This equipment is
intended for use in business environments and not to be used in homes.
USA (FCC) .................................................... 47 CFR 15 subpart B. This product is considered an exempt device per clause
15.103 (non FC versions only)
Radio Certification ......................................... (Contains) FCC ID: T68-FBLE, IC: 6627A-FBLE
Frequency ...................................................... 2402 MHz to 2480 MHz
Output Power ................................................. <100 mW