Fluke 07-653 User Manual

Page 1
Nuclear Associates 07-653
Fluoroscopic Imaging Test Phantom
Users Manual
Page 2
Fluke Biomedical
Radiation Management Services
6045 Cochran Road Cleveland, Ohio 44139
440.498.2564
www.flukebiomedical.com/rms
Page 3
Table of Contents
Section 1: General Information................................................................................... 1-1
1.1 Parts ........................................................................................................... 1-1
1.2 Assembly .................................................................................................... 1-1
1.3 Measurement Setup .................................................................................... 1-1
1.4 Procedure .................................................................................................... 1-3
i
Page 4
(Blank Page)
Page 5
1.1 Parts
Phantom Base Assembly
Thin Clip-On Attenuator (X.X mm Cu)
Thick Clip-On Attenuator (Y.Y mm Cu)
1.2 Assembly
Pediatric patient simulation – phantom without attenuators
Small adults – phantom with thin clip-on attenuator
General Information
Parts
1
Section 1
General Information
Larger adults – phantom with thick clip-on attenuator
Heavy adults – phantom with both clip-on attenuators
1.3 Measurement Setup
The image intensifier’s grid is either in-place or removed in accordance with the clinical protocol being simulated.
The entrance surface of the image intensifier is fragile. Additional attention is required if the grid is removed.
Fasten the phantom to the entrance surface of the image intensifier. (In some cases, the clips on the phantom will engage slots on the image intensifier assembly.)
Set systems with variable source to image receptor instance (SID) to a standard value (e.g. 100 cm).
Remove extraneous material from the x-ray beam.
The tabletop and /or patient mattress may be in place if desired.
CAUTION
1-1
Page 6
Nuclear Associates 07-653
Operators Manual
Substrate: uniform
density plate
Dual grey scale
steps to set video
level, gamma,
peak whites
Lead disks to set
black level
White dots to
check resolution,
preamp frequency
characteristics,
apeture correction
07-653
Large circles to
set sweep linearity
Y
X
Two absorber plates (not shown) to check
operation of automatic
brightness system
Figure 1-1.
Figure 1-2.
The fluoroscopic Imaging Test Phantom provides a test pattern, enabling the precise adjustment of many critical parameters of the Fluoroscopic System.
1-2
Page 7
General Information
Procedure
1.4 Procedure
Operate the system using fluoroscopic and acquisition modes under investigation.
Note the electrical setting on the control panel (e.g. kV, mA, ms).
Note the visibility of the small holes in each of the five cross targets.
Note the visibility of the six steps in each of the two white targets.
Note the visibility of the two lead (black) targets. These targets establish the black level for the
system.
Image analysis may be visual, video or digital.
1
1-3