Extech PRC50 Operating Manual

User Guide
Multifunction Process Calibrator
Model PRC50
Page 1 PRC50-EN-V2.1 11/11
1 Introduction
This multifunction process calibrator is a handheld, battery-operated instrument that measures and sources electrical and
Main Features:
z Measure: DC-voltage, DC-current, frequency and continuity
z Source: DC-voltage, DC-current, simulation transmitter, frequency and pulse
Additional features:
z Manual step source, auto-step and sweeping–step source
z Measure/source mA% display
z Wave-filter function
z Manual-hold function
2 Standard Accessories
Ensure that the package contains all of the accessories listed below. If items are damaged, please contact the vendor from
which the product was purchased.
z One (1) set of Industrial Test Leads
z One (1) set of Standard Test Leads
z One (1) set of Alligator clips
z User's Guide
z Two (2) Fuses 50mA/250V and 63mA/250V
Page 2 PRC50-EN-V2.1 11/11
3 Safety Information
For the correct and safe use of the instrument, be sure to follow the cautionary notes stated in this manual whenever
handling the instrument. Extech Instruments shall not be held liable for any damage resulting from use of the instrument in a
manner other than prescribed in the cautionary notes.
A
Warning identifies conditions and actions that pose hazards to the user; a Caution identifies conditions and actions that
may damage the meter or the equipment under test.
Refer to Table 1 for the international electric symbols adopted by the calibrator.
Table 1 - International Electrical Symbols
EARTH GROUND
Warning
To avoid possible electric shock or personal injury:
z Do not apply more than the rated voltage, as marked on the calibrator, between terminals or between any terminal and
earth ground;
z Before use, verify the meter’s operation by measuring a known voltage;
z Follow all equipment safety procedures;
z Do not connect the probe of the test leads with any live power when the other end has been inserted into the current
jack;
z Do not use the meter if it is damaged. Before using the meter, inspect the case. Look for cracks or missing plastic. Pay
particular attention to the insulation surrounding the connectors;
z Select the proper function and range for the measurement;
z Make sure the battery door is closed and latched before operating the meter;
z Remove test leads from the meter before opening the battery door;
z Inspect the test leads for damaged insulation or exposed metal. Check test lead continuity. Replace damaged test leads
before using the meter;
z When using the probes, keep fingers behind the finger guards on the probes;
z Connect the common test lead before connecting the live test lead. When disconnecting test leads, disconnect the live
test lead first;
z Do not use the meter if it operates abnormally. Protection may be impaired. When in doubt, have the meter inspected;
z Do not operate this instrument in areas where inflammable or explosive gases or vapor exists. It is extremely hazardous
to use the instrument under such environments;
z Do not operate the meter near explosive gas, vapor, or dust;
z Use only four (4) AAA 1.5V batteries, properly installed in the meter case, to power the meter;
z Disconnect the test leads before changing source or measurement functions;
z When servicing the meter, use only specified replacement parts.
z To avoid false readings, which could lead to possible electric shock or personal injury, replace the batteries as soon as
the low battery indicator (
) appears.
WARNING
INFORMATION
Page 3 PRC50-EN-V2.1 11/11
Caution
To avoid possible damage to meter or to equipment under test:
z Disconnect the power and discharge all high-voltage capacitors before testing resistance or continuity.
z Use the proper jacks, functions, and ranges for the measurement or source operation in question.
4 Calibrator Description
Refer to Figure 1
1. LCD display
2. Digit up/down adjust buttons
3. Control keypad
4. Measurement terminals
5. Source terminals
Figure 1 – Meter Description
4.1 Measurement/ Source Terminals
Figure 2 shows the calibrator’s measurement /source terminals. Table 2 explains their use.
Figure 2 - Measurement/ Source Terminals
Table 2 - Measurement/ Source Terminals
2
4
1
3
5
Terminal Function
1
2
3
4
5
6
7
8
Measurement Signals(+):DCmV、
Measurement Signals(+):DCV、FREQ
Measurement Signals(+):DCmA
Common (return ) (-) terminal for all measurement functions Source Signals:(-)DCmA Source Signals(+) Simulate mA Source Signals:(+)DCmA LOOP Terminal+24VDC Loop Power Terminal Common (return ) (-)terminal for source functions
Source Signals:(+)DCV、FREQ、PULSE
Page 4 PRC50-EN-V2.1 11/11
4.2 Keys
Figure 3 lists the calibrator’s function keys and Table 3 explains their use.
6
12
8910
7
3
4
5
11
13
15
17
19
21
23 2425 26
Figure 3 – Calibrator Function Keys
Table 3 – Function Key Descriptions
No. Name Function
1~5 Source value set keys Increment source set point 6~10 Source value set keys Decrement source set point 11 12 13 Power key Switch ON/OFF power 14 Backlight Switch ON/OFF backlight 15 16
17
18 19 Key Select measurement continuity function 20 21 22 23
24
25
26
MODE key ON key
V Key PULSE Key
mA Key
FREQ Key
RANGE Key UNIT Key LOOP key ZERO Key
START Key
Key
25/100% Key
Mode key (press to change the function mode) Switch ON/OFF measurement/ source function
Select measurement/source DCV function Select source pulse function
Select measurement/source mA function
Select measurement/source FREQ function
Select measurement/source range; mA /percentage toggle key not used Switch ON/OFF 24V Loop power Set the source value to its default value;
In pulse source function, set the pulse value Source auto-pulse value;
Switch ON the mA auto-stepping or sweeping function; HOLD (freeze) the measured value;
In DCmA source function ,select the auto-wave type; In frequency or pulse source, set the frequency value.
In mA source function, select 25% or 100% manual step output type; In pulse number mode, frequency source, set the amplitude value; Measure the average value.
12
14
16
18
20
22
Page 5 PRC50-EN-V2.1 11/11
4.3 Display Screen
Figure 4 shows a typical display screen.
Figure 4 - LCD display
a. Measurement
b. Source
c. Measurement/Source frequency
d. Not used
e. Battery level indicator
f. Measurement/Source function ON
g. Source function OFF
h. Measurement/Source voltage
i. Measurement/Source current
j. Source pulse
k. 24V Loop Power Supply ON
l. Not used
m. Beeper for measurement continuity
n. Not used
o. Measured value HOLD
p. Measurement average value
q. Not used
r. Main numerical display
s. Unit of measure for main display
t. DCmA 25% or 100% manual step source
u. Start source number of pulses / DCmA auto-stepping or sweeping source function
v. 24V Loop Power Supply ON
w. DCmA auto-stepping or sweeping source
x. Unit of measure for secondary display
y. Secondary numerical display
m
r
t
u
abcd e
g
f
h
j
no p
wv y
kl
q
s
x
Page 6 PRC50-EN-V2.1 11/11
Source/Measurement Preparations
Precautions for Safe Use of the Instrument
z When using the instrument for the first time, be sure to read the instructions given in Section 4 “Precautions for Safe
Use of the Instrument.”
z Do not open the instrument’s case. Contact the vendor for service.
z In case of failure: Should the instrument begin to emit smoke, give off an unusual odor, or show any other anomaly,
immediately switch OFF the POWER. Also switch OFF power to the object under test that is connected to the input
terminals. Contact the vendor.
General Handling Precautions
z Before transporting the instrument, switch OFF power to the object under test, and then power off the meter. Finally,
detach all lead cables from the instrument. Use a dedicated carry case when transporting the instrument.
z Do not bring any electrified object close to the input terminals, since this may damage the internal circuitry.
z Do not apply any volatile chemical to the instrument’s case or operation panel. Do not leave the instrument in contact
with any product made of rubber or vinyl for a prolonged period. Be careful not to let a soldering iron or any other heat-
emitting object come into contact with the operation panel (panel is made of thermoplastic resin).
z Use a soft, damp cloth to gently wipe the outer surfaces of the instrument. Water that is allowed into the instrument can
result in malfunction.
z For handling precautions regarding the batteries, see “Installing or Replacing the Batteries”.
z Never use the instrument with the cover of the battery holder opened.
Environmental Requirements
Use the instrument in locations that meet the following environmental requirements:
o
z Ambient temperature range: 32 to 122
z Place the instrument on flat and level areas
Do not use the instrument in locations that are:
z Exposed to direct sunlight or close to any heat source.
z Exposed to frequent mechanical vibration.
z Close to an electrical noise source, such as high-voltage equipment or motive power sources.
z Close to any source of intensive electric or electromagnetic fields.
z Exposed to large amounts of grease fumes, hot steam, dust or corrosive gases.
z Exposed to unstable areas such as where a risk of explosion due to the presence of flammable gases exists.
Notes:
z Use the instrument under the following environmental conditions when precise measurements are required:
o
Ambient temperature range: 74
F (23oC) ±5o; ambient humidity range: 20 to 80% RH (non-condensing)
z When using the instrument within a temperature range of 32 to 64
based on the temperature coefficient shown in the “Specifications” section to the given accuracy rating.
z When using the instrument at an ambient humidity of 30% or lower, prevent electrostatic charges by using an anti-static
mat or other alternative means.
z Condensation may occur if the instrument is relocated from places with low temperature and humidity to places with
high temperature and humidity, or if the instrument experiences a sudden temperature change. In these cases, leave
the instrument under the given ambient temperature for at least one hour to ensure that the instrument is free from
condensation before using the instrument.
F (0 to 50oC); ambient humidity range: 20 to 80% RH; non-condensing.
o
F (0 to 18oC) or 82 to 122oF (28 to 50oC), add a value
Page 7 PRC50-EN-V2.1 11/11
Indication of Battery Level
The battery replacement indicator shows the battery level in five stages.
Full battery:
The battery level is below 50% full:
The battery level is below 25% full:
Low battery:
The battery icon flashes in sequence when charging.
Note that the battery replacement indicator is driven by directly measuring the battery voltage when the calibrator is in actual
operation. Consequently, the indicator may read differently depending on the battery load condition (e.g., the load condition
of the source output or on/off state of the measurement function).
To replace batteries and/or fuses refer to Section 12 of this User Guide.
Switching Power ON
Press the Power key once to switch ON the calibrator. Press and hold the Power key for 2 seconds to switch it OFF.
Automatic Power OFF
When the calibrator is running on batteries and no key is pressed for approximately ten minutes, the calibrator switches OFF
automatically. The automatic power-off time can be programmed or set to the factory default state, see the “Global Settings”
section.
Switching the LCD Backlight ON and OFF
Press the backlight key to switch ON the backlight; press again to switch it OFF. Battery life is shortened by backlight
overuse.
Notes: The backlight automatically switches OFF after 10 seconds. The time can be programmed or reset to the factory
default condition, see the “Global Settings” section.
Page 8 PRC50-EN-V2.1 11/11
Loading...
+ 16 hidden pages