Fisher Biotech FB300-2 Electrophoresis Power Supply
Table of Contents
Safety Considerations 2
Introduction 3
Unpacking the Power Supply 3
Specifications 4
Getting Started 5
Using the Power Supply 5
Constant Voltage Operations 6
Constant Current Operations 6
Timed Operations 6
Automatic Power-fail Restart 7
Cleaning 7
Troubleshooting and Error Indications 8
FB300-1 Warranty Statement 9
Compliance 9
Replacement Parts 10
FB300-2 MANUAL Page 1 7608001 Rev A 8/02
Safety Considerations
Read and understand this manual completely before attempting to set up or
use this instrument.
This equipment has been designed and tested to conform to IEC1010-1 safety standards, as
applicable to laboratory instrumentation. This applies only to the FB300 when used as specified
in the documentation, in its intended applications, with Fisher Biotech approved electrophoresis
apparatus only. Usage in any other manor may not provide similar performance or safety
protection.
This equipment is provided with a 3-conductor, grounded AC line cord. The protective earth
ground is necessary for safe operation. Do not use any other AC line cord with this instrument.
The FB300 is a high voltage power supply capable of generating dangerous levels of voltage and
current during operation. Exercise caution when working around and with the electrical
connections of this equipment. Always check electrical connectors, wires, and associated
apparatus for any signs of wear or damage before using with this equipment. Be sure to use only
electrophoresis equipment that is suitably rated for the voltage and current capabilities of the
FB300 power supply.
The output of the FB300 power supply is intended for connection to electrically isolated
electrophoresis apparatus only. Use only with electrically isolated electrophoresis apparatus with
minimum isolation of 600V. Do not connect any terminal of the FB300 output to earth ground.
This may impair the safety protection provided by the equipment, or cause equipment damage.
The high voltage output of the FB300 power supply takes some amount of time to decay when
unloaded or lightly loaded. Wait a minimum of 60 seconds after stopping a run before touching
the power supply leads.
This equipment has a protective ground leakage current less than 0.5 mA using test methods
defined in CSA1010 and CSA151.
This equipment is for indoor use only.
FB300-2 MANUAL Page 2 7608001 Rev A 8/02
Introduction
Thank you for selecting a Fisher Scientific FB300 Electrophoresis power supply. This manual
describes the operation of the FB300. The power supply that you have purchased is the most
productive and easy-to-use unit available anywhere. This manual should answer any questions
that might arise in operating your power supply; however, don't hesitate to call our Fisher Lab
Equipment Technical Support Hotline at 1-800-943-2006 or 1-800-926-0505 if you need any
assistance.
The FB300 power supply is designed to provide constant voltage or constant current output to
apparatus used in electrophoresis applications. One to three sets of electrophoresis cells can be
connected in parallel and run simultaneously. The FB300 can deliver up to 75W of total output
power. When operating in constant voltage or constant current mode, the power supply
automatically limits the other parameter to either the power supply maximum, or a lower limit if
set by the user. If this non-constant limit is reached, the power supply will automatically switch
control modes, from constant voltage to constant current, or vice versa. In this way, the FB300
protects your electrophoresis cells from damaging over power conditions. The FB300 power
supply also provides for timed operation in either voltage or current modes, and allows an
automatic completion in the event of a power loss if enabled by the user.
• 10-300V, adjustable in 1 volt steps
• 4-400mA, adjustable in 1mA steps
• 75W maximum output
• Automatic control mode crossover
• 0-999 minute timed run
• Automatic restart if loss of AC power (if enabled)
• Bright 3-digit display
Unpacking the Power Supply
When unpacking your FB300 power supply, be sure you have received the following items.
• FB300 unit
• AC line cord
• This manual
Inspect your equipment and packaging material for signs of damage. Damage to the shipping
container may indicate rough handling which could cause internal damage to the power supply. If
you suspect shipping damage to the power supply, contact your carrier for instructions on filing a
claim. If you are missing any of the above items, contact your supplier for instructions.
10-300VDC, 75W Max
4-400mA, 75W Max
Ripple: ± 1%
Drift: ± 1%, after 30-minute warm-up
FB300-2 MANUAL Page 4 7608001 Rev A 8/02
Getting Started
Select a location that allows for 3" clearance behind the power supply, and comfortable reach of
the front panel controls and cell connections. Do not block the vented area of the case - on the
front bottom of the unit, or the fan area at the rear. Connect the power supply to a 3-prong
grounded AC outlet, using the AC cord provided with the unit only
electrophoresis apparatus to the power supply, making sure to match the red positive lead to the
red positive jack, and the black negative lead to the black negative jack. Similarly, connect one or
two more electrophoresis apparatus to the power supply if you plan to run more than one in
parallel. Power the unit on using the AC power switch located at the rear of the unit, next to the
AC line cord entry.
. Connect the
black negative jacks 3 digit LED display
up/down arrows
select key (Set)
constant key (K)
run key
Using the Power Supply
Press the gray key on the front center of the unit to enable the control logic. The 3-digit display
will illuminate and show the setpoint value of the last saved run setup. The FB300 preserves the
run settings each time you start a run. One set of conditions is saved for each mode, constant
voltage or constant current. The setpoint value, limit parameter, time duration, and power-fail
restart setting is saved. This allows convenient setup for repetitive runs.
To change the control mode, press the constant key. This key toggles control from constant
voltage to constant current. Each time, the FB300 will display the last saved setpoint value for
that particular mode.
After selecting a control mode, either constant voltage or current, and setting the limit and timed
run parameters if desired (see Constant Voltage/Current Operations) press the Run key to energize
the power supply output. The power supply output will ramp up to the appropriate setpoint, while
not allowing the limit parameter to be exceeded. If the limit parameter should be exceeded, then
the power supply will crossover control modes, making the limit parameter the new control
setpoint. The front panel constant mode LED indicator will change to indicate that a crossover has
occurred.
gray ON/OFF key red positive jacks
Figure 1. Front Panel Controls
FB300-2 MANUAL Page 5 7608001 Rev A 8/02
To stop a run in progress, press the Run key or the gray ON/OFF key. The display will change to
"OFF" signifying that the output is no longer energized. Pressing run again resumes the run from
the previous point (i.e. for a timed run). Pressing any other key will return to setup mode in either
constant voltage or constant current mode as appropriate.
During a run, the display will automatically cycle between voltage, current, and time displays.
Voltage and current is displayed as actually measured by the FB300. When displaying the
controlled setpoint, i.e. voltage for constant voltage control, the arrow keys can be used to modify
the setpoint. The limit parameter cannot be changed during a run. The time parameter displays
elapsed time for an untimed run, and time left for a timed run.
Constant Voltage Operations
Press the constant key (labeled K) until the LED indicating "Volts" is illuminated. Each time the
constant key is pressed, the FB300 will display the last saved setpoint value for that particular
mode. Use the UP and DOWN arrow keys to change the setpoint to the desired value (10-300
volts). The FB300 will not allow setting a voltage setpoint outside of the allowed range. Press
and hold either arrow key to quickly change the setpoint.
The limit parameter, in this case current, is normally set to the power supply maximum value of
400 mA. To set a lower limit value, press the Set key until the LED indicating "mAmps" is
illuminated. Use the UP and DOWN arrow keys to change the limit to the desired value (4-400
mA). The FB300 will not allow setting a current limit outside of the allowed range.
In addition, the FB300 will never exceed the maximum power output specification of 75W. When
starting a run, the power supply calculates a maximum limit parameter, above which the power
supply maximum power output rating would be exceeded. If this value is lower than the limit
setpoint entered, the FB300 will use this lower limit parameter.
Constant Current Operations
Press the constant key (labeled K) until the LED indicating "Amps" is illuminated. Each time the
constant key is pressed, the FB300 will display the last saved setpoint value for that particular
mode. Use the UP and DOWN arrow keys to change the setpoint to the desired value (4-400 mA).
The FB300 will not allow setting a current setpoint outside of the allowed range. Press and hold
either arrow key to quickly change the setpoint.
The limit parameter, in this case voltage, is normally set to the power supply maximum value of
300 volts. To set a lower limit value, press the Set key until the LED indicating "Volts" is
illuminated. Use the UP and DOWN arrow keys to change the limit to the desired value (10-300
volts). The FB300 will not allow setting a voltage limit outside of the allowed range.
In addition, the FB300 will never exceed the maximum power output specification of 75W. When
starting a run, the power supply calculates a maximum limit parameter, above which the power
supply maximum power output rating would be exceeded. If this value is lower than the limit
setpoint entered, the FB300 will use this lower limit parameter.
Timed Operations
It is possible to enter an amount of time for the power supply to provide power, after which it will
automatically shut off. Use this feature for timed runs.
Press the Set key until the LED indicating "Time" is illuminated. Use the UP and DOWN arrow
keys to change the time duration to the desired value (0-999 minutes). A time duration of zero
effectively disables timed run mode.
When running a timed run, the time parameter displays the time remaining in the run. When
running a non-timed run, the time parameter displays the elapsed time during the run
FB300-2 MANUAL Page 6 7608001 Rev A 8/02
Automatic Power-fail Restart
Automatic power-fail restart capability allows a timed run which is interrupted by loss of AC
power to be restarted automatically, so that the total time programmed for the time parameter will
be met. NOTE: Loss of AC power includes switching off the unit using the power switch located
at the rear of the unit. When performing timed runs with power-fail restart enabled, always use
the front panel controls to stop a run in progress.
To enable power-fail restart, press and hold the Set key when changing to the time parameter
(from the current parameter). While continuing to hold the Set key, simultaneously press and
release the UP arrow key. The display will show "PF" momentarily to indicate power-fail restart
is active. Whenever power-fail restart is enabled, "PF" is displayed briefly as the time parameter
is selected by pressing the Set key. The power-fail restart feature can be enabled either before or
after changing the time duration to the desired value (0-999 minutes). Setting the time duration to
zero will automatically disable power-fail restart (if you have enabled power-fail restart before
setting a time value, and want to disable it, then simply increment and then decrement the time
parameter back to zero to disable power-fail restart).
When AC power is restored during a timed run in which power-fail restart is enabled, the display
will show "PF" for approximately 10 seconds indicating power-fail restart pending. During this
time the output is not energized, to allow safely stopping the run (by pressing RUN). After the
power-fail restart pending delay is completed, the power supply output will ramp up to the
setpoint value and the time will continue from that point. Any number of power interruptions can
occur during the completion of a single run.
Cleaning
Before cleaning the unit, be sure to always turn off AC power using the switch at the rear of the
unit, and then disconnect the AC line cord. The front of the unit is sealed and can be wiped clean
with any mild detergent solution. Avoid harsh cleaners or agents as they may deteriorate the
surface of the tactile membrane keys.
FB300-2 MANUAL Page 7 7608001 Rev A 8/02
Troubleshooting and Error Indications
The FB300 detects and reports several events and conditions that are considered errors. The
FB300 will stop any run in process and display "EXX" where XX is one of the below listed errors.
When the FB300 is displaying an error indication, press the gray ON/OFF key to clear the error
and return to setup mode, or press the Run key to resume the run (after correcting the cause of the
error). Note that loss of AC power during a run is considered an error condition. Always stop a
run before turning off AC power to the FB300.
E'XX' Condition and possible remedy
00
01
03
04
05
06-99
Before servicing the unit, be sure to always turn off AC power using the switch at the rear of the
unit, and then disconnecting the AC line cord. The FB300 power supply requires no periodic
servicing and should provide years of trouble free operation. Should you need to replace the fuses
proceed as follows:
Minimum load current detected. The output is not connected, or the
electrophoresis apparatus is not set up properly. Check your setup and
connections.
Maximum load current exceeded. The output is short circuit, or the
electrophoresis apparatus is not set up properly. Check your setup and
connections
Loss of AC power; PF enabled and completed. The FB300 detected loss of
AC power, however the run was completed as programmed.
Loss of AC power; PF disabled and not completed. The FB300 detected loss
of AC power, the run was not completed since PF not enabled.
Loss of AC power; run not timed. The FB300 detected loss of AC power.
The FB300 detected an internal error condition. Contact technical support for
instructions.
Replacing a fuse
CAUTION: The FB300 uses double pole neutral fusing.
Turn off AC power using the switch at the rear of the unit, and then disconnecting the AC
line cord. Remove the fuse holder assembly using a small flat blade screwdriver.
Always replace both fuses with appropriate replacement fuses: 1.5A, 250V, 5X20mm,
type T fuse (T1.5A,250V) (Fisher catalog number FB-FUSE-1).
FB300-2 MANUAL Page 8 7608001 Rev A 8/02
FB300-2 Warranty Statement
The Fisher Scientific Company ("Fisher") warrants to the direct purchaser that the FB300 will be
free from defects in material or workmanship for a specified warranty period. During that period,
Fisher will repair or replace the product or provide credit, at its sole option, upon prompt
notification and compliance with its instructions. For FB300-2 power supplies that specified
period is 48 months from manufacturing date.
Unless otherwise agreed, the warranty is limited to the country in which the product is sold.
No Fisher employee, agent or representative has the authority to bind Fisher to any oral
representation or warranty concerning any product sold. Any oral representation or warranty
made prior to purchase of any product and not set forth in writing and signed by a duly authorized
officer of Fisher shall not be enforceable by the purchaser.
FISHER EXPRESSLY DISCLAIMS ALL OTHER WARRANTIES, EXPRESS OR IMPLIED,
INCLUDING THE IMPLIED WARRANTY OF MERCHANTABILITY AND FITNESS FOR A
PARTICULAR PURPOSE.
Fisher's sole responsibility and the purchaser's exclusive remedy for any claim arising out of the
purchase of any product listed above is repair, replacement or credit as described above, where
applicable. In no event: 1) shall the cost of the exclusive remedy exceed the purchase price: 2)
shall Fisher be liable for any special, indirect, incidental, consequential, or exemplary damages,
howsoever arising, even if Fisher has been advised of the possibility of such damages.
Each article that Fisher furnishes will conform to the written specifications given in this manual,
or those of a further improved model. Changes are made often to the information in the manual
and will be incorporated into future editions.
Compliance
Pertaining to models: FB300-1
FB300-2
EMC - This equipment when used in accordance with the instruction manual conforms to EMC
Directive 89/336/EEC and EMC Standard EN61326, as applicable to electrical equipment for
measurement control and laboratory use.
IEC - This equipment has been designed and tested to conform to Low Voltage Directive
73/23/EEC and IEC 1010-1 Safety Standards, as applicable to electrical equipment for
measurement control and laboratory use.
FB300-2 MANUAL Page 9 7608001 Rev A 8/02
Accessories
Adapter for Cells FBAD-1
Under Counter Bracket FBUB300
Replacement Parts
For replacement parts contact Fisher Scientific at:
Fisher Scientific
2000 Park Lane
Pittsburgh, PA 15275
1-800-766-7000
AC line cord, UK Plug FB-CORD-2
AC line cord, Cont. Euro Plug FB-CORD-3
Fuse, T1.5A,250V FB-FUSE-1
Technical Support : 1-800-943-2006 or 1-800-926-0505
FB300-2 MANUAL Page 10 7608001 Rev A 8/02
Alimentation électrique pour électrophorèse Fisher
Biotech FB300-2
Table des matières
Considérations de sécurité 2
Introduction 3
Déballage de l'alimentation électrique 3
Spécifications 4
Mise en route 5
Utilisation de l'alimentation électrique 5
Opérations à tension constante 6
Opérations à intensité constante 6
Opérations à durée déterminée 7
Redémarrage automatique en cas de coupure de courant 7
Nettoyage 7
Dépannage et indications d'erreur 8
Garantie du FB300 9
Conformité 9
Pièces de rechange 10
FB300-2 MANUAL Page 1 7608001 Rev A 8/02
Considérations de sécurité
Lisez et comprenez complètement ce manuel avant d'entreprendre d'installer
ou d'utiliser cet instrument.
Cet appareil a été conçu et testé pour être conforme aux normes de sécurité IEC1010-1, telles
qu'elles s'appliquent aux instruments de laboratoire. Ceci s'applique uniquement au FB300
lorsqu'il est utilisé de la manière spécifiée dans la documentation, dans le cadre des applications
auxquelles il est destiné, uniquement avec un appareil d'électrophorèse agréé par Fisher Biotech.
Une utilisation dans d'autres circonstances peut ne pas donner les mêmes performances ou la
même protection de sécurité.
Cet appareil est fourni avec un cordon d'alimentation en CA à 3 fils offrant une connexion à la
terre. Le fil de terre de protection est nécessaire à la sécurité lors du fonctionnement. N'utilisez
aucun autre cordon d'alimentation en CA avec cet instrument.
Le FB300 est une alimentation électrique à haute tension capable de générer des tensions et des
intensités dangereuses lors de son fonctionnement. Soyez prudent lorsque vous travaillez à
proximité et avec les connexions électriques de cet appareil. Vérifiez toujours les connecteurs
électriques, les fils et les appareils associés pour vous assurer qu'ils ne sont ni usés ni endommagés
avant de les utiliser avec cet appareil. Veillez à utiliser uniquement un appareil d'électrophorèse
qui corresponde aux valeurs nominales de tension et d'intensité de l'alimentation électrique FB300.
La sortie de l'alimentation électrique FB300 est prévue uniquement pour une connexion à un
appareil d'électrophorèse électriquement isolé. Utilisez-la uniquement avec un appareil
d'électrophorèse électriquement isolé muni d'une isolation minimum de 600 V. Ne connectez
aucune borne de sortie du FB300 à la terre. Ceci pourrait affecter la protection de sécurité fournie
par l'appareil ou endommager l'appareil.
La sortie à haute tension de l'alimentation électrique FB300 met un certain temps à décliner
lorsqu'elle est déchargée ou légèrement chargée. Attendez au minimum 60 secondes après l'arrêt
d'une exécution avant de toucher les fils de l'alimentation électrique.
Cet appareil a un courant de fuite à la terre de 2 mA. Bien que cela soit nettement inférieur à la
limite de 3,5 mA applicable aux instruments de laboratoire, certaines applications et certains
endroits exigent un courant de fuite inférieur à 0,5 mA. Vérifiez les besoins spécifiques de votre
application avant d'utiliser cet appareil.
Cet appareil ne doit pas être utilisé à l'extérieur.
FB300-2 MANUAL Page 2 7608001 Rev A 8/02