delay in display of variations in temperature detected by the probes
MAIN REGULATOR
working setpoint differential
minimum working setpoint
maximum working setpoint
locking of working setpoint calibration (using the procedure described in paragraph 5.2)
1 = YES
r40.099.0 °C/°F (1) not avail. not avail.0.00.0
r50.099.0 °C/°F (1)0.00.00.00.0
r60240 min30303030
r70.099.0 °C/°F (1) not avail. 10.010.010.0
increase in temperature during Energy Saving function; see also i5, i10, HE1 and HE2
decrease in temperature during Overcooling function; see also r6
duration of Overcooling function; see also r5
minimum difference “cell temperature - working setpoint” (when the instrument switches on) such as to provoke the exclusion of the
consequent value of the evaporator temperature among the ones used for the calculation of the relative average (for the defrost activation;
only if d8 = 3); also look at d17 (3)
PAR. MIN. MAX. UMEVX201 EVX203 EVX204/5 EVX214/5
C00240min 0000
C10240min 5555
C20240min 3333
C30240sec 0000
C40240 min10101010
C50240 min10101010
C60.0199.0 °C/°F (1) not avail. not avail. 80.080.0
C70.0199.0 °C/°F (1) not avail. not avail. 90.090.0
C801 5m innot avail. not avail.11
C10 09.999 h rnot avail.000
COMPRESSROR PROTECTION SYSTEM
delay in switching on of compressor after the insturment switches on (3) minimum time between two consecutive compressor start-ups;
also delay in compressor start-up after conclusion of cell probe error(code “Pr1”) (4) (5)
minimum duration of compressor switch off time (4)
minimum duration of compressor switch on time
duration of compressor switch off during cell probe error (code “Pr1”); see also C5
duration of compressor switch on during cell probe error (code “Pr1”); see also C4
condenser temperature is higher than that at which the condenser overheating alarm is activated (code “COH”) (6)
condenser temperature is higher than the limit at which the compressor blocked alarm is activated (code “CSd”)
compressor alarm delay locked (code “CSd”) (7)
number of operating hours is higher than the limit at which the need for maintenance is signaled.
0 = function absent
PAR. MIN. MAX. UMEVX201 EVX203 EVX204/5 EVX214/5
d0099hr 8888
DEFROSTING
if d8 = 0, 1 or 2, defrosting interval (8)
0 = interval defrosting will never be activated
if d8 = 3, maximum defrost interval
d102- - - -not avail.000
type of defrosting
0=
1=
2=
d2-99.0 99.0 °C/ °F (1) not avail.2. 02.02. 0
d3099min30303030
temperature at end of defrosting (only if P3 = 1); see also d3
se P3 = 0 or 2, defrosting duration
se P3 = 1, maximum defrosting duration; see also d2
0 = defrosting will not be activated
d401- - - - 0000
defrosting when instrument is switched on (only if d8 = 0, 1, 2 or 3) (3)
1 = YES
d5099min 0000
if d4 = 0, minimum time between switching on of instrument and activation of defrosting; see also i5 (3)
if d4 = 1, delay in activation of defrosting after instrument is switched on ; see also i5 (3)
d601- - - - 1111
temperature displayed during defrosting
0 = cell temperature
1 = if at the time of defrosting activation, the cell temperature is lower than the “working setpoint + r0”, at most “working
d701 5m innot avail.222
dripping duration (during dripping the compressor will remain switched off and the defrosting output will remain disactivated; if
d16 = 0, evaporator fan activity will depend on parameter F2; if d16 ≠ 0, the evaporator fan will remain switched off)
d804- - - - 0000
defrosting activation methods
0=
1=
2=
3=
4=IN REAL TIME - defrosting will be activated at the times established in parameters Hd1 ... Hd6 (visible in EVX214 and EVX215 only)
d9-99.0 99.0 °C/ °F (1) not avail.0. 00.00. 0
d1 1 01- - - -not avail.000
evaporator temperature is higher than that at which the defrost interval counter is suspended (only if d8 = 2)
defrosting alarm switches off once maximum time limit has been reached (code “dFd”; only if P3 = 1
and in absence of an evaporator probe (code “Pr2”)
1 = YES
d1 5 09 9mi nnot avail.000
d1 6 09 9mi nnot avail.000
minimum time that the compressor must be switched on before defrosting can be activated (only if d1 = 1) (11)
predripping duration (during predripping the compressor will remain switched off, the defrosting output will be activated and the
evaporator fan will remain switched off)
ELECTRIC - during defrosting the compressor will remain off and the defrosting output will be activated; evaporator fan activity will
depend on parameter F2
BY HOT GAS - during defrosting the compressor will be switched on and the defrosting output will be activated; evaporator fan activity
will depend on parameter F2
VIA STOPPING OF COMPRESSOR - during defrosting the compressor will remain switched off and the defrosting output will remain
disactivated; evaporator fan activity will depend on parameter F2
setpoint + r0”; if at the time of defrosting activation, the cell temperature is higher than the “working setpoint + r0”,
at most the cell temperature when defrosting is activated
AT INTERVALS - defrosting will be activated once the instrument has altogether been running for time d0
AT INTERVALS - defrosting will be activated once the compressor has altogether been switched on for time d0
AT INTERVALS - defrosting will be activated once the evaporator temperature has altogether been below temperature d9 for time d0
(visible in EVX203, EVX204, EVX214, EVX205 and EVX215 only) (10)
ADAPTABLE - defrosting will be activated when one of the following conditions is present (visible in EVX203, EVX204, EVX214, EVX205
and EVX215 only; also look at d0) (10):
- condition 1: the evaporator temperature will be below temperature d22 and the compressor will altogether be
switched on for time d18
- condition 2: the evaporator temperature will fall below temperature d19
Evco S.p.A. • Code 104X201E314 • page 6/8
d1 7 110- - - -not avail.111
number of evaporator temperature values used for the calculation of the relative average (for the defrost activation; only if d8 = 3); also look
at r7, i11 and i12
d18 03,000 minnot avail.404040
defrosting interval (only if d8 = 3 and for condition 1)
0 = defrosting for condition 1 will never be activated
d190.040.0 °C/°F (1) not avail.3 .03. 03.0
evaporator temperature above which the defrost is activated (relative to the evaporator temperatures average, or “evaporator temperatures
average - d19”) (only if d8 = 3 and for condition 2); also look at d17
d2 0 05 00minnot avail.18 018 018 0
minimum consecutive time the compressor must be switched on such as to provoke the defrost activation
0 = the defrost will never be activated because the compressor has been switched on
d2 1 05 00minnot avail.20 020 020 0
minimum consecutive time the compressor must be switched on after the insturment switches on (on condition that the difference “cell
temperature - working setpoint” is higher temperature r7) and after function Overcooling is activated such as to provoke the defrost
activation
0 = the defrost will never be activated because the compressor has been switched on
d220.010.0 °C/°F (1) not avail.2 .02. 02.0
evaporator temperature above which the defrosting interval is suspended (relative to the evaporator temperatures average, or “evaporator
temperatures average + d22”) (only if d8 = 3 and for condition 1); also look at d17
d230.010.0 °C/°F (1) not avail.1 .01. 01.0
PAR. MIN. MAX. UMEVX201 EVX203 EVX204/5 EVX214/5
A001- - - -not avail.000
evaporator temperatures average increase during function Energy Saving (for defrost activation; only if d8 = 3); also look at d17
TEMPERATURE ALARMS
temperature associated with the minimum temperature alarm (code “AL”)
0 = cell temperature
1 = evaporator temperature (12)
A1-99.0 99.0 °C/ °F (1)-10.0-10.0-10.0-10.0
A202- - - - 1111
temperature below that at which the minimum temperature alarm is activated (code “AL”); see also A0, A2 and A11
type of minimum temperature alarm (code “AL”)
0 = alarm absent
1 = relative to working setpoint (that is “working setpoint - A1”; consider A1 without sign)
2 = absolute (that is A1)
A4-99.0 99.0 °C/ °F (1)10.010.010.010.0
A502- - - - 1111
temperature higher than that at which the maximum temperature alarm is activated (code “AH”); see also A5 and A11
type of maximum temperature alarm (code “AH”)
0 = alarm absent
1 = relative to working setpoint (that is “working setpoint + A4”; consider A1 without sign))
2 = absolute (that is A4)
A60240 min120120120120
A70240 min15151515
A80240 min15151515
delay in maximum temperature alarm (code “AH”) after the instrument is switched on (3)
temperature alarm delay (code “AL” and code “AH”)
delay in maximum temperature alarm (code “AH”) following the conclusion of defrosting (in EVX201 only) and following the conclusion
of evaporator fan (in EVX203, EVX204, EVX214, EVX205 and EVX215 only) (13)
A90240 min15151515
A1 0 02 40minnot avail. not avail. not avail.1
delay in maximum temperature alarm (code “AH”) following the disactivation of the door microswitch input (14)
duration of interruption in the power supply that occurs when the instrument has been running for long enough to cause the storage of
the power interruption alarm when the power supply is restored. (code “PF”) (15)
A110.115.0 °C/°F (1)2.02.02.02.0
A1 2 02- - - -not avail. not avail. not avail.1
differential of parameters A1and A4
kind of signal for power interruption alarm (code “PF”); also look at A10
0 = the alarm will not be signalled
1 = the display will show the code “PF” flashing and the buzzer will be activated
2 = the display will show the code “PF” flashing and the buzzer will be activated (this last on condition that the power interruption duration
PAR. MIN. MAX. UMEVX201 EVX203 EVX204/5 EVX214/5
F005- - - -not avail.111
EVAPORTOR FAN
evaporator fan activity during normal operation
0 = switched off
1 = switched on; see also F13, F14 and i10
2 = in parallel with the compressor; see also F9, F13, F14 and i10
3 = dependent on F1 (16)
4 = switched off if the compressor is switched off, dependent on F1 if the compressor is switched on; see also F9 (16)
5 = dependent on F6; see also F9
F1-99.0 99.0 °C/°F (1) not avail.-1.0-1.0-1.0
F202- - - -not avail.000
evaporator temperature above the limit at which the evaporator fan is switched off (only if F0 = 3 or 4); see also F8
evaporator fan activity during defrosting and dripping
0 = switched off
1 = switched on (setting parameter d7 to 0 is recommended)
2 = dependent on F0
F301 5m innot avail.222
maximum duration of evaporator fan disactivation; see also F7 (during evaporator fan desactivation the compressor can be switched on,
the defrosting output will remain disactivated and the evaporator fan will remain switched off)
F402 40secnot avail.6 0606 0
time duration that evaporator fan is switched off during operation for a low percentage of relative humidity when the compressor is
switched off; see also F5 (only if F0 = 5)
F502 40secnot avail.1 0101 0
time duration that evaporator fan is switched on during operation for a low percentage of relative humidity when the compressor is
switched off; see also F4 (only if F0 = 5)
F601- - - -not avail.000
operation for low or high percentage of relative humidity (only if F0 = 5) (17)
0=
1=HIGH RELATIVE HUMIDITY - the evaporator fan will always be switched on
F7-99.0 99.0 °C/°F (1) not avail.5 .05. 05.0
evaporator temperature below limit at which the evaporator fan is disactivated (relative to working setpoint, that is “working setpoint + F7”);
see also F3
F80.115.0 °C/°F (1) not avail.2 .02. 02.0
F902 40secnot avail. not avail.00
F110.099.0 °C/°F (1) not avail. not avail.15.015.0
parameter F1differential
delay in the switching off of evaporator fan following the switching off of the compressor (only if F0 = 2, 4 and 5)
condenser temperature above that at which the condenser fan is switched on (“F11 + 2.0 °C/4 °F, only if u1 and/or u11 = 6 and provided
the compressor is switched on); see also F12 (18)
F12024 0 secnot avail. not avail.3 03 0
F13024 0 m innot avail.555
F14024 0 m innot avail.555
PAR. MIN. MAX. UMEVX201 EVX203 EVX204/5 EVX214/5
i005- - - - 1233
delay in switching off of the condenser fan following the switching off of the condenser (only if u1 and/or u11 = 6); see also F11
time the evaporator fan remains turned off during function Energy Saving; see also F14 and i10 (only if F0 = 1 or 2)
time the evaporator fan remains turned on during function Energy Saving; see also F13 and i10 (only if F0 = 1 or 2)
DIGITAL INPUTS
effect caused by the activation of the door microswitch input; see also i4
0 = no effect
1 = the compressor and evaporator fan (evaporator fan in EVX203, EVX204, EVX214, EVX205 and EVX215 only) will be switched off (at
2 = the evaporator fan will be switched off (at most for time i3 or until the input is disactivated) (visible in EVX203, EVX204, EVX214, EVX205
3 = the cell light will be switched on (only if u1 and/or u11 = 0, until the input is disactivated) (only visible in EVX204, EVX214, EVX205
4 = the compressor and evaporator fan will be switched off (at most for time i3 or until the input is disactivated) and the cell light will be
5 = the evaporator fan will be switched off (at most until time i3 or until the input has been disactivated) and the cell light will be switched
i101- - - - 0000
type of door microswitch input contact
0 = normally open (active input with closed contact)
1 = normally closed (active input with open contact)
i2-1120 min30303030
delay in signaling of door microswitch input alarm (code “id”)
-1 = the alarm will not be signaled
is higher than time A10)
LOW RELATIVE HUMIDITY - the evaporator fan will operate in parallel with the compressor; see also F4 and F5
most for time i3 or until the input is disactivated (19)
and EVX215 only)
and EVX215)
switched on (only if u1 and/or u11 = 0, until the input is disactivated) (only visible in EVX204, EVX214, EVX205 and EVX215) (19)
on (only if u1 and/or u11 = 0, until the input is disactivated) (only visible in EVX204, EVX214, EVX205 and EVX215)
Evco S.p.A. • Code 104X201E314 • page 7/8
i3-1120 min15151515
maximum duration of the effect caused by activation of the door microswitch on the compressor and the evaporator fan (the evaporator
fan in EVX203, EVX204, EVX214, EVX205 and EVX215 only)
-1 = the effect will last until the input is disactivated
i401- - - - 0000
storage of door microswitch input alarm (code “id”) (20)
1 = YES
i506- - - -not avail. not avail.22
effect caused by the activation of the multifunction input
0 = no effect
1=
2=
3=
4=
5=
6=SWITCHING OFF THE INSTRUMENT - the instrument will be switched off (until the input is disactivated)
i601- - - -not avail. not avail.00
type of mutlifunction input contact
0 = normally open (active input with closed contact)
1 = normally closed (active input with open contact)
i701 20minnot avail. not avail.00
if i5 = 3, multifunction input alarm delay (code “iA”)
if i5 = 4, delay in compressor switching on after the disactivation of the multifunction input (21)
i801 5- - - -not avail. not avail.00
number of multifunction input alarms (code “iA”) such to cause a pressure switch alarm (code “iSd”) (if i5 = 4)
0 = alarm absent
i919 99minnot avail. not avail.2402 40
i1 009 99 mi nnot avail.000
time that must pass in absence of mutlifunction output alarms (code “iA”) so that the alarm counter is reset (only if i5 = 4)
time without activations of the door switch input (on condition that the cabinet temperature has reached the working setpoint) in order
that function Energy Saving is activated automatically (it has effect on the evaporator fan only if F0 = 1 or 2)
0 = the function will never automatically be activated
i1 102 40 snot avail.1 5151 5
minimum time the door switch input must be activated such as to provoke the exclusion of the consequent value of the evaporator
temperature among the ones used for the calculation of the relative average (for the defrost activation; only if d8 = 3); also look at d17
i1 202 40 snot avail.6 0606 0
minimum time the door switch input must be activated altogether such as to provoke the exclusion of the consequent value of the
evaporator temperature among the ones used for the calculation of the relative average (for the defrost activation; only if d8 = 3); also look
at d17
i1 3024 0 - - - -not avail.18 01 80180
number of door switch input activations such as to provoke the defrost activation
0 = the defrost will never be activated because of the door switch input activation
i1 402 40 mi nnot avail.3 23 23 2
minimum duration of the door switch imput activation such as to provoke the defrost activation
0 = the defrost will never be activated because of the door switch input activation
PAR. MIN. MAX. UMEVX201 EVX203 EVX204/5 EVX214/5
u106- - - -not avail. not avail.00
DIGITAL OUTPUTS
operation controlled by fourth output (22)
0=
1=
2=
3=
4=
5=
6=CONDENSER FAN - in this case parameters P4, F11 and F12 will be activated
u201- - - -not avail. not avail.00
enabling of manual switch on/switch off of the cell light or the auxiliary output when the instrument is switched off
(only if u1 and/or u11 = 0 or 2) (23)
1 = YES
u401- - - -not avail. not avail.11
enabling of alarm output disactivation with the silencing of the buzzer (only if u1 and/or u11 = 3)
1 = YES
u5-99.0 99.0 °C/°F (1) not avail. not avail.-1.0-1.0
u611 20minnot avail. not avail.55
u70.099.0 °C/°F (1) not avail. not avail.2. 02.0
cell temperature below that at which the door resistors are switched on (“u5 - 2.0 °C/4 °F, only if u1 and/or u11 = 4) (6)
operating time of demistor resistors (only if u1 and/or u11 = 1)
cell temperature below that at which the evaporator valve is disactivated (relating to the working setpoint, that is “working setpoint + u7”)
(only if u1 and/or u11 = 5) (6)
u801- - - -not avail. not avail.00
type of evaporator valve contact (only if u1 and/or u11 = 5)
0 = normally open (valve active with contact closed)
1 = normally closed (valve active with contact open)
u901- - - - 1111
enabling of buzzer
1 = YES
u1 1 06- - - -not avail. not avail.33
operation controlled by fifth output (22)
0=
1=
2=
3=
4=
5=
6=CONDENSER FAN - in this case parameters P4, F11 and F12 will be activated
PAR. MIN. MAX. UMEVX201 EVX203 EVX204/5 EVX214/5
HE1 00:00 23:59 hr:minnot avail. not avail. 00:0000:00
HE2 00:00 23:59 hr:minnot avail. not avail. 00:0000:00
ENERGY SAVING IN REAL TIME
time of activation of the Energy Saving in real time function; see also r4 and HE2
duration of the Energy Saving in real time function; see also r4 and HE1
00:00= the Energy Saving in real time function will not be activated
SYNCHRONISATION OF DEFROSTING PERIODS - once time d5 has passed defrosting will be activated
ACTIVATION OF ENERGY SAVING FUNCTION - the Energy Saving function will be activated (until the input is disactived), provided the
Overcooling function is running; see also r4
ACTIVATION OF MULTIFUNCTION INPUT ALARM - once time i7 has passed the display will show the flashing code “iA” and the buzzer
will be activated (until the input is disactivated)
ACTIVATION OF THE PRESSURE SWITCH ALARM - the compressor will be switched off, if u1 and/or u11 = 6 the condenser fan will be
switched on, the display will show the flashing code “iA” and the buzzer will be activated (until the input is disactivated): when the
input has been activated the number of times established with parameter i8 the regulators will be switched off, if u1 and/or u11 = 6
the condenser fan will be switched on, the display will show the flashing code “iSd” and the buzzer will be activated (until the input
is disactivated and the instrument is switched off and re-started or until the power supply is interrupted); see also i7 and i9
SWTICHING ON THE AUXILIARY OUTPUT - the auxiliary output will be switched on (only if u1 and/or u11 = 2, until the input is
disactivated)
CELL LIGHT - in this case the AUXILIARY key and parameters i0 and u2 will be activated
DEMISTER RESISTORS - in this case the AUXILIARY key and parameter u6 will be activated
AUXILIARY OUTPUT - in this case the AUXILIARY key and parameters i5 and u2 will be activated
ALARM OUTPUTS - in this case parameter u4 will be activated
DOOR RESISTORS - in this case parameter u5 will be activated
EVAPORATOR VALVE - in this case parameters u7 and u8 will be activated
CELL LIGHT - in this case the AUXILIARY key and parameters i0 and u2 will be activated
DEMISTER RESISTORS - in this case the AUXILIARY key and parameter u6 will be activated
AUXILIARY OUTPUT - in this case the AUXILIARY key and parameters i5 and u2 will be activated
ALARM OUTPUTS - in this case parameter u4 will be activated
DOOR RESISTORS - in this case parameter u5 will be activated
EVAPORATOR VALVE - in this case parameters u7 and u8 will be activated
Evco S.p.A. • Code 104X201E314 • page 8/8
(1) the unit of measurement depends on P2
(2) Properly set the parameters corresponding to the regulators after modifying parameter P2
(3) the parameter has effect even after an interruption in the power supply that occurs while the instrument is switched on
(4) the time established with the parameter is counted even when the instrument is switched off
(5) if parameter C1 is set to 0, the delay after the end of the cell probe error will be 2 min
(6) the parameter differential is 2.0 °C/4 °F
(7) if when the instrument is switched on, the condenser temperature is already above that established in parameter C7, then parameter C8 will not have effect
(8) the instrument stores the defroster interval count every 30 min; the modification of parameter d0 takes effect following the end of the preceeding interval or following the activation of manual defrosting.
(9) the display returns to normal operation when, at the end of defrosting (EVX201 only) or at the end of evaporator fan disactivation (in EVX203, EVX204, EVX214, EVX205 and EVX215 only), the cell temperature falls
below that at which the display was initally blocked (or if a temperature alarm is signaled)
(10) if parameter P3 is set to 0 or 2, the instrument will function as if parameter d8 were set to 0
(11) if when defrosting is activated, the operating duration of the compressor is less than the time established with parameter d5, the compressor will remain on for the amount of time necessary to compelete defrosting.
(12) if parameter P3 is set to 0, the instrument will function as if parameter A0 were set to 0 but it will not store the alarm
(13) during defrosting and dripping and when the evaporator fan is stopped, the temperature alarms are absent, provided that these were signaled after the activation of defrosting
(14) during activation of the door microswitch input, the maximum temperature alarm is absent, provided the alarm was signaled after the activation of the input
(15) when power is restored, the alarm will always be signaled
(16) if parameter P3 is set to 0, the instrument will function as if parameter F0 were set to 2
(17) the parameter can also be modified using the procedure described in paragraph 4.8
(18) if parameter P4 is set to 0, the condenser fan will function in parallel with the compressor
(19) the compressor is switched off 10 sec after the activation of the input; if the input is activated during defrosting or when the evaporator fan is disactived, the activation will not have any effect on the compressor
(20) the instrument stores the alarm once the time established in parameter i2 has expired; if parameter i2 is set to -1, the instrument will not store the alarm
(21) make sure that the time established with parameter i7 is less than that established with parameter i9
(22) to avoid damaging the unit connected to the instrument, change the parameter setting when the instrument is switched off
(23) if parameter u2 is set to 0, switching off the instrument may cause the cell light and/or the auxiliary output to switch off (the next time the instrument is switched on the unit connected will remain switched off);
if parameter u2 is set to 1, switching off the instrument will not cause the cell light and/or the auxiliary output to switch off (the next time the instrument is switched on the unit connected will remain switched on).
13ELECTRIC CONNECTION
13.1 Preliminary notes
With reference to the electrical wiring diagrams:
• the unit connected to and operated by the fourth outpt depends on parameter u1 (EVX204, EVX214, EVX205 and EVX215 only)
• the unit connected to and operated by the fifth outpt depends on parameter u11 (EVX205 and EVX215 only)
• the serial port is the port for communicating with the monitoring system (via serial interface, via TTL, MODBUS communications protocol) or by programming key ;
the port must not be used for two different purposes
simultaneously.
13.2 EVX201 electrical wiring
13.3 EVX203 electrical wiring
13.4 EVX204, EVX214, EVX205 and EVX215 electrical wiring
13.5 Electrical wiring warnings
• do not use electric or pneumatic screwers on the terminal boards
• if the instrument was brought from a cold place to a hot one, humidity may condense inside the instrument; wait approximately one hour before switching on the power
• make sure that the power tension, frequency, and the electric operating power of the instrument are compatible with the those of the local power voltage
• disconnect the power before performing any type of maintenance operation
• do not use the instrument as a safety device
• for information about the instrument and repairs contact a member of the Evco sales network
EVCO S.p.A.This document belongs to Evco; unless you are authorized by Evco, you can not publish this document.
Via Mezzaterra 6, 32036 Sedico Belluno ITALYEvco does not take any responsibility about features, technical data and possible mistakes related in this document.
Phone +39-0437-852468 • Fax +39-0437-83648Evco does not take any responsibility about damages coming by the non-observance of additional information.
info@evco.it • www.evco.itEvco reserves the right to make any change without prior notice and at any time without prejudice the basic safety and operating features.
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.