Fluke 721EX-1650, 721EX-3610, 721EX-1610, 721EX-3605, 721EX-1601 User manual

...
March 2015
© 2015 Fluke Corporation. All rights reserved. Specifications are subject to change without notice. All product names are trademarks of their respective companies.
721Ex
Pressure Calibrator

Users Manual

This Fluke product will be free from defects in material and workmanship for three years from the date of purchase. This warranty does not cover fuses, disposable batteries, or damage from accident, neglect, misuse, alteration, contamination, or abnormal conditions of operation or handling. Resellers are not authorized to extend any other warranty on Fluke’s behalf. To obtain service during the warranty period, contact your nearest Fluke authorized service center to obtain return authorization information, then send the product to that Service Center with a description of the problem.
THIS WARRANTY IS YOUR ONLY REMEDY. NO OTHER WARRANTIES, SUCH AS FITNESS FOR A PARTICULAR PURPOSE, ARE EXPRESSED OR IMPLIED. FLUKE IS NOT LIABLE FOR ANY SPECIAL, INDIRECT, INCIDENTAL OR CONSEQUENTIAL DAMAGES OR LOSSES, ARISING FROM ANY CAUSE OR THEORY. Since some states or countries do not allow the exclusion or limitation of an implied warranty or of incidental or consequential damages, this limitation of liability may not apply to you.
11/99

Table of Contents

Title Page
Introduction .................................................................................................................... 1
Contact Fluke ................................................................................................................. 1
Safety Information .......................................................................................................... 2
Ex-Hazardous Areas ................................................................................................. 4
Measuring Pressure .................................................................................................. 4
Symbols ......................................................................................................................... 5
Standard Equipment....................................................................................................... 6
Product Features ............................................................................................................ 6
Display ....................................................................................................................... 9
Language Selection ................................................................................................... 11
Home Menu Functionality .......................................................................................... 11
Backlight Use ............................................................................................................ 11
Zero Function ............................................................................................................ 11
Menus ........................................................................................................................ 12
SWITCHTEST ........................................................................................................... 12
i
721Ex
Users Manual
%ERROR .................................................................................................................. 15
MINMAX ................................................................................................................... 18
SET UNITS ............................................................................................................... 19
CONTRAST .............................................................................................................. 20
Lock and Unlock Configurations (CFG) .................................................................... 20
AUTO OFF ................................................................................................................ 21
RESOLUTION ........................................................................................................... 22
PROBE TYPE ........................................................................................................... 22
DAMP ....................................................................................................................... 24
Measure Pressure ......................................................................................................... 25
Media Compatibility ....................................................................................................... 25
Measurements ............................................................................................................... 26
Transmitter Calibration .................................................................................................. 27
mA Input Function ..................................................................................................... 27
Pressure-to-Current Transmitter Calibration ............................................................. 27
Ranges and Resolution ............................................................................................. 29
Maintenance .................................................................................................................. 30
Replace the Batteries ................................................................................................ 30
Clean the Product ..................................................................................................... 31
User-Replaceable Parts and Accessories ..................................................................... 32
Specifications ................................................................................................................ 34
Environmental ........................................................................................................... 34
Electrical and Temperature Measurement (1 year) ................................................... 34
Physical .................................................................................................................... 35
ii

List of Tables

Table Title Page
1. Symbols ................................................................................................................................. 5
2. Product Features ................................................................................................................... 8
3. Display Functions .................................................................................................................. 10
4. Ranges and Resolutions ....................................................................................................... 29
5. Approved Batteries ................................................................................................................ 30
6. User-Replaceable Parts and Accessories ............................................................................. 32
iii
721Ex
Users Manual
iv

List of Figures

Figure Title Page
1. Product Interface ................................................................................................................... 7
2. Display................................................................................................................................... 10
3. Pressure Switch Connection ................................................................................................. 13
4. Percent Error Function Connection ....................................................................................... 16
5. Temperature Measurement with RTD Probe ......................................................................... 24
6. Measure Current ................................................................................................................... 26
7. Pressure-to-Current Transmitter Connections ....................................................................... 28
8. Battery Replacement ............................................................................................................. 31
9. User-Replaceable Parts and Accessories ............................................................................. 33
v
721Ex
Users Manual
vi

Introduction

The 721Ex Pressure Calibrator (the Product) is a simple to use and versatile pressure calibrator. The two internal pressure sensors are configured with different pressure measurement ranges. One range for lower pressure (P1) and one range for higher pressure (P2). The Product features inputs for mA, switch contacts, and an RTD probe.
1
721Ex
Users Manual

Safety Information

A Warning identifies conditions and procedures that are dangerous to the user. A Caution identifies conditions and procedures that can cause damage to the Product or the equipment under test.
 Warning
To prevent possible electrical shock, fire, or personal injury:
Only assemble and operate high-pressure
systems if you know the correct safety procedures. High-pressure liquids and gases are hazardous and the energy from them can be released without warning.
Read all safety information before you use
the Product.
Carefully read all instructions.
Only operate the Product in non-hazardous
areas or classified areas where this device is certified to operate.
Use the correct terminals, function, and
range for measurements.
Do not apply more than the rated voltage,
between the terminals or between each terminal and earth ground.
Do not touch voltages > 30 V ac rms, 42 V
ac peak, or 60 V dc.
Remove all probes, test leads, and
accessories before the battery door is opened.
Do not exceed the Measurement Category
(CAT) rating of the lowest rated individual component of a Product, probe, or accessory.
Do not use and disable the Product if it is
damaged.
Remove the input signals before you clean
the Product.
Use only specified replacement parts.
Have an approved technician repair the
Product.
Use the Product only as specified, or the
protection supplied by the Product can be compromised.
Only change batteries in an area known to
be non-hazardous.
2
Pressure Calibrator
Safety Information
The battery door must be closed and
locked before you operate the Product.
Do not improperly apply pressure. Vacuum
should not be applied to any gauge pressure sensor. The Product display shows “OL” when an inappropriate pressure is applied. If “OL” is shown on any pressure display, the pressure should be reduced or vented immediately to prevent Product damage or possible personnel injury. “OL” is shown when the pressure exceeds 110 % of the nominal range of the sensor or when a vacuum in excess of 2 PSI is applied on gauge range sensors.
Push the ZERO button to zero the pressure
sensor when vented to atmospheric pressure.
Check entity parameters before making
any connections to this Product.
Remove the batteries if the Product is not
used for an extended period of time, or if stored in temperatures that exceed the battery manufacturer’s specifications. If the batteries are not removed, battery leakage can damage the Product.
Replace the batteries when the low battery
indicator shows to prevent incorrect measurements.
Be sure that the battery polarity is correct
to prevent battery leakage.
Repair the Product before use if the battery
leaks.
Use only specified replacement parts.
3
721Ex
Users Manual

Ex-Hazardous Areas

An Ex-hazardous area as used in this manual refers to an area made hazardous by the potential presence of flammable or explosive vapors. These areas are also referred to as hazardous locations.
The Product has been designed for use in Ex-Hazardous Areas. These are areas where potentially flammable or explosive vapors may occur. These areas are referred to as hazardous (classified) locations in the United States, as Hazardous Locations in Canada, as Potentially Explosive Atmospheres in Europe and as Explosive Gas Atmospheres by most of the rest of the world. The Product is designed as intrinsically safe. This means that connecting the Product to equipment that is used within intrinsically-safe circuits will not cause an ignition-capable arc as long as the entity parameters are suitably matched.
 Warning
To prevent possible personal injury, check entity parameters before making any connections to this device.

Measuring Pressure

Warning
For safe operation and maintenance of the Product, Refer to the table of ranges and resolutions at the back of the manual for information about overpressure and burst pressure ratings. Pressure sensors can be damaged and personnel injury can occur due to improper application of pressure. Vacuum should not be applied to any gauge pressure sensor. The Product display shows “OL” when an inappropriate pressure is applied. If “OL” is seen on any pressure display, reduce or vent pressure immediately to prevent equipment damage or possible personnel injury. “OL” is shown when pressure >120 % of the nominal range of the sensor or when a vacuum >2 PSI is applied on gauge range sensors.
4

Symbols

Symbols used on the Product or in this manual are shown in Table 1.
Table 1. Symbols
Symbol Description Symbol Description
Pressure Calibrator
Symbols
Risk of Danger. Important information. See Manual.
Hazardous voltage. Risk of electric shock.
Conforms to European Union directives.
Conforms to ATEX requirements
Conforms to relevant Australian EMC standards.
Battery
Conforms to relevant South Korean EMC Standards.
This product complies with the WEEE Directive (2002/96/EC) marking requirements. The affixed label indicates that you must not discard this electrical/electronic product in domestic household waste. Product Category: With reference to the equipment types in the WEEE Directive Annex I, this product is classed as category 9 "Monitoring and Control Instrumentation" product. Do not dispose of this product as unsorted municipal waste. Go to Fluke’s website for recycling information.
5
721Ex
Users Manual

Standard Equipment

Make sure that the Product shipment is complete. It must include:
The Product
Product Manuals CD
Getting Started Manual
Quick Reference Guide
Test leads
Carry case
Calibration certificate

Product Features

Figure 1 and Table 2 shows the location of the buttons, pressure controls, connection ports, and electrical inputs.
6
Pressure Calibrator
Product Features
2
F1 F2 F3
1
ZERO
6
5
8
7
3
11
10
4
9
Figure 1. Product Interface
hvf001.eps
7
721Ex
Users Manual
Table 2. Product Features
Item Description
Power button. Turns on and off the Product Function buttons. Used to configure the Product. These keys correspond to messages on the display. Zero button. Zeros pressure measurements. Backlight button. Push to turn the backlight on or off. COMMON input Input terminals to measure current and a contact closer for switch test. Low pressure port [P1] High pressure port [P2] RTD probe connector Firmware programming connector (for factory use only) Battery door
8
Pressure Calibrator
Product Features
Notes
When is pushed to turn on the Product, a short startup self-check routine is run. During that routine, the display shows the current firmware revision level, auto-shutdown status, and the range of the internal pressure sensors.
A maximum of 5 minutes warm-up is necessary for the product to reach rated accuracy. A longer warm-up period can be necessary for large changes in ambient temperature. See the “Zero Function Use” section for more about zeroing the pressure sensor displays. It is recommended that pressure ranges be zeroed each time the Product is started.

Display

The display has two main regions:
The menu bar (located at the bottom of the screen) is
used with the function buttons to access the Product menu.
The main display has a maximum of three process
measurement sub-regions.
These sub-regions will be referred to as the UPPER, MIDDLE and LOWER displays. Figure 2 shows the location of the different display fields. Table 3 describes them.
9
721Ex
Users Manual
Table 3. Display Functions
Figure 2. Display
hvf007.eps
Item
Number
Units
Primary Parameters
Span Indicator
Pressure Units
e Description
Nam
Shows what is being measured.
Shows the percent of the 4 mA to 20 mA span. (For mA functions only.)
Shows one of 17 pressure units available for display.
Shows the unit of measure for the display.
10
Pressure Calibrator
Product Features

Language Selection

The user interface is available in three languages:
English
Norwegian
German
To select a language:
1. Turn the Product off.
2. Hold down , , and  simultaneously.
3. As the Product powers up, the display shows the language in the top left corner of the display. Repeat the procedure to show each subsequent language. Once the necessary language is shown, the Product user interface stays in that language until another language is chosen.

Home Menu Functionality

There are three options for the Home Menu:
P1/P2
mA/RTD
MENU
These options are shown across the bottom of the display.
From anywhere within the menu structure, push to get back to the Home Menu.

Backlight Use

Push to toggle on and off the backlight.

Zero Function

In pressure mode and when the pressure is within the zero limit, The Product zeros pressure on any port that is currently shown on the display. The zero limits are within 10 % of the full-scale range of the selected sensor. If the display shows “OL,” the zero function will not operate.
11
721Ex
Users Manual

Menus

There are 11 sub menus that can be accessed from (Main Menu). Push to go to the next menu selection. For the last menu, push (DONE) to return to the Main Menu.
The 11 sub-main menus are:
SWITCHTEST
%ERROR
MINMAX
SET UNITS
CONTRAST
LOCK CFG
AUTO OFF
RESOLUTION
HART
PROBE TYPE
DAMP
Push , , or , depending on the menu, to toggle through each parameter of an active menu. The individual menus are explained in the subsequent sections.

SWITCHTEST

To access the SWITCHTEST menu, push . SWITCHTEST is shown on the menu bar.
Connect a pressure switch to the Product as shown in Figure 3.
For all connections in this manual:
 Warning
To prevent possible electrical shock, fire, or personal injury, check entity parameters before making any connections to this Product.
12
Pressure Calibrator
Product Features
F1 F2 F3
ZERO
Use low volume tubing
when possible
Figure 3. Pressure Switch Connection
Pressure
Switch
hvf017.eps
To do a switch test:
1. Use the P1 or P2 pressure input connections to connect the Product to the input of the switch. Connect the contact output of the switch to the COM and mA terminals of the Product. The polarity of the terminals does not matter.
2. Connect the pump to the Product and the pressure switch.
3. Make sure the vent on the pump is open.
4. Push  to select P1 or P2 depending on how the Product is connected.
5. Zero the Product if necessary.
6. Close the vent after the Product is zeroed.
7. If connected to a normally closed switch, the top of the display will read “CLOSE”.
8. Slowly apply pressure with the pump until the switch opens.
Note
In the SWITCHTEST mode the display update rate is increased to help capture changing pressure inputs. Even with this enhanced sample rate pressurizing, the test should be done slowly to ensure accurate readings.
13
721Ex
Users Manual
9. Once the switch is open, “OPEN” will be shown. Bleed the pump slowly until the pressure switch closes.
At the top of the display it will now read, “SW OPENED AT” and show the pressure at which the switch opened.
hix044.eps
hix043.eps
14
10. Push the “NEXT” option to view when the switch closed, and the dead band.
Pressure Calibrator
Product Features
hix046.eps
hix045.eps
11. Push the “NEW TEST” option to clear the data and do another test.
12. Push  to en
d the test and return to the main menu.

%ERROR

The Product features a unique function which can calculate pressure vs. milliamp error as a percentage of the 4 mA - 20 mA loop span. The %ERROR mode uses all three screens and has a unique menu structure. It simultaneously shows pressure, mA, and percent error. See Figure 4.
15
721Ex
Users Manual
-
F1 F2 F3
ZERO
External
Loop Supply
16
Figure 4. Percent Error Function Connection
hvf019.eps
Pressure Calibrator
Product Features
Example:
A pressure transmitter under test is 30 psi (2 Bar) Full­Scale and outputs a corresponding 4 mA to 20 mA signal. Program a 0 psi to 30 psi pressure span into the Product and the Product calculates and shows the deviation or %Error from the expected 4 mA to 20 mA output. This eliminates the need for manual calculations and helps when it is difficult to set an exact pressure.
To use the %ERROR function:
1. Push  to access the menus.
2. Push  to step through the menu until %ERROR is
shown on the menu bar.
3. Push  to open the %ERROR screen.
4. Push  to scroll through the port choices (P1 or P2).
5. Push  to configure the pressure range settings.
6. Use the arrow keys to set the 100 % point of the desired pressure range, select DONE SET when finished.
7. Use the arrow keys to set 0 % point and select DONE SET. The % ERROR mode is now ready to use.
Note
The 0 % and 100 % point will be saved in non­volatile memory until they are changed again by the user.
hmq054.eps
17
721Ex
Users Manual

MINMAX

The Product has a min/max feature to capture the minimum and maximum values of any displayed parameter. To use the MINMAX menu:
1. Push  to access the menus.
2. Push  to step through the menu until MINMAX is
shown on the menu bar.
3. Push  to toggle the display through the min and
max values that are stored in the min/max registers. These readings are live so that the new min/max values will be recorded while in this mode.
hix055.eps
hix056.eps
To reset the min/max registers, push for “CLEAR”. These registers are also cleared at power-up or when the configuration is changed. Push to exit MIN MAX and see live measurements.
18
Pressure Calibrator
Product Features

SET UNITS

Use the SET UNITS menu to select the measurement units for each port. To use this menu:
1. Push  to access the menus.
2. Push  to step through the menu until SET UNITS is shown on the menu bar.
3. Push  to select the necessary unit. The choices are:
inHg 0 °C
mmHg 0 °C
kg/cm
mmH
mmH
ftH
2
O 4 °C
2
O 20 °C
2
O 60 °F
2
psi
inH2O 4 °C
inH
inH
cmH
cmH
O 20 °C
2
O 60 °F
2
O 4 °C
2
O 20 °C
2
bar
mbar
MPa
kPa
19
721Ex
Users Manual
4. Push  to move through each function (P1, P2, or
RTD) and change to the desired units.
5. Push  when finished with units selection.

CONTRAST

Use the Contrast menu to adjust the display contrast.
1. Push  to access the menus.
2. Push  to step through the menu until CONTRAST
is shown on the menu bar.
3. Push  to access the Contrast menu.
hix024.eps
4. Push  and  repeatedly to adjust the display contrast to the necessary level. Push to finish the adjustment and go home as shown below.
hix025.eps

Lock and Unlock Configurations (CFG)

Use the LOCK CFG or UNLOCK CFG options of the Configuration Lock Menu (CONFIG), shown below, to lock or unlock the display configuration.
hix026.eps
When the LOCK CFG option is pushed, the menu display goes home and the menu configuration option on the Main Menu is locked. All menus are locked with the exception of:
MINMAX
CONTRAST
CONFIG
20
Pressure Calibrator
Product Features
You will also notice that some menu choices disappear when the LOCK CFG is used.
When the UNLOCK CFG option is selected, the configuration is unlocked and the menu display goes to the subsequent menu.

AUTO OFF

The Product can be set to automatically power off after a chosen number of minutes. This function can also be disabled. To set the auto off parameters:
1. Push  to access the menus.
2. Push  to step through the menu until AUTO OFF is
shown on the menu bar.
3. Push  on the Auto Off Main Menu shown below.
hix031.eps
4. Push  or  to select the number of minutes before
the Product turns off or scroll down to 0 to disable Auto Off as shown below.
hix032.eps
5. Push  to set the parameters and go to the main
menu. The auto off time is reset when a key is pushed.
21
721Ex
Users Manual

RESOLUTION

To choose between a low or high resolution display:
1. Push  to access the menus.
2. Push  to step through the menu until RESOLUTION is shown on the menu bar.
3. Push  to select the Resolution menu.
4. Push  or  to turn low resolution on or off.
5. Push  when finished.
hmq062.eps

PROBE TYPE

To select an external RTD probe for use with the Product:
1. Push  to access the menus.
2. Push  to step through the menu until PROBE is
shown on the menu bar.
hix035.eps
3. Push  on to select the Probe type. The probe
choices are:
P100-385
P100-392
P100-JIS
22
Pressure Calibrator
Product Features
4. Push  to select the necessary probe type (see the figure below). Push to store the change and go to the Main Menu.
Note
The default probe type is PT100-385.
SELECT RTD PROBE TYPE
hix036.eps
5. Connect the RTD Probe.
The standard probe has a 10-inch insertion depth with a ¼-inch diameter stainless steel sheath. See Figure 5.
Note
The factory default type is PT100-385 so if the Product is used with the Fluke 720 RTD Probe (pn 4366669), it is not necessary to set the probe type. Connect the probe to the Product and configure the display to read temperature.
The display shows “OL” when the measured temperature is outside the nominal measurement range of the RTD function (below
°
C or above 150 °C).
-40
23
721Ex
Users Manual
F1 F2 F3
ZERO
RTD Probe

DAMP

Turn on or off Damping with the DAMP menu selection. When damping is on, the Product shows a running average from ten measurements. The Product makes approximately three indications per second.
To use the Damping function:
1. Push  to access the menus.
2. Push  to step through the menu until DAMP is
shown on the menu bar.
3. Push  to select the DAMP menu.
4. Push  or  to turn on or off the DAMP function.
5. Push  when finished.
hvf016.eps
Figure 5. Temperature Measurement with RTD Probe
24
hmq064.eps
Pressure Calibrator
Measure Pressure

Measure Pressure

To measure pressure, connect the Product with the correct fitting and select a pressure port. The Product has two internal sensors. Make sure to choose the sensor based on working pressures and accuracy.
Warning
To prevent personal injury:
Pressure sensors can be damaged
and/or personnel injury can occur due to improper application of pressure. Refer to Table 4 for information on overpressure and burst pressure ratings. Vacuum should not be applied to any gauge pressure sensor. The Product display shows “OL” when an inappropriate pressure is applied. If “OL” is shown on any pressure display, the pressure should be reduced or vented immediately to prevent Product damage or possible personnel injury. “OL” is shown when the pressure exceeds 110 % of the nominal range of the sensor or when a vacuum in excess of 2 PSI is applied on gauge range sensors.
Push Z to zero the pressure sensor
when vented to atmospheric pressure.
Note
To ensure accuracy of the Product, the Product must be zeroed before a device is calibrated. See the “Zero Function Use” section.

Media Compatibility

The Product has a media-isolated sensor to prevent sensor contamination. Whenever possible, clean dry air is the media of choice. If this is not possible, make sure the media is compatible with the sensor. For the 16 PSIG and 36 PSIG range, use media compatible with Silicon, Pyrex, RTV, Gold, Ceramic, Nickel, and Aluminum. For all other ranges, use media compatible with 316 Stainless Steel.
25
721Ex
Users Manual

Measurements

Use the input terminals on the front of the Product to measure current. Current is measured in mA and percentage of range. The range on the Product is set at 0 % at 4 mA and 100 % at 20 mA.
Use the RTD connector and an RTD probe to measure temperature.
From the main menu, push to select mA or RTD. This function will only work on the LOWER screen.
Note
The display shows “OL” when the measured current is more than the nominal range of current measurement (24 mA).
F1 F2 F3
ZERO
4 to 20mA
Figure 6. Measure Current
Device
Under
Te st
hvf011.eps
26
Pressure Calibrator
Transmitter Calibration

Transmitter Calibration

mA Input Function

The mA input function reads back the 4 mA to 20 mA output from the device being calibrated. This can be done passively. The device under test directly generates 4 mA to 20 mA and can be read by the Product.
hvf047.eps

Pressure-to-Current Transmitter Calibration

To calibrate a pressure-to-current transmitter (P/I):
1. Connect the Product and the pump to the transmitter. See Figure 7.
2. Apply pressure with the pump.
3. Measure the current output of the transmitter.
4. Ensure the reading is correct. If not, adjust the transmitter as necessary.
Note
Use low-volume tubing when possible.
27
721Ex
F1 F2 F3
ZERO
Users Manual
External
Loop Supply
Figure 7. Pressure-to-Current Transmitter Connections
-
hvf018.eps
28
Pressure Calibrator
Transmitter Calibration

Ranges and Resolution

Ranges and resolutions for the Product are shown in Table 4.
Table 4. Ranges and Resolutions
Range (PSI) 16 36 100 300 500 1000 1500 3000 5000 Burst Pressure 60 120 400 1200 2000 4000 6000 9000 10000 Proof Pressure
(PSI) Engineering Unit Factor psi 1 16 36 100 300 500 1000 1500 3000 5000 bar 0.06894757 1.1032 2.4821 6.8947 20.684 34.474 68.947 103.42 206.84 344.74 mbar 68.94757 1103.2 2482.1 6894.8 20684 34474 68948 N/A N/A N/A kPa 6.894757 110.32 248.21 689.48 2068.4 3447.4 6894.8 10342 20684 34474 MPa 0.00689476 0.1103 0.2482 0.6894 2.0684 3.4474 6.8948 10.342 20.684 34.474 kg/cm2 0.07030697 1.1249 2.5311 7.0307 21.092 35.153 70.307 105.46 210.92 351.53 cmH20 @ 4 °C cmH20 @ 20 °C mmH20 @ 4 °C mmH20 @ 20 °C inH20 @ 4 °C inH20 @ 20 °C inH20 @ 60 °F mmHg @ 0 °C inHg @ 0 °C
Proof pressure - maximum allowable pressure without a shift in calibration.
Burst pressure - sensor damaged or destroyed; some risk of personnel injury.
70.3089 1124.9 2531.1 7030.9 21093 35154 70309 N/A N/A N/A
70.4336 1126.9 2535.6 7043.4 21130 35217 70434 N/A N/A N/A
703.089 11249 25311 70309 N/A N/A N/A N/A N/A N/A
704.336 11269 25356 70434 N/A N/A N/A N/A N/A N/A
27.68067 442.89 996.50 2768.1 8304.2 13840 27681 41521 83042 N/A
27.72977 443.68 998.27 2773.0 8318.9 13865 27730 41595 83189 N/A
27.70759 443.32 997.47 2770.8 8312.3 13854 27708 41561 83123 N/A
51.71508 827.44 1861.7 5171.5 15515 25858 51715 77573 N/A N/A
2.03602 32.576 73.297 203.60 610.81 1018.0 2036.0 3054.0 6108.1 10180
35 70 200 600 1000 2000 3000 6000 7000
29
721Ex
Users Manual

Maintenance

Replace the Batteries

If the batteries discharge too far, the Product automatically shuts down to prevent battery leakage.
 Warning
To prevent possible electrical shock, fire, or personal injury:
Remove the batteries if the Product is
not used for an extended period of time, or if stored in temperatures above 50 °C. If the batteries are not removed, battery leakage can damage the Product.
Replace the batteries when the low
battery indicator shows to prevent incorrect measurements.
Be sure that the battery polarity is
correct to prevent battery leakage.
Repair the Product before use if the
battery leaks.
The battery door must be closed and
locked before you operate the Product.
Only change batteries in an area known
to be non-hazardous.
To change the batteries, see Figure 8:
1. Turn off the Product.
2. Turn the Product so that the display is down.
3. With a flat-head screwdriver, remove the battery door screw.
4. Replace the four AA batteries with new batteries. Make sure that the polarity on the batteries is correct. See Table 5 for a list of approved batteries.
5. Replace the battery door.
6. Tighten the battery door screw.
Table 5. Approved Batteries
Battery Manufacturer
(All Batteries Alkaline 1.5 V)
Duracell MN1500
Rayovac Max Plus 815
Eveready (Energizer) E91
Panasonic LR6XWA
Type
30
Pressure Calibrator
Maintenance

Clean the Product

Caution
To avoid damaged to the Product:
Do not use solvents or abrasive
cleansers.
Do not allow water into the case.
Clean the Product with a soft cloth dampened with water or water and mild soap.
Figure 8. Battery Replacement
hvf061.eps
31
721Ex
Users Manual

User-Replaceable Parts and Accessories

 Warning
To prevent possible electrical shock, fire, or personal injury, use only specified replacement parts.
User-replaceable parts are listed in Table 6 and shown in Figure 9 . For more information about these items, contact a Fluke representative. See the “Contact Fluke” section of this manual.
Table 6. User-Replaceable Parts and Accessories
Item Description
Red Holster
TL7X Probe Cap, Red 3986579
TL7X Probe Cap, Black 3986568
TPAK80-4-2002, Magnet Strap
TPAK80-4-8001, Strap 9 inches
Part
Number
4497306
669952
669960
Item Description
Not Shown
Not Shown
Not Shown
Not Shown
Not Shown
AA Alkaline Batteries 376756
Safety Sheet 4561164
Quick Reference Guide 4561158
Users Manual CD 4561173
Rubber Feet 4364579 Fluke-720RTD Probe for 721
and 719Pro Test Lead Set Variable
Alligator Clip, Red Variable
Alligator Clip, Black Variable
Part
Number
4366669
[1] See www.fluke.com for more information about the
test leads and alligator clips available for your region.
[1]
[1]
[1]
32
Pressure Calibrator
User-Replaceable Parts and Accessories
1
3
2
6
4
5
7
8
hvf065.eps
Figure 9. User-Replaceable Parts and Accessories
33
721Ex
Users Manual

Specifications

(15 °C to 35 °C unless otherwise noted)

Environmental

Operating Temperature ............................ -10 °C to +45 °C (14 °F to +113 °F)
Storage
With Batteries ........................................ Per battery manufacturer’s specification, not to exceed storage specification without batteries.
Without Batteries ................................... -20 °C to +60 °C (-4 °F to +140 °F)
Altitude ...................................................... 2000 m
Power Requirements ............................... 6 V dc
Batteries .................................................... 4 AA batteries (alkaline)
Battery Life ................................................ >35 hours, typical usage

Electrical and Temperature Measurement (1 year)

Function Range Resolution Accuracy
mA Measure 0 to 24 mA 0.001 mA
*Temperature measure (RTD/Ohms)
-40 °C to 150 °C (-40 °F to 302 °F) 0.01 °C, 0.01 °F
*Temperature measurement requires optional 720RTD Pt-100 RTD probe available as an accessory.
±0.015 % of rdg ±0.002 mA ±0.015 % of rdg ±0.02 Ω; ±0.1 °C (±0.2 °F)
±0.25 °C (± 0.45 °F) combined uncertainty when us
720 RT
D probe accessory
ing
34

Physical

Dimensions (with holster) ......................... (H x W x D) (20 x 11 x 5.8) cm, (7.9 x 4.3 x 2.3) in
Weight (with holster) ................................ 0.539 kg (1 lb 3 oz)
Ingress Protection ..................................... IEC 60529 - IP40
Connectors/Ports
Pressure .............................................. Two, 1/8 in NPT
RTD ..................................................... RTD probe
Temperature Effect (all functions) ............. No effect on accuracy on all functions from 15 °C to 35 °C
Add ±0.002
% Full Scale/°C for temps outside of 15 °C to 35 °C
Pressure Calibrator
Specifications
35
721Ex
Users Manual
1-year Specifications Low Pressure Sensor High Pressure Sensor
Model
721Ex­1601
721Ex­1603
721Ex­1605
721Ex­1610
721Ex­1615
721Ex­1630
721Ex­1650
Calibrator
Description
16 PSIG, 100 PSIG
16 PSIG, 300 PSIG
16 PSIG, 500 PSIG
16 PSIG, 1000 PSIG
16 PSIG, 1500 PSIG
16 PSIG, 3000 PSIG
16 PSIG, 5000 PSIG
Range
Sensor 1
-14 psi + 16 psi
-0.97 bar to
1.1 bar
Resolution
Sensor 1
0.001 psi,
0.0001 bar
Accuracy
Sensor 1
0.025 % of full scale
Range
Sensor 2
-12 psi to +100 psi
-0.83 bar to
6.9 bar
-12 psi to +300 psi
-0.83 bar to 20 bar
-12 psi to +500 psi
-0.83 bar to
34.5 bar
0 psi to +1000 psi
0.00 bar to 69 bar
0 psi to +1500 psi
0.00 bar to
103.4 bar
0 psi to +3000 psi
0.00 bar to 200 bar
0 psi to +5000 psi
0.00 bar to 345 bar
Resolution
Sensor 2
0.01 psi
0.0001 bar
0.01 psi
0.001 bar
0.01 psi
0.001 bar
0.1 psi
0.001 bar
0.1 psi
0.001 bar
0.1 psi
0.01 bar
0.1 psi
0.01 bar
Accuracy
Sensor 2
0.025 % of full scale
0.035 % of full scale
36
Pressure Calibrator
1-year Specifications Low Pressure Sensor High Pressure Sensor
Specifications
Model
721Ex­3601
721Ex­3603
721Ex­3605
721Ex­3610
721Ex­3615
721Ex­3630
721Ex­3650
Calibrator
Description
36 PSIG, 100 PSIG
36 PSIG, 300 PSIG
36 PSIG, 500 PSIG
36 PSIG, 1000 PSIG
36 PSIG, 1500 PSIG
36 PSIG, 3000 PSIG
36 PSIG, 5000 PSIG
Range
Sensor 1
-14 psi + 36 psi
-0.97 bar to
2.48 bar
Resolution
Sensor 1
0.001 psi,
0.0001 bar
Accuracy
Sensor 1
0.025 % of full scale
Range
Sensor 2
-12 psi to +100 psi
-0.83 bar to
6.9 bar
-12 psi to +300 psi
-0.83 bar to 20 bar
-12 psi to +500 psi
-0.83 bar to
34.5 bar
0 psi to +1000 psi
0.00 bar to 69 bar
0 psi to +1500 psi
0.00 bar to
103.4 bar
0 psi to +3000 psi
0.00 bar to 200 bar
0 psi to +5000 psi
0.00 bar to 345 bar
Resolution
Sensor 2
0.01 psi
0.0001 bar
0.01 psi
0.001 bar
0.01 psi
0.001 bar
0.1 psi
0.001 bar
0.1 psi
0.01 bar
0.1 psi
0.01 bar
0.1 psi
0.01 bar
Accuracy
Sensor 2
0.025 % of full scale
0.035 % of full scale
37
721Ex
Users Manual
Electromagnetic Compatibility (EMC) ........... IEC 61326-1 (Portable); IEC 61326-2-2, CISPR 11, Group 1, Class A
Group 1 equipment: group 1 has intentionally generated and/or use conductively coupled radio­frequency energy which is necessary for the internal functioning of the equipment itself.
Class A equipment is equipment suitable for use in all establishments other than domestic and those directly connected to a low voltage power supply network which supplies buildings used for domestic purposes. Caution - There may be potential difficulties in ensuring electromagnetic compatibility in other environments, due to conducted and radiated disturbances.
USA (FCC) – 47 CFR 15 subpart B, this product is considered an exempt device per clause
15.103
Applies to use in Korea only. Class A Equipment (Industrial Broadcasting & Communication Equipment)
[1] This product and the seller or environments and is not to be used in homes.
Ex-Haza
Entity Parameters ......................................... MEASUREMENT JACKS:
rdous Area Markings .......................
Ex ia IIB T3 Gb (Ta= -10… +45 °C)
KEMA 10 ATEX 0168X Ex ia IIB T3 Gb (Ta= -10…+45 °C)
II 2 G IECEX CSA 10.0013X Manufactured by Martel Electronics, Inc., 3 Corporate Park Dr. Derry, NH, USA
Ui = 30 V; Li = 80 mA; Pi = 750 mW; Ci = 0 μF; Li = 0 mH Uo = 7, 14 V; lo = 1,12 mA; Po = 2 mW: Co = 240 μF; Lo = 1 H Circular Connector: FOR USE WITH LTP100A RTD PROBE ONLY
[1]
meets requirements for industrial (Class A) electromagnetic wave e user should take notice of it. This equipment is intended for use in bu
quipment
siness
38
Loading...