Extech PRC30 Specifications

PRC30 Multifunction
Process Calibrator
Precision source and measure for thermocouples (8-Types), mA, mV, and V devices
Applications include:
• Lab and Field Calibration and Verifi cation Operations of Sensors, Thermocouple Probes, Controllers, Panel meters & Testing tools.
• Temperature Control Process Troubleshooting
• Process devices calibration and simulation: Transducers, Transmitters, Indicators, Controllers, and Recorders
• Field test and service
• Bench research and development
Features
Palm-sized double molded housing and
large dot-matrix backlit LCD
Measures DC process signals
High accuracy calibration function simulates precision
thermocouple outputs for use in calibrating thermometers,
transmitters, controllers or recorders
Up to fi ve user adjustable calibration presets
Displays output in terms of millivolts or temperature
based on thermocouple tables
24V Loop Power
Zero adjustment, auto zero when power on
and change mode
Standard banana I/O ports and mini thermocouple inputs
Large battery bank for extended work cycle
External power adapter for continuous work cycle
Complete with test leads, alligator clips, Thermocouple calibration cable with subminiature connector, universal calibration cable with subminiature terminal, 100V-240V AC Universal Adaptor with 4 plugs, six 1.5V AA batteries, and hard carrying case
PRODUCT DATASHEET PRODUCT DATASHEET
Ordering
PRC30 .............. ... Multifunction Process Calibrator
PRC30-NIST ........PRC30 with Certifi cate of Calibration Traceable to NIST
Specifi cations subject to change without notice. Copyright © 2012-2014 FLIR Systems, Inc. All rights reserved including the right of reproduction in whole or in part in any form. Rev. 4/2/14
www.extech.com
Specifi cations
DC Source
Current 0 to 24mA, -25% to +125% ±(0.01% + 1 digit)
Max. Load 1000Ω @ 24mA
Voltage 0 to 2000mV, 0 to 20V ±(0.01% + 1 digit)
mV/Temp -5 to 55mV ±(0.01% + 1 digit)
Type J -58 to 1832°F (-50 to 1000°C) ±(0.05% + 2°F or 1°C)
Type K -58 to 2498°F (-50 to 1370°C) ±(0.05% + 2°F or 1°C)
Type T -184 to 752°F (-120 to 400°C) ±(0.05% + 2°F or 1°C)
Type E -58 to 1382°F (-50 to 750°C) ±(0.05% + 2°F or 1°C)
Type C, R, S 32 to 3182°F (0 to 1750°C) ±(0.05% + 2°F or 1°C)
Type N -58 to 2372°F (-50 to 1300°C) ±(0.05% + 2°F or 1°C)
DC Measure
Current 0 to 50mA, -25% to +230% ±(0.01% + 1 digit)
Voltage 0 to 1999mV / 2 to 20V ±(0.01% + 1 digit) / auto range
mV/Temp -10 to 60mV ±(0.01% + 1 digit)
Type J -58 to 1832°F (-50 to 1000°C) ±(0.05% + 1.8°F or 1°C)
Type K -58 to 2498°F (-50 to 1370°C) ±(0.05% + 1.8°F or 1°C)
Type T -184 to 752°F (-120 to 400°C) ±(0.05% + 1.8°F or 1°C)
Type E -58 to 1382°F (-50 to 750°C) ±(0.05% + 1.8°F or 1°C)
Type C, R, S 32 to 3182°F (0 to 1750°C) ±(0.05% + 2°F or 1°C)
Type N -58 to 2372°F (-50 to 1300°C) ±(0.05% + 2°F or 1°C)
Loop Power 24V
Meter Dims (LxWxH) 6.3" x 3.2" x 1.7" (159 x 80 x 44mm)
Meter Wt 8.4oz (236g) - not including battery weight
Range Basic Accuracy
Specifi cations subject to change without notice. Copyright © 2012-2014 FLIR Systems, Inc. All rights reserved including the right of reproduction in whole or in part in any form.
www.extech.com