Portable Pressure
Calibrator
Features
• Measure and generate pressures from vacuum to 2 MPa (300 psi, 20 bar)
• Internal pump can generate vacuum to -80 kPa (-12 psi,
-0.8 bar) or pressure to 2 MPa (300 psi, 20 bar)
• Supply pressure connection allowing the use of external gas supply up to 2 MPa (300 psi, 20 bar)
• Includes variable volume for ne adjustment of pressures
• Electrical measurement and 24 volt supply for close looped calibrations
• Measure or generate 4 to 20 mA
• Measure 0 to 30 V dc
• Powered by internal, rechargeable, high capacity NiMH battery or universal ac mains adapter
• Compatible with Fluke 700P and 750P Pressure Modules
Technical Data
The 3130 Portable Pneumatic Pressure Calibrator is ideal for calibrating pressure transmitters, transduc­ers, gauges and similar devices. The 3130 contains everything you need to generate, control and measure pressure, as well as read the output of the device under test (DUT).
Pressure generation and control
The internal, motorized pump provides an easy, efficient alternative to time consuming, tiring hand pumps. For applications that involve filling a large volume with pressure, the 3130 allows for connec­tion to an external gas supply such as compressed plant air. The pressure can be fine-tuned using the variable volume.
Pressure measurement
The 3130 features an onboard pressure sensor with a full scale of 2 MPa (300 psi, 20 bar) and an accuracy of ± 0.025 % reading + 0.01 % FS (includes precision, one-year stability, uncertainty of the calibration standard, and temperature effects). In addition, the 3130 can be used with the Fluke 700P or 750P pressure module series to improve measure­ment performance across the entire range.
Electrical measurement
The 3130 offers electrical measurement capabilities for calibrating pressure transducers and transmitters, including measurements of 4 to 20 mA or 0 to 30 V dc. In addition, the 3130 provides 24 V dc supply to power the DUT and can generate current in the 4 to 20 mA range.
Portability
The 3130 is built into a ruggedized case with internal storage for power supply, test leads, and fittings. In addition, it includes a rechargeable, NiMH battery, allowing for approximately 50 hours of operation. When using the internal pump, the battery is sufficient to provide up to 100 calibration cycles to 300 psi. The battery is recharged by simply plugging the unit in using the included universal mains adapter.
Specications
Ordering Information
Models
3130-G2M Portable Pressure Calibrator
with US, EUR, UK, and China/ Australia line cords (standard calibration)
3130-G2M/C Portable Pressure Calibrator
with US, EUR, UK, and China/ Australia line cords (ISO 17025 accredited calibration)
3130-G2M-2 Portable Pressure Calibrator
with Brazil, Italy, Switzerland, and India line cords (standard calibration)
3130-G2M-2/C Portable Pressure Calibrator
with Brazil, Italy, Switzerland, and India line cords (ISO 17025 accredited calibration)
Environmental
Operating temperature -10 °C to +50 °C Storage temperature -20 °C to +60 °C Power requirements 12 V dc (Universal ac adapter/charger supplied) Battery Internal 3800 mAh advanced NiMH pack
Operating time
On full charge Approx 50 hours (Measure only or external air; no
Using internal pump Approx 100 calibration cycles to 300 psi
Physical
Dimensions 15.25 in L x 12 in W x 7 in D Weight ~7 kg (15 lb) EMI/RFI conformance EN61326:2006 Annex A Connectors/ports 1/8 in NPT (External supply port and test port) Included accessories Manual, NIST-traceable certificate, test leads, univer-
Ranges
Pressure (internal pump) -80 kPa to 2 MPa (-12 to 300 psi, -0.8 to 20 bar) Pressure (external air) 0 to 2 MPa (0 to 300 psi, 0 to 20 bar) mA 0 to 24.000 mA Volts 0 to 30.000 V dc Engineering units psi, bar, mbar, kPa, MPa, kgf/cm2, mmH2O @ 4 °C,
Instrumental measurement uncertainty
Pressure 0.025 % of reading ± 0.01 % FS mA 0.015 % of reading ± 0.002 mA Volts 0.015 % of reading ± 0.002 V
Temperature effect (all functions)
No effect on accuracy on all functions from 15 °C to 35 °C Add ± 0.002 % F.S./°C for temps outside of 15 °C to 35 °C
pump)
sal ac adapter/charger
mmH2O @ 20 °C, cmH2O @ 4 °C, cmH2O @ 20 °C, inH2O @ 4 °C, inH2O @ 20 °C, inH2O @ 60 °F, mmHg @ 0 °C, inHg @ 0 °C
2 Fluke Calibration 3130 Portable Pressure Calibrator
Fluke Calibration. Precision, performance, confidence.
Fluke Calibration PO Box 9090, Everett, WA 98206 U.S.A.
For more information call: In the U.S.A. (877) 355-3225 or Fax (425) 446-5116 In Europe/M-East/Africa +31 (0) 40 2675 200 or Fax +31 (0) 40 2675 222 In Canada (800)-36-FLUKE or Fax (905) 890-6866 From other countries +1 (425) 446-5500 or Fax +1 (425) 446-5116 Web access: http://www.flukecal.com
©2014 Fluke Calibration. This information is preliminary and subject to change. Printed in U.S.A. 2/2014 4265091C_EN Pub_ID: 12033-eng
Modification of this document is not permitted without written permission from Fluke Calibration.
Fluke Europe B.V.
PO Box 1186, 5602 BD Eindhoven, The Netherlands
Loading...