Manual Title: 5500A Operator Supplement Issue: 1
Print Date: December 1994 Issue Date: 1/07
Revision/Date: 11, 7/06 Page Count: 1
This supplement contains information necessary to ensure the accuracy of the
above manual. This manual is distributed as an electronic manual on the
following CD-ROM:
Manual Supplement
CD Title: 5500A/5520A
CD Rev. & Date: 2, 6/
CD PN: 1627768
2006
© 2007 Fluke Corporation. All rights reserved. Printed in U.S.A. x

5500A Operator Manual Supplement
Change #1, 39296
On page 1-21, under Temperature Calibration (RTD) Specifications, under RTD Type
change:
From: Pt 395, 100 Ω
To: Pt 385, 100 Ω
Change #2
On page 4-7, under Zeroing the Calibrator, add the following note between the first and
second paragraph:
Note
If the 5500A is operated outside the range of t
General Specifications, Chapter 1 of this manual, must be calculated and added to the absolute
uncertainties. Zeroing the 5500A is still required.
± 5 ºC, then the temperature coefficient defined in the
cal
1/07 1