Fluke 5500A User Manual

®
Introduction
Instruction Sheet
Model 5500A/COIL
50-Turn Current Coil
UNFUSED. Exceeding current and voltage ratings can cause burn or fire hazard.
Using the Coil
Clamp-type current meters operate as current transformers, with differing degrees of magnetic coupling between primary and secondary that vary from meter to meter. The position of the clamp meter with re­spect to the cable also affects the magnetic coupling between primary and secondary of the current trans­former, which causes variation in reading of the current meter. This is important to under­stand in order to make the most accurate and repeatable measurements. The base of the 5500A/COIL was designed so the current clamp can be centered carefully on the coil, minimizing operator error for best repeatabil­ity. Calibration accuracy to specifications is guaranteed only when proper clamp alignment is made. The clamp current meter should be centered as much as possible on the base dur­ing calibration and verification. If the clamp­type current meter has alignment marks, the alignment marks should align the clamp with the center bundled wire of the 50-Turn 5500A/COIL. (See Figure 1).
PWARNING
Figure 1. Positioning the Clamp
PN 600616 August 1995 Rev.8, 1/09
©1995-2009 Fluke Corporation. All rights reserved. Printed in U.S.A. Product names are trademarks of their respective companies.
Specifications
Number of Turns 50
Maximum Current 11A rms, continuous
20.5A rms, 2 minutes
Maximum Duty Cycle Derating < 11A, continuous
> 11A, 2 minutes ON, 8 minutes OFF
Maximum Voltage 7 V dc or 7 V rms ac
Minimum Inner Diameter of Clamp Jaws 2.54 cm [1”]
Specification for Clamp Meter/Coil interaction
Specification for 5500A/COIL
Coil Specifications [clamp/coil interaction]
Calibrator Output
Magnitude Frequency Amp-turns ±(% of output + Amps) ±(% of output + Amps)
0.2 - 0.33 A DC 10 - 16.4999 0.50 0.02
0.33 - 2.9999 A DC 16.5 - 149.999 0.5 0.14
3.0 - 20.5 A DC 150 - 1025
0.2 - 0.33 A 45 Hz to 65 Hz 10 - 16.4999 0.28 0.003 0.56 0.03
0.33 - 2.9999 A 45 Hz to 65 Hz 16.5 - 149.999 0.28 0.025 0.56 0.25
3.0 - 20.5 A 45 Hz to 65 Hz 150 - 1025 0.28 0.09 0.56 0.9
0.2 - 0.33 A 65 Hz to 440 Hz 10 - 16.4999 0.79 0.003 1.00 0.03
0.33 - 2.9999 A 65 Hz to 440 Hz 16.5 - 149.999 0.79 0.027 1.00 0.25
3.0 - 20.5
0.2 - 20.5 A Above 440 Hz 10 - 1025 Usable but unspecified
**Note: Some calibrators may not have adequate compliance voltage to drive coil in this range.
**
A 65 Hz to 440 Hz 150 - 1025
Effective
Current Output
Toroidal-type Clamps
[such as the Fluke 80i-600 and 80I-1000]
Not applicable
0.79 0.1 1.00 0.9
'other-type' of Clamps
[such as the Fluke 80i-KW, -400, 80i-410, -500, -1010 and 30-series]
0.5 0.5
Calculating Total Specification
The total specification of the effective current that the clamp-type meter measures is a function of the clamp/coil interaction and of the current calibrator. To determine the total specification use the following formula:
total
2
coil
2
USS +=
source
Example:
Assume we are driving the coil with the Fluke 5500A at 4A, 60 Hz (the clamp meter will see an effective current of 200A, 60 Hz) and we are calibrating a toroidal-type clamp meter. The Calibrator’s 1 year specification at 4 Amps is ± (0.06% + 2 mA), so the effective current in the coil bundle will have a speci­fication of ± (0.06% + 0.1A). Next, we find the total specification of the calibrator and the coil as a per­centage of the output:
Specification of effective calibrator current in coil bundle = ± (0.06% + 0.1A) = 0.11% Specification due to clamp Meter/Coil Interaction = ± (0.28% + 0.09A) = 0.325%
The RSS of these two specifications determines the total specification of the clamp/source combination:
total
22
=+=S
%343.0%11.0%325.0
Loading...