Fluid Components International CMU User Manual

Coriolis mass flowmeter
CMU
CT
Operating Manual
Please read the instructions carefully and store them in a safe place
CMU & CT OPERATING MANUAL Part No 06EN003380 rev 1.6
Page 1 of 112
INTRODUCTION ............................................................................................................ 9
I. Shipping and storage; product inspection............................................................................................. 9
II. Warranty ............................................................................................................................................ 9
III. Application domain the operating manual .........................................................................................9
IV. Measures to be taken before sending your device to the manufacturer for repair............................ 9
V. Supplementary operating instructions regarding the HART® interface ............................................ 9
1. STEPS PRIOR TO OPERATION......................................................................... 10
1.1 Installation and servicing ................................................................................................................. 11
1.2 Safety advisory for the user ............................................................................................................. 11
1.3 Hazard warnings.............................................................................................................................. 11
1.3.1 Danger...................................................................................................................................... 11
1.3.2 Warning .................................................................................................................................... 11
1.3.3 Caution ..................................................................................................................................... 12
1.3.4 Note .......................................................................................................................................... 12
1.4 Proper use of the device.................................................................................................................. 12
1.5 Returning your flowmeter for servicing or calibration ...................................................................... 12
1.6 Replacement of the transmitter electronics ..................................................................................... 13
2. IDENTIFICATION ................................................................................................ 13
3. THE CMU SENSOR ............................................................................................ 14
3.1 Application domain of the CMU sensor ...........................................................................................14
3.2 Mode of operation............................................................................................................................ 14
3.2.1 Measuring principle .................................................................................................................. 14
3.2.2 System configuration................................................................................................................ 14
3.2.3 Input.......................................................................................................................................... 14
3.3 Custody transfer operations ............................................................................................................ 14
3.4 Performance characteristics of the CMU sensor............................................................................. 15
3.4.1 Reference conditions................................................................................................................ 15
3.4.2 CMU flow ranges ...................................................................................................................... 15
3.4.3 Density measurement............................................................................................................... 16
3.4.4 Accuracy................................................................................................................................... 16
3.4.5 Pressure loss CMU................................................................................................................... 17
3.4.6 Ambient temperature ................................................................................................................ 18
3.4.7 Ambient temperature range...................................................................................................... 18
Page 2 of 112
CMU & CT OPERATING MANUAL
3.4.8 Storage temperature ................................................................................................................ 18
3.4.9 Climatic category...................................................................................................................... 18
3.4.10 Ingress protection..................................................................................................................... 18
3.5 Operating conditions........................................................................................................................ 18
3.5.1 Installation ................................................................................................................................ 18
3.5.2 Installation positions ................................................................................................................. 19
3.5.3 Assessment of installation position .......................................................................................... 20
3.5.4 Pressure surges ....................................................................................................................... 21
3.5.5 Using the device with hazardous fluids.................................................................................... 21
3.5.6 Vibration stability ...................................................................................................................... 21
3.6 Process conditions .......................................................................................................................... 23
3.6.1 Process temperature ................................................................................................................ 23
3.6.2 Physical state ........................................................................................................................... 23
3.6.3 Viscosity ................................................................................................................................... 23
3.6.4 Gas content .............................................................................................................................. 23
3.6.5 Process temperature range...................................................................................................... 23
3.6.6 Process pressure range ........................................................................................................... 23
3.6.7 Outlet pressure......................................................................................................................... 23
3.7 Connection to the transmitter .......................................................................................................... 23
3.7.1 Integral mount configuration..................................................................................................... 23
3.7.2 Remote mount configuration .................................................................................................... 23
3.8 Construction details ......................................................................................................................... 24
3.8.1 Dimensions and weight ............................................................................................................ 24
3.8.2 Dimension drawing for the types CMU-C to CMU-H................................................................ 26
3.8.2.1 Standard version dimension drawing................................................................................ 26
3.8.2.2 Integral mount version up to 150 °C (302 °F) ................................................................... 26
3.8.2.3 Remote mount version dimension drawing....................................................................... 27
3.8.2.4 Remote mount version dimension drawing up to 180 °C (356 °F) ................................... 27
3.8.2.5 Remote mount version dimension drawing up to 260 °C (500 °F) ................................... 28
3.8.3 Dimension drawing for the types CMU-J to CMU-R ................................................................ 28
3.8.3.1 Standard version dimension drawing................................................................................ 28
3.8.3.2 Integral mount configuration up to 180 °C (356°F) ........................................................... 29
3.8.3.3 Remote mount version dimension drawing....................................................................... 29
3.8.3.4 Remote mount version dimension drawing up to 180 °C (356°F) .................................... 30
3.8.3.5 Remote mount version up to 260 °C (500°F).................................................................... 30
3.8.4 Heater dimension drawings for CMU-J up to CMU-R.............................................................. 31
3.8.4.1 Heater for standard version .............................................................................................. 31
3.8.4.2 Heater for integral mount version up to 150 °C (302°F) ................................................... 31
3.8.4.3 Heater for remote mount version ...................................................................................... 32
3.8.4.4 Heater for remote mount version up to 180 °C (356°F) .................................................... 32
3.8.4.5 Heater for remote mount version up to 260 °C (500 °F) ................................................... 33
3.8.5 Material..................................................................................................................................... 33
3.9 Sensor CMU approvals ................................................................................................................... 34
3.9.1 Explosion protection ................................................................................................................. 34
3.9.2 CE marking............................................................................................................................... 34
3.9.3 Custody transfer operations .....................................................................................................34
4. COMMISSIONING................................................................................................34
4.1 Zero point calibration ....................................................................................................................... 34
CMU & CT OPERATING MANUAL Part No 06EN003380 rev 1.6
Page 3 of 112
4.2 Startup conditions ............................................................................................................................ 34
5. APPLICATION DOMAIN OF THE CT TRANSMITTER........................................ 35
6. CT TRANSMITTER: MODE OF OPERATION AND CONFIGURATION ............. 35
6.1 Measuring principle.......................................................................................................................... 35
6.2 System configuration ....................................................................................................................... 36
6.2.1 DSB data memory module ....................................................................................................... 36
7. INPUT.................................................................................................................. 37
7.1 Measured variable ........................................................................................................................... 37
7.2 Measuring range.............................................................................................................................. 37
8. OUTPUT.............................................................................................................. 38
8.1 Output signal.................................................................................................................................... 38
8.2 Failure signal ................................................................................................................................... 38
8.3 Load ................................................................................................................................................. 39
8.4 Damping........................................................................................................................................... 39
8.5 Low flow cutoff ................................................................................................................................. 39
9. CT PERFORMANCE CHARACTERISTICS......................................................... 39
9.1 Reference conditions....................................................................................................................... 39
9.2 Measured error ................................................................................................................................ 39
9.3 Repeatability error ........................................................................................................................... 39
9.4 Influence of ambient temperature.................................................................................................... 39
10. CT OPERATING CONDITIONS........................................................................... 40
10.1 Installation conditions and cable glands ...................................................................................... 40
10.2 Environmental conditions............................................................................................................. 40
10.2.1 Ambient temperature ................................................................................................................ 40
10.2.2 Ambient temperature range...................................................................................................... 40
10.2.3 Storage temperature................................................................................................................. 40
10.2.4 Ingress protection ..................................................................................................................... 40
10.3 Process conditions....................................................................................................................... 41
10.3.1 Fluid temperature ..................................................................................................................... 41
Page 4 of 112
CMU & CT OPERATING MANUAL
10.3.2 Physical state ........................................................................................................................... 41
10.3.3 Viscosity ................................................................................................................................... 41
10.3.4 Fluid temperature limit.............................................................................................................. 41
10.3.5 Flow rate limit ........................................................................................................................... 41
10.3.6 Pressure loss............................................................................................................................ 41
11. CONSTRUCTION DETAILS.................................................................................42
11.1 Type of construction/dimensions .................................................................................................42
11.2 Weight.......................................................................................................................................... 43
11.3 Material ........................................................................................................................................ 43
11.4 End connection ............................................................................................................................ 43
11.5 Electrical connection.................................................................................................................... 44
11.5.1 CT connections ........................................................................................................................ 44
11.5.2 Wiring diagram ......................................................................................................................... 46
11.5.2.1 Wiring diagram for the integral mount configuration of sensor and CT ............................ 46
11.5.2.2 Wiring diagram for the remote mount configuration of sensor and CT............................. 47
11.5.3 HART®...................................................................................................................................... 48
11.5.4 Communication via SensorPort................................................................................................ 48
12. CONTROL UNIT BE2 ...........................................................................................49
12.1 Introduction .................................................................................................................................. 49
12.2 Display ......................................................................................................................................... 49
12.3 Operating modes ......................................................................................................................... 50
12.4 Operation ..................................................................................................................................... 50
12.4.1 Operator interface .................................................................................................................... 50
12.4.2 The keys and their functions ....................................................................................................51
12.4.3 Functional classes, functions and parameters......................................................................... 52
12.4.3.1 Selection window/make a selection .................................................................................. 52
12.4.3.2 Input window/modify a value ............................................................................................. 53
12.4.3.3 Passwords......................................................................................................................... 53
13. CT TRANSMITTER FUNCTIONS.........................................................................54
13.1 MEASURED VALUES functional class........................................................................................ 55
13.1.1 Mass flow.................................................................................................................................. 56
13.1.2 Volume flow.............................................................................................................................. 56
13.1.3 Counter forward........................................................................................................................ 56
13.1.4 Counter reverse........................................................................................................................ 56
13.1.5 Density...................................................................................................................................... 57
13.1.6 Temperature............................................................................................................................. 57
13.1.7 Elapsed time............................................................................................................................. 57
13.1.8 Mass flow + Counter forward ................................................................................................... 57
13.1.9 Mass flow + Density ................................................................................................................. 58
13.1.10 Mass flow + Temperature..................................................................................................... 58
13.1.11 Volume flow + Counter forward ............................................................................................ 58
CMU & CT OPERATING MANUAL Part No 06EN003380 rev 1.6
Page 5 of 112
13.1.12 Volume flow + Density .......................................................................................................... 58
13.1.13 Display mode during startup ................................................................................................. 59
13.1.14 Raw values ........................................................................................................................... 59
13.2 PASSWORD functional class ...................................................................................................... 60
13.2.1 Customer password.................................................................................................................. 60
13.2.2 Change customer password..................................................................................................... 61
13.2.3 Service password ..................................................................................................................... 61
13.3 COUNTER functional class.......................................................................................................... 62
13.3.1 Unit of counters ........................................................................................................................ 63
13.3.2 Reset counters ......................................................................................................................... 63
13.4 MEASUREMENT PROCESSING functional class ...................................................................... 64
13.4.1 Damping ................................................................................................................................... 65
13.4.2 Low flow cut-off......................................................................................................................... 65
13.4.3 Low flow cut-off hysteresis ....................................................................................................... 65
13.4.4 Zero point calibration................................................................................................................ 66
13.5 FLOW functional class ................................................................................................................. 67
13.5.1 Mass flow QM unit .................................................................................................................... 68
13.5.2 Mass flow QM range................................................................................................................. 68
13.5.3 Mass flow QM limit MIN............................................................................................................ 69
13.5.4 Mass flow QM limit MAX........................................................................................................... 69
13.5.5 Mass flow QM limit hysteresis .................................................................................................. 69
13.5.6 Volume flow QV unit ................................................................................................................. 70
13.5.7 Volume flow QV range.............................................................................................................. 70
13.6 DENSITY functional class............................................................................................................ 71
13.6.1 Density measurement on/off .................................................................................................... 72
13.6.2 Density unit............................................................................................................................... 73
13.6.3 Density lower-range value........................................................................................................ 74
13.6.4 Density upper-range value .......................................................................................................74
13.6.5 Density limit MIN....................................................................................................................... 74
13.6.6 Density limit MAX ..................................................................................................................... 75
13.6.7 Density limit hysteresis ............................................................................................................. 75
13.6.8 Density limit for empty pipe ...................................................................................................... 75
13.6.9 Fixed density ............................................................................................................................ 76
13.6.10 Reference/process density display ....................................................................................... 76
13.6.11 Temperature coefficient........................................................................................................ 76
13.6.12 Reference temperature ......................................................................................................... 77
13.6.13 Reference pressure ..............................................................................................................77
13.7 TEMPERATURE functional class ................................................................................................ 78
13.7.1 Temperature unit ...................................................................................................................... 79
13.7.2 Temperature lower-range value ............................................................................................... 79
13.7.3 Temperature upper-range value............................................................................................... 79
13.7.4 Temperature limit MIN.............................................................................................................. 80
13.7.5 Temperature limit MAX............................................................................................................. 80
13.7.6 Max. measured temperature .................................................................................................... 80
13.8 PULSE OUTPUT functional class................................................................................................ 81
13.8.1 Pulse or frequency output ........................................................................................................ 81
13.8.2 Pulse output unit....................................................................................................................... 82
13.8.3 Pulse value ............................................................................................................................... 82
13.8.4 Pulse width ............................................................................................................................... 83
Page 6 of 112
CMU & CT OPERATING MANUAL
13.9 STATUS functional class ............................................................................................................. 84
13.9.1 Status output active state ......................................................................................................... 85
13.9.2 Status output 1 assignment......................................................................................................85
13.9.3 Status output 2 assignment......................................................................................................86
13.9.4 Binary input assignment........................................................................................................... 86
13.10 CURRENT OUTPUTS functional class ....................................................................................... 87
13.10.1 Current output I1 0/4 to 20 mA ............................................................................................. 88
13.10.2 Current output I1 alarm.........................................................................................................88
13.10.3 Current output I1 assignment ............................................................................................... 88
13.10.4 Current output I2 0/4 to 20 mA ............................................................................................. 89
13.10.5 Current output I2 alarm.........................................................................................................90
13.10.6 Current output I2 assignment ............................................................................................... 90
13.11 SIMULATION functional class .....................................................................................................91
13.11.1 Simulation on/off ................................................................................................................... 92
13.11.2 Direct simulation ................................................................................................................... 92
13.11.3 Measured value simulation................................................................................................... 93
13.11.3.1 Simulation mass flow QM abs ......................................................................................... 93
13.11.3.2 Density simulation............................................................................................................ 93
13.11.3.3 Temperature measurement simulation............................................................................ 93
13.11.4 Direct simulation of outputs .................................................................................................. 94
13.11.4.1 Status output simulation .................................................................................................. 94
13.11.4.2 Pulse output simulation.................................................................................................... 94
13.11.4.3 Simulation current output I1............................................................................................. 94
13.11.4.4 Simulation current output I2............................................................................................. 94
13.12 SELF-TEST function class........................................................................................................... 95
13.12.1 Sensor test on/off.................................................................................................................. 96
13.12.2 Max. deviation of excitation .................................................................................................. 96
13.12.3 Self-test calibration ............................................................................................................... 96
13.12.4 Monitoring of sensor amplitude and excitation current......................................................... 96
13.13 UMC TRANSMITTER SETTINGS functional class ..................................................................... 97
13.13.1 Language .............................................................................................................................. 98
13.13.2 Serial number ....................................................................................................................... 98
13.13.3 Software version................................................................................................................... 98
13.13.4 Reset system error ............................................................................................................... 98
13.13.5 Profibus/Modbus device address ......................................................................................... 99
13.14 SENSOR SETTINGS functional class ....................................................................................... 100
13.14.1 Sensor constant C .............................................................................................................. 100
13.14.2 Sensor material................................................................................................................... 101
13.14.3 Flow direction...................................................................................................................... 102
14. USE OF THE CT FOR CUSTODY TRANSFER OPERATIONS.........................103
14.1 Programming the transmitter ..................................................................................................... 103
14.2 Binary input (resetter) ................................................................................................................ 103
14.3 Self-test error ............................................................................................................................. 103
14.4 Verification stamp/stamp position .............................................................................................. 104
CMU & CT OPERATING MANUAL Part No 06EN003380 rev 1.6
Page 7 of 112
14.5 HART® communication in Custody transfer mode..................................................................... 104
15. CT TRANSMITTER ERROR MESSAGES......................................................... 105
15.1 Standard operating mode .......................................................................................................... 105
15.2 Custody transfer mode............................................................................................................... 105
15.3 List of error messages................................................................................................................ 105
15.3.1 Display of self-test errors........................................................................................................ 105
15.3.2 Display of system error........................................................................................................... 108
16. CERTIFICATES AND APPROVALS.................................................................. 110
17. STANDARDS AND AUTHORIZATIONS............................................................ 110
17.1 General standards and directives .............................................................................................. 110
17.2 Electromagnetic compatibility .................................................................................................... 110
17.3 Ex-Approval transmitter.............................................................................................................. 110
18. DECONTAMINATION CERTIFICATE FOR DEVICE CLEANING ..................... 111
Page 8 of 112
CMU & CT OPERATING MANUAL

Introduction

I. Shipping and storage; product inspection

Shipping and storage
The device is to be safeguarded against dampness, dirt, impact and damage. Product inspection
Upon receipt of the product, check the contents of the box and the product particulars against the infor­mation on the delivery slip and order form so as to ensure that all ordered components have been sup­plied. Notify us of any shipping damage immediately upon receipt of the product. Any damage claim re­ceived at a later time will not be honored.
II. Warranty
Your flowmeter was manufactured in accordance with the highest quality standards and was thoroughly tested prior to shipment. However, in the event any problem arises with your device, we will be happy to resolve the problem for you as quickly as possible under the terms of the warranty which can be found in the terms and conditions of delivery. Your warranty will only be honored if the device was installed and operated in accordance with the instructions for your device. Any mounting, commissioning and/or main­tenance work is to be carried out by qualified and authorized technicians only.

III. Application domain the operating manual

The present manual applies to Coriolis mass flowmeters that are operated in conjunction with the CT transmitter.
IV. Measures to be taken before sending your device to the manufacturer for repair
It is important that you do the following before shipping your flowmeter to Fluid Components Intl for repair:
Enclose a description of the problem with your device. Describe in as much detail as possible the
application and the physical and chemical properties of the fluid.
Remove any residues from the device and be sure to clean the seal grooves and recesses thor-
oughly. This is particularly important if the fluid is corrosive, toxic, carcinogenic, radioactive or other-
wise hazardous. The operator is liable for any substance removal or personal damage costs arising from inadequate cleaning of a device that is sent for repair.
V. Supplementary operating instructions regarding the HART® interface
For information regarding operation of the transmitter using the HART tion of the CT transmitter using the HART
®
hand-held terminal.”
®
hand-held terminal, see “Opera-
CMU & CT OPERATING MANUAL
Page 9 of 112
1. Steps prior to operation
It is essential that you read these operating instructions before installing and operat­ing the device. The device is to be installed and serviced by a qualified technician only. The CT transmitter is to be used exclusively to measure mass and volume flow, as well as liquid and gas density and temperature, in conjunction with a CMM, CMB or CMU sensor.
Downloading of the present document from our web site
and printing out this document is allowed only for purposes of using our mass flow­meters. All rights reserved. No instructions, wiring diagrams, and/or supplied software, or any portion thereof, may be produced, stored, in a retrieval system or transmitted by any means, electronic, me­chanical, photocopying or otherwise, without the prior written permission of FCI.
Although the materials in the present document were prepared with extreme care, errors cannot be ruled out. Hence, neither the company, the programmer nor the author can be held legally or otherwise re­sponsible for any erroneous information and/or any loss or damage arising from the use of the informa­tion enclosed.
FCI extends no express or implied warranty in regard to the applicability of the present document for any purpose other than that described.
We plan to optimize and improve the products described and in so doing will incorporate not only our own ideas but also, and in particular, any suggestions for improvement made by our customers. If you feel that there is any way in which our products could be improved, please send your suggestions to the following address:
FCI Fluid Components International LLC Coriolis Product Manager
1755 La Costa Meadows Drive San Marcos, CA 92078
or: via fax: 760 - 736 - 6250 via E-mail:
www.fluidcomponents.com/, where you will also find contact information for the FCI representative for
your area. For factory direct questions, contact us at
mailto:techsupport@fluidcomponents.com
We reserve the right to change the technical data in this manual in light of any technical pro­gress that might be made. For updates regarding this product, visit our website at
mailto:info@fluidcomponents.com.
www.fluidcomponents.com/
Page 10 of 112
CMU & CT OPERATING MANUAL

1.1 Installation and servicing

The devices described in this manual are to be installed and serviced by qualified technical personnel.

Warning

Before servicing the device, it must be completely switched off, and disconnected from all peripheral devices. The technician must also check to ensure that the device is com­pletely off-circuit. Only original replacement parts are to be used.
Fluid Components International accepts no liability for any loss or damage of any kind arising from improper operation of any product, improper handling or use of any replacement part, or from external electrical or mechanical effects, overvoltage or lightning. Any such improper opera­tion, use or handling shall automatically invalidate the warranty for the product concerned.
In the event a problem arises with your device, please contact us at one of the following numbers to ar­range to have your device repaired:
Phone: 760 – 744 - 6950 Fax: 760 – 736 - 6250
Contact our customer service department if your device needs repair or if you need assistance in diag­nosing a problem with your device

1.2 Safety advisory for the user

The present document contains the information that you need in order to operate the CMU Series Coriolis mass flowmeter properly. The document is intended for use by qualified personnel. This means person­nel who are qualified to operate the device described herein safely, including trical engineers, or service technicians who are conversant with the safety regulations pertaining to the use of electrical and automated technical devices and with the applicable laws and regulations in their own country.
Such personnel must be authorized by the facility operator to install, commission and service the product described, and are to read and understand the contents of the present operating instructions before work­ing with the device.
electronics engineers, elec-
1.3 Hazard warnings
The purpose of the hazard warnings listed below is to ensure that device operators and maintenance personnel are not injured and that the flowmeter and any devices connected to it are not damaged.
The safety advisories and hazard warnings in the present document that aim to avoid placing operators and maintenance personnel at risk and to avoid material damage are prioritized using the terms listed below, which are defined as follows in regard to these instructions and the advisories pertaining to the device itself.

1.3.1 Danger

means that failure to take the prescribed precautions stantial material damage.
will result in death, severe bodily injury, or sub-
1.3.2 Warning
means that failure to take the prescribed precautions stantial material damage.
CMU & CT OPERATING MANUAL Part No 06EN003380 rev 1.6
could result in death, severe bodily injury, or sub-
Page 11 of 112
1.3.3 Caution
means that the accompanying text contains important information about the product, handling the product or about a section of the documentation that is of particular importance.
1.3.4 Note
means that the accompanying text contains important information about the product, handling the product or about a section of the documentation that is of particular importance.

1.4 Proper use of the device

Warning
The operator is responsible for ensuring that the material used in the sensor and housing is suitable and that such material meets the requirements for the fluid be­ing used and the ambient site conditions. The manufacturer accepts no responsi­bility in regard to such material and housing.
Warning In order for the device to perform correctly and safely, it must be shipped, stored, set up, mounted operated and maintained properly.

1.5 Returning your flowmeter for servicing or calibration

Before sending your flowmeter back to us for servicing or calibration, make sure it is completely clean. Any residues of substances that could be hazardous to the environment or human health are to be re­moved from all crevices, recesses, gaskets, and cavities of the housing before the device is shipped.
Warning The operator is liable for any loss or damage of any kind, including personal injury, decontamination measures, removal operations and the like that are attributable to inadequate cleaning of the device.
The device is to be accompanied by a document describing the problem with the device. Please include in this document the name of a contact person that our technical service department can get in touch with so that we can repair your device as expeditiously as possible and therefore minimize the cost of repair­ing it.
Any device sent in for servicing is to be accompanied by a certificate as specified in Section 18 Decontamination certificate for device cleaning.
Page 12 of 112
CMU & CT OPERATING MANUAL

1.6 Replacement of the transmitter electronics

Before replacing the transmitter electronics, read the safety instructions in Section 1.1 Installation and servicing on page 11.
Warning Make sure that you abide by the applicable standards and regulations pertain­ing to electrical devices, device installation and process technology when re­placing the transmitter electronics. The highly integrated electronic components in the device carry the risk of ESD hazards and are only protected when in­stalled in the device pursuant to EMC standards.
Before dismantling the DAB data memory module (see Section 6.2.1 DSB on page 36) remove it from the device and plug into the replacement part. To remove the electronics insert, first remove the four fasten­ing screws. Then slowly slide in the replacement part, making sure that it is oriented the same way as the original part, until the component reaches the floor of the housing. Be careful not to damage the contact strip. Then reinstall the four fastening screws.
Caution The complete insert is to be replaced with all of its printed boards (except for the memory module). This is particularly important for the explosion-proof transmitter. The specified precision and interchangeability of the electronics are only guaranteed if the complete insert is replaced.
2. Identification
Manufacturer FCI Fluid Components International LLC
1755 La Costa Meadows Drive San Marcos, CA 92078
Phone: 760 – 744 – 6950 Fax: 760 – 736 – 6250
Internet: E-mail: mailto:techsupport@fluidcomponents.com
European Office:
Persephonestraat 3-01 5047 TT Tilburg Netherlands
Phone: +31 – 13 – 515 9989 Fax: +31 – 13 – 579 9036
Product type Mass flowmeter for liquid and gaseous products
Product name Sensor type CMU
Transmitter type CT, suitable for CMM, CMB and CMU Coriolis mass flow­meters
Version no. 1.6, dated April 12, 2006
CMU & CT OPERATING MANUAL Part No 06EN003380 rev 1.6
http://www.fluidcomponents.com
Page 13 of 112
[
×⋅⋅
=
ω
3. The CMU sensor

3.1 Application domain of the CMU sensor

The CMU sensor is intended for use solely for direct and continuous mass flow measurement of liquids and gases, irrespective of their conductivity, density, temperature, pressure, or viscosity. The sensor is also intended for use for the direct and continuous mass flow measurement of chemical fluids, suspen­sions, molasses, paint, varnish, lacquer, pastes and similar materials.

3.2 Mode of operation

3.2.1 Measuring principle
The Coriolis mass flowmeter is based on the principle whereby in a rotating system a force (known as the Coriolis force) is exerted on a mass at a rotation point that is moving towards or away from this point.
]
2
3.2.2 System configuration
The flowmeter consists of a sensor that is mounted in a pipe, and a transmitter (see Section 5 Application domain of the CT on pp. 35), that can be directly mounted on the sensor or installed sepa-
rately (e.g. on a wall).
The transmitter oscillates the flow tubes in the sensor over a excitation coil and picks up, via the sensor coil, the measuring signal which is proportional to the mass flow. After being temperature compensated, the measuring signal is converted into an analog output signal that is consistent with the measuring range setting.
C
vmF
3.2.3 Input
Measured variables: mass flow, density, temperature; volume flow is calculated

3.3 Custody transfer operations

Units designated for custody transfer operation may be certified in accordance to the local or national ordinance. Transmitters ordered for custody transfer applications incorporate special tamper-proof soft­ware, sealed and certified, that prevents the reset of the internal totalizer.
Page 14 of 112
CMU & CT OPERATING MANUAL

3.4 Performance characteristics of the CMU sensor

3.4.1 Reference conditions
Established flow profile
Inlet section has to correspond to mounting length
Operation is to be realized in the presence of downstream control valves
Measurement is to be realized in the absence of any gas bubbles
Flow tubes are to be kept clean at all times
Process temperature is to be regulated as specified in Section 3.6.1 Process temperature on
page 23
Process pressure is to be regulated as specified in Section 3.6.6 Process pressure range on
page 23
Ambient temperature is to range from + 10 °C to + 30 °C (50 °F to 86 °F)
Warm-up period: 15 minutes
Standard calibration is to be realized at 20 %, 50 % and 100 % (three times each)
High-frequency interference is to be regulated as specified in Section 17.2 Electromagnetic
compatibility on page 110

3.4.2 CMU flow ranges

Mass flow
Min.
measuring range
Model
CMU-C 60 [2.2] 600 [22.0] 330 [12.1] 0.06 [0.002] CMU-D 250 [9.2] 2,500 [91.9] 1,150 [42.3] 0.25 [0.01]
CMU-E 1,200 [44.1] 12,000 [440.9] 5,250 [192.9] 1.2 [0.04] CMU-G 3,000 [110.2] 30,000 [1,102.3] 20,000 [734.9] 3 [0.1] CMU-H 6,000 [220.5] 60,000 [2,204.6] 55,000 [2,020.9] 6 [0.2]
CMU-J 20,000 [734.9] 80,000 [2,939.4] 74,000 [2,719.0] 8 [0.3] CMU-K 25,000 [918.6] 120,000 [4,409.2] 118,000 [4,335.7] 12 [0.4]
CMU-L 30,000 [1,102.3] 200,000 [7,348.6] 200,000 [7,348.6] 20 [0.7] CMU-N 60,000 [2,204.6] 460,000 [16,901.8] 460,000 [16,901.8] 46 [1.7] CMU-P 150,000 [5,511.5] 700,000 [25,720.2] 700,000 [25,720.2]* 70 [2.6] CMU-Q 300,000 [11,022.9] 1,500,000 [55,114.6] 1,350,000 [49,603.2] 150 [5.5] CMU-R 400,000 [14,697.2] 2,200,000 [80,834.8] 1,900,000 [69,811.9] 220 [8.1]
Reference conditions: in conformity with IEC 770: Temperature: 20 °C, relative humidity: 65 %, air pressure: 101.3 kPa Fluid: water
kg/h [lbs/min] kg/h [lbs/min] kg/h [lbs/min] kg/h [lbs/min]
Max.
measuring range
Nominal
(¬p=1bar)
* (
Zero point stability
(of range)
p=0,6bar)
CMU & CT OPERATING MANUAL
Page 15 of 112
3.4.3 Density measurement
The attainable accuracy depends on the selected calibration type.
Without calibration no density measurement is possible and the empty pipe recognition is not available!
Density accuracy
Model
CMU-C 5 g/l 2 g/l CMU-D 5 g/l 2 g/l CMU-E 5 g/l 1 g/l CMU-G 5 g/l 1 g/l CMU-H 5 g/l 1 g/l
CMU-J 5 g/l 2 g/l
CMU-K 5 g/l 2 g/l
CMU-L 5 g/l 2 g/l CMU-N 5 g/l 2 g/l CMU-P 5 g/l 2 g/l CMU-Q 5 g/l 2 g/l CMU-R 5 g/l 2 g/l
without 3-Point 5-Point
no measurement of density
3.4.4 Accuracy
Mass flow
Accuracy CMU-C to CMU-H ± 0.1% of actual flow + zero point stability (see Section 3.4.2 CMU
flow ranges)
Accuracy CMU-J to CMU-R ± 0.15% of actual flow + zero point stability (see Section 3.4.2 CMU
flow ranges)
Repeatability error ± 0.05% of actual flow (sensor with transmitter)
Additional measured values
Volume flow ± 0.2 % of actual value + zero point stability Temperature ± 0.5 °C Hysteresis n/a Settling time 1 to 15 seconds Startup drift 15 minutes Long-term drift ± 0.02 % of upper-range value per year Influence of ambient temperature ± 0.005 % per K Influence of fluid temperature Compensated Influence of fluid pressure For fluids: too small to be relevant
Page 16 of 112
CMU & CT OPERATING MANUAL

3.4.5 Pressure loss CMU

Model
CMU-C 60 kg/h 600 kg/h
CMU-E 1200 kg/h
CMU-G
CMU-H 6000 kg/h
CMU-J 20000 kg/h
CMU-K 25000 kg/h
CMU-L 30000 kg/h
CMU-N 60000 kg/h
CMU-P 150000 kg/h
CMU-Q 300000 kg/h
CMU-R 400000 kg/h
Min.
measuring range
3000 kg/h
measuring range
12000 kg/h
30000 kg/h
60000 kg/h
80000 kg/h
120000 kg/h
200000 kg/h
460000 kg/h
700000 kg/h
1500000 kg/h
2200000 kg/h
2500 kg/hCMU-D 250 kg/h
Max.
Pressure loss [water (20°C), 1 mPas]
60 kg/h 150 kg/h 300 kg/h 450 kg/h 600 kg/h
0.05 bar 0.25 bar 0.84 bar 1.70 bar 2.82 bar 250 kg/h 625 kg/h 1250 kg/h 1875 kg/h 2500 kg/h
0.07 bar 0.35 bar 1.18 bar 2.39 bar 3.95 bar
1200 kg/h 3000 kg/h 6000 kg/h 9000 kg/h 12000 kg/h
0.07 bar 0.37 bar 1.23 bar 2.51 bar 4.15 bar
3000 kg/h 7500 kg/h 15000 kg/h 22500 kg/h 30000 kg/h
0.04 bar 0.21 bar 0.70 bar 1.43 bar 2.36 bar
6000 kg/h 15000 kg/h 30000 kg/h 45000 kg/h 60000 kg/h
0.02 bar 0.10 bar 0.32 bar 0.65 bar 1.08 bar
20000 kg/h 35000 kg/h 50000 kg/h 65000 kg/h 80000 kg/h
0.09 bar 0.25 bar 0.46 bar 0.74 bar 1.06 bar
25000 kg/h 48750 kg/h 72500 kg/h 96250 kg/h 120000 kg/h
0.06 bar 0.20 bar 0.39 bar 0.64 bar 0.95 bar
30000 kg/h 72500 kg/h 115000 kg/h 157500 kg/h 200000 kg/h
0.03 bar 0.15 bar 0.34 bar 0.58 bar 0.89 bar
60000 kg/h 160000 kg/h 260000 kg/h 360000 kg/h 460000 kg/h
0.03 bar 0.14 bar 0.33 bar 0.58 bar 0.89 bar
150000 kg/h 287500 kg/h 425000 kg/h 562500 kg/h 700000 kg/h
0.04 bar 0.13 bar 0.25 bar 0.41 bar 0.60 bar
300000 kg/h 600000 kg/h 900000 kg/h 1200000 kg/h 1500000 kg/h
0.07 bar 0.25 bar 0.51 bar 0.84 bar 1.24 bar
400000 kg/h 850000 kg/h 1300000 kg/h 1750000 kg/h 2200000 kg/h
0.06 bar 0.23 bar 0.48 bar 0.81 bar 1.21 bar
Model
CMU-C 2.2 lbs/min 22.0 lbs/min
CMU-D 9.2 lbs/min 91.9 lbs/min
CMU-G 1102.3 lbs/min110.2 lbs/min
CMU-J 734.9 lbs/min
CMU-L 1102.3 lbs/min 7348.6 lbs/min
CMU-P 5511.5 lbs/min 25720.2 lbs/min
Min.
measuring range
Max.
measuring range
440.9 lbs/minCMU-E 44.1 lbs/min
2204.6 lbs/minCMU-H 220.5 lbs/min
2939.4 lbs/min
4409.2 lbs/minCMU-K 918.6 lbs/min
16901.8 lbs/minCMU-N 2204.6 lbs/min
2.2 lbs/min 5.5 lbs/min 11.0 lbs/min 16.5 lbs/min 22.0 lbs/min
Pressure loss [water (20°C), 1 mPas]
0.73 psi 3.61 psi 12.15 psi 24.70 psi 40.87 psi
9.2 lbs/min 23.0 lbs/min 45.9 lbs/min 68.9 lbs/min 91.9 lbs/min
1.02 psi 5.07 psi 17.05 psi 34.67 psi 57.35 psi
44.1 lbs/min 110.2 lbs/min 220.5 lbs/min 330.7 lbs/min 440.9 lbs/min
1.07 psi 5.32 psi 17.91 psi 36.41 psi 60.24 psi
110.2 lbs/min 275.6 lbs/min 551.1 lbs/min 826.7 lbs/min 1102.3 lbs/min
0.61 psi 3.02 psi 10.17 psi 20.68 psi 34.21 psi
220.5 lbs/min 551.1 lbs/min 1102.3 lbs/min 1653.4 lbs/min 2204.6 lbs/min
0.28 psi 1.39 psi 4.67 psi 9.49 psi 15.69 psi
734.9 lbs/min 1286.0 lbs/min 1837.2 lbs/min 2388.3 lbs/min 2939.4 lbs/min
1.36 psi 3.61 psi 6.74 psi 10.67 psi 15.35 psi
918.6 lbs/min 1791.2 lbs/min 2663.9 lbs/min 3536.5 lbs/min 4409.2 lbs/min
0.88 psi 2.84 psi 5.69 psi 9.34 psi 13.74 psi
1102.3 lbs/min 2663.9 lbs/min 4225.5 lbs/min 5787.0 lbs/min 7348.6 lbs/min
0.47 psi 2.18 psi 4.89 psi 8.48 psi 12.88 psi
2204.6 lbs/min 5878.9 lbs/min 9553.2 lbs/min 13227.5 lbs/min 16901.8 lbs/min
0.36 psi 2.03 psi 4.75 psi 8.39 psi 12.88 psi
5511.5 lbs/min 10563.6 lbs/min 15615.8 lbs/min 20668.0 lbs/min 25720.2 lbs/min
0.59 psi 1.83 psi 3.62 psi 5.92 psi 8.68 psi
CMU & CT OPERATING MANUAL
Page 17 of 112
A
3.4.6 Ambient temperature
40 °C to + 60 °C (-40 °F to 140 °F)

3.4.7 Ambient temperature range

40 °C to + 60 °C (-40 °F to 140 °F); a special cable is required for temperatures below 20 °C (-4 °F)
3.4.8 Storage temperature
25 °C to + 60 °C (-13 °F to 140 °F), 40 °C (-40°F) available as special version
3.4.9 Climatic category
In conformity with IEC 654-1. Unsheltered class D locations with direct open-air climate.
3.4.10 Ingress protection
Standard version: IP 66 (NEMA 6); special version IP 68 (NEMA 6P) DIN EN 60529, if suitable and tightly screwed down cable glands are used.
3.5 Operating conditions
3.5.1 Installation
The sensor is to be protected, wherever possible, against valves, manifolds and similar fittings that gen­erate turbulence. The sensor is to be installed in accordance with the following instructions.
Diagram showing flowmeter installation
BC C C
Flowmeter installation: A = sensor, B = valve, C = pipe clamps and supports
Under no circumstances is the sensor to be used to support a pipe.
C
Page 18 of 112
CMU & CT OPERATING MANUAL
Do not install the sensor in sus­pended pipes.
3.5.2 Installation positions
Standard installation position
Do not adjust the position of a pipe by pulling or grasping the sensor.
Installation position A
Installation position B
CMU & CT OPERATING MANUAL
Page 19 of 112
Installation position C

3.5.3 Assessment of installation position

Type of fluid Position Assessment
Pure liquids Standard installation
Self-draining flow tubes
position
Position A or B OK Position C Liquid residue remains in pipe Liquids with gas bub-
bles
Standard installation position
Self-draining flow tubes, gas bubbles do not accumulate in flowmeter
Position A Not recommended owing to gas bubble accumulation in
flowmeter
Position B Gas bubbles may accumulate in the presence of low flow
velocities
Position C No gas bubble accumulation in flowmeter, liquid residues
may remain in device after discharge
Liquids containing substances that could
Standard installation position
Self-draining flow tubes, no deposit formation
form deposits Position A OK
Position B Substances in the liquid could form deposits at low flow
velocities
Position C Not recommended owing to presence in flowmeter of
substances that could form deposits Liquids containing gas bubbles, as well
Standard installation position
Self-draining flow tubes, no accumulation of gases or
substances that could form deposits as gas bubbles con­taining substances that could form de­posits Position A Not recommended owing to gas bubble accumulation in
flowmeter Position B Gas bubbles or substances that could form deposits at
low flow velocities Position C Not recommended owing to presence in flowmeter of
substances that could form deposits Gases that do not form a condensate
Standard installation position,
Any of these installations positions can be used
Position A, B or C
Page 20 of 112
CMU & CT OPERATING MANUAL
Type of fluid Position Assessment
Gas, condensate­forming gas/liquid,
Standard installation position
Flow direction should be from top to bottom so that any
condensate that forms can flow out efficiently moisture Position A OK
Position B Condensate might form in flowmeter Position C Not recommended owing to condensate accumulation in
flowmeter Slurries Standard installation
Optimal installation position
position
Position A High density substances could accumulate in the flow-
meter Position B Gas bubbles could accumulate
Position C Gas bubbles or high density substances could accumu-
late in the flowmeter
3.5.4 Pressure surges
Pressure surges in a pipe could be provoked by a sudden decrease in flow caused by rapid closing of a valve or similar factors. This change in pressure can lead to underpressure downstream from a valve that has been closed rapidly, and to outgasing. If the valve is mounted directly on the inlet section of the flowmeter, a gas bubble can form in the flow tube that can cause a measuring signal disturbance that would shift the zero point of the output signal. In extreme cases, a pressure surge could cause mechani­cal damage to the sensors and/or flow tube.
Whenever possible, quick-closing valves should be mounted downstream from the sensor. If this is not feasible, such valves are to be mounted a minimum of 10 x DIA (Φ) from the nearest sensor. Alterna­tively, valve closing speed can be reduced.

3.5.5 Using the device with hazardous fluids

The sealing technology used in the standard CMU mass flowmeter renders the device unsuitable for use with hazardous fluids. Only sensors that meet the standards for safety instruments are suitable for use with hazardous fluids.
The pathway between the sensor and transmitter must be pressure-tight so as to prevent fluid from leak­ing out of a sensor in the event a sensor develops a defect.
In the case of welded components, a colored liquid penetration test should be performed on the welds, or one joint (only the first one) should be x-rayed. Alternatively, an internal pressure monitoring device can be used to detect any defect.
3.5.6 Vibration stability
The sensors are insensitive to vibration; vibration stability has been validated in accordance with DIN IEC 68-2-6, for up to 1 g at 10 to 150 Hz.
If pipe vibration is greater than 1 g in the 10-150 Hz range, an additional fastening is to be mounted as shown in the following drawings. This fastening will prevent vibration from affecting the device’s mechani­cal configuration and/or measurement readings. The following drawings are valid for a sensor with a nominal size of approximately 2 inches [DN 040]. Installation is to be realized as shown in this drawing.
CMU & CT OPERATING MANUAL
Page 21 of 112
Installation using wall supports
Foot-mounted installation
Page 22 of 112
CMU & CT OPERATING MANUAL
3.6 Process conditions
3.6.1 Process temperature
40 °C to + 260 °C (-40 °F to 500 °F); rating plate range must be observed
3.6.2 Physical state
Liquid product (maximum density 2 kg/l) Gaseous product (minimum density 0.002 kg/l in operating state)
3.6.3 Viscosity
0.3 up to 50,000 mPas (0.3 to 50,000 cP)
3.6.4 Gas content
The use of products containing gas is not allowed for custody transfer operations. In other applications, the presence of gas will increase false readings. In order for the readings of products containing gas to be valid, small gas bubbles must be homogeneously distributed in the fluid. Large gas bubbles will auto­matically provoke extremely false readings and will shift the zero point. Thus, the extent to which readings are false is determined by the process conditions. A rule of thumb in this regard is as follows: A 1 % gas component will increase false readings by 1 %. The gas component is not to exceed 5 %.

3.6.5 Process temperature range

+ 260 °C (500 °F)

3.6.6 Process pressure range

According to PN16 pressure rating: 232 psig [16 bar] and PN40: 580 psig [40 bar]
3.6.7 Outlet pressure
Outlet pressure must be greater than the vapor pressure Ps (static pressure) of the measured product.

3.7 Connection to the transmitter

3.7.1 Integral mount configuration

When the transmitter is mounted directly on the sensor, no cable connection between the two compo­nents is needed. This connection is integrated at the factory.

3.7.2 Remote mount configuration

If the transmitter is dards are to be adhered to. The maximum cable length is 300 m (1000ft). See Section 11.5.2 Wiring dia­gram on page 46 for information regarding the connection and cable specifications.
not mounted directly on the sensor, installation regulations and applicable legal stan-
CMU & CT OPERATING MANUAL
Page 23 of 112
3.8 Construction details

3.8.1 Dimensions and weight

Standard versions:
End
Model
CMU-C SW10 --
connection inch [mm]
connection inch [mm]
End
SW12 --
End
connection inch [mm]
DN10 [360]14.2
A
End
connection inch [mm]
End
connection inch
½" NPT (f)¼" NPT (f) [300]11.8
11.8
CMU-D
CMU-E ½" NPT (f) --
CMU-G - --
CMU-H
CMU-J
CMU-K
CMU-L
CMU-N
CMU-P
CMU-Q
CMU-R
SW12 -
---
---
DN40 [1,018]40.1
DN50 [1,176]46.3
DN80 [1,370]53.9
DN100 [1,726]68.0
DN150 [2,184]86.0
DN250 [2,913]114.7
-
DN15 [515]20.3
DN25 [632]24.9
DN40
DN50 [1,024]40.3
DN80
DN100 [1,358]53.5
DN150
DN200 [2,198]86.5
DN300
[770]30.3
[1,196]47.1
[1,732]68.2
[2,284]89.9 DN300 [1,900]DN200 [2,268]89.3 DN250
[2,925]115.2
DN15 [396]15.615.4 [390]DN10
DN25 [520]20.5
DN40 25.3 [642] ¾" 150lb 25.9 [657] 1" 150lb 26.1
DN50 [776]30.6
DN80 [1,044]41.1
DN100 [1,184]46.6
DN150 [1,090]42.9
DN200
DN300 [1,864]73.4
DN350 [2,933]115.5
[1,448]57.0
74.8
½" NPT (f) [300]11.8
- - - 1½" 150lb 31.7
1½" 150lb 41.3 [1,050] 2" 150lb 41.5
2" 150lb 47.5 [1,207] 3" 150lb 48.0
3" 150lb 54.6 [1,388] 4" 150lb 55.1
4" 150lb 69.7 [1,770] 6" 150lb 70.7
6" 150lb 88.6 [2,250] 8" 150lb 89.4
8" 150lb 92.4 [2,348] 10" 150lb 92.4
10" 150lb 117.2 [2,976] 12" 150lb 117.9
½" 150lb 16.4
½" 150lb 21.1
B
Integral mount transmitter Remote mount transmitter
-40°C - 260°C
(-40°F to 500°F)
Model
-40°C - 100°C
(-40°F to 212°F)
-40°C - 150°C
(-40°F to 302°F)
-40°C - 100°C
(-40°F to 212°F)
-40°C - 180°C
(-40°F to 356°F)
inch [mm] inch [mm] inch [mm] inch [mm] inch [mm] inch [mm] inch [mm] CMU-C 12.9 [328] 16.9 [430] 8.9 [225] 12.9 [327] 16.8 [427] 3.3 [85] 1.6 [40] CMU-D 13.5 [343] 17.5 [445] 9.4 [240] 13.5 [342] 17.4 [442] 3.9 [100] 1.6 [40] CMU-E 15.6 [395] 19.6 [497] 11.5 [292] 15.5 [394] 19.4 [494] 5.8 [148] 1.9 [48]
CMU-G 18.1 [460] 22.1 [562] 14.1 [357] 18.1 [459] 22.0 [559] 7.9 [200] 2.9 [74]
CMU-H 20.8 [528] 24.8 [630] 16.7 [425] 20.7 [527] 24.7 [627] 10.0 [255] 4.0 [101]
CMU-J 39.8 [1,010] 43.8 [1,112] 35.7 [907] 39.7 [1,009] 43.7 [1,109] 24.2 [615] 9.1 [230]
CMU-K 47.6 [1,210] 51.7 [1,312] 43.6 [1,107] 47.6 [1,209] 51.5 [1,309] 31.5 [800] 9.8 [250]
CMU-L 48.4 [1,230] 52.4 [1,332] 44.4 [1,127] 48.4 [1,229] 52.3 [1,329] 32.1 [815] 10.6 [270] CMU-N 61.4 [1,560] 65.4 [1,662] 57.4 [1,457] 61.4 [1,559] 65.3 [1,659] 42.1 [1,070] 15.0 [380] CMU-P 67.7 [1,720] 71.7 [1,822] 63.7 [1,617] 67.7 [1,719] 71.6 [1,819] 47.6 [1,210] 15.7 [400]
CMU-Q 73.2 [1,860] 77.2 [1,962] 69.2 [1,757] 73.2 [1,859] 77.1 [1,959] 51.2 [1,300] 21.7 [550]
CMU-R 73.4 [1,865] 77.4 [1,967] 69.4 [1,762] 73.4 [1,864] 77.3 [1,964] 55.1 [1,400] 20.1 [510]
C
G
Page 24 of 112
CMU & CT OPERATING MANUAL
Weight:
Weight
Sensor Transmitter
Model
CMU-C 3.5 [7.7] CMU-D 4 [8.8] CMU-E 7 [15.4] CMU-G 15 [33.1] CMU-H 29 [63.9]
CMU-J 140 [308.6] 4.5 [9.9]
CMU-K 200 [440.9]
CMU-L 250 [551.2] CMU-N 470 [1036.2] CMU-P 750 [1653.5] CMU-Q 850 [1873.9] CMU-R 900 [1984.1]
Heated versions:
kg [lbs] kg [lbs]
KLM
Model
inch [mm] inch [mm] inch [mm]
CMU-J 24.0 [610] 26.8 [680] 9.4 [240] CMU-K 31.5 [800] 34.4 [875] 9.8 [250]
CMU-L 23.6 [600] 30.9 [785] 10.6 [270] CMU-N 42.5 [1080] 46.9 [1190] 12.8 [325] CMU-P 47.2 [1200] 52.4 [1330] 13.2 [335]
CMU & CT OPERATING MANUAL Part No 06EN003380 rev 1.6
Page 25 of 112

3.8.2 Dimension drawing for the types CMU-C to CMU-H

3.8.2.1 Standard version dimension drawing
Integral mount configuration that is suitable for process temperatures up to 100 °C (212°F):
For all the dimensions and weight, see Section 3.8.1 Dimensions and weight on page 24.
3.8.2.2 Integral mount version up to 150 °C (302 °F)
Integral mount configuration that is suitable for process temperatures up to 150 °C (302°F):
For all the dimensions and weights, see Section 3.8.1 Dimensions and weight on page 24.
Page 26 of 112
CMU & CT OPERATING MANUAL
3.8.2.3 Remote mount version dimension drawing
Remote mount configuration with junction box that is suitable for process temperatures up to 100 °C (212 °F):
For all the dimensions and weights, see Section 3.8.1 Dimensions and weight on page 24.
3.8.2.4 Remote mount version dimension drawing up to 180 °C (356 °F)
Remote mount configuration with junction box that is suitable for process temperatures up to 180 °C (356 °F):
For all the dimensions and weights, see Section 3.8.1 Dimensions and weight on page 24.
CMU & CT OPERATING MANUAL Part No 06EN003380 rev 1.6
Page 27 of 112
3.8.2.5 Remote mount version dimension drawing up to 260 °C (500 °F)
Remote mount configuration with junction box that is suitable for process temperatures up to 260 °C (500 °F):
For all the dimensions and weights, see Section 3.8.1 Dimensions and weight on page 24.

3.8.3 Dimension drawing for the types CMU-J to CMU-R

3.8.3.1 Standard version dimension drawing
Integral mount configuration that is suitable for process temperatures up to 100 °C (212 °C):
For all the dimensions and weights, see Section 3.8.1 Dimensions and weight on page 24.
Page 28 of 112
CMU & CT OPERATING MANUAL
3.8.3.2 Integral mount configuration up to 180 °C (356°F)
Integral mount configuration that is suitable for process temperatures up to 180 °C (356 °F):
For the dimensions and weights, see Section 3.8.1 Dimensions and weight on page 24.
3.8.3.3 Remote mount version dimension drawing
Remote mount configuration (with junction box) that is suitable for process temperature up to 100 °C (212 °F):
For the dimensions and weights, see Section 3.8.1 Dimensions and weight on page 24.
CMU & CT OPERATING MANUAL Part No 06EN003380 rev 1.6
Page 29 of 112
3.8.3.4 Remote mount version dimension drawing up to 180 °C (356°F)
Remote mount configuration (with junction box) that is suitable for process temperatures up to 180 °C (356°F):
For the dimensions and weights, see Section 3.8.1 Dimensions and weight on page 24.
3.8.3.5 Remote mount version up to 260 °C (500°F)
Remote mount configuration (with junction box) that is suitable for process temperatures up 260 °C (500 °F):
For all the dimensions and weights, see Section 3.8.1 Dimensions and weight on page 24
Page 30 of 112
CMU & CT OPERATING MANUAL
Loading...
+ 82 hidden pages