Fluke 725 Brochure

Page 1 of 2, Document #3023 725 Multifunciton Calibrator ©2000 Fluke Corporation Rev. C-7/2000
Fluke 725 Multifunction
Simultaneous Function Capability Channel A Channel B
24.000 mA DC M M or S
24.000 mA DC with 24V loop supply M
100.00 mV DC M or S
30.000V DC Measure M
Pt100, 200, 500, 1000 (385) M or S Pressure (requires Fluke 700PXX Modules) M M used as S Frequency; Squarewave, 1 CPM to10 kHz;
fixed amplitude 5V p-p M or S
M = Measure S = Source/Simulate
Calibrator
MULTIFUNCTION
725
PROCESS CALIBRATOR
V mA LOOP
V mA
MEAS
SOURCE
TC
RTD
STORE SETUP
RECALL
C / F
100%
Simply Powerful!
The new Fluke 725 Multifunction Process Calibrator is a powerful yet easy-to-use field calibrator. Use the measure and source functions to test and calibrate almost any process parameter.
Small, streamlined shape makes it easy
to carry Rugged, reliable design stands up to field use
Easy to read measure/source screen lets you
view input and output simultaneously Measure volts, mA, RTDs, thermocouples,
frequency, and ohms to test sensors and transmitters Source/simulate volts, mA, thermocouples, RTDs,
frequency, and ohms to calibrate transmitters Measure/source pressure using any of 28
25%
25%
0%
Fluke 700Pxx Pressure Modules Source mA with simultaneous pressure
measurement to conduct valve and I/P tests Support flow meter testing with frequency
and CPM functions Perform fast linearity tests with auto step and
auto ramp features Power transmitters during test using loop
supply with simultaneous mA measurement Store frequently-used test setups for later use
Backlight lets you work in poor light
Remote interface allows benchtop automated
operations Large battery capacity of four AA cells
Battery door for easy changes
Fluke. Keeping your world
up and running.
Ordering information
Fluke 725 Multifunction Process Calibrator
Each calibrator includes: TL75 Test Leads, AC70A Test Clips, one pair of stackable test leads, Users Manuals appropriate to country of destination (English, plus three of: French, German, Spanish, Italian, Dutch, Norwegian, Danish, Swedish, Finnish, Portuguese, Korean, Chinese, and Japanese), Statement of Quality Assurance Practices; CE and CSA markings.
Page 2 of 2, Document #3023 725 Multifunciton Calibrator ©2000 Fluke Corporation Rev. C-7/2000
Specifications
Summary specifications (18°C to 28°C for one year)
Function Measure or Source
Voltage 0 to 100 mV 0.01 mV .02% Rdg + 2 LSD Max load, 1 mA
mA 0 to 24 0.001 mA .02% Rdg + 2 LSD Max load, 1000
mV (TC terminals) -10.00 mV to +75.00 mV .01 mV .025% of range + 1 LSD Resistance 15 to 3200 0.01 to 0.1 0.10 to 1.0 Frequency 2.0 to 1000.0 CPM 0.1 CPM ±.05% For frequency source,
Loop Supply 24V dc N/A 10%
Temperature coefficient, -10°C to 18°C, 28°C to 55°C, ±.005% of range per °C.
Range Resolution Accuracy Notes
0 to 10V (source) 0.01V 0 to 30V (measure) 0.01V
1 to 1000 Hz 1 Hz ±.05% waveform is 5V p-p
1.0 to 10.0 kHz 0.1 kHz ±.25% squarewave, -0.1V offset
Thermocouple accuracy specifications
Thermocouple Measure or Source
J -200 to 0°C1.0°C
K -200 to 0°C1.2°C
T -200 to 0°C1.0°C
E -200 to 0°C0.9°C
R -20 to 0°C2.5°C
S -20 to 0°C2.5°C
B 600 to 800°C2.2°C
L -200 to 0°C0.85°C
U -200 to 0°C1.1°C
N -200 to 0°C1.5°C
Resolution
J, K, T, E, L, N, U 0.1°C, 0.1°F B, R, S 1°C, 1°F
Notes
Accuracy specifications include 0.2°C cold junction uncertainty.
0 to 1200°C0.7°C
0 to 1370°C0.8°C
0 to 400°C0.8°C
0 to 950°C0.7°C
0 to 500°C1.8°C
500 to 1750°C1.4°C
0 to 500°C1.8°C
500 to 1750°C1.5°C
800 to 1000°C1.8°C
1000 to 1800°C1.4°C
0 to 900°C0.7°C
0 to 400°C0.75°C
0 to 400°C0.9°C
RTD ranges and accuracy specifications
RTD Types, Ranges and Accuracies
Measure
(4 wire) Source
Ni 120 -80°C to 260°C0.2°C0.2°C
Pt 100 - 385 -200°C to 800°C0.33°C0.33°C Pt 100 - 3926 -200°C to 630°C0.3°C0.3°C Pt 100 - 3916 (JIS) -200°C to 630°C0.3°C0.3°C Pt 200 - 385 -200°C to 250°C0.2°C0.2°C
Pt 500 - 385 -200°C to 500°C0.3°C0.3°C
Pt 1000 - 385 -200°C to 100°C0.2°C0.2°C
Resolution
RTD 0.1°C, 0.1°F
250°C to 630°C0.8°C0.8°C
500 to 630°C0.4°C0.4°C
100 °C to 630°C0.3°C0.2°C
General specifications
Maximum voltage: 30V Storage temperature: -40°C to 71°C Operating temperature: 10°C to 55°C Relative humidity: 95% (10°C to 35°C); 75% (30°C to 40°C);
45% (40°C to 50°C); 35% (50°C to 55°C)
Shock: 30g, 11ms, half-sine shock (or 1meter drop test) Vibration: Random, 2g, 5-500 Hz Safety: CSA C22.2 No. 1010.1:1992 EMC: EN50082-1:1992 and EN55022:1994 Class B Size/weight: 96 x 200 x 47 mm (3.8 x 7.9 x 1.9 inches)
650g (23 oz)
Battery: Four AA alkaline batteries. Battery life: 25 hours typical Warranty: Three years