The Model 35080M kVp Divider is an alternative to high-voltage divider tanks. It provides a safe, noninvasive, fast measurement of kVp without endangering personnel and/or equipment. Its use is primarily
intended for X-ray machine service and calibration.
High-speed precision logarithmic circuitry allows for wide variations in mA and distance without a sacrifice
of accuracy or response speed. The Model 35080M kVp Divider provides an analog output voltage
proportional to the instantaneous voltage applied to the X-ray tube. When the kVp divider is connected to
a storage oscilloscope, the kV waveform may be viewed directly. If a digital display of the kVp value is
desired, then the Model 35080M kVp Divider may be connected to the Model 35050AT Dosimeter/kVp
Readout.
One Wide Range Filter Pack (37617) allows high-accuracy measurements from 50 to 150kV. The kVp
divider is placed on the X-ray table with the filter pack in place. A storage oscilloscope is connected to
the kVp divider. The X-ray beam is collimated to the area of the filter pack. An exposure is made and the
kV waveform is displayed on the storage oscilloscope. At no time are the high-voltage supply cables
disturbed.
The Model 35080M kVp Divider is highly portable -- a contrast to bulky, heavy high-voltage divider tanks.
The instrument is accurate to ± 2%, exclusive of linearity and filtration effects.
The user should note, however, that as with an invasive divider, the accuracy of readings will depend on
the accuracy of the oscilloscope or readout device used in the measurements.
1.2 Manual Addenda
Any improvements or changes concerning the instrument or manual will be explained in an addenda
included with the manual. Be sure to note these changes and incorporate them into the manual.
1.3 Safety Symbol and Terms
The following symbols and terms are found on the instrument or used in this manual.
Indicates that the user should refer to the operating instructions located in the manual.
WARNING
Explains dangers that might result in personal injury or death. Always read the
associated information very carefully before performing the indicated procedure.
Explains hazards that could damage the instrument. Such damage may
N
invalidate the warranty.
1-1
35080M
Operators Manual
The equipment described in this manual is intended to be used for the detection and measurement of
ionizing radiation. It should be used only by persons who have been trained in the proper interpretation of
its readings and the appropriate safety procedures to be followed in the presence of radiation.
Although the equipment described in this manual is designed and manufactured in compliance with all
applicable safety standards, certain hazards are inherent in the use of electronic and radiometric
equipment.
WARNINGS and CAUTIONS are presented throughout this document to alert the user to potentially
hazardous situations. A WARNING is a precautionary message preceding an operation that has the
potential to cause personal injury or death. A CAUTION is a precautionary message preceding an
operation that has the potential to cause permanent damage to the equipment and/or loss of data.
Failure to comply with WARNINGS and CAUTIONS is at the user’s own risk and is sufficient cause to
terminate the warranty agreement between Fluke Biomedical and the customer.
Adequate warnings are included in this manual and on the product itself to cover hazards that may be
encountered in normal use and servicing of this equipment. No other procedures are warranted by Fluke
Biomedical. It shall be the owner’s or user’s responsibility to see to it that the procedures described here
are meticulously followed, and especially that WARNINGS and CAUTIONS are heeded. Failure on the
part of the owner or user in any way to follow the prescribed procedures shall absolve Fluke Biomedical
and its agents from any resulting liability.
Indicated battery and other operational tests must be performed prior to each use to assure that the
instrument is functioning properly. If applicable, failure to conduct periodic performance tests in
accordance with ANSI N323-1978 (R1983) Radiation Protection Instrumentation Test and Calibration, paragraphs 4.6 and 5.4, and to keep records thereof in accordance with paragraph 4.5 of the
same standard, could result in erroneous readings or potential danger. ANSI N323-1978 becomes, by
this reference, a part of this operating procedure.
1-2
35080M
Operators Manual
1.4 Specifications
Range
50-150 kVp with the Wide Range Filter Pack (37617)
Accuracy
± 2% of reading, exclusive of linearity, filtration, and gain effects
Calibration
Internally generated signal provides a calibration check
Permitted Tube Current Range
1 mA to 3000 mA, dependent on waveform and distance
Response Time
150 microseconds 10-90%
Environment
Operating temperature 15° to 35°C; storage temperature -35° to 50°C; relative humidity to 90%
Orientation
The long axis of the Model 35080M kVp Divider is oriented perpendicular to the axis of the X-ray tube to
eliminate heel effect.
Table 1-1. Permitted Tube Current Settings (less than ± 1 kV effect)
Minimum Tube Current (12" Focal
Spot to kVp divider)
Maximum Tube Current (22" Focal
Spot to kVp divider)
Factory calibration performed at 200 mA, 2.5 mm Al inherent filtration, 22 inches focal
spot to kVp divider. Six-point calibration check.
Single-Phase
Half - Wave
1 mA 2 mA 4 mA
750 mA 1500 mA 3000 mA
Single-Phase
Full - Wave
Three-Phase
Batteries
One Eveready 9V alkaline type 522; life approximately 50 hours
Dimensions and Weight
25 mm high x 130 mm wide x 65 mm deep (1 in x 5-1/8 in x 2-9/16 in)
net weight 180 g (6.3 oz), including battery
Standard Filter Pack
One Wide Range Filter Pack* (37617) is provided for measurement from 50 to 150 kVp.
1-3
General Information
Specifications
1
Table 1-2. The Wide Range Filter Pack (37617) Specifications
Range: 50 to 150 kVp
Sensitivity: 0.1 V/10 kV
Cal: 0.400 V
Zero: 40 kV
Cal: 80 kV
Accuracy: ± 2% exclusive of linearity and filtration
effects
Linearity and filtration effects can be corrected using
curves from Section 2 of this instruction manual.
Optional Filter Packs
A K-edge Mammo Filter Pack (37355)* and a Linear Mammo Filter Pack (37351) used for precision
measurements from 22 to 40 kVp, a CT Filter Pack (33551) used for CT system kVp measurements from
70 to 140 kVp, a Mobile Filter Pack Plus (37946) used for measurements on X-ray systems with kV
waveforms above 1 kHz with substantial ripple, and a Low Range Filter Pack (38237) used for
measurements from 30 kVp to 90 kVp where the kVp range or sensitivity of other filter packs is not
adequate.
* Patent numbers 4,843,619 and 4,916,727
1-3
35080M
Operators Manual
1.5 Nomenclature
1
2
3
4
5
BAT
STATUS
kVp
DIVIDER
RUN
CAL/BAT CHK
1) Filter Pack Input Jacks
2) Battery Status LED
3) Power ON / OFF Button
4) BNC output connector
5) Function Switch
1-4
Operation
Battery Check
2
Section 2
Operation
Prior to taking measurements with the Model 35080M kVp Divider,
it is suggested that the battery and internal calibration be checked.
NOTE
2.1 Turning the Unit ON and OFF
The 35050M has a miniature power button (located near the BNC output connector) that becomes
depressed (ON) while a BNC cable is attached and released (OFF) when the cable is removed. When
the unit is not in use, remove the cable to preserve battery life. Do not cover the BNC connector with a
dust cap as this will turn on power and consume battery life as well.
Do not attach any TRIAX type cable into the BNC
connector. The BNC connector can only
accommodate a 2-pin COAX type cable. Attempting
to use a TRIAX cable could result in shock hazard or
damage to the unit.
WARNING
2.2 Battery Check
Set the function switch to the “CAL / BAT CHK” position. Depress the power button, either by hand or by
attaching a BNC cable to the output connector. The LED will indicate whether the battery is usable
(green, or above 7.0 volts) or needs to be replaced (red, or below 7.0 volts).
The battery must be replaced if the voltage is less than 6.5 volts as the instrument will not function
properly when the voltage supply is below this minimum level. To preserve battery life, do not run the unit
in the CAL mode for long periods of time.
2.3 Internal Calibration Check
Connect the output of the Model 35080M kVp Divider to a DVM with ± 0.1% accuracy. Remove any filter
pack that may be inserted in the kVp Divider. Turn the function switch to the “CAL” position. After a short
delay, a voltage of 0.750 ± 1% will appear. This reading verifies the stability of the kVp Divider ratio
electronics. A 1% error in the kVp Divider gain will cause a maximum kV reading error of 0.75% of
reading. Refer to Section 9.5 for detailed information on calibration.
2.4 Filter Pack Selection
One Wide Range Filter Pack (37617) is provided as standard for the Model 35080M kVp Divider. See
Sections 4 through 8 of this manual for use of optional filter packs available from Fluke Biomedical,
Radiation Management Services.
2-1
35080M
Operators Manual
2.5 Oscilloscope Zero and Offsets
It is assumed that the operator is familiar with the use of a
storage oscilloscope. The following examples are for
oscilloscopes with 8 major divisions for vertical scales and
sensitivity selections of 0.1 V/div and 0.2 V/div. The selection
of time scales or triggering are not discussed in this manual.
Figures 2-1A & 2-1B show suggested oscilloscope settings for the 50-150 kV range using the Wide
Range Filter Pack (37617).
1. Insert the Wide Range Filter Pack (37617) into the Model 35080M kVp Divider and connect the kVp
divider to the oscilloscope.
2. Set the oscilloscope gain to 0.1 V/div, DC Coupled. This setting corresponds to a 10 kV/div
sensitivity.
3. If making kV measurements from 50 to 120 kV, set the oscilloscope coupling from DC to GND and
adjust the vertical position control so that the oscilloscope trace coincides with the bottom horizontal
line on the screen, which corresponds to 40 kV. Then set the oscilloscope coupling back to DC and
the kVp Divider function switch to the ‘‘CAL’’ position. Verify that after a delay of several seconds
the trace on the oscilloscope coincides with middle horizontal line on the screen, which corresponds
to 80 kV. When switching to the ‘‘RUN’’ position, a slight offset may be noted on the oscilloscope.
Do NOT readjust the oscilloscope trace to compensate for this offset. Refer to Figure 2-1A.
4. If making kV measurements from 120 to 150 kV, set the function switch of the kVp divider to the
‘‘CAL’’ position and adjust the oscilloscope vertical position control so that the oscilloscope trace
coincides with the bottom line on the screen, which corresponds to 80 kV. Refer to Figure 2-1B.
NOTE
NOTE
If the accuracy of the Model 35080M kVp Divider is in
question, the output voltage may be checked with a ± 0.1%
DVM. The proper CAL voltage is on the filter pack label. The
allowed tolerance is ± 1%. A 1% error will cause at most a
0.75% error in the kV reading.
2-2
Operation
1A)
120 kV
110 kV
100 kV
90 kV
80 kV
70 kV
60 kV
50 kV 90 kV
40 kV ZERO at 40 80 kV CAL at 80
Sensitivity 10 kV/div
Scope setting 0.1 V/div
CAL at 80
1B)
160
kV
150
kV
140
kV
130
kV
120
kV
110
kV
100
kV
Sensitivity 10 kV/div
Scope setting 0.1 V/div
Oscilloscope Zero and Offsets
2
Figure 2-1A and 1B. Suggested Oscilloscope Settings for Use with the Wide Range Filter Pack
(37617)
As an alternative to reading directly in kV, the output of the Model 35080M kVp Divider may be read
directly in volts. The formula below is used to calculate the kV:
kV = Model 35080M voltage x 100 + ZERO kV
Example: A peak voltage of 0.300 volts is read. The ZERO kV of the Wide Range Filter Pack (37617) is
40 kV. Therefore:
kV = 0.300 x 100 + 40
= 30 + 40
= 70 kV
2.5 Geometry of Setup
The filter pack is aligned on the beam center, with the long axis of the Model 35080M kVp Divider
perpendicular to the X-ray tube axis, and the beam normal to the tabletop (see Figure 2-2). The X-ray
beam must be large enough to cover the active area of the filter pack. Distance is not critical, but a 22inch distance from focal spot to kVp divider is suggested.
The kVp divider must be inverted for under-table X-ray head evaluation (refer to Figure 2-3). Again, the
filter pack should be on beam center, perpendicular to the axis of the X-ray tube and the beam normal to
the table. The beam area must be large enough to cover the active area of the filter pack.
2-3
35080M
Operators Manual
In positioning the kVp divider, the long axis of the instrument should be perpendicular to the X-ray tube
axis to eliminate “heel effect.” The beam must strike the filter pack side of the kVp divider; the unit is
designed to be turned over for under-table X-ray heads. In setting the X-ray generator, 200 mA at 22
inches is recommended, but the limits of operation allow higher or lower settings and distances. Refer to
Section 1-4, Specifications, for low current limits.
After correctly positioning the kVp Divider, connect the oscilloscope, place the kVp Divider in the “RUN”
mode and make an exposure.
X-RAY TABLE
TUBE AXIS
ABOVE TABLE
X-RAY TUBE
LONG AXIS OF 35080M
TUBE AXIS
X-RAY TABLE
UNDER TABLE X-RAY TUBE
LONG AXIS OF 35080M
Figure 2-2. Geometry of Setup for Above-Table Figure 2-3. Geometry of Setup for Below-Table
X-ray Tube X-ray Tube
2-4
Operation
Filtration Effect Corrections
2
2.6 Filtration Effect Corrections
Corrections for filtration are made with the correction curves in Figure 2-4. Factory calibration of the
Model 35080M kVp Divider is performed with 2.5 mm Al total filtration, so any other amount of filtration in
the beam which is greater than or less than 2.5 mm Al will require some correction of the apparent
reading.
Example: Suppose the kVp Divider reads 90 kV, using the Wide Range Filter Pack (37617). The X-ray
machine has 5.0 mm Al total filtration. The filtration effect curve for this filter pack indicates a filtration
effect of 0.5 kV per mm Al at 90 kVp. To determine the actual reading, the filtration in the X-ray machine
is subtracted from the standard calibration filtration (2.5 mm Al) and the resulting number is multiplied by
the filtration effect: (2.5-5.0) x 0.5 = -1.25 kV. The resulting number, -1.25 kV, is the correction which is
added to the apparent reading. Therefore, the actual reading is 90 kV + (-1.25 kV) = 88.75 kV.
kVp Per mm
Filtration
Effect
Model 35080M Reading
Figure 2-4. The Wide Range Filter Pack (37617) Filtration Effect Correction Curve
2-5
35080M
y
p
Operators Manual
2.7 Linearity Corrections
Corrections for linearity in using the Wide Range Filter Pack (37617) are made by taking the apparent
output of the Model 35080M and adding (or subtracting) the value in kV shown on the following curve.
For the greatest accuracy possible, readings should first be corrected for filtration before linearity
corrections are made. For most applications, however, linearity corrections alone may be adequate,
especially if the difference in total filtration from the calibration standard of 2.5 mm Al is small. See
Section 2.6 for information about filtration correction.
Wide Range Filter Pack (37617)
Linearit
Correction Curve
kVp Correction
Model 35080M Reading kV
Figure2-5. The Wide Range Filter Pack (37617) Linearity Correction Curve. The kVp Correction Value
read from the vertical scale should be added to the Model 35080M kVp Divider reading to
produce the corrected kVp result.
2-6
Theory of Operation
Introduction
3
Section 3
Theory of Operation
3.1 Introduction
When used with the Wide Range Filter Pack (37617) the Model 35080M kVp Divider produces an output
voltage that closely follows the voltage applied to the X-ray tube. It does this with a 10-90% response
time of 150 microseconds, so the waveform can be used not only for determining kVp, but also for making
precise time determinations and for diagnosing generator malfunctions.
3.2 Measurement Theory
In this measurement scheme, two X-ray detectors are used (refer to Figures 3-1 and 3-2). The X-ray
beam passes through a relatively heavy filter prior to striking detector A, and through a lighter filter prior to
striking detector B. The ratio of the currents (I/I) is transformed to a voltage by a custom-designed ratio
module.
Typically, this ratio voltage increases as the voltage applied to the X-ray tube increases. The ratio
voltage is virtually independent of the current applied to the X-ray tube and the distance to the X-ray tube.
3.3 Wide Range Filter Pack (37617)
The Wide Range Filter Pack (37617) is a standard filter pack supplied with the Model 35080M kVp
Divider. This filter pack gives linear performance over a wide 50-150 kVp range. Detailed characteristics
of the filter pack are given in Section 1, General Information, and Section 2, Operating Instructions, of this
instruction manual.
Figure 3-1. Simplified Functional Block Diagram of the Model 35080M kVp Divider
3-1
35080M
Operators Manual
1.1
1
0.9
0.8
0.7
0.6
0.5
0.4
0.3
Model 35080M Output in Volts
0.2
0.1
0
405060708090100110120130140150
kV
Figure 3-2 Output Ratio Plotted as a Function of kV
VoltageWaveForm
Single-Phase
Model 35080M Wave
Form Single-Phase
Voltage WaveForm
Three-Phase
Model 35080M WaveForm
Output Three-Phase
Figure 3-3. Output of the Model 35080M kVp Divider Compared to Voltage Applied to X-ray Tube
3-2
Theory of Operation
Output
3
3.4 Output
In the Model 35080M kVp Divider, the intensity ratio is taken with a high-speed logarithmic ratio module.
Thus, the output of the Model 35080M kVp Divider follows the kVp waveform applied to the X-ray tube
within the linear range of the filter pack. The waveform differs from the waveform output of a resistive
divider in that information below the kV baseline of the filter is not displayed (see Figure 3-3).
3.5 Intensity Cutoff
An examination of the Model 35080M kVp Divider waveforms in Figure 3-3 shows that as the potential of
the X-ray beam waveform increases for the first time, the Model 35080M kVp Divider follows the signal as
soon as it comes into the linear range of the filter pack. However, as the X-ray signal decays, the signal
presented to the oscilloscope will suddenly switch to the filter pack’s baseline. This effect is caused by a
special circuit that prevents the presentation of a distorted signal to the oscilloscope. This circuit
automatically resets about 10 seconds after an exposure is made.
3.6 Internal Calibration
In order to assure that the ratio circuit in the Model 35080M kVp Divider is stable, a “CAL” position is
provided on the function switch. When the Model 35080M kVp Divider is switched into the “CAL” position,
a signal ratio of 0.25 is electronically switched into the ratio circuit. This produces an output of 0.75 volts
with no filter pack inserted. This value should be within ± 1% of 0.75 volts. When the Wide Range Filter
Pack (37617) is inserted, a CAL voltage of 0.40 volts ± 1% is produced at the Model 35080M kVp Divider
output. The CAL voltage is inscribed on each of the other filter packs. The value in kV is also given and
this may be used to check the oscilloscope calibration.
NOTE
If the Model 35080M kVp Divider is exposed to
an X-ray beam while the function switch is left
in the “CAL” position, the resulting scope
presentation will be badly distorted.
3-3
Loading...
+ 44 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.