The ST75 and ST75V (Vortab) can be specified with integral or remote electronics. The flow element has a serial number etched into
the side of the HEX as shown on figure A. In addition, the tag on the enclosure will also denote serial number and model number. The
associated transmitter circuit card has a serial number noted on the board as shown in figure B. The flow sensor and transmitter
circuit have been calibrated as a matched set and should be paired together in service unless otherwise approved by a factory technician.
Flow Direction Alignment
All sensor elements have a flow arrow indicator marked on the element assembly at the reference HEX flat. These flow elements have
been calibrated in a particular direction and are designed to be used in service with the flow arrow going in the same direction as flow
in the pipe stream. The flow element has been calibrated directly in the pipe tee or tube tee with careful consideration for orientation
and insertion depth. Removal from the tee section while physically possible is not recommended unless authorized by a factory
technician.
Flow Direction
FLOW
Serial Number
Figure A
Recommended Straight Run
To optimize flow meter system performance, FCI recommends installation with a minimum of 20 diameters upstream straight run and
10 pipe diameters of downstream straight run. Where straight run limitations significantly reduce the available pipe diameters, FCI
utilizes flow conditioners to produce a transferable flow profile from the calibration installation to actual field installations. FCI’s
proprietary AVAL software is available to make flow meter installation evaluations where straight run limitations are considered. See
Figure C for recommended installation.
FLUID COMPONENTS
INTERNATIONAL LLC
0.000 SFPS
32.2 DEGREES F
Figure B
This page is subject to proprietary rights statement on last page
Figure C - Recommended Straight Run Condition
FLUID COMPONENTS INTERNATIONAL LLC ST75/ST75V MASS FLOW
Specifications
Media Compatibility:
Air, compressed air, nitrogen, oxygen, argon, CO2, ozone,
other inert gases, natural gas, and other hydrocarbon
gases, hydrogen.
Min. SCFM 0.01 0.05 0.25
Min [NCMH] [0.01] [0.09] [0.42]
Max. SCFM 3.02 21.15 99.08
Max. [NCMH] [5.14] [35.94] [168.33]
* actual range subject to gas type and specific conditions.
Accuracy:
ST75: ± 2% of reading, ± 0.5% of full scale
Optional: ± 1% of reading, ± 0.5% of full scale
ST75V: ± 1% of reading, ± 0.5% of full scale
Repeatability:
Temperature Compensation:
Standard: 40 to 100°F [4 to 38°C]
Optional: 0 to 250°F [-18 to 121°C]
± 0.5% of reading
Transmitter
Enclosure:
with either 1/2" Female NPT or M20x1.5 entries. Epoxy coated.
Output Signals:
(2) 4-20 mA user assignable to flow rate and/or temperature
(1) 0-1 kHz pulse for total flow
Communication Port:
to PDA with digital display models
Input Power:
DC: 18 Vdc to 36 Vdc (6 watts max.)
AC:
Operating Temperature Range:
Digital Display:
Two-line x 16 characters LCD. Displays measured value and
engineering units. Top line assigned to flow rate. Second line is
user assignable to temperature reading, flow totalizer or
alternating. Display can be rotated in 90° increments for
optimum viewing orientation.
NEMA 4X [IP67], aluminum, dual conduit ports
RS232C standard. Optional wireless IR
85 Vac to 265 Vac (12 watts max.; CE Mark Approval for 100
Vac to 240 Vac)
0 to 140°F [-18 to 60°C]
(Optional)
Turndown Ratio: 10:1 to 100:1
Agency Approvals:
ATEX/IEC Ex: II 2 G Ex d IIC T6...T3
II 2 D Ex tD A21 IP67 T90°C...T121°C
FM, CSA:Class I, II, III, Div. 1, Groups B-G
Class I, Div. 2, Groups A-D
CRN No.:0F0303
Warranty:
One year
Flow Element
Installation:
ST75: In-line “T”, NPT or tube
Material of Construction:
probe element with Hastelloy-C thermowells; 316 stainless
steel NPT and tube fittings
This page is subject to proprietary rights statement on last page
2 Doc. No. 06EN003368 Rev. C
ST75/ST75V MASS FLOWFLUID COMPONENTS INTERNATIONAL LLC
FCI Flow Meters may be installed with less than the recommended straight run, but may have performance limitations. FCI offers
VORTAB flow conditioners for use in applications that have significant straight run limitations. FCI uses the AVAL application
modeling software to predict meter performance in each installation. AVAL outputs are available to review prior to order placement
and will indicate performance expectations both with and without flow conditioning.
Flow Element Installation
Warning: The element is shipped specifically installed in the TEE oriented for inline installation. DO not remove the
sensing element from the TEE section during installation as calibrated performance can be effected.
Process Connections
The ST75 is available in pipe Tee configurations with NPT threads and tubing tees with a compression fitting suitable to clamp down on
concentric smooth surface tubing. The pipe Tee versions are standard 150# class rated tees suitable for service up to 150 PSIG at the
process temperature maximum of 250°F (121°C). The compression fitting material offered in the tube type configuration is rated for
250 PSIG service.
Pipe Tee Installation: With pipe extensions properly cut to length and the appropriate sealing materials used on the threads, install flow
element section by slowly rotating the configuration until firmly secure on the pipe section. Complete by installing opposing end pipe
section using care to firmly secure the element assembly either in a top mount or side mount position.
Tube Tee Installation: Clean all mating surfaces of the tee fitting, ferrules and the flow tube. Insert the flow tubing into the tee fitting.
Make sure the tubing rests firmly in the fitting counter bore seat. Tighten the nut on both ends of the tee by hand. Hold the fitting body
steady with a backup wrench, tighten the fitting nuts 1-1/4 turns, from hand tight baseline.
The ST75V is available in with flow tube configurations offering male and female NPT threads, ANSI flanges and DIN flanges. The
flow tube assemblies are rated for service up to 240 PSIG at the process temperature maximum of 250°F (121°C).
NPT Flow Tube Installation: With pipe extensions properly cut to length and the appropriate sealing materials used on the threads,
install flow element section by slowly rotating the configuration until firmly secure on the pipe section. Complete by installing opposing
end pipe section using care to firmly secure the element assembly either in a top mount or side mount position.
Flanged Installation: Clean all mating surfaces. Install appropriate sealing gasket between mating flanges. Tighten flange mating hardware to meet system sealing requiments.
See Appendix C for instrument outline dimentional details.
This page is subject to proprietary rights statement on last page
3 Doc. No. 06EN003368 Rev. C
FLUID COMPONENTS INTERNATIONAL LLC ST75/ST75V MASS FLOW
Instrument Wiring
Before the instrument is opened to connect power and signal, FCI recommends that the following ESD precautions be observed:
Use a wrist band or heel strap with a 1 megaohm resistor connected to ground. If the instrument is in the shop setting, there should be
a static conductive mat on the work table or floor with a 1 megaohm resistor connected to ground. Connect the instrument to ground.
Apply antistatic agents such as Static free made by Chemtronics (or equivalent) to hand tools to be used on the instrument. Keep
high static producing items away from the instrument.
The above precautions are minimum requirements. The complete use of ESD precautions can be found in the U.S. Dept of defense
handbook 263.
Warning: Only Qualified personnel are to wire or test this instrument. The operator assumes all responsibility for safe
practices while wiring and trouble shooting.
FCI recommends installing and input power disconnect switch and fuse near the instrument to interrupt power during
installation and maintenance. Operator must have power disconnected before wiring.
See Safety instructions in Appendix A for the use of the ST75/ST75V series (AC and DC versions) in Hazardous Areas
Category II (Zone 1). Approval, KEMA 08ATEX0045/IECEx KEMA08.0012 for Category 2 GD protection Ex d IIC
T6..T1, Ex tD A21 IP67 T 90°C...Τ 121°C.
Input Power
The ST75/ST75V is available with both VDC and VAC input power configurations. Customers selecting VDC input power will
have a VDC input board only. Similarly, the VAC power board is supplied only with VAC powered units. In addition, both boards are
marked for either AC or DC power. Only connect the power specified on the wiring module as shown on Figures E and F respectively. Both VAC and VDC inputs require a Gnd wire to be connected. Input power terminal blocks are rated for 14-26 AWG.
To wire the instrument, ensure that the power is off. Pull the power and signal output wires through the port, using care not to
damage wires. FCI recommends using crimp lugs on the output wires to ensure proper connection with the terminal strip. Connect
the output wires as shown on figures E and F. Note that when the 4-20mA outputs are used simultaneously, a single return lead is
used.
Analog Output
4-20mA: The instrument is provided with a standard set up, of two 4-20mA outputs. Output 1 configured for flow and Output 2
configured for temperature. Terminal blocks rated for 14-28 AWG., 500 ohm max load per output.
This page is subject to proprietary rights statement on last page
4 Doc. No. 06EN003368 Rev. C
ST75/ST75V MASS FLOWFLUID COMPONENTS INTERNATIONAL LLC
Pulse Output Activation
The ST75/ST75V provides a pulse output feature. Instruments ordered with volumetric or mass flow units will be factory set with
totalizer and pulse output activated. The mode can be changed in the field. Wiring either sink or source mode is shown in Figures E
and F below. Though only one configuration is shown with the VAC and VDC power supplies, the source or sink can be utilized with
either power input.
Sink Mode: 40 VDC Max, 150 mA max. Customer supplied power source
Source Mode: 15 VDC output, 50 mA max
VDC Power Connection
RS 232 Connection
VAC Power Connection
RS 232 Connection
Figure E
VDC Power
As Shown:
18-36VDC power connected with gnd
4-20mA connected for flow and temperature
Pulse Out in source mode
Note: In source mode, 15VDC Output max, 50mA max.
Power Disspation
AC Version
Power dissipation values under nominal conditions:
Instrument (Electronics + Sensor): 11.6 Watts
Sensor only: 0.25 Watts
Figure F
VAC Power
As Shown:
85-265 VAC power connected with gnd
4-20mA connected for flow and temperature
Pulse Out in sink mode
Note: In sink mode, 40VDC max, 150mA max customer
supplied power source.
DC Version
Power dissipation values under nominal conditions:
Instrument (Electronics + Sensor): 4.5 Watts
Sensor only: 0.25 Watts
Power dissipation values under maximum load conditions:
Instrument (Electronics + Sensor): 12 Watts
Sensor only: 0.30 Watts
This page is subject to proprietary rights statement on last page
Power dissipation values under maximum load conditions:
Instrument (Electronics + Sensor): 6 Watts
Sensor only: 0.30 Watts
5 Doc. No. 06EN003368 Rev. C
FLUID COMPONENTS INTERNATIONAL LLC ST75/ST75V MASS FLOW
Setup Interface
All parameters on this meter are set through the RS232 interface connection (P3 plug) or PDA IR interface. A jumper selection
determines which communication mode is active. The factory default communication mode is set for the RS232 interface. This
setting allows the instrument to be setup with either a FC88 hand held communicator or a computer. The FC88 is powered through the
meter and comes with the serial interface cable. If a computer interface is used, an adapter (RJ to 9 pin Computer Serial Port) is
required and may be obtained from FCI: Part No. 014108-02.
Using Windows Terminal (usually located in Accessories) execute the program by double-clicking on the Terminal Icon.
1. Go to Settings.
2. Click on Communication.
3. Set for COM1 or COM2, 9600 Baud, 8 Bit, and No Parity. Press OK
4. Press the ENTER key to see the Input Mode? prompt.
5. Enter any of the meters single letter commands to execute a function (reference complete function menu in Appendix B).
If the PDA IR interface is used for communication, then jumper JP5 needs to be moved to the alternate position, see Figure G and H.
See PDA IR Communication Interface section for more details.
An additional command line interface (CLI) is available through the RS232 port. This interface is accessed with the “Y” command
using a computer or FC88. The command line password is “357”. See Appendix B - Table 6 for command line details.
JP5
Figure G
JP5 factory set for RS232 interface
This page is subject to proprietary rights statement on last page
Figure H
JP5 set for PDA IR interface
Infra Red Communications Window
Flow Rate, Total and Temperature Indicator
6 Doc. No. 06EN003368 Rev. C
ST75/ST75V MASS FLOWFLUID COMPONENTS INTERNATIONAL LLC
Start up and Commissioning
1.Verify all Input power and output signal wiring is correct and ready for initial power start up.
2.Apply power to instrument. The instrument will initialize in the Normal Operation Mode. All outputs will be active and instruments with the display option will indicate flow with the factory set flow unit. Allow 10 minutes for the instrument to warm up
and come to the thermal equilibrium.
The following FC88 commands are typical commands that are used during start up and commissioning:
CommandNameDescription
TNormal Operation ModeAll outputs are active
ZFlow Unit Set-UpSelect Flow Units(4 English, 4
Metric)Pipe Dimensions
WTotalizerEnable/Disable
VOutput ConfigurationSelect one of 4 Configurations:
Pulse and/or AlarmPulse factor and/or
setpoint
FK-Factor (default=1)Flow factor
NWarm Re-setRe-initialize C/B
STotalizer MenuEnables W menu (Option)
If the instrument is installed, and the process flow is zero, the instrument should indicate 0.000.
Flow Unit Modification
Example: SCFM Flow Units and 3 inch Sch 40 round pipe size set up:
EnterDisplayDescription
Entermenu: >From Normal Operation Mode
ZE for English M for Metric > Flow Unit Set-Up menu
E
0=SFPS, 1=SCFM, 2=SCFH, 3=LB/H,
4=GPM #
1R round duct or S rectangular>Select Standard Ft
RSelect Round Duct
Dia.: 4.0260000
Change? (Y/N)>
YEnter value: #
3.0683 inch Sch. 40 pipe I.D.
N
Y
area: 7.3926572 CMinflow: 0.0000000
Change? (Y/N)>
Maximum flow: 462.04
Enter to continue
Cmaxflow: 462.04
Change? (Y/N)>
Y#
462.04
N
N
N
CMintemp (F): -40.00000
Change? (Y/N)>
CMaxtemp (F): 250.00000
Change? (Y/N)>
Percent of Range is: OFF
Change to ON?>
LCD Mult Factor x1
Change? (Y/N)>
N100.0 SCFM
English units
/Min (SCFM)
3
Instrument will end up in Normal
Operation Mode
This page is subject to proprietary rights statement on last page
7 Doc. No. 06EN003368 Rev. C
FLUID COMPONENTS INTERNATIONAL LLC ST75/ST75V MASS FLOW
RS232 / FC88
Menu Control and Organization
Most entries require at least two key strokes; a Capital letter and the [ENTER] key, or one or more numbers and the [ENTER] key. All
user entries begin at the input mode prompt “>”,except when the instrument is in the Main Function Mode (just press the desired
function letter and [ENTER] to make an entry).
Backspaces are made using the backspace [BKSP] key. Some entries are case sensitive between numbers and letters. Be sure the
SHIFT key is pressed to indicate the correct case. A square after the prompt caret indicates the FC88 is in lower case. A slightly
raised rectangle in the same spot indicates the FC88 is in the upper case.
It is recommended that the FC88 be plugged into the instrument before power is applied. If the FC88 is plugged in while the
instrument power is on and the FC88 does not respond, press [ENTER], if there is still no response Press [N] or cycle the power.
Note:The Zero and Span may be changed from the original calibration, provided the new values are within the original calibrated
range. i.e. If the original calibration was 1 to 100 SCFM (4-20mA), the new zero (4mA) must be equal to or greater than 1
SCFM, the new span (20mA) must be equal to or less than 100 SCFM.
Some entries require a Factory pass code. If this occurs contact FCI Field Service to continue programming the instrument. The
instrument will prompt the user when this is necessary. Do not change any parameters that require this code unless there is an absolute
understanding of the instrument's operation. The user can not exit some routines unless all entries are completed or the power is
recycled.
The top level of the menu is shown in Appendix B - Table 5. Enter the large letter in the tables below to activate a command. The user
may exit a command at any time entering “Q” [ENTER] in the menus: D, K, V, W, or Z.
CCalibration Information
Display only: A/D, Delta-R, Ref-R data values
DDiagnostics
Display only: List of unit prameters.
K Factory Calibration Settings
Display only: Cal. parameters, i.e. linearization
and temperature compensation coefficients.
R Factory Reset
Replaces user data with factory calibration data
Table 1. Diagnostics and Factory Settings
Analog out
Select1234
4-20mA out 1FlowFlowTempTemp
4-20mA out 2TempFlowFlowTemp
ST75/ST75V MASS FLOWFLUID COMPONENTS INTERNATIONAL LLC
“V” Menu Output Configuration Set Up
NOTE: The display comes up to the last setting saved and stays for 2 seconds. If N or [ENTER] is entered, the menu
proceeds to the Pulse out. If Y is entered, the display moves to the selection options and/or asks for confirmation. If you
miss the option, select [Enter] repeatedly to loop around.
Pressing [V] [ENTER] will display“Output Mode Selected”followed by:
“4-20mA #1 = Flow”“ 4-20mA #2 = Temp”followed by
“Change? (Y/N)”
Press [ENTER] (no change).
The last saved mode will display at this point. i.e.,
“Source: Pulse”“Sink: Pulse”followed by,
“Change? (Y/N)”Select Y [Enter].The display reads,
“Source: Pulse” “Sink: Pulse”followed by,
“Enter 1 to make the selction #.”Select [ENTER].The next display reads,
“Source: Pulse” “Sink: Alarm”followed by,
“Enter 2 to make the selction #.”Select 2 and [ENTER].The next prompt reads,
“PFactor: 1.000”“Change? (Y/N)>”(this factor can be anywhere from 0.001 to 1000 - A
pulse factor of 1.000 will output 1 pulse per unit of
flow.)
If no change, select N and/or [ENTER] to continue.
The next prompt is, “Sample Period”“Change? (Y/N)>”(this value may be set from 0.5 to 5 seconds)
If no change, select N and/or [ENTER] to continue.
The next prompt is, “Source state: ”“High to Low” Change to “Low to High?>”(this selection toggles the pulse
signal normally high or normally
low).
[ENTER] to read display.
“Switchpt1”“0.0000000”the current set point.
“Change? (Y/N)>” enter Y [ENTER] and enter #______ . Set Point Value , i.e. 50 (value is in same units as the flow and
must be within the calibrated range). [ENTER]. The next prompt is,
“Sink state: ”“High to Low” Change to “Low to High?>”. Set the output signal to be normally “High” or
normally “Low.” Pressing [Y][ENTER] toggles the current setting. Pressing [ENTER] resumes normal operation.
same units as the flow
or temp.
This page is subject to proprietary rights statement on last page
9 Doc. No. 06EN003368 Rev. C
FLUID COMPONENTS INTERNATIONAL LLC ST75/ST75V MASS FLOW
PDA IR Communication Interface
The IR interface software is an optional accessory kit and can ordered using FCI part number 019819-01. The software is compatible
with PALM OS 4.1 or greater. If the software was ordered with the instrument, a CD should be located with the instrument documentation.
The factory has verified the following 3 PDA models. All commands meet their intended purpose and function properly.
1.Palm, Tungsten E, E2:Palm OS 5.2.1, 5.4.7
2.Palm, Zire 71, Palm OS 5.2.1
3.ecom instruments, m 515-EX, Intrinsically-safe. Palm OS 4.1
Procedure:
1.Down load the software into the target PDA. When complete, a yellow and blue FCI icon will be available.
2.Verify JP5 jumper is set in the PDA IR interface position, see Figure H.
3.Select FCI icon on PDA device.
4.The opening menu is displayed, select start.
5.Five menu groups are displayed.
Process:displays current process variables (Flow and Temperature)
ID-Unit:displays model, firmware version, serial no. …
Set-up:allows access to the following areas
UnitsK Factor
Line sizeTemp/Flow min/max
TotalizerOutput Cal
LCDOutput Config
Diagnostics:A/D values
Utilities:allows access to the following areas
Reset
Parameter memory
Calibration coefficients
Factory restore
Process and System Faults
6. After entering into specific menu areas, point the PDA IR port towards the Instrument display. Begin with the PDA device within
5 feet of the instrument display. Select the “Get All” or “Get” button to retrieve information from the instrument. If a value needs
to be changed, the value must first be retrieved.
Example: reading standard process variable information
1.Verify instrument and PDA are functioning.
2.Select FCI icon on the PDA.
3.Select the start button on the opening screen.
4.Select the “Process” button.
5.Point the PDA at the instrument display, start with the PDA no futher that 5 feet from the instrument.
6.Select the “Get Data” button.
7.Flow and temperature Data will begin streaming to the PDA.
8.If the IR link is interrupted, a “Command response timed out” message will be displayed.
9.Repeat the process if the link is interrupted.
This page is subject to proprietary rights statement on last page
10Doc. No. 06EN003368 Rev. C
ST75/ST75V MASS FLOWFLUID COMPONENTS INTERNATIONAL LLC
Maintenance
The FCI instrument requires little maintenance. There are no moving parts or mechanical parts subject to wear in the instrument. The
sensor assembly which is exposed to the process media is composed of 316 SS and Hastelloy C.
Without detailed knowledge of the environmental parameters of the application surroundings and process media, FCI cannot make
specific recommendations for periodic inspection, cleaning, or testing procedures. However, some suggested general guidelines for
maintenance steps are offered below. Use operating experience to establish the frequency of each type of maintenance.
Calibration
Periodically verify the calibration of the output and recalibrate if necessary. FCI recommends every 18 months at a minimum.
Electrical Connections
Periodically inspect cable connections on terminal strips and terminal blocks. Verify that terminal connections are tight and physically
sound with no sign of corrosion.
Remote Enclosure
Verify that the moisture barriers and seals protecting the electronics in the local enclosure is adequate and that no moisture is entering
the enclosure.
Electrical Wiring
FCI recommends occasional inspection of the system’s interconnecting cable, power wiring and flow element wiring on a “common
sense” basis related to the application environment. Periodically the conductors should be inspected for corrosion and the cable
insulation checked for signs of deterioration.
Flow Element Connections
Verify that all seals are performing properly and that there is no leakage of the process media. Check for deterioration of the gaskets
and environmental seals used.
Insertion Type Flow Element Assembly
Periodically remove the flow element for inspection based on historical evidence of debris, foreign matter, or scale build-up and
appropriate plant shutdown schedules and procedures. Check for corrosion, stress cracking, and/or build-up of oxides, salts, or foreign
substances. The thermowells must be free of excessive contaminants and be physically intact. Any debris or residue build-up could
cause inaccurate flow indication. Clean the flow element, as necessary, with a soft brush and available solvents (compatible with
Stainless Steel).
This page is subject to proprietary rights statement on last page
11Doc. No. 06EN003368 Rev. C
FLUID COMPONENTS INTERNATIONAL LLC ST75/ST75V MASS FLOW
Troubleshooting
Application Verification
After verifying that the flow meter is functioning, review the application parameters as shown below to verify the calibration matches
the process media.
Equipment Needed
Flow Instrument Calibration Data
Process Parameters and Limits
Check Serial Numbers
Verify that the serial number of the flow element and the flow transmitter electronics are the same. The flow element and the flow
transmitter are a matched set and cannot be operated independently of each other.
Check the Instrument Installation
Verify correct mechanical and electrical installation. Verify the flow element is mounted at least 20 diameters downstream and 10
diameters upstream from any bends or interference in the process pipe or duct.
Check for Moisture
Check for moisture on the flow transmitter. Moisture may cause intermittent operation. Check for moisture on the flow element. If a
component of the process media is near its saturation temperature it may condense on the flow element. Place the flow element where
the process media is well above the saturation temperature of any of the process gases.
Check Application Design Requirements
Application design problems may occur with first time application instruments, although the design should also be checked on
instruments that have been in operation for some time. If the application design does not match field conditions, errors occur.
1.Review the application design with plant operation personnel and plant engineers.
2.Ensure that plant equipment such as pressure and temperature instruments conform to the actual conditions.
3.Verify operating temperature, operating pressure, line size, and gas medium.
Verify Standard Versus Actual Process Conditions
The flowmeter measures the mass flow rate. The mass flow rate is the mass of the gas flowing through a pipe per time. Other flow
meters, such as an orifice plate or a pitot tube, measure the volumetric flow rate. The volumetric flow rate is the volume of gas per
time. If the readings displayed do not agree with another instrument, some calculations may be necessary before comparing them. To
calculate the mass flow rate, the volumetric flow rate, and the pressure and temperature, the point of measurement must be known. Use
the following equation to calculate the mass flow rate (Standard Volumetric Flow rate) for the other instrument:
Equation:
QQ
=× ×
SA
Where:used for pressure and temperature.)
QA= Volumetric FlowQS= Standard Volumetric Flow
PA= Actual PressureTA= Actual Temperature
PS= Standard PressureTS=
PSIA and °R are used for pressure and temperature units.
This page is subject to proprietary rights statement on last page
1
)
(
19.7 PSIA
580° R
12Doc. No. 06EN003368 Rev. C
)
(
530° R_
14.7 PSIA
)
=1485 SCFM
ST75/ST75V MASS FLOWFLUID COMPONENTS INTERNATIONAL LLC
Calibration Parameters Verification
The instrument uses a set of predetermined calibration parameters to process flow signals. Most of these parameters should not
change. A data package located with this manual contains the “ST75/ST75V Delta R Data Sheet”. This contains the calibration
parameters stored in the flow transmitter at the factory. To verify that these parameters have not changed, complete the following:
1.Identify the appropriate Delta R Data sheets by serial number of the instrument.
2.Press [D] [ENTER] to examine each of the parameters. The [ENTER] key allows scrolling one message at a time. Use Table
4 to verify parameters with the Delta R Data sheet ST75/ST75V Parameters.
S/W Version:dR Min:T SpanIDAC 0:
Flow Factor:dR Max:T ZeroIDAC 0:
Cmin Flow:Cal Ref:T SpanIDAC 1:
Cmax Flow:Tcslp:T ZeroIDAC 1:
Eng Units:Tcslp 0:State 0:
Line Size 0:Tcslp 2:Switch Pt 0:
Line Size 1:Tot Menu:State 1:
Cmin Temp:Tot Flag:Switch Pt 1:
Cmax Temp:Totalizer:K factor 1:
Min Flow:Rollover Cnt:K factor 2:
Max Flow:Fix Pt Flag:K factor 3:
Density:Pulse Factor:K factor 4:
*C1 [1]:Pulse Ou t:I factor:
*C1 [2]:Hours:Temp Flag:
*C1 [3]:Sample Period:Out Mode:
*C1 [4]:dR Slope:Boxcar Max:
*C1 [5]:dr Off Set:RTD-SLP-385:
Break Pt:Refr Slope:% of Range:
*C2 [1]:Re fr Off Set:Us er Name:
*C2 [2]:SpanIDAC 0:Shop Order #:
*C2 [3]:ZeroIDA C 0:Serial No .:
*C2 [4]:SpanIDAC 1:Model#:
*C2 [5]:Z ero IDAC 1:
Table 4. Diagnostic Test Sequence on Display
If parameters that have an asterisk (*) have changed, this may indicate a problem. Customer Service should be contacted. If the
parameters have not changed, continue with the next section.
This page is subject to proprietary rights statement on last page
13Doc. No. 06EN003368 Rev. C
FLUID COMPONENTS INTERNATIONAL LLC ST75/ST75V MASS FLOW
Hardware Verification
Equipment Required:
Digital Multimeter
Screw Driver
The ST75/ST75V Flowmeter is comprised of 4 basic components:
1.Sensor element.
2.Customer interface circuit board
3.Control circuit assembly circuit board module.
4.Electronics enclosure.
Step 1
Verify fuse (F1) located on the customer interface circuit board is in normal working condition.
Remove power from the instrument. Open the electronics enclosure exposing the customer interface circuit board. This circuit board
is located under the shorter enclosure lid along with all of the power and input/output connections. Unscrew the clear cover on the
fuse and pull the fuse out of the fuse holder. Check the fuse for continuity. If fuse reads open, replace with equivalent component
(FCI part no. 019933-01), Wickmann Inc. series 374, amp code 1160, package 41.
Fuse (F1)
Ac power customer interface circuit board shown. Fuse (F1) on DC power customer interface circuit board located in similar
position.
Step 2
Verify interconnecting cable from the customer interface board and the control circuit board assembly module are correctly seated
into the appropriate header.
Remove power from the instrument. Open the electronics enclosure exposing the customer interface circuit board. This circuit board
is located under the shorter enclosure lid along with all of the power and input/output connections. Remove the 2 screws securing the
interface circuit board to the electronics enclosure. Carefully lift the interface face board exposing the interconnecting cable between
the interface board and the control circuit assembly. Verify cable is seated firmly at both ends of the cable header.
Interface Board
This page is subject to proprietary rights statement on last page
Control Circuit
Assembly
Cable Header
14Doc. No. 06EN003368 Rev. C
ST75/ST75V MASS FLOWFLUID COMPONENTS INTERNATIONAL LLC
Step 3
Verify sensor element continuity and resistance.
Remove sensor element cable from the bottom of the control circuit assembly. Note that 2 of the wires have a red stripe and are
located closest to the interconnecting cable header. Using an ohm meter verify that resistance between the 2 red striped wires is
approximately 1100 ohms +/- 20. This resistance is temperature dependant. The resistance at 70 degrees F should be 1082 ohms.
Verify the resistance between the 2 natural colored wires are approximately the same.
Sensor Element Cable
FCI provides full in-house technical support. Additional technical representation is also provided by FCI field representatives. Before
contacting a field or in-house representative, please perform the troubleshooting techniques outlined in this document. If problems
persist, contact the FCI Customer Service department at 1-800-854-1993 or 1-760-744-6950.
If the instrument is to be returned to FCI, please obtain an Return Authorization. The form contains a declaration of decontamination
cleaning information that the instrument must comply with before it is shipped to FCI.
This page is subject to proprietary rights statement on last page
15Doc. No. 06EN003368 Rev. C
FLUID COMPONENTS INTERNATIONAL LLC ST75/ST75V MASS FLOW
Transmitter Circuit Calibration Check (Delta R Verification)
Small flat Blade Screwdriver, 3/32 inches wide blade
FCI Normalization Cable, FCI part number 006407
Procedure
1.Verify all “D” mode calibration parameters are correct according to the meters Delta R Data Sheet, before starting verification.
2.Turn power off
3.Mark all sensor element wires connected to the circuit board, so they may be reconnected to the proper terminals. Disconnect the
wires.
4.Connect the resistance decade box to the electronics as per the appropriate diagram for the ST75/ST75V.
NOTE:Interconnector wiring (resistance decade box to electronics) must be 24 AWG and 45 inches long, to avoid any
inaccuracies in the Delta R verification, caused by improper wire lengths or wire gauges.
5.Set both decade boxes for the nominal resistance value (1000 ohms) +/- .01%
6.Connect DVM to the meters output termination and monitor the meter output.
7.Turn the power ON and allow the instrument 5 mins. To stabilize
8.With the FC88 connected, Press [T] [Enter] to view the Normal Operating Mode.
9.Adjust the Active Decade Box (Reference decade box remains fixed @ 1000 ohms) to achieve the appropriated Delta R for the
displayed flow value and output, noted on the meters Delta R Data Sheet.
10. Note the [C] mode and verify the meters displayed TCDR and REFR values corresponding to the displayed flow rate as per the
meters Delta R Data Sheet.
11. Return to the [T] mode to continue the verification.
ACTIVE
1433-W DECADE RISISTOR GENERAL RADIO USA
ORANGE
ORANGE
BLACK
BLACK
NORMALIZATION CABLE
FCI P/N 006407
10K
H
0
1
2
3
L
4
5
X
6
9
7
8
G
1K STEPS
23mA MAX
REFERENCE
1433-W DECADE RISISTOR GENERAL RADIO USA
10K
H
0
1
2
3
L
4
5
X
6
9
7
8
G
1K STEPS
23mA MAX
0
1
2
3
4
5
6
100 STEPS
80mA MAX
1
2
3
4
5
6
100 STEPS
80mA MAX
1
2
3
4
5
X
6
9
7
8
0
1
2
3
4
5
X
6
9
7
8
This page is subject to proprietary rights statement on last page
16Doc. No. 06EN003368 Rev. C
ST75/ST75V MASS FLOWFLUID COMPONENTS INTERNATIONAL LLC
Appendix A - Approval Information
EC
Information
This page is subject to proprietary rights statement on last page
17Doc. No. 06EN003368 Rev. C
FLUID COMPONENTS INTERNATIONAL LLC ST75/ST75V MASS FLOW
Safety Instructions for the use the ST51/75 flowmeter in Hazardous Areas
Approval KEMA 08ATEX0045/IEC KEM08.0012 for:
Category II 2 G for Gas protection Ex d IIC T6…T3
Category II 2 D for Dust protection Ex tD A21 IP67 T90ºC…T121ºC
The ST51/75 series consist of a sensing element and associated integral or remote mounted electronics mounted
in a type “d” flameproof enclosure.
Relation between ambient temperature, process temperature and temperature class is as follows:
Ambient temperature range (Ta): T6 [85ºC] for : -40ºC <Ta< + 65ºC
Process temperature range (Tp): T6 [85ºC] for : -40ºC <Tp< + 65ºC
T5 [100ºC] for : -40ºC <Tp< + 100ºC
T4 [135ºC] for : -40ºC <Tp< + 135ºC
T3 [200ºC] for : -40ºC <Tp< + 200ºC
Electrical data: Power supply: 85 to 265 VAC, 50/60 Hz, 12 Watt max; 24 VDC, 12 VA Max
Dansk Sikkerhedsforskrifter Italiano Normative di sicurezza
Deutsch Sicherheitshinweise Nederlands Veiligheidsinstructies
English Safety instructions Português Normas de segurança
ΥπΥπ_δεί_ειςασφαλείας
Español Instrucciones de seguridad
Suomi Turvallisuusohjeet Svenska Säkerhetsanvisningar
Français Consignes de sécurité
DK Dansk- Sikkerhedsforskrifter
Disse sikkerhedsforskrifter gælder for Fluid Components, ST51/75 EF-typeafprøvningsattest-nr. KEMA 08ATEX0045/IEC KEM08.0012
(attestens nummer på typeskiltet) er egnet til at blive benyttet i eksplosiv atmosfære kategori II 2 GD.
1) Ex-anlæg skal principielt opstilles af specialiseret personale.
2) ST51/75 skal jordforbindes.
3) Klemmerne og elektronikken er monteret i et hus, som er beskyttet af en eksplosionssikker kapsling med følgende noter:
• Gevindspalten mellem huset og låget er på en sådan måde, at ild ikke kan brede sig inden i det.
• Ex-„d“ tilslutningshuset er forsynet med et 1/2" NPT og/eller M20x1.5 gevind for montering af en Ex-„d“ kabelindføring, der er attesteret iht.
IEC/EN 60079-1
• Det er vigtigt at sørge for, at forsyningsledningen er uden spænding eller eksplosiv
atmosfære ikke er til stede, før låget åbnes og når låget er åbent på „d” huset (f.eks. ved tilslutning eller servicearbejde).
• Låget på „d” huset skal være skruet helt ind, når apparatet er i brug. Det skal sikres ved at dreje en af låseskruerne på låget ud.
D A Deutsch-Sicherheitshinweise
Diese Sicherheitshinweise gelten für die Fluid Components, ST51/75 flowmeter gemäß der EG-Baumusterprüfbescheinigung Nr. KEMA
08ATEX0045/IEC KEM08.0012 (Bescheinigungsnummer auf dem Typschild) der Kategorie II 2 GD.
1) Die Errichtung von Ex-Anlagen muss grundsätzlich durch Fachpersonal vorgenommen werden.
2) Der ST51/75 muß geerdet werden.
3) Die Klemmen und Elektroniken sind in einem Gehäuse in der Zündschutzart druckfeste Kapselung („d“ ) eingebaut.
• Der Gewindespalt zwischen dem Gehäuse und dem Deckel ist ein zünddurchschlagsicherer Spalt.
• Das Ex-“d“ Anschlussgehäuse besitzt ein 1/2“ NPT und/oder M20x1.5 Gewinde für den Einbau einer nach IEC/EN 60079-1 bescheinigten
Ex-“d“ Kabeleinführung.
• Es ist sicherzustellen, dass vor dem Öffnen und bei geöffnetem Deckel des „d“ Gehäuses (z.B. bei Anschluss oder Service- Arbeiten)
entweder die Versorgungsleitung spannungsfrei oder keine explosionsfähige Atmosphäre vorhanden ist.
• Der Deckel des „d“ Gehäuses muss im Betrieb bis zum Anschlag hineingedreht sein. Er ist durch eine der Deckelarretierungsschrauben zu
sichern.
This page is subject to proprietary rights statement on last page
18Doc. No. 06EN003368 Rev. C
ST75/ST75V MASS FLOWFLUID COMPONENTS INTERNATIONAL LLC
GB IRL English- Safety instructions
These safety instructions are valid for the Fluid Components, ST51/75 flowmeter to the EC type approval certificate no KEMA 08ATEX0045/IEC
KEM08.0012 (certificate number on the type label) for use in potentially explosive atmospheres in Category II 2 GD.
1) The installation of Ex-instruments must be made by trained personnel.
2) The ST51/75 must be grounded.
3) The terminals and electronics are installed in a flame proof and pressure-tight housing with following notes:
• The gap between the housing and cover is an ignition-proof gap.
• The Ex-“d“ housing connection has a 1/2” NPT and/or M20x1.5 cable entry for mounting an Ex-d cable entry certified acc. to IEC/EN 60079-1.
• Make sure that before opening the cover of the Ex”d“ housing, the power supply is disconnected or there is no explosive atmosphere present (e.g.
during connection or service work).
• During normal operation: The cover of the “d“ housing must be screwed in completely and locked by tightening one of the cover locking screws.
GR Υπ_δεί_εις ασφαλείας
Αυτές οι οδηγίες ασφαλείας ισχύουν για τα Ροόµετρα της Fluid Components τύπου ST51/75 που φέρουν Πιστοποιητικό Εγκρίσεως Ευρωπαϊκής
Ένωσης, µε αριθµό πιστοποίησης KEMA 08ATEX0045/IEC KEM08.0012 (ο αριθµός πιστοποίησης βρίσκεται πάνω στην ετικέτα τύπου του
οργάνου) για χρήση σε εκρηκτικές ατµόσφαιρες της κατηγορίας II 2 GD.
1) Η εγκατάσταση των οργάνων µε αντιεκρηκτική προστασία πρέπει να γίνει από εξειδικευµένο προσωπικό.
• Το “Ex-d” αντιεκρηκτικό περίβληµα, έχει ανοίγµατα εισόδου καλωδίου µε διάµετρο ½” NPT ή/και M20x1.5, κατάλληλα για τοποθέτηση υποδοχής αντιεκρηκτικούκαλωδίουπιστοποιηµένηςκατά IEC/EN 60079-1
• Βεβαιωθείτεότιπριντοάνοιγµακαλύµµατος του του “Ex-d” αντιεκρηκτικού περιβλήµατος, η τάση τροφοδοσίας είναι
αποσυνδεδεµένη ή ότι δεν υφίσταται στη περιοχή εκρηκτική ατµόσφαιρα (π.χ. κατά τη διάρκεια της σύνδεσης ή εργασιών
συντήρησης)
• Κατά τη διάρκεια οµαλής λειτουργίας: Το κάλυµα του “d” καλύµµατος αντιεκρηκτικού περιβλήµατος πρέπει να είναι εντελώς βιδωµένο και
ασφαλισµένο, σφίγγοντας µία από τις βίδες ασφαλείας του περιβλήµατος.
FIN Suomi - Turvallisuusohjeet
Nämä turvallisuusohjeet koskevat Flud Components, ST51/75 EY-tyyppitarkastustodistuksen nro. KEMA 08ATEX0045/IEC KEM08.0012
(todistuksen numero näkyy tyyppikilvestä) käytettäessä räjähdysvaarallisissa tiloissa luokassa II 2GD.
1) Ex-laitteet on aina asennettava ammattihenkilökunnan toimesta.
2) ST51/75 on maadoitettava.
3) Syöttöjännitteen kytkemisessä tarvittavat liittimet ja elektroniikka on asennettu koteloon jonka rakenne kestää räjähdyspaineen seuraavin
lisäyksin :
• Kotelon ja kannen välissä on räjähdyksen purkausväli.
• Ex-d liitäntäkotelossa on 1/2“ NPT ja/tai M20x1.5 kierre IEC/EN 60079-1 mukaisen Ex-d kaapeliläpiviennin asennusta varten
• Kun “d“-kotelon kansi avataan (esim. liitännän tai huollon yhteydessä), on varmistettava, että joko syöttöjohto on jännitteetön tai ympäristössä ei
ole räjähtäviä aineita.
• “d“ -kotelon kansi on kierrettävä aivan kiinni käytön yhteydessä ja on varmistettava kiertämällä yksi kannen lukitusruuveista kiinni.
This page is subject to proprietary rights statement on last page
19Doc. No. 06EN003368 Rev. C
FLUID COMPONENTS INTERNATIONAL LLC ST75/ST75V MASS FLOW
F B L Consignes de sécurité
Ces consignes de sécurité sont valables pour le modèle ST51/75 de la société Fluid Components (FCI) conforme au certificat d’épreuves de type
KEMA 08ATEX0045/IEC KEM08.0012 (numéro du certificat sur l’étiquette signalétique) conçu pour les applications dans lesquelles un matériel
de la catégorie II2GD est nécessaire.
1) Seul un personnel spécialisé et qualifié est autorisé à installer le matériel Ex.
2) Les ST51/75 doivent être reliés à la terre.
3) Les bornes pour le branchement de la tension d’alimentation et l’électronique sont logées dans un boîtier à enveloppe antidéflagrante
avec les notes suivantes :
• Le volume entre le boîtier et le couvercle est protégé en cas d’amorçage.
• Le boîtier de raccordement Ex-d dispose d’un filetage1/2’’ NPT et/ou M20x1.5 pour le montage d’un presse-étoupe Ex-d certifié selon la IEC/EN
60079-1.
• Avant d’ouvrir le couvercle du boîtier « d » et pendant toute la durée où il le restera (pour des travaux de raccordement, d’entretien ou de
dépannage par exemple), il faut veiller à ce que la ligne d’alimentation soit hors tension ou à ce qu’il n’y ait pas d’atmosphère explosive.
• Pendant le fonctionnement de l’appareil, le couvercle du boîtier « d » doit être vissé et serré jusqu’en butée. La bonne fixation du couvercle doit
être assurée en serrant une des vis d’arrêt du couvercle.
I Italiano - Normative di sicurezza
Queste normative di sicurezza si riferiscono ai Fluid Components, ST51/75 secondo il certificato CE di prova di omologazione n° KEMA
08ATEX0045/IEC KEM08.0012 (numero del certificato sulla targhetta d’identificazione) sono idonei all’impiego in atmosfere esplosive applicazioni
che richiedono apparecchiature elettriche della Categoria II 2 GD.
1) L’installazione di sistemi Ex deve essere eseguita esclusivamente da personale specializzato.
2) I ST51/75 devono essere collegati a terra.
3) I morsetti per il collegamento e l’elettronica sono incorporati in una custodia a prova di esplosione („d“) con le seguenti note:
• La sicurezza si ottiene grazie ai cosidetti „interstizi sperimentali massimi“, attraverso i quali
una eventuale accensione all’interno della custodia non può propagarsi all’esterno oraggiungere altre parti dell’impianto.
• La scatola di collegamento Ex-d ha una filettatura 3/4“ e/o 1“ NPT per il montaggio di un passacavo omologato Ex-d secondo IEC/EN 60079-1.
• Prima di aprire il coperchio della custodia „d“ (per es. durante operazioni di collegamento o di manutenzione) accertarsi che l’apparecchio sia
disinserito o che non si trovi in presenza di atmosfere esplosive.
• Avvitare il coperchio della custodia „d“ fino all’arresto. Per impedire lo svitamento del coperchio é possibile allentare una delle 2 viti esagonali
poste sul corpo della custodia, incastrandola nella sagoma del coperchio.
NL B Nederlands - Veiligheidsinstructies
Deze veiligheidsinstructies gelden voor de Fluid Components, ST51/75 overeenkomstig de EG-typeverklaring nr. KEMA 08ATEX0045/IEC
KEM08.0012 (nummer van de verklaring op het typeplaatje) voor gebruik in een explosieve atmosfeer volgens Categorie II 2GD.
1) Installatie van Ex-instrumenten dient altijd te geschieden door geschoold personeel.
2) De ST51/75 moet geaard worden.
3) De aansluitklemmen en de electronika zijn ingebouwd in een drukvaste behuizing met de volgende opmerkingen:
• De schroefdraadspleet tussen de behuizing en de deksel is een ontstekingsdoorslagveilige
spleet.
• De Ex-d aansluitbehuizing heeft een 1/2” of een M20x1.5 schroefdraad voor aansluiting van een volgens IEC/EN 60079-1
goedgekeurde Ex- ‘d’ kabelinvoer.
• Er moet worden veilig gesteld dat vóór het openen bij een geopende deksel van de ‘d’ behuizing (bijv. bij aansluit- of servicewerkzaamheden)
hetzij de voedingsleiding spanningsvrij is, hetzij geen explosieve atmosfeer aanwezig is.
• De deksel van de ‘d’ behuizing moet tijdens bedrijf tot aan de aanslag erin geschroefd zijn.
Hij moet door het eruit draaien van een van de dekselborgschroeven worden geborgd.
This page is subject to proprietary rights statement on last page
20Doc. No. 06EN003368 Rev. C
ST75/ST75V MASS FLOWFLUID COMPONENTS INTERNATIONAL LLC
P Português - Normas de segurança
Estas normas de segurança são válidas para os Fluid Components, ST51/75 conforme o certificado de teste de modelo N.º KEMA
08ATEX0045/IEC KEM08.0012 (número do certificado na plaqueta com os dados do equipamento)
são apropriados para utilização em atmosferas explosivas categoria II 2 GD.
1) A instalação de equipamentos em zonas sujeitas a explosão deve, por princípio, ser executada por técnicos qualificados.
2) Os ST51/75 Flexmasster precisam ser ligados à terra.
3) Os terminais e a electrónica para a conexão da tensão de alimentação estão instalados num envólucro com protecção contra ignição á prova de
sobrepressão com as seguintes notas :
• A fenda entre o envólucro e a tampa deve ser á prova de passagem de centelha.
• O envólucro de conexão Ex-“d“ possui uma rosca 1/2“ NPT e/ou M20x1.5 para a entrada de cabos Ex-“d“ certificado conforme a norma IEC/EN
60079-1.
• Deve-se assegurar que, antes de abrir a tampa do armário „d“ ( por exemplo, ao efectuar a conexão ou durante trabalhos de manutenção), o cabo
de alimentação esteja sem tensão ou que a atmosfera não seja explosíva.
• Durante a operação, a tampa do envólucro „d“ deve estar aparafusada até o encosto. A tampa deve ser bloqueada, por um dos parafusos de
fixação.
E Español - Instrucciones de seguridad
Estas indicaciones de seguridad son de aplicación para el modelo ST51/75 de Fluid Components, según la certificación CE de modelo Nº KEMA
08ATEX0045/IEC KEM08.0012 para aplicaciones en atmósferas potencialmente explosivas según la categoría II 2 GD (el número decertificación
se indica sobre la placa informativa del equipo).
1) La instalación de equipos Ex tiene que ser realizada por personal especializado.
2) Los ST51/75 tienen que ser conectados a tierra.
3) Los bornes de conexión y la unidad electrónica están montados dentro de una caja con protección antideflagrante y resistente a presión,
considerándose los siguientes puntos:
• La holgura entre la rosca de la tapa y la propia de la caja está diseñada a prueba contra ignición.
• La caja tiene conexiones eléctricas para entrada de cables con rosca 1/2” NPTy/o M20x1.5, donde deberán conectarse prensaestopas
certificados Exd según IEC/EN60079-1.
• Antes de la apertura de la tapa de la caja "Exd“ (p. ej. durante los trabajos de conexionado o de puesta en marcha) hay que asegurar que el
equipo se halle sin tensión o que no exista presencia de atmósfera explosiva.
• Durante el funcionamiento normal: la tapa de la caja antideflagrante tiene que estar cerrada, roscada hasta el tope, debiendose asegurar
apretando los tornillos de bloqueo.
S Svenska - Säkerhetsanvisningar
Säkerhetsanvisningarna gäller för Fluid Components, Flödesmätare typ ST51/75 enligt EG-typkontrollintyg nr KEMA 08ATEX0045/IEC
KEM08.0012 (intygsnumret återfinns på typskylten) är lämpad för användning i explosiv gasblandning i kategori II 2 GD.
1) Installation av Ex- klassade instrument måste alltid utföras av fackpersonal.
2) ST51/75 måste jordas.
3) Anslutningsklämmorna och elektroniken är inbyggda i en explosions och trycktät kapsling med följande kommentar:
• Spalten mellan kapslingen och lockets gänga är flamsäker.
• Ex-d kapslingen har en 1/2“ NPT och / eller M20x1.5 gänga för montering av en IEC/EN 60079-1 typkontrollerad
Ex- „d“ kabel förskruvning
• När Ex- „d“-kapslingens lock är öppet (t.ex. vid inkoppling - eller servicearbeten) ska man se till att enheten är spänningslös eller att ingen explosiv
gasblandning förekommer.
• Under drift måste Ex - d“-kapslingens lock vara iskruvad till anslaget. För att säkra locket skruvar
man i en av lockets insex låsskruvar .
This page is subject to proprietary rights statement on last page
21Doc. No. 06EN003368 Rev. C
FLUID COMPONENTS INTERNATIONAL LLC ST75/ST75V MASS FLOW
INTENTIONALLY LEFT BLANK
This page is subject to proprietary rights statement on last page
22Doc. No. 06EN003368 Rev. C
ST75/ST75V MASS FLOWFLUID COMPONENTS INTERNATIONAL LLC
Appendix B - List Commands
COMMAND
MNEMONIC
A R AvgDelta_r, AvgRef
B R Delta_r, Ref_r
C R Tcdelta_r, Ref_r
D R Diagnostics
F R/W Kfactors
G R/W
K R/W Cal Parameters
L R/W Output Cal
N W Warm Restart
R W Factory Restore
S R/W Totalizer Menu On/Off
T R Normal Mode
V R/W Output Config
W R/W Totalizer
Y W Command Line Interface
Z W
COMMAND
FUNCTION COMMAND DECRIPTION
Clear FlashEE, Boxcar Count,
ADC to Ohms Cal
Flow units, Pipe Size, and LCD
Scaling
Table 5. ST75/ST75V List of Single Letter Commands
COMMAND COMMAND
MNEMONICFUNCTION
BKR/WBreak Po in tFlo at
BMR/WBoxcar Filter MaxInteger
CMR/WCminflo wFlo at
CRR/WCalibration RefFloat
CXR/WCmaxflowFloa t
C1[1-5]R/WCoefficient s set1Float
C2[1-5]R/WCoefficient s set2Float
DIRDiagnosticsNull
DMR/WDeltaR MinimumFloat
DNR/WDensityFloat
DRRDelta RFloat
DXR/WDeltaR MaximumFloat
DSR/WDeltaR SlopeFloat
DFR/WDeltaR OffsetFloat
EUR/WEngineering UnitsInteger
FFR/WFlow FactorFloat
FPR/WFix Point FlagInteger
F0R/WPulse Out State0Integer
F1R/WPulse Out State1Integer
HRR/WTot Dump Hours Cnt rInt eger
IFR/WI FactorFloat
COMMAND DESCRIPTIONDATA TYPE
This page is subject to proprietary rights statement on last page
Table 6. ST75/ST75V List of CLI Commands
23Doc. No. 06EN003368 Rev. C
FLUID COMPONENTS INTERNATIONAL LLC ST75/ST75V MASS FLOW
Table 6. ST75/ST75V List of CLI Commands, Cont.
COMMAND COMMAND
MNEMONICFUNC TION
K[1-4]R/WK FactorsFloat
L0R/WLin e S ize 0Flo at
L1R/WLin e S ize 1Flo at
MNR/WMinflowFloat
MXR/WMaxflowFloat
OMR/WOutmodeInteger
PFR/WPulse FactorFloat
PLR/WPulse OutInteger
PSR/WPulse Sample PeriodFloat
PWR/WPulse W idthFloat
P0R/WSwitch Po int0Integer
P1R/WSwitch Po int1Integer
ROR/WRollOver CntrLong
RRRReference RFloat
RSR/WRefR SlopeFloat
RFR/WRefR OffsetFloa t
SFRSFPS FlowFloat
SNR/WSerial NumberString (16 chars max.)
SOR/WShop Order NumberString (16 chars max.)
S0R/WSpanDAC0 for 4-20mA #1Integer
S3R/WSpanDAC1 for 4-20mA #2Integer
S2WSave FACTORYN/A
TCRTCdeltarFloat
TDR/WTcs lpFloat
TFR/WTotalizer OFF/ON FlagInteger
TMR/WCmintempFloat
TPR/WTotalizer Temperature FlagInteger
TTR/WTotalizer ValueFloat
TXR/WCmaxtempFloat
TZRTemperatureFloat
T0R/ WT c s lp0Flo at
T2R/ WT c s lp2Flo at
T3R/WTSp an DA C0 for 4-20mA # 1Integer
T7R/WTSp an DA C1 for 4-20mA # 2Integer
T5R/WTZero DAC0 fo r 4-20mA #1In teger
T8R/WTZero DAC1 fo r 4-20mA #2In teger
UFRUser FlowFloat
UKRUs er FlowKFloat
UNR/WUser Name String (16 chars max.)
VNRVersion NumberString (16 chars max.)
XXR/WTes t Flow Rate (SFPS)Float
XYWDelete Test Flow RateFloat
Z0R/WZeroDA C0 for 4-20mA # 1Integ er
Z2R/WZeroDA C1 for 4-20mA # 2Integ er
Command Line Password: 357
COMMAND DES CRIPTIONDATA TYPE
This page is subject to proprietary rights statement on last page
24Doc. No. 06EN003368 Rev. C
ST75/ST75V MASS FLOWFLUID COMPONENTS INTERNATIONAL LLC
NOTE: When invoking a Write Function, there must be a space separating the Command characters and the data value. All Read
and Write Functions are completed with a <CR>. To exit CLI, press <CR> following the last Command <CR>.
This page is subject to proprietary rights statement on last page
25Doc. No. 06EN003368 Rev. C
FLUID COMPONENTS INTERNATIONAL LLC ST75/ST75V MASS FLOW
INTENTIONALLY LEFT BLANK
This page is subject to proprietary rights statement on last page
26Doc. No. 06EN003368 Rev. C
ST75/ST75V MASS FLOWFLUID COMPONENTS INTERNATIONAL LLC
Appendix C - Drawings
FLUID COMPONENTS
INTERNATIONAL LLC
0.000 SFPS
32.2 DE GREES F
FLUID COMPONENTS
INTERNATIONAL LLC
0.000 SFPS
32.2 DEGREES F
This page is subject to proprietary rights statement on last page
27Doc. No. 06EN003368 Rev. C
FLUID COMPONENTS INTERNATIONAL LLC ST75/ST75V MASS FLOW
FLOW
FLOW
FLUID COMPONENTS
INTERNATIONAL LLC
This page is subject to proprietary rights statement on last page
FLUID COMPONENTS
INTERNATIONAL LLC
28Doc. No. 06EN003368 Rev. C
ST75/ST75V MASS FLOWFLUID COMPONENTS INTERNATIONAL LLC
FLOW
FLOW
FLUID COMPONENTS
INTERNATIONAL LLC
This page is subject to proprietary rights statement on last page
FLUID COMPONENTS
INTERNATIONAL LLC
29Doc. No. 06EN003368 Rev. C
FLUID COMPONENTS INTERNATIONAL LLC ST75/ST75V MASS FLOW
FLOW
FLOW
FLUID COMPON ENTS
INTERNATIONAL LLC
This page is subject to proprietary rights statement on last page
FLUID COMPONENTS
INTERNATIONAL LLC
30Doc. No. 06EN003368 Rev. C
ST75/ST75V MASS FLOWFLUID COMPONENTS INTERNATIONAL LLC
FLUID COMPONENTS
INTERNATIONAL LLC
This page is subject to proprietary rights statement on last page
FLUID COMPONENTS
INTERNATIONAL LLC
31Doc. No. 06EN003368 Rev. C
FLUID COMPONENTS INTERNATIONAL LLC ST75/ST75V MASS FLOW
O
O
O
O
O
O
FLUID COMPONENTS
INTERNATIONAL LLC
This page is subject to proprietary rights statement on last page
FLUID COMPONENTS
INTERNATIONAL LLC
32Doc. No. 06EN003368 Rev. C
ST75/ST75V MASS FLOWFLUID COMPONENTS INTERNATIONAL LLC
Appendix D - Customer Service
Customer
Service/
Technical
Support
FCI provides full in-house technical support. Additional technical representation is also provided by
FCI field representatives. Before contacting a field or in-house representative, please perform the
troubleshooting techniques outlined in this document.
By Mail
Fluid Components International LLC
1755 La Costa Meadows Dr.
San Marcos, CA 92078-5115 USA
Attn: Customer Service Department
By Phone
Contact the area FCI regional representative. If a field representative is unable to be contacted or if a
situation is unable to be resolved, contact the FCI Customer Service Department toll free at
1 (800) 854-1993.
By Fax
To describe problems in a graphical or pictorial manner, send a fax including a phone
or fax number to the regional representative. Again, FCI is available by facsimile if all possibilities
have been exhausted with the authorized factory representative. Our Fax number is 1 (760) 736-6250;
it is available 7 days a week, 24 hours a day.
By E-Mail
FCI Customer Service can be contacted by e-mail at: techsupport@fluidcomponents.com.
Describe the problem in detail making sure a telephone number and best time to be contacted is stated
in the e-mail.
International Support
For product information or product support outside the contiguous United States, Alaska, or Hawaii,
contact your country’s FCI International Representative or the one nearest to you.
After Hours Support
For product information visit FCI's Worldwide Web at www.fluidcomponents.com. For product
support call 1 (800) 854-1993 and follow the prerecorded instructions.
Point of Contact
The point of contact for service, or return of equipment to FCI is your authorized FCI sales/service
office. To locate the office nearest you, please go to www.fluidcomponents.com.
This page is subject to proprietary rights statement on last page
33Doc. No. 06EN003368 Rev. C
FLUID COMPONENTS INTERNATIONAL LLC ST75/ST75V MASS FLOW
Warranty Repairs or Returns
FCI prepays ground transportation charges for return of freight to the customer’s door. FCI
reserves the right to return equipment by the carrier of our choice.
International freight, handling charges, duty/entry fees for return of equipment are paid by the
customer.
Non-Warranty Repairs or Returns
FCI returns repaired equipment to the customer either collect or prepaid and adds freight charges to
the customer invoice.
Return to Stock Equipment
The customer is responsible for all shipping and freight charges for equipment that is returned to
FCI stock from the customer site. These items will not be credited to customer’s account until
either all freight charges are cleared or until the customer agrees to have any freight costs incurred
by FCI deducted, along with applicable return to stock charges, from the credit invoice.
(Exceptions are made for duplicate shipments made by FCI.)
If any repair or return equipment is received at FCI, freight collect, without prior factory consent,
FCI bills the sender for these charges.
Field Service Procedures
Contact an FCI field representative to request field service.
A field service technician is dispatched to the site from either the FCI factory or one of the FCI
representative offices. After the work is complete, the technician completes a preliminary field
service report at the customer site and leaves a copy with the customer.
Following the service call, the technician completes a formal, detailed service report. The formal
report is mailed to the customer within five days of the technician’s return to the factory or office.
Field Service Rates
All field service calls are billed at the prevailing rates as listed in the FCI Price Book unless
specifically excepted by the FCI Customer Service Manager. FCI reserves the right to bill for
travel times at FCI’s discretion.
Customers are charged for shipping costs related to the transfer of equipment to and from the job
site. They are also invoiced for field service work and travel expenses by FCI’s Accounting
Department.
This page is subject to proprietary rights statement on last page
34Doc. No. 06EN003368 Rev. C
ST75/ST75V MASS FLOWFLUID COMPONENTS INTERNATIONAL LLC
Fax No: _____________________________________ Direct Phone: _______________________________
From: ______________________________________No. of Pages: _______________________________
To expedite repairs the following Return Authorization Request form must be completed and faxedback to FCI before a Return Authorization Number will be issued. The Decontamination Statement must be signed and Applicable MSDS Sheets must be included with the shipment.
FCI will either fax or telephone you with the Return Authorization
Number upon receipt of the completed form.
Packing Procedures
1. Electronics should be wrapped in an anti-static or static-resistant bag (FCI can provide), then wrapped in protective
bubble wrap and surrounded with appropriate dunnage* in a
box. Instruments weighing up to 50 lbs., should be wrapped
in protective wrap and surrounded with appropriate dunnage*
in a box. Instruments weighing more than50 lbs., orex-tending more than four feet should be secured in wooden
crates by bolting the assemblies in place.
2. The sensorelement must be protected with tubing or other
sturdy wrapping or, when applicable, retracted completely and
secured into the Packing Gland Assembly. FCI will supply
suitable probe protectors at your request (see diagram - right).
3. FCI can supply crates at a nominal fee.
4. No more than four (4) small units packaged in each carton.
5. Packages weighing in excess of 70 lbs., or with the com-bined length and girth of more than 138” cannot be shipped
via UPS and should be shipped via carriers who specialize in
the transport of industrialized instrumentation.
6. FCI will not be held liable for damage caused during shipping.
7. To ensure immediate processing mark the RA number on the outside of the box. Items without
an RA number marked on the box or crate may be delayed.
8. Freight Must be “PrePaid” to FCI receiving door.
Anti-Static Bag
PVC
Probe protector
Factory Return Shipping Address:Fluid Components International LLC
1755 La Costa Meadows Drive
San Marcos, CA 92078-5115
Attn:Repair Department,
RA #____________
*
Approriate dunnage as defined by UPS, will protect package contents from a drop of 3 feet.
This message is intended for the use of the individual or entity to whom it is addressed and may contain proprietary data or
confidential business or financial information that can only be used, copied, or disclosed as authorized by Fluid Components.
Visit FCI on the Worldwide Web: w ww.fluidcomponents.com
1755 La Costa Meadows Drive, San Marcos, California 92078 USA ‡ Phone: 760-744-6950 ‡ 800-854-1993 ‡ Fax: 760-736-6250
This page is subject to proprietary rights statement on last page
35Doc. No. 06EN003368 Rev. C
FLUID COMPONENTS INTERNATIONAL LLC ST75/ST75V MASS FLOW
1755 La Costa Meadows Drive, San Marcos, CA 92078-5115 USA
Web Site: www.fluidcomponents.com / E-mail: techsupport@fluidcomponents.com
FCI Factory Technical Service Contact: ___________________________________________________________
Reason For Return Sensor Element Electronics As FoundTesting Credit
4.
Recalibrate (Old Data) Recalibrate (New Data) Other
(Note: A new Application Data Sheet (ADS) must be submitted for all recalibrations and re-certifications)
Payment Via Faxed Purchase Order
5.
(Note: A priced quotation is provided for all Non-Warranty repairs after equipment has been evaluated.
All Non-Warranty repairs are subject to a minimum evaluation charge)
Sensor Element Protector Requested Electronics Anti-Static Bag Requested
Decontamination Information !This Section Must Be Completed !
Exposure to hazardous materials is regulated by Federal, State (California), County and City laws and regulations.
These laws provide FCI’s employees with the “Right to Know” the hazardous or toxic materials or substances in which
they may come in contact while handling returned products. Consequently, our employees must have access to data
regarding the hazardous or toxic materials or substances which the equipment has been exposed to in your process(es).
Accordingly, prior to returning your instrument for evaluation/repair, please read then sign the certification below and
thoroughly comply with the applicable instructions.
I certify that the returned item(s) has(have) been thoroughly and completely cleaned. If the returned item(s) has(have)
been exposed to hazardous or toxic materials or substances, the undersigned attests that the attached Material Safety
Data Sheet(s) (MSDS) which cover said materials or substances are complete and accompany the returned item(s).
Furthermore, I understand that this Certificate, or providing a MSDS, shall not waive our responsibility to provide a
neutralized, decontaminated, and clean product for evaluation/repair to FCI. Cleanliness of a returned item or the
acceptability of the MSDS shall be at the sole discretion of FCI. Any item returned which does not comply with these
instructions shall be returned to your location Freight Collect and at your risk.
Process Flow Media ___________________________________________________________________________________
Authorized Signature ________________________________ Date _____________________________________________
FCI Document No. 05CS000004A
This page is subject to proprietary rights statement on last page
36Doc. No. 06EN003368 Rev. C
ST75/ST75V MASS FLOWFLUID COMPONENTS INTERNATIONAL LLC
Instrument Warranty
Goods furnished by the Seller are to be within the limits and of the sizes published by the Seller and subject to the Seller’s
standard tolerances for variations. All items made by the Seller are inspected before shipment, and should any of said
items prove defective due to faults in manufacture or performance under Seller approved applications, or fail to meet the
written specifications accepted by the Seller, they will be replaced or repaired by Seller at no charge to Buyer provided
return or notice of rejection of such material is made within a reasonable period but in no event longer than one (1) year for
non-calibration defects and one (1) year for calibration defects from date of shipment to Buyer, and provided further, that
an examination by Seller discloses to Seller’s reasonable satisfaction that the defect is covered by this warranty and that
the Buyer has not returned the equipment in a damaged condition due to Buyer’s or Buyer’s employees’, agents’, or
representatives’ negligence and Buyer has not tampered, modified, redesigned, misapplied, abused, or misused the goods
as to cause the goods to fail. In addition, this warranty shall not cover damage caused by Buyer’s exposure of the goods to
corrosive or abrasive environments. Moreover, Seller shall in no event be responsible for (1) the cost or repair of any work
done by Buyer on material furnished hereunder (unless specifically authorized in writing in each instance by Seller), (2) the
cost or repair of any modifications added by a Distributor or a third party, (3) any consequential or incidental damages,
losses, or expenses in connection with or by reason of the use of or inability to use goods purchased for any purpose, and
Seller’s liability shall be specifically limited to free replacement, or refund of the purchase price, at Seller’s option, provided
return or rejection of the goods is made consistent with this paragraph, and the Seller shall in no event be liable for
transportation, installation, adjustment, loss of good will or profits, or other expenses which may arise in connection with
such returned goods, or (4) the design of products or their suitability for the purpose for which they are intended or used.
Should the Buyer receive defective goods as defined by this paragraph, the Buyer shall notify the Seller immediately,
stating full particulars in support of his claim, and should the Seller agree to a return of the goods, the Buyer shall follow
Seller’s packaging and transportation directions explicitly. In no case are the goods to be returned without first obtaining a
return authorization from the Seller. Any repair or replacement shall be at Seller’s factory, unless otherwise directed, and
shall be returned to Seller transportation prepaid by Buyer. If the returned goods shall prove defective under this clause
they will be replaced or repaired by Seller at no charge to Buyer provided the return or rejection of such material is made
within a reasonable period, but in no event longer than (1) year from the date of shipment of the returned goods or the
unexpired terms of the original warranty period whichever is later. If the goods prove to be defective under this paragraph,
the Buyer shall remove the goods immediately from the process and prepare the goods for shipment to Seller. Continued
use or operation of defective goods is not warranted by Seller and damage occurring due to continued use or operation
shall be for Buyer’s account. Any description of the goods contained in this offer is for the sole purpose of identifying them,
and any such description is not part of the basis of the bargain, and does not constitute a warranty that the goods will
conform to that description. The use of any sample or model in connection with this offer is for illustrative purposes only, is
not part of the basis of the bargain, and is not to be construed as a warranty that the goods will conform to the sample or
model. No affirmation of that fact or promise made by the Seller, whether or not in this offer, will constitute a warranty that
the goods will conform to the affirmation or promise. THIS WARRANTY IS EXPRESSLY IN LIEU OF ANY AND ALL
OTHER EXPRESS OR IMPLIED WARRANTIES WITH RESPECT TO THE GOODS OR THEIR INSTALLATION, USE,
OPERATION, REPLACEMENT OR REPAIR, INCLUDING ANY IMPLIED WARRANTY OF MERCHANTABILITY OR
FITNESS OF PURPOSE; AND THE GOODS ARE BEING PURCHASED BY BUYER “AS IS”. SELLER WILL NOT BE
LIABLE BY VIRTUE OF THIS WARRANTY OR OTHERWISE FOR ANY SPECIAL, INCIDENTAL OR CONSEQUENTIAL
LOSS OR DAMAGE RESULTING FROM THE USE OR LOSS OF USE OF THE GOODS.
This page is subject to proprietary rights statement on last page
37Doc. No. 06EN003368 Rev. C
FLUID COMPONENTS INTERNATIONAL LLC ST75/ST75V MASS FLOW
Notes
This page is subject to proprietary rights statement on last page
38Doc. No. 06EN003368 Rev. C
ST75/ST75V MASS FLOWFLUID COMPONENTS INTERNATIONAL LLC
Notes
This page is subject to proprietary rights statement on last page
39Doc. No. 06EN003368 Rev. C
ST75/ST75V MASS FLOWFLUID COMPONENTS INTERNATIONAL LLC
FCI’s Complete Customer Commitment. Worldwide
ISO 9001:2000 and AS9100 Certified
Notice of Proprietary Rights
This document contains confidential technical data, including trade secrets and proprietary information which is the property of Fluid Components
International LLC (FCI). Disclosure of this data to you is expressly conditioned upon your assent that its use is limited to use within your company only
(and does not include manufacture or processing uses). Any other use is strictly prohibited without the prior written consent of FCI.
Visit FCI on the Worldwide Web: www.fluidcomponents.com