Fluke 3458-HFL, 8508A User Manual

Page 1
Multimeter Selection Guide Fluke 8508A & Agilent 3458/HFL
When comparing two products features and specification using sales information it is often difficult to determine which product will meet the needs of your specific application. This is mostly because each supplier will define specifications using alternate methods. This Selection Guide will help you compare the specifications and features of the Fluke 8508A Reference multimeter with the nearest equivalent digital multimeter the Agilent 3458/HFL. Both the 8508A and the special 3458/HFL multimeters are available through the Fluke sales channel.
The following pages summarize & segment the features of each meter into the following three categories.
Accuracy
Functionality & Versatility
Ease of Use
The 'Comments' column found in the following table provides a brief explanation against each feature and its relevance to the intended metrology applications.
Further reference information:
1) Agilent 3458A/HFL Technical Data - 1275558 D-ENG-N
2) Fluke 8508A Extended Specifications - 1887212 D-ENG-N rev A
3) 1281Operators Handbook – 850090 issue 6
TECHNICAL INFORMATION
Page 2
DC Voltage
Calibration Interval 365 day, ±1°C of last TCAL
±(ppm of setting + ppm of range), relative to calibration standards used.
Range 1281 HP3458/HFL 8508A
120.00000mV - 4 + 3 -
199.999999mV 6 + 0.5 - 2.7 + 0.5
1.20000000V - 3 + 0.3 -
1.99999999V 3 + 0.2 - 2.7 + 0.2
12.000000V - 3 + 0.05 -
19.9999999V 3 + 0.2 - 2.7 + 0.2
120.00000V - 5 + 0.3 -
199.999999V 6 + 0.1 - 4.0 + 0.2
1050.0000V 6 + 0.1 5 + 0.1 4.0 + 0.5
Page 3
DC Current
Calibration Interval 365 day, ±1°C of last TCAL
±(ppm of setting + ppm of range), relative to calibration standards used.
Range 1281 HP3458/HFL 8508A
120.000nA - 30 + 400 -
1.200000µA - 20 + 40 -
12.000000µA - 20 + 10 -
120.00000µA - 20 + 8 -
199.99999µA 25 + 2 - 6.5 + 2.0
1.2000000mA - 20 + 5 -
1.9999999mA 25 + 2 - 6.5 + 2.0
12.000000mA - 20 + 5 -
19.999999mA 25 + 2 - 8.0 + 2.0
120.00000mA - 35 + 5 -
199.99999mA 50 + 5 - 33 + 4.0
1.0500000A - 110 + 10 -
1.9999999A 150 + 10 - 170 + 8.0
19.999999A - - 380 + 20
Page 4
Resistance
Calibration Interval 365 day, ±1°C of last TCAL
±(ppm of setting + ppm of range), relative to calibration standards used.
Range 1281 HP3458/HFL 8508A
1.99999999 Ohm - - 10 + 2.0
12.00000 Ohm - 10 + 5 -
19.9999999 Ohm 12 + 1.0 - 7.0 + 0.7
120.00000 Ohm - 8 + 5 -
199.999999 Ohm 8 + 0.3 7.0 + 0.25
1.2000000 kOhm - 7.5 + 0.5 -
1.99999999 kOhm 6 + 0.3 - 7.0 + 0.25
12.000000 kOhm - 7.5 + 0.5 -
19.9999999 kOhm 6 + 0.3 - 7.0 + 0.25
120.00000 kOhm - 7.5 + 0.5 -
199.999999 kOhm 6 + 0.3 - 7.0 + 0.25
1.2000000 MOhm - 15 + 2 -
1.99999999 MOhm 10 + 0.7 - 7.0 + 0.5
12.000000 MOhm - 50 + 10 -
19.9999999 MOhm 20 + 4.0 - 9.0 + 5.0
120.00000 MOhm - 500 + 10 -
199.999999 MOhm 200 + 45 - 30 + 50
1.2000000 GOhm - 5000 + 10 -
1.99999999 GOhm 2000 + 450 - 500 + 500
Page 5
AC Current ±(ppm of setting + ppm of range), relative to calibration standards used.
Calibration Interval 365 day, ±1°C of last TCAL
Range Frequency 1281 HP3458/HFL 8508A
120.0000µA 10Hz - 20Hz - 4000 + 200 -
199.9999µA 1Hz - 10Hz - - 250 + 100
1.2000000mA 10Hz - 20Hz - 4000 + 200 -
1.999999mA 1Hz - 10Hz - - 250 + 100
12.00000mA 10Hz - 20Hz - 4000 + 200 -
19.99999mA 1Hz - 10Hz - - 250 + 100
120.0000mA 10Hz - 20Hz - 4000 + 200 -
199.9999mA 1Hz - 10Hz - - 250 + 100
1.050000A 10Hz - 20Hz - 4000 + 200 -
1.999999A 10Hz - 1kHz 500 + 200 - -
19.99999A 10Hz - 2kHz - - 800 + 100
20Hz - 45Hz - 1500 + 200 ­45Hz - 100Hz - 600 + 200 ­100Hz - 5kHz - 600 + 200 -
10Hz - 5kHz 200 + 200 - ­10Hz - 10kHz - - 250 + 100 10kHz - 30kHz - - 600 + 100 30kHz - 100kHz - - 4000 + 100
20Hz - 45Hz - 1500 + 200 ­45Hz - 100Hz - 600 + 200 ­100Hz - 5kHz - 300 + 200 ­5kHz - 20kHz - 600 + 200 ­20kHz - 50kHz - 4000 + 400 ­50kHz - 100kHz - 5500 + 1500 -
10Hz - 5kHz 200 + 100 - ­10Hz - 10kHz - - 250 + 100 10kHz - 30kHz - - 600 + 100 30kHz - 100kHz - - 4000 + 100
20Hz - 45Hz - 1500 + 200 ­45Hz - 100Hz - 600 + 200 ­100Hz - 5kHz - 300 + 200 ­5kHz - 20kHz - 600 + 200 ­20kHz - 50kHz - 4000 + 400 ­50kHz - 100kHz - 5500 + 1500 -
10Hz - 5kHz 200 + 100 - ­10Hz - 10kHz - - 250 + 100 10kHz - 30kHz - - 600 + 100 30kHz - 100kHz - - 4000 + 100
20Hz - 45Hz - 1500 + 200 ­45Hz - 100Hz - 600 + 200 ­100Hz - 5kHz - 300 + 200 ­5kHz - 20kHz - 600 + 200 ­20kHz - 50kHz - 4000 + 400 ­50kHz - 100kHz - 5500 + 1500 -
10Hz - 5kHz 200 + 100 - ­10Hz - 10kHz - - 250 + 100 10kHz - 30kHz - - 600 + 100
20Hz - 45Hz - 1600 + 200 ­45Hz - 100Hz - 800 + 200 ­100Hz - 5kHz - 1000 + 200 ­5kHz - 20kHz - 3000 + 200 ­20kHz - 50kHz - 10000 + 400 -
10Hz - 2kHz - - 600 + 100 1kHz - 5kHz 1500 + 400 - ­2kHz - 10kHz - - 700 + 100 10kHz - 30kHz - - 3000 + 100
2kHz - 10kHz - - 2500 + 100
Page 6
AC Voltage ±(ppm of setting + ppm of range), relative to calibration standards used.
Calibration Interval 365 day, ±1°C of last TCAL
Range Frequency 1281 HP3458/HFL 8508A
12.00000mV 1Hz - 40Hz - 300 + 300 ­40Hz - 1kHz - 200 + 110 ­1kHz - 20kHz - 300 + 110 ­20kHz - 50kHz - 1000 + 110 ­50kHz - 100kHz - 5000 + 110 ­100kHz - 300kHz - 40000 + 200 -
120.0000mV 1Hz - 40Hz - 70 + 40 ­40Hz - 1kHz - 70 + 20 ­1kHz - 20kHz - 140 + 20 ­20kHz - 50kHz - 300 + 20 ­50kHz - 100kHz - 800 + 20 ­100kHz - 300kHz - 3000 + 100 ­300kHz - 1MHz - 10000 + 100 ­1MHz - 2MHz - 15000 + 100 -
199.9999mV 1Hz - 10Hz - - 120 + 70 10Hz - 40Hz - - 120 + 70 40Hz - 100Hz 100 + 20 - 100 + 20 100Hz - 2kHz 100 + 20 - 100 + 10 2kHz - 10kHz 100 + 20 - 100 + 20 10kHz - 30kHz 300 + 40 - 300 + 40 30kHz - 100kHz 700 + 100 - 700 + 100
1.2000000V 1Hz - 40Hz - 70 + 40 ­40Hz - 1kHz - 70 + 20 ­1kHz - 20kHz - 140 + 20 ­20kHz - 50kHz - 300 + 20 ­50kHz - 100kHz - 800 + 20 ­100kHz - 300kHz - 3000 + 100 ­300kHz - 1MHz - 10000 + 100 ­1MHz - 2MHz - 15000 + 100 -
1.999999V 1Hz - 10Hz - - 100 + 60 10Hz - 40Hz - - 100 + 10 40Hz - 100Hz 80 + 10 - 80 + 10 100Hz - 2kHz 60 + 10 - 60 + 10 2kHz - 10kHz 80 + 10 - 80 + 10 10kHz - 30kHz 200 + 20 - 200 + 20 30kHz - 100kHz 500 + 100 - 500 + 100 100kHz - 300kHz 3000 + 1000 - 3000 + 1000 300kHz - 1MHz 10000 + 10000 - 10000 + 10000
12.00000V 1Hz - 40Hz - 70 + 40 ­40Hz - 1kHz - 70 + 20 ­1kHz - 20kHz - 140 + 20 ­20kHz - 50kHz - 300 + 20 ­50kHz - 100kHz - 800 + 20 ­100kHz - 300kHz - 3000 + 100 ­300kHz - 1MHz - 10000 + 100 ­1MHz - 2MHz - 15000 + 100 -
Page 7
AC Voltage ±(ppm of setting + ppm of range), relative to calibration standards used.
Calibration Interval 365 day, ±1°C of last TCAL
Range Frequency 1281 HP3458/HFL 8508A
19.99999V 1Hz - 10Hz - - 100 + 60
10Hz - 40Hz - - 100 + 10 40Hz - 100Hz 80 + 10 - 80 + 10 100Hz - 2kHz 60 + 10 - 60 + 10 2kHz - 10kHz 80 + 10 - 80 + 10 10kHz - 30kHz 200 + 20 - 200 + 20 30kHz - 100kHz 500 + 100 - 500 + 100 100kHz - 300kHz 3000 + 1000 - 3000 + 1000 300kHz - 1MHz 10000 + 10000 - 10000 + 10000
120.0000V 1Hz - 40Hz - 200 + 40 -
40Hz - 1kHz - 200 + 20 ­1kHz - 20kHz - 200 + 20 ­20kHz - 50kHz - 350 + 20 ­50kHz - 100kHz - 1200 + 20 ­100kHz - 300kHz - 4000 + 100 ­300kHz - 1MHz - 15000 + 100 -
199.9999V 1Hz - 10Hz - - 100 + 60
10Hz - 40Hz - - 100 + 10 40Hz - 100Hz 80 + 10 - 80 + 10 100Hz - 2kHz 60 + 10 - 60 + 10 2kHz - 10kHz 80 + 10 - 80 + 10 10kHz - 30kHz 200 + 20 - 200 + 20 30kHz - 100kHz 500 + 100 - 500 + 100 100kHz - 300kHz 3000 + 1000 - 3000 + 1000 300kHz - 1MHz 10000 + 10000 - 10000 + 10000
1050.000V 1Hz - 10Hz - 400 + 40 100 + 70
10Hz - 40Hz - 400 + 40 100 + 20 40Hz - 1kHz - 400 + 20 ­40Hz - 10kHz 80 + 10 - 80 + 20 1kHz - 20kHz - 600 + 20 ­10kHz - 30kHz 200 + 20 - 200 + 40 20kHz - 50kHz - 1200 + 20 ­30kHz - 100kHz 500 + 100 - 500 + 200 50kHz - 100kHz - 3000 + 20 -
Page 8
Fluke 8508A 3458/HFL
8508A vs 3458/HFL
Functional & Versatility
Comments
Parameters
Full Scale (dc & ac Voltage) 200mV, 2V, 20V, 200V, 1050Vrms
Full Scale (dc & ac current)
AC Measurement (Bandwidth)
Full Scale (Resistance)
200 Ohm Resistance Range.
200uA, 2mA, 20mA, 200mA, 2A & 20A
AC Voltage - 1MHz (2%) AC Current - 100kHz (0.4%)
2Ω, 20Ω, 200Ω, 2kΩ, 20kΩ, 200k, 2M, 20M, 200M 2G, 20G(eleven ranges)
Selectable excitation current ­10mA & 1mA
Ω,
120mV, 1.2V, 12V, 120V, 1050Vdc (ac = 700Vrms)
100nA, 1uA, 120uA, 1.2mA, 12mA, 120mA, 1.05A
AC Voltage - 1MHz (1%) AC Voltage - 10MHz (15%) AC Current - 100kHz (0.7% Typical)
12, 120, 1.2k, 12k, 120k, 1.2M, 12M, 120M, 1.2G(nine ranges)
Fixed excitation current ­10mA
8508A maintains >10Gohm input impedance to 20Vdc. This compares with 12V on 3458/HFL. The 8508A 1050Vrms measurement is often used to calibrate calibrators at 1000 Vrms output. 3458A/HFL has lower current ranges with a sensitivity of 1pA compared with 10pA on the 8508A – the 3458/HFL datasheet does state that lower ranges have ‘typical’ uncertainty only. 8508A states typical above 10kHz and 2GOhm for resistance.
A 20Amp range on the 8508A improves calibration of Multi-product calibrators. This reduces the need for additional external shunts. A DMM with 10 MOhm input impedance is generally not useful for precision ac measurement above 1MHz. A 50 Ohm termination at these frequencies is more desirable. The 8508A high frequency Current uncertainty will help maintain good TURs when calibrating Multi-Product calibrators.
8508A has a wide resistance measurement range. 8508A measurement resolution of
10n
this compares with 10u on
3458/HFL Errors due to self-heating within the resistor
under test are minimized when the multimeter low excitation current is maintained. Use in conjunction with RTD or low resistance measurements.
Page 9
8508A vs 3458/HFL
Functional & Versatility
Parameters
Comments
High Resistance Measurement (compliance voltage)
Tru-Ohms or Offset Compensated Resistance technique.
Store RTD temperature coefficients
ACAL Not Required
Auto-zero (A-zero) Range zero is enabled by operator
Low impedance low Current measurement technology
ACV measurement technique
Test Voltage or Full Scale stimulus = 200V
Current Reversal Switched Current
Supports a wide variety of PRT & SPRT sensors
Yes No
Analog True rms
Test Voltage or Full Scale stimulus @ 1.2G= 5V
Not Available
Required every 24hrs or
of last ACAL
A-zero should be enabled to maintain specification
Analog, Synchronous and Random sampled ac
±1º C
Reduces errors when performing high resistance measurement (feature available 8508A - 20M to 20Grange only)
Current reversal reduces errors caused by thermal emfs as does switched current but reversal technique has the added benefit of reducing errors due to self heating
PRT calibration coefficients are stored (max 100 probes) within the 8508As internal memory. 8508A - Precision without the need for internal drift corrections. ACAL will overwrite last stored value and cannot be retrieved. This might be important for users that require ‘as found data’. Auto-zero must be enabled on 3458A/HFL if
environment exceeds zero. Zero/nulling meters before measurement is accepted metrology practice. 8508A is less intrusive when compared with conventional meters when measuring low magnitude current sources. The 8508A input impedance on the 200uA range is 60 Ohm. This low impedance mimics technology normally found in Electrometers and picoampmeters. This compares with 5k – 10k Ohm on other DMMs. Digitize sub sampled techniques are useful when measuring low frequency <1Hz input signals.
±1°C of last Input
Page 10
ACV & DCV 4 wire input.
Remote switched GUARD Terminal
Read rate
Calibration technique
Selectable 2 and 4-wire measurement.
IEEE controlled Manual Switch
High Speed - 150 readings/sec @
5.5 digit resolution
Individual Function and Range calibration.
Unavailable
High Speed - 100,000 reading/sec @ 4.5 digit resolution
Artifact Calibration & Individual Function and Range calibration.
4-wire sense measurement support improved accuracy when calibrating calibrators with 4 wire sense output.
IEEE control of GUARD terminal helps fully automate calibration without need for manual switching.
High read rates are useful for some ATE applications where accuracy is not of high priority.
Both techniques have their merits. 8508A and associated calibration technique provides full results traceability. ‘As found’ data is difficult to determine using 3458/HFL ‘artifact’ calibration. Artifact calibration provides a low cost solution for adjustment.
Page 11
8508A vs 3458/HFL
Ease of Use Parameters
Rear input switched remotely using IEEE 488 commands Instrument simultaneously displays Frequency as well as the AC voltage & current.
Two channel Ratio reference measurement
Able to load SPRT/PRT calibration linearization coefficients and display temperature and resistance simultaneously.
Constant energization current on Resistance Ratio function.
GPIB Remote interface IEEE-488.2 IEEE-488.1
Self-cal or ACAL function Not Required Required to meet specification
Yes (8508A/01 only)
Yes (up to100) No
Comments
Yes No
Yes Selectable
Yes, but limited to 12Vdc on
the Reference input
Yes
No (the 3458/HFL does not
measure resistance ratio)
The 3458/HFL must be switched manually using a front panel button.
Requires further commands or key presses to determine frequency.
8508A can accept and compare dc to dc, ac to ac, dc to ac, Resistance to Resistance & Temperature to Temperature. Auto-ranging allows inputs of different magnitude to be connected and compared. Feature supports the replacement of Kelvin Varley dividers, resistance bridge and ac/dc transfer standards
8508As displays resistance and temperature simultaneously.
When measuring Resistance in Ratio mode not only is the current the same but it is reversed to help reduce thermal offsets.
8508A components are selected for minimum drift over time and temperature. Therefore does not require frequent internal calibration when conditions exceed TCAL specification.
Page 12
8508A Options and Accessories:
Model Description 8508A
8508A/01
Accessories
8508A-SPRT
8508A-PRT
8508A-LEAD
Y8508
Y8508S
8508A-7000K
1883673
1256990
Model Description 3458/HFL/E 3458/HFLM/E 3458/HFL/E/AC 3458/HFLM/E/120/AC
8.5 digit Reference Multimeter
8.5 digit Reference Multimeter with Rear Input Ratio measurement
Standard Platinum Resistance Thermometer
Platinum Resistance Thermometer
Comprehensive Measurement Lead Kit
Rack Mount Kit
Rack Mount Kit Slides
Calibration kit comprising 1GOhm Standard and connecting leads.
UKAS Accredited Certification
NVLAP Accredited Certification
Special Lab DMM, English Special Lab DMM, English, with -001 Extended Memory Option Special Lab DMM, English, + Accredited Calibration Special Lab DMM, English, with -001 Extended Memory Option + Accredited Calibration