This file contains remote control and programming information
for the above-mentioned models with use of the PM9080
Optically Isolated RS232 Adapter/Cable.
It consists of the following chapters:
1. INSTALLING THE PM9080
2. INTRODUCTION TO PROGRAMMING
3. COMMAND REFERENCE
APPENDIXES
APPENDIX A ACKNOWLEDGE DATA
APPENDIX B STATUS DATA
APPENDIX C WAVEFORM DATA
APPENDIX D ASCII CODES
- Connect the PM9080 to the RS232 port of the computer.
If necessary, use a 9-pin to 25-pin adapter and
25-pin gender changer.
- Hook the PM9080 cable to the Power Quality Analyzer.
- Turn on the computer and the Power Quality Analyzer.
- Make sure that the communication settings match for the
RS232 port of the computer and the Power Quality Analyzer.
After power-on, the default settings of the Power Quality
Analyzer are as follows:
1200 baud, No parity, 8 data bits, 1 stop bit
You can modify the baud rate with the PC (Program
Communication) command. See chapter 3 COMMAND REFERENCE.
Other settings are fixed.
You can modify the computer RS232 port settings to match
the above settings with the following DOS command:
MODE COM1:1200,N,8,1
This command assumes that COM1 is the RS232 port used on
the computer. Replace COM1 in the above command with COM2,
COM3, or COM4 if one of these ports is used. You can place
this command in the computer startup file AUTOEXEC.BAT so
that the default settings for the computer are the same as
for the Power Quality Analyzer. If you want to use a higher
data transfer speed (baud rate), let your QBASIC program
change the settings for both the computer and the Power
Quality Analyzer. See the example under the PC (Program
Communication) command in chapter 3 COMMAND REFERENCE.
When you have installed the PM9080 as described in the
previous chapter, you can control the Power Quality Analyzer
from the computer with simple communication facilities, such
as GWBASIC, QuickBASIC and QBASIC (programming languages from
Microsoft Corporation).
All examples given in this manual are in the QBASIC language
but will also run in QuickBASIC. QuickBASIC allows you to
make executable files from programs so you can start such
programs directly from DOS.
It is assumed that you have knowledge of these programming
languages. QBASIC is supplied with Microsoft MS-DOS 5.0 and
higher and Windows 95, 98, and NT, including ’on-line’ Help.
Features of the syntax and protocol for the Power Quality
Analyzer are as follows:
- Easy input format with a ’forgiving’ syntax:
All commands consist of two characters that can be
UPPER or lower case.
Parameters that sometimes follow the command may be
separated from it by one or more separation characters.
- Strict and consistent output format:
Alpha character responses are always in UPPERCASE.
Parameters are always separated by a comma
("," = ASCII 44, see Appendix D).
Responses always end with the carriage return code
(ASCII 13). Because the carriage return code is a
non-visible character (not visible on the screen or on
paper), this character is represented as <cr> in the
command syntax.
- Synchronization between input and output:
After receipt of every command, a Power Quality Analyzer
returns an acknowledge character (digit) followed by the
carriage return code (ASCII 13). This indicates that the
command has been successfully received and executed.
The computer program must always read this acknowledge
response before sending the next command to the
Power Quality Analyzer.
Page 2.2
** Commands sent to the Power Quality Analyzer **
All commands for the Power Quality Analyzer consist of a
header made up of two alpha characters sometimes followed
by parameters. Example:
RI This is the Reset Instrument command. It
resets the Power Quality Analyzer.
Some of the commands are followed by one or more parameters
to give the Power Quality Analyzer more information.
Example:
SS 8 This is the Save Setup command. It saves the
present acquisition settings in memory. The SS
header is followed by a separator (space),
then followed by the parameter "8" to
indicate where to store the settings. The
meaning of this parameter is described in
Chapter 3 COMMAND REFERENCE.
Some commands require several parameters.
Example:
WT 9,50,30 This is the Write Time command.
This command requires three parameters. The
parameters are separated by a comma, which is
called the Program Data Separator. You may
use only one comma between the parameters.
Also refer to the section ’Data Separators’.
A code at the end of each command tells the Power Quality
Analyzer that the command is ended. This is the carriage
return code (ASCII 13) and is called the Program Message
Terminator. This code is needed to indicate to the Power
Quality Analyzer that the command is completed so it can
start executing the command. Also refer to the section
’Command and Response Terminators’.
Page 2.3
** Responses received from the Power Quality Analyzer **
After each command sent to the Power Quality Analyzer there
is an automatic response from it, indicated as <acknowledge>
(which you MUST input), to let the computer know whether or
not the received command has been successfully executed.
Refer to the ’Acknowledge’ section below.
There are several commands that ask the Power Quality
Analyzer for response data. Such commands are called Queries.
Example:
ID This is the IDentification query, which asks for
the model number and the software version of the
Power Quality Analyzer.
When the Power Quality Analyzer has received a query, it
sends the <acknowledge> reply as it does after any command,
but now it is followed by the queried response data.
The format of the response data depends upon which query is
sent. When a response consists of different response data
portions, these are separated with commas (ASCII code 44).
Also refer to the section ’Data Separators’.
All response data, <acknowledge> as well as following
(queried) response data are terminated with the carriage
return code (<cr> = ASCII 13). Also refer to the section
’Command and Response Terminators’.
Page 2.4
** Acknowledge **
After receiving of a command, the Power Quality Analyzer
automatically returns the <acknowledge> response to let the
computer know whether or not the received command has been
successfully executed.
This response is a one-digit number followed by <cr> as
response terminator. If <acknowledge> is 0, it indicates
that the Power Quality Analyzer has successfully executed the
command. If the command was a query, the <acknowledge><cr>
response is immediately followed by the queried response data
terminated with <cr>.
If <acknowledge> is 1 or higher, it indicates that the
Power Quality Analyzer has not executed the command
successfully. In that case, if the command was a query, the
<acknowledge><cr> response is NOT followed by any further
response data.
There can be several reasons for a non-zero <acknowledge>
response. For more information see Appendix A.
In case of an error you can obtain more detailed status
information by using the ST (STATUS) query.
Note: YOU MUST ALWAYS INPUT <acknowledge>, EVEN WHEN
THE COMMAND WAS NOT A QUERY.
Page 2.5
** Data Separators **
Data Separators are used between parameters sent to the
instrument and between values and strings received
from the instrument. Comma (",") is used as program
data separator as well as response data separator:
- Program Data Separator
Name Character ASCII Value Comments
Decimal
--------------------------------------------------------- comma , 44 Single comma allowed
** Command and Response Terminators **
(Message Terminators)
- Command (Program Message) Terminators
A code is needed at the end of each command to tell the
instrument that the command is ended, and that it
can start executing the command. This code is called the
Program Message Terminator. The code needed for the
instrument is carriage return (ASCII code 13 decimal).
Notes:
1. The carriage return code is a non-visible ASCII
character. Therefore this code is represented as <cr>
in the Command Syntax and Response Syntax lines given
for each command.
2. The QBASIC programming language, which is used for
all program examples, automatically adds a carriage
return to the end of the command output. (In the QBASIC
language, this is the PRINT #.... statement.)
After <cr> is recognized by the instrument, the
entered command is executed. After EACH command the
instrument returns <acknowledge><cr> to the
computer to signal the end of the command processing (also
see the section ’Acknowledge’.)
- Response (Message) Terminators
The response from the instrument ends with a
carriage return (ASCII 13). This is indicated as <cr> in
the Response Syntax for each command.
Page 2.7
** Typical program sequence **
An example
A typical program sequence consists of the following user
actions:
1. Set the communication parameters for the RS232 port of
the computer to match the instrument settings.
2. Output a command or query to the instrument.
3. Input the acknowledge response from the instrument.
If the response value is zero, go to step 4.
If the response value is non-zero, the instrument did not
execute the previous command. Read the error message from
the following acknowledge subroutine, recover the error,
and repeat the command or query. (This is not shown in the
following program example.)
4. If a query was output to the instrument, input its
response.
5. The sequence of points 2, 3, and 4 may be repeated for
different commands or queries.
6. Close the communication channel.
Refer to the program example on the next page.
Page 2.8
’Example of a typical program sequence:
’***************** Begin example program ****************
OPEN "COM1:1200,N,8,1,CS,DS,RB2048" FOR RANDOM AS #1
’This QBASIC program line sets the parameters for the
’RS232 port (COM1 on the Computer) to match the
’instrument power-on default settings. It also opens a
’communication channel (assigned #1) for input or output
’through the COM1 port. Your instrument must be connected
’to this port. "RB2048" sets the size of the computer
’receive buffer to 2048 bytes to prevent buffer overflow
’during communication with the instrument.
PRINT #1, "ID"
’Outputs the IDENTITY command (query) to the instrument.
GOSUB Acknowledge
’This subroutine inputs the acknowledge response from
’the instrument and displays an error message if the
’acknowledge value is non-zero.
INPUT #1, Response$
’This inputs the response data from the IDENTITY query.
PRINT Response$
’Displays the queried data.
CLOSE #1
’This closes the communication channel.
END
’This ends the program.
’ Page 2.9
’**************** Acknowledge subroutine ******************
’Use this subroutine after each command or query sent to the
’instrument. This routine inputs the acknowledge
’response from the instrument. If the response is non-zero,
’the previous command was not correct or was not correctly
’received by the instrument. Then an error message is
’displayed and the program is aborted.
Acknowledge:
INPUT #1, ACK ’Reads acknowledge from instrument.
IF ACK <> 0 THEN
PRINT "Error "; ACK; ": ";
SELECT CASE ACK
CASE 1
PRINT "Syntax Error"
CASE 2
PRINT "Execution Error"
CASE 3
PRINT "Synchronization Error"
CASE 4
PRINT "Communication Error"
CASE IS < 1
PRINT "Unknown Acknowledge"
CASE IS > 4
PRINT "Unknown Acknowledge"
END SELECT
PRINT "Program aborted."
END
END IF
RETURN
’****************** End example program *****************
Each command description starts on a new page with a header
for quickly finding the command. This header indicates the
command name and the two-character header used for the
command syntax. Example:
===========================================================
AUTO SETUP AS
Where AUTO SETUP is a descriptive name for the command
(this is no syntax!),
and AS are the first two characters used for
the command syntax (not the complete
syntax).
- Purpose:
Explains what the command does or what it is used for.
- Command Syntax:
Shows the syntax for the command. Parameters are separated
by commas. Commands are terminated by <cr> (carriage
return).
- Response Syntax:
Shows the format of the response from the instrument.
Responses are terminated by <cr> (carriage return).
Each Response Syntax starts with the <acknowledge>
response, followed by the query response if the syntax
relates to a query.
- Example:
This is an example QBASIC program which shows how you can
use the command. The example may also include some other
commands to show the relation with these commands.
The following two comment lines (start with ’) successively
indicate the beginning and the end of an example program.
’***************** Begin example program ****************
’****************** End example program *****************
Page 3.2
Use an MS-DOS Editor and copy the complete program between
these two lines to a file name with the .BAS extension.
Start QBASIC and open this file from the FILE menu.
Long programs (longer than 55 lines) include page breaks.
Such page breaks are preceded by the ’ (remark) character
to prevent the QBASIC interpreter from interpreting them as
an incorrect statement.
When you have connected the Power Quality Analyzer as
indicated in the PM9080 Instruction Manual, you can start
the program from the RUN menu.
Page 3.3
** Syntax conventions **
The Command Syntax and the Response Syntax may contain the
following meta symbols and data elements:
UPPERCASE These characters are part of the syntax.
For commands, lower case is also allowed.
<...> An expression between these brackets is a
code, such as <cr> (carriage return) that can
not be expressed in a printable character, or
it is a parameter that is further specified.
Do not insert the brackets in the command!
[...] The item between these brackets is optional.
This means that you may omit it for the
command, or for a response it may not appear.
Do not insert the brackets in the command!
| This is a separator between selectable items.
This means that you must choose only one of
the items (exclusive or).
{...} Specifies an element that may be repeated 0 or
more instances.
<signed_integer> = <binary_character><binary_character>
Two bytes representing a signed
integer value. The first byte is the
most significant and contains the
sign bit (bit 7).
<signed_long> = four <binary_character>’s
<unsigned_integer>= <binary_character><binary_character>
Two bytes representing an unsigned
integer value. The first byte is the
most significant.
<unsigned_long> = four <binary_character>’s
** Overview of commands for the Power Quality Analyzer **
COMMAND PAGE
COMMAND NAME HEADER NUMBER
------------------------------------------------------ AUTO SETUP AS 3.5
ARM TRIGGER AT 3.7
CLEAR MEMORY CM 3.9
CPL VERSION QUERY CV 3.11
DEFAULT SETUP DS 3.13
GET DOWN GD 3.15
GO TO LOCAL GL 3.17
GO TO REMOTE GR 3.20
HOLD HO 3.21
IDENTIFICATION ID 3.23
INSTRUMENT STATUS IS 3.25
PROGRAM COMMUNICATION PC 3.28
PROGRAM SETUP PS 3.30
QUERY HARMONICS QH 3.34
QUERY MEASUREMENT QM 3.36
QUERY PRINT QP 3.40
QUERY SETUP QS 3.44
QUERY WAVEFORM QW 3.45
READ DATE RD 3.62
RESET INSTRUMENT RI 3.64
RECALL SETUP RS 3.66
READ TIME RT 3.69
SWITCH ON SO 3.71
SAVE SETUP SS 3.72
STATUS QUERY ST 3.73
TRIGGER ACQUISITION TA 3.76
WRITE DATE WD 3.78
WRITE TIME WT 3.80
Page 3.5
=============================================================
AUTO SETUP AS
Invokes an automatic setup for the active mode. The result
of this command is the same as pressing the AUTO key
on the instrument.
Note: You can select the items that are affected by the
AUTO SET procedure via the USER OPTIONS key on
the instrument.
Command Syntax:
AS<cr>
Response Syntax:
<acknowledge><cr>
Example:
The following example program sends an AUTO SETUP command to
the instrument. Connect a repetitive signal on
INPUT 1 to see the effect of AUTO SETUP.
’ Page 3.6
’***************** Begin example program *****************
CLS ’Clears the PC screen.
OPEN "COM1:1200,N,8,1,CS,DS,RB2048" FOR RANDOM AS #1
PRINT #1, "AS" ’Sends AUTO SETUP command.
GOSUB Acknowledge ’Input acknowledge from instrument.
CLOSE #1
END
’**************** Acknowledge subroutine ******************
’Use this subroutine after each command or query sent to the
’instrument. This routine inputs the acknowledge
’response from the instrument. If the response is non-zero,
’the previous command was not correct or was not correctly
’received by the instrument. Then an error message is
’displayed and the program is aborted.
Acknowledge:
INPUT #1, ACK ’Reads acknowledge from instrument.
IF ACK <> 0 THEN
PRINT "Error "; ACK; ": ";
SELECT CASE ACK
CASE 1
PRINT "Syntax Error"
CASE 2
PRINT "Execution Error"
CASE 3
PRINT "Synchronization Error"
CASE 4
PRINT "Communication Error"
CASE IS < 1
PRINT "Unknown Acknowledge"
CASE IS > 4
PRINT "Unknown Acknowledge"
END SELECT
PRINT "Program aborted."
END
END IF
RETURN
’****************** End example program ******************
Page 3.7
=============================================================
ARM TRIGGER AT
Resets and arms the trigger system for a new acquisition.
This command is used for single shot measurements.
When the AT command is given while an acquisition is in
progress, this acquisition is aborted and the trigger
system is rearmed.
Command Syntax:
AT<cr>
Response Syntax:
<acknowledge><cr>
Example:
The following example program arms the trigger system of
the instrument with the AT command.
This means that after this command the instrument starts an
acquisition when a trigger occurs from the signal (when
exceeding the trigger level) or from a TA (Trigger
Acquisition) command.
After the AT command it is assumed that the signal amplitude
is sufficient to trigger the acquisition. If it is not, you
can use the TA (TRIGGER ACQUISITION) command to force the
acquisition to be triggered. But this is not useful if you
want the acquisition to be started on a signal edge for
synchronization purposes.
Also see the example program for the IS command, which also
uses the AT command for a single shot application.
’***************** Begin example program *****************
OPEN "COM1:1200,N,8,1,CS,DS,RB2048" FOR RANDOM AS #1
PRINT #1, "AT" ’Sends the ARM TRIGGER command.
GOSUB Acknowledge ’Input acknowledge from instrument.
CLOSE #1
END
’
’ Page 3.8
’**************** Acknowledge subroutine ******************
’Use this subroutine after each command or query sent to the
’instrument. This routine inputs the acknowledge
’response from the instrument. If the response is non-zero,
’the previous command was not correct or was not correctly
’received by the instrument. Then an error message is
’displayed and the program is aborted.
Acknowledge:
INPUT #1, ACK ’Reads acknowledge from instrument.
IF ACK <> 0 THEN
PRINT "Error "; ACK; ": ";
SELECT CASE ACK
CASE 1
PRINT "Syntax Error"
CASE 2
PRINT "Execution Error"
CASE 3
PRINT "Synchronization Error"
CASE 4
PRINT "Communication Error"
CASE IS < 1
PRINT "Unknown Acknowledge"
CASE IS > 4
PRINT "Unknown Acknowledge"
END SELECT
PRINT "Program aborted."
END
END IF
RETURN
’****************** End example program ******************
Page 3.9
=============================================================
CLEAR MEMORY CM
Clears all saved setups, waveforms, and screens from
memory.
Command Syntax:
CM<cr>
Response Syntax:
<acknowledge><cr>
Example:
’ Page 3.10
’***************** Begin example program *****************
OPEN "COM1:1200,N,8,1,CS,DS,RB2048" FOR RANDOM AS #1
PRINT #1,"CM" ’Sends the Clear Memory command.
GOSUB Acknowledge ’Input acknowledge from instrument.
CLOSE #1
END
’**************** Acknowledge subroutine ******************
’Use this subroutine after each command or query sent to the
’instrument. This routine inputs the acknowledge
’response from the instrument. If the response is non-zero,
’the previous command was not correct or was not correctly
’received by the instrument. Then an error message is
’displayed and the program is aborted.
Acknowledge:
INPUT #1, ACK ’Reads acknowledge from instrument.
IF ACK <> 0 THEN
PRINT "Error "; ACK; ": ";
SELECT CASE ACK
CASE 1
PRINT "Syntax Error"
CASE 2
PRINT "Execution Error"
CASE 3
PRINT "Synchronization Error"
CASE 4
PRINT "Communication Error"
CASE IS < 1
PRINT "Unknown Acknowledge"
CASE IS > 4
PRINT "Unknown Acknowledge"
END SELECT
PRINT "Program aborted."
END
END IF
RETURN
’****************** End example program ******************
Page 3.11
=============================================================
CPL VERSION QUERY CV
<version> is an ASCII string representing the year this
version has been created.
Example:
’ Page 3.12
’***************** Begin example program *****************
OPEN "COM1:1200,N,8,1,CS,DS,RB2048" FOR RANDOM AS #1
PRINT #1,"CV" ’Sends CPL VERSION query.
GOSUB Acknowledge ’Input acknowledge from instrument.
INPUT #1,VERSION$ ’Inputs queried data.
PRINT "CPL Version "; VERSION$ ’Displays version data.
END
’**************** Acknowledge subroutine ******************
’Use this subroutine after each command or query sent to the
’instrument. This routine inputs the acknowledge
’response from the instrument. If the response is non-zero,
’the previous command was not correct or was not correctly
’received by the instrument. Then an error message is
’displayed and the program is aborted.
Acknowledge:
INPUT #1, ACK ’Reads acknowledge from instrument.
IF ACK <> 0 THEN
PRINT "Error "; ACK; ": ";
SELECT CASE ACK
CASE 1
PRINT "Syntax Error"
CASE 2
PRINT "Execution Error"
CASE 3
PRINT "Synchronization Error"
CASE 4
PRINT "Communication Error"
CASE IS < 1
PRINT "Unknown Acknowledge"
CASE IS > 4
PRINT "Unknown Acknowledge"
END SELECT
PRINT "Program aborted."
END
END IF
RETURN
’****************** End example program ******************
Resets the instrument to the factory settings at
delivery, except for the RS232 communication settings such
as baud rate, to keep the communication alive.
A Master Reset (refer to the Users Manual) performs the
same, but also resets the RS232 communication settings to
the default values.
Command Syntax:
DS<cr>
Response Syntax:
<acknowledge><cr>
Note: Wait for at least 2 seconds after the
<acknowledge> reply has been received, to let
the instrument settle itself before you send the
next command.
Example:
’ Page 3.14
’***************** Begin example program *****************
OPEN "COM1:1200,N,8,1,CS,DS,RB2048" FOR RANDOM AS #1
CLS
PRINT #1, "DS" ’Sends DEFAULT SETUP command.
GOSUB Acknowledge ’Input acknowledge from instrument.
SLEEP 2 ’Delay (2 s) necessary after "DS".
PRINT #1, "ID" ’Sends the IDENTIFICATION query.
GOSUB Acknowledge ’Input acknowledge from instrument.
INPUT #1, ID$ ’Inputs identity data from instrument.
PRINT ID$ ’Displays identity data.
CLOSE #1
END
’**************** Acknowledge subroutine ******************
’Use this subroutine after each command or query sent to the
’instrument. This routine inputs the acknowledge
’response from the instrument. If the response is non-zero,
’the previous command was not correct or was not correctly
’received by the instrument. Then an error message is
’displayed and the program is aborted.
Acknowledge:
INPUT #1, ACK ’Reads acknowledge from instrument.
IF ACK <> 0 THEN
PRINT "Error "; ACK; ": ";
SELECT CASE ACK
CASE 1
PRINT "Syntax Error"
CASE 2
PRINT "Execution Error"
CASE 3
PRINT "Synchronization Error"
CASE 4
PRINT "Communication Error"
CASE IS < 1
PRINT "Unknown Acknowledge"
CASE IS > 4
PRINT "Unknown Acknowledge"
END SELECT
PRINT "Program aborted."
END
END IF
RETURN
’****************** End example program ******************
Page 3.15
=============================================================
GET DOWN GD
Switches the instrument’s power off. If a power adapter
is connected, you can use the SO command to switch power
on again. If there is no power adapter connected, the
instrument can only be switched on manually by pressing
the Power ON/OFF key.
Command Syntax:
GD<cr>
Response Syntax:
<acknowledge><cr>
Example:
’ Page 3.16
’***************** Begin example program *****************
OPEN "COM1:1200,N,8,1,CS,DS,RB2048" FOR RANDOM AS #1
CLS
PRINT #1, "GD" ’Sends the GET DOWN command.
GOSUB Acknowledge ’Input acknowledge from instrument.
PRINT "The GET DOWN command switched the instrument off."
PRINT "Press any key on the PC keyboard to switch "
PRINT "the instrument on again."
SLEEP
PRINT #1, "SO" ’Sends the SWITCH ON command.
GOSUB Acknowledge ’Input acknowledge from instrument.
CLOSE #1
END
’**************** Acknowledge subroutine ******************
’Use this subroutine after each command or query sent to the
’instrument. This routine inputs the acknowledge
’response from the instrument. If the response is non-zero,
’the previous command was not correct or was not correctly
’received by the instrument. Then an error message is
’displayed and the program is aborted.
Acknowledge:
INPUT #1, ACK ’Reads acknowledge from instrument.
IF ACK <> 0 THEN
PRINT "Error "; ACK; ": ";
SELECT CASE ACK
CASE 1
PRINT "Syntax Error"
CASE 2
PRINT "Execution Error"
CASE 3
PRINT "Synchronization Error"
CASE 4
PRINT "Communication Error"
CASE IS < 1
PRINT "Unknown Acknowledge"
CASE IS > 4
PRINT "Unknown Acknowledge"
END SELECT
PRINT "Program aborted."
END
END IF
RETURN
’****************** End example program ******************
Page 3.17
=============================================================
GO TO LOCAL GL
Sets the instrument in the local operation mode
so the keypad is enabled.
Also refer to the GR (Go to Remote) command.
Command Syntax:
GL<cr>
Response Syntax:
<acknowledge><cr>
Example:
The following example uses the GR (GO TO REMOTE) command
(refer to the description for this command) to set the
instrument in the REMOTE state so that the keypad
is disabled. After that, the GL (GO TO LOCAL) command
is sent so that the keypad is enabled again.
’ Page 3.18
’***************** Begin example program *****************
CLS ’Clears the PC screen.
OPEN "COM1:1200,N,8,1,CS,DS,RB2048" FOR RANDOM AS #1
PRINT #1, "GR" ’Sends GO TO REMOTE command.
GOSUB Acknowledge ’Input acknowledge from instrument.
PRINT "All instrument keys (except the Power ON/OFF key)
PRINT "are now disabled by the GR (GO TO REMOTE) command."
PRINT "Check this."
PRINT
PRINT "Press any key on the PC keyboard to continue."
SLEEP
PRINT
PRINT #1, "GL" ’Sends GO TO LOCAL command.
GOSUB Acknowledge ’Input acknowledge from instrument.
PRINT "The instrument keys are now enabled again by the "
PRINT "GL (GO TO LOCAL) command."
PRINT "Check this."
CLOSE #1
END
’
’ Page 3.19
’**************** Acknowledge subroutine ******************
’Use this subroutine after each command or query sent to the
’instrument. This routine inputs the acknowledge
’response from the instrument. If the response is non-zero,
’the previous command was not correct or was not correctly
’received by the instrument. Then an error message is
’displayed and the program is aborted.
Acknowledge:
INPUT #1, ACK ’Reads acknowledge from instrument.
IF ACK <> 0 THEN
PRINT "Error "; ACK; ": ";
SELECT CASE ACK
CASE 1
PRINT "Syntax Error"
CASE 2
PRINT "Execution Error"
CASE 3
PRINT "Synchronization Error"
CASE 4
PRINT "Communication Error"
CASE IS < 1
PRINT "Unknown Acknowledge"
CASE IS > 4
PRINT "Unknown Acknowledge"
END SELECT
PRINT "Program aborted."
END
END IF
RETURN
’****************** End example program ******************
Page 3.20
=============================================================
GO TO REMOTE GR
Sets the instrument in the remote operation mode
so that the keypad is disabled.
You can use the following methods to return to the
local operation mode so that the keypad is enabled:
1. Sending the GL (Go to Local) command.
Command Syntax:
GR<cr>
Response Syntax:
<acknowledge><cr>
See an example for this command under GO TO LOCAL (GL).
Page 3.21
=============================================================
HOLD HO
Sets the instrument in the Hold mode. In other words, the
instrument stops sampling the input channels and
calculating measurement results.
Command Syntax:
HO<cr>
Response Syntax:
<acknowledge><cr>
Example:
’ Page 3.22
’***************** Begin example program *****************
OPEN "COM1:1200,N,8,1,CS,DS,RB2048" FOR RANDOM AS #1
CLS
PRINT #1, "HO" ’Sends the HOLD command.
GOSUB Acknowledge ’Input acknowledge from instrument.
PRINT "The HOLD command has put the instrument in HOLD."
PRINT "Check on the instrument screen."
PRINT "Press any key on the PC keyboard to continue and"
PRINT "enable acquisition again."
SLEEP
PRINT #1, "AT" ’Sends the ARM TRIGGER command to
’enable acquisition again.
GOSUB Acknowledge ’Input acknowledge from instrument.
CLOSE #1
END
’**************** Acknowledge subroutine ******************
’Use this subroutine after each command or query sent to the
’instrument. This routine inputs the acknowledge
’response from the instrument. If the response is non-zero,
’the previous command was not correct or was not correctly
’received by the instrument. Then an error message is
’displayed and the program is aborted.
Acknowledge:
INPUT #1, ACK ’Reads acknowledge from instrument.
IF ACK <> 0 THEN
PRINT "Error "; ACK; ": ";
SELECT CASE ACK
CASE 1
PRINT "Syntax Error"
CASE 2
PRINT "Execution Error"
CASE 3
PRINT "Synchronization Error"
CASE 4
PRINT "Communication Error"
CASE IS < 1
PRINT "Unknown Acknowledge"
CASE IS > 4
PRINT "Unknown Acknowledge"
END SELECT
PRINT "Program aborted."
END
END IF
RETURN
’****************** End example program ******************
Page 3.23
=============================================================
IDENTIFICATION ID
Returns the instrument model identification information.
Command Syntax:
ID<cr>
Response Syntax:
<acknowledge><cr>[<identity><cr>]
where,
<identity> is an ASCII string containing the following
data elements:
<model_number>;<software_version>;
<creation_date>;<languages>
Example:
The following example program queries the identity data of
the instrument and displays this data on the PC screen.
’ Page 3.24
’***************** Begin example program *****************
CLS ’Clears the PC screen.
OPEN "COM1:1200,N,8,1,CS,DS,RB2048" FOR RANDOM AS #1
PRINT #1, "ID" ’Sends IDENTIFICATION query.
GOSUB Acknowledge ’Input acknowledge from instrument.
INPUT #1, IDENT$ ’Inputs the queried data.
PRINT IDENT$ ’Displays queried data.
CLOSE #1
END
’**************** Acknowledge subroutine ******************
’Use this subroutine after each command or query sent to the
’instrument. This routine inputs the acknowledge
’response from the instrument. If the response is non-zero,
’the previous command was not correct or was not correctly
’received by the instrument. Then an error message is
’displayed and the program is aborted.
Acknowledge:
INPUT #1, ACK ’Reads acknowledge from instrument.
IF ACK <> 0 THEN
PRINT "Error "; ACK; ": ";
SELECT CASE ACK
CASE 1
PRINT "Syntax Error"
CASE 2
PRINT "Execution Error"
CASE 3
PRINT "Synchronization Error"
CASE 4
PRINT "Communication Error"
CASE IS < 1
PRINT "Unknown Acknowledge"
CASE IS > 4
PRINT "Unknown Acknowledge"
END SELECT
PRINT "Program aborted."
END
END IF
RETURN
’****************** End example program ******************
Page 3.25
=============================================================
INSTRUMENT STATUS IS
Queries the contents of the instrument’s status register.
The returned value reflects the present operational status
of the instrument. This is a 16-bit word, presented as an
integer value, where each bit represents the Boolean value
of a related event.
Command Syntax:
IS<cr>
Response Syntax:
<acknowledge><cr>[<status><cr>]
where,
<status> = integer value 0 to 65535
<status>
Bit Value Status Description
--------------------------------------------------------- 0 1 Maintenance mode
1 2 Charging
2 4 Recording
3 8 AutoRanging
4 16 Remote
5 32 Battery Connected
6 64 Power (Net) Adapter connected
7 128 Calibration necessary
8 256 Instrument Held (HOLD status)
9 512 Pre Calibration busy
10 1024 Pre Calibration valid
11 2048
12 4096 Triggered
13 8192 Instrument On
14 16384 Instrument Reset occurred
15 32768 Next <status> value available
Example:
Page 3.26
’***************** Begin example program *****************
CLS ’Clears the PC screen
OPEN "COM1:1200,N,8,1,CS,DS,RB2048" FOR RANDOM AS #1
PRINT #1, "IS" ’Sends the INSTRUMENT STATUS query
GOSUB Acknowledge ’Input acknowledge from instrument
INPUT #1, Status$ ’Input Instrument Status
StV = VAL(Status$) ’Decimal value of Instrument Status
PRINT "Instrument Status : "; StV
IF (StV AND 1) = 1 THEN PRINT " instrument in Maintenance mode."
IF (StV AND 2) = 2 THEN PRINT " instrument charging."
IF (StV AND 4) = 4 THEN PRINT " instrument recording."
IF (StV AND 8) = 8 THEN PRINT " AutoRanging active"
IF (StV AND 16) = 16 THEN PRINT " instrument remote."
IF (StV AND 32) = 32 THEN PRINT " Battery connected."
IF (StV AND 64) = 64 THEN PRINT " Power Adapter connected."
IF (StV AND 128) = 128 THEN PRINT " Calibration necessary."
IF (StV AND 256) = 256 THEN PRINT " instrument in HOLD."
IF (StV AND 512) = 512 THEN PRINT " Pre-calibration busy."
IF (StV AND 1024) = 1024 THEN PRINT " Pre-calibration valid."
IF (StV AND 4096) = 4096 THEN PRINT " instrument triggered."
IF (StV AND 8192) = 8192 THEN
PRINT " instrument On."
ELSE
PRINT " instrument Off."
END IF
IF (StV AND 16384) = 16384 THEN PRINT " Reset Instrument occurred."
END
’
’ Page 3.27
’**************** Acknowledge subroutine ******************
’Use this subroutine after each command or query sent to the
’instrument. This routine inputs the acknowledge
’response from the instrument. If the response is non-zero,
’the previous command was not correct or was not correctly
’received by the instrument. Then an error message is
’displayed and the program is aborted.
Acknowledge:
INPUT #1, ACK ’Reads acknowledge from instrument.
IF ACK <> 0 THEN
PRINT "Error "; ACK; ": ";
SELECT CASE ACK
CASE 1
PRINT "Syntax Error"
CASE 2
PRINT "Execution Error"
CASE 3
PRINT "Synchronization Error"
CASE 4
PRINT "Communication Error"
CASE IS < 1
PRINT "Unknown Acknowledge"
CASE IS > 4
PRINT "Unknown Acknowledge"
END SELECT
PRINT "Program aborted."
END
END IF
RETURN
’****************** End example program ******************
Page 3.28
=============================================================
PROGRAM COMMUNICATION PC
Restores a complete setup, previously saved with the SS
(Save Setup) command and queried with the QS (Query Setup)
command and saved in a string variable or to a file.
Command Syntax 1:
PS [<saved_setup_no>]<cr>
where,
<saved_setup_no> = 0 to 10
This is the register number where a
setup is stored. Also see the
description of the Save Setup (SS)
command (0 is actual setup).
Response Syntax 1:
<acknowledge><cr>
Command Syntax 2:
<queried_setup><cr>
<queried_setup> = The data returned with the QS command.
(<omit the <acknowledge><cr> response).
Response Syntax 2:
<acknowledge><cr>
Note: Wait for at least two seconds after the
<acknowledge> reply has been received, to let
the instrument settle itself before you send the
next command.
Remarks:
The instrument sends the <acknowledge> reply
after it has executed the setup from the PS command.
You must send the <setup> string as a whole, exactly as
returned from the QS (Query Setup) command.
If you do not follow this rule, the instrument
may crash. A Reset may then be necessary to recover
the instrument. (Refer to the instrument Users Manual.)
Example:
The following example program demonstrates the use of the
QS (QUERY SETUP) and the PS (PROGRAM SETUP) commands.
The present setup is queried from instrument and saved to
file. The program asks you to change the instrument settings.
Then the original setup is read from file and sent back
to the instrument.
’ Page 3.31
’***************** Begin example program *****************
OPEN "COM1:1200,N,8,1,CS,DS,RB2048" FOR RANDOM AS #1
CLS
GOSUB ClearPort ’Clears pending data from port.
PRINT #1, "QS" ’Queries the actual setup data.
GOSUB Acknowledge ’Input acknowledge from instrument.
GOSUB Response ’Writes the setup data to file.
PRINT "Present setup data are stored in the file SETUP0"
PRINT "This setup will now be retrieved from the file and"
PRINT "sent back to the instrument."
PRINT "To see if this works, change the present settings and"
PRINT "verify if the instrument returns to the previous"
PRINT "settings."
PRINT
PRINT "Press any key on the PC keyboard to continue."
SLEEP
CLS
PRINT #1, "PS" ’Program header for programming
’the setup data to the instrument.
GOSUB Acknowledge ’Input acknowledge from instrument.
OPEN "SETUP0" FOR INPUT AS #2
’Opens file SETUP0 for data retrieval.
DO WHILE NOT EOF(2)
SUCHR$ = INPUT$(1, #2) ’Reads setup data from file
PRINT #1, SUCHR$; ’Programs instrument with the"
’setup data stored in SETUP0$.
LOOP
PRINT #1, CHR$(13); ’Program message terminator
CLOSE #2 ’Close file SETUP0.
GOSUB Acknowledge ’Input acknowledge from instrument.
END
’
’ Page 3.32
’**************** Acknowledge subroutine ******************
’Use this subroutine after each command or query sent to the
’instrument. This routine inputs the acknowledge
’response from the instrument. If the response is non-zero,
’the previous command was not correct or was not correctly
’received by the instrument. Then an error message is
’displayed and the program is aborted.
Acknowledge:
INPUT #1, ACK ’Reads acknowledge from instrument.
IF ACK <> 0 THEN
PRINT "Error "; ACK; ": ";
SELECT CASE ACK
CASE 1
PRINT "Syntax Error"
CASE 2
PRINT "Execution Error"
CASE 3
PRINT "Synchronization Error"
CASE 4
PRINT "Communication Error"
CASE IS < 1
PRINT "Unknown Acknowledge"
CASE IS > 4
PRINT "Unknown Acknowledge"
END SELECT
PRINT "Program aborted."
END
END IF
RETURN
’******* Clears pending data from the RS232 port *********
ClearPort:
WHILE LOC(1) > 0
Dummy$ = INPUT$(1, #1)
WEND
RETURN
’
’ Page 3.33
’****************** Response subroutine *********************
’This subroutine reads bytes from the RS232 buffer as long
’as they enter. When no bytes enter for 1 second, the program
’assumes that the instrument has terminated its response.
’All bytes that enter the buffer are appended to the string
’Resp$.
Response:
start! = TIMER
’Wait for bytes (maximum 1 s) to enter RS232 buffer
WHILE ((TIMER < (start! + 1)) AND (LOC(1) = 0))
WEND
IF LOC(1) > 0 THEN ’If RS232 buffer contains bytes
OPEN "Setup0" FOR OUTPUT AS #2 ’File for setup data
DO
’ LOC(1) gives the number of bytes waiting:
ScopeInput$ = INPUT$(LOC(1), #1) ’Input bytes
PRINT #2, ScopeInput$;
start! = TIMER
WHILE ((TIMER < (start! + 1)) AND (LOC(1) = 0))
WEND
LOOP WHILE LOC(1) > 0 ’Repeat as long as bytes enter
CLOSE #2
END IF
RETURN
’****************** End example program ******************
<check_sum> = <binary_character>
One binary character which represents the sum
of all the <binary_character>’s sent after
the <block_length> and before the <check_sum>.
<block_length>= <unsigned_integer>
This (2-bytes) value gives the number of bytes
that are transmitted after the <block_length>
and before the <check_sum>.
<nbr_of_samples>= <unsigned_integer>
Total number of harmonics samples that follow.
<samples> = {<harm_amplitude><harm_phase>}
<harm_amplitude>= <float>
<harm_phase> = <signed integer>
<check_sum> = <binary_character>
One binary character which represents the sum
of all the <binary_character>’s sent after
the <block_length> and before the <check_sum>.
Example: The method for reading and analyzing harmonics data (QH)
is similar for reading and analyzing waveforms (QW).
Queries for active readings (see Syntax 1) or measurement
results from the instrument (see Syntax 2).
Command Syntax 1:
QM<cr>
Command Syntax 2:
QM <no>{,<no>}<cr>
where in VOLTS/AMPS/HERTZ mode:
<no> = 11 : Voltage rms (ac + dc)
21 : Ampere rms (ac + dc)
31 : Line Frequency on trigger channel
41 : Crest on channel 1 (Fluke 43B only)
51 : Crest on channel 2 (Fluke 43B only)
where in POWER mode: (relative to %r or %f)
<no> = 11 : Real power
21 : Apparent power
31 : Reactive power
41 : Total Power Factor (TPF) (Fluke 43B only)
51 : Displacement Power Factor (DPF or cos phi)
61 : Line Frequency
where in HARMONICS mode: (Volt/Ampere/Watt mode)
<no> = 11 : Total Harmonic Distortion (THD, not Watts mode)
21 : Total rms reading
31 : K-factor (not in Volts mode)
41 : Frequency of selected harmonic component
51 : Absolute rms of selected harmonic component
61 : Relative rms (% of Total rms) of selected
harmonic component
71 : Phase of selected harmonic component
where in SAGS & SWELLS mode:
<no> = 11 : Voltage rms (ac + dc)
21 : Ampere rms (ac + dc)
where in TRANSIENTS mode: (at cursor position of the
selected transient)
<no> = 11 : Maximum Voltage peak
21 : Minimum Voltage peak
31 : Maximum Ampere peak
41 : Minimum Ampere peak
where in TRANSIENTS Learn mode:
<no> = 11 : Voltage rms (ac + dc)
21 : Line Frequency on Voltage channel
where in Motor INRUSH CURRENT mode:
<no> = 11 : Ampere amplitude at first cursor position
21 : Ampere amplitude at second cursor position
where in Meter OHMS/DIODE/CONTINUITY/CAPACITANCE
and TEMPERATURE mode:
<no> = 11 : Meter reading
where in SCOPE mode:
<no> = 11 : Reading of channel 1 (can be switched off)
21 : Reading of channel 2 (can be switched off)
Notes: - Maximum 10 readings per command.
- If one of the readings <no> is non-valid, no readings
will be returned.
- Only active (valid) readings will be returned.
Page 3.37
Response Syntax 1:
<acknowledge><cr>[<reading>{,<reading>}<cr>]
where,
<reading> = <no>,<valid>,<source>,<unit>,<type>,<pres>,<resol>
<no> see Command Syntax 2
<valid> validity of the reading:
1 reading valid
0 reading non-valid
<source> source of the reading:
1 Voltage channel: Input A (Scope mode)
2 Ampere channel: Input B (Scope mode)
3 Input external: COM & V/Ohm/Diode (Meter mode)
12 Input_AB (Phase A over B): A>B (Scope mode)
21 Input_BA (Phase B over A): B>A (Scope mode)
<unit> unit of the reading:
0 None (off)
1 Volt
2 Ampere
3 Ohm
4 Watt
5 Farad
6 Kelvin
7 seconds
8 hours
9 days
10 Hertz
11 Degrees
12 Celsius
13 Fahrenheit
14 percentage (%)
15 dBm 50 Ohm
16 dBm 600 Ohm
17 dB Volt
18 dB Ampere
19 dB Watt
20 Volt * Ampere Reactive (VAR)
21 Volt * Ampere (VA)
<type> reading characteristic of the measurement:
0 None
1 Mean
2 Rms
3 True rms
4 Peak peak
5 Peak maximum
6 Peak minimum
7 Crest factor
8 Period
9 Duty cycle negative
10 Duty cycle positive
11 Frequency
12 Pulse width negative
13 Pulse width positive
14 Phase
15 Diode
16 Continuity
18 Reactive Power
19 Apparent Power
20 Real Power
21 Harmonic Reactive Power
22 Harmonic Apparent Power
23 Harmonic Real Power
24 Harmonic rms
25 Displacement Power Factor
26 Total Power Factor
27 Total Harmonic Distortion
28 Total Harmonic Distortion with respect
to Fundamental
29 K Factor (European definition)
30 K Factor (US definition)
31 Line Frequency
32 AC average
<pres> presentation value of the reading:
0 Absolute value
1 Relative value
2 Logarithmic value
3 Linear value
4 Fahrenheit
5 Celsius
<resol> resolution of the reading as <float> to
determine the least significant digit
Note: Only displayed results are available for output.
Page 3.38
Example:
’***************** Begin example program *****************
CLS ’Clears the PC screen.
OPEN "COM1:1200,N,8,1,CS,DS,RB2048" FOR RANDOM AS #1
PRINT #1, "QM" ’Queries for active readings
GOSUB Acknowledge ’Input acknowledge from instrument.
’*** Examines only the 7 inputs of the first reading <no> 11.
INPUT #1, reading.no’1st <decimal_number>
IF reading.no = 11 THEN
PRINT "Measurement reading 1";
ELSEIF reading.no = 21 THEN
PRINT "Measurement reading 2";
ELSE
PRINT "Unknown measurement reading";
END IF
INPUT #1, validity’2nd <decimal_number>
IF validity = 1 THEN
PRINT " is valid"
ELSE
PRINT " is ’not’ valid"
END IF
INPUT #1, source’3rd <decimal_number>
PRINT "Source of reading = ";
IF source = 1 THEN
PRINT "Voltage channel Input A"
ELSEIF source = 2 THEN
PRINT "Ampere channel Input B"
ELSEIF source = 3 THEN
PRINT "Input External"
ELSE
PRINT "Unknown source?"
END IF
INPUT #1, unit’4th <decimal_number>
PRINT "Unit of reading = ";
IF unit = 1 THEN
PRINT "Volt"
ELSEIF unit = 2 THEN
PRINT "Ampere"
ELSEIF unit = 3 THEN
PRINT "Ohm"
ELSE
PRINT "Unexpected unit?"
END IF
INPUT #1, types’5th <decimal_number>
PRINT "Type of reading = ";
IF types = 1 THEN
PRINT "Mean value"
ELSEIF types = 2 THEN
PRINT "Rms value"
ELSEIF types = 3 THEN
PRINT "True rms value"
ELSE
PRINT "Unexpected characteristic?"
END IF
INPUT #1, presentation’6th <decimal_number>
PRINT "Presentation of reading= ";
IF presentation = 0 THEN
PRINT "Absolute value"
ELSEIF presentation = 1 THEN
PRINT "Relative value"
ELSEIF presentation = 2 THEN
PRINT "Logarithmic value"
ELSE
PRINT "Unexpected value?"
END IF
INPUT #1, resolution’7th <decimal_number>
PRINT "Resolution of reading ="; resolution
GOSUB ClearReadings ’Clears rest of readings data from port
’
PRINT #1, "QM 11" ’Queries Measurement reading 1 or
’Meter absolute reading (Meter mode).
GOSUB Acknowledge ’Input acknowledge from instrument.
INPUT #1, result
PRINT "Measurement value ="; result; "V"
CLOSE #1
END
’
’ Page 3.39
’**************** Acknowledge subroutine ******************
’Use this subroutine after each command or query sent to the
’instrument. This routine inputs the acknowledge
’response from the instrument. If the response is non-zero,
’the previous command was not correct or was not correctly
’received by the instrument. Then an error message is
’displayed and the program is aborted.
Acknowledge:
INPUT #1, ACK ’Reads acknowledge from instrument.
IF ACK <> 0 THEN
PRINT "Error "; ACK; ": ";
SELECT CASE ACK
CASE 1
PRINT "Syntax Error"
CASE 2
PRINT "Execution Error"
CASE 3
PRINT "Synchronization Error"
CASE 4
PRINT "Communication Error"
CASE IS < 1
PRINT "Unknown Acknowledge"
CASE IS > 4
PRINT "Unknown Acknowledge"
END SELECT
PRINT "Program aborted."
END
END IF
RETURN
’******* Clears pending data from the RS232 port *********
ClearReadings:
LINE INPUT #1, dummy$
WHILE LOC(1) > 0
LINE INPUT #1, dummy$
WEND
RETURN
’****************** End example program ******************
<printer_data>
This data can directly be sent to the printer to get a
screen copy on paper.
Example:
The following program reads the instrument screen (print)
data and copies this data to the file Qpfile. Hereafter, you
can copy this file to the printer port LPT1, for example.
The Read Buffer length for the PC is set to 7500 bytes to
prevent buffer overflow during input from the instrument.
The data transfer speed (baud rate) is set to 19200 and after
the output it is set back to 1200 (default baud rate).
’ Page 3.41
’***************** Begin example program *****************
CLS
OPEN "COM1:1200,N,8,1,CS,DS,RB7500" FOR RANDOM AS #1
’Programs COM1 port parameters to
’match with the instrument power-on
’defaults.
PRINT #1, "PC 19200" ’Programs instrument to the maximum
’baud rate.
GOSUB Acknowledge ’Input acknowledge from instrument.
CLOSE #1
OPEN "COM1:19200,N,8,1,CS,DS,RB7500" FOR RANDOM AS #1
’Programs COM1 port parameters to
’match with the new instrument
’settings.
PRINT #1, "QP 0,1" ’Sends QUERY PRINT data command.
’(actual screen for LaserJet print)
GOSUB Acknowledge ’Input acknowledge from instrument.
PRINT
PRINT "Busy reading print data !"
PRINT
GOSUB Response
PRINT #1, "PC 1200" ’Programs instrument back to the
’default baud rate.
GOSUB Acknowledge ’Input acknowledge from instrument.
PRINT "Print data copied to file ’QPFILE’."
PRINT "You can copy the file contents to the Laser Printer."
PRINT "DOS-example: COPY Qpfile LPT1"
CLOSE ’Close all files.
END
’
’ Page 3.42
’**************** Acknowledge subroutine ******************
’Use this subroutine after each command or query sent to the
’instrument. This routine inputs the acknowledge
’response from the instrument. If the response is non-zero,
’the previous command was not correct or was not correctly
’received by the instrument. Then an error message is
’displayed and the program is aborted.
Acknowledge:
INPUT #1, ACK ’Reads acknowledge from instrument.
IF ACK <> 0 THEN
PRINT "Error "; ACK; ": ";
SELECT CASE ACK
CASE 1
PRINT "Syntax Error"
CASE 2
PRINT "Execution Error"
CASE 3
PRINT "Synchronization Error"
CASE 4
PRINT "Communication Error"
CASE IS < 1
PRINT "Unknown Acknowledge"
CASE IS > 4
PRINT "Unknown Acknowledge"
END SELECT
PRINT "Program aborted."
END
END IF
RETURN
’
’ Page 3.43
’****************** Response subroutine *********************
’This subroutine reads bytes from the RS232 buffer as long
’as they enter. When no bytes enter for 1 second, the program
’assumes that the instrument has terminated its response.
’All bytes that enter the buffer are appended to the string
’Resp$.
Response:
start! = TIMER
’Wait for bytes (maximum 2 s) to enter RS232 buffer
WHILE ((TIMER < (start! + 2)) AND (LOC(1) = 0))
WEND
IF LOC(1) > 0 THEN ’If RS232 buffer contains bytes
Resp$ = ""
OPEN "Qpfile" FOR OUTPUT AS #2 ’File for print data
DO
’ LOC(1) gives the number of bytes waiting:
ScopeInput$ = INPUT$(LOC(1), #1) ’Input bytes
PRINT #2, ScopeInput$;
start! = TIMER
WHILE ((TIMER < (start! + 2)) AND (LOC(1) = 0))
WEND
LOOP WHILE LOC(1) > 0 ’Repeat as long as bytes enter
CLOSE #2
END IF
RETURN
’****************** End example program ******************
Queries the present acquisition setup data from the
instrument.
Command Syntax:
QS [<setup_no>]<cr>
where,
<saved_setup_no> = 0 to 10
This is the register number where a
setup is stored. Also see the
description of the Save Setup (SS)
command (0 is actual setup).
Response Syntax:
<acknowledge><cr>[#0{<node>}<cr>]
where,
<node> = <node_header><node_identifier><node_length>
[<node_data>]<check_sum>
<node_header> = <binary_character>
Possible values:
20 hex All nodes except the last node
A0 hex End node
<node_identifier> = <binary_character>
Unique number for each specific node.
<node_length> = <unsigned_integer>
Specifies the number of <binary_character>
fields that follow in the <node_data>
field.
<node_data> = {<binary_character>}
The contents of <node_data> depends on the
<node_identifier> and the selected setup.
<check_sum> = <binary_character>
Contains the sum of all the binary bytes
in the <node_dat> field.
Note: Also see the Program Setup (PS) command.
See an example for this command under PROGRAM SETUP (PS).
Queries the trace data (administration and/or sample data)
related to the waveform from the instrument.
When a waveform is queried that is still under processing,
the processing is finished first (no half traces returned).
Command Syntax:
QW returns list of available trace numbers
QW <trace_no>[,V|S]
<trace_no> = <decimal number> assigned to the following
Trace Sources:
<trace_no> Trace Source:
-------------------------------------------- 10 MinMax voltage trace INPUT 1
Modes: VOLTS/AMPS/HERTZ
POWER
TRANSIENTS
SCOPE
11 24-bits Min/Max/Avg record trace 1
Modes: VOLTS/AMPS/HERTZ (see Note)
POWER (see Note)
HARMONICS (see Note)
SAGS & SWELLS
SCOPE (see Note)
Meter (see Note)
20 MinMax current trace INPUT 2
Modes: VOLTS/AMPS/HERTZ
POWER
TRANSIENTS
INRUSH CURRENT
SCOPE (see Note)
21 24-bits Min/Max/Avg record trace 2
Modes: VOLTS/AMPS/HERTZ (see Note)
POWER (see Note)
HARMONICS (see Note)
SAGS & SWELLS
SCOPE (see Note)
Note: Availability depends on other instrument
setup variables.
V | v Trace values (samples) only
S | s Setup (administration) data only. When V or S is
omitted, trace vales and setup data are returned.
<trace_data> = <trace_admin>,<trace_samples><cr>
This includes the complete information about the trace
(waveform). For detailed descriptions about the waveform
structure, refer to Appendix C.
If option V or v (value only) is given:
<trace_data> = <trace_samples><cr>
For detailed descriptions about the waveform structure,
refer to Appendix C.
If option S or s (Setup data only) is given:
<trace_data> = <trace_admin><cr>
where,
<trace_admin> = string of hexadecimal characters,
representing the setup related to the given
<trace_no>.
Page 3.47
Example:
’***************** Begin example program *****************
’
’***** If an error occurs in the waveform data,
’***** the program stops.
’
C65536 = 65536 ’2-bytes Maximum constant
C32768 = 32768 ’2-bytes Sign-bit constant
C256 = 256 ’1-byte Maximum constant
C128 = 128 ’1-byte Sign-bit constant
OPEN "COM1:1200,N,8,1,CS,DS,RB2048" FOR RANDOM AS #1
CLS
GOSUB ClearPort ’Clears pending data from port
’
Query$ = "QW 10" ’Queries voltage trace INPUT 1
’
’*****
’* A normal trace is a series of waveform samples consisting
’* of single waveform points.
’* A min/max trace is a series of waveform samples consisting
’* of minimum and maximum waveform points.
’* A min/max/average trace is a series of waveform samples
’* consisting of minimum, maximum, and average waveform points.
’*****
PRINT #1, Query$ ’Response = <trace_admin>,<trace_samples>
GOSUB Acknowledge ’Inputs acknowledge from instrument
Resp$ = "" ’Clears the total Response string
GOSUB Response ’Writes waveform data to Resp$ & files
GOSUB Interpret.Admin ’Interprets waveform administration data
’See also Appendix C
GOSUB Interpret.Samples ’Interprets waveform sample data
GOSUB Create.CSV ’Creates Wave.CSV file from waveform data
’as input for Excel, for example.
END
’
’ Page 3.48
’**************** Acknowledge subroutine ******************
’Use this subroutine after each command or query sent to the
’instrument. This routine inputs the acknowledge
’response from the instrument. If the response is non-zero,
’the previous command was not correct or was not correctly
’received by the instrument. Then an error message is
’displayed and the program is aborted.
Acknowledge:
INPUT #1, ACK ’Reads acknowledge from instrument.
IF ACK <> 0 THEN
PRINT "Error "; ACK; ": ";
SELECT CASE ACK
CASE 1
PRINT "Syntax Error"
CASE 2
PRINT "Execution Error"
CASE 3
PRINT "Synchronization Error"
CASE 4
PRINT "Communication Error"
CASE IS < 1
PRINT "Unknown Acknowledge"
CASE IS > 4
PRINT "Unknown Acknowledge"
END SELECT
PRINT "Program aborted."
END
END IF
RETURN
’******* Clears pending data from the RS232 port *********
ClearPort:
WHILE LOC(1) > 0
Dummy$ = INPUT$(1, #1)
WEND
RETURN
’
’ Page 3.49
’****************** Response subroutine *********************
’This subroutine reads bytes from the RS232 buffer as long
’as they enter. When no bytes enter for 1 second, the program
’assumes that the instrument has terminated its response. All
’bytes that enter the buffer are appended to the string Resp$
’and are written to the following files:
’File Waveform : the waveform data bytes
’File Waveresp : the waveform ASCII values
’
Response:
start! = TIMER
’Wait for bytes (maximum 1 s) to enter RS232 buffer
WHILE ((TIMER < (start! + 1)) AND (LOC(1) = 0))
WEND
IF LOC(1) > 0 THEN ’If RS232 buffer contains bytes
OPEN "WaveForm" FOR OUTPUT AS #2
’File to contain the waveform data bytes
docount = 1
total.count& = 0
DO
’ LOC(1) gives the number of bytes waiting:
total.count& = total.count& + LOC(1)
ScopeInput$ = INPUT$(LOC(1), #1) ’Input bytes
PRINT #2, ScopeInput$;
PRINT total.count&;
Resp$ = Resp$ + ScopeInput$
start! = TIMER
WHILE ((TIMER < (start! + 1)) AND (LOC(1) = 0))
WEND
docount = docount + 1
LOOP WHILE LOC(1) > 0 ’Repeat as long as bytes enter
CLOSE #2
PRINT
END IF
’
’***** Write the total Response string to file WaveResp
’
OPEN "WaveResp" FOR OUTPUT AS #3
PRINT "Response data length = "; LEN(Resp$)
PRINT #3, "Response data length = "; LEN(Resp$)
FOR i = 1 TO LEN(Resp$)
PRINT #3, ASC(MID$(Resp$, i, 1));
NEXT i
CLOSE #3: RETURN
’
’ Page 3.51
’
DIM expscale(2)’Exponents for Y/X.Scale
DIM YXscale#(2)’Values for Y/X.Scale
’
’***** 3 bytes <y_scale> = <mantissa_high><mantissa_low><exponent>
’***** <mantissa> = <mantissa_high> * 256 + <mantissa_low>
’***** <y_scale> = <sign><mantissa> E <sign><exponent>
’***** Example: +123E-4 = 123 / 10000 = 0.0123
FOR i = 0 TO 2
SumCheck1% = (SumCheck1% + ASC(MID$(Resp$,Resp.Count+i,1))) MOD 256
NEXT i
nb = ASC(MID$(Resp$, Resp.Count, 1))
IF nb >= 128 THEN
nb = - (256 - nb) * 256’Negative value
nb = nb + ASC(MID$(Resp$, Resp.Count + 1, 1))
ELSE
nb = nb * 256’Positive value
nb = nb + ASC(MID$(Resp$, Resp.Count + 1, 1))
END IF
expscale(1) = ASC(MID$(Resp$, Resp.Count + 2, 1))
YXscale#(1) = nb
Resp.Count = Resp.Count + 3
’*****
’* Further calculation after ’Signed.Samples’ determination
’*****
’***** 3 bytes <x_scale> = <mantissa_high><mantissa_low><exponent>
’***** <mantissa> = <mantissa_high> * 256 + <mantissa_low>
’***** <x_scale> = <sign><mantissa> E <sign><exponent>
’***** Example: +123E-4 = 123 / 10000 = 0.0123
FOR i = 0 TO 2
SumCheck1% = (SumCheck1% + ASC(MID$(Resp$,Resp.Count+i,1))) MOD 256
NEXT i
nb = ASC(MID$(Resp$, Resp.Count, 1))
IF nb >= 128 THEN
nb = - (256 - nb) * 256’Negative value
nb = nb + ASC(MID$(Resp$, Resp.Count + 1, 1))
ELSE
’ Page 3.52
’
DIM exponent(6)’Exponents for Y/X.Zero & Y/X.Resol & Y/X.At.0
DIM YXvalue#(6)’Values for Y/X.Zero & Y/X.Resol & Y/X.At.0
’
’***** 3 bytes <y_zero> = <mantissa_high><mantissa_low><exponent>
’***** <mantissa> = <mantissa_high> * 256 + <mantissa_low>
’***** <y_zero> = <sign><mantissa> E <sign><exponent>
’***** Example: +123E-4 = 123 / 10000 = 0.0123
FOR i = 0 TO 2
SumCheck1% = (SumCheck1% + ASC(MID$(Resp$,Resp.Count+i,1))) MOD 256
NEXT i
nb = ASC(MID$(Resp$, Resp.Count, 1))
IF nb >= 128 THEN
nb = - (256 - nb) * 256’Negative value
nb = nb + ASC(MID$(Resp$, Resp.Count + 1, 1))
ELSE
nb = nb * 256’Positive value
nb = nb + ASC(MID$(Resp$, Resp.Count + 1, 1))
END IF
exponent(1) = ASC(MID$(Resp$, Resp.Count + 2, 1))
YXvalue#(1) = nb
Resp.Count = Resp.Count + 3
’*****
’* Further calculation after ’Signed.Samples’ determination
’*****
’***** 3 bytes <x_zero> = <mantissa_high><mantissa_low><exponent>
’***** <mantissa> = <mantissa_high> * 256 + <mantissa_low>
’***** <x_zero> = <sign><mantissa> E <sign><exponent>
’***** Example: +123E-4 = 123 / 10000 = 0.0123
FOR i = 0 TO 2
SumCheck1% = (SumCheck1% + ASC(MID$(Resp$,Resp.Count+i,1))) MOD 256
NEXT i
nb = ASC(MID$(Resp$, Resp.Count, 1))
IF nb >= 128 THEN
nb = - (256 - nb) * 256’Negative value
nb = nb + ASC(MID$(Resp$, Resp.Count + 1, 1))
ELSE
nb = nb * 256’Positive value
nb = nb + ASC(MID$(Resp$, Resp.Count + 1, 1))
END IF
exponent(2) = ASC(MID$(Resp$, Resp.Count + 2, 1))
YXvalue#(2) = nb
Resp.Count = Resp.Count + 3
’*****
’* Further calculation after ’Signed.Samples’ determination
’*****
’***** 3 bytes <y_resolution> = <mantissa_high><mantissa_low><exponent>
’***** <mantissa> = <mantissa_high> * 256 + <mantissa_low>
’***** <y_resolution> = <sign><mantissa> E <sign><exponent>
’***** Example: +123E-4 = 123 / 10000 = 0.0123
FOR i = 0 TO 2
SumCheck1% = (SumCheck1% + ASC(MID$(Resp$,Resp.Count+i,1))) MOD 256
NEXT i
nb = ASC(MID$(Resp$, Resp.Count, 1))
IF nb >= 128 THEN
nb = - (256 - nb) * 256’Negative value
nb = nb + ASC(MID$(Resp$, Resp.Count + 1, 1))
ELSE
’
’ Page 3.53
’
nb = nb * 256’Positive value
nb = nb + ASC(MID$(Resp$, Resp.Count + 1, 1))
END IF
exponent(3) = ASC(MID$(Resp$, Resp.Count + 2, 1))
YXvalue#(3) = nb
Resp.Count = Resp.Count + 3
’*****
’* Further calculation after ’Signed.Samples’ determination
’*****
’***** 3 bytes <x_resolution> = <mantissa_high><mantissa_low><exponent>
’***** <mantissa> = <mantissa_high> * 256 + <mantissa_low>
’***** <x_resolution> = <sign><mantissa> E <sign><exponent>
’***** Example: +123E-4 = 123 / 10000 = 0.0123
FOR i = 0 TO 2
SumCheck1% = (SumCheck1% + ASC(MID$(Resp$,Resp.Count+i,1))) MOD 256
NEXT i
nb = ASC(MID$(Resp$, Resp.Count, 1))
IF nb >= 128 THEN
nb = - (256 - nb) * 256’Negative value
nb = nb + ASC(MID$(Resp$, Resp.Count + 1, 1))
ELSE
nb = nb * 256’Positive value
nb = nb + ASC(MID$(Resp$, Resp.Count + 1, 1))
END IF
exponent(4) = ASC(MID$(Resp$, Resp.Count + 2, 1))
YXvalue#(4) = nb
Resp.Count = Resp.Count + 3
’*****
’* Further calculation after ’Signed.Samples’ determination
’*****
’***** 3 bytes <y_at_0> = <mantissa_high><mantissa_low><exponent>
’***** <mantissa> = <mantissa_high> * 256 + <mantissa_low>
’***** <y_at_0> = <sign><mantissa> E <sign><exponent>
’***** Example: +123E-4 = 123 / 10000 = 0.0123
FOR i = 0 TO 2
SumCheck1% = (SumCheck1% + ASC(MID$(Resp$,Resp.Count+i,1))) MOD 256
NEXT i
nb = ASC(MID$(Resp$, Resp.Count, 1))
IF nb >= 128 THEN
nb = - (256 - nb) * 256’Negative value
nb = nb + ASC(MID$(Resp$, Resp.Count + 1, 1))
ELSE
nb = nb * 256’Positive value
nb = nb + ASC(MID$(Resp$, Resp.Count + 1, 1))
END IF
exponent(5) = ASC(MID$(Resp$, Resp.Count + 2, 1))
YXvalue#(5) = nb
Resp.Count = Resp.Count + 3
’*****
’* Further calculation after ’Signed.Samples’ determination
’*****
’***** 3 bytes <x_at_0> = <mantissa_high><mantissa_low><exponent>
’***** <mantissa> = <mantissa_high> * 256 + <mantissa_low>
’***** <x_at_0> = <sign><mantissa> E <sign><exponent>
’***** Example: +123E-4 = 123 / 10000 = 0.0123
’
’ Page 3.54
’
FOR i = 0 TO 2
SumCheck1% = (SumCheck1% + ASC(MID$(Resp$,Resp.Count+i,1))) MOD 256
NEXT i
nb = ASC(MID$(Resp$, Resp.Count, 1))
IF nb >= 128 THEN
nb = - (256 - nb) * 256’Negative value
nb = nb + ASC(MID$(Resp$, Resp.Count + 1, 1))
ELSE
’ Page 3.56
’
Resp.Count = Resp.Count + 1
PRINT "Signed.Samples = ";
PRINT #4, "Signed.Samples = ";
IF Signed.Samples = 1 THEN
PRINT "TRUE "; : PRINT #4, "TRUE"
ELSE
PRINT "FALSE "; : PRINT #4, "FALSE"
END IF
PRINT "Sample.Format = ";
PRINT #4, "Sample.Format = ";
IF MinMax.Samples = 0 THEN
PRINT "Single"
PRINT #4, "Single"
ELSEIF MinMax.Samples = 1 THEN
PRINT "Min/Max"
PRINT #4, "Min/Max"
ELSEIF MinMax.Samples = 2 THEN
PRINT "Min/Max/Ave"
PRINT #4, "Min/Max/Ave"
ELSE
PRINT "Unknown: "; OCT$(Sample.Format); " octal"
PRINT #4, "Unknown: "; OCT$(Sample.Format); " octal"
END IF
PRINT "Number of Sample.Bytes ="; Sample.Bytes
PRINT #4, "Number of Sample.Bytes ="; Sample.Bytes
’*****
’* Further calculation now that ’Signed.Samples’ is determined
’*****
FOR j = 1 TO 2
IF expscale(j) > 127 THEN ’Negative exponent
expscale(j) = 256 - expscale(j)
FOR i = 1 TO expscale(j)
YXscale#(j) = YXscale#(j) / 10
NEXT i
ELSE ’Positive exponent
FOR i = 1 TO expscale(j)
YXscale#(j) = YXscale#(j) * 10
NEXT i
END IF
NEXT j
Y.Scale = YXscale#(1)
X.Scale = YXscale#(2)
PRINT "<y_scale> ="; Y.Scale,
PRINT " <x_scale> ="; X.Scale
’
FOR j = 1 TO 6
IF exponent(j) > 127 THEN ’Negative exponent
exponent(j) = 256 - exponent(j)
FOR i = 1 TO exponent(j)
YXvalue#(j) = YXvalue#(j) / 10
NEXT i
ELSE ’Positive exponent
FOR i = 1 TO exponent(j)
YXvalue#(j) = YXvalue#(j) * 10
NEXT i
Sample.Byte = - (256 - Sample.Byte)
END IF
Overload& = Sample.Byte
FOR i = 2 TO Sample.Bytes
Sample.Byte = ASC(MID$(Resp$, Resp.Count + i - 1, 1))
SumCheck2% = (SumCheck2% + Sample.Byte) MOD 256
Overload& = Overload& * 256 + Sample.Byte
NEXT i
IF (Signed.Samples = 0) OR (Overload& < CLimit) THEN
Overload.Value = Overload& * Y.Resol’Positive value
ELSE’Negative value
Overload.Value = - ((CMaxim - Overload&) * Y.Resol)
END IF
Resp.Count = Resp.Count + Sample.Bytes
PRINT "Overload sample value ="; Overload&; Overload.Value
PRINT #4, "Overload sample value ="; Overload&; Overload.Value
’
’***** <Sample.Bytes> bytes <underload> value
Sample.Byte = ASC(MID$(Resp$, Resp.Count, 1))
SumCheck2% = SumCheck2% + Sample.Byte
IF (Signed.Samples = 1) AND (Sample.Byte >= 128) THEN
Sample.Byte = - (256 - Sample.Byte)
END IF
Underload& = Sample.Byte
FOR i = 2 TO Sample.Bytes
Sample.Byte = ASC(MID$(Resp$, Resp.Count + i - 1, 1))
SumCheck2% = (SumCheck2% + Sample.Byte) MOD 256
Underload& = Underload& * 256 + Sample.Byte
NEXT i
IF (Signed.Samples = 0) OR (Underload& < CLimit) THEN
Underload.Value = Underload& * Y.Resol’Positive value
ELSE’Negative value
Underload.Value = - ((CMaxim - Underload&) * Y.Resol)
END IF
Resp.Count = Resp.Count + Sample.Bytes
PRINT "Underload sample value ="; Underload&; Underload.Value
PRINT #4, "Underload sample value ="; Underload&; Underload.Value
’
’ Page 3.58
’
’***** <Sample.Bytes> bytes <invalid> value
Sample.Byte = ASC(MID$(Resp$, Resp.Count, 1))
SumCheck2% = SumCheck2% + Sample.Byte
IF (Signed.Samples = 1) AND (Sample.Byte >= 128) THEN
Sample.Byte = - (256 - Sample.Byte)
END IF
Invalid& = Sample.Byte
FOR i = 2 TO Sample.Bytes
Sample.Byte = ASC(MID$(Resp$, Resp.Count + i - 1, 1))
SumCheck2% = (SumCheck2% + Sample.Byte) MOD 256
Invalid& = Invalid& * 256 + Sample.Byte
NEXT i
IF (Signed.Samples = 0) OR (Invalid& < CLimit) THEN
Invalid.Value = Invalid& * Y.Resol’Positive value
ELSE’Negative value
Invalid.Value = - ((CMaxim - Invalid&) * Y.Resol)
END IF
Resp.Count = Resp.Count + Sample.Bytes
PRINT "Invalid sample value ="; Invalid&; Invalid.Value
PRINT #4, "Invalid sample value ="; Invalid&; Invalid.Value
’
’***** 2 bytes <nbr_of_samples>
Sample.Byte = ASC(MID$(Resp$, Resp.Count, 1))
SumCheck2% = (SumCheck2% + Sample.Byte) MOD 256
Nbr.Of.Samples = Sample.Byte
Sample.Byte = ASC(MID$(Resp$, Resp.Count + 1, 1))
SumCheck2% = (SumCheck2% + Sample.Byte) MOD 256
Nbr.Of.Samples = Nbr.Of.Samples * 256 + Sample.Byte
IF MinMax.Samples = 1 THEN ’Min/Max pair of samples
Nbr.Of.Samples = Nbr.Of.Samples * 2
END IF
IF MinMax.Samples = 2 THEN ’Min/Max/Ave samples
Nbr.Of.Samples = Nbr.Of.Samples * 3
END IF
Resp.Count = Resp.Count + 2
PRINT "Number of samples ="; Nbr.Of.Samples
PRINT #4, "Number of samples ="; Nbr.Of.Samples
’
’***** <Sample.Bytes> bytes <sample_value>’s
’
DIM Sample.Value(Nbr.Of.Samples) AS LONG
FOR i = 1 TO Nbr.Of.Samples’Sample loop
Sample.Byte = ASC(MID$(Resp$, Resp.Count, 1))
SumCheck2% = (SumCheck2% + Sample.Byte) MOD 256
IF (Signed.Samples = 1) AND (Sample.Byte >= 128) THEN
Sample.Value&(i) = Sample.Value&(i) * 256 + Sample.Byte
NEXT j
END IF
Resp.Count = Resp.Count + Sample.Bytes
’
’ Page 3.59
’
IF i=1 OR i=2 OR i = Nbr.Of.Samples-1 OR i = Nbr.Of.Samples THEN
IF (Signed.Samples = 0) OR (Sample.Value&(i) < CLimit) THEN
Ampl.Value = Sample.Value&(i) * Y.Resol’Positive value
ELSE’Negative value
Ampl.Value = - ((CMaxim - Sample.Value&(i)) * Y.Resol)
END IF
PRINT "Sample"; i; "="; Sample.Value&(i); Ampl.Value
END IF
PRINT #4, "Sample"; i; "="; Sample.Value&(i); Ampl.Value
NEXT i
’
’***** 1 byte <check_sum>
Check.Sum% = ASC(MID$(Resp$, Resp.Count, 1))
IF Check.Sum% <> (SumCheck2% MOD 256) GOTO Wave2.Error
Resp.Count = Resp.Count + 1
PRINT "<check_sum> ="; Check.Sum%; " & ";
PRINT "SumCheck2 MOD 256 ="; SumCheck2% MOD 256
PRINT #4, "<check_sum> ="; Check.Sum%; " & ";
PRINT #4, "SumCheck2 MOD 256 ="; SumCheck2% MOD 256
’
’***** 1 byte CR
C.R = ASC(MID$(Resp$, Resp.Count, 1))
IF C.R <> 13 GOTO Wave2.Error
Resp.Count = Resp.Count + 1
CLOSE #4: RETURN
Wave2.Error:
PRINT "Waveform sample error at byte :"; Resp.Count
PRINT "Waveform decimal byte value ="; ASC(MID$(Resp$,Resp.Count,1))
PRINT "SumCheck so far (MOD 256) ="; SumCheck2% MOD 256
CLOSE: END
’
’ Page 3.60
Create.CSV:
’
’*****
’***** Convert the total Response string to file Wave.CSV
’***** as input file for Excel (spreadsheet), for example.
’*****
’
OPEN "Wave.CSV" FOR OUTPUT AS #4
PRINT #4, "Title , ";
IF MID$(Query$, 4, 2) = "10" THEN
PRINT #4, "Voltage trace"
ELSEIF MID$(Query$, 4, 2) = "11" THEN
PRINT #4, "Record trace 1"
END IF
IF Trace.Result = 0 OR Trace.Result = 1 THEN
PRINT #4, "ID ,"; Trace.Result ’Acquisition trace
PRINT #4, "Type , "; "Acquisition trace"
ELSEIF Trace.Result = 2 THEN
PRINT #4, "ID ,"; 2’Record trace
PRINT #4, "Type , "; "Record trace"
END IF
PRINT #4, "Date , "; Month$+"/"+Day$+"/"+MID$(Year$,3,2)
PRINT #4, "Time , "; Hours$+":"Minutes$+":"+Seconds$
’
’***** X.Scale = time per division (over 10 divisions)
PRINT #4, "X Scale ,"; X.Scale
PRINT #4, "X At 0% ,"; X.Zero
PRINT #4, "X Resolution ,"; X.Resol
PRINT #4, "X Size ,"; Nbr.Of.Samples
PRINT #4, "X Unit , ";
IF X.Unit = 7 THEN PRINT #4, "s"
IF X.Unit = 10 THEN PRINT #4, "Hz"
PRINT #4, "X Label ,";
IF X.Unit = 7 THEN PRINT #4, X.Scale; "s/Div"
IF X.Unit = 10 THEN PRINT #4, X.Scale; "Hz/Div"
’
PRINT #4, "Y Scale ,"; Y.Scale
PRINT #4, "Y At 50% ,"; Y.Zero
PRINT #4, "Y Resolution ,"; Y.Resol
PRINT #4, "Y Size ,";
IF Sample.Bytes = 1 THEN’1-byte samples
PRINT #4, 256
END IF’Range = 256
IF Sample.Bytes = 2 THEN’2-byte samples
PRINT #4, 65536
END IF’Range = 256*256
PRINT #4, "Y Unit , ";
IF Y.Unit = 1 THEN PRINT #4, "V"
IF Y.Unit = 2 THEN PRINT #4, "A"
IF Y.Unit = 3 THEN PRINT #4, "Ohm"
IF Y.Unit > 3 OR Y.Unit = 0 THEN PRINT #4, "No print"
PRINT #4, "Y Label ,";
IF Y.Unit = 1 THEN PRINT #4, Y.Scale; "V/Div"
IF Y.Unit = 2 THEN PRINT #4, Y.Scale; "A/Div"
IF Y.Unit = 3 THEN PRINT #4, Y.Scale; "Ohm/Div"
IF Y.Unit > 3 OR Y.Unit = 0 THEN PRINT #4, Y.Scale; "No print"
PRINT #4,
’
’ Page 3.61
’
’***** Sample values x,y (time,amplitude)
Time.Value = X.Zero’Start at x-offset
MinMax.Flag = MinMax.Samples’Switch flag (2, 1, 0)
FOR i = 1 TO Nbr.Of.Samples
IF (Signed.Samples = 0) OR (Sample.Value&(i) < CLimit) THEN
’Positive value
Amplit.Value = Sample.Value&(i) * Y.Resol
ELSE
’Negative value
Amplit.Value = - ((CMaxim - Sample.Value&(i)) * Y.Resol)
END IF
IF MinMax.Samples = 2 THEN’Min/Max/Ave waveform
IF MinMax.Flag = 2 THEN
MinMax.Flag = MinMax.Flag - 1
PRINT #4, Time.Value; ","; Amplit.Value; ",";
ELSEIF MinMax.Flag = 1 THEN
MinMax.Flag = MinMax.Flag - 1
PRINT #4, Amplit.Value; ",";
ELSE
MinMax.Flag = 2
PRINT #4, Amplit.Value
Time.Value = Time.Value + X.Resol
END IF
END IF
IF MinMax.Samples = 1 THEN’Min/Max waveform
IF MinMax.Flag = 1 THEN
MinMax.Flag = 0
PRINT #4, Time.Value; ","; Amplit.Value; ",";
ELSE
MinMax.Flag = 1
PRINT #4, Amplit.Value
Time.Value = Time.Value + X.Resol
END IF
END IF
IF MinMax.Samples = 0 THEN’Single waveform
PRINT #4, Time.Value; ","; Amplit.Value
Time.Value = Time.Value + X.Resol
END IF
NEXT i
CLOSE #4: RETURN
’
’****************** End example program ******************
Page 3.62
=============================================================
READ DATE RD
<date> = string of the following format:
<year>,<month>,<day>
e.g. 2001,8,14
Example:
The following example program reads the date setting from
the instrument.
’ Page 3.63
’***************** Begin example program *****************
CLS
OPEN "COM1:1200,N,8,1,CS,DS,RB2048" FOR RANDOM AS #1
PRINT #1, "RD" ’Sends the READ DATE query.
GOSUB Acknowledge ’Input acknowledge from instrument.
INPUT #1, SMYear$, SMMonth$, SMDay$ ’Inputs the date string.
PRINT "Date "; SMYear$; "-"; SMMonth$; "-"; SMDay$
’Displays the date string.
END
’**************** Acknowledge subroutine ******************
’Use this subroutine after each command or query sent to the
’instrument. This routine inputs the acknowledge
’response from the instrument. If the response is non-zero,
’the previous command was not correct or was not correctly
’received by the instrument. Then an error message is
’displayed and the program is aborted.
Acknowledge:
INPUT #1, ACK ’Reads acknowledge from instrument.
IF ACK <> 0 THEN
PRINT "Error "; ACK; ": ";
SELECT CASE ACK
CASE 1
PRINT "Syntax Error"
CASE 2
PRINT "Execution Error"
CASE 3
PRINT "Synchronization Error"
CASE 4
PRINT "Communication Error"
CASE IS < 1
PRINT "Unknown Acknowledge"
CASE IS > 4
PRINT "Unknown Acknowledge"
END SELECT
PRINT "Program aborted."
END
END IF
RETURN
’****************** End example program ******************
Page 3.64
=============================================================
RESET INSTRUMENT RI
Resets the entire instrument, including the CPL interface.
The baud rate remains unchanged.
Command Syntax:
RI<cr>
Response Syntax:
<acknowledge><cr>
Note: Wait for at least 2 seconds after the
<acknowledge> reply has been received, to let
the instrument settle itself before you send the
next command.
Example:
The following example resets the instrument and waits for 2
seconds to let the instrument execute the reset and become
ready for next commands.
The instrument is queried for the identification data; this
data is input and displayed on the PC screen.
’ Page 3.65
’***************** Begin example program *****************
CLS ’Clears the PC screen.
OPEN "COM1:1200,N,8,1,CS,DS,RB2048" FOR RANDOM AS #1
PRINT #1, "RI" ’Sends the RESET INSTRUMENT command.
GOSUB Acknowledge ’Input acknowledge from instrument.
SLEEP 2 ’Delay (2 s) necessary after reset.
GOSUB ClearPort ’Clears pending data from port.
PRINT #1, "ID" ’Sends IDENTIFICATION query.
GOSUB Acknowledge ’Input acknowledge from instrument.
INPUT #1, IDENT$ ’Inputs the queried data.
PRINT IDENT$ ’Displays queried data.
CLOSE #1
END
’**************** Acknowledge subroutine ******************
’Use this subroutine after each command or query sent to the
’instrument. This routine inputs the acknowledge
’response from the instrument. If the response is non-zero,
’the previous command was not correct or was not correctly
’received by the instrument. Then an error message is
’displayed and the program is aborted.
Acknowledge:
INPUT #1, ACK ’Reads acknowledge from instrument.
IF ACK <> 0 THEN
PRINT "Error "; ACK; ": ";
SELECT CASE ACK
CASE 1
PRINT "Syntax Error"
CASE 2
PRINT "Execution Error"
CASE 3
PRINT "Synchronization Error"
CASE 4
PRINT "Communication Error"
CASE IS < 1
PRINT "Unknown Acknowledge"
CASE IS > 4
PRINT "Unknown Acknowledge"
END SELECT
PRINT "Program aborted."
END
END IF
RETURN
’******* Clears pending data from the RS232 port *********
ClearPort:
WHILE LOC(1) > 0
Dummy$ = INPUT$(1, #1)
WEND
RETURN
’****************** End example program ******************
Recalls an internally stored setup. This setup must have
been stored in the instrument manually or with the SS
(Save Setup) command.
The effect of the RS command is that the instrument setup
is recalled and the instrument forced to running state.
Command Syntax:
RS <setup_reg><cr>
where,
<setup_reg> =
1 to 10 : Screen/Setup memories
Response Syntax:
<acknowledge><cr>
Note: The new setup is active when you have received the
<acknowledge> response from the instrument.
Example:
The following example program saves the present setup in
setup memory 8. You are requested to change the present
settings. Then the original settings are recalled from
setup memory 8 and made the actual setting.
’ Page 3.67
’***************** Begin example program *****************
CLS ’Clears the PC screen.
OPEN "COM1:1200,N,8,1,CS,DS,RB2048" FOR RANDOM AS #1
PRINT #1, "SS 8" ’Sends SAVE SETUP command.
’Setup saved in setup memory 8.
GOSUB Acknowledge ’Input acknowledge from instrument
PRINT "The present setup data are stored in setup memory 8."
PRINT "The remainder of this program will restore these."
PRINT "To test if this works, change the present settings"
PRINT "and verify if the instrument returns to the original"
PRINT "settings after continuing the program."
PRINT
PRINT "Press any key on the PC keyboard to continue."
SLEEP
PRINT #1, "RS 8" ’Sends RECALL SETUP command.
’Setup recalled from register 8.
GOSUB Acknowledge ’Input acknowledge from instrument.
PRINT
PRINT "Original settings restored"
CLOSE #1
END
’
’ Page 3.68
’**************** Acknowledge subroutine ******************
’Use this subroutine after each command or query sent to the
’instrument. This routine inputs the acknowledge
’response from the instrument. If the response is non-zero,
’the previous command was not correct or was not correctly
’received by the instrument. Then an error message is
’displayed and the program is aborted.
Acknowledge:
INPUT #1, ACK ’Reads acknowledge from instrument.
IF ACK <> 0 THEN
PRINT "Error "; ACK; ": ";
SELECT CASE ACK
CASE 1
PRINT "Syntax Error"
CASE 2
PRINT "Execution Error"
CASE 3
PRINT "Synchronization Error"
CASE 4
PRINT "Communication Error"
CASE IS < 1
PRINT "Unknown Acknowledge"
CASE IS > 4
PRINT "Unknown Acknowledge"
END SELECT
PRINT "Program aborted."
END
END IF
RETURN
’****************** End example program ******************
Page 3.69
=============================================================
READ TIME RT
<time> = string of the following format:
<hours>,<minutes>,<seconds>
e.g. 15,4,43
Example:
The following example program reads the time setting from
the instrument.
’ Page 3.70
’***************** Begin example program *****************
OPEN "COM1:1200,N,8,1,CS,DS,RB2048" FOR RANDOM AS #1
PRINT #1,"RT" ’Sends the READ TIME query.
GOSUB Acknowledge ’Input acknowledge from instrument.
INPUT #1,SMhour$,SMmin$,SMsec$ ’Inputs the time strings.
PRINT "Time "; SMhour$;":";SMmin$;":";SMsec$
’Displays the time string.
END
’**************** Acknowledge subroutine ******************
’Use this subroutine after each command or query sent to the
’instrument. This routine inputs the acknowledge
’response from the instrument. If the response is non-zero,
’the previous command was not correct or was not correctly
’received by the instrument. Then an error message is
’displayed and the program is aborted.
Acknowledge:
INPUT #1, ACK ’Reads acknowledge from instrument.
IF ACK <> 0 THEN
PRINT "Error "; ACK; ": ";
SELECT CASE ACK
CASE 1
PRINT "Syntax Error"
CASE 2
PRINT "Execution Error"
CASE 3
PRINT "Synchronization Error"
CASE 4
PRINT "Communication Error"
CASE IS < 1
PRINT "Unknown Acknowledge"
CASE IS > 4
PRINT "Unknown Acknowledge"
END SELECT
PRINT "Program aborted."
END
END IF
RETURN
’****************** End example program ******************
Page 3.71
=============================================================
SWITCH ON SO
Queries the error status of the instrument.
This is a 16-bit word, presented as an integer value,
where each bit represents the Boolean value of a related
error event. After the reply or after a RI (Reset
Instrument) command, the value is reset to zero.
A complete description of the status word is given in
Appendix B.
Command Syntax:
ST<cr>
Response Syntax:
<acknowledge><cr>[<status>
where,
<status> = integer value 0 to 32767
Example:
The following example program sends a wrong command to the
instrument to test the Acknowledge subroutine and to check
the status returned from the ST query.
The acknowledge subroutine contains a GOSUB Status.display
to input the status data from the instrument when the
acknowledge response is non-zero (ACK <> 0).
’ Page 3.74
’***************** Begin example program *****************
CLS ’Clears the PC screen.
OPEN "COM1:1200,N,8,1,CS,DS,RB2048" FOR RANDOM AS #1
PRINT #1, "PC 12345" ’Sends a baud rate value that is
’ out of range for the instrument.
GOSUB Acknowledge.Status ’Input acknowledge from instrument
’and the status value if the
’acknowledge value is non-zero.
END
’************* Acknowledge + Status subroutine ***********
’This subroutine inputs the acknowledge value from the
’instrument. If the acknowledge value is non-zero,
’the ST query is used to get further status information from
’the instrument with respect to the error.
’In case of an error the program is aborted.
Acknowledge.Status:
INPUT #1, ACK ’Reads acknowledge from instrument.
IF ACK <> 0 THEN
PRINT "Error "; ACK; ": ";
SELECT CASE ACK
CASE 1
PRINT "Syntax Error"
CASE 2
PRINT "Execution Error"
CASE 3
PRINT "Synchronization Error"
CASE 4
PRINT "Communication Error"
CASE IS < 1
PRINT "Unknown Acknowledge"
CASE IS > 4
PRINT "Unknown Acknowledge"
END SELECT
GOSUB Status.display ’Further specifies the error.
PRINT "Program aborted."
END
END IF
RETURN
’
’ Page 3.75
’************** Displays instrument status *****************
’This subroutine gives you further information if the
’acknowledge reply from the instrument is non-zero.
Status.display:
PRINT #1, "ST" ’Sends the STATUS query.
GOSUB Acknowledge.Status ’Inputs acknowledge from instrument.
INPUT #1, STAT ’Inputs status value.
PRINT "Status " + STR$(STAT) + ": ";
IF STAT = 0 THEN PRINT "No error"
IF (STAT AND 1) = 1 THEN PRINT "Illegal Command"
IF (STAT AND 2) = 2 THEN
PRINT "Data format of parameter is wrong"
END IF
IF (STAT AND 4) = 4 THEN PRINT "Parameter out of range"
IF (STAT AND 8) = 8 THEN
PRINT "Invalid command in this CPL interface"
END IF
IF (STAT AND 16) = 16 THEN PRINT "Command not implemented"
IF (STAT AND 32) = 32 THEN
PRINT "Invalid number of parameters"
END IF
IF (STAT AND 64) = 64 THEN
PRINT "Wrong number of data bits"
END IF
IF (STAT AND 512) = 512 THEN
PRINT "Conflicting instrument settings"
END IF
IF (STAT AND 16384) = 16384 THEN
PRINT "Checksum error"
END IF
RETURN
’****************** End example program ******************
Page 3.76
=============================================================
TRIGGER ACQUISITION TA
Triggers an acquisition. This command acts as a
hardware trigger to start a new acquisition.
In SINGLE shot acquisition mode the trigger system
must have been armed with the AT (Arm Trigger)
command.
Command Syntax:
TA<cr>
Response Syntax:
<acknowledge><cr>
Example:
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.