Fluid Components International RF83 Troubleshooting

CHAPTER 5 - TROUBLESHOOTING FLUID COMPONENTS INTL
5. Troubleshooting
Tools Needed
Digital Multimeter (DMM)
Quick Check
Check the jumper positions of J12, J13, J16 and J17. Jumper J12 and J17 energize the relays at flow or wet. Jumpers J13 and J16 energize the relays at no flow.
Check that the control circuit is firmly seated into it's socket. Check if power is present and customer fuses are good, if they are used. Follow the trouble shooting flow chart in Figure 5-1 near the end of this chapter.
Warning:
Only qualified personnel should attempt to test this instrument. The operator assumes all responsibilities for safe practices while troubleshooting.
The control circuit contains electrostatic discharge (ESD) sensitive devices. Use standard ESD precautions when handling the control circuit. See Chapter 2, Installation for ESD details.
Non-maintenance Observations
At this point, observe the system setup to verify operation. No disassembly or testing is required at this time.
Check Serial Numbers
Verify that the serial number of the flow element and the control circuit are the same when the instrument is used with the remote enclosure option. The flow element and the control circuit are a matched set and cannot be operated independently of each other.
Check Input Power
Verify that the correct power source is turned on and connected.
Check the Instrument Installation
Review the information on instrument installation in Chapter 2 to verify correct mechanical and electrical installation.
At the time of order the flow element placement should have been determined. However, if not, the flow element should be mounted at least 20 diameters downstream and 20 diameters upstream from any bends or interference in the process pipe or duct.
Check for Moisture
Check for moisture on the control circuit. Moisture on the control circuit may cause intermittent operation. Check for moisture on the flow element. If a component of the process media is near its saturation temperature it
may condense on the flow element. Place the flow element where the process media is well above the saturation temperature of any of the process gases.
Doc. No. 003147 Rev. B 5 - 1
Model RF83 Flow Switch
FLUID COMPONENTS INTL CHAPTER 5 - TROUBLESHOOTING
Check Application Design Requirements
Application design problems may occur with first time application instruments, although the design should also be checked on instruments that have been in operation for some time. If the application design does not match field conditions, errors occur.
1. Review the application design with plant operation personnel and plant engineers.
2. Ensure that plant equipment such as pressure and temperature instruments conform to the actual conditions.
3. Verify operating temperature, operating pressure, line size, and gas medium.
Control Circuit Dash Number Specification
Verify that wiring is connected per the correct wiring diagram. To find the correct diagram look at the control circuit dash number printed on the instrument name plate. Compare the dash number to Appendix A, Figure A-10. The figure will then guide the user to find the appropriate wiring diagrams. If there is a space within the dash number there is a standard option associated with the blank. See Note 4 in Figures A-10.
Verification of Flow Element Resistance
The measurements are based on a standard (4K ohm RTD at 70°F, or 21°C) flow element. Variation of ±100 ohms from nominal is to be expected, depending on temperature. The maximum allowable difference in resistance between matched RTD's is 1% at ambient temperature (immersed in water). The heater resistance should be 430±5 ohms for each heater. The heaters are connected in parallel, therefore the combined parallel resistance is 215 ohms. Be sure to subtract the cable resistance to get the true resistance.
Note:
When the flow element is connected to a remote enclosure control circuit, a shielded cable is to be used. Be sure that the cable shield is ONLY connected to the control circuit socket, terminal number 7. The other end of the shield should be left floating.
1. Turn off the operating power to the instrument.
2. Gently remove (pull straight out) the control circuit from the socket. Using a DMM, measure the resistance of sensor A and sensor B sensing elements.
3. Measure the resistances as found in Table 5-1 to determine if the flow element is functional. For remote instruments, if there is a faulty reading at the control circuit socket, (control circuit pulled out) make
the resistance measurements found in Table 5-2.
For the remote configuration (the control circuit is in a separate enclosure from the flow element), if an open circuit exists for one RTD and a value of twice the resistance exists for the other RTD then there is probably a miswiring or mislabeling problem.
For the local configuration (the control circuit is in the same enclosure as the flow element), if there is an indication of an open or short in the RTD, the flow element will need to be replaced.
After replacing the flow element, it will be necessary to follow the set point adjustment procedures found in the Operation Chapter before returning the instrument to service. If a millivolt output option is provided on the control circuit, the zero adjustment procedure must be performed.
If the flow element resistance is correct, proceed to the voltage verification test.
Table 5-1. Local Flow Element Resistances In Ohms
From Terminal (Pin) To Terminal (Pin) Expected Ohms* Heater (7) Heater (10) 215 Sen A (9) Com Sen (8) 4000 Sen A (9) Sen B (7) 8000 Sen B (7) Com Sen (8) 4000
Model RF83 Flow Switch 5 - 2 Doc. No. 003147 Rev. B
(control circuit enclosure)