This electrophoresis power supply is a high technology instrument available in several versions. As it is capable
of giving dangerous voltage levels by which high power is involved, we suggest that you take a few moments to
read this manual thoroughly. Although this instrument is equipped with all necessary safety features against
abuse and other accidental failures, caution should be exercised when working with high voltage equipment.
• Avoid to touch the outlets with any conducting object and make sure there is
a second person present for your safety in case of any severe electric shock.
• Never touch any part of the assembly (power supply, leads or tank) before hav
ing switched off.
• Never manipulate with wet hands.
• Do not ground any of the outputs or the buffer in the tank.
• Connect the outlets only to an insulated electrophoresis tank with safety
cover.
• Never make any other connections, such as e.g. putting several power supplies
in series or in parallel.
• In order to prevent electric shock, never open the back plate nor remove the
cover.
• Do not expose the unit to rain or any other liquid.
• Do not spill liquid or insert metal objects inside the unit.
• Take care so that the power supply is not dropped to avoid damaging the cabi
net which defeats safeguards or injuring yourself.
• If the unit has been dropped or the cabinet has been damaged, unplug it
and have it checked by an authorised service technician to restore the safe
guards.
• The fact that the unit operates satisfactorily does not imply that the unit is
properly earthed or that it is completely safe. If in any doubt about the effec
tive earthing of the unit, contact a qualified electrician.
• Never block the ventilation holes or place the unit in any enclosure unless
proper ventilation is provided.
• Never place the unit near or over a radiator, heat register or stove.
• Avoid locations where the instrument is exposed directly to the sun light.
This instrument is manufactured with the latest technology and needs no particular maintenance. Consort certifies that this instrument was thoroughly inspected
and tested at the factory prior to shipment and found to meet all requirements
defined by contract under which it is furnished. However, dimensions and other
physical characteristics may differ.
The normal operating temperature should be between 0° and 40°C. Never use
the instrument in a room with high humidity (>95 %) or at very low temperatures
(condensation water!).
Connect the instrument only to an earthed power line. The required power
source is indicated on the label at the back of the instrument. Do not cut and
splice the power cord. When removing the power cord from the wall outlet, be
sure to unplug by holding the plug attachment and not by pulling the cord. Do not
hold the plug by wet hand.
This instrument (excluding all accessories) is warranted against defective
material and workmanship for a period of thirty-six (36) months from the date of
shipment ex factory. Consort will repair all defective equipment returned to it
during the warranty period without charge, provided the equipment has been
used under normal laboratory conditions and in accordance with the operating
limitations and maintenance procedures in this instruction manual and when not
having been subject to accident, alteration, misuse or abuse. A return
authorisation must be obtained before returning any product for warranty repair
on a freight prepaid basis!
BIOTEC-FISCHER or Consort is not liable for consequential damages arising out
of the use or handling of its products.
Introduction
Contact
Warranty
In the event of this instrument being returned for servicing, the owner is
requested to remove the power supply lead and NOT to send the following items
unless they are suspect:
Manual
Cables
Accessories
If serious malfunctioning occurs, stop using the unit immediately and consult your
local dealer.
éê = Button for entering a value or for selecting a function.
SET/ENTER
RUN/STOP
= Button for programming the desired parameters.
= Button to start or stop an experiment.
Servicing
Keyboard
SET-UP
1. 1. Select SET-UP and press SET.
2. Select LANGUAGE and press SET.
3. Select the desired language and press SET.
4. Select CONTRAST and press SET.
5. Adjust the contrast of the display and press SET.
6. Select ALARM and press SET.
7. Decide whether or not the output voltage should shut down at a too
low output current and press SET.
8. Select POWER FAIL and press SET.
9. Decide whether or not the instrument should proceed with the
experiment for the remaining time after a power failure and press
SET.
10. Select STAND-BY and press SET to return.
• Select INFO and press SET to see general information about the electrophoresis power supply.
---- STAND-BY ---SET-UP
---- SET-UP ---LANGUAGE
CHOOSE LANGUAGE:
ENGLISH
---- SET-UP ----
CONTRAST
---- SET-UP ----
ALARM
CURRENT<1mA
ALARM: YES
---- SET-UP ----
POWER FAIL
POWER FAIL
DETECTION: YES
---- SET-UP ----
STAND-BY
---- SET-UP ----
INFO
Method programming permits to store up to 9 different methods, each with
9 steps, of frequently used parameters in the non-volatile memory for
future recall. Each step is able to recall a next one, providing a flexible
multiple step function for special techniques. The method mode also
permits to program a linear voltage gradient for any step provided the
limiting current or power is not attained. Parameters can be changed
temporarily without interrupting the run.
1. Verify if the instrument is switched off.
2. Connect the electrophoresis tank to the instrument.
3. Switch the instrument on.
4. Select the METHOD mode and press SET.
5. Select the desired method and press SET.
6. Select the desired step and press SET.
7. Decide whether or not a voltage gradient has to be performed and
press SET.
8. Enter the desired voltage limit and press SET (= the voltage end
point in case of a gradient step).
9. Enter the desired current limit and press SET (this limit is automati
cally set to maximum and cannot be changed in case of a gradient
step).
10. Enter the desired power limit and press SET (this limit is automati
cally set to maximum and cannot be changed in case of a gradient
step).
11. Enter the desired time-units (h of kVh) and press SET.
12. Enter the desired time and press SET.
13. Decide if the experiment should automatically proceed with a next
step (resuming from point 6) or stop after this step. Enter the
desired procedure and press SET.
14. Press RUN to start the experiment.
• At any time, you can return in the stand-by mode by holding SET
pressed during 2 seconds.
• At any time, you can avoid programming unnecessary parameters.
To do so, simply press RUN instead of SET after entering the last
relevant value and the instrument will immediately start the exper
iment keeping the other parameters to their previous values.
• To ensure that the desired parameter is kept constant, we suggest
to pre-set all other parameters to maximum.
• A blinking unit of the controlling parameter appears when the regu
lation is complete.
• The timer will always count down.
If the timer is still running after switching on, the power supply will
automatically proceed with the experiment for the remaining time.
Press STOP if you want to go in stand-by during the run
Method Programming
STAND-BY
METHOD
ENTER 1 ∫ 1
METHOD: 1
ENTER 1 ∫ 1
STEP: 1
ENTER 1 ∫ 1
GRADIENT: NO
ENTER 1 ∫ 1
VOLTAGE: 400V
ENTER 1 ∫ 1
CURRENT 300mA
ENTER 1 ∫ 1
POWER 50W
ENTER 1 ∫ 1
TIME UNIT h
ENTER 1 ∫ 1
TIME 99:59h
ENTER 1 ∫ 1
END METHOD: YES
V
time
STEP 2 STEP 3 STEP 4 STEP 5
RUN 1 ∫ 1 99:59h
400V 300mA 50W
Manual Programming
Manual programming permits to set voltage, current, power and time limits
for a simple routine electrophoresis run. Parameters can be changed
temporarily without interrupting the run.
1. Verify if the instrument is switched off.
2. Connect the electrophoresis tank to the instrument.
3. Switch the instrument on.
4. Select the MANUAL mode and press SET.
5. Enter the desired voltage limit and press SET.
6. Enter the desired current limit and press SET.
7. Enter the desired power limit and press SET.
8. Enter the desired time-units (hours or kilovolthours) and press SET.
9. Enter the desired time and press SET (pre-set to [--:--] if no timing
is required).
10. Press RUN to start the experiment.
• At any time, you can return in the stand-by mode by holding SET
pressed during 2 seconds,
• At any time, you can avoid programming unnecessary parameters.
To do so, simply press RUN instead of SET after entering the last
relevant value and the instrument will immediately start the exper
iment keeping the other parameters to their previous values.
• To ensure that the desired parameter is kept constant, we suggest
to pre-set all other parameters to maximum.
• A blinking unit of the controlling parameter appears when the regu
lation is complete.
• The timer will always count down. However, when no time limit has
been programmed it will count up.
• If the timer is still running after switching on, the power supply will
automatically proceed with the experiment for the remaining time.
Press STOP if you want to go in stand-by during the run.
only the parameters or the running step can be changed. A gradient step
can never be changed.
1 . Press SET to change temporarily the parameters without interrupting
the run.
2. Change the actual voltage limit and press SET.
3. Change the actual current limit and press SET.
4. Change the actual power limit, and press SET.
5. Change the actual time and press SET.
6. The power supply will now continue the run.
• Temporarily changed parameters are not been kept in memory!
• The instrument will automatically return to the run mode when no button
has been pressed during 10 seconds.
---- STAND-BY ---MANUAL
ENTER
VOLTAGE 400V
ENTER
CURRENT 300mA
ENTER
POWER 50W
ENTER
TIME UNIT h
ENTER
TIME 99:59h
RUN
400V 300mA 50W
Temporarity changing parameters
SET 1 ∫ 1 99:59h
400V 300mA 50W
SET 1 ∫ 1 99:59h
400V 300mA 50W
SET 1 ∫ 1 99:59h
400V 300mA 50W
SET 1 ∫ 1 99:59h
400V 300mA 50W
Data Logging
1.
Select SET-UP and press SET.
2. Select DATA and press SET.
3. Select DURATION and press SET.
4. Select the desired time for the data-logging and press SET.
5. Select LOG and press SET.
6. Decide whether or not all data (voltage, current, power, time) during
the next experiments should be logged and press SET.
7. Select SET-UP and press SET to return.
• Starting a new experiment will erase the previous data in memory.
• Stored data can be transferred to a computer via the RS232 output
using optionally available software.
• At any time, you can return in the stand-by mode by holding SET
pressed during 2 seconds.
---- STAND-BY ---SET-UP
---- SET-UP ----
DATA
---- DATA ----
DURATION
LOG DATA
FOR: 10h
---- DATA ---LOG
LOG
DATA: YES
---- DATA ----
SET-UP
Alarms
Ground leakage A dangerous ground leakage is present (check
thoroughly the complete assembly). Your power supply
is NOT defective! See Frequently Asked Questions at
the end of this manual.
Short Circuit Connected tank is short-circuited or has a very low
resistance (check thoroughly the complete assembly).
Your power supply is NOT defective! See Frequently
Asked Questions at the end of this manual.
Overload Connected tank has a too low resistance (check
specifications). Your power supply is NOT defective! See
Frequently Asked Questions at the end of this manual.
Current <1mA The output current is lower than 1 mA (switch the alarm
off in the SET-UP menu). Your power supply is NOT
defective! See Frequently Asked Questions at the end of
this manual.
Call Service Failure in the electronics. See Support on
www.consort.be.
You could also try a general RESET: switch the instrument on
while holding SET pressed. Attention! All stored programs and data will be
erased!
*** ERROR ***
GROUND LEAKAGE
*** ERROR ***
SHORT CIRCUIT
*** ERROR ***
OVERLOAD
*** ERROR ***
CURRENT <1mA
*** ERROR ***
CALL SERVICE
Mode d’emploi
Electrophoresis power supplies Model EV202
EV215
EV231
EV232
EV233
EV243
EV261
EV262
EV265
Loading...
+ 20 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.