336 and 337
Clamp Meter
Instruction Card
Warning
Read the 333, 334, 335, 336, 337 Clamp Meter
Safety Sheet before using this meter.
Lisez la fiche de sécurité 333, 334, 335, 336, 337
Clamp Meter Safety Sheet avant d'utiliser cet appareil.
Prima di usare questo strumento, leggere
il documento 333, 334, 335, 336, 337 Clamp Meter
Safety Sheet.
Vor der Verwendung des Meßgeräts die
Sicherheitsinformationen zum 333, 334, 335, 336, 337
Clamp Meter lesen
Lea la hoja de instrucciones de seguridad "333, 334,
335, 336, 337 Clamp Meter Safety Sheet" antes de
utilizar este medidor.
Leia "333, 334, 335, 336, 337 Clamp Meter Safety
Sheet" antesde usar este medidor.
Lees het 333, 334, 335, 336, 337 Clamp Meter Safety
Sheet voordat u deze meter gebruikt
Les 333, 334, 335, 336, 337 Clamp Meter Safety Sheet
før måleinstrumentet tas i bruk.
Man skal læse "333, 334, 335, 336, 337 Clamp Meter
Safety Sheet" inden man tager instrumentet i brug.
Lue 333, 334, 335, 336, 337 Clamp Meter Safety
Sheet ennen tämän laitteen käyttöä.
Läs säkerhetsdatabladet "333, 334, 335, 336, 337
Clamp Meter Safety Sheet" innan du använder
detta instrument.
Meter SafetySheet
Clamp Meter Safety Sheet
Clamp Meter Safety Sheet
Clamp Meter Safety Sheet
333, 334, 335, 336, 337 Clamp Meter Safety Sheet
PN 1549876
April 2001 Rev.3, 6/05
2001-2005 Fluke Corporation All rights reserved.
www.fluke.com
333, 334, 335, 336, 337 Clamp
333, 334, 335, 336, 337
333, 334, 335, 336, 337
333, 334, 335, 336, 337
333, 334, 335, 336, 337
On/Off
A
V
V
OFF
MIN
MAX
Automatic Off
20 Min
A
V
V
OFF
Backlight On/Off
Hz
A
A
A
A
V
OFF
Volt s
Volt s
DC
Amps
Hz
AC
DC
CAT
A
V
OFF
Volt s
DC
V
V
ZERO
TRUE RMS CLAM
COM
600V
1000A
Tactile Barrier
Collerette de protection (Fem.)
Hz
A
A
Nervatura di sicurezza
Griffschutz
Protección dactilar
CAT
Barreira tátil
Følebarriere
OFF
Amps
AC
MIN/MAX
INRUSH
337
P METER
CAT
V
600V
Tastbare barrière
Berøringsgrensen
Kosketuseste
Känselbarriär
V
V
ZERO
TRUE RMS CLAM
COM
600V
1000A
A
Hz
A
A
OFF
DC
INRUSH
P METER
CAT
600V
ZERO
Amps
MIN/MAX
337
V
CAT
1
V
V
ZERO
TRUE RMS CLAM
COM
600V
1000A
Hz
A
A
OFF
Amps
DC
MIN/MAX
INRUSH
337
P METER
CAT
V
600V
2
Battery Replacement
1
V
V
ZERO
TRUE RMS C
COM
600V
1000A
Hz
A
A
OFF
Amps
AC
MIN/MAX
INRUSH
337
LAMP METER
CAT
V
600V
Hz
A
INRUSH
TRUE RMS
R
E
ET
335
M
V
P
LAM
C
T
CA
600V
COM
CAT
2
Currents Cancel
Les courants s'annulent, Le correnti si cancellano,
Ströme abbrechen, Se cancelan las corrientes,
Cancelamento de correntes, Stromen compenseren
elkaar, Strømutjevning, Strømstyrkeannullering
Strømstyrker udbalancerer, Virta pois, Avbryt ström
,
,
,,
V,
0 - 600.0 V ac
0 - 600.0 V dc
0 - 6000
V
V
V
V
MAXMIN
ZERO
TRUE RMS CLAM
COM
600V
1000A
Hz
A
A
OFF
Amps
Volts
Hz
AC
DC
MIN/MAX
INRUSH
337
P METER
CAT
V
600V
CAT
Continuity
CAT
CAT
A
V
V
OFF
ZERO
INRUSH
TRUE RMS CLAMP METER
CAT
COM
600V
A
V
V
OFF
ZERO
INRUSH
TRUE RMS CLAMP METER
CAT
COM
600V
Inrush Current
Appel de courant, Corrente di spunto,
Hold
Hz (Amps Only)
337
Einschaltstoßstrom,Corriente de entrada, Corrente
de partida, Inkomende stroom,
Innkoplingsstrømspiss, Sekundær
tomgangsstrømstyrke,Tasoitettu virta, Inström,
600V
1000A
Hz
A
, , ,
< 30
A
V
V
ZERO
TRUE RMS CLAMP METER
COM
A
OFF
600V
1000A
A
Hz
A
OFF
Amps
AC
INRUSH
MIN/MAX
INRUSH
337
CAT
V
600V
600V
1000A
Hz
MIN/MAX
337
V
600V
1000A
Hz
A
MIN/MAX
337
V
CAT
CAT
A
V
V
,
CAT
Off
1
1
A
V
V
OFF
ZERO
INRUSH
TRUE RMS CLAMP METER
CAT
COM
600V
600V
1000A
Hz
A
Amps
Amps
AC
MIN/MAX
337
V
AC
CAT
A
V
V
OFF
ZERO
INRUSH
TRUE RMS CLAMP METER
CAT
COM
600V
600V
1000A
Hz
A
Hz
MIN/MAX
337
V
Min Max
337
Amps
INRUSH
INRUSH
MIN/MAX
MIN/MAX
MIN/MAX
AC
Amps
AC
Amps
AC
1
MIN/MAX
2
MIN/MAX
3
2 s
2
HOLD
On
3
HOLD
MAX
2
600V
A
V
V
OFF
ZERO
INRUSH
TRUE RMS CLAMP METER
CAT
COM
600V
1000A
ZERO
Hz
A
MIN
Amps
Amps
AC
MIN/MAX
337
V
AC
ZERO
MIN MAX
Amps
AC
CAT
100 ms
4
3
Specifications
*@ 23 ˚C ± 5 ˚C, 0 - 90 % RH 336 337
A
Inrush Integration
A
V
(20 - 400 Hz)
V
Continuity
Hz-Amps Only
Trigger Level
10-100 Hz 5 A
5-10, 100-400
Hz 10 A
Storage
Temperature
Operating
Temperature
Altitude
EMC - instrument unspecified for use in EMC field 0.5 V/m
CAT III 600V, pollution degree II:
CAT III equipment is designed to protect against transients in
equipment in fixed - equipment installations, such as distribution
panels, feeders and short branch circuits, and lighting systems
in large buildings.
* < 18 ˚C, > 28 ˚C add 0.1 x (specified accuracy)/ ˚C
Range 0 - 600.0 A 0 - 999.9 A
Accuracy 2 % ± 5 counts (10 - 100 Hz)
Crest Factor
add 2% for
CF > 2
6 % ± 5 counts (100 - 400 Hz)
3 @ 500 A
2.5 @ 600 A
3 @ 500 A
2.5 @ 600 A
1.42 @ 1000 A
AC Response rms
Time
100 ms
Range 0 - 600.0 A 0 - 999.9 A
Accuracy 2 % ± 5 counts
Range 0 - 600.0 V
Accuracy
1 % ± 5 counts (20 - 100 Hz)
6 % ± 5 counts (100 - 400 Hz)
AC Response rms
Range 0 - 600.0 V
Accuracy 1 % ± 5 counts
Range 0 - 600.0 Ω, 600 - 6000 Ω
Accuracy 1.5 % ± 5 counts
≤ 30 Ω
Range
Accuracy
5.0 - 400.0 Hz
0.5 % ± 5 counts
-40 ˚C to 60 ˚C
-10 ˚C to 50 ˚C
2500 m