EXFO IQS-2100 User Manual

Page 1
IQS-2100
Light Source for IQS Platforms
User Guide
Page 2
Copyright © 2003–2008 EXFO Electro-Optical Engineering Inc. All rights reserved. No part of this publication may be reproduced, stored in a retrieval system or transmitted in any form, be it electronically, mechanically, or by any other means such as photocopying, recording or otherwise, without the prior written permission of EXFO Electro-Optical Engineering Inc. (EXFO).
Information provided by EXFO is believed to be accurate and reliable. However, no responsibility is assumed by EXFO for its use nor for any infringements of patents or other rights of third parties that may result from its use. No license is granted by implication or otherwise under any patent rights of EXFO.
EXFO’s Commerce And Government Entities (CAGE) code under the North Atlantic Treaty Organization (NATO) is 0L8C3.
The information contained in this publication is subject to change without notice.
Trademarks
EXFO’s trademarks have been identified as such. However, the presence or absence of such identification does not affect the legal status of any trademark.
Units of Measurement
Units of measurement in this publication conform to SI standards and practices.
Version number 2.0.0
ii IQS-2100
Page 3
Contents
Contents
Certification Information ........................................................................................................v
1 Introducing the IQS-2100 Light Source ....................................................... 1
Main Features .........................................................................................................................1
Available Models ....................................................................................................................2
Typical Applications ................................................................................................................2
Conventions ............................................................................................................................3
2 Safety Information ....................................................................................... 5
3 Getting Started with Your Light Source ..................................................... 7
Inserting and Removing Test Modules ....................................................................................7
Starting the Light Source Application ...................................................................................11
Entering Values Using Sliders and Numeric Boxes .................................................................15
Exiting the Application .........................................................................................................16
4 Setting Up Your Light Source .................................................................... 17
Selecting a Wavelength ........................................................................................................18
Setting the Attenuation ........................................................................................................20
Selecting a Modulation Frequency ........................................................................................22
Saving and Recalling Configurations .....................................................................................23
5 Operating Your Light Source ..................................................................... 25
Cleaning and Connecting Optical Fibers ...............................................................................25
Installing the EXFO Universal Interface (EUI) .........................................................................26
Activating or Deactivating Light Emission .............................................................................27
Viewing Results ....................................................................................................................28
6 Monitoring Light Source Modules ............................................................ 29
Using Monitor Windows .......................................................................................................29
Using QuickTools ...................................................................................................................32
7 Maintenance ............................................................................................... 35
Cleaning Fixed Connectors ....................................................................................................36
Cleaning EUI Connectors ......................................................................................................38
Recalibrating the Unit ...........................................................................................................40
Recycling and Disposal (Applies to European Union Only) ....................................................41
Light Source iii
Page 4
Contents
8 Troubleshooting ..........................................................................................43
Solving Common Problems ...................................................................................................43
Viewing Online Documentation ............................................................................................44
Contacting the Technical Support Group ..............................................................................45
Transportation ......................................................................................................................47
9 Warranty ......................................................................................................49
General Information .............................................................................................................49
Liability .................................................................................................................................49
Exclusions .............................................................................................................................50
Certification ..........................................................................................................................50
Service and Repairs ...............................................................................................................51
EXFO Service Centers Worldwide ..........................................................................................52
A Technical Specifications ..............................................................................53
B SCPI Command Reference ..........................................................................55
Quick Reference Command Tree ...........................................................................................56
Product-Specific Commands—Description ............................................................................57
Index .................................................................................................................73
iv IQS-2100
Page 5

Certification Information

Certification Information
F.C.C. Information
Electronic test equipment is exempt from Part 15 compliance (FCC) in the United States. However, compliance verification tests are systematically performed on most EXFO equipment.
 Information
Electronic test equipment is subject to the EMC Directive in the European Union. The EN61326 standard prescribes both emission and immunity requirements for laboratory, measurement, and control equipment. This unit has undergone extensive testing according to the European Union Directive and Standards.
IMPORTANT
Use of shielded remote I/O cables, with properly grounded shields and metal connectors, is recommended in order to reduce radio frequency interference that may emanate from these cables.
Light Source v
Page 6
Certification Information
DECLARATION OF CONFORMITY
Application of Council Directive(s): 73/23/EEC - The Low Voltage Directive
Manufacturer’s Name: EXFO Electro-Optical Engineering Inc. Manufacturer’s Address: 400 Godin Avenue
Equipment Type/Environment: Test & Measurement / Industrial Trade Name/Model No.: IQS-2100
Standard(s) to which Conformity is Declared:
EN 61010-1:2001 Safety Requirements for Electrical Equipment for M easurement,
EN 60825-1:1994 +A11:1996 +A2: 2001 + A1: 2002
EN 61326:1997 +A1: 1998 +A2: 2001
EN 55022: 1998 +A1: 2000 Limits and Methods of Measurement of Radio Di sturbance
I, the undersigned, hereby declare that the equipment specified above conforms to the above Directive and Standards.
Manufacturer
Signature:
Control, and Laboratory Use, Part 1: General Requirements.
Safety of laser products – Part 1: Equipment classifications, requirements, and user’s guide
Electrical Equipment for M easurement, Control and Laboratory Use - EMC Requirements
Characteristics of Information Technology Equipment.
89/336/EEC - The EMC Directive And their amendments
Quebec, Quebec Canada, G1M 2K2 (418) 683-0211
Light Source
Full Name: Stephen Bull, E. Eng
Position: Vice-President Research and
Address: 400 Godin Avenue, Quebec (Quebec),
Date: January 7, 2002
Development
Canada, G1M 2K2
vi IQS-2100
Page 7
1 Introducing the IQS-2100 Light
Source

Main Features

The IQS-2100 Light Source is designed for scientific and industrial applications using the IQS Platforms. The IQS-2100 Light Source includes a choice of multimode LEDs and temperature-stabilized single-mode Fabry-Perots laser emitters, in single- and dual-wavelength configurations, for both singlemode and multimode test applications.
ACTIVE LED
LED push button
The IQS-2100 Light Source is designed for optimal stability. Steady drive circuitry maximizes optical output power and maintains excellent stability, while precision optical components ensure efficient, low-loss, narrow-beam output coupling. Laser sources are stabilized by thermoelectric Peltier coolers that regulate submount internal temperature.
ACTIVE
IQS-2100
LIGHT SOURCE
Source port
Retaining screw knob
Light Source 1
Page 8
Introducing the IQS-2100 Light Source

Available Models

A single-operation display screen lets you activate sources and select signal wavelength, power, and modulation characteristics. The emitter’s central wavelength is also displayed. For easy repeat access, the software stores multiple power level and modulation configurations. Its Windows-based software allows the IQS-2100 Light Source to integrate easily into any test system.
The IQS-2100 Light Source supports local control (via the IQS Manager software) and remote control (through GPIB, RS-232, or Ethernet TCP/IP using SCPI commands or the provided LabVIEW drivers). For more information, refer to the IQS platform user guide.
Available Models
The IQS-2100 Light Source offers different models, which are distinguished by characteristics such as the type of LED or laser used:
³ Fabry-Perot laser
³ Surface-Emitting LED

Typical Applications

This light source is suitable for quality control, calibration, acceptance testing, and loss and return loss testing in laboratory and manufacturing environments.
2 IQS-2100
Page 9
Introducing the IQS-2100 Light Source

Conventions

Conventions
Before using the product described in this manual, you should understand the following conventions:
WARNING
Indicates a potentially hazardous situation which, if not avoided, could result in death or serious injury. Do not proceed unless you understand and meet the required conditions.
CAUTION
Indicates a potentially hazardous situation which, if not avoided, may result in minor or moderate injury. Do not proceed unless you understand and meet the required conditions.
CAUTION
Indicates a potentially hazardous situation which, if not avoided, may result in component damage. Do not proceed unless you understand and meet the required conditions.
IMPORTANT
Refers to information about this product you should not overlook.
Light Source 3
Page 10
Page 11

2 Safety Information

WARNING
Do not install or terminate fibers while a light source is active. Never look directly into a live fiber and ensure that your eyes are protected at all times.
WARNING
Use of controls, adjustments and procedures for operation and maintenance other than those specified herein may result in hazardous radiation exposure.
Your instrument is a Class 1 laser or LED product in compliance with standards IEC 60825-1 Amendment 2: 2001 and 21 CFR 1040.10. Laser radiation may be encountered at the output port.
The following labels indicate that a product contains a Class 1 source:
CLASS 1
LASER PRODUCT
Note: Labels shown for information purposes only. They are not affixed to your
product.
Light Source 5
CLASS 1
LED PRODUCT
Page 12
Page 13
3 Getting Started with Your
Light Source

Inserting and Removing Test Modules

CAUTION
Never insert or remove a module while the controller unit and its expansion units are turned on. This will result in immediate and irreparable damage to both the module and unit.
To insert a module into the controller or expansion unit:
1. Exit IQS Manager and turn off all your units.
2. Remove the protective cover from the desired unused module slot.
2a. Pull the retaining screw knob firmly towards you and release the
bottom of the cover.
2b. Gently pull the top of the protective cover downwards, to remove
it from the unit grooves.
Protective cover
Retaining screw
knob
2a
3. Position the module so that its front panel is facing you and the top and bottom protruding edges are to your right.
Light Source 7
2b
Page 14
Getting Started with Your Light Source
Inserting and Removing Test Modules
4. Insert the protruding edges of the module into the grooves of the unit’s module slot.
5
Retaining screw knob
Retaining screw
Protruding edges (right side of module)
5. Push the module all the way to the back of the slot, until the retaining screw makes contact with the unit casing.
6. While applying slight pressure to the module, turn the retaining screw knob (located at the bottom of the panel) clockwise until the knob is horizontal.
This will secure the module into its “seated” position.
6
8 IQS-2100
Page 15
Getting Started with Your Light Source
Inserting and Removing Test Modules
The module is correctly inserted when its front panel is flush with the front panel of the controller or expansion unit.
When you turn on the controller unit, the startup sequence will automatically detect your module.
Note: You can insert IQ modules into your controller or expansion unit; the
IQS Manager software will recognize them. However, the IQS-2100 locking mechanism (retaining screw) will not work for IQ modules.
To remove a module from your controller or expansion unit:
1. While pulling gently on the knob, turn it counterclockwise until it stops.
The module will slowly be released from the slot.
Retaining screw knob
2. Place your fingers underneath the module or hold it by the retaining screw knob (NOT by the connector) and pull it out.
Light Source 9
Page 16
Getting Started with Your Light Source
Inserting and Removing Test Modules
Connector
NO
YES
Retaining screw
knob
CAUTION
Pulling out a module by a connector could seriously damage both the module and connector. Always pull out a module by the retaining screw knob.
3. Cover empty slots with the supplied protective covers.
3a. Slide the top of the protective cover into the upper grooves of the
unit.
3b. Snap the cover into place by pushing the retaining screw knob.
CAUTION
Failure to reinstall protective covers over empty slots will result in ventilation problems.
10 IQS-2100
Page 17
Getting Started with Your Light Source

Starting the Light Source Application

Starting the Light Source Application
Your IQS-2100 Light Source module can be configured and controlled from its dedicated IQS Manager application.
Note: For details about IQS Manager, refer to the IQS platform user guide.
To start the application:
1. From the Current Modules function tab select the module to use.
It will turn white to indicate that it is highlighted.
2. Click Start Application.
OR
Press the green LED push button on the front of the corresponding module.
You can also double-click its row.
Light Source 11
Page 18
Getting Started with Your Light Source
Starting the Light Source Application
Note: Pressing the LED push button will not activate or turn on the module.
Note: To start the corresponding monitor window at the same time, click Start
App. & Monitor. The window opens on the Monitors function tab.
The main window (shown below) contains all the commands required to control the Light Source:
Title bar
Data
display
Function bar
Control
Center
Status bar
12 IQS-2100
Page 19
Getting Started with Your Light Source
Starting the Light Source Application
Title Bar
The title bar is located at the top of the main window. It displays the module name and its position in the controller or expansion unit. The module position is identified as follows:
Controller unit or expansion unit (1 to 999) housing the module
Slot number in which module is inserted (0 identifies first slot)
[ 999 – 1 ]
Status Bar
The status bar, located at the bottom of the main window, identifies the operational status of the IQS-2100 Light Source.
Control mode
Local: Module controlled locally only.
Remote: Module controlled remotely, but
local commands can also be used.
Lockout: Module controlled remotely only.
Current date and timeModule/unit status
For more information about automating or remotely controlling the IQS-2100 Light Source, refer to your platform user guide.
Light Source 13
Page 20
Getting Started with Your Light Source
Starting the Light Source Application
Data Display
In addition to the Control Center, the main window also contains the data display, where the source status indicator, as well as values for wavelength, modulation, and attenuation controls are shown (see figure below).
Source
status
indicator
Signal wavelength
Modulation value
indicator
Attenuation value indicator
14 IQS-2100
Page 21
Getting Started with Your Light Source

Entering Values Using Sliders and Numeric Boxes

Entering Values Using Sliders and Numeric Boxes
Many parameters in IQS Manager and module applications can be set using the following tools.
Slider
Fine-tuning
box
Navigation buttons
³ Slider: Drag it to the desired value on the scale below.
³ Navigation buttons: Click either buttons to move the slider. The slider
moves by steps corresponding to the number in the fine-tuning box, which you can change by using the up and down arrow buttons next to the box. You cannot change the list of fine-tuning values from here.
³ Numeric box: Click it to display the on-screen numeric pad, which you
can use to enter a powe value.
Numeric box button
To enter a value using the numeric box:
1. Use the button to clear the entry display.
Parameter identification
Entry display
Clear button
Toggle button
Backspace button (to correct a value)
2. Enter the value.
3. Click OK to confirm the value.
Light Source 15
Page 22
Getting Started with Your Light Source

Exiting the Application

Exiting the Application
Closing any application that is not currently being used is a good way to free system memory.
To close the application from the main window:
Click in the top right corner of the main window.
OR
Click the Exit button located at the bottom of the function bar.
To close all currently running applications:
From IQS Manager, click Close All Applications.
16 IQS-2100
Page 23

4 Setting Up Your Light Source

Light source features are controlled with the Windows-compatible IQS Manager software. Please refer to the IQS Platforms user guides for information regarding the IQS-500/600 Controller Units and IQS Manager software conventions.
The source is controlled and operated from within the Control Center. Depending on your source model, some or all of the following operations are available:
³ selecting wavelength (for dual-wavelength modules only)
³ setting attenuation
³ selecting modulation
³ saving and recalling a configuration
Note: You do not need to turn on the IQS-2100 Light Source or connect it to a DUT
to set it up. To turn on the source, see Operating Your Light Source on page 25.
Light Source 17
Page 24
Setting Up Your Light Source

Selecting a Wavelength

Selecting a Wavelength
If you are using a dual-wavelength module, you can select a wavelength to perform your tests. When the source is activated, the wavelength appears in the top right corner.
18 IQS-2100
Page 25
Setting Up Your Light Source
Selecting a Wavelength
To select the desired wavelength:
Click the appropriate button in the Wavelength panel. A green light identifies the selected wavelength.
To indicate a successful wavelength selection, the message Stabilization in progress... appears on the data display (only if the source is an active laser).
You can also select a wavelength from the QuickTools utility. For details, see Monitoring Light Source Modules on page 29.
Note: The wavelengths specified in the Wav e l ength panel are the measured
wavelengths of the optical sources (to the nearest 1 nm for FP lasers, and to the nearest 10 nm for LED sources).
Light Source 19
Page 26
Setting Up Your Light Source

Setting the Attenuation

Setting the Attenuation
You can modify the power of the IQS-2100 Light Source output by attenuating the emitted signal.
With some IQS-2100 Light Source models, the Attenuation control is grayed out during the stabilization period, which occurs after you have selected a wavelength if the source is activated.
To set the source attenuation:
1. Select the Instrument function tab.
2. Adjust the attenuation of the laser signal emitted by the source from
the Attenuation (dB) panel. For details, see Entering Values Using Sliders and Numeric Boxes on page 15.
The attenuation value increases or decreases by increments of 0.1 dB.
Attenuation value indicator
Focus indicator
20 IQS-2100
Page 27
Setting Up Your Light Source
Setting the Attenuation
The attenuation value in the data display will then indicate the increased or decreased attenuation.
As you change the attenuation value, you will notice that the numbers change from green to orange. After you release the attenuation controls, they revert to their original green color to indicate that the change has been applied to the source (when it is active). The focus indicator rectangle on the control turns blue when setting the attenuation value.
The attenuation value indicator on the data display (see preceding figure) indicates the level of attenuation in dB, selected using the attenuation controls.
You can also set attenuation from the QuickTools utility. For details, see Monitoring Light Source Modules on page 29.
Note: To obtain maximum output power, attenuation should be set to 0.0 dB and
modulation should be set to None.
Light Source 21
Page 28
Setting Up Your Light Source

Selecting a Modulation Frequency

Selecting a Modulation Frequency
You can modulate the laser output of the IQS-2100 Light Source to simulate data transfer. A number of modulation frequencies are available.
To select a modulation frequency:
1. Select a modulation frequency by choosing a value in the Modulation
panel, as shown in the following figure.
A green light identifies the selected modulation frequency.
To indicate a successful selection, Stabilization in progress appears in the data display (only if the source is an active laser).
The Modulation value indicator in the data display indicates the source modulation frequency selected.
Modulation value indicator
2. You can also select a modulation frequency from the QuickTools utility. For details, see Monitoring Light Source Modules on page 29.
Note: The internal modulation is full on/off modulation at a 50 % duty cycle.
22 IQS-2100
Page 29
Setting Up Your Light Source

Saving and Recalling Configurations

Saving and Recalling Configurations
Once you have set the IQS-2100 Light Source parameters, you can save your custom configuration and recall it at any time. You can also recall the factory-defined settings.
Saved configurations include all parameters set in the Control Center (Instrument function tab) and in the Settings function tab (if present).
To save a configuration:
1. Select the Configuration function tab.
To always use the last saved parameters when starting.
To save parameters being used just before shutting down, overwriting the previous file.
2. In the Current Module Configuration panel, enter the name you wish to use for your configuration file.
It will be saved in D:\IQS Manager\Configuration Files\(your_module)\.
3. Click Save.
Light Source 23
Page 30
Setting Up Your Light Source
Saving and Recalling Configurations
To recall a configuration:
1. Select the Configuration function tab.
2. Click Open.
3. Select the configuration file you wish to recall and confirm your action.
You are returned to the application and the new parameters are set.
To revert to factory settings:
1. Select the Configuration function tab.
2. Click the Reset Module to Factory Settings button.
Reverting to the factory settings will interrupt any module operation in progress.
The operation may take a few seconds to complete.
IMPORTANT
IMPORTANT
24 IQS-2100
Page 31

5 Operating Your Light Source

Once you have set your source parameters, you are ready to use the source in a test setup.
To operate the source, you must perform the following steps:
³ Connect the source to other test components.
³ Activate or deactivate the source.

Cleaning and Connecting Optical Fibers

IMPORTANT
To ensure maximum power and to avoid erroneous readings:
³ Always clean fiber ends as explained below before inserting
them into the port. EXFO is not responsible for damage or errors caused by bad fiber cleaning or handling.
³ Ensure that your patchcord has appropriate connectors. Joining
mismatched connectors will damage the ferrules.
To connect the fiber-optic cable to the port:
1. Clean the fiber ends as follows:
1a. Gently wipe the fiber end with a lint-free swab dipped in isopropyl
alcohol.
1b. Use compressed air to dry completely.
1c. Visually inspect the fiber end to ensure its cleanliness.
2. Carefully align the connector and port to prevent the fiber end from
touching the outside of the port or rubbing against other surfaces.
If your connector features a key, ensure that it is fully fitted into the port’s corresponding notch.
3. Push the connector in so that the fiber-optic cable is firmly in place, thus ensuring adequate contact.
Light Source 25
Page 32
Operating Your Light Source

Installing the EXFO Universal Interface (EUI)

If your connector features a screwsleeve, tighten the connector enough to firmly maintain the fiber in place. Do not overtighten, as this will damage the fiber and the port.
Note: If your fiber-optic cable is not properly aligned and/or connected, you will
notice heavy loss and reflection.
Installing the EXFO Universal Interface (EUI)
The EUI fixed baseplate is available for connectors with angled (APC) or non-angled (UPC) polishing. A green border around the baseplate indicates that it is for APC-type connectors.
Green border
indicates APC
option
Bare metal
(or blue border)
indicates UPC
option
To install an EUI connector adapter onto the EUI baseplate:
1. Hold the EU connector adapter so the dust cap opens downwards.
2 3 4
2. Close the dust cap in order to hold the connector adapter more firmly.
3. Insert the connector adapter into the baseplate.
4. While pushing firmly, turn the connector adapter clockwise on the
baseplate to lock it in place.
26 IQS-2100
Page 33
Data display
Red arrows
Source
status
indicator
Active LED
Source
On/Off
button
Operating Your Light Source

Activating or Deactivating Light Emission

Activating or Deactivating Light Emission
Before activating the source, read carefully Safety Information on page 5. Upon source activation, the set parameter values are used. Therefore, make sure the source setup is correct before activation.
To activate or deactivate light emission:
³ Select ON to activate the light emission. The ACTIVE LED on the
module’s front panel lights up, indicating that the source is active.
The data display lights up and two red arrows appear beside the status indicator on the data display, indicating that the source is on (see preceding figure).
³ Select OFF to deactivate the light emission. The ACTIVE LED on the
module’s front panel turns off, indicating that the source is off.
The status indicator on the data display is dimmed and the two red arrows disappear from the data display indicating that the source has been deactivated.
Light Source 27
Page 34
Operating Your Light Source

Viewing Results

Note: You should let the source warm up for 30 minutes to obtain optimum
wavelength stability.
You can also activate or deactivate the light source from the QuickTools utility. For more information, see Monitoring Light Source Modules on page 29.
Viewing Results
You cannot view results directly using the IQS-2100 Light Source software. To view results, you must use modules and systems which perform tests. For more information, refer to test modules or system user guides, or call EXFO.
28 IQS-2100
Page 35
6 Monitoring Light Source
Modules
When using your IQS-2100 Light Source module, either alone or with other modules in a test setup, you can view module data and status using its monitor window in IQS Manager.

Using Monitor Windows

Monitor windows display basic data about modules. A combination of resizable windows allows you to create an integrated data display (refer to the platform user guide).
From the monitor window, you can change module parameters either by:
³ opening the module application to access all the functions
OR
³ using the QuickTools utility, which provides frequently used functions
from the application.
Light Source 29
Page 36
Monitoring Light Source Modules
Using Monitor Windows
To select modules and display their monitor windows:
1. On the Current Modules function tab, select the controller or
expansion unit containing the modules you want to monitor.
Selected modules
(checked)
2. In the Monitor column, select the box next to each module you want to monitor.
If you want to monitor all the modules in the current unit, click Select
All Monitors. If you want to clear your choices, click Deselect All Monitors.
3. Click Start Monitor to apply your selection.
IQS Manager will display the selected monitor windows on the Monitors function tab.
Note: To start t h e highlighted module’s corresponding application at the same
time, click Start App. & Monitor. The application will appear in a different window.
30 IQS-2100
Page 37
Remove Monitor
button
Rearrange
Monitors
button
(1 or 2
columns)
Monitoring Light Source Modules
Using Monitor Windows
Monitor window
arrow buttons
Light Source 31
Close All button
Page 38
Monitoring Light Source Modules

Using QuickTools

Using QuickTools
With QuickTools, you can fine-tune your module directly, while keeping an eye on your entire test setup.
Note: You can only access QuickTools if the module’s monitor window is selected
from the Monitors function tab and is currently active.
To start QuickTools:
1. From the Monitors function tab, elect the monitor window of the
module you wish to control.
2. Using the arrow button in the upper left corner, select QuickTools.
The corresponding monitor window flashes when QuickTools is activated.
Note: If you want to open the actual application for your module rather than
QuickTools, click Show Controller.
32 IQS-2100
Page 39
Monitoring Light Source Modules
Using QuickTools
For the IQS-2100 Light Source, one of the two QuickTools utility versions will be displayed, depending on the selected module, as shown in the following figure.
Single-
wavelength
source
QuickTools
utility
Dual-
wavelength
source
QuickTools
utility
To control a specific source with QuickTools:
Make sure that the IQS-2100 Light Source monitor window is selected (its title bar should be displayed in the same color as the sidebar buttons).
³ Turn the source on (for more information, see Activating or
Deactivating Light Emission on page 27).
³ From the Attenuation section, click the -0.1 dB and +0.1 dB buttons
to increase or lower the attenuation by increments of 0.1 dB. The current Attenuation value is displayed between the button (for details, see Setting the Attenuation on page 20).
Note: To obtain maximum output power, the attenuation should be set to 0.0 dB
and modulation should be set to None.
³ From the Wavelength section, you can select a wavelength by clicking
the selection arrows on both sides of the list (for details, see Selecting a Wavelength on page 18).
Note: For single-source IQS-2100 modules, the Wav e l e ngth panel is not displayed.
Light Source 33
Page 40
Monitoring Light Source Modules
Using QuickTools
³ From the Modulation section, select a modulation frequency by
clicking the selection arrows on both sides of the list (for details, see Selecting a Modulation Frequency on page 22).
To c l os e Q u ic k Too ls :
³ Click the Close button located at the top of the window.
OR
³ Click outside the QuickTools window.
To close a monitor window:
Click the button on the upper left of the monitor window and select Remove Monitor.
OR
Click the Close All button at the bottom of the window.
34 IQS-2100
Page 41

7 Maintenance

To help ensure long, trouble-free operation:
³ Always clean fiber-optic connectors before using them.
³ Keep the unit free of dust.
³ Clean the unit casing and front panel with a cloth slightly dampened
with water.
³ Store unit at room temperature in a clean and dry area. Keep the unit
out of direct sunlight.
³ Avoid high humidity or significant temperature fluctuations.
³ Avoid unnecessary shocks and vibrations.
³ If any liquids are spilled on or into the unit, turn off the power
immediately and let the unit dry completely.
Use of controls, adjustments, and procedures for operation and maintenance other than those specified herein may result in hazardous radiation exposure.
WARNING
Light Source 35
Page 42
Maintenance

Cleaning Fixed Connectors

Cleaning Fixed Connectors
Regular cleaning of connectors will help maintain optimum performance. Do not try to disassemble the unit. Doing so would break the connector.
To clean fixed connectors:
1. Fold a lint-free wiping cloth in four to form a square.
2. Moisten the center of the lint-free wiping cloth with only one drop of
isopropyl alcohol.
Alcohol may leave traces if used abundantly. Avoid contact between the tip of the bottle and the wiping cloth, and do not use bottles that distribute too much alcohol at a time.
3. Gently wipe the connector threads three times with the folded and moistened section of the wiping cloth.
IMPORTANT
IMPORTANT
Isopropyl alcohol takes approximately ten seconds to evaporate. Since isopropyl alcohol is not absolutely pure, evaporation will leave microscopic residue. Make sure you dry the surfaces before evaporation occurs.
4. With a dry lint-free wiping cloth, gently wipe the same surfaces three times with a rotating movement.
5. Throw out the wiping cloths after one use.
6. Moisten a cleaning tip (2.5 mm tip) with only one drop of isopropyl
alcohol.
36 IQS-2100
Page 43
Maintenance
Cleaning Fixed Connectors
IMPORTANT
Alcohol may leave traces if used abundantly. Avoid contact between the tip of the bottle and the cleaning tip, and do not use bottles that distribute too much alcohol at a time.
7. Slowly insert the cleaning tip into the connector until it reaches the ferrule inside (a slow clockwise rotating movement may help).
7
8
9
8. Gently turn the cleaning tip one full turn.
9. Continue to turn as you withdraw the cleaning tip.
10. Repeat steps 7 to 9, but this time with a dry cleaning tip (2.5 mm tip
provided by EXFO).
Note: Make sure you don’t touch the soft end of the cleaning tip and verify the
cleanliness of the cotton tip.
11. Throw out the cleaning tips after one use.
Light Source 37
Page 44
Maintenance

Cleaning EUI Connectors

Cleaning EUI Connectors
Regular cleaning of EUI connectors will help maintain optimum performance. There is no need to disassemble the unit.
If any damage occurs to internal connectors, the module casing will have to be opened and a new calibration will be required.
To clean EUI connectors:
1. Remove the EUI from the instrument to expose the connector
baseplate and ferrule.
IMPORTANT
Turn
Push
2. Moisten a 2.5 mm cleaning tip with one drop of isopropyl alcohol (alcohol may leave traces if used abundantly).
3. Slowly insert the cleaning tip into the EUI adapter until it comes out on the other side (a slow clockwise rotating movement may help).
Pull
3
4
5
4. Gently turn the cleaning tip one full turn, then continue to turn as you withdraw it.
38 IQS-2100
Page 45
Cleaning EUI Connectors
5. Repeat steps 3 to 4 with a dry cleaning tip.
Note: Make sure you don’t touch the soft end of the cleaning tip.
6. Clean the ferrule in the connector port as follows:
6a. Deposit one drop of isopropyl alcohol on a lint-free wiping cloth.
IMPORTANT
Isopropyl alcohol may leave residues if used abundantly or left to evaporate (about 10 seconds).
Avoid contact between the tip of the bottle and the wiping cloth, and dry the surface quickly.
6b. Gently wipe the connector and ferrule.
6c. With a dry lint-free wiping cloth, gently wipe the same surfaces to
ensure that the connector and ferrule are perfectly dry.
6d. Verify connector surface with a portable fiber-optic microscope
(for example, EXFO’s FOMS) or inspection probe (for example, EXFO’s FIP).
Maintenance
WARNING
Verifying the surface of the connector WHILE THE UNIT IS ACTIVE WILL result in permanent eye damage.
7. Put the EUI back onto the instrument (push and turn clockwise).
8. Throw out cleaning tips and wiping cloths after one use.
Light Source 39
Page 46
Maintenance

Recalibrating the Unit

Recalibrating the Unit
Manufacturing and service center calibrations are based on the ISO/IEC 17025 Standard, which states that calibration documents must not contain a recommended calibration interval, unless this has been previously agreed upon with the customer.
Validity of specifications depends on operating conditions. For example, the calibration validity period can be longer or shorter depending on the intensity of use, environmental conditions and unit maintenance. You should determine the adequate calibration interval for your unit according to your accuracy requirements.
Under normal use, EXFO recommends calibrating your unit every year.
40 IQS-2100
Page 47
Maintenance

Recycling and Disposal (Applies to European Union Only)

Recycling and Disposal (Applies to European Union Only)
Recycle or dispose of your product (including electric and electronic accessories) properly, in accordance with local regulations. Do not dispose of it in ordinary garbage receptacles.
This equipment was sold after August 13, 2005 (as identified by the black rectangle).
³ Unless otherwise noted in a separate agreement between EXFO and a
customer, distributor or commercial partner, EXFO will cover costs related to the collection, treatment, recovery and disposal of end-of-lifecycle waste generated by electronic equipment introduced after August 13, 2005 to an European Union member state with legislation regarding Directive 2002/96/EC.
³ Except for reasons of safety or environmental benefit, equipment
manufactured by EXFO, under its brand name, is generally designed to facilitate dismantling and reclamation.
For complete recycling/disposal procedures and contact information, visit the EXFO Web site at www.exfo.com/recycle.
Light Source 41
Page 48
Page 49

8 Troubleshooting

Solving Common Problems

The following is a list of common problems along with their possible causes and some recommended actions to solve them.
Note: In all cases, if problem persists after performing the recommended actions,
contact EXFO.
Problem Possible Cause Recommended Action
LED push button does not light up.
Pushing the LED push button does not open the module main window.
Impossible to open a window.
Power not on. Check AC power cord and turn
on the IQS-500/600 Controller or Expansion Unit
Module is not properly inserted.
Computer is locked up. Reboot the IQS-500/600
LED is burnt. Call EXFO.
Computer is locked up. Reboot the IQS-500/600
Too many windows are open at the same time.
Turn off the IQS-500/600 Controller or Expansion Unit, then remove and reinstall the module.
Controller Unit.
Controller Unit.
Close unused windows, then try to open the needed window again.
Light Source 43
Page 50
Troubleshooting

Viewing Online Documentation

Viewing Online Documentation
An online version of the IQS-2100 Light Source user guide is available at all times from the application.
To access the online user guide:
Click Help in the function bar.
44 IQS-2100
Page 51
Troubleshooting

Contacting the Technical Support Group

Contacting the Technical Support Group
To obtain after-sales service or technical support for this product, contact EXFO at one of the following numbers. The Technical Support Group is available to take your calls from Monday to Friday, 8:00 a.m. to 7:00 p.m. (Eastern Time in North America).
For detailed information about technical support, visit the EXFO Web site at
www.exfo.com.
Technical Support Group
400 Godin Avenue Quebec (Quebec) G1M 2K2 CANADA
To accelerate the process, please have information such as the name and the serial number (see the product identification label—an example is shown below), as well as a description of your problem, close at hand.
1 866 683-0155 (USA and Canada) Tel.: 1 418 683-5498 Fax: 1 418 683-9224
support@exfo.com
P/N
**************** A
542392-3D
S/N Made in Canada QST442B
January 2020
Ver.
Mfg. date
465 Godin Avenue Vanier (Quebec) G1M 3G7 CANADA
IQS-21XXXXX-XX
IQS light source code
Connector code or Universal interface
You may also be requested to provide software and module version numbers. This information, as well as technical support contact information, can be found in the About function tab.
Light Source 45
Page 52
Troubleshooting
Contacting the Technical Support Group
³ Select the Technical Support tab to view phone numbers and active
Internet links to EXFO’s Technical Support Group. Use these links to send an information request by email or to access EXFO’s web site.
³ Select the Module Information tab to view the module identification,
serial number and firmware version.
46 IQS-2100
Page 53
Troubleshooting

Transportation

Transportation
Maintain a temperature range within specifications when transporting the unit. Transportation damage can occur from improper handling. The following steps are recommended to minimize the possibility of damage:
³ Pack the unit in its original packing material when shipping.
³ Avoid high humidity or large temperature fluctuations.
³ Keep the unit out of direct sunlight.
³ Avoid unnecessary shocks and vibrations.
Light Source 47
Page 54
Page 55

9 Warranty

General Information

EXFO Electro-Optical Engineering Inc. (EXFO) warrants this equipment against defects in material and workmanship for a period of two years from the date of original shipment. EXFO also warrants that this equipment will meet applicable specifications under normal use.
During the warranty period, EXFO will, at its discretion, repair, replace, or issue credit for any defective product, as well as verify and adjust the product free of charge should the equipment need to be repaired or if the original calibration is erroneous. If the equipment is sent back for verification of calibration during the warranty period and found to meet all published specifications, EXFO will charge standard calibration fees.
THIS WARRANTY IS IN LIEU OF ALL OTHER WARRANTIES EXPRESSED, IMPLIED, OR STATUTORY, INCLUDING, BUT NOT LIMITED TO, THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE. IN NO EVENT SHALL EXFO BE LIABLE FOR SPECIAL, INCIDENTAL, OR CONSEQUENTIAL DAMAGES.

Liability

EXFO shall not be liable for damages resulting from the use of the product, nor shall be responsible for any failure in the performance of other items to which the product is connected or the operation of any system of which the product may be a part.
EXFO shall not be liable for damages resulting from improper usage or unauthorized modification of the product, its accompanying accessories and software.
Light Source 49
Page 56
Warranty

Exclusions

Exclusions
EXFO reserves the right to make changes in the design or construction of any of its products at any time without incurring obligation to make any changes whatsoever on units purchased. Accessories, including but not limited to fuses, pilot lamps, batteries and universal interfaces (EUI) used with EXFO products are not covered by this warranty.
This warranty excludes failure resulting from: improper use or installation, normal wear and tear, accident, abuse, neglect, fire, water, lightning or other acts of nature, causes external to the product or other factors beyond EXFO’s control.

Certification

EXFO certifies that this equipment met its published specifications at the time of shipment from the factory.
50 IQS-2100
Page 57
Warranty

Service and Repairs

Service and Repairs
EXFO commits to providing product service and repair for five years following the date of purchase.
To send any equipment for service or repair:
1. Call one of EXFO’s authorized service centers (see EXFO Service
Centers Worldwide on page 52). Support personnel will determine if the equipment requires service, repair, or calibration.
2. If equipment must be returned to EXFO or an authorized service center, support personnel will issue a Return Merchandise Authorization (RMA) number and provide an address for return.
3. If possible, back up your data before sending the unit for repair.
4. Pack the equipment in its original shipping material. Be sure to include
a statement or report fully detailing the defect and the conditions under which it was observed.
5. Return the equipment, prepaid, to the address given to you by support personnel. Be sure to write the RMA number on the shipping slip. EXFO
will refuse and return any package that does not bear an RMAnumber.
Note: A test setup fee will apply to any returned unit that, after test, is found to
meet the applicable specifications.
After repair, the equipment will be returned with a repair report. If the equipment is not under warranty, you will be invoiced for the cost appearing on this report. EXFO will pay return-to-customer shipping costs for equipment under warranty. Shipping insurance is at your expense.
Routine recalibration is not included in any of the warranty plans. Since calibrations/verifications are not covered by the basic or extended warranties, you may elect to purchase FlexCare Calibration/Verification Packages for a definite period of time. Contact an authorized service center (see EXFO Service Centers Worldwide on page 52).
Light Source 51
Page 58
Warranty

EXFO Service Centers Worldwide

EXFO Service Centers Worldwide
If your product requires servicing, contact your nearest authorized service center.
EXFO Headquarters Service Center
400 Godin Avenue Quebec (Quebec) G1M 2K2 CANADA
EXFO Europe Service Center
Omega Enterprise Park, Electron Way Chandlers Ford, Hampshire S053 4SE ENGLAND
EXFO China Service Center/ Beijing OSIC
Beijing New Century Hotel Office Tower, Room 1754-1755 No. 6 Southern Capital Gym Road Beijing 100044 P. R . C H I N A
1 866 683-0155 (USA and Canada) Tel.: 1 418 683-5498 Fax: 1 418 683-9224
quebec.service@exfo.com
Tel.: +44 2380 246810 Fax: +44 2380 246801
europe.service@exfo.com
Tel.: +86 (10) 6849 2738 Fax: +86 (10) 6849 2662
beijing.service@exfo.com
52 IQS-2100
Page 59

A Technical Specifications

SPECIFICATIONS
TEC Fabry-Perot Laser Specifications
a
Model 02BLC 03BLC 04BLC 23BLC 34BLC
Wavelength b(nm) 1310 +20/—30 1550 ± 20 1625 ± 15 1310 +20/—30 1550 ± 20
1550 ± 20 1625 ± 15 Spectral width (rms) c(nm) 2 5 10 2/5 5/10 Output power (dBm) * 0 * 0 * —4 * —1 * —4 Stability d(dB) (D/2)
15 min ± 0.003 ± 0.003 ± 0.01 ± 0.005 ± 0.01
8 h ± 0.03 ± 0.03 ± 0.05 ± 0.05 ± 0.05 Temperature sensitivity e(dB) ± 0.25 ± 0.25 ± 0.25 ± 0.25 ± 0.25 Modulation 270 Hz, 1 kHz, 2 kHz (50 % duty cycle)
Model 02ORL 03ORL 04ORL 23ORL 34ORL
Wavelength b(nm) 1310 +20/—30 1550 ± 20 1625 ± 15 1310 +20/—30 1550 ± 20
1550 ± 20 1625 ± 15 Spectral width (rms) c(nm) 2 5 10 2/5 5/10 Output power (dBm) * —2 * —2 * —6 * —3 * —6 Stability d(dB) (D/2)
15 min ± 0.01 ± 0.01 ± 0.01 ± 0.01 ± 0.01 8 h ± 0.03 ± 0.03 ± 0.03 ± 0.05 ± 0.03
Temperature sensitivity
e
(dB) ± 0.25 ± 0.25 ± 0.25 ± 0.25 ± 0.25
SURFACE-EMITTING LED SPECIFICATIONS
a
Model 01C/D 02C/D 12C 12D
Wavelength b(nm) 850 ± 25 1300 +45/—60 850 ± 25 850 ± 25
1300 +45/—60 1300 +45/—6 0
Spectral width (FWHM)
f, g
(nm) 50 145 50/145 50/145 Output power (dBm) * —17/* —14 * —21/* —17 * —18/—22 * —15/—18 Stability d(dB) (D/2)
15 min ± 0.003 ± 0.003 ± 0.005 ± 0.005
8 h ± 0.03 ± 0.03 ± 0.05 ± 0.05 Temperature sensitivity e(dB) ± 0.4 ± 0.4 ± 0.4 ± 0.4 Modulation 270 Hz, 1 kHz, 2 kHz (50 % duty cycle)
IQS-2100 GENERAL SPE CIFICATIONS
Size (H x W x D) 125 mm x 36 mm x 282 mm (4 15/16 in x 1 7/16 in x 11 1/8 in) Weight 0.5 kg (1.1 lb) Temperature
Operating 0 °C to 40 °C (32 °F to 104 °F) Storage —35 °C to 70 °C (—31 °F to 158 °F)
Relative humidity 0 % to 95 % non-condensing
IMPORTANT
The following technical specifications can change without notice. The information presented in this section is provided as a reference only. To obtain this product’s most recent technical specifications, visit the EXFO Web site at www.exfo.com.
Light Source 53
Page 60
Page 61

B SCPI Command Reference

This appendix presents detailed information on the commands and queries supplied with your IQS-2100 Light Source.
IMPORTANT
Since the IQS controllers and expansion units can house many instruments, you must explicitly specify which instrument you want to remotely control.
You must add the following mnemonic at the beginning of any command or query that you send to an instrument (except for IEEE 488.2 and platform commands):
LINStrument<LogicalInstrumentPos>: where <LogicalInstrumentPos> corresponds to the identification number of the instrument.
IQS controller or expansion unit
identification number (for example, 001)
XXXY
Instrument slot number (0 to 9)
For information on modifying unit identification, refer to your platform user guide.
Light Source 55
Page 62
SCPI Command Reference

Quick Reference Command Tree

Quick Reference Command Tree
Command Param eter (s) P.
SNUMber? 57
SOURce[1..n] AM INTernal FREQuency <ModulationFreq[<wsp>HZ]> 58
FREQuency? 59
POWer ATTenuation <Attenuation[<wsp>DB]>|MAXim
ATTenuation? [MAXimum|MINimum|DEFault] 62
STATe <PowerState> 64
STATe? 65
WAVel engt h UPPe r|LO Wer 66
WAVel engt h? 67
WAVelength COUNt? 68
LOWer? 69
UPPer? 70
STATus? 71
um|MINimum|DEFault
60
56 IQS-2100
Page 63
SCPI Command Reference
Product-Specific Commands—Description
Product-Specific Commands—Description
:SNUMber?
Description This query returns a value indicating the serial
number of the module.
Syntax :SNUMber?
Parameter(s) None
Response Syntax <SerialNumber>
Response(s) SerialNumber:
The response data syntax for <SerialNumber> is defined as a <STRING RESPONSE DATA> element.
The <SerialNumber> response represents a string containing the serial number of the module.
Example(s) SNUM? Returns "123456-AB"
Light Source 57
Page 64
SCPI Command Reference
Product-Specific Commands—Description
:SOURce[1..n]:AM:INTernal:FREQuency
Description This command selects the internal modulation
Syntax :SOURce[1..n]:AM:INTernal:FREQuency<wsp>
Parameter(s) ModulationFreq:
frequency. The internal modulation is 50 % duty cycle at the selected frequency.
*RST sets the modulation frequency to 0 Hz (CW).
<ModulationFreq[<wsp>HZ]>
The program data syntax for <ModulationFreq> is defined as a <DECIMAL NUMERIC PROGRAM DATA> element followed by an optional <SUFFIX PROGRAM DATA> element. The allowed <SUFFIX PROGRAM DATA> element is HZ.
The <ModulationFreq> parameter is the new modulation frequency: 270, 1000, 2000, or 0 (CW).
Example(s) SOUR:POW:STAT ON
SOUR:AM:INT:FREQ 2000Hz
See Also SOURce[1..n]:AM:INTernal:FREQuency?
58 IQS-2100
Page 65
SCPI Command Reference
Product-Specific Commands—Description
:SOURce[1..n]:AM:INTernal:
FREQuency?
Description This query returns a value indicating the current
internal modulation frequency. If the source is in CW mode, the function will return 0.
*RST sets the modulation frequency to 0 Hz (CW).
Syntax :SOURce[1..n]:AM:INTernal:FREQuency?
Parameter(s) None
Response Syntax <ModulationFrequency>
Response(s) ModulationFrequency:
The response data syntax for <ModulationFrequency> is defined as a <NR3 NUMERIC RESPONSE DATA> element.
The <ModulationFrequency> response is the internal modulation frequency of the source, in Hz. If the source is in CW mode, the returned value is 0.
Example(s) SOUR:POW:STAT ON
SOUR:AM:INT:FREQ 270 SOUR:AM:INT:FREQ?
See Also SOURce[1..n]:AM:INTernal:FREQuency
Light Source 59
Page 66
SCPI Command Reference
Product-Specific Commands—Description
:SOURce[1..n]:POWer:ATTenuation
Description This command changes the internal attenuation
Syntax :SOURce[1..n]:POWer:ATTenuation<wsp><Att
Parameter(s) Attenuation:
of the source. The source power is at its maximum when the attenuation is set to 0.0 dB.
*RST sets the attenuation to 0 dB.
enuation[<wsp>DB]>|MAXimum|MINimum|D EFault
The program data syntax for <Attenuation> is defined as a <numeric_value> element followed by an optional <SUFFIX PROGRAM DATA> element. The allowed <SUFFIX PROGRAM DATA> element is DB. The <Attenuation> special forms MINimum, MAXimum and DEFault are accepted on input.
MINimum allows to set the instrument to the smallest supported value. MAXimum allows to set the instrument to the greatest supported value.
DEFault allows the instrument to select a value for the <Attenuation> parameter.
The <Attenuation> parameter is the new power attenuation in dB. The power attenuation is always a positive value.
60 IQS-2100
Page 67
SCPI Command Reference
Product-Specific Commands—Description
:SOURce[1..n]:POWer:ATTenuation
Example(s) SOUR:POW:STAT ON
SOUR:POW:ATT 2
Notes Instrument must be in the Ready state to execute
this command.
See Also SOURce[1..n]:POWer:ATTenuation?
Light Source 61
Page 68
SCPI Command Reference
Product-Specific Commands—Description
:SOURce[1..n]:POWer:ATTenuation?
Description This query returns a value corresponding to the
Syntax :SOURce[1..n]:POWer:ATTenuation?[<wsp>MA
Parameter(s) Parameter 1:
internal power attenuation of the source.
*RST sets the attenuation to 0 dB.
Ximum|MINimum|DEFault]
The program data syntax for the first parameter is defined as a <CHARACTER PROGRAM DATA> element. The allowed <CHARACTER PROGRAM DATA> elements for this parameter are: MAXimum|MINimum|DEFault.
MINimum is used to retrieve the instrument's smallest supported value. MAXimum is used to retrieve the instrument's greatest supported value. DEFault is used to retrieve the instrument's default value.
Response Syntax <Attenuation>
62 IQS-2100
Page 69
SCPI Command Reference
Product-Specific Commands—Description
:SOURce[1..n]:POWer:ATTenuation?
Response(s) Attenuation:
The response data syntax for <Attenuation> is defined as a <NR3 NUMERIC RESPONSE DATA> element.
The <Attenuation> response is the power attenuation of the source, in dB.
Example(s) SOUR:POW:STAT ON
SOUR:POW:ATT 3 SOUR:POW:ATT?
See Also SOURce[1..n]:POWer:ATTenuation
Light Source 63
Page 70
SCPI Command Reference
Product-Specific Commands—Description
Description This command turns the optical source on or off.
Syntax :SOURce[1..n]:POWer:STATe<wsp><PowerStat
Parameter(s) PowerState:
:SOURce[1..n]:POWer:STATe
When source is on, the red LED (Active) on the front of the instrument lights up.
*RST sets the optical source to OFF.
e>
The program data syntax for <PowerState> is defined as a <Boolean Program Data> element. The <PowerState> special forms ON and OFF are accepted on input for increased readability. ON corresponds to 1 and OFF corresponds to 0.
The <PowerState> parameter is the new power state of the source.
Example(s) SOUR:POW:STAT ON
Notes Instrument must be in the Ready state to execute
this command.
See Also SOURce[1..n]:POWer:STATe?
64 IQS-2100
Page 71
SCPI Command Reference
Product-Specific Commands—Description
:SOURce[1..n]:POWer:STATe?
Description This query returns a value indicating the state of
the optical source (on or off).
*RST sets the optical source to OFF.
Syntax :SOURce[1..n]:POWer:STATe?
Parameter(s) None
Response Syntax <PowerState>
Response(s) PowerState:
The response data syntax for <PowerState> is defined as a <NR1 NUMERIC RESPONSE DATA> element.
The <PowerState> response corresponds to the power state of the source, as follows:
0, the optical source is off. 1, the optical source is on.
Example(s) SOUR:POW:STAT OFF
SOUR:POW:STAT?
See Also SOURce[1..n]:POWer:STATe
Light Source 65
Page 72
SCPI Command Reference
Product-Specific Commands—Description
:SOURce[1..n]:POWer:WAVelength
Description This command selects a wavelength when using
Syntax :SOURce[1..n]:POWer:WAVelength<wsp>UPPe
Parameter(s) SelectedSources:
a dual-wavelength source instrument.
*RST sets the selected wavelength to the LOW value.
r|LOWer
The program data syntax for the first parameter is defined as a <CHARACTER PROGRAM DATA> element. The allowed <CHARACTER PROGRAM DATA> elements for this parameter are: UPPer|LOWer.
UPPer, switches to the highest wavelength. LOWer, switches to the lowest wavelength.
Example(s) SOUR:POW:WAV LOW
Wai t 3 s e co n ds . SOUR:POW:STAT ON
Notes Instrument must be in the Ready state to execute
this command. This command can cause the instrument to enter the Stabilizing state.
See Also SOURce[1..n]:POWer:WAVelength?
SOURce[1..n]:POWer:WAVelength:LOWer? SOURce[1..n]:POWer:WAVelength:UPPer?
66 IQS-2100
Page 73
SCPI Command Reference
Product-Specific Commands—Description
:SOURce[1..n]:POWer:WAVelength?
Description This query returns a value indicating which
wavelength is currently selected.
*RST sets the selected wavelength to the LOW value.
Syntax :SOURce[1..n]:POWer:WAVelength?
Parameter(s) None
Response Syntax <SelectedSources>
Response(s) SelectedSources:
The response data syntax for <SelectedSources> is defined as a <CHARACTER RESPONSE DATA> element.
LOWer, the lower wavelength is activated. UPPer, the upper wavelength is activated.
Example(s) SOUR:POW:WAV?
See Also SOURce[1..n]:POWer:WAVelength
SOURce[1..n]:POWer:WAVelength:LOWer? SOURce[1..n]:POWer:WAVelength:UPPer?
Light Source 67
Page 74
SCPI Command Reference
Product-Specific Commands—Description
:SOURce[1..n]:POWer:WAVelength:
Description This query returns the number of available
Syntax :SOURce[1..n]:POWer:WAVelength:COUNt?
Parameter(s) None
Response Syntax <NbWavelength>
Response(s) NbWavelength:
COUNt?
wavelengths on the instrument.
*RST has no effect on this command.
The response data syntax for <NbWavelength> is defined as a <NR1 NUMERIC RESPONSE DATA> element.
The <NbWavelength> response is the number of wavelengths available on the instrument:
1, one wavelength available 2, two wavelengths available
Example(s) SOUR:POW:WAV:COUN?
68 IQS-2100
Page 75
SCPI Command Reference
Product-Specific Commands—Description
:SOURce[1..n]:POWer:WAVelength:
LOWer?
Description This query returns a value indicating the lower
wavelength.
*RST has no effect on this command.
Syntax :SOURce[1..n]:POWer:WAVelength:LOWer?
Parameter(s) None
Response Syntax <LowerWavelength>
Response(s) LowerWavelength:
The response data syntax for <LowerWavelength> is defined as a <NR3 NUMERIC RESPONSE DATA> element.
The <LowerWavelength> response is the lowest source wavelength value in meters.
Example(s) SOUR:POW:WAV:LOW?
Notes If you do not have a dual-wavelength instrument,
use this function to determine the wavelength of your source.
See Also SOURce[1..n]:POWer:WAVelength
SOURce[1..n]:POWer:WAVelength? SOURce[1..n]:POWer:WAVelength:UPPer?
Light Source 69
Page 76
SCPI Command Reference
Product-Specific Commands—Description
:SOURce[1..n]:POWer:WAVelength:
Description This query returns a value indicating the upper
Syntax :SOURce[1..n]:POWer:WAVelength:UPPer?
Parameter(s) None
Response Syntax <UpperWavelength>
Response(s) UpperWavelength:
UPPer?
wavelength.
*RST has no effect on this command.
The response data syntax for <UpperWavelength> is defined as a <NR3 NUMERIC RESPONSE DATA> element.
The <UpperWavelength> response is the highest source wavelength value in meters.
Example(s) SOUR:POW:WAV:UPP?
Notes Not available on single-source instruments.
See Also SOURce[1..n]:POWer:WAVelength
SOURce[1..n]:POWer:WAVelength? SOURce[1..n]:POWer:WAVelength:LOWer?
70 IQS-2100
Page 77
SCPI Command Reference
Product-Specific Commands—Description
:STATus?
Description This query returns a value indicating the status of
the module (READY, BUSY, etc.).
Syntax
Parameter(s) None
Response Syntax
Response(s) Status:
Example(s) STAT? Returns READY (Module is ready.)
Light Source 71
Page 78
Page 79

Index

Index
A
About function tab ..................................... 45
activating the light source........................... 27
active LED...................................................... 1
after-sales service ........................................ 45
application
contacting EXFO support from .............. 45
exiting ................................................... 16
main window description ................ 11, 12
setting up .............................................. 17
starting, single-module.......................... 11
viewing results....................................... 28
see also light source
attenuation setting, light source ................. 20
B
Busy, module status .................................... 13
C
calibration
certificate............................................... 40
interval .................................................. 40
caution
of personal hazard................................... 3
of product hazard.................................... 3
certification information ................................v
cleaning
EUI connectors....................................... 38
fiber ends .............................................. 25
fixed connectors .................................... 36
front panel............................................. 35
UI connectors ........................................ 38
closing monitor window ............................. 34
configuration
recall...................................................... 23
save ....................................................... 23
connectors, cleaning ............................. 36, 38
contact information, EXFO ..........................45
conventions, safety........................................ 3
customer service.................................... 45, 51
D
deactivating the light source ....................... 27
E
equipment returns....................................... 51
EUI
baseplate ...............................................26
connector adapter ................................. 26
EUI connectors, cleaning .............................38
EXFO service centers.................................... 52
EXFO support e-mail.................................... 45
EXFO universal interface. see EUI
EXFO Web site ............................................. 45
exiting application....................................... 16
F
fiber ends, cleaning ..................................... 25
firmware version, module............................ 45
front panel description, light source.............. 1
front panel, cleaning ................................... 35
H
help. see online user guide
I
identification label....................................... 45
identification, slot ....................................... 13
inserting a module ........................................ 7
IQS-2100 Light Source. see light source
Light Source 73
Page 80
Index
L
label, identification ..................................... 45
LabVIEW drivers ............................................ 2
LED push button
location ................................................... 1
problem with......................................... 43
light source
activating/deactivating .......................... 27
available models ...................................... 2
connecting optical fiber......................... 25
front panel .............................................. 1
operating............................................... 25
operating with QuickTools..................... 33
selecting a wavelength .................... 17, 18
setting the attenuation.......................... 20
setting up .............................................. 17
M
main window, application..................... 11, 12
maintenance
EUI connectors....................................... 38
fixed connectors .................................... 36
front panel ............................................ 35
general information............................... 35
UI connectors ........................................ 38
modulation selection, light source .............. 22
module
insertion .................................................. 7
monitoring ............................................ 29
removal ................................................... 7
status..................................................... 13
see also light source module information
firmware version number ...................... 45
module identification number ............... 45
serial number......................................... 45
module position.......................................... 13
monitor window
closing ................................................... 34
description............................................. 29
opening ................................................. 30
monitoring modules.................................... 29
mounting EUI connector adapter ................ 26
O
online user guide......................................... 44
opening monitor window............................ 30
operating the light source ..................... 25, 33
P
PDF. see online user guide
port, source ................................................... 1
position, module ......................................... 13
product
identification label ................................. 45
specifications ......................................... 53
Q
QuickTools utility ................................... 29, 32
R
Ready, module status .................................. 13
recalibration ................................................ 40
recalling a configuration ............................. 23
remote control
methods .................................................. 2
SCPI commands ..................................... 55
removing a module ....................................... 7
results, viewing............................................ 28
retaining screw knob location .......................1
return merchandise authorization (RMA) .... 51
S
safety
caution .................................................... 3
conventions .............................................3
warning ...................................................3
saving configuration.................................... 23
selecting
modulation, light source........................ 22
74 IQS-2100
Page 81
wavelength, light source ....................... 18
serial number, module ................................ 45
service and repairs....................................... 51
service centers............................................. 52
setting the attenuation, light source ........... 20
setting up the light source .......................... 17
shipping to EXFO ........................................ 51
slot number................................................. 13
software. see application
source port.................................................... 1
source. see light source
specifications, product ................................ 53
status bar .................................................... 13
storage requirements .................................. 35
symbols, safety.............................................. 3
T
technical specifications................................ 53
technical support ........................................ 45
temperature for storage.............................. 35
title bar ....................................................... 13
transportation requirements ................. 35, 47
Index
U
UI connectors, cleaning............................... 38
unit recalibration......................................... 40
user guide. see online user guide
V
viewing results, using light source .............. 28
W
warranty
certification ........................................... 50
exclusions .............................................. 50
general .................................................. 49
liability................................................... 49
wavelength selection, light source .............. 18
window, impossible to open ....................... 43
Light Source 75
Page 82
P/N: 1052814
www.EXFO.com · info@exfo.com
CORPORATE HEADQUARTERS 400 Godin Avenue Quebec (Quebec) G1M 2K2 CANADA
EXFO AMERICA 3701 Plano Parkway, Suite 160 Plano TX, 75075 USA
EXFO EUROPE Omega Enterprise Park,
EXFO ASIA-PACIFIC 151 Chin Swee Road
TOLL-FREE (USA and Canada) 1 800 663-3936
© 2008 EXFO Electro-Optical Engineering Inc. All rights reserved. Printed in Canada (2008-05)
Electron Way
#03-29, Manhattan House
Tel.: 1 418 683-0211 · Fax: 1 418 683-2170
Tel.: 1 972 907-1505 · Fax: 1 972 836-0164
Chandlers Ford, Hampshire S053 4SE ENGLAND Tel.: +44 2380 246810 · Fax: +44 2380 246801
SINGAPORE 169876 Tel.: +6563338241 · Fax: +6563338242

Loading...