Fluke 190 Series II Technical Data

®
20 years of ScopeMeter
Test Tool Innovation
Introducing the complete
190 Series II
190 Series II ScopeMeter Oscilloscopes—the first high­performance scopes built for harsh industrial environments
Introducing the first high-performance portable oscilloscopes with 2 or 4 independently insulated input channels, an IP51 dust- and dripwater proof rating and a CAT III 1000 V/CAT IV 600 V safety rating. Choose from 500 MHz, 200 MHz, 100 MHz or 60MHz bandwidth models. Now plant maintenance engineers can take a 2- or 4-channel scope into the harsh world of industrial electronics.
The 190 Series II include these capabilities:
Up to four independent floating isolated inputs, up to 1000 V
Up to 5 GS/s real time sampling (Depending on model and channels used)
Deep memory: 10,000 points per trace waveform capture (scope mode)
CAT III 1000 V/CAT IV 600 V safety rated instrument for industrial
environments
Up to seven hours of battery operation using BP291
Isolated USB host port for direct data storage to a USB memory device;
USB device port for easy PC communication
Easy access battery door for quick battery swaps in the field
Compact and only 2.2 kg (4.8 lb)
Security slot: lock down oscilloscope with Kensington
unattended
IP 51 rating, dust- and drip-proof
Connect-and-View
slow and even complex signals
Frequency Spectrum using FFT-analysis
Automatic capture and REPLAY of 100 screens
ScopeRecord
frequency signal analysis
TrendPlot
paperless recorder mode with deep memory for long-term
automatic measurements
5,000 count DMM included in the 2-channel models
triggering for intelligent, automatic triggering on fast,
Roll mode gives 30,000 points per input channel for low
New
500 MHz
®
lock while
Technical Data
E
G
M
A
E
N
N
A
T
M
S
Y
Y
T
S
I
T
L
E
A
M
U
®
Q
C
E
R
T
1
0
0
9
I
F
O
I
E
S
I
D
T
T
E
O
E
M
Oscilloscope modes
190-062 190-102 190-202 190-502 190-104 190-204
Vertical deflection
Number of channels 2 2 2 2 4 4 Bandwidth 60 MHz 100 MHz 200 MHz 500 MHz 100 MHz 200 MHz Rise time 5.8 ns 3.5 ns 1.7 ns 0.7 ns 3.5 ns 1.7 ns Number of scope inputs 2 input channels plus external trigger 4 input channels Channel architecture All inputs fully insulated from each other and from ground
Input coupling AC or DC, with ground level indicator Input sensitivity 2 mV/div to 100 V/div, plus variable attenuation Bandwidth limiter User selectable: 20 kHz, 20 MHz or full bandwidth Normal/invert/variable On each input channel, switched separately
Input voltage CAT III 1000 V/CAT IV 600 V rated, see General Specifications for further details Vertical resolution 8 bit Accuracy ± (2.1 % of reading + 0.04 x range/div) @ 5 mV/div to 100 V/div Input impedance 1 MΩ ± 1 % // 14 pF ± 2 pF
Horizontal
Maximum real-time sample rate (sampled simultaneously)
Record length Up to 10,000 samples per channel Time base range 10 ns/div
Maximum record length 10,000 samples per channel in scope mode;
Timing accuracy ± (0.01 % of reading + 1 pixel) Glitch capture 8 ns peak detect on each channel
Display and acquisition
Display 153 mm (6 in) full-color LCD with LED backlight Display modes Any combination of channels; average on/off; replay Visible screen width 12 divisions horizontally in scope mode Digital persistence modes off/short/medium/long/infinite and envelope mode Waveform mathematics One mathematical operation on any 2 input channels: add/subtract/multiply; X-Y-mode
Acquisition modes Normal, Averaged, Auto, Single Shot, ScopeRecord™ roll, glitch capture,
Trigger and delay
Source Input A, B or External (via meter input) Input A, B, C or D Modes Automatic Connect-and-View
Connect-and-View Advanced automatic triggering that recognizes signal patterns, automatically sets up and continuously adjusts
Video triggering (on ch.A) NTSC, PAL, PAL+, SECAM; Includes field 1, field 2 and line select High-res, non-interlaced video Non-interlaced video with line-select, for line frequencies in the range 14 kHz up to 65 kHz Pulse width triggering
(on channel A) Time delay 1 full screen of pre-trigger view or up to 100 screens (=1,200 divisions) of post-trigger delay Dual slope triggering Triggers on both rising and falling edges alike N-cycle triggering Triggers on N-th occurrence of a trigger event; N to be set in the range 2 to 99
Inputs may be activated in any combination
625 MS/s for each channel
to 4 s/div Time base in a 1-2-4-sequence
Slower time/division settings using ScopeRecord
30,000 points per channel in ScopeRecord
(using real time sampling and data compression, at any timebase setting)
Frequency Spectrum using FFT analysis
waveform compare with automatic “Pass/Fail testing”; Replay
video, video line, selectable pulsewidth (channel A only), N-cycle
triggering, time base and amplitude. Automatically displays stable waveforms of complex and dynamic signals like motor drive and control signals. Can be switched off if preferred.
Pulse width qualified by time Allows for triggering <t, >t, =t, ≠ t, where t is selectable in minimum steps of 0.01 div or 50 ns
1.25 GS/s for each channel
5 ns/div to 4 s/div
, free run, single shot, edge, delay, dual slope,
2.5 GS/s (2ch) 5 GS/s (single
2 ns/div to 4 s/div
Roll mode (see ‘Recorder mode’)
channel) or 2.5 GS/s (on 2ch)
1 ns/div. to 4 s/div.
Roll mode (see ‘Recorder mode’)
1.25 GS/s for each channel
5 ns/div to 4 s/div
2.5 GS/s (2ch)
1.25 GS/s (4ch)
2 ns/div to 4 s/div
2 Fluke Corporation ScopeMeter 190 Series II2 Fluke Corporation ScopeMeter 190 Series II
Automatic capture of 100 screens
When in oscilloscope mode, the instrument ALWAYS memorizes the last 100 screens—no specific user setup required. When an anomaly is seen, the REPLAY button can be pressed to review the full sequence of screen events over and over. Instrument can be set up for triggering on glitches or intermittent anomalies and will operate in “baby-sit” mode capturing 100 specified events
Replay Manual or continuous replay. Displays the captured 100 screens as a “live” animation, or under manual
control. Each screen has date and time-stamp.
Replay storage Two sets of 100 screens each can be saved internally for later recall and analysis.
Direct storage of additional sets on external flash memory drive through USB host port.
FFT—frequency spectrum analysis
Shows frequency content of oscilloscope waveform using Fast Fourier Transform Window Automatic, Hamming, Hanning or None Automatic window Digitally re-samples acquired waveform to get optimum frequency resolution in FFT resultant Vertical scale Linear/Logarithmic (in volts or amps) Frequency axis Frequency range automatically set as a function of timebase range of oscilloscope
Waveform compare and pass/fail testing
Waveform Compare Provides storage and display of a reference waveform for visual comparison with newly acquired
Pass/Fail Testing In waveform compare mode, the oscilloscope can be set up to store only matching (“Pass”) or only
waveforms. Reference is derived from an acquired waveform and can be modified in the oscilloscope
non-matching (“Fail”) acquired waveforms in the replay memory bank for further analysis
Automatic scope measurements
V dc, V ac rms, V ac+dc, Vpeak max, Vpeak min, Vpeak to peak, A ac, A dc, A ac+dc, frequency (in Hz), risetime (using cursors), falltime (using cursors), Power Factor (PF), Watts, VA, VA reactive, phase (between any 2 inputs), pulsewidth (pos./neg.), dutycycle (pos./neg.), temperature °C, temperature °F (not for Japan), dBV, dBm into 50 I and 600 I, V and frequency inverters, V/Hz ration (190-xx2 only)
Advanced power and motor drive functions
V/Hz ratio (190-x02 only), Power Factor (PF), Watts, VA, VA reactive, V measurement on pulsewidth modulated motordrives and frequency inverters
PWM
ac and V
(ac+dc) for measurement on pulsewidth modulated motordrives
PWM
ac and V
PWM
(ac+dc) for
PWM
Advanced functions mA*s (current-over-time, between cursors);
V*s (voltage over time, between cursors); W*s (energy, between cursors)
Cursor measurements
Source On any input waveform or on mathematical resultant waveform (excl. X-Y-mode) Dual horizontal lines Voltage at cursor 1 and at cursor 2, voltage between cursors Dual vertical lines Time between cursors, 1/T between cursors (in Hz), voltage between markers, risetime with markers,
Single vertical line Min-Max and Average voltage at cursor position; frequency and rms-value of individual frequency
ZOOM Ranges from full record overview to zoom in up to sample level, at any record length
falltime with markers; Vrms between cursors, Watts between cursors
component in the FFT Resultant
ScopeMeter 190 Series II Fluke Corporation 3
Meter modes
190-062 190-102 190-202 190-502 190-104 190-204
Meter inputs Via 4 mm banana inputs, fully isolated from
Number of readings One at a time Up to 4 simultaneously Maximum resolution 5,000 counts 999 counts Input impedance 1 MΩ ± 1 % // 14 pF ± 2 pF Advanced meter functions Auto/manual ranging, relative measurements (Zero reference), TrendPlot
scope inputs and scope ground
The specified accuracy is valid over the temperature range 18 °C to 28 °C Add 10 % of specified accuracy for each degree C below 18 °C or above 28 °C
Voltage
V dc accuracy ± (0.5 % + 5 counts) ± (1.5 % + 5 counts) V ac true rms accuracy 15 Hz to 60 Hz: ± (1.5 % + 10 counts) 60 Hz to 1 kHz: 60 Hz to 20 kHz: ± (2.5 % + 15 counts) V ac+dc true rms accuracy 15 Hz to 60 Hz: ± (1.5 % + 10 counts) 60 Hz to 1 kHz: 60 Hz to 20 kHz: ± (2.5 % + 15 counts) Voltmeter ranges 500 mV, 5 V, 50 V, 500 V, 1,000 V
± (1 % + 10 counts) ± (2.5 % + 15 counts)
± (1 % + 10 counts) ± (2.5 % + 15 counts)
Resistance
Ranges 500 Ω, 5 kΩ, 50 kΩ, 500 kΩ, 5 MΩ, 30 MΩ — Accuracy ± (0.6 % + 5 counts)
Other meter functions
Continuity Beeper on < 50 Ω (± 30 Ω) — Diode test Up to 2.8 V — Current (A) A dc, A ac, A ac+dc using an optional current clamp or shunt
Temperature With optional accessories. Scale factors 1 °C/mV or 1 °F/mV
Scaling factors: 0.1 mV/A, 1 mV/A to 100 V/A and 400 mV/A
Via BNC scope inputs
recording
4 Fluke Corporation ScopeMeter 190 Series II
Recorder modes
190-062 190-102 190-202 190-502 190-104 190-204
ScopeRecord
Dual or multiple input waveform storage mode, using deep memory Source and display Input A, Input B, Dual.
Bandwidth 20 MHz or 20 kHz, user selectable Memory depth 30,000 data points, each holding min/max pair of information Min/max values Min/max values are created at samples that are measured at high sample rate
Recording modes Single sweep, continuous roll,
Stop-on-trigger ScopeRecord mode can be stopped by an individual trigger event, or by an interruption
Horizontal scale Time from start, time of day Zoom Ranges from full record overview to zoom in up to sample level, at any record length Memory Two multiple input ScopeRecord waveforms can be saved internally for later recall and analysis
ScopeRecord™ Roll mode sample rate and recording timespan
Time base range 5 ms/div ~ 2 min/div Recorded timespan 6 sec ~ 48 hr Time/division in ‘view all’ mode 0.5 s/div ~ 4 h/div Glitch capture 8 ns Sample rate 125 MS/s Resolution 200 μsec ~ 4.8 sec
Trendplot™ Recording
Multiple channel electronic paperless recorder. Graphically plots, displays and stores results of up to four automatic scope measurements or a DMM-reading over time.
Source and display Any combination of scope measurements, made on any of the input channels, or DMM reading
Memory depth 18,000 points (sets) per measurement. Each recorded sample point contains a minimum, a maximum and an
Ranges Normal view: 5 s/div to 30 min/div
Recorded time span Up to 22 days, with a resolution of 102 seconds Recording mode Continuous recording, starting at 5 s/div with automatic record compression Measurement speed 3 automatic measurements per second or more Horizontal scale Time from start, time of day Zoom Up to 64x zoom-out for full record overview, up to 10x zoom-in for maximum detail Memory Two multiple input TrendPlot records can be saved internally for later recall and analysis
Cursor measurements—all recorder modes
Source Any waveform trace in any waveform display mode (Scope, ScopeRecord or TrendPlot) Dual vertical lines Cursors may be used to identify Min, Max or Average value of any datapoint in a record,
Roll Mode
All channels sampled simultaneously.
ensuring capture and display of glitches
Start-on-Trigger (through external), Stop-on-Trigger (through external)
of a repetitive trigger signal, through any input channel (through External on 190-XX2 Series)
Direct storage on external flash memory drive through USB host port
(2-channel instruments)
average value, plus a date- and timestamp.
In view-all mode: 5 min/div to 48 hr/div (overview of total record)
Direct storage on external flash memory drive through USB host port
with time between cursors, time from start or absolute time.
Any combination of inputs, up to 4 channels. All channels sampled simultaneously.
Single sweep, continuous roll, Start-on-Trigger (through any channel), Stop-on-Trigger (through any channel)
ScopeMeter 190 Series II Fluke Corporation 5
General Specifications
190-062 190-102 190-202 190-502 190-104 190-204
Input voltage range
Rated maximum floating voltage CAT III 1000 V/CAT IV 600 V (maximum voltage between any contact and earth-ground voltage level) Probe input voltage VPS410 CAT III 1000 V/CAT IV 600 V (Maximum voltage between 10:1 probe tip and reference lead) Probe input voltage VPS510 CAT III 300 V (Maximum voltage between 10:1 probe tip and reference lead) Maximum BNC input voltage CAT IV 300 V (maximum voltage on BNC input directly) Maximum voltage
on meter input
Memory save and recall
Memory locations (internal) 30 waveform memories plus 10 recording memories plus 9 screen copy memories (190-XX, 2 channel
15 waveform memory locations Stores Scope-trace waveform data (2 or 4 traces each) plus screen-copy plus corresponding setup Two recording memories Each may contain:
External data storage
Screencopies
Volatility Measurement data is initially stored in RAM, which is maintained by the main battery with a 30 seconds
Real-time clock Provides date and time stamp information for ScopeRecord, for 100 Screen Replay sequences and
Case
Design Rugged, shock-proof with integrated protective holster. Handstrap and hangstrap included as standard
Drip and dust proof IP 51 according to IEC529 Shock and vibration Shock 30 g, vibration (sinusoidal) 3 g according to MIL-PRF-28800F Class 2 Display size 127 mm x 88 mm (153 mm/6.0 in diagonal) LCD Resolution 320 x 240 pixels Contrast and brightness User adjustable, temperature compensated Brightness 200 cd/m
Mechanical data
Size 265 mm x 190 mm x 70 mm (10.4 in x 7.5 in x 2.8 in) Weight (including battery) 2.1 kg (4.6 lb) 2.2 kg (4.8 lb)
Power
Line power Mains adapter/battery charger BC190 included, version depending of country Battery power Re-chargeable double capacity Li-Ion battery (included). Battery swappable through
Battery type (incl.) and capacity [+opt. battery]
Battery charge indicator Battery has built-in status indicator for use with external charger,
Battery operating time (with backlight low)
Battery charging time 2½ hours using BP290; 5 hours using BP291 Five hours BP291 Battery power saving functions Auto ‘power down’ with adjustable power down time; Auto ‘Display off’ with adjustable power down time;
Safety
Compliance EN61010-1-2001, Pollution Degree 2;
CAT III 1000 V/CAT IV 600 V (safety designed banana input connectors)
models); 15 waveforms memories plus 2 recording memories plus 1 screen copy memory (190-XX, 4 channel models)
a 100 Screen Replay sequence, or
• a ScopeRecord Roll-mode recording (2 or 4 traces), or
• a TrendPlot recording of up to 4 measurements
On PC, using FlukeView Software, or
Direct storage on external flash memory drive (maximum 2 GB) through USB host port On PC, using FlukeView Software, or
Internally (in instrument) which can be copied on to external flash memory drive as .BMP-file, through USB host port
back-up when battery is exchanged When storing data, this is written in non-volatile flash-ROM
for TrendPlot recordings
Kensington lock supported to lock down instrument when left unattended
2
typ. using power adapter, 90 cd/m2 typical using battery power
easily accessible battery door at the rear of the instrument BP290; 2400 mAh
[BP291 (4800 mAh) optional]
next to battery status indicator on instrument screen Up to four hours using BP290 (included),
Up to eight hours using BP291 (optional)
On-screen battery power indicator
CAN/CSA C22.2, No. 61010-1-04, with approval; UL61010B; ANSI/ISA-82.02.01
BP291; 4800 mAh
Up to seven hours using BP291 (included)
6 Fluke Corporation ScopeMeter 190 Series II
190-062 190-102 190-202 190-502 190-104 190-204
Environmental
Operating temperature 0 °C ~ +40 °C; +40 °C ~ +50 °C excl. battery Storage temperature -20 °C ~ +60 °C Humidity +10 °C ~ +30 °C: 95
+30 °C ~ +40 °C: 75 +40 °C ~ +50 °C: 45
Maximum operating altitude Up to 2,000 m (6666 ft) for CAT IV 600 V, CAT III 1000 V;
% RH non-condensing; % RH non-condensing; % RH non-condensing
up to 3,000 m (10,000 ft) for CAT III 600 V, CAT II 1000 V Maximum storage altitude 12 km (40,000 ft) Electro-Magnetic-
EN 61326 (2005-12) for emission and immunity Compatibility (EMC)
Interfaces Two USB-ports provided. Ports are fully insulated from instrument’s floating measurement circuitry USB-host
port directly connects to external flash memory drive (up to 2 GB) for storage of waveform data, complete
datasets in which data and setup information is included, instrument settings and screen copies
A mini-USB-B is provided which allows for interconnection to PC for remote control and data transfer under
PC-control Probe calibration output Dedicated probe-cal output with reference contact provided,
fully insulated from any measurement input channel Warranty Three years (parts and labor) on main instrument, one year on accessories
Included accessories
Battery charger/mains adapter BC190 Li-Ion battery pack BP290 (2400 mAh) BP291 (4800 mAh) Voltage probe sets. Each set
includes ground lead, hook clip, ground spring and probe tip insulation sleeve.
Test leads TL175 (one red, one black) with test pins (N/A) Voltage Probes VPS410-x: each set includes: Ground lead, hook clip, ground spring and probe tip insulation sleeve.
Other Li-Ion battery (BP290 or BP291, see above); Battery charger (BC190); Hangstrap; Handstrip
VPS410 (one red, one blue) VPS410 (one red, one grey, one
blue, one green)
VPS510-x: each set includes: Ground lead, hook clip, ground spring, probe tip insulation sleeve
and BNC-to probe tip adapter.
(user selectable for left- or right hand use); Multi language users manuals on CD-ROM;
FlukeView
®
demo package (with restricted functionality); USB interface cable for PC connectivity.
ScopeMeter 190 Series II Fluke Corporation 7
Ordering information
Models
Fluke 190-502 Color ScopeMeter, 500 MHz, 2 channels plus DMM/Ext.input Fluke 190-502/S
Fluke 190-204 Fluke 190-204/S
Fluke 190-104 Fluke 190-104/S
Fluke 190-202 Fluke 190-202/S Color ScopeMeter, 200 MHz, 2 channels plus DMM/Ext.input,
Fluke 190-102 Fluke 190-102/S Color ScopeMeter, 100 MHz, 2 channels plus DMM/Ext.input,
Fluke 190-062 Fluke 190-062/S Co
Accessories
BC190 Mains adapter/battery charger BP290 BP291 EBC290
HH290 VPS510-R VPS510-G VPS510-B Electronic Voltage Probe set, 10:1, 500 MHz, one set blue VPS510-V VPS410-R VPS410-G VPS410-B VPS410-V Industrial Voltage Probe set, 10:1, one set green VPS420-R
SW90W C290 SCC290
TL175 TRM50 AS400 RS400 RS500
Co
lor ScopeMeter, 500 MHz, 2 channels plus DMM/Ext.input,
with SCC-290 kit included
lor ScopeMeter, 200 MHz, 4 channels
Co
Co
lor ScopeMeter, 200 MHz, 4 channels, with SCC-290 kit
included
Co
lor ScopeMeter, 100 MHz, 4 channels
Co
lor ScopeMeter, 100 MHz, 4 channels, with SCC-290 kit
included
Co
lor ScopeMeter, 200 MHz, 2 channels plus DMM/Ext.input
with SCC-290 kit included
Co
lor ScopeMeter, 100 MHz, 2 channels plus DMM/Ext.input
with SCC-290 kit included
Co
lor ScopeMeter, 60 MHz, 2 channels plus DMM/Ext.input lor ScopeMeter, 60 MHz, 2 channels plus DMM/Ext.input,
with SCC-290 kit included
-ion battery pack, 2400 mAh
Li
-ion battery pack, 4800 mAh
Li
ternal battery charger for BP290 and BP291 (uses BC190
Ex mains adapter)
nging Hook for 190 Series II instruments
Ha
ectronic Voltage Probe set, 10:1, 500 MHz, one set red
El
ectronic Voltage Probe set, 10:1, 500 MHz, one set grey
El
ectronic Voltage Probe set, 10:1, 500 MHz, one set green
El
dustrial Voltage Probe set, 10:1, one set red
In
dustrial Voltage Probe set, 10:1, one set grey
In Industrial Voltage Probe set, 10:1, one set blue
gh working voltage ruggedized probe set, 100:1, 150 MHz
Hi (bicolored, red/black)
ukeView ScopeMeter Software package (full version)
Fl
rd shell protective carrying case for 190 Series II
Ha
ukeView ScopeMeter Software package (full version)
Fl and C290 Carrying Case kit for 190-series II
istGuard™ safety designed Test Leads set (1 red, 1 black)
Tw
C Feedthrough 50 I terminator (set of 2 pieces, black)
BN
obe Accessory Extension Set for VPS400-series probes
Pr
obe Accessory Replacement Set for VPS400-series probes
Pr
obe Accessory Replacement Set for VPS500-series probes
Pr
8 Fluke Corporation ScopeMeter 190 Series II
Loading...