Extech PRC20 Specifications

PRC20 Thermocouple
Calibrator
Source and measure 8 thermocouple type devices (J, K, T, E, C, R, S and N) with one meter that provides precision output displayed as °C/°F or mV.
Applications include:
• Lab and Field Calibration and Verifi cation Operations of Sensors, Thermocouple Probes, Controllers, Panel meters & Testing tools.
• Engineering Research and Design
• Manufacturing and Test Engineering
• HVAC
• Temperature Control Process Troubleshooting
Features
Palm-sized double molded housing and large
dot-matrix digital backlit LCD with thermocouple
type indication
High accuracy calibration function simulates
precision thermocouple outputs for use in calibrating
thermometers, transmitters, controllers or recorders
Displays output in terms of millivolts or temperature
based on thermocouple tables
Standard mini-thermocouple input
Large battery bank for extended work cycle
Up to fi ve user adjustable calibration presets
External power adapter for continuous work cycle
Specifi cations
Thermocouple
Type J -58 to 1832°F (-50 to 1000°C) 0.05% rdg ±1°
Type K -58 to 2498°F (-50 to 1370°C) 0.05% rdg ±1°
Type T -184 to 752°F (-120 to 400°C) 0.05% rdg ±1°
Type E -58 to 1382°F (-50 to 750°C) 0.05% rdg ±1°
Type C 32 to 3182°F (0 to 1750°C) 0.05% rdg ±1°
Type R/S 32 to 3182°F (0 to 1750°C) 0.05% rdg ±1°
Type N -58 to 2372°F (-50 to 1300°C) 0.05% rdg ±1°
Voltage -10.00mV to +60.00mV 10μV 10μV ±1d
Resolution 0.1° (up to 999.9) or 1° (over 999.9°) for J, K, T, E ; 1° for R, N; 0.5° for C, S
Input Impedance 10 MΩ
Sampling Time 4 times/second
Meter Dims (LxWxH) 6.3" x 3.2" x 1.7" (159 x 80 x 44mm)
Meter Wt 8oz (225g) - not including battery weight
Range Basic Accuracy
PRODUCT DATASHEET PRODUCT DATASHEET
Ordering
PRC20 ............ ...Thermocouple Calibrator
PRC20-NIST ... ...PRC20 with Certifi cate of Calibration Traceable to NIST
Specifi cations subject to change without notice. Copyright © 2012-2014 FLIR Systems, Inc. All rights reserved including the right of reproduction in whole or in part in any form. Rev. 4/2/14
Complete with Thermocouple calibration cable with subminiature connectors, standard calibration cable with subminiature connectors, 100V-240V AC Universal Adaptor with 4 plugs, 6 x 1.5V AA batteries, and hard carrying case
www.extech.com