Care must be taken to ensure radiation beam is off
prior to placing the unit into position.
To avoid erroneous results, the unit must be
positioned in the center of the beam. Failure to
position the unit in the center of the beam would
cause unexpected results do to the difference in the
beam spectrum at various locations of the beam.
This device is intended for use only by persons who
have been trained in the proper interpretation of its
readings and in the appropriate safety procedures
to be followed in the presence of radiation.
Use of instrument while "LO BAT" indicator is
displayed may result in erroneous measurements
due to the loss of an accurate voltage reference.
WARNING
CAUTION
1.1 Product Description
The Model 07-492 Mammography kVp Meter is designed to measure the effective peak potential applied
to mammographic x-ray tubes. The measurement is made without direct connection to the x-ray unit
being tested.
The kVp Meter is a battery powered, self-contained instrument designed for portable use. Two modes of
operation are available:
•Mo/Mo (Phase 1) - For use with single-phase x-ray machines using moly target (anode) and moly
filtration.
•Mo/Mo (Phase 3) - For use with three phase x-ray machines using moly target (anode) and moly
filtration.
1-1
Page 6
Nuclear Associates 07-492
Operators Manual
NOTE
Correction factors for different target/filter
combinations are provided in Section 5, Calibration
and Adjustment.
The kVp meter uses two differentially filtered detectors whose ratio of integrated outputs is calibrated over
the range of 24 to 40 kVp. Measurement results are presented on an easy to read 3½ digit Liquid
Crystal Display. Status is indicated by decimal point positioning on the readout. Low battery condition is
indicated by LO BAT displayed on the readout.
1.2 Specifications
Range Mo: 24 kVp to 40 kVp
Accuracy ± 1.5 kV
Reproducibility ± 0.5 kV
Resolution 0.1 kV
Distance Dependence Negligible for SSD of 20 cm or greater
Angle Dependence Negligible for 10 degrees or less
Position Dependence Negligible for misalignment of 5 cm or less at 30 cm.
Calibration Period One year
Filtration Effects 1.5 kV per mmAl at 25 kVp
4.0 kV per mmAl at 35 kVp
Minimum Requirements 75 mA @ 24 kVp, 25 cm
Power Requirements 9V Alkaline Battery, MN1604A or equivalent
Display 3 ½ Digit LCD
Low Battery Indication: LO BAT
High kV Indication: Left Blinking Decimal
Low kV Indication: Middle blinking Decimal
High Intensity Indication: Right blinking decimal and non-zero display.
Low mAs Indication: Right blinking decimal and zero display.
LED Indicator Auto Reset, new exposure detected, and last reading cleared.
Controls Front Panel: Power switch.
Rear Panel: Phase selection switch
Output Connector BNC Connector on rear panel for waveform output.
Dimensions (H x W x D) 2.5 in. x 8 in. x 6 in. (6.4 cm x 20.3 cm x 15.2 cm)
Weight 2.1 lbs. (.95 kg)
Temperature Limits Operating: +18° to + 40° C (+64° to + 104° F)
Maximum 90% relative humidity, non-condensing
Storage: -18° to + 55° C (0° to + 132° F)
1-2
Page 7
General Information
Receiving Inspection
1.3 Receiving Inspection
1. Check the packaging and its contents for damage. If damage is evident, file a claim with the carrier
and contact Fluke Biomedical, Radiation Management Services at 440.248.9300 for further
instructions.
2. Remove the packaging and visually inspect the unit for damage.
3. Verify that the following items have been received and are in good condition:
• One Model 07-492 Mammography kVp Meter, Part No. 164005
• One MN1604A (or equivalent) 9V Alkaline Battery, Part No. 16-29
• One 07-492 Operator's Manual, Part No. 164001
NOTE
If items are missing or damaged, contact Fluke
Biomedical at 440.248.9300 for further instructions.
1.4 Storage
1
If necessary, the kVp meter may be stored in an area free of corrosive materials, temperature and
humidity fluctuations, vibration, and shock. In addition, the storage area should satisfy the environmental
specifications of the unit (refer to Section 2, Indicators and Controls).
1.5 Routine Cleaning
Do not immerse the Model 07-492 Mammography
Digital kVp Meter. The unit is not waterproof.
Liquid could damage the circuits. The unit should
be kept clean and free from dirt and contamination.
The unit may be cleaned by wiping with a damp
cloth using any commercially available cleaning or
decontaminating agent.
CAUTION
1.6 Procedures, Warnings, and Cautions
The equipment described in this manual is intended to be used for the detection and measurement of
ionizing radiation. It should be used only by persons who have been trained in the proper interpretation of
its readings and the appropriate safety procedures to be followed in the presence of radiation.
Although the equipment described in this manual is designed and manufactured in compliance with all
applicable safety standards, certain hazards are inherent in the use of electronic and radiometric
equipment.
WARNINGS and CAUTIONS are presented throughout this document to alert the user to potentially
hazardous situations. A WARNING is a precautionary message preceding an operation which has the
potential to cause personal injury or death. A CAUTION is a precautionary message preceding an
operation which has the potential to cause permanent damage to the equipment and/or loss of data.
Failure to comply with WARNINGS and CAUTIONS is at the user’s own risk and is sufficient cause to
1-3
Page 8
Nuclear Associates 07-492
Operators Manual
terminate the warranty agreement between Fluke Biomedical, Radiation Management Services and the
customer.
Adequate warnings are included in this manual and on the product itself to cover hazards that may be
encountered in normal use and servicing of this equipment. No other procedures are warranted by Fluke
Biomedical. It shall be the owner’s or user’s responsibility to see to it that the procedures described here
are meticulously followed, and especially that WARNINGS and CAUTIONS are heeded. Failure on the
part of the owner or user in any way to follow the prescribed procedures shall absolve Fluke Biomedical
and its agents from any resulting liability.
Indicated battery and other operational tests must be performed prior to each use to assure that the
instrument is functioning properly. If applicable, failure to conduct periodic performance tests in
accordance with ANSI N323-1978 (R1983) Radiation Detection Instrumentation Test and Calibration,
paragraphs 4.6 and 5.4, and to keep records thereof in accordance with paragraph 4.5 of the same
standard, could result in erroneous readings or potential danger. ANSI N323-1978 becomes, by this
reference, a part of this operating procedure.
1.7 Inspection and Material Return Instructions
Instruments should be examined and tested as soon as received by the purchaser. Claims for damage, if
any, should be filed at once with the carrier. Any material returned for repair must be accompanied by a
valid customer purchaser order, identifying the work to be done. A Material Return Form is provided at
the back of this manual. Send the completed form with items returned for repair to enable our Sales
Personnel to process the order as quickly as possible.
Material valued at $200.00 or more and/or weighing more than twenty pounds should be shipped the best
way prepaid and fully insured.
Fluke Biomedical suggests that any instrument weighing over twenty pounds be wrapped in heavy Kraft
paper and packed in a double corrugated carton or wooden box. Protect the instrument on all sides with
at least three inches of excelsior or similar padding. Mark the case plainly with suitable caution warnings
to insure careful handling.
1-4
Page 9
Indicators and Controls
Front Panel
Section 2
Indicators and Controls
2.1 Front Panel
See Figure 2-1 for front panel control and indicator location.
2.2 Numeric Readout
The front panel 3 ½ numeric digit LCD readout displays the measurement data. Also indicated on the
display are low battery, overrange kVp, underrange kVp, high mAs, and low mAs conditions as listed in
Table 2-1.
Table 2-1. Displayed Indications
2
Condition Indication
Low Battery LO BAT in the upper left corner of the display.
Overrange kVp Flashing decimal point directly above A.
Underrange kVp Flashing decimal point directly above B.
High mAs Flashing decimal point directly above C/D and a non-zero numeric display.
Low mAs
Figure 1-1. Front Panel Indicators and Controls
An overrange kVp condition indicates that the measured kVp is above the calibrated range. An
underrange kVp condition indicates that the measured kVp is below the calibrated range.
Flashing decimal point directly above C/D and a display numerically equal to
zero.
A high mAs condition indicates that the beam intensity is too high and the measurement is completed with
less than 50 ms of the radiation waveform. The resulting measurements may be less accurate than those
obtained with longer sample times. This is especially true in the case of single-phase x-ray machines
where the radiation spectrum changes dramatically throughout the duration of a pulse. For example, if
the measured waveform contains less than six pulses of single-phase rectification, then a partial pulse
can have a significant contribution to the value of the average of the pulses. The effect is not as great for
three phase machines because the radiation spectrum remains relatively constant throughout the
2-1
Page 10
Nuclear Associates 07-492
Operators Manual
exposure. To improve accuracy, the user should either increase distance or decrease the mA, both of
which will reduce the beam intensity.
A low mAs condition indicates that the exposure is less than the required value to make a measurement.
The user should either increase time or mA or decrease distance for the particular kVp setting. Refer to
Section 3, Battery Installation and Setup, for a discussion of minimum mAs requirements.
The auto reset on the front panel will turn on whenever the kVp meter detects an x-ray exposure. The
LED will remain lit for 0.5 seconds or for the duration of the exposure, whichever is greater. The previous
reading is cleared when the LED lights.
2.3 Switches
The power switch is located on the front panel. The power switch is used to turn the unit on or off. A
phase switch is located on the rear panel which allows selection for single phase and three phase
applications.
2.4 Top Panel
The top panel of the kVp meter contains the following information:
• A brief set of operating instructions.
• An indication as to where to center the beam, referred to as the target.
• Display guide for A, B, C, and D decimal point indications as discussed above.
2.5 Rear Panel
The battery access panel and a BNC connector are located on the rear panel. Refer to Section 3, Battery
Installation and Setup, for battery replacement procedures. The BNC connector allows the user to
connect an oscilloscope to the kVp meter to view the signal from the radiation detection diodes. The
signal can be observed on the oscilloscope for the duration of the exposure. A storage scope or camera
is necessary to view the signal for extended periods of time.
The phase selector switch is also located on the rear panel of the kVp meter. The switch has two
positions, one phase and three phase; position is determined by the type of x-ray machine being tested.
2-2
Page 11
Battery Installation and Setup
Installing/Replacing the Battery
3
Section 3
Battery Installation and Setup
3.1 Installing/Replacing the Battery
Use the following procedure to install/replace the 9 V Alkaline (P/N 16-29) battery used to power the kVp
meter:
1. Be sure the power switch located on the front panel is in the OFF position.
2. Open the battery cover door, located on the rear panel, by pulling on the access tab until it opens.
3. If applicable, remove the used battery. Connect the new battery to the terminal connector following
standard polarity conventions.
4. Place the battery into the battery compartment.
5. Close the battery compartment door.
3.2 Setting the Phase Switch (Rear Panel)
The Phase Switch position is determined by the type of x-ray machine to be tested as follows:
• The single-phase (1) position should be selected for testing single-phase x-ray machines.
• The three-phase (3) position should be selected for testing three phase x-ray machines.
3.3 Detector Positioning
The kVp meter should be positioned so that the red detector area on the top panel of the unit is centered
in the x-ray beam. The x-ray beam should cover at least the red portion of the case top to assure an
accurate measurement. Placement of the kVp meter relative to the x-ray tube is shown in Figure 3-1.
3-1
Page 12
Nuclear Associates 07-492
Operators Manual
Figure 3-1. Detector Positioning
3.4 Determining mAs Requirements
The mAs minimum requirements as a function of distance and kVp are indicated in Figure 3-2. When the
mA has been determined from the chart as discussed in this section, the mAs should be calculated using
an exposure time of 50 ms (1/20 second) or greater.
The curves in Figure 3-2 are intended to guide the user in selecting the mAs required for a particular kVp
measurement. Any mAs beyond the user-designated amount will not be included in the measurement.
(Refer to Section 4, Theory of Operation, for a detailed discussion.)
Use the following guidelines to determine the required mAs from the graph in Figure 3-2:
1. Locate the kVp to be measured on the horizontal axis.
2. From the kVp point on the axis, move vertically to the selected FDD curve.
3. From the kVp position on the appropriate FDD curve, move horizontally to the vertical axis and the
read the required mAs.
4. If necessary, interpolate between distances using the inverse square law.
The curves in Figure 3-2 are derived from single-phase measurements. Three phase machines will
generally require 70% of the mAs determined from the curves.
mAs requirements may vary considerably depending on the following conditions:
1. Detector sensitivity
2. Tube filtration
3. Radiation waveform
4. FDD
5. mAs errors in the x-ray machine
If the determined kVp fails to produce a result, double the time (or decrease the FDD by 30%) and try
again.
3-2
Page 13
Battery Installation and Setup
Determining mAs Requirements
Typically, the largest apparent variation in mAs requirements will be due to a kVp error. If doubling the
mAs fails to produce a result, increase the kVp setting (or decrease the FDD) until a result is achieved.
1000
7
5
4
3
2
Digital kVp Meter:
Typical single phase mA Requirements
May vary on different machines
mA
100
10
7
5
4
3
2
7
5
4
3
2
FDD = 40 cm
80 cm
60 cm
1
22 24 26 28 30 32 34 36 38 40 42 44
kVp
Figure 3-2. Minimum Requirements (mA vs. kVp)
3
3.5 Initial and Periodic Operation Checks
Generally it is not easy to verify the accuracy of the kVp meter in normal field usage. If care is taken,
either of the methods outlined in the following sections can be used to verify performance of the meter.
Both methods discussed require a careful collection of installation data and periodic measurement of that
data.
3.6 Voltage Divider Method
Use the following procedure to perform initial and periodic operation checks of the kVp meter if an x-ray
voltage divider is available:
1. Choose two x-ray systems suited for the test, preferably two different systems, i.e. three phase vs.
one phase, falling load vs. not, etc. Choosing two dissimilar x-ray machines for the check reduces
the possibility of systematic errors.
2. Make comparative measurements of the x-ray systems using the voltage divider and the kVp meter.
3-3
Page 14
Nuclear Associates 07-492
Operators Manual
NOTE
If an oscilloscope is used to measure the voltage
divider, an independent voltmeter should be used
to measure scope sensitivity before each operation
check.
Data with a kVp test cassette would be useful. Note positioning error and film usage when applying the
cassette.
3. At installation, record all parameters of the measurement in a bound notebook. Also record
geometry, cable connections, tube I.D., generator I.D., divider and kVp meter oscilloscope
waveform pictures, oscilloscope and voltmeter I.D, and any other information that may prove useful
for periodic operation checks.
4. Periodically (every six months), or whenever a serious discrepancy in kVp operation is noted, the
above tests should be repeated to determine if the unit is operating properly. The kVp meter,
divider, cassette result, and x-ray settings should be examined for consistency.
3.7 Multiple X-Ray Machine Method
If a voltage divider is not available, use the following procedure to perform initial and periodic operation
checks of the kVp meter:
1. Choose several x-ray systems suited for the test. If possible, choose at least two different systems,
i.e. three phase vs. one phase, falling load vs. not, etc. Choosing dissimilar x-ray machines for the
check reduces the possibility of systematic errors.
2. Make comparative measurements of the x-ray systems using the kVp meter.
Data with a kVp test cassette would be useful. Note positioning error and film usage when applying the
cassette.
3. At installation, record all parameters of the measurement in a bound notebook. Also record
geometry, cable connections, tube I.D., generator I.D., kVp meter oscilloscope waveform pictures,
oscilloscope I.D, and any other information that may prove useful for later periodic operation
checks.
4. Periodically (every six months), or whenever a serious discrepancy in kVp operation is noted, the
above tests should be repeated to determine if the unit is operating properly. The kVp meter,
cassette result, and x-ray settings should be examined for consistency. If the kVp meter suddenly
produces different results from previous tests on a specific x-ray machine, then there is a strong
probability that the kVp meter is operating properly. However, if all of the x-ray measurements
suddenly change, then the kVp meter may be malfunctioning.
3-4
Page 15
Theory of Operation
General
Section 4
Theory of Operation
4.1 General
The kVp measurement is computed basically from a measurement of the linear absorption coefficient
(MU) of the hardened x-ray beam. As the kV increases, the linear absorption coefficient increases.
An x-ray beam is composed primarily of two parts, the bremsstrahlung radiation and the characteristic
radiation. For a sufficiently hardened beam, if a plot is made of the log of the kV vs. the ratio of the
bremsstrahlung part of the beam, a nearly straight line would result.
4.2 Filtration Effects
A slight change in the beam spectrum being measured will cause a change in the linear absorption
coefficient. Such a change can be caused by filtration differences with respect to the calibration beam.
With lower filtration, the x-ray beam will not be as hard as the calibration beam and the results will be
lower. With more filtration, the beam will be harder than the calibration beam and the results will be
higher.
4
4.3 Waveform Effects
Beam spectrum changes occur with different waveforms. A single-phase waveform, as opposed to a
three-phase waveform, displays an approximate 5% change in readings. Correction for this is made by
selecting the appropriate phase switch position. The Phase switch is located on the rear panel. If loading
on the generator causes the waveform to appear in the classic manner, the switch should be positioned in
the three-phase position. In some situations, the waveform may be somewhat between the single-phase
and the three-phase positions and the results may be compromised.
4.4 Low Battery
The low battery indication is displayed when the battery voltage drops below a predetermined value. This
value is selected such that there is not a noticeable effect on the measurement results. Use of the
instrument below this value (i.e. while the LO BAT indication is displayed) may result in erroneous
measurements due to the loss of an accurate voltage reference.
4.5 Positioning Error
The kVp meter should be positioned in the center of the beam. The measurement of the meter is two
square inches. If measurements are made in other parts of the beam, different results will occur since the
beam spectrum is different at different locations. The accuracy of the kVp meter will be compromised.
The heel effect is an example of such a beam spectrum change. To view the effects of beam positioning
on the measurement, the user may want to position the meter in various parts of the beam for illustration
purposes.
4-1
Page 16
Nuclear Associates 07-492
Operators Manual
(Blank page)
Page 17
Calibration and Adjustment
Lo Battery Threshold Adjustment
Section 5
Calibration and Adjustment
The Model 07-492 kVp Meter has been factory calibrated. Any calibration should be done by the factory
or a qualified calibration facility. The unit should be calibrated once a year by adjusting the five variable
resistors as discussed in the following sections.
5.1 Lo Battery Threshold Adjustment
Resistor R25 adjusts the threshold for the LO BAT indication on the LCD Display. Use the following
procedure to adjust R25:
1. Remove the battery.
2. Connect a variable DC power supply to the instrument at the battery connector cable.
3. Connect a voltmeter between battery plus (TP3) and analog ground (TP4).
4. Turn on the power supply and adjust it to 9 Volts.
5. Turn on the kVp meter. The voltmeter should read between 2.4 and 3.2 V.
6. Slowly decrease the power supply until the voltage across the test points just returns to the value it
had prior to dropping off.
7. Turn R25 until LO BAT disappears from the display.
8. Turn R25 slowly in the opposite direction just until LO BAT reappears on the display.
9. Connect the voltmeter between analog ground and the power supply negative terminal.
10. Increase the power supply voltage until the voltmeter reads 4.5 V while being careful not to exceed
9V from the power supply. LO BAT should be off at this point.
5
5.2 kVp Calibration Adjustments
The following are some guidelines to ensure proper calibration of the kVp meter:
1. An independent accurate means of measuring the x-ray machine is necessary.
2. The kVp meter must be calibrated at 26 kVp, 30 kVp, and 40 kVp (± 1 kVp) if it is to be within
specification over its entire range.
3. The potentiometers must be adjusted interactively as detailed below.
4. Be sure to set the phase switch, located the rear panel of the unit, appropriately.
5. Be sure to follow the procedure outlined below in the order listed.
Use the following procedure for kVp calibration:
1. Set a calibrated Moly Anode/Moly Filter x-ray machine to 26 kVp, with sufficient mAs to obtain a
reading.
2. Locate the Model 07-492 kVp meter in the center of the beam with the cover removed from the
meter.
5-1
Page 18
Nuclear Associates 07-492
Operators Manual
NOTE
Be sure the phase switch on the rear panel of the
kVp meter is in the appropriate position.
3. Make an exposure.
4. Adjust R106 (offset adjustment) until the kVp meter reads 26 kVp.
5. Set the x-ray machine to 30 kVp.
6. Make an exposure.
7. Adjust R105 (slope adjustment) until the kVp reads 30 kVp.
8. Repeat Steps 1 through 7 until the readings are accurate.
9. Set the x-ray machine to 40 kVp.
10. Make an exposure.
11. Adjust R104 (high end compensation adjustment) until the kVp reads 40 kVp.
12. Repeat Steps 5 through 11 until the readings are accurate.
13. Repeat Steps 1 through 11 until the readings are accurate.
After the Model 07-492 kVp Meter has been calibrated using the moly/moly filter combination, the
following correction table can be used when operating with different filter combinations.
Target/Filter Energy in kVp Correction Factor
Rh/Rh 25.0 0.903
30.0 0.993
35.0 1.057
39.9 1.097
44.9 1.126
48.9 1.144
Rh/AI 25.0 0.856
30.0 0.912
35.0 0.951
39.9 0.988
44.9 1.027
48.9 1.055
Mo/Rh 23.0 0.911
25.0 0.943
30.0 1.007
35.0 1.039
40.0 1.047
Mo/AI 24.0 0.888
27.1 0.890
30.0 0.902
35.0 0.926
40.0 0.955
45.0 0.981
49.0 1.004
5-2
Page 19
(Blank page)
Page 20
Fluke Biomedical
Radiation Management Services
6045 Cochran Road
Cleveland, Ohio 44139
440.498.2564
120 Andrews Road
Hicksville, New York 11801
516.870.0100
www.flukebiomedical.com/rms
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.