When comparing two products features and specification using sales information
it is often difficult to determine which product will meet the needs of your specific
application. This is mostly because each supplier will define specifications using
alternate methods. This Selection Guide will help you compare the specifications
and features of the Fluke 8508A Reference multimeter with the nearest
equivalent digital multimeter the Agilent 3458/HFL. Both the 8508A and the
special 3458/HFL multimeters are available through the Fluke sales channel.
The following pages summarize & segment the features of each meter into the
following three categories.
• Accuracy
• Functionality & Versatility
• Ease of Use
The 'Comments' column found in the following table provides a brief explanation
against each feature and its relevance to the intended metrology applications.
Further reference information:
1) Agilent 3458A/HFL Technical Data - 1275558 D-ENG-N
2) Fluke 8508A Extended Specifications - 1887212 D-ENG-N rev A
3) 1281Operators Handbook – 850090 issue 6
TECHNICAL INFORMATION
Page 2
DC Voltage
Calibration Interval 365 day, ±1°C of last TCAL
±(ppm of setting + ppm of range), relative to calibration standards used.
Range1281HP3458/HFL8508A
120.00000mV-4 + 3-
199.999999mV6 + 0.5-2.7 + 0.5
1.20000000V-3 + 0.3-
1.99999999V3 + 0.2-2.7 + 0.2
12.000000V-3 + 0.05-
19.9999999V3 + 0.2-2.7 + 0.2
120.00000V-5 + 0.3-
199.999999V6 + 0.1-4.0 + 0.2
1050.0000V6 + 0.15 + 0.14.0 + 0.5
Page 3
DC Current
Calibration Interval 365 day, ±1°C of last TCAL
±(ppm of setting + ppm of range), relative to calibration standards used.
Range1281HP3458/HFL8508A
120.000nA-30 + 400-
1.200000µA-20 + 40-
12.000000µA-20 + 10-
120.00000µA-20 + 8-
199.99999µA25 + 2-6.5 + 2.0
1.2000000mA-20 + 5-
1.9999999mA25 + 2-6.5 + 2.0
12.000000mA-20 + 5-
19.999999mA25 + 2-8.0 + 2.0
120.00000mA-35 + 5-
199.99999mA50 + 5-33 + 4.0
1.0500000A-110 + 10-
1.9999999A150 + 10-170 + 8.0
19.999999A--380 + 20
Page 4
Resistance
Calibration Interval 365 day, ±1°C of last TCAL
±(ppm of setting + ppm of range), relative to calibration standards used.
Range1281HP3458/HFL8508A
1.99999999 Ohm--10 + 2.0
12.00000 Ohm-10 + 5-
19.9999999 Ohm12 + 1.0-7.0 + 0.7
120.00000 Ohm-8 + 5-
199.999999 Ohm8 + 0.37.0 + 0.25
1.2000000 kOhm-7.5 + 0.5-
1.99999999 kOhm6 + 0.3-7.0 + 0.25
12.000000 kOhm-7.5 + 0.5-
19.9999999 kOhm6 + 0.3-7.0 + 0.25
120.00000 kOhm-7.5 + 0.5-
199.999999 kOhm6 + 0.3-7.0 + 0.25
1.2000000 MOhm-15 + 2-
1.99999999 MOhm10 + 0.7-7.0 + 0.5
12.000000 MOhm-50 + 10-
19.9999999 MOhm20 + 4.0-9.0 + 5.0
120.00000 MOhm-500 + 10-
199.999999 MOhm200 + 45-30 + 50
1.2000000 GOhm-5000 + 10-
1.99999999 GOhm2000 + 450-500 + 500
Page 5
AC Current ±(ppm of setting + ppm of range), relative to calibration standards used.
8508A maintains >10Gohm input impedance
to 20Vdc. This compares with 12V on
3458/HFL. The 8508A 1050Vrms
measurement is often used to calibrate
calibrators at 1000 Vrms output.
3458A/HFL has lower current ranges with a
sensitivity of 1pA compared with 10pA on
the 8508A – the 3458/HFL datasheet does
state that lower ranges have ‘typical’
uncertainty only. 8508A states typical above
10kHz and 2GOhm for resistance.
A 20Amp range on the 8508A improves
calibration of Multi-product calibrators. This
reduces the need for additional external
shunts.
A DMM with 10 MOhm input impedance is
generally not useful for precision ac
measurement above 1MHz. A 50 Ohm
termination at these frequencies is more
desirable.
The 8508A high frequency Current
uncertainty will help maintain good TURs
when calibrating Multi-Product calibrators.
8508A has a wide resistance measurement
range. 8508A measurement resolution of
10n
Ω this compares with 10uΩ on
3458/HFL
Errors due to self-heating within the resistor
under test are minimized when the
multimeter low excitation current is
maintained. Use in conjunction with RTD or
low resistance measurements.
Page 9
8508A vs 3458/HFL
Functional & Versatility
Parameters
Comments
High Resistance
Measurement (compliance
voltage)
Tru-Ohms or Offset
Compensated Resistance
technique.
Store RTD temperature
coefficients
ACAL Not Required
Auto-zero (A-zero) Range zero is enabled by operator
Low impedance low Current
measurement technology
ACV measurement
technique
Test Voltage or Full Scale
stimulus = 200V
Current Reversal Switched Current
Supports a wide variety of PRT &
SPRT sensors
Yes No
Analog True rms
Test Voltage or Full Scale
stimulus @ 1.2GΩ = 5V
Not Available
Required every 24hrs or
of last ACAL
A-zero should be enabled to
maintain specification
Analog, Synchronous and
Random sampled ac
±1º C
Reduces errors when performing high
resistance measurement (feature available
8508A - 20MΩ to 20GΩ range only)
Current reversal reduces errors caused by
thermal emfs as does switched current but
reversal technique has the added benefit of
reducing errors due to self heating
PRT calibration coefficients are stored (max
100 probes) within the 8508As internal
memory.
8508A - Precision without the need for
internal drift corrections. ACAL will overwrite
last stored value and cannot be retrieved.
This might be important for users that
require ‘as found data’.
Auto-zero must be enabled on 3458A/HFL if
environment exceeds
zero. Zero/nulling meters before
measurement is accepted metrology
practice.
8508A is less intrusive when compared with
conventional meters when measuring low
magnitude current sources. The 8508A
input impedance on the 200uA range is 60
Ohm. This low impedance mimics
technology normally found in Electrometers
and picoampmeters. This compares with 5k
– 10k Ohm on other DMMs.
Digitize sub sampled techniques are useful
when measuring low frequency <1Hz input
signals.
±1°C of last Input
Page 10
ACV & DCV 4 wire input.
Remote switched GUARD
Terminal
Read rate
Calibration technique
Selectable 2 and 4-wire
measurement.
IEEE controlled Manual Switch
High Speed - 150 readings/sec @
5.5 digit resolution
Individual Function and Range
calibration.
Unavailable
High Speed - 100,000
reading/sec @ 4.5 digit
resolution
Artifact Calibration & Individual
Function and Range
calibration.
4-wire sense measurement support
improved accuracy when calibrating
calibrators with 4 wire sense output.
IEEE control of GUARD terminal helps fully
automate calibration without need for
manual switching.
High read rates are useful for some ATE
applications where accuracy is not of high
priority.
Both techniques have their merits. 8508A
and associated calibration technique
provides full results traceability. ‘As found’
data is difficult to determine using 3458/HFL
‘artifact’ calibration.
Artifact calibration provides a low cost
solution for adjustment.
Page 11
8508A vs 3458/HFL
Ease of Use Parameters
Rear input switched remotely
using IEEE 488 commands
Instrument simultaneously
displays Frequency as well as
the AC voltage & current.
Two channel Ratio reference
measurement
Able to load SPRT/PRT
calibration linearization
coefficients and display
temperature and resistance
simultaneously.
Constant energization current on
Resistance Ratio function.
GPIB Remote interface IEEE-488.2 IEEE-488.1
Self-cal or ACAL function Not Required Required to meet specification
Yes (8508A/01 only)
Yes (up to100) No
Comments
Yes No
Yes Selectable
Yes, but limited to 12Vdc on
the Reference input
Yes
No (the 3458/HFL does not
measure resistance ratio)
The 3458/HFL must be switched manually
using a front panel button.
Requires further commands or key presses
to determine frequency.
8508A can accept and compare dc to dc, ac
to ac, dc to ac, Resistance to Resistance &
Temperature to Temperature. Auto-ranging
allows inputs of different magnitude to be
connected and compared. Feature supports
the replacement of Kelvin Varley dividers,
resistance bridge and ac/dc transfer
standards
8508As displays resistance and temperature
simultaneously.
When measuring Resistance in Ratio mode
not only is the current the same but it is
reversed to help reduce thermal offsets.
8508A components are selected for
minimum drift over time and temperature.
Therefore does not require frequent internal
calibration when conditions exceed TCAL
specification.
Page 12
8508A Options and Accessories:
Model Description
8508A
8508A/01
Accessories
8508A-SPRT
8508A-PRT
8508A-LEAD
Y8508
Y8508S
8508A-7000K
1883673
1256990
Model Description
3458/HFL/E
3458/HFLM/E
3458/HFL/E/AC
3458/HFLM/E/120/AC
8.5 digit Reference Multimeter
8.5 digit Reference Multimeter with Rear Input Ratio measurement
Standard Platinum Resistance Thermometer
Platinum Resistance Thermometer
Comprehensive Measurement Lead Kit
Rack Mount Kit
Rack Mount Kit Slides
Calibration kit comprising 1GOhm Standard and connecting leads.
UKAS Accredited Certification
NVLAP Accredited Certification
Special Lab DMM, English
Special Lab DMM, English, with -001 Extended Memory Option
Special Lab DMM, English, + Accredited Calibration
Special Lab DMM, English, with -001 Extended Memory Option + Accredited Calibration