1Getting started .....................................................................................................................................................9
1.1Installing the instrument ..................................................................................................................................... 9
Important ............................................................................................................................................................. 5
4.2LED Communication (on port 1) ...................................................................................................................... 23
5Technical data ...................................................................................................................................................25
5.1Technical data ...................................................................................................................................................25
Important
Important
Read these instructions carefully before installing and using the instrument and follow all additional information for
installation and electrical connection; keep these instructions close to the instrument for future consultations.
The instrument must be disposed according to the local legislation about the collection for electrical and electronic
equipment.
5EVD120
Important
[ EVD120 ]
Introduction
[ EVD120 ]
Introduction
EVD120 is a two analog channels and two digital inputs acquisition module.
Each analog channel can work with one of the following kind of probe (universal analog channels):
•PTC/NTC probe
•J/K thermocouple
•2/3 wires Pt 100 and Pt 1000 probe
•2/3 wires 4-20 mA passive current transducer
•0-10/2-10 V and 0-20/4-20 mA active transducer.
The instrument has the following communication ports:
•RS-485 port with MODBUS communication protocol, to allow the connection of the instrument in RS-485 net-
works with MODBUS communication protocol
•serial port for the communication with the programming key EVKEY.
The case of EVD120 is 3 DIN modules.
7EVD120
Introduction
Getting started
[ EVD120 ]
1Getting started
1.1Installing the instrument
On DIN rail; dimensions in mm (in).
9EVD120
Getting started
Fig. 1 - Dimensions of EVD120.
Fig. 2 - Installation of EVD120.
Additional information for installation:
•working conditions (working temperature, humidity, etc.) must be between the limits related in the technical data
•do not install the instrument close to heating sources (heaters, hot air ducts, etc.), devices provided with big magne-
tos (big speakers, etc.), locations subject to direct sunlight, rain, humidity, dust, mechanical vibrations or bumps
•according to the safety legislation, the protection against electrical parts must be ensured by a correct installation of
the instrument; the parts that ensure the protection must be installed so that you can not remove them if not by using
a tool.
10
EVD120
Getting started
1.2Wiring diagram
Fig. 3 - Wiring diagram of EVD120.
With reference to the wiring diagram:
•port 1 is the port for the serial communication (via RS-485, with MODBUS communication protocol)
•port 2 is the serial port for the communication with the programming key.
Additional information for electrical connection:
•do not operate on the terminal blocks with electrical or pneumatic screwers
•if the instrument has been moved from a cold location to a warm one, the humidity could condense on the inside;
wait about an hour before supplying it
•test the working power supply voltage, working electrical frequency and working electrical power of the instrument;
they must correspond with the local power supply
•disconnect the local power supply before servicing the instrument
•provide the probes with a protection able to protect them against contacts with metal parts or use insulated probes
•do not use the instrument as safety device
•for repairs and information on the instrument please contact Evco sales network.
Programming
[ EVD120 ]
2Programming
2.1Preliminary remarks
EVD120 supports the MODBUS communication protocol.
The kind of network EVD120 can work in is RS-485 (look at Fig. 4); for further information on the realization of the
network, consult the Installation manual of RICS for Windows® (the document is available in the CD of RICS).
According to the standard settings:
•the instrument address has value 247
•the baud rate has value 9,600 baud
•the parity has value even.
These parameters are hereinafter called “network parameters”.
11EVD120
Programming
(1) use a twisted pair
(2) pair, as far as possible, the ca-
ble to the ones you have already con-
nected
2.2Programming an EVD120 having standard network parameters using the software RICS (adding an EVD120 having standard network parameters to an
existing and working network of instruments managed by RICS)
Operate as follows:
•make sure the network is made of instruments that support the MODBUS communication protocol; to learn the
communication protocol supported by the instruments, consult the List of compatible instruments and supported
communication protocol (the document is available in the internet site www.evco.it)
•make sure the network is made of less than 32 instruments
•make sure no instrument in the network has value 247 as instrument address; to learn the instrument address,
consult the instructions (the document is available in the internet site www.evco.it)
•make sure the baud rate of the instruments in the network has value 9,600 baud; to learn the baud rate of the
instrument, consult the instructions (the document is available in the internet site www.evco.it)
•make sure the parity of the instruments in the network has value even; to learn the parity of the instrument, consult
the instructions (the document is available in the internet site www.evco.it)
•quit RICS
•disconnect the power supply of the instruments in the network
•connect terminal “ + ” of the hardware driver of the last instrument in the network to terminal “ 1 ” of EVD120
•connect terminal “ -- ” of the hardware driver of the last instrument in the network to terminal “ 2 ” of EVD120
•connect terminal “ S ” of the hardware driver of the last instrument in the network to terminal “ 3 ” of EVD120
•to reduce the reflections on the signal transmitted through the cable, plug in the termination of the first and last
element of the network
(1)
(1)
(2)
Fig. 4 - Example of RS-485 network.
12
EVD120
Programming
(3) gain access to RICS with the
account of the administrator or of the
installer
(4) the availability of the functions
of EVD120 depends on the RICS
version; consult the List of compat-
ible instruments and supported com-
munication protocol (the document
is available in the internet site
www.evco.it)
•operate the remaining connections related in paragraph 1.2
•connect the power supply of the instruments in the network
•start RICS
(3) (4)
.
Keep operating as follows:
1. Quit possible dialog boxes.
2. Right-click on the desktop, then use the quick menu to choose Add Device.
A new picture will be shown.
3. Right-click on the picture, then use the quick menu to choose Setup.
The dialog box Setup Device will be shown.
4. Click the card Network, then digit 247 in the pop-up menu Address.
5. Choose OK.
6. Right-click on the picture, then use the quick menu to choose Examine > Parameters.
The dialog box Parameters will be shown.
7. Change the configuration parameters according to your requirements.
8. Change the instrument address.
9. Quit the dialog box Parameters.
10. Switch off/on the power supply of EVD120.
11. Repeat steps 3, 4 and 5.
2.3Programming an EVD120 having standard network parameters using the
software Params Manager
Operate as follows:
•make sure Params Manager is installed in the Personal Computer and the connection between the serial interface
and the Personal Computer works; for further information on the installation of Params Manager and on the realiza-
tion of the connection between the serial interface and the Personal Computer, consult the User manual of Params
Manager (the document is available in the CD of RICS and in the internet site www.evco.it)
•make sure the serial interface supports the MODBUS communication protocol; to learn the communication protocol
supported by the serial interfaces, consult the User manual of Params Manager (the document is available in the
CD of RICS and in the internet site www.evco.it)
•quit Params Manager
•connect terminal “ + ” of the RS-485 output of the serial interface to terminal “ 1 ” of EVD120
•connect terminal “ -- ” of the RS-485 output of the serial interface to terminal “ 2 ” of EVD120
•connect terminal “ S ” of the RS-485 output of the serial interface to terminal “ 3 ” of EVD120
•switch on the power supply of EVD120
•start Params Manager.
Keep operating as follows:
1. Choose Program setup.
The dialog box Setup will be shown.
2. Use the pop-up menu COM port to choose the COM number of the Personal Computer where you have connected
the serial interface.
3. Use the pop-up menu External HW/Protocol to choose DIRECT.
4. Use the pop-up menu Baud rate to choose 9600.
5. Use the pop-up menu Parity to choose Even.
6. Click the card Device Settings, then use the pop-up menu Protocol to choose MODBUS (DIRECT).
7. Digit 247 in the text field Address.
8. Choose Apply.
The frame Information will be shown.
9. Choose OK, then Exit.
10. Start Params Manager again.
11. Choose Link source device.
The dialog box Source device map will be shown.
12. Change the configuration parameters according to your requirements.
13. Change the instrument address.
Programming
14. Choose Exit, then Exit again.
15. Switch off the power supply of EVD120.
16. Disconnect terminals “1”, “2” and “3” of EVD120 from the RS-485 output of the serial interface.
2.4Restoring well-known network parameters of EVD120 temporarily using the
programming key EVKEY (in order to change them using the software Params
Manager)
Operate as follows:
•make sure to have the programming key EVKEY
•switch on the power supply of EVD120 and wait 4 s at least since the power supply has been switched on
•connect EVKEY to port 2 of EVD120: the LED of EVKEY will shed green light
13EVD120
(5) if you switch off the power
supply of EVD120 before having
changed its network parameters, you
will restore its not well-known net-
work parameters
Fig. 5 - Connecting EVKEY to port 2 of EVD120.
•press the button of EVKEY 1 s: LED Prog of EVD120 will flash and the LED of EVKEY will shed green light and
red light alternatively.
The temporary network parameters are the following ones:
•the instrument address has value 248
•the baud rate has value 9,600 baud
•the parity has value even.
Keep operating as follows:
•disconnect EVKEY from port 2 of EVD120
•operate as related in paragraph 2.3 without switching off the power supply of EVD120 before having changed its
network parameters and holding in consideration the instrument address has value 248 (step 7.)
(5)
.
2.5Learning/Programming the instrument address of EVD120 using the programming key EVKEY
2.5.1Learning the instrument address of EVD120 using the programming key EVKEY
Operate as follows:
•make sure to have the programming key EVKEY
•switch on the power supply of EVD120 and wait 4 s at least since the power supply has been switched on
•connect EVKEY to port 2 of EVD120: the LED of EVKEY will shed green light
Fig. 5 - Connecting EVKEY to port 2 of EVD120.
14
EVD120
Programming
(6) the value to consider is not the
instrument address; the instrument
address is the sum of all the values
to consider in the procedure
•press the button of EVKEY 1 s: LED Prog of EVD120 will flash and the LED of EVKEY will shed green light and
red light alternatively
•press and release the button of EVKEY: the LED of EVD120 will go out; if the LED of EVKEY sheds green light
consider value 0, if it sheds red light consider value 1
(6)
•press and release the button of EVKEY: LED Prog of EVD120 will light up; if the LED of EVKEY sheds green light
consider value 0, if it sheds red light consider value 2
(6)
•press and release the button of EVKEY: LED Communication of EVD120 will light up; if the LED of EVKEY sheds
green light consider value 0, if it sheds red light consider value 4
(6)
•press and release the button of EVKEY: LED Communication and LED Prog of EVD120 will light up; if the LED of
EVKEY sheds green light consider value 0, if it sheds red light consider value 8
(6)
•press and release the button of EVKEY: LED Power/Status of EVD120 will light up; if the LED of EVKEY sheds
green light consider value 0, if it sheds red light consider value 16
(6)
•press and release the button of EVKEY: LED Power/Status and LED Prog of EVD120 will light up; if the LED of
EVKEY sheds green light consider value 0, if it sheds red light consider value 32
(6)
Programming
•press and release the button of EVKEY: LED Power/Status and LED Communication of EVD120 will light up; if
the LED of EVKEY sheds green light consider value 0, if it sheds red light consider value 64
(6)
•press and release the button of EVKEY: all the LED of EVD120 will light up; if the LED of EVKEY sheds green
light consider value 0, if it sheds red light consider value 128
(6)
•press and release the button of EVKEY: the procedure will be quitted, LED Prog of EVD120 will flash and the LED
of EVKEY will shed green light
•disconnect EVKEY from port 2 of EVD120
•switch off/on the power supply of EVD120
•the instrument address is the sum of all the values to consider in the procedure.
To quit the procedure early operate as follows:
•disconnect EVKEY from port 2 of EVD120
•switch off/on the power supply of EVD120.
15EVD120
(7) do not assign the instrument
address the following values: 0, 248,
249, 250, 251, 252, 253, 254 or 255;
if you assign the values 0 or 255,
LED Power/Status of EVD120 will
flash and the LED of EVKEY will
shed red light
(8) if you press and release the
button of EVKEY, the procedure will
be quitted without saving the modi-
fication
2.5.2Programming the instrument address of EVD120 using the programming key EVKEY
Operate as follows:
•operate as related in paragraph 2.5.1
•to the steps “if the LED of EVKEY sheds green light consider value ... , if it sheds red light consider value ... ”, press
the button of EVKEY 4 s: the LED of EVKEY will change colour
(7)
To continue the procedure operate as follows:
•press and release the button of EVKEY.
To save the modification operate as follows:
•after the step “if the LED of EVKEY sheds green light consider value 0, if it sheds red light consider value 128”,
press and release the button of EVKEY: all the LED of EVD120 will flash
•press the button of EVKEY 4 s: the procedure will be quitted, LED Prog of EVD120 will flash and the LED of
EVKEY will shed green light
(8)
•disconnect EVKEY from port 2 of EVD120
•switch off/on the power supply of EVD120.
To quit the procedure early operate as follows:
•disconnect EVKEY from port 2 of EVD120
•switch off/on the power supply of EVD120.
16
EVD120
Programming
(9) the copy of the configuration
parameters takes 10 s at most; if in
this time EVKEY does not signal the
operation has successfully been com-
pleted (or the LED of EVKEY fin-
ishes shedding red light to shed green
light again), it takes to switch off the
power supply of EVD120, discon-
nect EVKEY from port 2 of EVD120
and repeat the copy
2.6Copying the configuration parameters of EVD120 in the programming key
EVKEY using the software Params Manager
Operate as follows:
•make sure to have the programming key EVKEY
•switch on the power supply of EVD120 and wait 4 s at least since the power supply has been switched on
•connect EVKEY to port 2 of EVD120: the LED of EVKEY will shed green light
Fig. 5 - Connecting EVKEY to port 2 of EVD120.
•press the button of EVKEY 1 s: LED Prog of EVD120 will flash and the LED of EVKEY will shed green light and
red light alternatively.
•operate as related in paragraph 2.3 till step 12 inclusive holding in consideration the instrument address has value
248 (step 7.).
Keep operating as follows:
1. Choose Store to key.
The configuration parameters of EVD120 will be copied in EVKEY; during the copy the LED of EVKEY sheds red
light.
2. Wait the LED of EVKEY finishes shedding red light
3. Choose Exit, then Exit again.
4. Switch off the power supply of EVD120.
5. Disconnect EVKEY from port 2 of EVD120.
6. Disconnect terminals “1”, “2” and “3” of EVD120 from the RS-485 output of the serial interface.
2.6.1Copying the configuration parameters of EVD120 in the programming key EVKEY using the
software Params Manager (in order to program several EVD120 that are different of the instru-
ment address)
Operate as follows:
•operate as related in paragraph 2.6 making sure parameter Enable AutoIncremental Address Key has value 1.
The first EVD120 programmed with the procedure related in paragraph 2.7 will have the same configuration
parameters of the EVD120 used to copy the configuration parameters in EVKEY; the second EVD120 pro-
grammed with the procedure related in paragraph 2.7 will have the same configuration parameters of the EVD120
used to copy the configuration parameters in EVKEY but parameter instrument address (its value will be in-
creased of one unit); the third EVD120 programmed with the procedure related in paragraph 2.7 will have the
same configuration parameters of the EVD120 used to copy the configuration parameters in EVKEY but param-
eter instrument address (its value will be increased of two units) and so on.
(9)
.
(10) the copy of the configuration
parameters in EVD120 is only al-
lowed if the firmwares of the instru-
ments coincide; if they do not coin-
cide, LED Power/Status and Prog of
EVD120 will flash and the LED of
EVKEY will shed red light.
Programming
2.7Reading the configuration parameters of EVD120 from the programming key
EVKEY
Operate as follows:
•make sure to have the programming key EVKEY
•switch off the power supply of EVD120
•connect EVKEY to port 2 of EVD120
•switch on the power supply of EVD120 and wait 4 s at least since the power supply has been switched on: LED Prog
of EVD120 will flash and the LED of EVKEY will shed green light
•press the button of EVKEY 4 s: the configuration parameters of EVKEY will be copied in EVD120; during the copy
the LED of EVKEY sheds red light
•wait the LED of EVKEY finishes shedding red light
•switch off the power supply of EVD120
•disconnect EVKEY from port 2 of EVD120.
(10)
(9)
17EVD120
Configuration parameters
[ EVD120 ]
3Configuration parameters
3.1Preliminary remarks
Configuration parameters are visible using one of the following software:
•RICS
•Params Manager.
3.2Configuration parameters
TEXT FIELDMIN.MAX. U.M.DEF.ANALOG CHANNELS
(11) if parameters “Probe kind 1”
and/or “Probe kind 2” have value 8,
9, 10 or 11, parameter “Celsius Fahr-
enheit” will have no effect
(12) switch off/on the power sup-
ply of the instrument after the modi-
fication of the parameter
Celsius Fahrenheit01- - -0unit of measure temperature
0= °C
1= °F
Enable Probe 101- - -1enabling the analog channel 1
1 =YES
Enable Probe 201- - -1enabling the analog channel 2
1 =YES
Probe kind 1014- - -1kind of probe analog channel 1
0 =PTC
1 =NTC
2= J
3= K
4 =reserved
5 =3 wires Pt 100
6 =2 wires Pt 100
7 =reserved
8 =reserved
9 =3 wires Pt 1000
10 =2 wires Pt 1000
11 =4-20 mA
12 =0-20 mA
13 =2-10 V
14 =0-10 V
Probe kind 2014- - -1kind of probe analog channel 2
0 =PTC
1 =NTC
2= J
3= K
4 =reserved
5 =3 wires Pt 100
6 =2 wires Pt 100
7 =reserved
8 =reserved
9 =3 wires Pt 1000
19EVD120
Configuration parameters
(11)
(12)
(12)
20
EVD120
Configuration parameters
(13) the unit of measure depends on
parameters “Celsius Fahrenheit” (°C
or °F) and “Probe kind 1” or “Probe
kind 2” (temperature or variable of
the range of the transducer).
10 =2 wires Pt 1000
11 =4-20 mA
12 =0-20 mA
13 =2-10 V
14 =0-10 V
Probe1 Offset-25.025.0
Probe2 Offset-25.025.0
DigitalFilter probe 106- - -3reading speed analog channel 1 (0 = fast ... 6 = slow)
DigitalFilter probe 206- - -3reading speed analog channel 2 (0 = fast ... 6 = slow)
(13)
0.0offset analog channel 1
(13)
0.0offset analog channel 2
(12)
(12)
MinTarI 1-1500.0 1500.0 points 0.0minimum value of the range of the transducer of the ana-
log channel 1 corresponding to 0/4 mA or 0/2 V
MinTarI 2-1500.0 1500.0 points 0.0minimum value of the range of the transducer of the ana-
log channel 2 corresponding to 0/4 mA or 0/2 V
MaxTarI 1-1500.0 1500.0 points 100.0maximum value of the range of the transducer of the ana-
log channel 1 corresponding to 20 mA or 10 V
MaxTarI 2-1500.0 1500.0 points 100.0maximum value of the range of the transducer of the ana-
log channel 2 corresponding to 20 mA or 10 V
Saturation03- - -0values the signal of the transducer of the analog channel 1
Configuration 1saturates (or finishes being converted by the instrument)
0 =function not enabled
1 =below “MinTarI 1” and above “MaxTarI 1”
2 =below 0.0 and above 100.0
3 =below 0.0 and above 1000.0
Saturation03- - -0values the signal of the transducer of the analog channel 2
Configuration 2saturates (or finishes being converted by the instrument)
0 =function not enabled
1 =below “MinTarI 2” and above “MaxTarI 2”
2 =below 0.0 and above 100.0
3 =below 0.0 and above 1000.0
TEXT FIELDMIN.MAX. U.M.DEF.DIGITAL INPUTS
Polarity Digital Input 101- - -0kind of contact digital input 1
0 =NO (input active if you close the contact)
1 =NC (input active if you open the contact)
Polarity Digital Input 201- - -0kind of contact digital input 2
0 =NO (input active if you close the contact)
1 =NC (input active if you open the contact)
TEXT FIELDMIN.MAX. U.M.DEF.SERIAL COMMUNICATION (MODBUS)
Address1247- - -247instrument address
Baud rate03- - -2baud rate
(12)
(12)
0 =2,400 baud
1 =4,800 baud
2 =9,600 baud
3 =19,200 baud
Parity02- - -2parity
(12)
0 =none
1 =odd
2 =even
Configuration parameters
TEXT FIELDMIN.MAX. U.M.DEF.PROGRAMMING
Enable AutoIncremental 01- - -1enabling the function of automatic increase (of one unit)
Address Keyof the instrument address (parameter “Address”) during
the copy of the parameter set from the programming key
to the instrument
21EVD120
Signals
23EVD120
Signals
[ EVD120 ]
4Signals
4.1LED Power/Status
If it is lit, the instrument will be supplied.
If it flashes, an analog channel 1 error and/or an analog channel 2 error will be running.
Remedies:
•look at parameters “Probe kind 1” and/or “Probe kind 2”
•check the integrity of the probe
•check the connection instrument-probe
•check the variable read by the probe.
If it is out, the instrument will not be supplied.
4.2LED Communication (on port 1)
If it is mainly lit, the serial communication will correctly be running.
24
EVD120
Signals
If it flashes, the instrument will read some data but it will not be the recipient.
Remedies:
•look at parameters “instrument address” of the instruments of the RS-485 network.
If it is mainly out, the instrument will read some not correct data.
Remedies:
•look at parameters “baud rate” and “parity” of the instruments of the RS-485 network.
If it is out, the serial communication will not be running.
4.3LED Prog
If it is lit or if it flashes, the programming of the instrument will be running.
Technical data
[ EVD120 ]
5Technical data
5.1Technical data
Box:
self-extinguishing grey.
Frontal protection:
IP 40.
Connections:
•screw terminal blocks (power supply, inputs and port for the serial communication)
•6 poles connector (port for the communication with the programming key).
Working temperature:
from 0 to 55 °C (32 to 131 °F, 10 ... 90% of relative humidity without condensate).
Power supply:
230 VAC, 50/60 Hz, 1.5 VA (approximate).
Analog channels:
2 (analog channel 1 and analog channel 2) for:
•PTC/NTC probes
•J/K thermocouples
•2/3 wires Pt 100 and Pt 1000 probes
•2/3 wires 4-20 mA passive current transducers
•0-10/2-10 V and 0-20/4-20 mA active transducers
(universal analog channels).
Digital inputs:
2 (digital input 1 and digital input 2) for NO/NC contact (free of voltage, 12 V 5 mA approximate).
Working range:
•from -50 to 150 °C (-50 to 300 °F) for PTC probe
•from -40 to 110 °C (-40 to 230 °F) for NTC probe
•from -100 to 800 °C (-140 to 1,450 °F) for J thermocouple
•from -100 to 1,300 °C (-140 to 2,350 °F) for K thermocouple
•from -100 to 650 °C (-140 to 1,200 °F) for 2/3 wires Pt 100 probe
•from -100 to 650 °C (-140 to 1,200 °F) for 2/3 wires Pt 1000 probe.
Resolution:
•0.1 for 2/3 wires 4-20 mA passive current transducers and for 0-10/2-10 V and 0-20/4-20 mA active transducers
•0.1 °C/1 °F otherwise.
Serial ports:
•port for the serial communication (via RS-485, with MODBUS communication protocol)
•port for the communication with the programming key.
25EVD120
Technical data
User manual of EVD120.
Version 1.00 of November 2006.
File EVD120_user_manual_eng_v1.00.pdf.
PT.
This document belongs to Evco; unless you are authorized by Evco, you can not publish it. Evco does not take any responsibility about features, technical data and possible mistakes related in this document
or coming by its use. Evco does not take any responsibility about damages coming by the non-observance of the additional information. Evco reserves the right to make any change without prior notice and
at any time without prejudice the basic safety and operating features.
MODBUS/JBUS IMPLEMENTATION TABLE
Applied to the following products:
EVD120
File name: ModBUS_PRJ314AA06_v1.02.pdf
Version: 1.02
Date of release: 12
h
December 2006
Evco S.r.l.
1 INTRODUCTION
This document describes the resources of the device that can be accessed via the serial port. The protocol is
MODBUS RTU/JBUS.
2 LIST OF DEVICES
This document applys to :
Device Name/Family Notes
EVD120M7
3 ADDRESSING CONVENTIONS
Please note that according to MODBUS specs:
the first register is called register 1
register x must be read ad address x-1
According to JBUS:
the first register is called register 0
register x must be read ad address x
The JBUS convention (zero based) has been used in this document, addresses are shown in hexadecimal
with the leading symbol $.
4 IMPLEMENTED MODBUS FUNCTION CODES
Command Function Code Notes
READ HOLDING REGISTERS $03 Maximum 30 registers at once
WRITE SINGLE REGISTER $06
WRITE MULTIPLE HR $10 Maximum 30 registers at once
READ AND WRITE MULTIPLE HR $17 Maximum 30 registers at once
ModBUS PRJ314AA06 v1.02 12
th
December 2006 page 2 of 8
Holding register addresses table:
Description
Evco S.r.l.
Address
(zero
AccessNotes
based)
DIGITAL INPUTS
Digital inputs register $0101 R
PROBES
Probe 1 (1) $0201 R 1 decimal digit
Probe 2 (1) $0202 R 1 decimal digit
(1) Probes
Reserved Value Meaning
$8000 Probe failure
$8001 Data not available
Bit0 = Digital input 1
Bit1= Digital input 2
ModBUS PRJ314AA06 v1.02 12
th
December 2006 page 3 of 8
Evco S.r.l.
Address
Description
(zero
AccessNotes
based)
PARAMETERS
Meas. unit (0 = Celsius, 1 = Fahrenheit) $0601 R/W
Modbus nodes address $0610 R/W
Baud Rate $0611 R/W
Parity $0612 R/W
Enable Auto-incremental address $0613 R/W
Kind of digital input 1 (0 = NO, 1 = NC) $061F R/W