Fluke 91 Series II, 92 Series II, 96 Series II, 105 Series II User Manual

Page 1
Date 950401
© 1995 Fluke Corporation, All rights reserved. All product names are trademarks of their respective companies.
Fluke91/92/96/105
ScopeMeter Series II
Users Manual
i
Page 2
TABLE OF CONTENTS
CHAPTER 1 INTRODUCING YOUR SCOPEMETER TEST TOOL
SCOPEMETER TEST TOOL FEATURES
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2
USING THE HOLSTER AND THE TILT STAND
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-3
POWERING THE SCOPEMETER TEST TOOL
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-4
MINIMIZING SIGNAL NOISE
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-5
CHARGING THE BATTERY
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-6
LOOKING AT ALL MEASUREMENT CONNECTIONS
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-7
READING THE DISPLAY
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-8
USING THE KEYS
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-10
STEPPING THROUGH A WINDOW
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-12
USING ON-LINE INFORMATION
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-13
CHAPTER 2 GETTING STARTED
ADJUSTING THE DISPLAY
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-3
SELECTING THE PROBE TYPE
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-4
POWER-ON CONFIGURATIONS
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-4
PERFORMING AN EASY SETUP
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-5
USING THE DEMONSTRATION (DEMO) BOARD
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-6
CHAPTER 3 MAKING MEASUREMENTS
MAKING CONNECTIONS
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-2
USING THE MEASUREMENT FUNCTIONS
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-3
Page 3
II
91/92/96/99/105
Users Manual
USING A SHORTCUT TO MOST COMMONLY MEASUREMENTS
. . . . . . . . . . . . . . . . . . . . . . . . . . . 3-9
ORDERING MEASUREMENT READINGS
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-10
MEASURING FUNCTIONS NOT AVAILABLE FROM THE MEASURE MENU
. . . . . . . . . . . . . . . . . 3-10
USING THE MEASUREMENTS MAP
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-11
CHAPTER 4 USING THE DUAL DISPLAY MODE FUNCTIONS
MAKING CONNECTIONS
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-2
SELECTING A MAIN MODE
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-3
SELECTING RANGES (MANUAL/AUTO RANGE)
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-7
HOLDING A STABLE MEASUREMENT (TOUCH HOLD
)
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-8
DISPLAYING MINIMUM AND MAXIMUM READINGS WITH RELATED TRENDPLOT
. . . . . . . . . . . 4-9
SELECTING THE SCOPEMETER KEY SUBMENU
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-11
MAKING MEASUREMENTS IN V AND EXT.mV MODE
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-12
TAKING RELATIVE READINGS (SCALING)
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-12
CHAPTER 5 USING IN SCOPE MODE
MAKING CONNECTIONS
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-2
MAKING AN EASY SETUP
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-3
CONTROLLING INPUTS A AND B
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-3
ADJUSTING THE AMPLITUDE
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-4
ADJUSTING THE TIME BASE
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-5
POSITIONING THE WAVEFORM ON THE DISPLAY
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-6
ACQUIRING WAVEFORMS
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-7
SELECTING THE SCOPE SUBMENU
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-9
TRIGGERING
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-11
DISPLAYING THE MINIMUM AND MAXIMUM OF A WAVEFORM
. . . . . . . . . . . . . . . . . . . . . . . . . . 5-14
MAKING COMBINATIONS WITH INPUT A AND INPUT B
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-15
Page 4
III
DISPLAYING MULTIPLE WAVEFORMS
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-15
ADAPTING THE DISPLAY
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-16
MORE INFORMATION
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-17
CHAPTER 6 USING ADDITIONAL CAPABILITIES
MAKING MEASUREMENTS USING THE CURSORS
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-2
DELETING MEMORIES
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-6
SAVING TO MEMORY
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-7
RECALLING FROM MEMORY
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-8
USING WAVEFORM MATH FUNCTIONS
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-11
USING A PRINTER
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-14
USING THE WAVEFORM GENERATOR
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-17
TESTING COMPONENTS
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-17
CHANGING THE RESET CONFIGURATIONS
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-19
ALTERING THE CONTINUOUS AUTO SET CONFIGURATION
. . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-20
CHAPTER 7 MEASURING EXAMPLES
MEASURING TEMPERATURE
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-2
MEASURING CURRENT
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-4
MEASURING POWER WITH MATH FUNCTION
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-7
MEASURING THREE-PHASE ON A DUAL INPUT
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-9
MEASURING PHASE USING THE CURSORS
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-12
MEASURING PULSE RESPONSE OF AN AMPLIFIER
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-14
Page 5
IV
91/92/96/99/105
Users Manual
CHAPTER 8 SCOPEMETER TEST TOOL TUTORIAL
TESTING POTENTIOMETERS
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-4
MAKING LOW FREQUENCY VOLTAGE MEASUREMENTS WITH THE METER
. . . . . . . . . . . . . . . 8-5
ASSISTING CONTINUOUS AUTO SET ON COMPLEX WAVEFORMS
. . . . . . . . . . . . . . . . . . . . . . 8-10
MAKING LOW FREQUENCY MEASUREMENTS WITH THE SCOPE
. . . . . . . . . . . . . . . . . . . . . . . 8-17
MAKING SINGLE SHOT MEASUREMENTS WITH THE SCOPE
. . . . . . . . . . . . . . . . . . . . . . . . . . . 8-20
FINDING HIDDEN WAVEFORM DETAILS WITH THE SCOPE
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-22
MAKING SCOPE CURSOR MEASUREMENTS
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-26
USING THE MATH FUNCTIONS
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-32
CHAPTER 9 USER MAINTENANCE
CLEANING
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-2
KEEPING BATTERIES IN OPTIMAL CONDITION
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-2
REPLACING AND DISPOSING OF BATTERIES
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-3
REPLACING FUSES
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-5
CALIBRATING THE PROBES
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-5
CHAPTER 10 APPENDIXES
10A SPECIFICATIONS
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-3
10B PARTS AND ACCESORIES
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-11
10C PM8907 INFORMATION
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-15
10D RECOMMENDED PRINTER SETUP PARAMETERS
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-17
10E WARRANTY AND SERVICE CENTERS
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-19
10F TERMINOLOGY
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-25
10G MENU MAP
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-31
Page 6
V
Page 7
VI
91/92/96/99/105
Users Manual
ABOUT THIS MANUAL
Chapter 1 Introducing the ScopeMeter Test Tool
This chapter introduces features and capabilities of your ScopeMeter test tool.
Chapter 2 Getting Started
This chapter provides a 15-minute demonstration intended for those who are not familiar with the ScopeMeter test tool.
Chapter 3 Making Measurements
This chapter explores all measurements and specifies the use of the direct Measure Menu key. At any time, you can choose over 30 measurements to get an immediate reading on the display.
Chapter 4 Using the Dual Display Mode
This chapter addresses the use of four Dual Display Mode Functions of the ScopeMeter test tool: , , , and
. You will learn how to set up the test tool for ac and dc voltage measurements, diode tests, resistance measurements, and trend plotting. Chapter 5 contains all the Scope mode information.
Chapter 5 Using the Scope Mode
This chapter explores the specific use of the test tool as a digital storage oscilloscope. You will learn how to make, store, and compare measurements.
Chapter 6 Using Additional Capabilities
This chapter explores the additional capabilities available with your ScopeMeter test tool. The chapter covers subjects, such as Scope measurements using the cursors, saving and recalling screens, waveforms, or setups, and making a hard copy on your printer.
Chapter 7 Measuring Examples
This chapter outlines step-by-step procedures necessary to make some typical measurements.
Chapter 8 ScopeMeter Test Tool Tutorial
This chapter looks at special capabilities of your ScopeMeter test tool. You will become familiar with all aspects by using the demonstration board.
Chapter 9 User Maintenance
This chapter describes the cleaning of the ScopeMeter test tool and proper use and replacement of the battery pack. Periodic probe calibration is also covered here.
Page 8
VII
Chapter 10 Appendixes
A. Specifications: Operating characteristics. B. Parts and Accessories: Model numbers and replacement
codes for all parts and accessories delivered with your
ScopeMeter test tool. C. PM8907 Power Adapter/ Battery Charger. D. Advised Printer Setup Parameters: How to set up your
printer. E. Warranty Information and Service Centers: Warranty
terms and Service Center addresses. F. Terminology: Glossary of special terms. G. Menu Map: Graphic view of the various menus.
Index
The index at the end of the manual lists words alphabetically. Consult this list to find an item quickly.
NOTE
Throughout this manual an trough a model number
()
indicates that the function being discussed does not
apply to that specific model number.
91
Page 9
VIII
91/92/96/99/105
Users Manual
WARNING
READ "SAFETY" CAREFULLY BEFORE USING YOUR SCOPEMETER TEST TOOL.
SAFETY
The instrument described in this manual is designed to be used only by qualified personnel.
Safety Precautions
To use this instrument safely, it is essential that operating and servicing personnel follow both generally accepted safety procedures and the safety precautions specified in this manual. Specific warning and caution statements, where they apply, will be found throughout the manual. Where necessary, the warning and caution statements and/or symbols are marked on the instrument.
A CAUTION identifies conditions and actions that may damage the test tool.
A WARNING IDENTIFIES CONDITIONS AND ACTIONS THAT POSE HAZARD(S) TO THE USER.
International electrical symbols used are explained below.
The terms "Isolated" or "Electrically floating" are used in this manual to indicate a measurement in which the ScopeMeter test tool COM (common, also called ground) is connected to a voltage different from earth ground. The term "Grounded" is used in this manual to indicate a measurement in which the ScopeMeter test tool COM (common) is connected to an earth ground potential.
The ScopeMeter test tool COM (common) inputs (red INPUT A BNC shield, grey INPUT B BNC shield, and black 4-mm banana jack) are connected internally via self­recovering fault protection. The input connectors have no exposed metal and are fully insulated to protect against electrical shock. The black 4-mm banana jack COM (common) can be connected to a voltage above earth ground for isolated (electrically floating) measurements
Caution (see explanation in manual)
DOUBLE INSULATION (Protection Class)
Common (Lo) input symbol, equal potential
Recycling symbol
High BNC input symbol
DC-Direct Current
Earth AC-Alternating
Current
Page 10
IX
and is rated up to 60 0V rms above earth g roun d. The ScopeMeter test tool uses a three-lead connection sys-
tem for dual input, isolated (electrically floating) measure­ments. The connections for isolated and grounded measurements are shown in the following illustration.
Figure 1 Common (Ground) Connections
NOTE
Fluke 91 has a two lead connection system.
Page 11
X
91/92/96/99/105
Users Manual
WARNING
DO THE FOLLOWING TO AVOID ELECTRICAL SHOCK IF A SCOPEMETER TEST TOOL
COM (COMMON)
INPUT IS CONNECTED TO >42V PEAK (30V RMS):
1. USE ONLY THE TEST LEAD/PROBE SET SUPPLIED WITH THE SCOPEMETER TEST TOOL (OR SAFETY-DESIGNED EQUIVALENTS WITHOUT EXPOSED METAL CONNECTORS).
2. DO NOT USE CONVENTIONAL EXPOSED METAL BNC OR BANANA PLUG CONNECTORS IF THE
COM (COMMON) IS >42V PEAK (30V RMS).
3. USE ONLY ONE COM (COMMON) CONNECTION (THE 4-MM BLACK BANANA JACK).
4. REMOVE ALL PROBES AND TEST LEADS THAT ARE NOT IN USE.
5. USE 600V RATED PROBE TIP ADAPTERS. "600V" IS PRINTED ON EQUIPMENT SO RATED.
6. CONNECT THE PM8907 POWER ADAPTER TO THE
AC OUTLET BEFORE CONNECTION TO THE SCOPEMETER TEST TOOL.
Page 12
XI
If Safety-Protection is Impaired Use of equipment in a manner not specified may
impair the protection provided by the equipment. Before use, inspect probes for mechanical damage and replace damaged probes!
Whenever it is likely that safety has been impaired, the instrument must be turned off and disconnected from line power. The matter should then be referred to qualified personnel. Safety is likely to be impaired if, for example, the instrument fails to perform the intended measurements or shows visible damage.
MAKING ISOLATED MEASUREMENTS
Isolating from Earth Ground
The ScopeMeter test tool uses a three-lead connection system for dual input, isolated (electrically floating) measurements. Use a test lead connected to the black 4-mm banana jack as common ground.
There are only three wires to connect for dual input isolated ground operation, useful for three-phase industrial voltage measurements. You can connect the three leads on the test tool (INPUT A, INPUT B, and common) to a three­phase electrical power line in any order. Use the A-B mode (see Chapter 5) to display all three phases of electrical power at the same time.
For three-phase electrical or other isolated ground industrial measurements, we strongly recommend that you use the black Industrial Alligator Clip and the red and grey Industrial Probe Alligator Clips (see Figure 4). These parts are included with original purchase ScopeMeter test tools and are also available as accessories from your FLUKE representative.
Isolating from Input to Input
The ScopeMeter test tool is not isolated between inputs but uses the black 4-mm banana jack for measurements isolated from earth ground.
Complete isolation between INPUT A and INPUT B is useful for some floating measurement applications, such as measuring different isolated transformer windings at the same time.
For those applications requiring true
isolated grounds (commons) between inputs, we recommend that you make two separate measurements or use two ScopeMeter test tools.
91
Page 13
XII
91/92/96/99/105
Users Manual
PREPARING TO USE THE DEMONSTRATION (DEMO) BOARD
A Demo Board with 9V battery is included in your ScopeMeter test tool kit. It generates different types of waveforms that can be measured at test points TP1 to TP5. All descriptions and measurement examples refer to a waveform that can be measured on the Demo Board. Do not connect the Demo Board yet, but when you make measurements, do the following:
Place the Demo Board on a non-conductive surface.
Connect the 9-volt battery, and make sure that polarity is correct.
The displays and measurements shown in this manual are typical. They will vary depending on the condition of the Demo Board battery.
Page 14
XIII
Figure 2 Demo Board
Non-conductive
surface
Page 15
XIV
91/92/96/99/105
Users Manual
Page 16
XV
Single Input Connection
with Mini Test Hook
Single Input Connection
with HF Adapter
Dual Input Connections
with Mini Test Hooks
Dual Input Connections
with HF Adapters
Figure 3 Input Connections for High Frequency Electronic Measurement
Page 17
XVI
91/92/96/99/105
Users Manual
Page 18
XVII
Single Input Connection
with High Voltage
Test Pin
Single Input Connection
with High Voltage
Test Pin & AC20
Dual Input Connections with
Industrial Alligator Clips
(PM9084/001 & AC20)
Single Input Connection with
Industrial Alligator Clips
(PM9084/001 & AC20)
Figure 4 Input Connections for Isolated Heavy Duty Industrial Measurements
Page 19
SCOPEMETER TEST TOOL FEATURES . . . . . . . . . 1-2
USING THE HOLSTER AND THE TILT STAND . . . . 1-3
POWERING THE SCOPEMETER TEST TOOL . . . . . 1-4
MINIMIZING SIGNAL NOISE . . . . . . . . . . . . . . . . . . . 1-5
CHARGING THE BATTERY . . . . . . . . . . . . . . . . . . . . 1-6
Saving Battery Life . . . . . . . . . . . . . . . . . . . . . . . . . 1-6
LOOKING AT ALL
MEASUREMENT CONNECTIONS . . . . . . . . . . . . . . . 1-7
Common Ground, Inputs A B . . . . . . . . . . . . . . . . . 1-7
Input/Output Terminal Ratings . . . . . . . . . . . . . . . 1-7
RS-232 Optical Interface Connection . . . . . . . . . . 1-7
READING THE DISPLAY . . . . . . . . . . . . . . . . . . . . . 1-8
Reading a Dual Display . . . . . . . . . . . . . . . . . . . . . 1-9
Reading a Scope Display . . . . . . . . . . . . . . . . . . . 1-9
Reading a Window Display . . . . . . . . . . . . . . . . . . 1-9
USING THE KEYS . . . . . . . . . . . . . . . . . . . . . . . . . . 1-10
STEPPING THROUGH A WINDOW . . . . . . . . . . . . 1-12
USING ON-LINE INFORMATION . . . . . . . . . . . . . . 1-13
Chapter 1
Introducing your ScopeMeter Test Tool
Page 20
1 - 2 91/92/96/99/105 Users Manual
SCOPEMETER TEST TOOL FEATURES
Your ScopeMeter test tool combines the capabilities of an easy-to-use digital storage oscilloscope with the versatility of a digital multimeter. When you select a measurement function, the test tool automatically chooses the best setup to analyze and compare complex waveforms, or simply to read voltage levels.
ScopeMeter test tool features include:
Digital Storage Oscilloscope.
3 2/3-digit, 5-MHz Digital MultiMeter (DMM).
Battery power.
Measure menu for quick setup.
Continuous Auto Set function for hands-off signal probing.
Large memory for front panel Setups, Waveforms, and Screens.
Min Max TrendPlot function with time stamp for long term recording.
Optical to RS-232 interface port, 600V isolation.
Instant Information key.
Page 21
Introducing The ScopeMeter Test Tool 1 - 3
USING THE HOLSTER AND THE TILT STAND
The ScopeMeter test tool is cradled in a holster that provides shock protection during rough handling. All keys and connections are accessible with the holster in place. You will need to remove the holster only to replace the batteries or access the Quick Operating Guide. (See Chapter 9 for battery replacement instructions.)
The test tool is also equipped with a multiposition tilt stand, allowing viewing from different angles. The stand can also be used to hang the test tool at a convenient viewing position. Simply push up on the quick release and tilt the stand. The tilt stand/bracket is fully usable with the holster in place. Typical positions are shown in Figure 1-1.
Figure 1-1 Multiposition Stand
RELEASE
HERE
Page 22
1 - 4 91/92/96/99/105 Users Manual
POWERING THE SCOPEMETER TEST TOOL
The test tool can be powered from any of the following sources. Refer to Figure 1-2.
Internal Battery Pack (PM9086) A rechargeable NiCad Battery Pack comes installed in every test tool.
C Cell Batteries You can use four alkaline batteries in place of the NiCad Battery Pack. (The charger is defeated when standard C cells are installed).
Power Adapter (PM8907) The Power Adapter/Battery Charger powers the test tool from a standard ac outlet. The test tool can be used during battery charging. Verify that your local line voltage is appropriate before using the Power Adapter/Battery Charger to power the test tool and/or charge the battery pack.
See Appendix 10C for more information.
Automotive Adapter (PM9087, optional) The Automotive Adapter powers and charges from a standard 12V dc automotive accessory (lighter) outlet.
Refer to Chapter 9 for battery replacement instructions.
Figure 1-2 Power Connections
EXTERNAL POWER SOCKET
Page 23
Introducing The ScopeMeter Test Tool 1 - 5
MINIMIZING SIGNAL NOISE
In general, using your test tool on battery power only will minimize noise pickup. Using the 10:1 probe will help in noise rejection.
If you use the Power Adapter (PM8907), connect a test lead from the black 4-mm banana jack COM (common) to the measurement common (ground) of the system under test. This technique reduces or eliminates any power line related noise.
Page 24
1 - 6 91/92/96/99/105 Users Manual
CHARGING THE BATTERY
WARNING
TO AVOID ELECTRICAL SHOCK, USE ONLY A BATTERY CHARGER THAT IS AUTHORIZED FOR USE WITH SCOPEMETER TEST TOOL.
Use the following procedure to charge the battery pack and power the instrument:
1. Connect the Power Adapter/Battery Charger to line voltage.
2. Insert the Power Adapter/Battery Charger low voltage plug into the Power Adapter connector of the ScopeMeter test tool. The test tool can now be used while the NiCad batteries charge slowly. If the test tool is turned off, the batteries charge more quickly.
During operation, when the batteries are low, a blinking battery symbol appears on the top right of the display. When this occurs there is typically more than 30 minutes of operating time left.
3. The Power Adapter/Battery Charger uses a trickle charge for the batteries, so no damage can occur if you leave it charging for long periods, e.g., through the weekend. Typically a 21-hour recharge provides 5 hours (Fluke 91, 92, 96 and 99 : 4 hours) of use.
Saving Battery Life
When operated only on batteries, the test tool conserves power by shutting itself down. If no new keys have been pressed for 5 minutes or if the battery level is too low, the test tool beeps and displays a message. This prompts you to turn off the test tool or to continue. If no key is pressed during the next 5 minutes, the test tool turns itself off automatically .
When the is pressed ON, the last configuration prior to the automatic shutdown will be restored.
Automatic power shutdown will not occur if Min Max is on or if any key is pressed. Although recording will continue if the batteries are low, memory retention is not jeopardized.
If the POWER ADAPTER is connected, there is no automatic power shutdown.
Page 25
Introducing The ScopeMeter Test Tool 1 - 7
LOOKING AT ALL MEASUREMENT CONNECTIONS
The ScopeMeter test tool provides four signal connection points: two safety BNC jack inputs (red INPUT A and grey INPUT B ) and two safety 4-mm banana jack inputs (COM and
EXT.mV). This arrangement is shown in Figure 1-3. All connections are positioned within a protective recess at the top of the instrument. The banana jacks are also used as Waveform Generator Output.
Figure 1-3 Measurement Connections
In mode, you can use the red INPUT A BNC jack and the grey INPUT B BNC jack as input. In mode, only the red INPUT A BNC jack is used as input. In , , and modes, use the red and black 4-mm banana jacks.
For low frequency measurements (up to about 2 MHz), ground can be connected to the black 4-mm banana jack. For higher frequencies you must use the HF Adapter or Mini Testhook on the probe.
The red 4-mm banana jack also serves as an input connection for an external trigger in Scope mode or as an output connection for the waveform generator. This waveform generator can provide voltage or current output.
Common Ground, Inputs A B
The test tool uses a three-lead connection system for dual input, isolated (electrically floating) measurements.
Input/Output Te rminal Ratings
Maximum voltage ratings are defined near the related terminal. Refer to the Specifications in Appendix 10A for complete terminal rating information.
91
91 92 96
INPUT B is not available for FLUKE 91
91
91
Page 26
1 - 8 91/92/96/99/105 Users Manual
RS-232 Optical Interface Connection
The Optically Isolated RS-232 Adapter/Cable (PM9080) can be connected to the test tool for printer output and computer interface. Using the FlukeView software for DOS and Windows adds the following functionality to your ScopeMeter test tool:
Storing of measurements in memory for later retrieval.
Comparing of measurements with reference examples.
Storing, analyzing, and documenting measurements.
The following table shows the printing and communication features for each model:
READING THE DISPLAY
The display provides a great deal of information. The major part of the display is always devoted to meter readings or the scope waveform. A Dual display appears in ,
, , and modes, and is divided into two areas: Main display and Bottom display. A Scope display appears in mode, and is divided into three areas: Top display, Main display, and Bottom display.
Refer to Figure 1-4 during the following discussions.
Figure 1-4 ScopeMeter Test Tool Display
91 92 96 99 105 Screen dump using FlukeView Direct output to printer Remote control via computer software
Page 27
Introducing The ScopeMeter Test Tool 1 - 9
Reading a Dual Display
Main Display: Displays the numeric readings combined with a full screen-width waveform display.
Bottom Display: Displays the menu that provides choices available through the function keys and shows which are active.
Reading a Scope Display
Top Display: Identifies the voltage range, type of signal coupling, selected probe type for INPUT A and INPUT B, time scale, trigger source, and trigger slope. The top right area displays the present status of the test tool, AUTO, MANUAL, RUN, HOLD, etc.
Main Display: Displays the actual or stored waveforms (Scope mode).
Bottom Display: Displays the menu that provides choices available through the function keys and shows which are active.
Reading a Window Display
When you change a setup, a part of the Main display area is used for a window. It displays choices accessed with the related function key from which you make a selection by pressing (SELECT ITEM). The waveform area is compressed to about 50% (amplitude) size, and the window replaces the bottom half of the Main display.
Page 28
1 - 10 91/92/96/99/105 Users Manual
USING THE KEYS
Figure 1-5 The Keypad
The keys with a predefined function, regardless of the test tool's mode, are called keys or hard keys. Usally they give access to a menu, where items can be selected.
The two yellow keys and provide the possible ways to find your measurement.
gives immediate access to a list of measurements. When you choose a measurement, the test tool automati­cally selects the related main mode.
gives access to the Main menu. Here you can choose from the five main modes: , , , , and . Pressing gives you more control over the selected main mode.
Page 29
Introducing The ScopeMeter Test Tool 1 - 11
Figure 1-6 shows the basic navigation of the test tool.
Figure 1-6 Basic Navigation
The five blue keys
,
,
,
, and are called function keys. These keys change function based on the present active menu. Actual function key definitions appear on the bottom display. Function keys and windows work together to provide a complete menu-driven user interface.
is the MORE function key, which opens the window assigned to the key. When you press the Submenu key for any main function, the , , , and keys allow immediate access to the more common functions that are also under the (MORE function key).
The
keys are used to choose an item in a box, or directly from the window. These keys are indicated by two arrows.
Yellow Yellow Dark grey
Hard key
SubMenu Structure
Most common Functions
also found under
Dark grey
Page 30
1 - 12 91/92/96/99/105 Users Manual
STEPPING THROUGH A WINDOW
When a window appears, the assignments to the function keys change as follows:
is the 'CLOSE' key, which closes the active window. is the 'CANCEL' key, which ignores changed
selections and closes the active window.
is the 'SELECT ITEM' key, which opens and closes a
list box and selects the highlighted item. Stepping through a window has a fixed sequence as
shown in Figure 1-7.
Figure 1-7 Sequence to Go Through a Window
The fixed sequence is as follows:
¿
Press a MORE function key. This opens a window.
(Press (NEXT PAGE) or (PREVIOUS PAGE) to choose an item on page 2 or page 3.)
¡
or Choose the item that has to be changed.
¬
Open the list box.
Ð
or Choose the new parameter.
ƒ
This selects the new parameter and closes the list box.
Repeat steps
¡
through ƒ for more items.
Ý
Close the window and continue measurements.
Page 31
Introducing The ScopeMeter Test Tool 1 - 13
USING ON-LINE INFORMATION
You can get information about functions at any time by pressing . The short descriptions will help you under­stand how the test tool functions.
When you operate the ScopeMeter test tool, the display provides information about the present condition or explains procedures taking place and asks for confirmation. These messages are always displayed in a box.
More information is available by pressing . This displays one or more pages of extended on-line information. More information is available under the following conditions:
- In a menu (F1 to F5) you always can get information on the functions for every function key.
- In a window you can get more information about the highlighted function when the icon is displayed.
- In a message you can get more information about the conflict with the present setup when the icon is dis­played. If necessary, the test tool asks for confirmation and disables the conflicting situation.
Figure 1-8 hows an example of a "function info" screen. You can set the help level for on-line information to high or
low. This is explained in Chapter 6.
Figure 1-8 Using On-Line Information
When you have read the displayed information, press
to exit the information mode.
This returns the test tool to the latest setup before you pressed , and you can continue your measurement.
=== ST7535 pcx ===
Page 32
ADJUSTING THE DISPLAY . . . . . . . . . . . . . . . . . . . . 2-3
Adjusting the Contrast . . . . . . . . . . . . . . . . . . . . . . 2-3
Turning the Backlight On and Off . . . . . . . . . . . . . . 2-3
SELECTING THE PROBE TYPE . . . . . . . . . . . . . . . . 2-4
POWER-ON CONFIGURATIONS . . . . . . . . . . . . . . . . 2-4
Master Reset (Default Startup) . . . . . . . . . . . . . . . 2-4
Power-on/save memory . . . . . . . . . . . . . . . . . . . . . 2-5
PERFORMING AN EASY SETUP . . . . . . . . . . . . . . . 2-5
USING THE DEMONSTRATION (DEMO) BOARD . . 2-6
Quick Measurement Demonstration. . . . . . . . . . . . . 2-7
Performing Some Meter Operations . . . . . . . . . . . 2-10
Performing Some Scope Operations . . . . . . . . . . . 2-12
Chapter 2
Getting Started
Page 33
2 - 2 91/92/96/99/105 Users Manual
This chapter provides a 15-minute demonstration intended for those who are not familiar with the ScopeMeter test tool. It gives some hands-on experience, with an emphasis on learning by seeing and doing. Those who are familiar with the test tool can skip the following pages and continue to Chapter 3.
Press to turn on the test tool.
Operation begins in the last-known configuration. Factor y default settings are used at the first power-on. Subsequent power-ons maintain configuration changes made with keystrokes or through recall of a setup memory during the previous session.
Page 34
Getting Started 2 - 3
ADJUSTING THE DISPLAY
Adjusting the Contrast
Use
to adjust the contrast of the display to your preference. Note that new contrast adjustment of the display is stored in memory until a new adjustment is made.
Figure 2-1 Contrast adjustment
Contrast adjustment can also be made after you have pressed , and you have highlighted the contrast with
, and pressed to select. is then
visible in the right bottom of the display.
Turning the Backlight On and Off
1. Press to turn on the backlight. The actual power source,
BATTERIES
or POWER ADAPTER appears for
a moment on the display.
NOTE
Using the backlight shortens battery power operation time by about
1
/2 hour.
2. Press to turn off the backlight again.
Page 35
2 - 4 91/92/96/99/105 Users Manual
SELECTING THE PROBE TYPE
To prevent measurement errors, ensure that your test tool is adapted to your probe at power-on. Note that the PM8918 probes, delivered with the test tool, have an attenuation of 10:1.
For accurate measurement, periodic probe calibrations are necessary. This is explained in Chapter 9, "User Mainte­nance".
POWER-ON CONFIGURATIONS
There are two different ways to reset the ScopeMeter test tool: the Master Reset, and the power-on/save memory.
Master Reset (Default Startup)
CA UTION
A Master Reset clears everything saved in memory. Never perform a Master Reset unless you are prepared to lose everything you saved in memory.
1. Turn the ScopeMeter test tool off.
2. Press and hold .
3. Press and release . The test tool turns on, and you should hear a double beep, indicating the Master Reset was successful. The screen should display input: A PROBE 10:1 on top of the waveform area.
4. Release .
Perfor m a Master Reset only to make sure that your test tool is in the initial settings condition. Master Reset sets the test tool for using 10:1 probes.
See Chapter 6 for more information about Master Reset conditions.
Page 36
Getting Started 2 - 5
Power-on/save memory
This procedure also resets the ScopeMeter test tool and sets the probe range to 10:1. The Reset does not clear the memories.
1. Turn the test tool off.
2. Press and hold .
3. Press and release . The test tool turns on, and you should hear a double beep. input: A PROBE 10:1 appears on the display.
4. Release .
Use this feature when you have saved a setup, screen, or waveforn in memory.
PERFORMING AN EASY SETUP
For quick operation, the test tool is equipped with a Continuous Auto Set function. This function optimizes the range and the time in all modes and assures a stable picture on nearly all waveforms. You only need to press
to activate. Once activated this function permits
hands-off probing. You can redefine the Continuous Auto Set configuration to
your own preference. Refer to Chapter 6 for more informa­tion.
Page 37
2 - 6 91/92/96/99/105 Users Manual
USING THE DEMONSTRATION (DEMO) BOARD
During the Quick Measurements, Quick Meter Operations, and Quick Scope Operations, you use the Demo Board that is included in your ScopeMeter test tool kit. Before doing any measurement, review the safety precautions in the beginning of this manual. Refer to Figure 2-2 for the setup of this demonstration.
Connect the red scope probe to TP2.
Continue with the Quick Measurements Demonstration for more hands-on training using active waveforms.
NOTE
The displays and measurements are typical. They will var y depending on the condition of the Demo Board battery.
Figure 2-2 Demonstration setup
Page 38
Getting Started 2 - 7
1. + Perform a Reset at Power ON. At power-on, all five function key definitions in the bottom display are assigned to the Main menu. Notice that METER is highlighted.
Ensure the probe range is 10:1, as indicated on the display. The screen displays about "03.43 Vrms AC" and "+02.15 V DC" in large numbers, and
34.44 Hz in smaller numbers. (The values can var y depending on the condition of the Demo Board battery.) An additional scope trace gives a graphical representation of the waveform.
2. Open the Measure menu; the five function keys now are assigned to: MORE MEASURE, V, Hz, dB, or TIME.
Continued on next page
QUICK MEASUREMENTS DEMONSTRATION
Page 39
2 - 8 91/92/96/99/105 Users Manual
3. Open the VOLT measurement menu; a list of the voltage measurements appears on the display
4. +
Use these keys to choose a measurement. For now, highlight Vrms AC+DC.
5. Press this function key to select Vrms AC+DC. The value of about "4.00 Vrms AC+DC" appears as the main top reading on the display. The previous readings are now shifted down one. Note that the Hz has been removed.
6. Open the TIME measurement menu; a list of the time-related measurements appears on the display.
7. +
Choose another measurement. For now, highlight rise time.
91 92
Continued on next page
QUICK MEASUREMENTS DEMONSTRATION
Page 40
Getting Started 2 - 9
8. Some measurements are done in another mode. This message appears in a double-lined bo x on the display: "This measurement is possible in SCOPE mode. Change to SCOPE mode now ?".
9. The test tool changes automatically to the Scope mode. You can now find the Rise Time measurement result below the waveform area.
10. The left and right cursors move across the display. The intersection of waveform and cursor mark the level for the two horizontal cursors.
This completes the Quick Measurements Demonstration; continue with the Meter Operations.
QUICK MEASUREMENTS DEMONSTRATION
Page 41
2 - 10 91/92/96/99/105 Users Manual
If you have not already done so, perform a + reset.
1. Set the Display to 3V manual range, about "3.412 Vrms AC" and "+2.127 V DC" appear, and the amplitude of the graphical representation increases.
2. Press this key once to measure a stable display . TOUCH HOLD appears in the top right of the display.
3. Press this key once again to resume new measurements.
4. Press this key. The display changes to 1V range. The voltage readings are overloaded, and "OL Vrms AC" and "+OL V DC" appear.
5. The test tool automatically assumes a new setup that is optimized for measur ing the actual input signal.
Input A
Input A
Continued on next page
PERFORMING SOME METER OPERATIONS
Page 42
Getting Started 2 - 11
6. Press this key once to start recording the maximum, average, minim u m, and max-min measure­ments. At the same time the trend plot representation appears at the lower area of the display.
7. Press this key again to stop Min Max recording and trend plot. But first a message "All MIN­MAX recordings will be lost. Are you sure?" appears to alert for the consequences.
8. Continue with your measurements.
Continued on next page
PERFORMING SOME METER OPERATIONS
Page 43
2 - 12 91/92/96/99/105 Users Manual
9. Open the Meter-submenu mode. The five function keys are now assigned to the meter­submenu identifiers.
10. Open the More Meter window. The window enables you to change several settings in Meter mode.
11. Go to the next page and find more settings that you can change.
12. Close the window again, without changing the setup configuration of the test tool.
This completes the Meter Operations demo; continue with the Scope Operations.
PERFORMING SOME METER OPERATIONS
Page 44
Getting Started 2 - 13
If you have not already done so perform a + reset.
1. Activate the Scope Mode, then press . An image of the waveform appears on the display. The top display shows the attenuator, probe, time base, and trigger information.
2. The waveform amplitude decreases. The attenuator scale in the top display changes to a less sensitive range.
3. The number of periods of the wavefor m increases. The time base scale in the top display changes to a slower time/DIV.
4. The waveform shifts upwards.
5. The waveform shifts to the right of the display.
Input A
Input A
Continued on next page
PERFORMING SOME SCOPE OPERATIONS
Page 45
2 - 14 91/92/96/99/105 Users Manual
6. The test tool automatically selects attenuation, time base, and triggering to give you a readable display of the waveform.
7. Press this to look at the Trigger menu now.
8. This begins to display as the waveform rises above the trigger level (+SLOPE) or falls below the trigger level (-SLOPE). Watch the display; the wavefor m starts at a different place as trigger + or - is selected. Continue with +SLOPE selected.
9. This activates the ADJUST LEVEL function. The level value is displayed in the right bottom, and is visualized by the " " icon on the left edge of the display. Ensure also that RUN is displayed in the top right of the display as an indication that your test tool is triggered.
a
Continued on next page
PERFORMING SOME SCOPE OPERATIONS
Page 46
Getting Started 2 - 15
10. Adjust the trigger level: the " " icon moves upward as you change the trigger level. When the level is more than the waveform value, RUN changes into NOTRIG, and the display becomes unstable; your test tool is no longer triggered.
11. Adjust the trigger level to about 50% of the waveform amplitude, so that your test tool is triggered again.
12. Turn your ScopeMeter test tool off.
You are now able to operate basic functions in routine applications. Continue with Chapters 3 and 4 for a more detailed discussion of the ScopeMeter test tool functions.
a
PERFORMING SOME SCOPE OPERATIONS
Page 47
MAKING CONNECTIONS . . . . . . . . . . . . . . . . . . . . . . 3-2
USING THE MEASUREMENT FUNCTIONS . . . . . . . 3-3
USING A SHORTCUT TO MOST
COMMONLY MEASUREMENTS . . . . . . . . . . . . . . . . 3-9
ORDERING MEASUREMENT READINGS . . . . . . . 3-10
MEASURING FUNCTIONS NOT
AVAILABLE FROM THE MEASURE MENU . . . . . . 3-10
Additional measurements for Meter
and EXT.mV modes . . . . . . . . . . . . . . . . . . . . . . 3-10
Additional measurements for the Scope mode . . 3-11
USING THE MEASUREMENTS MAP . . . . . . . . . . . 3-12
Chapter 3
Making Measurements
Page 48
3 - 2 91/92/96/99/105 Users Manual
MAKING CONNECTIONS
ScopeMeter test tool measurement connections are illustrated in Figure 3-1.
Use the red INPUT A BNC jack or the red and black 4 mm banana jacks for measurements. The ScopeMeter test tool informs you when to use INPUT A or the banana jacks.
The red 4-mm banana jack is used for diode test, continuity test, Ohm measurement, and when you use a Current or Temperature probe with banana jacks.
When you use the PM8918 probe, select the 10:1 probe type for correct decimal readout.
Figure 3-1 Measurement Connections
Page 49
Making Measurements 3 - 3
USING THE MEASUREMENT FUNCTIONS
Use the Measure menu to choose one or more automatic measurements. Simply choose your desired measurement and the test tool sets itself up and displays the result.
Regardless of the mode you are in, press to access the Measure menu.
Press (MORE MEASURE) to access a multipaged (seven pages) list of all possible measurements. Press
(PREVIOUS PAGE) and (NEXT PAGE) to step
through the pages in the More Measure window.
On a page, use and to highlight the desired measurement, then press to select. This activates your chosen measurement and closes the window. The measurement result is instantly displayed on the screen.
Measure Menu
Page 50
3 - 4 91/92/96/99/105 Users Manual
See Figure 3-2 for screen examples in Dual Display mode and Scope mode.
Figure 3-2 Measurement results in Dual Display mode
and Scope mode
ATTENTION
The Scope Mode measurements are not valid for the FLUKE 91 and 92
The measurements in the More Measure window are listed below.
Page 1 of 7: OHM
Measure the resistor value () of a resistor
connected to the red and black 4-mm banana jacks. DIODE Test a diode that is connected to the red and
black 4-mm banana jacks. The red banana jack is the high input while the black one is the low. The result is displayed as the forward or reverse voltage of a diode.
CONTINUITY
Detect an open or closed circuit that is connected to the red and black 4-mm banana jacks. A beeper draws your attention when the circuit is closed.
°C PROBE EXT.mV Detect a temperature in degrees Celsius with a Temperature probe (optional) that is connected to the red and black 4-mm banana jacks.
°F PROBE EXT.mV Detect a temperature in degrees Fahrenheit with a Temperature probe (optional) that is connected to the red and black 4-mm banana jacks.
ACCESSORY (mV) EXT.mV Measure millivolts with an accessory that is connected to the red and black 4-mm banana jacks.
Page 51
Making Measurements 3 - 5
Page 2 of 7:
NOTE
INPUT A in Meter mode can make separate V DC and Vrms AC measurements on the same signal at the same time. V DC and Vrms AC readings are normally displayed together.
V DC METER Measure the DC (direct) voltage of the input signal in Meter mode.
Vrms AC METER Measure the RMS (Root Mean Square) voltage of the input signal in Meter mode. This is done with DC- coupled input. Allows AC component of signal to be displayed
Vrms AC+DC METER Measure the true RMS (Root Mean Square) voltage of the input signal in Meter mode. If a DC-offset is present, this measurement will read higher or lower than Vrms AC. If there is no DC offset present, this measurement will be the same as Vrms AC.
V DC (mean) SCOPE Measure the mean (average of readings) value of the waveform between the (vertical) cursors in Scope mode.
Vrms SCOPE Measure the RMS (Root Mean Square) voltage of the waveform between the (vertical) cursors in Scope mode.
dV SCOPE Measure the voltage difference between the (horizontal) cursors in Scope mode.
Page 3 of 7: V peak/peak SCOPE Measure the difference between
highest and lowest voltage value of the waveform between the (vertical) cursors in Scope mode.
Vmax peak SCOPE Measure the maximum peak value of the waveform between the cursors in Scope mode.
Vmin peak SCOPE Measure the minimum peak value of the waveform between the cursors in Scope mode.
dBV DC METER (decibel Volts) Measure the voltage ratio of the DC component of a signal with respect to a reference voltage in Meter mode. This is done with DC­coupled input.
dBV AC METER (decibel Volts) Measure the voltage ratio of the AC component of a signal with respect to a reference voltage in Meter mode. This is done with DC­coupled input.
For dBV, you can select a range of references by pressing
or . Choose a voltage between 50 mV and
9.99V. The default value is 1V.
Page 52
3 - 6 91/92/96/99/105 Users Manual
Page 4 of 7: A DC METER Measure the DC current in Meter mode
with a Current probe (available as an option) that is connected to the INPUT A BNC. This is done with DC­coupled input. A list box enables you to select the proper range: 1, 10, 100 mV/A, or 1 V/A.
A AC METER Measure the AC current in Meter mode with a Current probe (available as an option) that is connected to the INPUT A BNC. This is done with DC­coupled input. A list box enables you to select the proper range: 1, 10, 100 mV/A, or 1 V/A.
A AC+DC METER Measure the AC+DC current in Meter mode with a Current probe (available as an option) that is connected to the INPUT A BNC. This is done with DC-coupled input. A list box enables you to select the proper range: 1, 10, 100 mV/A, or 1 V/A.
A DC EXTmV Measure the DC current in EXT.mV mode with a Current probe (available as an option) that is connected to the red and black 4-mm banana jacks. This is done with DC-coupled input. A list box enables you to select the proper range: 1, 10, 100 mV/A, or 1 V/A.
A AC EXTmV Measure the AC current in EXT.mV mode with a Current probe (available as an option) that is connected to the red and black 4-mm banana jacks. This is done with DC-coupled input. A list box enables you to select the proper range: 1, 10, 100 mV/A, or 1 V/A.
A AC+DC EXTmV Measure the AC+DC current in EXT.mV mode with a Current probe (available as an option) that is connected to the red and black 4-mm banana jacks. This is done with DC-coupled input. A list box enables you to select the proper range: 1, 10, 100 mV/A, or 1 V/A.
Page 53
Making Measurements 3 - 7
Page 5 of 7: Hz METER Measure the frequency of the input signal in
Meter mode. Hz SCOPE Measure the frequency of the wavefor m in
Scope mode. At least 1
1
/2 cycles of this waveform must
occur between the cursors. duty cycle + METER Measure the ratio between the
time the signal is positive to the total period time in Meter mode. It is shown as a percentage figure.
duty cycle - METER Measure the ratio between the time the signal is negative to the total period time in Meter mode. It is shown as a percentage figure.
pulse width + METER Measure the duration of the average positive pulse of the waveform in Meter mode.
pulse width - METER Measure the duration of the average negative pulse of the waveform in Meter mode.
Page 6 of 7: dBm DC METER (decibel milliwatt) Measure the power
ratio of the DC component of a signal with respect to 1 mW in Meter mode. This is done with DC-coupled input.
You can choose 50, 60, 75, 93, 110, 125, 135, 150, 250, 300, 500, 600, 800, 900, 1000, or 1200 as reference impedance.
dBm AC METER (decibel milliwatt) Measure the powe r ratio of the AC component of a signal with respect to 1 mW in Meter mode. This is done with DC-coupled input. You can choose 50, 60, 75, 93, 110, 125, 135, 150, 250, 300, 500, 600, 800, 900, 1000, or 1200 as reference impedance.
dBW DC METER (decibel Watt) Measure the power ratio of the DC component of a signal with respect to 1W in Meter mode. This is done with DC-coupled input. You can choose 1, 2, 4, 8, 16, or 50Ω as reference impedance.
dBW AC METER (decibel Watt) Measure the power ratio of the AC component of a signal with respect to 1W in Meter mode. This is done with DC-coupled input. You can choose 1, 2, 4, 8, 16, or 50Ω as reference impedance.
WATT DC METER Measure the DC audio watts from the signal in Meter mode. This is done with DC-coupled input. You can choose 1, 2, 4, 8, 16, or 50 as reference impedance.
WATT AC METER Measure the AC audio watts from the signal in Meter mode. This is done with DC-coupled input. You can choose 1, 2, 4, 8, 16, or 50 as reference impedance.
For dBm , dBW, and WATT, you can select a range of references by pressing or .
Page 54
3 - 8 91/92/96/99/105 Users Manual
Page 7 of 7: RPM 1 METER (revolutions per minute) Measure one
count per revolution in Meter mode. RPM 2 METER (revolutions per minute) Measure one
count per two revolutions in Meter mode. dt (delta time) SCOPE Measure the time difference
between the (vertical) cursors in Scope mode. 1/dt (frequency) SCOPE Measure the relative
frequency of the waveform in Scope mode. It is shown as the reciprocal value of the time difference (dt) between the vertical cursors.
rise time SCOPE Measure the rise time between 10% and 90% points of the first rising slope of the wavefor m in Scope mode. First use and to determine the 0% and 100% points of the waveform. These are the intersection of the cursors with the waveform. Set the rising slope to at least four time divisions for a precise measurement. The rise time is now instantly displayed.
Figure 3-3 Rise Time Measurement
Page 55
Making Measurements 3 - 9
USING A SHORTCUT TO MOST COMMONLY MEASUREMENTS
From the Measure menu, you can also find many most commonly used measurements directly under one of the function keys , , , or . Press one of the four function keys to choose the type of measurement: V, Hz, dB, or TIME. This opens a single window with a list of measurements within the chosen type.
Press (V) to open a list with voltage measurements:
Press (Hz) to open a list with frequency measurements:
Volt List Box
Frequency List Box
Press (dB) to open a list with power measurements:
Press (TIME) to open a list with time measurements:
CANCEL ignores the TIME window selection and returns to the Measure menu again.
dB List Box
TIME List Box
Page 56
3 - 10 91/92/96/99/105 Users Manual
ORDERING MEASUREMENT READINGS
At default, three measurement readings are displayed at the same time, although you can display up to four readings. See Chapters 4 and 6 for more information.
When you exceed the number of selected readings, all readings shift down one and your latest chosen measurement is placed on top. Notice that the previous bottom reading is then removed.
MEASURING FUNCTIONS NOT AVAILABLE FROM THE MEASURE MENU
There are also a few more measurements that you can only find in the Scope, Meter or EXT.mV modes. These are listed below.
Chapter 4 describes how to select the Meter and EXT.mV mode measurements, and Chapter 5 the Scope mode measurements.
Additional measurements for Meter and EXT.mV modes Vrms AC (no DC) Measure the RMS (Root Mean
Square) voltage of the input signal in Meter mode. This is done with AC-coupled input so that the DC signal component is blocked and the r ange is optimized for the AC signal component. The V DC reading is inhibited due to AC coupling. In some cases this measurement gives one more digit of resolution over Vrms AC (DC coupled).
dBV AC+DC (decibel Volts) Measure the voltage ratio of the AC and DC components of a signal with respect to a reference voltage in Meter mode. This is done with DC­coupled input.
dBV AC (no DC) (decibel Volts) Measure the voltage ratio of the AC component of a signal with respect to a reference voltage in Meter mode. This is done with AC­coupled input so that the DC signal component is blocked and the range is optimized for the AC sinal component.
Page 57
Making Measurements 3 - 11
dBm AC (decibel milliwatt) Measure the power ratio of the AC component of a signal with respect to 1 mW in Meter mode. This is done with DC-coupled input.
dBm AC (no DC) (decibel milliwatt) Measure the power ratio of the AC and DC components of a signal with respect to 1 mW in Meter mode. This is done with AC-coupled input so that the DC signal component is blocked and the range is optimized for the AC sinal component.
dBW AC+DC (decibel Watt) Measure the power ratio of the AC and DC components of a signal with respect to 1W in Meter mode. This is done with DC-coupled input.
dBW AC (no DC) (decibel Watt) Measure the pow er r atio of the AC component of a signal with respect to 1W in Meter mode. This is done with AC-coupled input so that the DC signal component is block ed and the range is optimiz ed for the AC sinal component.
WATT AC+DC Measure the AC and DC audio watts from the signal in Meter mode. This is done with DC-coupled input. You can choose 1, 2, 4, 8, 16, or 50 as reference impedance.
WATT AC (no DC) Measure the AC audio watts from the signal in Meter mode. This is done with AC-coupled input so that the DC signal component is blocked and the range is optimized for the A C sinal component. You can choose 1, 2, 4, 8, 16, or 50Ω as reference impedance.
For dBV, dBm and dBW, you can select a range of references by pressing or .
dBV: Choose a voltage between 50 mV and 9.99V. The default value is 1V.
dBm: Choose one of the following reference impedances: 50, 60, 75, 93, 110, 125, 135, 150, 250, 300, 500, 600, 800, 900, 1000, or 1200Ω.
dBW: Choose one of the following references: 1, 2, 4, 8, 16, or 50Ω.
Additional measurements for the Scope mode TRIG to left Measure the time difference between the
trigger occurrence and the left cursor. TRIG to right Measure the time difference between the
trigger occurrence and the right cursor. V at left Measure the voltage amplitude where the left
cursor crosses the waveform. V at right Measure the voltage amplitude where the right
cursor crosses the waveform. phase Measure the phase shift between two waveforms. A
maximum of three phase shift measurements between the reference wa veform and three other waveforms are possible.
Page 58
3 - 12 91/92/96/99/105 Users Manual
USING THE MEASUREMENTS MAP
The measurements can be selected in various menus. The following Measurements map shows where you can find all measurements.
MEASURE
MENU
MAIN MENU
Voltage Measurements
V DC
Vrms AC
••
Vrms AC+DC
Vrms AC (No DC)
V DC (mean)
Vrms
••
dV
Vpeak/peak
••
V max peak
V min peak
••
V at left
V at right
ACCESSORY (mV)
Frequency Measurements
Hz (frequency)
duty cycle
••
RPM 1
RPM 2
••
Page 59
Making Measurements 3 - 13
Power Measurements
dBV DC
dBV AC
••
dBV AC+DC
dBV AC (no DC)
dBm DC
dBm AC
••
dBm AC+DC
dBm AC (no DC)
dBW DC
dBW AC
••
dBW AC+DC
dBW AC (no DC)
WATT DC
WATT AC
••
WATT AC+DC
WATT AC (no DC)
Time Measurements
pulse width
••
dt
1/dt
••
rise time
TRIG to left
TRIG to right
phase
MEASURE
MENU
MAIN MENU
Page 60
3 - 14 91/92/96/99/105 Users Manual
Ampere Measurements
A DC
••
A AC
A AC+DC
••
A AC (no DC)
Temperature Measurements
°
C
°
F
Resistance Measurements
Ohm
••
Continuity
Diode Measurements
Diode
MEASURE
MENU
MAIN MENU
Page 61
MAKING CONNECTIONS . . . . . . . . . . . . . . . . . . . . . . 4-2
SELECTING A MAIN MODE . . . . . . . . . . . . . . . . . . . . 4-3
SELECTING RANGES
(MANUAL/AUTO RANGE) . . . . . . . . . . . . . . . . . . . . . 4-7
HOLDING A STABLE MEASUREMENT (TOUCH HOLD
) . . . . . . . . . . . . . . 4-8
DISPLAYING MINIMUM AND MAXIMUM
READINGS WITH RELATED TRENDPLOT . . . . . . . . 4-9
Displaying a Min Max Record . . . . . . . . . . . . . . . . 4-9
Displaying the Min Max TrendPlot™ . . . . . . . . . . . 4-9
Freezing the Display . . . . . . . . . . . . . . . . . . . . . . 4-10
SELECTING THE SCOPEMETER KEY SUBMENU 4-11
Enabling and Disabling the
Change Alert™ Function . . . . . . . . . . . . . . . . . . . 4-11
Changing the Refresh Rate of the Display . . . . . . 4-11
MAKING MEASUREMENTS
IN METER AND EXT.mV MODE . . . . . . . . . . . . . . . 4-12
Changing the number of readings . . . . . . . . . . . . 4-12
TAKING RELATIVE READINGS (SCALING) . . . . . 4-12
Readings Relative to a Reference Point (Zero
) 4-12 Readings as a Percent Change from Reference Point (Zero %
) . . . . . . . . . . . . . . . . . 4-13
Readings Displayed as a Percent
of Scale (0%-100%) . . . . . . . . . . . . . . . . . . . . . . 4-13
Stop Scaling . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-14
Readings Relative to a Minimum or Maximum Value (Min Max Zero
) . . . . . . . . . . . 4-14
Scaling Using Minimum and Maximum Readings 4-15
Chapter 4
Using the Dual Display Mode Functions
Page 62
4 - 2 91/92/96/99/105 Users Manual
MAKING CONNECTIONS
ScopeMeter test tool measurement connections are illus­trated in Figure 4-1.
Use the red INPUT A BNC jack for Scope and Meter modes. Use the grey INPUT B BNC jack as a second input for Scope measurements . The red and black 4-mm banana jacks are used for , , and modes.
For low frequency measurements (up to about 2 MHz), ground can be connected to the black 4-mm banana jack. For higher frequencies you must use the HF Adapter or Mini Testhook on the probe.
When you use the PM8918 probe, select the 10:1 probe type for correct decimal readout.
Figure 4-1 Measurement Connections
91
91
Scope only
Page 63
Using the Dual Display Mode Functions 4 - 3
SELECTING A MAIN MODE
At power-on, the main menu at the bottom of the display is always shown. To get the main menu from another menu or screen, press .
Choose a main mode with any of the function keys:
, , , , or .
The measuring result is then displayed.
INPUT A and INPUT B are used to measure the signal wavefor ms. You can set the amplitude of both waveforms, time base speed, and trigger ing for an optimum trace display. See Chapter 5 for more information about SCOPE mode.
Signals supplied to INPUT A are measured. Two measurement results appear initially on the display in a 3 2/3-digit numeric display. At the same time you find the graphic representation of the wav ef orm in the lower displa y.
You can select the type and number of readings in the MORE METER window; press and to open the window.
Main menu
91
Page 64
4 - 4 91/92/96/99/105 Users Manual
If necessary, you can select the range of both the scope trace and meter display, and the time base speed.
Press to select the desired attenuation. Ranges are from 1V to 3 kV (Vrms: 2.5 kV) with 10:1 probe, or from 100 mV to 300V (Vrms: 250V) with direct 1:1 input.
Press to select the desired time base; ranges are from 1 µs to 100 ms/div.
Resistance mode. A resistor connected to the red and black 4-mm banana jack inputs is measured. The red banana jack is the high input, and the black banana jack (COM) is the low. The time base is fixed at 5 ms/DIV.
NOTE
In the diode and resistance modes, the black COMmon banana jack is not connected internally to the common of BNC INPUTS A and B.
Page 65
Using the Dual Display Mode Functions 4 - 5
The measurement result is displayed as a 3 2/3-digit display. The ranges are 30 (manual only), 300Ω, 3 kΩ, 30 kΩ, 300 k, 3 MΩ, and 30 M. A beeper alert is generated at <5% of a selected range in Manual Range mode. The beeper alert is turned on with ALERT ))).
Diode test mode. A diode connected to the red and black 4-mm banana jack inputs is tested. The measurement current through the diode is 0.5 mA. The red banana jack is the high input, and the black banana jack (COM) is the low.
The result is displayed as the forward or reverse voltage of the diode. A beeper alert is generated at forward voltage. The beeper alert is turned on with ALERT ))).
The time base is fixed at 5 ms/DIV and the attenuator at 3V.
NOTE
In the diode and resistance modes, the black COMmon banana jack is not connected internally to the common of BNC INPUTS A and B.
Page 66
4 - 6 91/92/96/99/105 Users Manual
EXTernal milliVolt function. Signals supplied to the red and black 4-mm banana jack inputs are measured. The red banana jack is the high input, and the black banana jack (COM) is the low. The ranges are 300 mV, and 3V. Time base ranges are from 1 µs to 100 ms/div.
Use this mode for low voltages and for the accessories available for the ScopeMeter test tool.
Page 67
Using the Dual Display Mode Functions 4 - 7
SELECTING RANGES (MANUAL/AUTO RANGE)
In manual range, the input attenuator range can be changed by pressing the INPUT A key or
, when these are assigned to . If in auto range, this switches automatically to manual range, and MANUAL appears in the top right of the display. The selected attenuator and time base ranges are shown on the display.
The desired time base range can be selected by pressing
Auto ranging (automatic setting of the attenuator and time base) is selected when you press .
Page 68
4 - 8 91/92/96/99/105 Users Manual
HOLDING A STABLE MEASUREMENT (TOUCH HOLD
)
The Touch Hold function captures the next stable measurement and a beep indicates that a stable measurement has been made. When the Touch Hold function is turned on, the numeric display is frozen (held) until a stable measurement is detected. Stable measurements are defined as:
Within +/-100 display counts for 1s.
Above 200 display counts for dc signals.
Above 300 display counts for ac signals.
Below overload (OL) in and diode test.
TOUCH HOLD appears in the top right of the display when the function is active.
NOTE
The Touch Hold function is not available in Min Max.
Figure 4-2 The Touch Hold function
The following procedure can be used for Touch Hold:
1. Press and measure the signal.
2. Wait until the ScopeMeter test tool produces a beeper signal to indicate a stable signal.
3. Remove the measuring leads and read the result from the display.
4. Measure a new signal, wait for the beeper, remove the leads, and read again.
5. Press to return to normal measurement mode.
Page 69
Using the Dual Display Mode Functions 4 - 9
DISPLAYING MINIMUM AND MAXIMUM READINGS WITH RELATED TRENDPLOT
Displaying a Min Max Record
Press to record the maximum (MAX), average (AVG), minimum (MIN), and difference (MAX-MIN) readings of the main (upper displayed) measurement. The test tool also continues to display the present reading. Time stamps appear next to the MAX, AVG, and MIN readings to show the elapsed time in hours, minutes, and seconds since the Min Max start to the most recent change in the respective reading. The AVG time stamp updates continuously. The MAX and MIN time stamps update only when new high or low readings are encountered.
NOTE
A beep occurs when a new MAX or MIN value is detected. New MIN, MAX, and AVG readings are then updated on the display.
Displaying the Min Max TrendPlot™
The test tool also continuously logs the maximum, aver age , and minimum readings to memory and displays all three as graphs. Vertical scaling and horizontal time compression are automatically resized to fit the TrendPlot on the display. The TrendPlot is slowly traced on the display from left to right until the screen is written and the time scale is resized. The initial time scale is 15s/DIV; after a full screen the scale is compressed to 30s, 1 min, 2 min, etc, to a maximum of 40 days over a whole display.
Page 70
4 - 10 91/92/96/99/105 Users Manual
ATTENTION
An overload of MAX or MIN will stop the Min Max record mode. The overloaded value (MAX or MIN) and the AVG are then displayed in reverse video, and the time stamp of the overloaded value is stopped. The TrendPlot will continue.
Freezing the Display
You can press at any time to freeze the displayed readings. HOLD appears in the top right of the display. Press a second time to resume the display updates.
Turn Min Max off again with one of the following actions:
Press a second time.
Press .
Select a new measurement.
Page 71
Using the Dual Display Mode Functions 4 - 11
SELECTING THE SCOPEMETER KEY SUBMENU
From any main mode, you can access the main mode sub­menu by pressing for more control over the selected mode. Figure 4-3 shows the METER mode submenu.
Figure 4-3 METER Mode Submenus
In , , , and modes you can choose the functions ALERT ))), Zero ∆, FAST/SMOOTH, and RESET READING directly with one of the function keys. Or you can find these functions in the Submenu MORE window. For more information about the SCOPE submenu, see Chapter 5.
The Submenu MORE window also offers advanced measurement adaptations, such as Scaling and Value Result. To open the window, press . Make your selections and press (CLOSE) to close the window.
Figure 4-4 More Meter Mode
Enabling and Disabling the Change Alert™ Function
Press to enable an audible alert signal. The test tool draws your attention with a beep when the measurement changes more than 100 digits. In mode the alert warns you when the measurement falls to less than 5% of the measurement range.
Changing the Refresh Rate of the Display
Select (FAST) to speed up the refresh rate in the measurement result display. A fast refresh rate is useful when you are making adjustments and must see the resulting reading as quickly as possible. Select (SMOOTH) to average readings over the last measurements (8 seconds at 5 ms/div). It reduces the influence of noise or unstable input signals. Select (no highlight) to get the normal readings.
Page 72
4 - 12 91/92/96/99/105 Users Manual
MAKING MEASUREMENTS IN METER AND EXT.mV MODE
In Meter and EXT.mV mode you can make the same measurements as found in the Measure Menu. To find these measurements, select ADD MAIN READING in the Submenu MORE window. This opens a list of all possible measurements. For more details of the measurements, see Chapter 3.
An easy way to return to the factory default readings of the active main mode is to press (RESET READING) from the Submenu.
Changing the number of readings
At the factory default setting the ScopeMeter test tool displays a combined numeric reading result and a wavefor m. You can customize the number of readings on the display from one to a maximum of four. On page 2 of the More window, highlight READINGS ON DISPLAY, select the new number 1, 2, 3, or 4, and press to select. Then press to remove the window again. In and modes the maximum number of readings is two.
TAKING RELATIVE READINGS (SCALING)
Scale Main Reading displays the present measurement result with respect to a defined range. From a Submenu, press to open a MORE functions window. Next, use
or to highlight SCALE READING. When you press , a list box appears on the display to select the scaling function: off, ZERO (see ), ZERO %, and 0%-100%. Highlight a scaling function, and press to select.
Figure 4-5 Scaling Functions
Readings Relative to a Reference Point (Zero ∆)
Select "ZERO " to set the present upper reading as the zero reference point. Subsequent readings are displayed as units of variation from this point. This feature is useful when you need to monitor input activity in relation to a known good value. The z ero function can also be used with dBV, dBm, or dBW.
You can also activate this function from the Submenu, and then press .
Page 73
Using the Dual Display Mode Functions 4 - 13
Readings as a Percent Change from Reference Point (Zero %∆)
Select "ZERO %∆" to set the present upper reading as the relative reference. Subsequent readings are displayed as a percentage of variation (delta) from this point. The figure below shows percentage change (tolerance) when measuring a resistor:
Readings Displayed as a Percent of Scale (0%-100%)
Scaling starts automatically when both 0% and 100% values have been entered as the scaled reference values. Use the following procedure (example for Meter mode):
1. From the SCALE READING list box, highlight 0%-100% and press .
2. Use
to highlight REFERENCE 100% and
press .
3. Measure the value that will represent 100%.
4. Highlight NOW (as the present 100% value) and press
to save the displayed value as 100%.
5. Repeat actions 2 to 4 for the 0% value.
6. Press (CLOSE) to activate scaling and remove the window.
The present measurement result with respect to the defined range with a 0% and 100% value appears now as a scaled % readout.
Page 74
4 - 14 91/92/96/99/105 Users Manual
A typical scaling display is shown in the figure below: Stop Scaling
To stop scaling do one of the following:
Choose another function.
Highlight SCALE READING in the More window, press
, highlight "off", and press .
Readings Relative to a Minimum or Maximum Value (Min Max Zero ∆)
Min Max Zero views the relative readings in units appro­priate for the function and range selected. When in Min Max recording mode, set the test tool as follows:
1. From a Submenu, press (ZERO ∆). Now the reading relative to the main value appears on the
display.
NOTE
In this example the Meter Min Max Zero is used.
Page 75
Using the Dual Display Mode Functions 4 - 15
NOTE
Scaling and Zero selections are mutually exclusive; selecting one deactivates the other.
Scaling Using Minimum and Maximum Readings
The MORE window also gives you access to a scaling function when the ScopeMeter test tool is in Min Max
recording mode. First press to activate Min Max recording. Next, from a Submenu MORE window, use
or to highlight SCALE READING. Then press to open the Scale list box.
NOTE
In this example the Ohm Min Max Scaling is used.
Figure 4-6 Min Max Scaling Functions
Min Max Scaling views the readings as relative percentage values. For example, you can determine how the existing reading relates as a percentage of a range defined by two of the following settings: MIN, MAX, MAX-MIN, and AVG. You would accomplish this from the Scale Reading list box as follows:
1. Use
to highlight 0%-100% and press .
2. Use
to highlight REFERENCE 100% and
press .
3. Measure the signal. The maximum value represents the 100% value.
4. Highlight MAX and press to set the recorded maximum value as 100% reference.
Page 76
4 - 16 91/92/96/99/105 Users Manual
5. Use
to highlight REFERENCE 0% and
press .
6. Measure the signal. The minimum value represents the 0% value.
7. Highlight MIN and press to set the recorded minimum value as 0% reference.
8. Press (CLOSE) to remove the window.
The Min Max Scaling is automatically activated. Since these entries set up a valid range, the ScopeMeter
test tool begins scaling with a display similar to the one in the following figure:
Page 77
MAKING CONNECTIONS . . . . . . . . . . . . . . . . . . . . . . 5-2
MAKING AN EASY SETUP . . . . . . . . . . . . . . . . . . . . 5-3
CONTROLLING INPUTS A AND B . . . . . . . . . . . . . . 5-3
Selecting the Input . . . . . . . . . . . . . . . . . . . . . . . . . 5-3
Selecting the Input Coupling . . . . . . . . . . . . . . . . . 5-3
Capturing Glitches . . . . . . . . . . . . . . . . . . . . . . . . . 5-4
Reversing the Polarity of the Displayed Waveform 5-4
ADJUSTING THE AMPLITUDE . . . . . . . . . . . . . . . . . 5-4
ADJUSTING THE TIME BASE . . . . . . . . . . . . . . . . . . 5-5
POSITIONING THE WAVEFORM ON THE DISPLAY 5-6
ACQUIRING WAVEFORMS . . . . . . . . . . . . . . . . . . . . 5-7
Choosing Recurrent or Single Acquisition . . . . . . . 5-7
Roll Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-7
Using "Zoom" to Increase or
Decrease Time Resolution . . . . . . . . . . . . . . . . . . . 5-8
SELECTING THE SCOPE SUBMENU . . . . . . . . . . . . 5-9
Capturing 10 or 20 Divisions . . . . . . . . . . . . . . . . . 5-9
Smoothing the Waveform (Average) . . . . . . . . . . 5-10
TRIGGERING . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-11
Trigger Basics . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-11
Selecting a Trigger Source . . . . . . . . . . . . . . . . . 5-12
Defining the Trace Start . . . . . . . . . . . . . . . . . . . 5-12
Selecting a Trigger Slope . . . . . . . . . . . . . . . . . . 5-12
Selecting a Trigger Level . . . . . . . . . . . . . . . . . . 5-13
Selecting Auto Level . . . . . . . . . . . . . . . . . . . . . . 5-13
Choosing a Trigger Delay . . . . . . . . . . . . . . . . . . 5-13
DISPLAYING THE MINIMUM
AND MAXIMUM OF A WAVEFORM . . . . . . . . . . . . 5-14
MAKING COMBINATIONS
WITH INPUT A AND INPUT B . . . . . . . . . . . . . . . . . 5-15
DISPLAYING MULTIPLE WAVEFORMS . . . . . . . . 5-15
ADAPTING THE DISPLAY . . . . . . . . . . . . . . . . . . . 5-16
Selecting a Dot Size . . . . . . . . . . . . . . . . . . . . . . 5-16
Selecting a Display Grid . . . . . . . . . . . . . . . . . . . 5-16
Adjusting the Trace Quality . . . . . . . . . . . . . . . . . 5-16
MORE INFORMATION . . . . . . . . . . . . . . . . . . . . . . . 5-17
Chapter 5
Using in Scope Mode
Page 78
5 - 2 91/92/96/99/105 Users Manual
MAKING CONNECTIONS
Scope connections are illustrated in Figure 5-1.
Maximum scope inputs on either BNC are as follows:
Maximum input signal voltage: 300V rms direct
Maximum input signal voltage: 600V rms when using
10:1 probe
Voltage to ground: 600V rms
Input impedance is 10 M/15 pF with the 10:1 probe.
Press and to select the Scope mode.
Figure 5-1 Scope Connections
91
91
Page 79
Using the Scope Mode 5 - 3
MAKING AN EASY SETUP
Continuous Auto Set automatically selects the optimum settings for connected input signals and keeps track of any variation of the signal. Start Continuous Auto Set by pressing . For most signals, the ScopeMeter test tool will automatically set inputs, amplitudes, time base setting, and triggering for an optimum display.
Pressing sets the following:
INPUT A and/or INPUT B is switched on depending on signal input. All other waveform displays are switched off.
Between two and five periods of the waveform with the lowest frequency are displayed.
The attenuation per input is set so that the waveform occupies about four vertical divisions.
The trigger source is chosen as the input signal with the lowest frequency.
If no triggerable waveform is found when you press , approximate settings are used. If the input signal subsequently changes, the test tool readjusts itself; you will always have a meaningful picture on the display.
Continuous Auto Set can be configured through the Menu. Refer to Chapter 6 for a complete description.
CONTROLLING INPUTS A AND B
Press or to open the input settings menu.
Figure 5-2 Input A Menu
Selecting the Input
Press on the appropriate input to turn it on and off.
Selecting the Input Coupling
Press on the appropriate input to select the input signal coupling. The selection is shown in the upper line of the top displa y. AC-coupling allows for reading ac voltages above 40 Hz only. DC-coupling allows for reading ac and dc voltages.
The GND selection disconnects the input signal and displays the ground or zero level as a horizontal line. Ground provides a useful tool for setting up the display to show a high dc level or voltage spike. For example, if you anticipate a high positive dc level, select GND and press
to move the ground level lower on the display. A small marker box ( ) on the right side of the display identifies the ground level.
91
Page 80
5 - 4 91/92/96/99/105 Users Manual
Capturing Glitches
Press on INPUT A to turn GLITCH DETECTion on. This function displays the events that occur between two samples on INPUT A. INPUT B is turned off. The events can be glitches or other asynchronous waveforms that are 40 ns (nanoseconds) or wider.
NOTE
GLITCH detection and AVERAGE are mutually exclusive.
Reversing the Polarity of the Displayed Waveform
Press to open the MORE INPUT window. Highlight INPUT A (or INPUT B) and press to open the list box. You can tur n on and off an input and invert the INPUT A and INPUT B display.
Invert reverses the polarity of the displayed waveform. This is identified by the trace identifier on the display. For example, a negative-going waveform would be displayed as positive-going, providing a more meaningful viewing perspective in some cases.
ADJUSTING THE AMPLITUDE
You can use
for INPUT A or INPUT B to adjust input waveform attenuation. Sixteen ranges are available. A 1:1 probe can accommodate 1 mV to a maximum of 100V per division, and a 10:1 probe can accommodate 10 mV to a maximum of 1 kV per division. The amplitude can be changed in increments of 1, 2, and 5 (e.g., 100 mV, 200 mV, 500 mV). The upper row of the top display shows the range and the type of probe for each input.
When a 10:1 probe is used with the 10 mV or 20 mV range, averaging is turned on automatically. (AVG 4 appears on the display.)
91
Page 81
Using the Scope Mode 5 - 5
ADJUSTING THE TIME BASE
Press to increase or decrease the number of periods. The time base readout is on the second row of the top display. Refer to Figure 5-3. Available settings depend on the acquisition mode in effect, and are as follows:
Recurrent 5ns (10ns)/DIV - 1s/DIV
*
Single 100ns/DIV - 1s/DIV
Roll 2s/DIV - 60s/DIV
*
Fluke 91, 92, 96 and 99 : 10ns/DIV Fluke 105 : 5ns/DIV
The time base can be changed in increments of 1, 2, and 5 (e.g., 10 ns, 20 ns, 50 ns).
Figure 5-3 Time Base and Amplitude Adjustment
Page 82
5 - 6 91/92/96/99/105 Users Manual
POSITIONING THE WAVEFORM ON THE DISPLAY
Considerable flexibility is offered in moving the trace(s) around the display. and reposition the trace up or down four divisions. This operation can be done separately for each input. The key simultaneously repositions all traces left or right from -4 to +16.5 divisions. Waveform positioning is demonstrated in Figure 5-4.
Figure 5-4 Positioning the Waveform
Page 83
Using the Scope Mode 5 - 7
ACQUIRING WAVEFORMS
Choosing Recurrent or Single Acquisition
You can instr uct the ScopeMeter test tool to continuously update its display, giving you a dynamic look at waveform activity as it occurs (RECURRENT). You can also tell the test tool to update its display only once, yielding a snapshot record of waveform activity (SINGLE), termed "Single­Shot" elsewhere in this manual.
From the Scope Submenu, press to highlight RECURRENT for continuous display updates. Pressing
stops and starts an acquisition. A HOLD on the top right of the display indicates that acquisition is stopped. The time base range is 5ns (10ns)/DIV to 1s/DIV.
Set up the ScopeMeter test tool for "Single-Shot" acquisi­tions with the following procedure:
1. Press to select SINGLE (highlighted).
NO TRIG begins flashing on the display. The instrument is waiting for a trigger.
RUN appears on the display when the single acquisi­tion is triggered. HOLD then reappears when acquisi­tion is completed.
2. Now press again and wait for another single
acquisition trigger.
The time base range is 100ns/DIV to 1s/DIV.
Roll Mode
Roll mode supplies a visual log of waveform activity and is especially useful when measuring lower frequency waveforms. To enter the Roll mode, press to select RECURRENT, then press to select 2s/DIV, 5s/DIV, 10s/DIV, 20s/DIV, or 60s/DIV.
The wavefor m is now slowly traced on the display from left to right until the full screen is written. Then the trace slowly shifts from right to left like a pen-recorder. The Roll mode is automatically turned off if a time base of 1s/DIV or faster is selected.
Roll is an untriggered mode. The Capture length and Single-Shot functions are not available. However, the wavefor m can be frozen on the display by pressing .
Page 84
5 - 8 91/92/96/99/105 Users Manual
Using "Zoom" to Increase or Decrease Time Resolution
Zoom displays a waveform with increased time resolution, giving up to 1000x magnification. Use the following procedure to Zoom in a waveform:
1. Press
2. Use the and keys to highlight "scope zoom".
3. Press to select. appears in the right bottom of the display.
4. Press to enlarge the waveform (speed up the time base) or press to shrink the wavefor m (slow down the time base).
The ScopeMeter test tool zooms in and out one step at a time and, if necessary, the trigger delay is recalculated. The focal point for this activity is waveform activity at the fourth time division. The zoom feature is useful as a quick examination of a live waveform. Zoom cannot be used on saved waveforms recalled from memory or when the waveform is frozen (HOLD).
Figure 5-5 Zoom in a Waveform
Page 85
Using the Scope Mode 5 - 9
SELECTING THE SCOPE SUBMENU
From the Scope mode, press to open the Scope mode submenu. This gives you more control over the Scope mode. Figure 5-6 shows the Scope Submenu.
Figure 5-6 Scope Submenu
Refer to Chapter 6 for descriptions of the (ZERO
)
and (CURSOR READING) function key. The MORE Scope window gives you access to all control
over the Scope mode. In addition to the functions shown in Figure 5-6, you will also find functions, such as Capture Length, Average Waveforms, and Scope measurements. To use these functions, press to open the MORE SCOPE window.
Figure 5-7 More Scope Window
Capturing 10 or 20 Divisions
This enables you to capture either 10 or 20 divisions of waveform detail. Selecting 10 divisions, or one full screen, is sufficient for most applications. For fast single shots, selecting 20 divisions enables you to capture more information, but on slow waveforms it lowers the screen update speed by a factor of 2. Now you can use
to move the acquired waveform right and left
across the display.
Figure 5-8 Capture length
Page 86
5 - 10 91/92/96/99/105 Users Manual
Smoothing the Waveform (Average)
The test tool can average multiple samples of a waveform, allowing for noise suppression without loss of bandwidth. To select average in the Scope mode, do the following:
1. From the More Scope window, highlight AVERAGE WAVEFORMS.
2. Press to open the list box.
3. Use to choose the number of waveforms to average for each display update. The maximum number is 256.
4. Press to select.
5. Press (CLOSE) to close the window.
To turn off averaging, set AVERAGE WAVEFORMS to be off again.
NOTE
AVERAGE WAVEFORMS: OFF changes to AVERAGE WAVEFORMS: by 4 in the 10 mV and 20 mV attenuator
ranges when a 10:1 probe is used. Averaging and Glitch detect (on INPUT A) are mutually exclusive.
Additional information about smoothing appears in Chapter 6. Waveform samples with and without averaging are shown in Figure 5-9.
Figure 5-9 Averaging
Page 87
Using the Scope Mode 5 - 11
TRIGGERING
Trigger Basics
Triggering tells the ScopeMeter test tool when to begin displaying the waveform. You can select which input signal should be used as the source, on which edge this should occur, and at what waveform level it should occur. Finally, you can tell the test tool to delay the waveform to be displayed with a specified time delay, number of cycles, or number of events.
The second top row on the display identifies the trigger parameters being used. Trigger icons on the display indicate the position of the trigger delay and the trigger level. Refer to Figure 5-10.
For example, 20 ms/DIV TRIG:A +120EVT -5DIV means that INPUT A is used as the trigger source with a
positive slope. The amount of trigger delay in time divisions is -5, and that the amount of trigger delay in ev ents is +120.
Figure 5-10 Display with All Trigger Information
Page 88
5 - 12 91/92/96/99/105 Users Manual
Press to open the Trigger menu.
Figure 5-11 Trigger Menu
Selecting a Trigger Source
Press to enter the MORE TRIGGER window, then highlight TRIGGER SOURCE and press to open the list box. Select one of the following trigger sources: INPUTA, INPUT B , EXTERN 2V (via external trigger input), EXTERN 0.2V (via external trigger input), or GENERATOR (via internal waveform generator).
NOTE
You can also find the trigger source selection under the appropriate INPUT A or INPUT B More window.
EXTERN 2V or EXTERN 0.2V are available as trigger source selections. You can connect an external trigger at the red 4-mm banana jack or use the PM9081 Banana-to-BNC adapter supplied. Make your selection for the appropriate TTL-compatible (Transistor-to­Transistor Logic) trigger level (0.2V or 2V).
91
91 92 96
GENERATOR
. See Chapter 6 for more
information.
NOTE
EXTERN and GENERATOR are mutually exclusive; selecting one deactivates the other.
Defining the Trace Star t
Highlight TRACE START and press to open the list box. When Wa it f or t rig ger is selected, the test tool needs a trigger to display a waveform. When free run is selected, and if no trigger is detected for 100 ms, the test tool provides its own trigger and the trace still appears on the display. Use this mode for a continuous display of a waveform.
Selecting a Trigger Slope
From the Trigger menu, press to select trigger ing on either the +SLOPE or -SLOPE of the chosen input.
91
92 96
Page 89
Using the Scope Mode 5 - 13
Selecting a Trigger Level
Press to open the ADJUST LEVEL box. If you have selected INPUT A or INPUT B as the trigger source, you can use and to adjust the trigger level continuously. Observe the level changes in the display. Press to select.
The trigger level is identified by the trigger icon (e.g., ) on the left of the display.
Selecting Auto Level
Press to find the optimal trigger level for the source waveform voltage.
Choosing a Trigger Delay
You can also begin to display the wavefor m at some time before or after the trigger point is detected. This is identified by the delay icon "T" on the top of the display.
From the second page of the MORE TRIGGER window, select RESET DELAY to reset all the selected trigger delays to their initial values (TIME DELAY : -2, TRIGGER SPECIAL DELAY: OFF).
Types of trigger delays are discussed below. In many instances, you can select more than one type of delay.
Press to enter the Time Delay list box. Use or
to set the trigger time delay in increments of the existing time base divisions. The range of divisions is -20 to +640. For example, if the time base is set to 2ms/DIV, +16 defines a delay of 16 time divisions (32 milliseconds) after the trigger is detected. Press to select.
From the second page of the MORE TRIGGER window, select TRIGGER SPECIAL DELAY (off, Events, or N-Cycle).
N-CYCLE The trigger waveform is divided by the number of cycles "N" to create a stable picture for waveforms with a cyclic character. (N-CYCLE can be used only on INPUT A.) Use or to select the number of cycles. A total of +2 to +255 cycles can be selected.
EVENTS Triggering via the external input is delayed for a number of e v ents that occur on INPUT A. Use or
to select the number of events (+1 to +1023). External trigger star ts the delay, and after the number of events on INPUT A is equal to the selected number, the test tool acquires the INPUT A signal.
Page 90
5 - 14 91/92/96/99/105 Users Manual
DISPLAYING THE MINIMUM AND MAXIMUM OF A WAVEFORM
The ScopeMeter test tool records the envelopes (minimum and maximum values) of the live waveforms A and B and A±B. (INPUT B not in FLUKE 91).
Press to activate the Min Max envelope mode. MIN MAX now appears on the top right of the display.
The minimum and maximum values are automatically saved in waveform memories. The display shows the resultant envelope in lower intensity (grey) together with the actual waveform. Any change in attenuation or time base will restart a new Min Max envelope action. The function is turned off with one of the following actions:
Press a second time.
Press .
You can use Min Max envelope to observe variations in time or amplitude of input signals over a longer period of time.
Page 91
Using the Scope Mode 5 - 15
MAKING COMBINATIONS WITH INPUT A AND INPUT B
Press on INPUT B to select A+B to show INPUTS A and B added together or to select A-B to show INPUT B subtracted from INPUT A.
You can also choose A+B and A-B from the MORE INPUT A or MORE INPUT B window. Highlight COMBINE A & B, and press to open the list box. You can also select A versus B to provide a plot with INPUT A on the vertical axis and INPUT B on the horizontal axis. Press to activate your selection.
DISPLAYING MULTIPLE WAVEFORMS
By displaying a combination of live and recalled w a v ef orms, the test tool can display a maximum of four waveforms simultaneously. For example, INPUTS A, B, A+B, and waveforms recalled from memor y could be displayed. You can display any combination of four waveforms.
If you attempt to display a fifth w aveform, a message on the display lets you know that you exceeded the maximum number of four waveforms.
NOTE
The A versus B display is not counted in the four-trace maximum.
91
Page 92
5 - 16 91/92/96/99/105 Users Manual
ADAPTING THE DISPLAY
You can adapt the display to your preference as follows:
1. Press .
2. Press to open the Display Menu window.
Selecting a Dot Size
3. Use to highlight DOT SIZE and press to open the list box.
4. Choose a dotsize: SMALL, MEDIUM, or LARGE and press to select. This selects the trace to be single, double, or triple width.
Selecting a Display Grid
5. Use to highlight SCOPE GRID and press
to open the list box.
6. Use to choose between FULL, QUAD­RANT, or BORDER. Press to select any of the three.
BORDER provides time and amplitude division marks around the display. QUADRANT adds horizontal and vertical division points as added reference points to the border display.
FULL adds a crosshatch pattern based on the horizontal time and vertical amplitude divisions on the display.
Figure 5-12 Grid Selection
Adjusting the Trace Quality
7. Use to highlight PERSISTENCE, and press to open the list box.
Persistence improves the trace quality when you measure on complex waveforms. Choose one of the five persistence states: short, normal, medium, high, or infinite to obtain the best image on the display. Press to select. Initially the ScopeMeter test tool is set to normal persistence.
NOTE
Contrast and Persistence work interactively. So when you readjust persistence, it may be necessary to choose another contrast again.
Page 93
Using the Scope Mode 5 - 17
In the DISPLAY window you can also choose the DOT JOIN function, which either shows the individual measure­ment samples or joins them together to give a continuous trace. DISPLAY MODE allows you to select normal or reverse video.
8. Press (CLOSE) to remove the window. The test
tool displays the USER OPTIONS menu again.
MORE INFORMATION
Chapter 8 (ScopeMeter Test Tool Tutorial) presents Scope mode applications that use many of the features discussed in this chapter. Refer to Chapter 6 for a discussion of additional capabilities.
Page 94
MAKING MEASUREMENTS USING THE CURSORS 6-2
Using the Markers . . . . . . . . . . . . . . . . . . . . . . . . . 6-3
Adding a Cursor Main Reading . . . . . . . . . . . . . . . 6-4
DELETING MEMORIES . . . . . . . . . . . . . . . . . . . . . . . 6-6
SAVING TO MEMORY . . . . . . . . . . . . . . . . . . . . . . . . 6-7
Saving a Screen . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-7
Saving Acquired Waveforms . . . . . . . . . . . . . . . . . 6-7
Saving a Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-7
Copying Waveforms or Settings . . . . . . . . . . . . . . . 6-8
RECALLING FROM MEMORY . . . . . . . . . . . . . . . . . . 6-8
Recalling a Screen . . . . . . . . . . . . . . . . . . . . . . . . . 6-8
Recalling a Waveform . . . . . . . . . . . . . . . . . . . . . . 6-9
Recalling a Setup Configuration . . . . . . . . . . . . . . 6-10
USING WAVEFORM MATH FUNCTIONS . . . . . . . . 6-11
Scaling the Display . . . . . . . . . . . . . . . . . . . . . . . . 6-11
Choosing a Math Operation . . . . . . . . . . . . . . . . . 6-11
USING A PRINTER . . . . . . . . . . . . . . . . . . . . . . . . . 6-14
Connecting to a Printer . . . . . . . . . . . . . . . . . . . . 6-14
Setting Print Parameters . . . . . . . . . . . . . . . . . . . 6-14
Selecting Printer Types . . . . . . . . . . . . . . . . . . . . 6-14
Selecting Print Options . . . . . . . . . . . . . . . . . . . . 6-14
Printing Grey in White or Black . . . . . . . . . . . . . . 6-16
Sending to a Printer . . . . . . . . . . . . . . . . . . . . . . 6-16
USING THE WAVEFORM GENERATOR . . . . . . . . 6-17
TESTING COMPONENTS . . . . . . . . . . . . . . . . . . . . 6-17
CHANGING THE RESET CONFIGURATIONS . . . . 6-19
ALTERING THE CONTINUOUS
AUTO SET CONFIGURATION . . . . . . . . . . . . . . . . 6-20
Using Manual Override . . . . . . . . . . . . . . . . . . . . 6-20
Changing Continuous Auto Set Options . . . . . . . 6-20
Chapter 6
Using Additional Capabilities
Page 95
6 - 2 91/92/96/99/105 Users Manual
MAKING MEASUREMENTS USING THE CURSORS
The Cursors allow you to make precise digital measurements of waveform activity between measurement areas. This can be done on live waveforms and on saved waveforms.
From Scope mode, press to turn on (and off) the cursor readings.
Active cursors appear on the display as two ver tical and two horizontal lines.
The intersection of vertical and horizontal lines of the left cursor marks a point on the wavefor m. The intersection of vertical and horizontal lines of the right cursor marks another point on the waveform. The measurement takes place between these two points.
Use and to move the left and right cur­sors which designates the area for the cursor measure­ment.
NOTE
You can also find many cursor measurements when using the
key. These measurements are explained in Chapter 3.
91
92
Page 96
Using Additional Capabilities 6 - 3
When the cursor readings are turned on, you can choose one or more measurements as follows:
1. From Scope mode, press to open to the More Scope window.
2. From the second screen page of the More Scope win­dow, highlight ADD MAIN READING. Press to open the list box. The ADD MAIN READING list box allows you to choose which cursor measurement read­ing to display. A list of possible measurements is shown in Figure 6-1.
Figure 6-1 Cursor Measurements Window
3. Choose the type of Scaling Reading: off, ZERO
, or %
off displays the measurement results in normal
values.
ZERO
displays a change to the reference value.
% displays a change to the reference value as
a percentage.
4. Choose the waveform source with Measure on Wave­form. The source selection is determined by active dis­play traces. The traces can be obtained from the active display (A, B, A±B) or from waveform memory.
5. Choose the number of Readings; the test tool can display up to five cursor readings at the same time. When you choose another main readings, all readings will shift down one, and your last chosen measurement is placed on top. Notice that the bottom reading is then removed from the display.
Using the Markers
Use the marker to identify the point on the waveform that a measurement has occurred or is expected to occur. The marker appears on the waveform as an "X".
6. From the third screen page of the More Scope window, press to open the Markers list box. A list box of markers in use is shown in Figure 6-2. Use or
to choose the marker. Press to select.
Figure 6-2 Markers List Box
Page 97
6 - 4 91/92/96/99/105 Users Manual
In this example, the markers are defining sample points for measuring frequency.
NOTE
Markers may be easier to see when you press and freeze the waveform.
7. Press to close the window.
Adding a Cursor Main Reading
From the ADD MAIN READING list box, use or to highlight the reading to display, then press to select. Repeat this for more readings.
In making cursor measurement calculations, the ScopeMe­ter test tool uses waveform activity between the cursors.
Figure 6-3 Cursor Main Readings List Box
The measurement results are displayed in the bottom dis­play area. A maximum of five measur ing functions can be turned on at the same time.
dV (for delta volts) Measures the voltage difference between the cursors.
dt (for delta time) Measures the time difference between the cursors.
1/dt (for relative frequency) Shows the reciprocal value of dt.
Page 98
Using Additional Capabilities 6 - 5
TRIG to left Measures the time difference between the trigger occurrence and the left cursor.
TRIG to right Measures the time difference between the trigger occurrence and the right cursor.
V at left Measures the voltage amplitude where the left cursor crosses the waveform.
V at right Measures the voltage amplitude where the right cursor crosses the waveform.
Vrms Measures the Root Mean Square (rms) voltage between the cursors.
V DC (mean) Measures the MEAN (average of read­ings) value of waveform between the cursors.
Vpeak/peak Measures the PEAK to PEAK value of the waveform between the cursors.
Vmax peak Measures the maximum peak value of the waveform between the cursors.
Vmin peak Measures the minimum peak value of the waveform between the cursors.
frequency Measures the frequency of the waveform between the cursors. At least 1
1
/2 cycles of this wave­form must occur between the cursors. You can scroll a waveform through the cursors by pressing . Read the frequency change in the bottom display.
rise time Measures the rise time between 10% and 90% points of the waveform. The 0% and 100% points are determined by the intersection of the wavefor m and the cursors. Set the rising slope to at least four time divisions for a precise measurement.
phase Measures the phase shift between two waveforms. You can select upto three waveforms for phase measurements to a reference waveform as selected with Measure on Waveform.
Page 99
6 - 6 91/92/96/99/105 Users Manual
DELETING MEMORIES
Open the Delete Menu at any time by pressing and
Figure 6-4 Delete Memory List Box
You can now delete all memories or selected memories (Screen memories, Waveform memories, or Setup memories). A message appears prompting you to confirm that you want to delete all memory.
CA UTION
A Master Reset destroys all memory contents and leaves initial setup with its default settings. Refer to "Master Reset configuration" further on in this chapter for a complete discussion of Initial Setup.
You can also delete a single memor y location. From the More Save window, first choose the type of memory (DELETE what?) you want to delete: Screen, Waveform, or Setup memory. Press (SELECT ITEM) to open a list box and choose the memory location you want to delete. Use or to highlight the memory number. Press
to delete the selected memory location.
91
92
Page 100
Using Additional Capabilities 6 - 7
SAVING TO MEMORY
You can save screens, waveforms, and setups to memor y, copy them between memories, and recall them from memory. Ten nonvolatile Screen memories, 20 nonvolatile Waveform memories, and 40 nonvolatile Setup memories are available. (For FLUKE 96 yo u can save five nonvolatile Screen memories, 10 nonvolatile Waveform memories, and 20 nonvolatile Setup memories).
Press to open the Save to Memory menu.
Saving a Screen
Use this function to save the current active display image for future reference. You can save a maximum of 10 screens (For FLUKE 96 you can save a maximum of five screens).
Press (SAVE SCREEN) to save the current screen in memory (memory #).
The test tool automatically uses the next available memory location. The display continues to show the actual input waveform.
NOTE
When Medium, Long, or Infinite persistence has been cho­sen, you will only sav e the actual waveform information.
Saving Acquired Waveforms
Use this function to save acquired waveforms on INPUT A, INPUT B, and A±B in Scope mode, needed for future mea­surements. You can save a maximum of 20 wav eforms (For FLUKE 96 you can save a maximum of 10 waveforms).
Press (SAVE WAVEFORM).
The test tool starts automatically with the next available memory locations.
Saving a Setup
Use this function to save your keystroke sequences. Once the test tool is set up to make the required measurement, you can save this configuration as one of a maximum of 40 setups (For FLUKE 96 you can save a maximum of 20 set­ups). A setup is simply the current operating configuration of your ScopeMeter test tool.
Press (SAVE SETUP).
The test tool starts automatically with the next available memory.
Copying Waveforms or Settings
To copy a waveform or setting from one memory to another ,
91
92
Save menu
Loading...