Intrinsically Safe
Fluke 718Ex 30G,
100G, and 300G
Pressure Calibrators
Pump Up the Pressure
The new Fluke 718Ex Intrinsically Safe Pressure Calibrator
is a powerful new intrinsically safe, self contained pressure
calibration solution that offers:
Wide range of selectable
ATEX II 1G Ex ia IIC
•
T4 compliance
I.S. Class I, Division
•
1 compliance
Built-in pressure/vacuum
•
hand pump, with vernier
and bleed valve
Pressure measurement to
•
0.05 % of full span, using an
internal pressure sensor
Pressure measurement to
•
3,000 psi/200 bar using
any of the 8 intrinsically
safe Fluke 700PEx Pressure
Modules
Pressure specifications (internal sensor)
18 °C to 28 °C, one year
•
measurement units for
pressure
Current measurement
•
with 0.02 % accuracy and
0.001 mA resolution
Pressure switch test function
•
Min/Max hold functions
•
Compact size and weight
•
Simple, push-button user
•
interface
Rugged and reliable, for
•
field use
Pressure switch test function
•
Technical Data
Pressure Input
Range Resolution Accuracy
-12 psi to 30 psi
(-82.7 kPa to 207 kPa)
-12 psi to 100 psi
(-82.7 kPa to 690 kPa)
-12 psi to 300 psi
(-82.7 kPa to 2068 kPa)
Temp Coeff. -10 °C to
18 °C, 28 °C to 55 °C
Engineering Units (in Hg)
PSI, in. H2O (4 °C), in. H2O (20 °C), kPa, cm H2O (4 °C),
cm H2O (20 °C), BAR, mBAR, kg/cm2, mmHg, in Hg
Media
Gasses (non corrosive)
Overpressure
3 x F.S. on 30G
2 x F.S. on 100G
375 PSI on 300G
.001 psi (0.01 kPa) .05 % FS
.01 psi (0.01 kPa) .05 % FS
.01 psi (0.01 kPa) .05 % FS
+ .01 % of range per °C
Ordering information
Model
Fluke-718Ex 30G Pressure Calibrator
Fluke-718Ex 100G Pressure Calibrator
Fluke-718Ex 300G Pressure Calibrator
Included accessories
Protective red holster, TL75 test leads, AC72 test clips, CD users
manuals (English, French, German, Spanish, Italian, Dutch,
Norwegian, Danish, Swedish, Finnish, Portuguese, Korean,
Chinese, Japanese), Fluke 718Ex CCD control drawing
I.S. Class I Div 1 Groups A-D T4
AEx ia IIC T4
II 1 G Ex ia IIC T4
KEMA 04 ATEX 1061
Ta = -10 °C to +55 °C
Specifications continued
Pressure display (external modules)
Pressure Module Readout
Pressure range Determined by pressure module
Resolution 5 digits
Accuracy Determined by pressure module
Engineering Units (in Hg)
Operating modes
Mode
Measure pressure (using internal sensor)
Display pressure module reading (automatically selected when pressure module
connected)
Measure current
PSI, in. H2O (4 °C), in. H2O (20 °C), Kpa, cm H2O (4 °C),
cm H2O (20 °C), BAR, mBAR, Kg/cm2, mmHg, in Hg
Current specifications
18 °C to 28 °C, one year
Current Measurement
Input/output
Pressure input 1/8 in. NPT pressure fitting
Pressure module input LEMO connector
Current input Shrouded banana jacks
Range 0 mA to 24 mA
Resolution .001 mA
mA ± 5 °C .02 % Rdg ± 2 count
Temp Coeff.
to 18 °C, ± .005 % of range per °C
to 55 °C
Battery
Power One 9-volt alkaline battery
Battery life 4 to 20 hours, depending on functions used
Zeroing modes
Mode
Gage and Pushing Zero button stores present pressure value as an offset and subtracts it
differential from the displayed value
Absolute Pushing Zero button causes nominal barometric pressure to be displayed. User corrects
to actual barometric pressure with up/down keys. Difference is used as zero offset calibration.
Pressure module specification
(All specifications in % of full span. Specifications reflect a confidence interval of 95 %.)
Range Reference High
Range/ (approx)/ uncertainty Stability Temperature Total
Model Resolution Resolution (23 ± 3 ºC) (1 year) (0 to 50 ºC) uncertainty media media material (x nominal)
Differential
FLUKE-700P01Ex 10 in. H
FLUKE-700P24Ex 15 psi/0.001 103 kPa/0.01 0.025 0.010 0.015 0.050 316 SS Dry 316 SS 3x
Gage
FLUKE-700P05Ex 30 psi/0.001 207 kPa/0.01 0.025 0.010 0.015 0.050 316 SS N/A 316 SS 3x
FLUKE-700P06Ex 100 psi/0.01 690 kPa/0.07 0.025 0.010 0.015 0.050 316 SS N/A 316 SS 3x
FLUKE-700P27Ex 300 psi / 0.01 2070 kPa / 0.1 0.025 0.010 0.015 0.050 316 SS N/A 316 SS 3x
FLUKE-700P09Ex 1500 psi/0.1 10 MPa/0.001 0.025 0.010 0.015 0.050 316 SS N/A 316 SS 2x
Absolute
FLUKE-700PA4Ex 15 psi/0.001 103 kPa/0.01 0.050 0.010 0.010 0.070 316 SS N/A 316 SS 3x
High
FLUKE-700P29Ex 3000 psi/0.1 20.7 M Pa/0.001 0.050 0.010 0.020 0.080 C276 N/A C276 2x
1
Total uncertainty, one year for temperature range 0 °C to +50 °C. Total uncertainty, 1.0 % of full span for temperature range -10 °C to 0 °C.
2
“Dry” indicates dry air or non-corrosive gas as compatible media. “316 SS” indicates media compatible with Type 316 Stainless Steel. “C276”
O/0.01 2.5 kPa/0.002 0.200 0.050 0.050 0.300 Dry Dry 316 SS 3x
2
indicates media compatible with Hastelloy C276.
2
1
side side Fitting pressure
Low2 Max over-
Fluke. Keeping your world
up and running.
®
Environmental and safety requirements
Operating temperature -10 °C to 55 °C
Non-operating temperature -40 °C to 71 °C
Relative humidity 95 % (10 °C to 30 °C)
(%RH operating 75 % (30 °C to 40 °C)
without condensation) 45 % (40 °C to 50 °C)
35 % (50 °C to 55 °C)
Size (LxWxD) 216 mm x 94 mm x 66 mm (8.50 in x 3.72 in x 2.60 in)
Weight 35 oz (992 g)
Vibration Random, 2g, 5-500 Hz
Shock 1 Meter Drop test
EMC EN61326 2002-02, Criteria C
Fluke Corporation
PO Box 9090, Everett, WA 98206 U.S.A.
Fluke Europe B.V.
PO Box 1186, 5602 BD
Eindhoven, The Netherlands
For more information call:
In the U.S.A. (800) 443-5853 or
Fax (425) 446-5116
In Europe/M-East/Africa +31 (0) 40 2675 200 or
Fax +31 (0) 40 2675 222
In Canada (800)-36-FLUKE or
Fax (905) 890-6866
From other countries +1 (425) 446-5500 or
Fax +1 (425) 446-5116
Web access: http://www.fluke.com
©2007-2010 Fluke Corporation.
Specifications subject to change without notice.
Printed in U.S.A. 7/2010 2116965F D-EN-N
Modification of this document is not permitted