Fluke 114, 115, 117 Data Sheet

Fluke 114, 115, 116 and 117
Digital Multimeters
Extended specifications
Technical Data
General specifications (all models)
Accuracy is specified for 1 year after calibration, at operating temperatures of 18 °C to 28 °C, with relative humidity of 0 % to 90 %.
Maximum voltage between any terminal and earth ground
Surge protection 6 kV peak per IEC 61010-1 600 V CAT III, Pollution Degree 2 W Fuse for A input 11 A, 1000 V FAST 17 kA Fuse (Fluke PN 803293) Display Digital: 6,000 counts, updates 4/sec; Bar Graph: 33 segments, updates 32/sec Temperature Operating: -10 °C to +50 °C; Storage: -40 °C to +60 °C Humidity 0 % to 90 % to 35 °C; 75 % to 40 °C; 45 % to 50 °C Temperature coefficient 0.1 x (specified accuracy/°C) (< 18 °C or > 28 °C) Operating altitude 2,000 meters Battery 9 Volt Alkaline, NEDA 1604A/IEC 6LR61 Battery life Alkaline: 400 hours typical, without backlight Safety compliances ANSI/ISA 82.02.01 (61010-1) 2004, CAN/CSA C22.2 No 61010-1-04, UL 6101B (2003) and IEC/EN
Certifications UL, CSA, TUV, N10140 , VDE IP rating (dust and water protection) IP42
61010-1 2nd Edition for measurement Category III, 600 V, Pollution Degree 2, EMC EN61326-1
Accuracy specifications (all models)
Function Range Resolution
DC millivolts 600.0 mV 0.1 mV 0.5 % + 2 114, 115, 116, 117 DC volts 6.000 V
Auto-V LoZ1 true-rms 600.0 V 0.1 V 2.0 % + 3 4.0 % + 3 114, 116, 117
AC millivolts1 true-rms 600.0 mV 0.1 mV 1.0 % + 3 2.0 % + 3 114, 115, 116, 117 AC volts1 true-rms 6.000 V
Continuity
Ohms 600.0 W
Diode test 2.000 V 0.001 V 0.9 % + 2 115, 116, 117 Capacitance 1000 nF
LoZ capacitance (power-up option) 1 nF to 500 µF 10 % + 2 typical 115, 116, 117
1
All ac ranges except Auto-V LoZ are specified from 1 % to 100 % of range. Auto-V LoZ is specified from 0.0 V. Because inputs below 1 % of
range are not specified, it is normal for this and other true-rms meters to display non-zero readings when the test leads are disconnected from a circuit or are shorted together. For volts, crest factor of 3 at 4000 counts, decreasing linearly to 1.5 at full scale. For amps, crest factor of 3. AC volts is ac-coupled. Auto-V LoZ, ac mV, and ac amps are dc-coupled.
60.00 V
600.0 V
60.00 V
600.0 V 600 W 1 W
6.000 kW
60.00 kW
600.0 kW
6.000 MW
40.00 MW
10.00 µF
100.0 µF 9999 µF
0.001 V
0.01 V
0.1 V
0.001 V
0.01 V
0.1 V
0.1 W
0.001 kW
0.01 kW
0.1 kW
0.001 MW
0.01 MW
1 nF
0.01 µF
0.1 µF 1 µF
Accuracy ± ([% of Reading]
+ [Counts]) Model
0.5 % + 2 114, 115, 116, 117
DC, 45 Hz to
500 Hz
45 Hz to 500 Hz 500 Hz to 1 kHz
1.0 % + 3 2.0 % + 3 114, 115, 116, 117
Beeper on < 20 W, off > 250 W;
detects opens or shorts of
100 µF to 1000 µF: 1.9 % + 2
> 1000 µF: 5 % + 20
500 Hz to 1 kHz
500 µs or longer
0.9 % + 2
0.9 % + 1
0.9 % + 1
0.9 % + 1
0.9 % + 1
1.5 % + 2
1.9 % + 2
1.9 % + 2
1.9 % + 2
114, 115, 116, 117
114, 115, 116, 117
115, 116, 117
Accuracy specifications (all models) cont.
Function Range Resolution
Temperature (K-Type thermocouple) -40 °C to 400 °C
AC amps true-rms1 (45 Hz to 500 Hz) 6.000 A
-40 °F to 752 °F
10.00 A 20 A overload for 30 seconds maximum
0.1 °C
0.2 °F
0.001 A
0.01 A
Accuracy ± ([% of Reading]
+ [Counts]) Model
1 % + 10 1 % + 18
2
2
1.5 % + 3 115, 117
116
AC µAmps true-rms1 (45 Hz to 1 kHz) 600.0 µA 0.1 µA 1.5 % + 3 (2.5 % + 3 > 500 Hz) 116 DC amps 6.000 A
10.00 A 20 A overload for 30 seconds maximum
0.001 A
0.01 A
1.0 % + 3 115, 117
DC µAmps true-rms 600.0 µA 0.1 µA 1.0 % + 2 116 Hz (V or A input)
Hz (V input)
1
All ac ranges except Auto-V LoZ are specified from 1 % to 100 % of range. Auto-V LoZ is specified from 0.0 V. Because inputs below 1 % of range
are not specified, it is normal for this and other true-rms meters to display non-zero readings when the test leads are disconnected from a circuit or are shorted together. For volts, crest factor of 3 at 4000 counts, decreasing linearly to 1.5 at full scale. For amps, crest factor of 3. AC volts is ac-coupled. Auto-V LoZ, ac mV, and ac amps are dc-coupled.
2
AC Volts Hz is ac-coupled and specified from 5 Hz to 50 kHz. AC Amps Hz is dc-coupled and specified from 45 Hz to 5 kHz. Amps input burden
voltage (typical): 6 A input 2 mV/A, 10 A input 37 mV/A.
3
Frequency is ac-coupled, 45 Hz to 50 kHz.
2
3
99.99 Hz
999.9 Hz
9.999 kHz
50.00 kHz
99.99 Hz
999.9 Hz
9.999 kHz
50.00 kHz
0.01 Hz
0.1 Hz
0.001 kHz
0.01 kHz
0.01 Hz
0.1 Hz
0.001 kHz
0.01 kHz
0.1 % + 2 115, 117
0.1 % + 2 116
Frequency counter sensitivity (models 115, 116, 117)
Typical sensitivity (rms sine wave)
Input range 5 Hz to 45 Hz 45 Hz to 5 kHz 5 kHz to 10 kHz 10 kHz to 50 kHz
Volts AC 6 V
AC Amps (115, 117 only)
60 V
6 A
10 A
0.2 V 2 V
20 V
N/A N/A
0.2 V to 0.3 V 2 V to 3 V
20 V to 30 V
0.4 A
0.5 A
0.3 V to 0.4 V 3 V to 4 V
30 V to 40 V
N/A N/A
0.4 V to 1.0 V 4 V to 10 V
40 V to 100 V
N/A N/A
Input characteristics (all models)
Function
Volts AC > 5 MW < 100 pF > 60 dB at dc, 50 or 60 Hz
Volts DC > 10 MW < 100 pF > 100 dB at dc, 50 or 60 Hz > 60 dB at 50 or 60 Hz
Auto-V LoZ ~3 kW < 500 pF > 60 dB at dc, 50 or 60 Hz
Ohms < 2.7 V dc
Diode Test < 2.7 V dc 2.000 V dc < 1.2 mA
Input impedance
(nominal) Common mode rejection ratio(1 kW unbalanced) Normal mode rejection
Open circuit test voltage Full scale voltage Short circuit current
To 6.0 MW 40 MW
< 0.7 V dc < 0.9 V dc
Fluke. Keeping your world up and running.
< 350 µA
For more information call: In the U.S.A. (800) 443-5853 or Fax (425) 446-5116 In Europe/M-East/Africa +31 (0) 40 2675 200 or
Fluke Corporation PO Box 9090, Everett, WA USA 98206
Fluke Europe B.V. PO Box 1186, 5602 BD Eindhoven, The Netherlands
Fax +31 (0) 40 2675 222 In Canada (800)-36-FLUKE or Fax (905) 890-6866 From other countries +1 (425) 446-5500 or Fax +1 (425) 446-5116 Web access: http://www.fluke.com
©2006 Fluke Corporation. All rights reserved. Printed in U.S.A. 11/2006 2793260 D-EN-N Rev A