Fluke 1732, 1734 User's Manual

Manual Supplement
Manual Title: 1732/1734 Users Supplement Issue: 3 Print Date: February 2017 Issue Date: 7/19 Revision/Date: 1, 7/17 Page Count: 9
©2017-2019 Fluke Corporation. All rights reserved. x
1732/1734 Users Manual Supplement
Change #1, 491
On page 6, replace Figure 1 with:
Figure 1. Adapter Installation
7/19 1
hcf069.eps
Manual Supplement 1732/1734 Users
Change #2, 594
On page 16, replace the Measurement Line Power Source section with:
Measurement Line Power Source:
XWWarning
To prevent injury, do not touch the metal parts of one test lead when the other is still connected to hazardous voltage.
WCaution
To prevent damage to the Product, make sure the measured voltage does not exceed the input rating of the power supply.
1. Attach the power supply to the Logger.
2. Move the slide-cover on the power supply to access the safety
sockets.
3. Connect the short test leads (see Figure 7B & 7C) with the power supply inputs. Make sure to use the non-stackable plugs. The test leads are rated for measurement/overvoltage CAT III 1000 V and CAT IV 600 V.
4. Connect the test leads with the voltage measurement inputs:
Connect A/L1 with one input of the power supply.
Connect N with the second input of the power supply.
OR
Connect A/L1 with one input of the power supply.
Connect B/L2 with the second input of the power supply.
5. Use the short fan out of the Voltage Test Lead, 3-phase + N. Plug the connector A/L1 into the socket A/L1 of the voltage measurement inputs of the Logger. Repeat this with B/L2, C/L3 and N.
2 7/19
Loading...
+ 7 hidden pages