Fluke 18-250 User Manual

Page 1
Nuclear Associates 18-250
Digital Stereotactic Breast Biopsy Accreditation Phantom
Users Manual
Page 2
Fluke Biomedical
Radiation Management Services
6045 Cochran Road Cleveland, Ohio 44139
440.498.2564
www.flukebiomedical.com/rms
Page 3
Table of Contents
Section 1: General Information................................................................................... 1-1
1.1 Introduction .................................................................................................. 1-1
1.2 Accreditation Images on Recumbent or Upright Digital Stereo Biopsy Units 1-2
i
Page 4
(Blank Page)
Page 5
General Information
Introduction
Section 1
General Information
1.1 Introduction
Digital Mammography Evaluation Phantom, Model 18-250*
Provides an easy way to test image quality on digital biopsy mammography units and qualify for
ACR accreditation.
Contains the same objects that are embedded in the Mammographic Accreditation Phantom (Model
18-220), which is also specified by the ACR.
Has an extended top edge that allows for easy positioning on recumbent biopsy units.
Enables the user to determine if the images are similar to or better than screen film.
Is fast, convenient and easy to use.
1
Model 18-250. Digital Mammography Evaluation Phantom
* Designed by Carol Mount, R.T. ® (M), and Joel E. Gray. Ph.D., Department of Diagnostic Radiology, Mayo Clinic, Rochester,
MN 55905. Manufactured under licensing agreement with Mayo Foundation for Medical Education and Research.
Model 18-250. Digital Image Demonstrating Image Evaluation Insert
1-1
Page 6
Nuclear Associates 18-250
Operators Manual
1.2 Accreditation Images on Recumbent or Upright Digital Stereo Biopsy Units
1. Center the phantom at the chest wall edge of the image receptor. Place the compression paddle in contact with the phantom to secure it.
2. If phototiming is available, make a phototimed exposure.
3. If the unit does not have phototiming, manually select the kVp and mAs routinely used to radiograph the Mammographic Accreditation Phantom (Model 18-220).
4. Make the exposure.
5. Analyze the image on the monitor.
6. To pass accreditation with the Digital Mammography Evaluation Phantom, you must be able to see
3 fibers, 3 groups of specks, and 2.5 masses.
7. Record your image on disk or film.
8. In accordance with the American College of Radiology, this test should be done weekly.
Wax Insert
Fibers
A. 0.93 mm nylon fiber E. 0.54 mm speck I. 0.25 mm (thickness) mass B. 0.74 mm nylon fiber F. 0.32 mm speck J. 0.50 mm (thickness) mass C. 0.54 mm nylon fiber G. 0.24 mm speck K.0.75 mm (thickness) mass D. 0.32 mm nylon fiber H. 0.20 mm speck L. 1.0 mm (thickness) mass
Specks Masses
1-2
Page 7
(Blank Page)
Page 8
Fluke Biomedical
Radiation Management Services
6045 Cochran Road Cleveland, Ohio 44139
440.498.2564
www.flukebiomedical.com/rms
Loading...