Fluke 321 Fluke 322 Fluke 333 Fluke 334 Fluke 335 Fluke 336 Fluke 337
Range 0 - 400.0 A 0 - 40.00 A 0 - 400.0 A 0 - 600.0 A 0 - 600.0 A 0 - 600.0 A 0 - 999.9 A
40.0 - 400.0 A
Accuracy 1.8% ± 5 counts 1.8% ± 5 counts 2% ± 5 counts 2% ± 5 counts 2% ± 5 counts 2% ± 5 counts 2% ± 5 counts
(50 - 60 Hz) (50 - 60 Hz) (50 - 60 Hz) (50 - 60 Hz) (50 - 60 Hz) (10 - 100 Hz) (10 - 100 Hz)
3.0% ± 5 counts 3.0% ± 5 counts 6% ± 5 counts 6% ± 5 counts
(60 Hz - 400 Hz) (60 Hz - 400 Hz) (100 - 400 Hz) (100 - 400 Hz)
Crest Factor ————2.4 @ 500 A3 @ 500 A 3 @ 500 A
(50/60 Hz) 2.0 @ 600 A 2.5 @ 600 A 2.5 @ 600 A
add 2% for C.F.>2 add 2% for C.F.>2 1.42 @ 1000 A
add 2% for C.F.>2
AC Response Averaging Averaging Averaging Averaging True-rms True-rms True-rms
Inrush Integration Time — — — 100 mS 100 mS 100 mS 100 mS
Range —————0 - 600.0 A 0 - 999.9 A
Accuracy —————2% + 3 counts 2% + 3 counts
Range 0 - 400.0 V 0 - 400.0 V 0 - 600.0 V 0 - 600.0 V 0 - 600.0 V 0 - 600.0 V 0 - 600.0 V
0 - 600 V 0 - 600 V
Accuracy 1.2% + 5 counts 1.2% + 5 counts 1% + 5 counts 1% + 5 counts 1% + 5 counts 1% + 5 counts 1% + 5 counts
(50 - 400 Hz) (50 - 400 Hz) (50/60 Hz) (50/60 Hz) (50/60 Hz) (20 - 100 Hz) (20 - 100 Hz)
6% + 5 counts 6% + 5 counts
(100 - 400 Hz) (100 - 400 Hz)
AC Response Averaging Averaging Averaging Averaging True-rms True-rms True-rms
Range — 0 - 400.0 V 0 - 600.0 V 0 - 600.0 V 0 - 600.0 V 0 - 600.0 V 0 - 600.0 V
0 - 600 V
Accuracy — 1% + 5 counts 1% + 5 counts 1% + 5 counts 1% + 5 counts 1% + 5 counts 1% + 5 counts
Range 0 - 400.0 Ω 0 - 400.0 Ω 0 - 600.0 Ω 0 - 600.0 Ω 0 - 600.0 Ω 0 - 600.0 Ω 0 - 600.0 Ω
601 - 6000 Ω 601 - 6000 Ω 601 - 6000 Ω 601 - 6000 Ω
Accuracy 1% ± 5 counts 1% ± 5 counts 1.5% ± 5 counts 1.5% ± 5 counts 1.5% ± 5 counts 1.5% ± 5 counts 1.5% ± 5 counts
Continuity ≤ 30 Ω≤30 Ω≤30 Ω≤30 Ω≤30 Ω≤30 Ω≤30 Ω
Range — — – — — — 5.0 - 400.0 Hz
Accuracy ——————0.5% ±5 counts
Trigger Level ——————10 - 100 Hz ≥ 5 A
5 - 10 Hz,
100 - 400 Hz ≥ 10 A
MIN/MAX No No No No No No Yes
Backlight No No No No Yes Yes Yes
Display Hold Yes Yes Yes Yes Yes Yes Yes
Height 7.5” 7.5” 9.375” 9.375” 9.375” 9.875” 9.875”
Size Width 2.5” 2.5” 3.125” 3.125” 3.125” 3.125” 3.125”
Depth 1.4” 1.4” 1.625” 1.625” 1.625” 1.625” 1.625”
Jaw Opening 1” 1” 1.2” 1.2” 1.2” 1.7” 1.7”
Weight 8 oz. 8 oz. 11 oz. 11 oz. 11 oz. 11 oz. 11 oz.
Need Technical Assistance?
Call 1-800-44-FLUKE toll free in the U.S. or
(800) 36-FLUKE in Canada or (425) 446-5500
in other countries.
Fluke Corporation
PO Box 9090, Everett, WA USA 98206
Fluke Europe B.V.
PO Box 1186, 5602 BD
Eindhoven, The Netherlands
For more information call:
In the U.S.A. (800) 443-5853 or
Fax (425) 446-5116
In Europe/M-East/Africa (31 40) 2 675 200 or
Fax (31 40) 2 675 222
In Canada (800) 36-FLUKE or
Fax (905) 890-6866
From other countries (425) 446-5500 or
Fax (425) 446-5116
Web access: http://www.fluke.com/
©2001 Fluke Corporation. All rights reserved.
Printed in U.S.A. 9/2001 1629613 D-ENG-N Rev A
Fluke. Keeping your world
up and running.
Models 333, 334, 335,
336 and 337
Operating temperature
-14 °F to 122 °F
(-10 °C to 50 °C)
Altitude
2,500 meters
Battery life
(2) AA alkaline approximately
150 hours continuous, automatic low battery indicator
Safety rating
IEC 1010-2-031, 600 V CAT III
Temperature de-rating
Add 0.1 x specified accuracy for
each ºC above 28 ºC or below
18 ºC
Storage temperature
-40 ºF to 140 ºF
(-40 ºC to 60 ºC)
Warranty
Three years
Models 321 and 322
Operating temperature
-14 °F to 122 °F
(-10 °C to 50 °C)
Altitude
2,000 meters
Battery life
(2) AAA carbon zinc approximately 100 hours continuous,
automatic low battery indicator
Safety rating
IEC 1010-2-031, 600 V CAT III
Temperature de-rating
< 18 °C, > 28 °C add 0.1 x
specified accuracy / °C
Storage temperature
-22 °F to 140 °F (-30 °C to 60 °C)
Warranty