345 Calibration Manual Supplement
Change #1
On page 26, under Power Check-Stage 1, replace the entire table with:
Calibrator Specification (k)W
Volts (V) Amps (A) Phase Shift (°) Min Nom Max
19 100 0.00 1820 1900 1980
21 100 0.00 2042 2100 2158
38 100 0.00 3700 3800 3900
44 100 0.00 4.24 4.40 4.56
380 100 0.00 37.00 38.00 39.00
440 100 0.00 42.4 44.0 45.6
Calibrator Specification (k)VA
Volts (V) Amps (A) Phase Shift (°) Min Nom Max
19 100 0.00 1848 1900 1943
21 100 0.00 2042 2100 2158
38 100 0.00 3700 3800 3900
44 100 0.00 4.24 4.40 4.56
380 100 0.00 37.00 38.00 39.00
440 100 0.00 42.4 44.0 45.6
Calibrator Specification PF
Volts (V) Amps (A) Phase Shift (°) Min Nom Max
19 100 0.00 0.998 1.00 1.00
21 100 0.00 0.998 1.00 1.00
38 100 0.00 0.998 1.00 1.00
44 100 0.00 0.998 1.00 1.00
380 100 0.00 0.998 1.00 1.00
440 100 0.00 0.998 1.00 1.00
3/12 1
Manual Supplement 345 Calibration
On page 27, under Power Check-Stage 2, replace the entire table with:
For 345 with Firmware V1.1
Calibrator Specification (k)W
Volts (V) Amps (A) Phase Shift (°) Min Nom Max
19.8681 100 73.00 501 581 661
21.9595 100 -73.00 562 642 722
39.2133 100 73.00 1066 1146 1226
46.0105 100 -73.00 1.27 k 1.35 k 1.43 k
397.363 100 73.00 10.83 k 11.62 k 11.96 k
460.105 100 -73.00 12.6 k 13.5 k
Calibrator Specification (k)VAR
Volts (V) Amps (A) Phase Shift (°) Min Nom Max
19.8681 100 73.00 1650 1900 2150
21.9595 100 -73.00 -2350 -2100 -1850
39.2133 100 73.00 3500 3750 4000
46.0105 100 -73.00 -4.56 k -4.40 k -4.24 k
397.363 100 73.00 37.00 k 38.00 k 39.00 k
460.105 100 -73.00 -45.6 k -44.0 k -42.4 k
Calibrator Specification PF
Volts (V) Amps (A) Phase Shift (°) Min Nom Max
19.8681 100 73.00 0.242 0.292 0.342
21.9595 100 -73.00 0.242 0.292 0.342
39.2133 100 73.00 0.242 0.292 0.342
13.8 k
46.0105 100 -73.00 0.242 0.292 0.342
397.363 100 73.00 0.242 0.292 0.342
460.105 100 -73.00 0.242 0.292 0.342
2 3/12
345 Calibration Manual Supplement
For 345 with Firmware V1.2 or greater
Calibrator Specification (k)W
Volts (V) Amps (A) Phase Shift (°) Min Nom Max
19.8681 100 73.00 501 581 661
21.9595 100 -73.00 562 642 722
39.2133 100 73.00 1066 1146 1226
46.0105 100 -73.00 1.27 k 1.35 k 1.43 k
397.363 100 73.00 10.83 k 11.62 k 11.96 k
460.105 100 -73.00 12.6 k 13.5 k 13.8 k
Calibrator Specification (k)VAR
Volts (V) Amps (A) Phase Shift (°) Min Nom Max
19.8681 100 73.00 -2150 -1900 -1650
21.9595 100 -73.00 1850 2100 2350
39.2133 100 73.00 -4000 -3750 -3500
46.0105 100 -73.00 4.24 k 4.40 k 4.56 k
397.363 100 73.00 -39.00 k -38.00 k -37.00 k
460.105 100 -73.00 42.4 k 44.0 k 45.6 k
345 Calibrator Specification PF
Volts (V) Amps (A) Phase Shift (°) Min Nom Max
19.8681 100 73.00 0.242 0.292 0.342
21.9595 100 -73.00 0.242 0.292 0.342
39.2133 100 73.00 0.242 0.292 0.342
46.0105 100 -73.00 0.242 0.292 0.342
397.363 100 73.00 0.242 0.292 0.342
460.105 100 -73.00 0.242 0.292 0.342
3/12 3
345 Calibration Manual Supplement
Change #2, 63545
On page 4, add the following to the Symbols table:
Conforms to relevant South Korean EMC Standards.
Conforms to CAN/CSA-C22.2 No. 61010-1, second edition, including Amendment 1.
Change #3, 65704
On page 9, under General Data:
Change: Battery Eliminator BE345 Input……110V / 230V, 50/60 Hz
Output….15 V dc, 300 mA
To: Battery Eliminator GFP151DA-120125B-1
Input……100 V to 240 V, 50/60 Hz, 0.36 A Output….12 V dc, 1.25 A
On page 10, in the Note:
Change: BE345 To: GFP151DA-120125B-1
On page 10, under the Warning remove the second bullet and in bullets one and three:
Change: BE345
To: GFP151DA-120125B-1
On page 10, remove Figure 2.
On page 12, under Battery and Battery Eliminator Check in step 1:
Change: BE345
To: GFP151DA-120125B-1
On page 15, Table 2:
Change: BE345 To: GFP151DA-120125B-1
On page 28, Table 4 in the part number column:
Change: BE345 To: GFP151DA-120125B-1
Change #4, 66326
On page 4, Table 1, replace CAT III and CAT IV, with:
CAT III
CAT IV
4 8/13
Measurement Category III is applicable to test and measuring circuits connected to the distribution part of the building’s low-voltage MAINS installation.
Measurement Category IV is applicable to test and measuring circuits connected at the source of the building’s low-voltage MAINS installation.
345 Calibration Manual Supplement
On page 9, replace Electrical Safety and EMC sections, with:
General Specifications
Maximum working voltage in CAT IV areas:
Current measurement: ............................... 600 V ac rms or dc between conductor & ground
Voltage measurement ................................ 600 V ac rms or dc between either input terminal and ground, or 825 V between
Maximum working voltage in CAT III areas .... 825 V ac rms or dc between either input terminal and ground
Safety ............................................................. IEC 61010-1: 600 V CAT IV, Pollution Degree 2
IP Rating ........................................................ IEC 60529: IP40
Electromagnetic Environment ..................... IEC 61326-1: Portable
Electromagnetic Compatibility .................... Applies to use in Korea only. Class A Equipment (Industrial Broadcasting &
energized phase voltages (delta power config.)
Communication Equipment) [1] This product meets requirements for industrial (Class A) electromagnetic wave
equipment and the seller or user should take notice of it. This equipment is intended for use in business environments and is not to be used in homes.
[1]
5 8/13