5.5 Replacement Parts ...................................................................................... 5-3
ii
Page 5
Introduction
Product Description
Section 1
Introduction
1.1 Product Description
The Model 1060 is a smart radiation detector. It is a printed circuit board that supports a wide variety of
"front end" radiation detectors and provides electronic digital and analog information about the radiation
environment to users.
The Model 1060 collects, interprets, analyzes, stores and communicates radiation detector data from
virtually all pulse generating radiation detectors such as Geiger Mueller tubes, proportional tubes, boron
tri-fluoride tubes, and photo-multiplier tube/scintillators. To support these radiation detectors the 1060
provides a wide range high voltage suitable to operate most radiation detectors.
In addition, it will simultaneously accept up to 8 analog and 16 digital inputs.
The 1060 provides digital outputs, and is a good citizen on a network. It operates independently or as a
member of a larger system.
1
It is configured by the user through serial software commands, either locally or remotely.
It defaults upon power up or reset to predetermined operational characteristics that can be established
and altered by the operator. Such defaults are stored in a non-volatile memory.
The Model 1060 is only 2.8 in x 8.8 in (7.11 cm x 22.29 cm) in size and therefore requires a very small
space for mounting. It can be mounted in a variety of enclosures for many different environments.
It provides, via digital communication, raw and conditioned data from any or all of the inputs. It performs a
wide variety of calculations on raw data, such as conversion to engineering units, and boxcar or
exponential smoothing of data.
Optionally, it can be programmed to make local decisions based upon input data. The available area
monitor models are shown below:
Clear Acrylic Enclosure
Model No.
1060AM-CL-ER Environmental range - 1 μR to 1000 (μR/hr (SI units)
1060AM-CL-LR Low range - 0.01 mR to 1000 mR/hr (SI units)
1060AM-CL-MR Medium range - 0.1 mR to 10,000 mR/hr (SI units)
1060AM-CL-HR High range - 1 mR to 100,000 mR/hr (SI units)
1060MHV-CL Area Monitor with external MHV probe connection
Description
1-1
Page 6
Victoreen 1060AM
Operators Manual
NEMA 4X Enclosure
Model No.
1060AM-NM-ER Environmental range - 1 μR to 1000 (μR/hr (SI units)
1060AM-NM-LR Low range - 0.01 mR to 1000 mR/hr (SI units)
1060AM-NM-MR Medium range - 0.1 mR to 10,000 mR/hr (SI units)
1060AM-NM-HR High range - 1 mR to 100,000 mR/hr (SI units)
1060MHV-NM-HR Area Monitor with external MHV probe connection
Description
1.2 Specifications
1.2.1 Digital Specifications
On-board ComputerIntel 80C196, 16 Bit, 16 MHz, Embedded CMOS Microcontroller 1 Serial Ports
3 asynchronous modes
Up to 187.5 K Baud @ 12 MHz
1 Buffer for input/output
HSI high-speed input capture Interrupt on either rising, falling or a transition
2 Timers
Timer 1, Free running, based on XTAL
Timer 2, Incremented by T2CLK or HSI.1
512 Bytes of RAM
8 Channel, 10 Bit, ADC w/ Sample and Hold Available at PCB connector
for external inputs. Two channels are dedicated to jam detection
and detector temperature sensing.
Extended temperature range -25°C to +85°C
MemoryPROM OTP (One-Time-Programmable) 16K X 16, (256K) Program/Data
48K X 16, (768K) Paged
RAM Static RAM, 64K X 16 (1Mb)
EEPROM Serial IK X 16, for conversion factors, setpoints, historical data (stored
upon loss of communications only), etc.
Serial Communications Dual UART: Two RS-485 and one RS-232
The RS-485 is the primary communications link to data acquisition system. The
RS-232 is a local diagnostics, local data logging, and set-up port.
Status Indicators 8 LEDs indicating operating status (i.e., communications in progress, normal
operation, high voltage ON/OFF)
Switches8 DIP switches used to initiate various local functions (i.e., high voltage, on/off,
etc.)
Digital Outputs16 latched TTL outputs
1-2
Page 7
Introduction
Specifications
Digital Inputs16 digital inputs with internal pull-up resistors
1
1.2.2 Analog Specifications
High Voltage Regulated 500 to 2500 VDC, < 1 mV ripple, digitally controlled with 1 V resolution
Max
500 microamps at 1400 V
Input Circuitry: Pulse mode input range 3 mV to 7 V suitable for BF3 detectors, proportional
counters, Geiger Mueller tubes, and photo multiplier tubes used with scintillators.
Polarity selection: (PMT input only)
High Disc., 0 - 7.50 VDC; set via software
Low Disc., 0 - 1.230 VDC; set via software
Jam detection (anti-jam)
1.2.3 Miscellaneous Specifications
Power Requirements Approximately 12 VDC @ 200 mA - Nominal Voltage range is 7 VDC (minimum)
to 16 VDC (maximum). With special adapter can accommodate 24 VDC.
EnclosuresTwo standard enclosures are available (all dimensions in inches unless other-
wise indicated):
An acrylic plastic tube housing for indoor applications.
Outside DimensionsMounting Hole Pattern
P/N
1060AM-CL-xx 3.5 dia x 10 -
(8.89 x 25 cm)
A plastic rectangular housing for NEMA 4 applications (outdoor or indoor).
P/N
(9.65 x 28.26 x 10.16 cm) (7.46 x 23.81 cm)
User Interface Via any IBM compatible Personal Computer
Model 1060 Circuit
Board Size Approximate size: 2.8 in x 8.775 in (7.11 x 22.29 cm) with mounting holes on
Auxiliary Voltage An auxiliary voltage of +/- 12 VDC @ 50 mA, non-inductive load, is available to
Temperature/Humidity
Range The humidity range is 5% to 95% rh, non-condensing over the entire temperature
Shock and Vibration Mechanical shock and vibration specifications are per ANSI N42.17A, section 8.4
Outside DimensionsMounting Hole Pattern
1060-NM-xx 3.8 x 11.125 x 4.0 2.9375 x 9.375
2.4 in x 8.375 in (6.1 x 21.27 cm) centers
provide power to preamplifiers.
range of 32° to 122°F (0° to 50°C).
and 8.5.
Software Specifications
Language "C" The software written in Intel "iC96", Version X243
Operating System Real-time, interrupt driven, embedded system
1-3
Page 8
Victoreen 1060AM
Operators Manual
System OperationAs part of a larger system, the 1060 will communicate via serial link to a host
computer. In this configuration, the 1060 will recognize its system address and
respond to commands.
Main Software ModulesThe main software modules include the following:
INITIALIZE
RC
MAIN
APPL.H
The Intel libraries include:
fpal 96, stdlib, KR, string
stdio, printf, C 96,
Software Quality
Assurance The 1060 software follows the quality assurance guidelines set forth in NRC
regulatory policies and IEEE standards for Software:
ANSI/IEEE Std 1016-1987 Software Design Specification
The Institute of Electrical and Electronic Engineers. "Software Engineering
Standards", c 1987.
ANSI/ANS-10.4-1987, "Guidelines for the Verification and Validation of Scientific
and Engineering Computer Programs for the Nuclear Industry"
IEEE Std 175-185 "Floating Point Mathematics".
Specific FunctionsThe software can perform the following functions:
1. Count pulses from a detector (analog /digital)
2. Calculate the average Counts Per Unit Time
3. Apply dead-time correction
4. Compensate for background radiation
5. Compensate for pressure (requires analog input information)
6. Compensate for flow (requires analog input information)
7. Convert to engineering units
8. Maintain all operating parameters, and data buffers in the event of a power
failure
9. Respond to serial communications from supervisory processors:
• Provide raw detector input
• Provide uncompensated radiation values
• Provide compensated radiation values
• Provide CPS, CPM, CPI
• Provide historical averages
• Provide for user alterable parameters
• Perform self-calibrations and store relevant parameters
• Provide information in engineering units
1-4
Page 9
Introduction
Specifications
• Provide digital input and analog inputs signals, such as for pressure and
flow
10. Responds to data entry switches.
11. Updates indicators, and digital and analog outputs, and responds to
serial communications within 250 mSec.
12.Task and prioritize Interrupts
1.3 Detector Types
The radiation detected and energy response are a function of the radiation detectors selected. The four
most common types are Geiger Mueller tubes used to detect: x-rays, gamma rays, and beta rays;
Proportional tubes used to detect alpha particles, and beta rays, BF
photomultiplier tubes/scintillators used to detect x-rays and gamma rays.
The standard detectors are listed as follows:
1.3.1 Geiger Mueller Tubes
Model 1B85 Geiger Mueller tubes are routinely supplied for environmental monitoring. These tubes are
rugged, relatively large in size, very sensitive to radiation, and are thin walled for enhanced low energy
sensitivity.
tubes to detect neutrons, and
3
1
Range Name Range in mR/hr CPS/mR/hr
P/N
380-49 Low .002 – 200 36.67
3
90-12 Low BKGND - 10
20
IE qualified Geiger Mueller tubes have demonstrated their reliability in use in various Nuclear Facilities.
The tubes supplied are summarized as follows:
Range Name Range in mR/hr CPS/mR/hr
P/N
3
857-210-20 Low 0.01 - 10
857-220-20 Medium 0.1- 10
857-230-20 High 1.0- 10
8.0
4
1.2
5
0.53
Many other custom Geiger Mueller tubes are available, upon special request.
14
C Beta 0.158 59 5.3
l33
Ba Gamma 0.356 72 6.5
129
I Beta/Gamma 0.029 49 4.4
1.3.2 BF3 Tubes
These tubes are used to detect neutrons, but must be configured into a detector assembly with the
appropriate moderator and attenuator. The part number for this detector assembly is RP-N. In the neutron
detector, the moderator/attenuator assembly is a polyethylene cylinder 9.5 in x 8.5 in diameter containing
proportional counter and neutron energy compensating materials. This arrangement is based upon
a BF
3
the standard reliable Snoopy designs for neutron energy response.
The operating characteristics of the BF
The BF
proportional counter operates at 1150 V.
3
Active length: 2 in (5.08 cm)
Gas pressure: 20 cm Hg
Fill Gas: enriched BF
3
tube are as follows:
3
, 96% Boron 10
1-5
Page 10
Victoreen 1060AM
Operators Manual
Resolving time: 1 microsecond
Plateau Slope: 2% per 100 V
10
Tube Life expectancy: >10
counts
Neutron Sensitivity: 2000 counts/mRem (nominal)
The neutron Energy Response is as follows:
The energy range is thermal 0.025 eV to 15 MeV. The accuracy is within 10% of the theoretical ICRP
dose rate.
The directionality effect is less than +/- 20% in 3 orthogonal directions.
The standard ranges and engineering units available are:
Rate: 1 μRem/hr to 75 Rem/hr 0 μSv/hr to .75 Sv/hr
0 μRad/hr to 9 Rad/hr 0 μGy/hr to .09 Gy/hr
0 CPM to 2.5 x 106 CPM 0 CPS to 41,660 CPS
Integrate: 0 μRem to 1000 Rem 0 μSv to 10 Sv
0 μRad to 125 Rad 0 μGy to 1.25 Gy
9
0 to 10
The tube is insensitive to gamma rays and shows no response in
Counts 0 to 109 Counts
l37
Cs gamma radiation in fields up to
500 mR/hr.
The overall dimensions of the assembly are: 12.50 in diameter and overall length of 10.25 in. The
assembly comes with two mounting brackets. The total weight of the assembly is 21 lbs.
The detector assembly operating range is -80°C to +80°C
The detector can be placed up to 10 ft (3 m) away from the Model 1060.
1.3.3 Photomultiplier Tubes/Scintillators
The standard scintillator is a Nal(T1) crystal on a photomultiplier tube. There are two standard detectors
configurations, however, many others models are available. Please see the Model 943 series detectors.
Nominally, the calibration would set the high voltage to provide pulse heights of 2mV/keV.
Specifications
Crystal Nal(T1)
Crystal size 1.5 in x 1.0 in
Housing Stainless Steel
End window Aluminum
Temperature range 32° - 122°F, 0° - 50°C
Extended range 32° -150°F, 0° - 65°C
FWHM less than 9%
Energy Range 70 keV to 3.0 MeV
Weight 3 lbs (1.4kg)
PMT 2 in photocathode, 10 stage
Max voltage 1400 Volts
1-6
Page 11
Victoreen 1060AM
Operators Manual
The Model 1060-xxx scintillation detector does not contain an integral preamplifier, the Model 943-36
does.
1.4 Receiving Inspection
Upon receipt of the package:
1. Inspect the cartons (s) and contents for damage. If damage is evident, file a claim with the carrier
and notify Fluke Biomedical, Radiation Management Services at 440.248.9300
2. Remove the contents from the packing material.
3. Verify that all items listed on the packing list have been received and are in good order.
1.5 Storage
If the unit is to be stored prior to use, pack it in the original container if possible, and store in an
environment free of corrosive materials, fluctuations in temperature, and humidity, and vibration and
shock.
1.6 Procedures, Warnings and Cautions
The equipment described in this manual is intended to be used for the detection and measurement of
ionizing radiation. It should be used only by persons who have been trained in the proper interpretation of
its readings and the appropriate safety procedures to be followed in the presence of radiation.
Although the equipment described in this manual is designed and manufactured in compliance with all
applicable safety standards, certain hazards are inherent in the use of electronic and radiometric
equipment.
Warnings and Cautions are presented throughout this document to alert the user to potentially
hazardous situations. A Warning is a precautionary message preceding an operation that has the
potential to cause personal injury or death. A Caution is a precautionary message preceding an operation
that has the potential to cause permanent damage to the equipment and/or loss of data. Failure to comply
with Warnings and Cautions is at the user's own risk and is sufficient cause to terminate the warranty
agreement between Fluke Biomedical and the customer.
Adequate warnings are included in this manual and on the product itself to cover hazards that may be
encountered in normal use and servicing of this equipment. No other procedures are warranted by Fluke
Biomedical. It shall be the owner's or user's responsibility to see to it that the procedures described here
are meticulously followed, and especially that Warnings and Cautions are heeded. Failure on the part of
the owner or user in any way to follow the prescribed procedures shall absolve Fluke Biomedical and its
agents from any resulting liability.
Indicated battery and other operational tests must be performed prior to each use to assure that the
instrument is functioning properly. If applicable, failure to conduct periodic performance tests in
accordance with ANSI N323-1978 (R1983) Radiation Protection Instrumentation Test and Calibration,
paragraphs 4.6 and 5.4, and to keep records thereof in accordance with paragraph 4.5 of the same
standard, could result in erroneous readings or potential danger. ANSI N323-1978 becomes, by this
reference, a part of this operating procedure.
1-6
Page 12
Victoreen 1060AM
Operators Manual
(Blank Page)
Page 13
Installation/Replacement
Installation
Section 2
Installation/Replacement
2.1 Installation
The device power must be turned OFF during installation/replacement to prevent possible damage.
2
Many components on the printed circuit board are
static sensitive. ESD precautions should be
observed when handling the printed circuit board
assembly.
If readings from a device's position indicates that
the radiation rate is relatively high, CAUTION
should be observed when replacing or working on
this unit. High rates can quickly cause the
operator's cumulative exposure to increase with the
potential for injury.
There is a potential shock hazard when the Model
1060 board is exposed and the unit is operating. Be
especially careful not to touch any components in
the high voltage area, or connections to the
detector.
CAUTION
CAUTION
CAUTION
2.1.1 Connections - 1060AM-NM (NEMA 4 Enclosure)
The following information lists the I/O connections for the Model 1060AM-NM, NEMA 4 Enclosure:
J1 Power connector (3-pin Male)
1 +12 VDC
2 GND
3 Chassis/Earth GND
J2 RS-232 Connector (5-pin Female)
1 Tx Data
2 Rx Data
3 DTR
4 DSR
5 Signal Ground
2-1
Page 14
Victoreen 1060AM
Operators Manual
J3 RS-485 Connector (4-pin Female)
J4 RS-485 Connector (4-pin Female)
1 A
2 B
3 GND
4 Termination (120Ω)
RS-485 - In multi-drop applications, it may be necessary to terminate the input furthest from the
computer system. To terminate the communications lines, install a jumper between Pin 2 and Pin
4.
J5 I/O Connector (6-pin Female)
1 4-20mA
2 GND
3 Meter +
4 GND
5 Alarm Lamp
6 GND
Figure 2-1. Contact Arrangements (View from wiring side/user connections)
2.1.2 Connections - 1060AM-CL Family
The following information lists the I/O connections for the Model 1060AM-CL family (clear acrylic cylinder
housing):
J1 Power connector (2mm Pin)
Center Pin +12 VDC
Sleeve GND
J2 RS-232 Connector (DB9F) (From 1060)
2 Tx Data
3 Rx Data
4 DTR
5 GND
6 DSR
J3 & J4 RS-485* Connector (RJ11 Modular)
1 Not Used
2 Not Used
3 CHAN 3-A
4 CHAN 3-B
*RS-485: In multi-drop applications, it may be necessary to terminate the input furthest from the
computer system. To terminate the communications lines, install a jumper on the 1060 PCB -J2 Pins 12 &
2-2
Page 15
Installation/Replacement
Installation
13. Some commercial RS-485 to RS-232 converters require the external communications device (1060)
to be connected A to A and B to B. Other units may require A to B and B to A. Ensure the correct
connection is made.
J5 I/O Connector (6-pin Female)
1 4-20 mA
2 GND
3 Meter +
4 GND
5 Alarm Lamp
6 GND
2
Figure 2-2. Contact Arrangements (View from wiring side/user connections)
2.2 Interface Requirements
The Model 1060 is designed to communicate remotely with a host computer and locally with a laptop
computer. The protocol of that communication is described as follows.
The communication interface to the model 1060 is serial, balanced, full or half-duplex, RS-485 signals
with 8 data bits, no parity, 1 stop bit at 9600 baud rate.
Computers with an RS-232 communication port(s) can be converted toRS-485 via adapters to permit
remote communication with multi-dropped Model 1060s.
2.3 Enclosures
See enclosure specifications, for size of standard enclosures and dimensions for mounting.
2.4 Detector Setup and Control
The model 1060, when supplied with a detector is factory setup for proper operation. The detector
parameters, found on the Calibration Data Sheet, are entered into the 1060 to provide a calibrated
system. Some applications may involve the replacement of the detector, or the use of a detector from
another manufacturer. Under these circumstances, caution must be exercised to prevent damage to the
detector and/or the 1060.
2-3
Page 16
Victoreen 1060AM
Operators Manual
2.4.1 Input Channel Selection
The Select Input Channel command (SIC) allows one of the four input processing circuits to be activated
as the input to the pulse processing circuitry. The following codes are used to select the active channel.
Input Channel Hex Entry TLA
GM input C 68
PMT with preamp input
Positive pulse A 42
Negative pulse A 43
Proportional D 50
Neutron input D 70
PMT w/o preamp B 44
2.4.2 High Voltage Adjustment
The Set High Voltage (SHV) command allows the user to adjust the high voltage for the appropriate
detector. Photo multiplier tube/scintillation detectors generally require that the overall gain be set by the
high voltage. The high voltage setting affects the input pulse amplitude and for this reason controls the
calibration of the detector. A detector/1060 provided as a system is factory calibrated. Typical 943 family
of detectors are set to obtain approximately 1 V pulse amplitude/0.5 MeV. When adjusting the high
voltage, caution must be exercised so as not to exceed the rating of the detector.
2.4.3 Low Discriminator Adjustment
The Set Low Discriminator (SLD) command controls the voltage threshold above which input pulses will
be processed. Input pulses with an amplitude less than the low discriminator setting will not be counted.
The typical setting is approximately 0.200V or 100 KeV.
2.4.4 High Discriminator Adjustment
The Set High Discriminator (SHD) command controls the voltage threshold below which input pulses will
be processed. Input pulses with amplitude that exceeds the high discriminator setting will not be counted.
The typical setting is approximately 7.00 V or 3.5 Mev.
2.4.5 Using other detectors
Typically, the high voltage setting and the low discriminator setting are used to calibrate the system.
These values may be found on the Calibration Data Sheet, or through the manufacturer of the detector.
Follow the manufacturers recommended calibration procedure when calibrating the system. Caution must
be exercised when setting the high voltage so as not to
exceed the detector rating.
Figure 2-3 illustrates the typical multi-drop interface block
diagram.
Figure 2-3. Block Diagram
2-4
Page 17
Operation
Communication Protocol
3
Section 3
Operation
3.1 Communication Protocol
Communication Port Command Entry Syntax
RS-232 Communication Port 1:
This port (9-pin connector) is used for local communication.
Commands from the computer are of the form:
CMD<cr>
Responses from the 1060 are of the form:
RSP<cr>
where: CMD = TLA (Three Letter Acronym) command
RSP = Response from 1060
<cr>= Carriage Return
RS-485 Communication Port 2 and Port 3:
These ports are used in multi-drop configuration; commands must be preceded with the 1060 unit ID:
ID<sp><CMD><cr>
where ID=1060 ID, <sp>Space, CMD one listed below, <cr>=Carriage Return
Responses from the 1060 with matching ID are of the form:
<ID><sp><RSP><cr>
where ID=1060 ID, <sp>Space, RSP=Response to command, <cr>=Carriage Return
3.2 Terminal Operation
Command Set
The following COMMAND SET is comprised of STANDARD, SETUP and RADIATION commands.
STANDARD commands are always available to all communication ports. SETUP commands (marked "s")
are used to configure the 1060 and are made available through Switch S2 as follows:
Port 1 (RS-232), S2, Position 2 ON = Setup commands available.
Port 2 (RS-485), S2, Position 3 ON = Setup commands available.
Port 3 (RS-485), S2, Position 4 ON = Setup commands available.
RADIATION commands (marked "r") are used to poll the 1060 for specific radiation data.
To enter Setup mode, enter the command ESM (Switch 2 must be ON). This will enable the SETUP and
disable the RADIATION commands. To exit Setup mode, enter XSM. This will disable the SETUP and
enable the RADIATION commands.
3-1
Page 18
Victoreen 1060AM
Operators Manual
3.2.1 COMMAND SET
The COMMAND SET is comprised of STANDARD, SETUP, TEST, and RADIATION commands.
STANDARD commands are always available to all communication ports.
SETUP commands (marked "s") are used to configure the 1060 and are made available through switch
S2 as follows:
Port 1 (R-S232), S2, Position 2 ON = Setup commands available
Port 2 (RS-485), S2, Position 3 ON = Setup commands available
Port 3 (RS-485), S2, Position 4 ON = Setup commands available
SETUP MODE disables RADIATION commands.
SETUP MODE cannot be entered while Pulse Output is active.
TEST commands (marked "t") are used to test parts of the hardware.
At this time, only the hardware that is exercised is the analog output.
TEST MODE and SETUP MODE can both be active.
TEST MODE can be entered while Pulse Output is active.
RADIATION commands (marked "r") are used to poll the 1060 for specific radiation data.
To enter Setup mode, enter the command ETM. This will enable the TEST and disable the RADIATION
commands. The analog output driver is disabled and the user can drive the output via commands ANS,
DAC, and LOG (see the following COMMAND SET list). To exit TEST mode, enter XTM. This will disable
the TEST and enable the RADIATION commands. The analog output driver is enabled.
ANS t Analog out per Counts per Second entered
CPH r Counts per Hour (read counts averaged over 60 min)
CPM r Counts per Minute (read counts averaged over 60 sec)
CPS r Counts per Second (read latest second count value)
DAC t DAC POT output per numeric entry 0 to 255
ESM r Enter Setup Mode (enable Setup Commands, disable Radiation
Commands) Pulse output cannot be active
ETM t Enter Test Mode (enable Test Commands and Radiation Commands)
Analog meter is controlled by Test Commands.
LOG t Drive analog meter numeric input, i.e. LOG 10, needle moves to position 10.
MHV Measure High Voltage through Analog to Digital converter)
RxH r Counts per Hour (using Conversion Factor x, where x=0 to 9)
RxM r Counts per Minute (using Conversion Factor x, where x=0 to 9)
RxS r Counts per Second (using Conversion Factor x, where x=0 to 9)
RAB Reset Averaging Buffers (Resets counting buffers)
RCx Read Conversion Factor x, where x=0 to 9
RCR Read Conversion Factor CPM to mR/Hr
RDI Read Digital Input (Reads 16 digital inputs)
RDO Read Digital Output (Reads 16 digital outputs)
RGT Read GM Threshold setting from EEPROM
RHD Read High Discriminator Voltage setting from EEPROM
RHS Read High Scale analog output (shown in units of mR/hr.)
RHV Read High Voltage (Reads HV setting from EEPROM)
RIC Read Input Control byte (detector input channel)
RID Read 1060 unit Identification
RLD Read Low Discriminator Voltage setting from EEPROM
RLS Read Low Scale analog output (shown in units of mR/hr.)
RRH r Read mR/Hr (averaged over 60 minutes)
RRM r Read mR/Hr (averaged over 60 seconds)
3-2
Page 19
RRS r Read mR/Hr (latest second mR/Hr value)
RSH r Read Sieverts (averaged over 60 minutes)
RSM r Read Sieverts (averaged over 60 seconds)
RSS r Read Sieverts (latest second Sievert value)
RTA Read TAU (deadtime)
SCx s Set Conversion Factor x, where x = 0 to 9
*SCR s Set Conversion Factor CPM to mR/Hr (enter CPM per mR/hr)
SDO Set Digital Output (Sets 16 digital outputs)
SFD s Set output Format Decimal (integer)
SFE s Set output Format Engineering (default, x.xxExx)
SFH s Set output Format Hex (integer, 16 hex digits)
SGT s Set GM Threshold (Sets Thresh., stores in EEPROM)
SHD s Set High Discriminator (Sets Hi Discrim., stores in EEPROM)
SHS s Set High Scale analog output (entered in units of mR/hr)
SHV s Set High Voltage (Sets HV, stores setting in EEPROM)
SIC s Set Input Control
SID s Set unit Identification
SLD s Set Low Discriminator (Sets Lo Discrim., stores in EEPROM)
SLS s Set Low High Scale analog output (enter in units of mR/hr)
*STA s Set TAU (deadtime correction);
VER Read firmware Version and Revision level
XSM s Exit Setup Mode
XTM t Exit Test Mode
NOTE
Commands SCR and STA must be used to enter
the CPM to mR/hr conversion factor and deadtime
correction. These values must be entered
whenever a new detector is installed. These values
can be found on the detector calibration sheet.
8. ON = Terminal vs Computer communication format (Reserved)
3.5 Data Output
The standard communication is a block of 8-bit ASCII characters. Standard network protocol is defined in
Section 3.1. Terminal data protocol and commands are defined in Section 3.2.
3.6 Operational Checkout
The LEDs indicate normal operation and are defined in Section 4.1.10. A visual review of their status will
provide a quick check of operability of the Model 1060.
3.7 Counts
3.7.1 Counts per second
The number of counts from the detector is accumulated for 0.1 seconds. Ten of these intervals are
summed to arrive at a Counts Per Second (CPS) total.
3.7.2 Counts per minute
Each Counts Per Second (CPS) value is stored in a buffer location. Sixty CPS values are summed to
arrive at a Counts Per Minute (CPM) total.
3.7.3 Counts per hour
Each Counts Per Minute (CPM) value is stored in a buffer location. Sixty CPM values are summed to
arrive at a Counts Per Hour (CPH) total.
3.7.4 Counts in mR/Hr
Counts in mR/Hr are provided using CPM to mR/Hr conversion factor.
3.7.5 Counts in Sieverts
Counts in Sieverts are provided using mR/Hr to Sieverts conversion factor.
3.7.6 Counts per Correction Factor
Counts per correction factor are provided using correction factors 0 through 9.
3-4
Page 21
Theory of Operation
General
4
Section 4
Theory of Operation
4.1 General
4.1.1 Detector Input
The Model 1060 has four separate input terminal points, one for each type of detector: Geiger Mueller
tube, proportional tube, BF
circuits that condition the signal pulse from the detector. The conditioned pulse is then input into the
"Clock2" input of the Intel 80C196 microprocessor. The microprocessor counts both the rising edge and
the trailing edge of the signal pulse.
4.1.2 High Voltage Supply
The Model 1060 takes the +5 VDC power and utilizes a switching power supply and a step up transformer
to produce a range of voltages from 500 to 2500 VDC. A 1000:1 divider provides feedback for regulation
of the high voltage. The high voltage is digitally controlled with a resolution of 1 V. A 12-bit analog to
digital converter provides the high voltage setting to the microprocessor. The microprocessor has the
ability to completely shut off the high voltage.
tube, and photomultiplier/scintillator. These points are inputs to analog
3
4.1.3 Auxiliary DC Power
A +/- 12 VDC power is provided via a switching power supply from the +12 unregulated input power. This
voltage is available at an edge connector to provide power for detectors that contain preamplifiers. The
output of this supply is 50 mA. A non-inductive load is recommended.
4.1.4 Jam Detection
The low discriminator output is buffered and charges a capacitor through an RC network. This voltage is
then monitored by one of the 8 analog input channels of the microprocessor. At a predetermined voltage
threshold corresponding to a high-count rate, approximately equal to twice the detector maximum, and/or
approaching continuous discharge, the microprocessor will shut down the high voltage and indicate that a
jam condition has occurred.
4.1.5 Thresholds
An upper and lower threshold provide an energy window for pulses. The microprocessor will only
recognize those pulses within the magnitude established by the thresholds. The range of the lower
threshold is from 0 to 1.25 V.
The range of the upper threshold is from 0 to 7 V and must be higher than the lower threshold. The upper
threshold may be disabled if desired. Both thresholds are under software control. The resolution of the
upper and lower thresholds are 25 mV and 5 mV respectively.
4.1.6 Analog inputs
There are 8 analog inputs, but two are reserved, one for jam detection and one for local temperature measurement,
usually, the detector. The other 6 are made available to the user on an edge connector. The input range is 0 to 5 V DC,
with 10-bit resolution. Typical uses for these inputs are for process transducers. All are read by software.
4-1
Page 22
Victoreen 1060AM
Operators Manual
4.1.7 Digital inputs
There are 16 TTL level digital inputs with pull up resistors. They may be used to indicate the state of
various process conditions. All are read by software.
4.1.8 Digital outputs
There are 16TTL level digital outputs under software control. They may be used as input signals to drive
process equipment.
4.1.9 Communication
The software transmits parallel data from the microprocessor to the Dual UART that converts it to serial
data and controls the transmission to external devices. Two RS-485 serial channels are provided. These
channels are for multi-drop communication. There is one RS-232 port for local communication.
4.1.10 Engineering Units
The software will convert raw counts to any engineering unit that the Model 1060 has been programmed
and calibrated for. Raw counts as well as converted units can be combined to provide appropriate
engineering units.
4.1.11 Error Analysis
From the sensitivity of the detectors and the Gaussian distribution statistical model for radioactive decay
measurements, the measurement error can be calculated directly as a function of exposure rate and the
counting time for each detector.
For this distribution the standard deviation is the square root of the number of counts ( =√Total Counts).
The total number of counts is equal to the exposure rate times the tube characteristic times the total
counting time (Total Counts = Exp. Rate x Tube Char. x Time). The total counts is the mean or expected
measurement.
It is also given for Gaussian distributions that 99.73% of the actual measurements will fall within the mean
plus or minus three standard deviations (μ ± 3s). This is then chosen as the error range.
From the above discussion the accuracy of any detector can be computed.
4.1.12 Minimum Detectability
Minimum detectability, in the presence of a radiation background, is usually calculated at the 50% and
95% confidence level. A good first approximation to each is as follows: First, determine the mean and
standard deviation of the background. Look up the sensitivity of the detector to the nuclide. Factor in any
geometry efficiency. The 50% confidence level is the activity level, accounting for geometry and
sensitivity, of the nuclide at the mean plus three standard deviations of the background. For the 95%
confidence level use the mean plus 4.645 standard deviations of the background.
4.2 Functional Description
To be provided later.
4.3 Circuit Description
To be provided later.
4-2
Page 23
Calibration, Maintenance and Troubleshooting
Calibration
5
Section 5
Calibration, Maintenance and Troubleshooting
5.1 Calibration
5.1.1 Precautions
Only trained and qualified personnel operating within the protocols of an approved radiation protection
program should be permitted to calibrate any radiation detector.
5.1.2 Exposure
The Model 1060 is calibrated at the factory against NIST traceable standards. The calibration is detector
specific.
Calibrations other than to standard nuclides are available upon request. Model 1060s will always provide
raw counts, but if calibrated, will also provide engineering unit information.
5.2 Maintenance
The instrument power should be turned OFF any time that the instrument is opened to prevent possible
damage.
Many components on the printed circuit board are
static sensitive. ESD precautions should be
observed when handling the printed circuit board
assembly.
If readings from a device's position indicates that
the radiation rate is relatively high, CAUTION
should be observed when replacing or working on
this unit. High rates can quickly cause the
operator's cumulative exposure to increase with the
potential for injury.
There is a potential shock hazard when the Model
1060 board is exposed and the unit is operating. Be
especially careful not to touch any components in
the high voltage area, or connections to the
detector.
CAUTION
CAUTION
CAUTION
5-1
Page 24
Victoreen 1060AM
Operators Manual
5.3 Corrective Maintenance
5.3.1 Board Replacement
To be provided later.
5.3.2 Detector Replacement
To be provided later.
5.3.3 Circuit Troubleshooting
To be provided later.
5.4 Troubleshooting
Many components on the printed circuit board are
static sensitive. ESD precautions should be
observed when handling the printed circuit board
assembly.
If readings from a device’s position indicates that
the radiation rate is relatively high, CAUTION
should be observed when replacing or working on
this unit. High rates can quickly cause the
operator’s cumulative exposure to increase with the
potential for injury.
There is a potential shock hazard when the Model
1060 board is exposed and the unit is operating. Be
especially careful not to touch any components in
the high voltage area, or connections to the
detector.
380-49 1B85 GM tube for Environmental
857-210-20 Low Range 1E rated GM tube
857-220-20 Medium Range rated 1E GM tube
857-230-20 High Range 1E rated GM tube
90-12 Energy Compensated GM Probe
PR-N Neutron Detector Assembly
1060-xxx NaI(T1) Detector
943-36 NaI (T1) Detector with Preamp
Replacement Parts
5
5-3
Page 26
Fluke Biomedical
Radiation Management Services
6045 Cochran Road
Cleveland, Ohio 44139
440.498.2564
www.flukebiomedical.com/rms
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.