EVCO EVD120 User Manual

Summary
Important
Introduction
1 Getting started .....................................................................................................................................................9
1.1 Installing the instrument ..................................................................................................................................... 9
1.2 Wiring diagram ................................................................................................................................................. 10
3EVD120
Summary
[ EVD120 ]
Important ............................................................................................................................................................. 5
Introduction ......................................................................................................................................................... 7
Programming
2 Programming ..................................................................................................................................................... 11
2.1 Preliminary remarks .......................................................................................................................................... 11
2.2 Programming an EVD120 having standard network parameters using the software RICS
(adding an EVD120 having standard network parameters to an existing and working network of
instruments managed by RICS) ........................................................................................................................ 11
2.3 Programming an EVD120 having standard network parameters using the software Params Manager .......... 12
2.4 Restoring well-known network parameters of EVD120 temporarily using the programming key
EVKEY (in order to change them using the software Params Manager) .........................................................13
2.5 Learning/Programming the instrument address of EVD120 using the programming key EVKEY ................ 13
2.5.1 Learning the instrument address of EVD120 using the programming key EVKEY ....................................... 13
2.5.2 Programming the instrument address of EVD120 using the programming key EVKEY ................................ 15
2.6 Copying the configuration parameters of EVD120 in the programming key EVKEY using the software
Params Manager ................................................................................................................................................ 16
2.6.1 Copying the configuration parameters of EVD120 in the programming key EVKEY using the software
Params Manager (in order to program several EVD120 that are different of the instrument address) ........... 16
2.7 Reading the configuration parameters of EVD120 from the programming key EVKEY ............................... 17
Configuration parameters
3 Configuration parameters ................................................................................................................................. 19
3.1 Preliminary remarks ..........................................................................................................................................19
3.2 Configuration parameters ................................................................................................................................. 19
Signals
4 Signals ............................................................................................................................................................... 23
4.1 LED Power/Status ............................................................................................................................................. 23
4.2 LED Communication (on port 1) ...................................................................................................................... 23
4.3 LED Prog ...........................................................................................................................................................24
Technical data
5 Technical data ...................................................................................................................................................25
5.1 Technical data ...................................................................................................................................................25
Important
Important
Read these instructions carefully before installing and using the instrument and follow all additional information for
installation and electrical connection; keep these instructions close to the instrument for future consultations.
The instrument must be disposed according to the local legislation about the collection for electrical and electronic
equipment.
5EVD120
Important
[ EVD120 ]
Introduction
[ EVD120 ]
Introduction
EVD120 is a two analog channels and two digital inputs acquisition module.
Each analog channel can work with one of the following kind of probe (universal analog channels):
PTC/NTC probe
J/K thermocouple
2/3 wires Pt 100 and Pt 1000 probe
2/3 wires 4-20 mA passive current transducer
0-10/2-10 V and 0-20/4-20 mA active transducer.
The instrument has the following communication ports:
RS-485 port with MODBUS communication protocol, to allow the connection of the instrument in RS-485 net-
works with MODBUS communication protocol
serial port for the communication with the programming key EVKEY.
The case of EVD120 is 3 DIN modules.
7EVD120
Introduction
Getting started
[ EVD120 ]
1 Getting started
1.1 Installing the instrument
On DIN rail; dimensions in mm (in).
9EVD120
Fig. 1 - Dimensions of EVD120.
Fig. 2 - Installation of EVD120.
Additional information for installation:
working conditions (working temperature, humidity, etc.) must be between the limits related in the technical data
do not install the instrument close to heating sources (heaters, hot air ducts, etc.), devices provided with big magne-
tos (big speakers, etc.), locations subject to direct sunlight, rain, humidity, dust, mechanical vibrations or bumps
according to the safety legislation, the protection against electrical parts must be ensured by a correct installation of
the instrument; the parts that ensure the protection must be installed so that you can not remove them if not by using
a tool.
10
EVD120
1.2 Wiring diagram
Fig. 3 - Wiring diagram of EVD120.
With reference to the wiring diagram:
port 1 is the port for the serial communication (via RS-485, with MODBUS communication protocol)
port 2 is the serial port for the communication with the programming key.
Additional information for electrical connection:
do not operate on the terminal blocks with electrical or pneumatic screwers
if the instrument has been moved from a cold location to a warm one, the humidity could condense on the inside;
wait about an hour before supplying it
test the working power supply voltage, working electrical frequency and working electrical power of the instrument;
they must correspond with the local power supply
disconnect the local power supply before servicing the instrument
provide the probes with a protection able to protect them against contacts with metal parts or use insulated probes
do not use the instrument as safety device
for repairs and information on the instrument please contact Evco sales network.
Programming
[ EVD120 ]
2 Programming
2.1 Preliminary remarks
EVD120 supports the MODBUS communication protocol.
The kind of network EVD120 can work in is RS-485 (look at Fig. 4); for further information on the realization of the
network, consult the Installation manual of RICS for Windows® (the document is available in the CD of RICS).
According to the standard settings:
the instrument address has value 247
the baud rate has value 9,600 baud
the parity has value even.
These parameters are hereinafter called “network parameters”.
11EVD120
Programming
(1) use a twisted pair
(2) pair, as far as possible, the ca-
ble to the ones you have already con-
nected
2.2 Programming an EVD120 having standard network parameters using the soft­ware RICS (adding an EVD120 having standard network parameters to an existing and working network of instruments managed by RICS)
Operate as follows:
make sure the network is made of instruments that support the MODBUS communication protocol; to learn the
communication protocol supported by the instruments, consult the List of compatible instruments and supported
communication protocol (the document is available in the internet site www.evco.it)
make sure the network is made of less than 32 instruments
make sure no instrument in the network has value 247 as instrument address; to learn the instrument address,
consult the instructions (the document is available in the internet site www.evco.it)
make sure the baud rate of the instruments in the network has value 9,600 baud; to learn the baud rate of the
instrument, consult the instructions (the document is available in the internet site www.evco.it)
make sure the parity of the instruments in the network has value even; to learn the parity of the instrument, consult
the instructions (the document is available in the internet site www.evco.it)
quit RICS
disconnect the power supply of the instruments in the network
connect terminal “ + ” of the hardware driver of the last instrument in the network to terminal “ 1 ” of EVD120
connect terminal “ -- ” of the hardware driver of the last instrument in the network to terminal “ 2 ” of EVD120
connect terminal “ S ” of the hardware driver of the last instrument in the network to terminal “ 3 ” of EVD120
to reduce the reflections on the signal transmitted through the cable, plug in the termination of the first and last
element of the network
(1)
(1)
(2)
Fig. 4 - Example of RS-485 network.
12
EVD120
Programming
(3) gain access to RICS with the
account of the administrator or of the
installer
(4) the availability of the functions
of EVD120 depends on the RICS
version; consult the List of compat-
ible instruments and supported com-
munication protocol (the document
is available in the internet site
www.evco.it)
operate the remaining connections related in paragraph 1.2
connect the power supply of the instruments in the network
start RICS
(3) (4)
.
Keep operating as follows:
1. Quit possible dialog boxes.
2. Right-click on the desktop, then use the quick menu to choose Add Device.
A new picture will be shown.
3. Right-click on the picture, then use the quick menu to choose Setup.
The dialog box Setup Device will be shown.
4. Click the card Network, then digit 247 in the pop-up menu Address.
5. Choose OK.
6. Right-click on the picture, then use the quick menu to choose Examine > Parameters.
The dialog box Parameters will be shown.
7. Change the configuration parameters according to your requirements.
8. Change the instrument address.
9. Quit the dialog box Parameters.
10. Switch off/on the power supply of EVD120.
11. Repeat steps 3, 4 and 5.
2.3 Programming an EVD120 having standard network parameters using the software Params Manager
Operate as follows:
make sure Params Manager is installed in the Personal Computer and the connection between the serial interface
and the Personal Computer works; for further information on the installation of Params Manager and on the realiza-
tion of the connection between the serial interface and the Personal Computer, consult the User manual of Params
Manager (the document is available in the CD of RICS and in the internet site www.evco.it)
make sure the serial interface supports the MODBUS communication protocol; to learn the communication protocol
supported by the serial interfaces, consult the User manual of Params Manager (the document is available in the
CD of RICS and in the internet site www.evco.it)
quit Params Manager
connect terminal “ + ” of the RS-485 output of the serial interface to terminal “ 1 ” of EVD120
connect terminal “ -- ” of the RS-485 output of the serial interface to terminal “ 2 ” of EVD120
connect terminal “ S ” of the RS-485 output of the serial interface to terminal “ 3 ” of EVD120
switch on the power supply of EVD120
start Params Manager.
Keep operating as follows:
1. Choose Program setup.
The dialog box Setup will be shown.
2. Use the pop-up menu COM port to choose the COM number of the Personal Computer where you have connected
the serial interface.
3. Use the pop-up menu External HW/Protocol to choose DIRECT.
4. Use the pop-up menu Baud rate to choose 9600.
5. Use the pop-up menu Parity to choose Even.
6. Click the card Device Settings, then use the pop-up menu Protocol to choose MODBUS (DIRECT).
7. Digit 247 in the text field Address.
8. Choose Apply.
The frame Information will be shown.
9. Choose OK, then Exit.
10. Start Params Manager again.
11. Choose Link source device.
The dialog box Source device map will be shown.
12. Change the configuration parameters according to your requirements.
13. Change the instrument address.
Programming
14. Choose Exit, then Exit again.
15. Switch off the power supply of EVD120.
16. Disconnect terminals “1”, “2” and “3” of EVD120 from the RS-485 output of the serial interface.
2.4 Restoring well-known network parameters of EVD120 temporarily using the programming key EVKEY (in order to change them using the software Params Manager)
Operate as follows:
make sure to have the programming key EVKEY
switch on the power supply of EVD120 and wait 4 s at least since the power supply has been switched on
connect EVKEY to port 2 of EVD120: the LED of EVKEY will shed green light
13EVD120
(5) if you switch off the power
supply of EVD120 before having
changed its network parameters, you
will restore its not well-known net-
work parameters
Fig. 5 - Connecting EVKEY to port 2 of EVD120.
press the button of EVKEY 1 s: LED Prog of EVD120 will flash and the LED of EVKEY will shed green light and
red light alternatively.
The temporary network parameters are the following ones:
the instrument address has value 248
the baud rate has value 9,600 baud
the parity has value even.
Keep operating as follows:
disconnect EVKEY from port 2 of EVD120
operate as related in paragraph 2.3 without switching off the power supply of EVD120 before having changed its
network parameters and holding in consideration the instrument address has value 248 (step 7.)
(5)
.
2.5 Learning/Programming the instrument address of EVD120 using the pro­gramming key EVKEY
2.5.1 Learning the instrument address of EVD120 using the programming key EVKEY
Operate as follows:
make sure to have the programming key EVKEY
switch on the power supply of EVD120 and wait 4 s at least since the power supply has been switched on
connect EVKEY to port 2 of EVD120: the LED of EVKEY will shed green light
Fig. 5 - Connecting EVKEY to port 2 of EVD120.
14
EVD120
Programming
(6) the value to consider is not the
instrument address; the instrument
address is the sum of all the values
to consider in the procedure
press the button of EVKEY 1 s: LED Prog of EVD120 will flash and the LED of EVKEY will shed green light and
red light alternatively
press and release the button of EVKEY: the LED of EVD120 will go out; if the LED of EVKEY sheds green light
consider value 0, if it sheds red light consider value 1
(6)
press and release the button of EVKEY: LED Prog of EVD120 will light up; if the LED of EVKEY sheds green light
consider value 0, if it sheds red light consider value 2
(6)
press and release the button of EVKEY: LED Communication of EVD120 will light up; if the LED of EVKEY sheds
green light consider value 0, if it sheds red light consider value 4
(6)
press and release the button of EVKEY: LED Communication and LED Prog of EVD120 will light up; if the LED of
EVKEY sheds green light consider value 0, if it sheds red light consider value 8
(6)
press and release the button of EVKEY: LED Power/Status of EVD120 will light up; if the LED of EVKEY sheds
green light consider value 0, if it sheds red light consider value 16
(6)
press and release the button of EVKEY: LED Power/Status and LED Prog of EVD120 will light up; if the LED of
EVKEY sheds green light consider value 0, if it sheds red light consider value 32
(6)
Programming
press and release the button of EVKEY: LED Power/Status and LED Communication of EVD120 will light up; if
the LED of EVKEY sheds green light consider value 0, if it sheds red light consider value 64
(6)
press and release the button of EVKEY: all the LED of EVD120 will light up; if the LED of EVKEY sheds green
light consider value 0, if it sheds red light consider value 128
(6)
press and release the button of EVKEY: the procedure will be quitted, LED Prog of EVD120 will flash and the LED
of EVKEY will shed green light
disconnect EVKEY from port 2 of EVD120
switch off/on the power supply of EVD120
the instrument address is the sum of all the values to consider in the procedure.
To quit the procedure early operate as follows:
disconnect EVKEY from port 2 of EVD120
switch off/on the power supply of EVD120.
15EVD120
(7) do not assign the instrument
address the following values: 0, 248,
249, 250, 251, 252, 253, 254 or 255;
if you assign the values 0 or 255,
LED Power/Status of EVD120 will
flash and the LED of EVKEY will
shed red light
(8) if you press and release the
button of EVKEY, the procedure will
be quitted without saving the modi-
fication
2.5.2 Programming the instrument address of EVD120 using the programming key EVKEY
Operate as follows:
operate as related in paragraph 2.5.1
to the steps “if the LED of EVKEY sheds green light consider value ... , if it sheds red light consider value ... ”, press
the button of EVKEY 4 s: the LED of EVKEY will change colour
(7)
To continue the procedure operate as follows:
press and release the button of EVKEY.
To save the modification operate as follows:
after the step “if the LED of EVKEY sheds green light consider value 0, if it sheds red light consider value 128”,
press and release the button of EVKEY: all the LED of EVD120 will flash
press the button of EVKEY 4 s: the procedure will be quitted, LED Prog of EVD120 will flash and the LED of
EVKEY will shed green light
(8)
disconnect EVKEY from port 2 of EVD120
switch off/on the power supply of EVD120.
To quit the procedure early operate as follows:
disconnect EVKEY from port 2 of EVD120
switch off/on the power supply of EVD120.
16
EVD120
Programming
(9) the copy of the configuration
parameters takes 10 s at most; if in
this time EVKEY does not signal the
operation has successfully been com-
pleted (or the LED of EVKEY fin-
ishes shedding red light to shed green
light again), it takes to switch off the
power supply of EVD120, discon-
nect EVKEY from port 2 of EVD120
and repeat the copy
2.6 Copying the configuration parameters of EVD120 in the programming key EVKEY using the software Params Manager
Operate as follows:
make sure to have the programming key EVKEY
switch on the power supply of EVD120 and wait 4 s at least since the power supply has been switched on
connect EVKEY to port 2 of EVD120: the LED of EVKEY will shed green light
Fig. 5 - Connecting EVKEY to port 2 of EVD120.
press the button of EVKEY 1 s: LED Prog of EVD120 will flash and the LED of EVKEY will shed green light and
red light alternatively.
operate as related in paragraph 2.3 till step 12 inclusive holding in consideration the instrument address has value
248 (step 7.).
Keep operating as follows:
1. Choose Store to key.
The configuration parameters of EVD120 will be copied in EVKEY; during the copy the LED of EVKEY sheds red
light.
2. Wait the LED of EVKEY finishes shedding red light
3. Choose Exit, then Exit again.
4. Switch off the power supply of EVD120.
5. Disconnect EVKEY from port 2 of EVD120.
6. Disconnect terminals “1”, “2” and “3” of EVD120 from the RS-485 output of the serial interface.
2.6.1 Copying the configuration parameters of EVD120 in the programming key EVKEY using the
software Params Manager (in order to program several EVD120 that are different of the instru-
ment address)
Operate as follows:
operate as related in paragraph 2.6 making sure parameter Enable AutoIncremental Address Key has value 1.
The first EVD120 programmed with the procedure related in paragraph 2.7 will have the same configuration
parameters of the EVD120 used to copy the configuration parameters in EVKEY; the second EVD120 pro-
grammed with the procedure related in paragraph 2.7 will have the same configuration parameters of the EVD120
used to copy the configuration parameters in EVKEY but parameter instrument address (its value will be in-
creased of one unit); the third EVD120 programmed with the procedure related in paragraph 2.7 will have the
same configuration parameters of the EVD120 used to copy the configuration parameters in EVKEY but param-
eter instrument address (its value will be increased of two units) and so on.
(9)
.
(10) the copy of the configuration
parameters in EVD120 is only al-
lowed if the firmwares of the instru-
ments coincide; if they do not coin-
cide, LED Power/Status and Prog of
EVD120 will flash and the LED of
EVKEY will shed red light.
Programming
2.7 Reading the configuration parameters of EVD120 from the programming key EVKEY
Operate as follows:
make sure to have the programming key EVKEY
switch off the power supply of EVD120
connect EVKEY to port 2 of EVD120
switch on the power supply of EVD120 and wait 4 s at least since the power supply has been switched on: LED Prog
of EVD120 will flash and the LED of EVKEY will shed green light
press the button of EVKEY 4 s: the configuration parameters of EVKEY will be copied in EVD120; during the copy
the LED of EVKEY sheds red light
wait the LED of EVKEY finishes shedding red light
switch off the power supply of EVD120
disconnect EVKEY from port 2 of EVD120.
(10)
(9)
17EVD120
Configuration parameters
[ EVD120 ]
3 Configuration parameters
3.1 Preliminary remarks
Configuration parameters are visible using one of the following software:
•RICS
Params Manager.
3.2 Configuration parameters
TEXT FIELD MIN. MAX. U.M. DEF. ANALOG CHANNELS
(11) if parameters “Probe kind 1”
and/or “Probe kind 2” have value 8,
9, 10 or 11, parameter “Celsius Fahr-
enheit” will have no effect
(12) switch off/on the power sup-
ply of the instrument after the modi-
fication of the parameter
Celsius Fahrenheit 0 1 - - - 0 unit of measure temperature
0= °C
1= °F
Enable Probe 1 0 1 - - - 1 enabling the analog channel 1
1 = YES
Enable Probe 2 0 1 - - - 1 enabling the analog channel 2
1 = YES
Probe kind 1 0 14 - - - 1 kind of probe analog channel 1
0 = PTC
1 = NTC
2= J
3= K
4 = reserved
5 = 3 wires Pt 100
6 = 2 wires Pt 100
7 = reserved
8 = reserved
9 = 3 wires Pt 1000
10 = 2 wires Pt 1000
11 = 4-20 mA
12 = 0-20 mA
13 = 2-10 V
14 = 0-10 V
Probe kind 2 0 14 - - - 1 kind of probe analog channel 2
0 = PTC
1 = NTC
2= J
3= K
4 = reserved
5 = 3 wires Pt 100
6 = 2 wires Pt 100
7 = reserved
8 = reserved
9 = 3 wires Pt 1000
19EVD120
Configuration parameters
(11)
(12)
(12)
20
EVD120
Configuration parameters
(13) the unit of measure depends on
parameters “Celsius Fahrenheit” (°C
or °F) and “Probe kind 1” or “Probe
kind 2” (temperature or variable of
the range of the transducer).
10 = 2 wires Pt 1000
11 = 4-20 mA
12 = 0-20 mA
13 = 2-10 V
14 = 0-10 V
Probe1 Offset -25.0 25.0
Probe2 Offset -25.0 25.0
DigitalFilter probe 1 0 6 - - - 3 reading speed analog channel 1 (0 = fast ... 6 = slow)
DigitalFilter probe 2 0 6 - - - 3 reading speed analog channel 2 (0 = fast ... 6 = slow)
(13)
0.0 offset analog channel 1
(13)
0.0 offset analog channel 2
(12)
(12)
MinTarI 1 -1500.0 1500.0 points 0.0 minimum value of the range of the transducer of the ana-
log channel 1 corresponding to 0/4 mA or 0/2 V
MinTarI 2 -1500.0 1500.0 points 0.0 minimum value of the range of the transducer of the ana-
log channel 2 corresponding to 0/4 mA or 0/2 V
MaxTarI 1 -1500.0 1500.0 points 100.0 maximum value of the range of the transducer of the ana-
log channel 1 corresponding to 20 mA or 10 V
MaxTarI 2 -1500.0 1500.0 points 100.0 maximum value of the range of the transducer of the ana-
log channel 2 corresponding to 20 mA or 10 V
Saturation 0 3 - - - 0 values the signal of the transducer of the analog channel 1
Configuration 1 saturates (or finishes being converted by the instrument)
0 = function not enabled
1 = below “MinTarI 1” and above “MaxTarI 1”
2 = below 0.0 and above 100.0
3 = below 0.0 and above 1000.0
Saturation 0 3 - - - 0 values the signal of the transducer of the analog channel 2
Configuration 2 saturates (or finishes being converted by the instrument)
0 = function not enabled
1 = below “MinTarI 2” and above “MaxTarI 2”
2 = below 0.0 and above 100.0
3 = below 0.0 and above 1000.0
TEXT FIELD MIN. MAX. U.M. DEF. DIGITAL INPUTS
Polarity Digital Input 1 0 1 - - - 0 kind of contact digital input 1
0 = NO (input active if you close the contact)
1 = NC (input active if you open the contact)
Polarity Digital Input 2 0 1 - - - 0 kind of contact digital input 2
0 = NO (input active if you close the contact)
1 = NC (input active if you open the contact)
TEXT FIELD MIN. MAX. U.M. DEF. SERIAL COMMUNICATION (MODBUS)
Address 1 247 - - - 247 instrument address
Baud rate 0 3 - - - 2 baud rate
(12)
(12)
0 = 2,400 baud
1 = 4,800 baud
2 = 9,600 baud
3 = 19,200 baud
Parity 0 2 - - - 2 parity
(12)
0 = none
1 = odd
2 = even
Configuration parameters
TEXT FIELD MIN. MAX. U.M. DEF. PROGRAMMING
Enable AutoIncremental 0 1 - - - 1 enabling the function of automatic increase (of one unit)
Address Key of the instrument address (parameter “Address”) during
the copy of the parameter set from the programming key
to the instrument
21EVD120
Signals
23EVD120
Signals
[ EVD120 ]
4 Signals
4.1 LED Power/Status
If it is lit, the instrument will be supplied.
If it flashes, an analog channel 1 error and/or an analog channel 2 error will be running.
Remedies:
look at parameters “Probe kind 1” and/or “Probe kind 2”
check the integrity of the probe
check the connection instrument-probe
check the variable read by the probe.
If it is out, the instrument will not be supplied.
4.2 LED Communication (on port 1)
If it is mainly lit, the serial communication will correctly be running.
24
EVD120
Signals
If it flashes, the instrument will read some data but it will not be the recipient.
Remedies:
look at parameters “instrument address” of the instruments of the RS-485 network.
If it is mainly out, the instrument will read some not correct data.
Remedies:
look at parameters “baud rate” and “parity” of the instruments of the RS-485 network.
If it is out, the serial communication will not be running.
4.3 LED Prog
If it is lit or if it flashes, the programming of the instrument will be running.
Technical data
[ EVD120 ]
5 Technical data
5.1 Technical data
Box:
self-extinguishing grey.
Frontal protection:
IP 40.
Connections:
screw terminal blocks (power supply, inputs and port for the serial communication)
6 poles connector (port for the communication with the programming key).
Working temperature:
from 0 to 55 °C (32 to 131 °F, 10 ... 90% of relative humidity without condensate).
Power supply:
230 VAC, 50/60 Hz, 1.5 VA (approximate).
Analog channels:
2 (analog channel 1 and analog channel 2) for:
PTC/NTC probes
J/K thermocouples
2/3 wires Pt 100 and Pt 1000 probes
2/3 wires 4-20 mA passive current transducers
0-10/2-10 V and 0-20/4-20 mA active transducers
(universal analog channels).
Digital inputs:
2 (digital input 1 and digital input 2) for NO/NC contact (free of voltage, 12 V 5 mA approximate).
Working range:
from -50 to 150 °C (-50 to 300 °F) for PTC probe
from -40 to 110 °C (-40 to 230 °F) for NTC probe
from -100 to 800 °C (-140 to 1,450 °F) for J thermocouple
from -100 to 1,300 °C (-140 to 2,350 °F) for K thermocouple
from -100 to 650 °C (-140 to 1,200 °F) for 2/3 wires Pt 100 probe
from -100 to 650 °C (-140 to 1,200 °F) for 2/3 wires Pt 1000 probe.
Resolution:
0.1 for 2/3 wires 4-20 mA passive current transducers and for 0-10/2-10 V and 0-20/4-20 mA active transducers
0.1 °C/1 °F otherwise.
Serial ports:
port for the serial communication (via RS-485, with MODBUS communication protocol)
port for the communication with the programming key.
25EVD120
Technical data
User manual of EVD120.
Version 1.00 of November 2006.
File EVD120_user_manual_eng_v1.00.pdf.
PT.
This document belongs to Evco; unless you are authorized by Evco, you can not publish it. Evco does not take any responsibility about features, technical data and possible mistakes related in this document
or coming by its use. Evco does not take any responsibility about damages coming by the non-observance of the additional information. Evco reserves the right to make any change without prior notice and
at any time without prejudice the basic safety and operating features.
MODBUS/JBUS IMPLEMENTATION TABLE
Applied to the following products:
EVD120
File name: ModBUS_PRJ314AA06_v1.02.pdf
Version: 1.02
Date of release: 12
h
December 2006
Evco S.r.l.
1 INTRODUCTION
This document describes the resources of the device that can be accessed via the serial port. The protocol is MODBUS RTU/JBUS.
2 LIST OF DEVICES
This document applys to :
Device Name/Family Notes
EVD120M7
3 ADDRESSING CONVENTIONS
Please note that according to MODBUS specs:
the first register is called register 1 register x must be read ad address x-1
According to JBUS:
the first register is called register 0 register x must be read ad address x
The JBUS convention (zero based) has been used in this document, addresses are shown in hexadecimal with the leading symbol $.
4 IMPLEMENTED MODBUS FUNCTION CODES
Command Function Code Notes
READ HOLDING REGISTERS $03 Maximum 30 registers at once
WRITE SINGLE REGISTER $06
WRITE MULTIPLE HR $10 Maximum 30 registers at once
READ AND WRITE MULTIPLE HR $17 Maximum 30 registers at once
ModBUS PRJ314AA06 v1.02 12
th
December 2006 page 2 of 8
Holding register addresses table:
Description
Evco S.r.l.
Address
(zero
Access Notes
based)
DIGITAL INPUTS
Digital inputs register $0101 R
PROBES
Probe 1 (1) $0201 R 1 decimal digit
Probe 2 (1) $0202 R 1 decimal digit
(1) Probes
Reserved Value Meaning
$8000 Probe failure
$8001 Data not available
Bit0 = Digital input 1 Bit1= Digital input 2
ModBUS PRJ314AA06 v1.02 12
th
December 2006 page 3 of 8
Evco S.r.l.
Address
Description
(zero
Access Notes
based)
PARAMETERS
Meas. unit (0 = Celsius, 1 = Fahrenheit) $0601 R/W
Modbus nodes address $0610 R/W
Baud Rate $0611 R/W
Parity $0612 R/W
Enable Auto-incremental address $0613 R/W
Kind of digital input 1 (0 = NO, 1 = NC) $061F R/W
msk $0001 = digital In 1 msk $0002 = digital In 2
bit0 = Enable bit1..3 = reserved
Probe 1 configuration $0620 R/W
bit4..7 = Digital filter (2) bit8..11 = probe kind (3) bit12..15 = reserved
Probe 1 offset $0621 R/W Default 0
Min range of linear input 1 (corresponding to 0/4 mA or 0/2 V)
Max range of linear input 1 (corresponding to 20 mA or 10 V)
$0622 R/W Default 0
$0623 R/W Defailt 1000
0 = no saturation 1 = saturation on min and max
Kind of saturation of linear input 1 $0624 R/W
range of linear input 1 2 = saturation on 0, 1000 3 = saturation on 0, 10000
bit0 = Enable bit1..3 = reserved
Probe 2 configuration $0630 R/W
bit4..7 = Digital filter (2) bit8..11 = probe kind (3) bit12..15 = reserved
Probe 2 offset $0631 R/W Default 0
Min range of linear input 2 (corresponding to 0/4 mA or 0/2 V)
$0632 R/W Default 0
ModBUS PRJ314AA06 v1.02 12
th
December 2006 page 4 of 8
Evco S.r.l.
Max range of linear input 2 (corresponding to 20 mA or 10 V)
$0633 R/W Default 1000
Kind of saturation of linear input 2 $0634 R/W
INFO
Firmware ID $FF08 R
Firmware Variation/revision $FF09 R
(2) Digital filter: 3 is the default value.
(3) Probe kind
Value Meaning
0 PTC
1 NTC
0 = no saturation 1 = saturation on min and max
range of linear input 2 2 = saturation on 0, 1000 3 = saturation on 0, 10000
2 TC J
3 TC K
4 Reserved
5 Pt 100 3 wires
6 Pt 100 2 wires
7 Rerserved
8 Rerserved
9 Pt 1000 3 wires
10 Pt 1000 3 wires
11 4-20 mA
12 0-20 mA
13 2-10 V
14 0-10 V
PAY ATTENTION:
ALL THE ADDRESSES NOT MENTIONED EXPRESSLY IN THIS DOCUMENT MUST BE CONSIDERED
RESERVED. IF YOU DO NOT RESPECT THIS CONDITION, YOU COULD ALTER THE OPERATION OF
THE INSTRUMENT.
th
ModBUS PRJ314AA06 v1.02 12
December 2006 page 5 of 8
Evco S.r.l.
ULTERIORI HOLDING REGISTER PER USO INTERNO
FOTOGRAFIA SULLO STATO DELLO STRUMENTO:
Tramite questi Holding Register è possibile eseguire una fotografia statica sul funzionamento dello
strumento; in tale fotografia vengono riportati i valori misurati dalle sonde, il minimo e il massimo assunto da
entrambe dal momento del precedente rinfresco, all’attuale lettura, lo stato degli ingressi digitali e degli
allarmi, la traccia di una loro eventuale attivazione o disattivazione precedente.
E’ possibile inoltre monitorare una eventuale informazione aggiuntiva da rilevare, tramite il cosiddetto
“variation register”: tale holding register contiene una serie di flag che vengono alzati nel momento in cuivi è
una qualunque variazione sul funzionamento dello strumento (per esempio se i parametri non vengono mai
modificato non ha senso perdere tempo a monitorarli tutti: è più comodo leggere di continuo un flag che
monitorizza una loro possibile modifica).
Address
Description
(zero
based)
Access
Notes
PARAMETERS
Kind of list $0500 R
Variation register 1 $0501 R/W
(*) Reading this holding register,
memory will be update.
Other flags $0502 R/W
Current probe 1 value $0503 R
Current probe 2 value $0504 R
Probe 1 value memory $0505 R
Probe 2 value memory $0506 R
Min probe 1 value $0507 R
Max probe 1 value $0508 R
Min probe 2 value $0509 R
Max probe 2 value $050A R
Current digital inputs $050B R
Digital inputs memory $050C R
msk $0001: current digital input 1 msk $0002: current digital input 2 msk $0001: digital input 1 msk $0002: digital input 2
msk $0001: memo digital input 1 max Digital input memory min max
$050D R
msk $0002: memo digital input 2 max
msk $0100: memo digital input 1 min
msk $0200: memo digital input 2 min
msk $0001: current probe failure 1 Current status of alarm $050E R
msk $0002: current probe failure 2
msk $0004: current calibration error
ModBUS PRJ314AA06 v1.02 12
th
December 2006 page 6 of 8
Evco S.r.l.
Alarm memory $050F R
Alarm memory min max $0510 R
Periodical request counter init
Cold junction current value (in °C)
$0511 R/W
$0512 R
UNUSED 1 $0513 R
UNUSED 2 $0514 R
UNUSED 3 $0515 R
Photo counter $0516 R
Refresh counter $0517 R
UNUSED 4 $0518 R
msk $0001: probe failure 1
msk $0002: probe failure 2
msk $0004: calibration error
msk $0001: probe failure 1 max
msk $0002: probe failure 2 max
msk $0004: calibration error max
msk $0100: probe failure 1 max
msk $0200: probe failure 2 max
msk $0400: calibration error max
Time for periodical request counter;
this value will be adjust with a random
time between 0 and 7.
when this counter become null,
instrument’s status memory will be
updated. Reading this holding
register, memory will be update.
UNUSED 5 $0519 R
UNUSED 6 $051A R
UNUSED 7 $051B R
UNUSED 8 $051C R
Periodical request counter
Refresh status memory command
$051D R/W
$0480 W
Counter for periodical request; when
this counter become null, bit 4 in the
variation register become 1.
write 1 to refresh status memory;
writing 1, status memory will be
updated.
(*)
bit15 bit14 bit13 bit12 bit11 bit10 bit9 bit8 bit7 bit6 bit5 bit4 bit3 bit2 bit1 bit0
bit0, bit1: allways 0;
bit2, bit3: reserved;
bit4: periodical request by timer;
bit5: some variation occurred;
bit6: reset occurred;
bit7: internal status changed
bit8: alarm status changed
ModBUS PRJ314AA06 v1.02 12
th
December 2006 page 7 of 8
bit9: analogical input changed
bit10: digital input changed
bit11: analogical output changed
bit12: digital input changed
bit13: parameters value changed
bit14:set point or other internal settings changed
bit 15: reserved
Evco S.r.l.
ModBUS PRJ314AA06 v1.02 12
th
December 2006 page 8 of 8
Loading...