APPENDIX E - CE CONFORMANCE FLUID COMPONENTS INTL
Appendix E. CE Conformance
Approved CE Marking For The MT91 Configuration
Approved Options
1) all insertion U-lengths
2) all specified application combinations
3) 85 to 260 VAC power input.
4) all metallic sensor elements
5) all cable jackets and lengths
6) Aluminum and 300 series stainless steel NEMA 4X local and remote applications.
7) All process connections and flanges
8) All process temperature ranges
Approved Part Numbers
1-15216 17-18
MT91
All options conform to CE marking
requirements.
All other configurations must be approved for CE marking conformity by FCI's Engineering Department.
X Y
Conditionally Approved CE Marking
24 VDC Factory Selectable Options
Input power of 24 VDC may be optionally selected. Since the manufacturer does not supply the power source or
transformer for these connections, the responsibility for the conditioning of these sources and associated compliance
to the EMC Directive shall be the responsibility of the user.
Panel and Rack Mounted Control Circuit Options
Conditionally approved models included in this family are supplied with panel or rack mounted electronic assemblies
without a remote enclosure (MT91 block 16, codes 1 or 3). Since the manufacturer does not supply an EMC
enclosure for these configurations the associated compliance to the EMC directive shall be the responsibility of the
user.
X or Y options only conform to CE marking
Installation Conformity Criteria
Grounding
The enclosure must be grounded to earth ground through a path of less than 1 ohm.
Interconnecting Cables
All interconnecting cables between the local enclosure, power source and monitoring device shall be enclosed in
metal conduit. AC power input cabling shall be enclosed separately in conduit before entrance into the instrument
and is not to be combined with switch or monitor output cabling.
Doc. No. 003185 Rev. B
E - 1 Model MT91 Multipoint Flowmeter
FLUID COMPONENTS INTL APPENDIX E - CE CONFORMANCE
Documentation Location
European Location
The technical documentation file part A resides at Fluid Components Intl, European Service Center, Beatrix De
Rijkweg 8, 5657 Eg Einhoven, Netherlands, Phone: 31-40-2-571-972 FAX: 31-40-2-517-809.
Manufacturer Location
The technical documentation file part B resides at the Configuration Management department of Fluid Components
Intl, 1755 La Costa Meadows Dr. San Marcos, CA. 92069 USA, Phone: 1 (800) 854-1993
FAX: 1 (619) 736-6250.
Model MT91 Multipoint Flowmeter E - 2 Doc. No. 003185 Rev. B