Essick Air EA1407, HD1409 User Manual

0 (0)

READ AND SAVE THESE INSTRUCTIONS

®

DUAL FAN CONSOLE

EVAPORATIVE

HUMIDIFIER

(14 GALLON SERIES)

USE AND CARE GUIDE

Adjustable HumidistatEasy fill bottlesFour Casters

EA1407 ; HD1409

Français ………….17

Español ………..…33

TO ORDER PARTS AND ACCESSORIES CALL 1.800.547.3888

1B72889

11/2014

PRINTED IN CHINA

IMPORTANT SAFEGUARDS

General Safety Instructions

READ BEFORE USING YOUR HUMIDIFIER

DANGER: means if the safety information is not followed someone will be seriously injured or killed.

WARNING: means if the safety information is not followed someone could be seriously injured or killed.

CAUTION: means if the safety information is not followed someone may be injured.

1.To reduce the risk of fire or shock hazard, this humidifier has a polarized plug (one blade is wider than the other). Plug humidifier directly into a 120V, A.C. electrical outlet. Do not use extension cords. If the plug does not fully fit

into the outlet, reverse plug. If it still does not fit, contact a qualified electrician to install the proper outlet. Do not change the plug in any way.

2.Keep the electric cord out of traffic areas. To reduce the risk of fire hazard, never put the electric cord under rugs, near heat registers, radiators, stoves or heaters.

3.Always unplug the unit before moving, cleaning or removing the fan assembly section from the humidifier, or whenever it is not in service.

4.Keep the humidifier clean. To reduce the risk of injury, fire or damage to humidifier, use only cleaners specifically recommended for humidifiers. Never use flammable, combustible or poisonous materials to clean your humidifier

5.Do not put foreign objects inside the humidifier.

6.Do not allow unit to be used as a toy. Close attention is necessary when used by or near children.

7.To reduce the risk of electrical hazard or damage to humidifier, do not tilt, jolt or tip humidifier while unit is running.

8.To reduce the risk of accidental electrical shock, do not touch the cord or controls with wet hands.

9.To reduce the risk of fire, do not use near an open flame such as a candle or other flame source.

WARNING: For your own safety, do not use humidifier if any parts are damaged or missing.

WARNING: To reduce risk of fire, electric shock, or injury always unplug before servicing or cleaning.

WARNING: To reduce the risk of fire or shock hazard, do not pour or spill water into control or motor area. If controls get wet, let them dry completely and have unit checked by authorized service personnel before plugging in.

2

INTRODUCTION

Your new humidifier adds invisible moisture to your home by moving dry inlet air through a saturated wick. As air moves through the wick, the water evaporates into the air, leaving behind any white dust, minerals, or dissolved and suspended solids in the wick. Because the water is evaporated, there is just clean and invisible moist air.

As the evaporative wick traps accumulated minerals from the water, its ability to absorb and evaporate water decreases. We recommend changing the wick at the beginning of every season. In hard water areas, more frequent replacement may be necessary to maintain your humidifier’s efficiency.

Use only AIRCARE® or Essick Air® / MoistAIR® brand replacement wicks and chemicals. To order parts, wicks and chemicals call 1-800-547-3888. This series of humidifiers uses wick # HDC12. Only AIRCARE® or Essick Air® / MoistAIR® evaporative wicks guarantee the certified output of your humidifier. Use of other brands of wicks voids the certification of output and may void your warranty.

KNOW YOUR HUMIDIFIER

Description

EA1407/HD1409

*Output per 24 hrs

14 gallons/53 Litres

Capacity of Each Bottle

2.375 gallons/ 9 Litres

*Sq. ft. coverage

+3,000 (Tight Const.)

2,900 (Avg. Const.)

 

Fan Speeds

3

Auto Fan Speed

Yes

Replacement Wick

HDC12 (Set of 4)

Refill Indicator

Yes

Adjustable Humidistat

Yes

Auto Shut Off

Yes

Controls

Digital

Casters

Yes (4)

ETL Listed

Yes

Volts

120 A.C.

Hertz

60

Amps (High Speed)

2.88

Watts (High Speed)

165

* Based on an area with average insulation and an 8 foot ceiling height. ANSI/AHAM HU-1-1997. SEC 8.1 Results may vary.

HOW YOUR HUMIDIFIER WORKS

Dry air is drawn into the humidifier though the back and moisturized as it passes through the evaporative wicks. It is then fanned out into the room.

3

Essick Air EA1407, HD1409 User Manual

ASSEMBLY

1.Unpack humidifier from carton. Remove manual.

2.Remove all packaging materials. Some packaging material is located under the fan assembly. Verify the following items are in the humidifier cabinet.

♦ 2 wick assemblies with

4 casters and 2 caster pads

 

4 wicks installed

 

 

2 bottles

♦ 2 bottle caps and 2 valve caps

Float with retainer ring

Manual & registration card

CASTERS

 

 

3. Turn the emptied cabinet upside down.

4. Insert each caster stem into one of the four caster holes manually Do not use tools. The casters should fit snugly and be inserted until the stem shoulder reaches the cabinet surface.

5. Turn the cabinet right side up.

REINSTALLING COMPONENTS

1.Reinstall float.

2.Reinstall wick assemblies into cabinet.

3.Reinstall water bottles

Float

Front

Right

Left

FRO

N

T

 

FRONT

FRONT

FRONT

FRONT

 

Front

Front

Front

NOTES ON LOCATION:

In order to get the most effective use from your humidifier, it is important to position the unit where the most humidity is needed or where the moist air will be circulated throughout the house such as near a cold air return. If the unit is positioned close to a window, condensation may form on the window pane. If this occurs, the unit should be repositioned in another location.

Place humidifier on a flat, level surface. Do NOT position the unit directly in front of a hot air duct or radiator. If placing on carpet, use caster pads to ensure level placement.

Due to release of cool, moist air from the humidifier, it is best to direct air away from thermostat and hot air registers. Place the unit on a level surface 2 inches from the wall.

Front of cabinet

Front caster

Caster pad

Carpet or uneven surface

If positioning on carpet or

uneven surface, place caster pads under front casters only to level (if necessary).

NOTE: The use of a buble level will ensure placement for proper bottle draining.

4

FILLING

WARNING: Disconnect power before filling, cleaning or servicing unit. Keep grille dry at all times.

WARNING: To reduce the risk of fire or shock hazard, DO NOT pour or spill water into control or motor area. If controls get wet, let them dry completely and have unit checked by authorized service personnel before plugging in

NOTE: Fill unit with cool, fresh (preferably unsoftened) water. If you only have softened water available in your home, you can use it, but mineral buildup will occur more quickly.

NOTE: On initial fill up, it may take approximately 30 minutes for the wicks to become fully saturated.

CAUTION: Use only EPA Registered Bacteriostat. Under no circumstances should you use Water Treatment products intended for Rotobelt or Ultrasonic humidifiers.

We recommend using AIRCARE® or Essick Air® Bacteriostat Treatment P/N 1970 when you refill the water reservoir to eliminate bacterial growth. Add bacteriostat according to the instructions on the bottle. Call 1-800-547-3888 to order.

NOTE: The humidifier contains water bottles. The weight of the bottles in the front of the cabinet may cause the front casters to sink deeper into the carpet. If this happens, the water level will be higher towards the front of the humidifier, and lower towards the back. The water level float (located at the rear of the cabinet) may not sense the water so the flashing “F” will be displayed and the unit will not energize. If this happens, the unit will need to be leveled in order for the humidifier to be energized. To level the humidifier, place the included caster pads under the front two casters of the humidifier only.

1. Open the pivoting lid of the humidifier.

Remove the water bottles by lifting them straight up and out of the cabinet.

Notice the water bottles are designed as right and left hand units. When refilling the bottles ensure correct placement for proper function.

The water bottles should rest in the front of the cabinet. Each “Side-Fill Cap” must be facing the rear of the unit.

Each water bottle is designed with a “Side-Fill Cap” and bottom valve cap. Before filling each bottle check that the bottle valve cap is on securely.

NOTE: A fill hose, P/N 4400, that attaches to a faucet can be purchased by calling1-800-547-3888.

5

2.To fill, remove the “Side-Fill Cap” and place the opening under a water faucet (the bottle should fit under a four-inch clearance faucet). Fill with recommended mixture of cold water and Bacteriostat (if necessary) only. Solid debris could make the valve leak. Install the “Side-Fill Cap” by twisting clockwise. Hand tighten only.

NOTE: Both bottle caps have rubber washers in them. Do Not Remove.

3.Check for possible leaks by performing the following:

a.Grasp bottle handle and hold the bottle upright over a sink.

b.Inspect the bottom valve for leaks.

c.Press up on the valve cap plunger for about one second to let out a little water.

d.Release the plunger.

e.Listen for air bubbles. Look for leaks around the“Side-Fill Cap”.

f.If you hear air bubbles or see leaks, retighten the “Side-Fill Cap” and repeat Steps a-e.

4. Wipe off excess water from the bottle.

5. Position the water bottle back into the unit by grasping the water bottle handle and lowering it gently into the

cabinet until the bottle sits on pads at bottom of cabinet. The “Side-Fill Cap”

must be facing the rear of the cabinet. When positioned properly, the plunger

will open and water will flow into the cabinet. The water level will

automatically remain at approximately 1/2” until the bottles are empty. Allow

the wick to saturate for 30 minutes before operation.

ABOUT HUMIDITY

Where you set your desired humidity levels depends on your personal comfort level, the outside temperature and the inside temperature.

NOTE: Recent CDC tests show that the chances of flu transmission are greatly reduced at levels of 43% humidity or higher.

You may wish to purchase a hygrometer* to measure the humidity level in your home. The following is a chart of recommended humidity settings.

*Model 1990 digital hygrometer is available for purchase by calling 1-800-547-3888.

NOTE: External hygrometer readings and humidistat readings may differ. Humidity levels can vary significantly even in one room.

6

IMPORTANT: Water damage may result if condensation starts to form on windows or walls. Humidity SET point should be lowered until condensation no longer forms. We recommend room humidity levels do not exceed 50%.

HUMIDISTAT

*When Outdoor

Recommended

Temperature is:

Indoor Relative

°F

° C

Humidity (RH)

 

 

is

-20

-30°

15 - 20%

-10°

-24°

20 - 25%

-18°

25 - 30%

10°

-12°

30 - 35%

20°

-6°

35 - 40%

30°

-1°

40 - 45%

* or higher

This unit has an humidistat located on the power cord. It

senses the percentage of humidity in the air and causes the humidifier to cycle on and off to maintain the selected humidity SET point. The fan will turn on if the

“ROOM” reading falls 3% below the SET point and will continue to run the humidifier until a reading of 1% above the SET point is achieved. The

fan will turn off until the ROOM reading falls 3% below the SET point again.

NOTE: Be sure the humidistat located on the power cord is at least two inches away from the wall, free from obstruction and away from any hot air source.

CONTROLS AND OPERATION

Where you set your desired humidity levels depends on your personal comfort level, outside temperature and inside temperature. A starting set point of 30%

to 40% ROOM humidity is recommended but there may be conditions that require a different setting.

NOTE: If excessive condensation occurs, reduce fan to a lower speed setting.

DIGITAL DISPLAY

Press the power button once to turn the unit on. Initially there is a slight delay while the electronics calibrate.Two bars (- -) appear during this time. Once calibration is complete a ROOM humidity reading between 20% - 95% will be displayed. A flashing 20 display indicates that the room humidity is less than 20%.

Fan Speed

®

Humidity Control

Power Button

Digital Display

NOTE: The default settings at initial startup are (LOW) F1 fan speed and 65%

SET humidity. After desired settings are selected by the user, they remain as long as the unit remains plugged in and there is no power interruption. When the unit is unplugged, the original default settings are reinstated.

7

FAN SPEED

This humidifier is equipped with a 3-speed fan control. Initially pressing the speed button will display the current speed setting. Each push of the speed button advances the mode through the settings of

(HIGH) F3, (MEDIUM) F2 , (LOW) F1 and A (AUTO).

HUMIDITY CONTROL

This button allows you to adjust the humidity to suit your needs. This humidifier will automatically cycle on and off as required to maintain the selected setting.

The readout displays both actual (ROOM) and desired (SET)

humidity percentage (%) settings when humidity control button is pressed. The ROOM readout has a range of 20% to 95% RH. If the room humidity level is equal to or less than 20% RH the display will show a flashing 20.

The SET readout has a range of 25% to 65% RH and is adjusted in increments of 5%. Initially depressing the humidity button will display the current humidity setting. Each additional press of the button increases the setting by 5%. A SET point of 65% RH operates the humidifier continuously regardless of the ROOM reading. The humidistat that senses the readings is located on the power cord. Insure that it is free from obstruction.

REFILL DISPLAY

For maximum run time both the bottle and the base should be filled. When both the bottle and the base are empty, the display alternates the room humidity reading and “F” (FILL), indicating additional water is required.

CHECK FILTER INDICATOR

The ability of any humidifier to efficiently supply humidity output relies heavily on the condition of the wick (filter). As the wick loads with impurities it gradually loses its wicking capability. When this happens, moisture output is reduced and the humidifier has to work longer to satisfy the selected setting. This humidifier has a check

filter reminder timed to appear after 720 hours of operation. When the display shows an alternating “CF” and “room humidity” setting at 5 second intervals it is a reminder to check the filter (wick) condition.

NOTE: Some areas have high water mineral content causing the need for more frequent wick changes.

Reset the Check Filter “CF” function each time you replace a wick by unplugging the unit from the power source for 1-2 minutes, then plug it in again.

Refer to “Replacement of Wick” section below. If a build-up of deposits or severe discoloration is evident, replace the wick to restore maximum efficiency.

CONTROL LOCK FEATURE

To avoid unwanted tampering with the humidifier settings the controls can be locked.

CONTROL LOCK ACTIVATION PROCEDURE

After the humidifier functions have been set up, hold the power

button down for 5 seconds. The display will show “CL” for 2 seconds; release the power button and the control will resume the display of the room humidity. The humidifier will continue to function with the locked in settings. If buttons are operated while “CL” is active, the “CL” is displayed and settings are unaffected.

8

Float with retainer clip (reference)
For normal operation,
float retainer clip should be in bottom section of middle slot.
Fasten float retainer clip into top section of middle slot for manual dryout

CONTROL LOCK DEACTIVATION PROCEDURE

To deactivate the “CL” function simply press and hold the power button for 5 seconds. “CL” will flash at the rate of 1 second on and 1 second off for a period of 2 seconds and then resume the display of room humidity.

MANUAL DRY OUT

Center Back of Cabinet

At the end of the humidification season or if

you do not plan to run the humidifier for one week or

more, a manual dry out

mechanism is provided to completely dry out the

humidifier cabinet. To do

this, simply slide the float retainer clip located at the

back of the humidifier

cabinet to the top of the middle slot,

as it is shown in the illustration. This will

hold the float in an upward position and the fan will run until you turn the power off.

NOTE: Be sure to reposition float retainer to the normal operating position (lowest portion of the slot) before using the humidifier again. See diagram for proper placement of the float retainer in the float retainer slot in the cabinet back.

REPLACEMENT OF WICKS

WARNING: To reduce the risk of shock or injury from moving parts, always unplug humidifier before removing or replacing any parts.

NOTE: This unit uses model HDC12 wicks. Replace the wick at the beginning of each season, and more often in areas with hard water. The wicks may be rinsed periodically in clear water only. Take care not to damage paper. To replace evaporative wicks in your humidifier, follow these instructions.

WICK REPLACEMENT

1.Unplug humidifier before removing the wick. We recommend moving the humidifier to an area where floor coverings are not susceptible to water damage, i.e. kitchen or bath areas.

2. Open the pivoting lid andremove water bottle s and set aside. Grasp the fan assembly front center and rear center. Left off and set aside on a clean flat surface. The wick housings are now accessible.

3. Lift wick housing out of cabinet and place where you may work on them comfortably. With a soft, absorbent cloth remove any remaining water or loose mineral deposits not trapped in the wick from the humidifier cabinet.

9

4.To open wick housing for easy access to wciks:

a.Deflect the two spring tabs on top of wick housing. Lift top cover off of the wick housing and set aside.

b.Remove the two wicks and dispose of properly. Clean internal parts of humidifier per the “Care and Maintenance” section (page 11-12) of your operator’s manual.

c.Install new wicks in the wick housing track with the glue beads of the wicking pad to the inside of the wick housing.

d.Return the top cover of the wick housing by installing cover over the fixed tabs on the rear of the lower wick housing and pivoting the cover down over the top of the wicks. Snap the cover over the two spring tabs.

e.Replace the wick housing in the cabinet.

NOTE: There is a front and back orientation. To position the wick housings correctly:

Be sure that the spring tabs of wick housings are towards front of cabinet.

Align the lower portion of the wick assembly over the six ribs located at the bottom rear of the humidifier cabinet.

f.Position fan assembly onto the cabinet. Make sure the fan venturi (lowest cylinder section below the fan) is placed inside the wick housing.

g.Replace water bottles.

h.Close the pivoting lid.

i.The unit is ready for use or to be stored away for next season.

Reinstall Top Onto Cabinet

Wick Housing Assembly

FR

O

N

T

Separate Upper and Front

Wick Housings

Remove Used Wicks and

Replace with New Wicks

Wick

Reassemble Wick Housing

Flat Portion of Wick Housing

Towards Back of Cabinet

FRONT

FRONT

Spring Tabs Toward Front of Cabinet

Water Bottle Placement

Right

Left

FRONT

FRONT

10

CARE AND MAINTENANCE

Cleaning your humidifier regularly helps eliminate odors and bacterial and fungal growth. Ordinary household bleach is a good disinfectant and can be used to wipe out the humidifier base and bottle/reservoir after cleaning .We recommend cleaning your humidifier at least once every two weeks to maintain optimum environmental conditions for your home. We also recommend using AIRCARE® or

Essick Air® Bacteriostat Treatment each time you refill your humidifier to eliminate bacterial growth. Add bacteriostat according to the instructions on the bottle. Please call 1-800-547-3888 to order Bacteriostat Treatment, part number1970.

CLEANING INSTRUCTIONS

WARNING: To reduce the risk of injury, fire or damage to humidifier, use only cleaners specifically recommended for humidifiers. Never use flammable, combustible or poisonous materials to clean your humidifier. To reduce the risk of scalds and damage to humidifier, never put hot water in humidifier.

CAUTION: Local water purity varies from area to area and under certain conditions water impurities and airborne bacteria may promote the growth of

microorganisms in the reservoir of the humidifier. To reduce bacteria growth that may cause odors and be harmful to your health, use only HDC12 replacement wicks. We also recommend using EPA approved bacteriostat, and humidifier cleaner, available at your local retailer.

NOTE: If it is necessary to use approved bacteriostat, follow directions correctly. Each bottle of the whole house humidifier holds 2.375 gallons/9 litres of water.

STEP 1:

To make cleaning easier, the humidifier cabinet should be empty and completely dried out. To accomplish this, use the following method:

a)On the back side of the humidifier cabinet, find the float retainer. In normal operating mode, the retainer will be positioned in the bottom of the slot in the cabinet. Push the retainer up to the top slot position in the cabinet until it locks in place. (See “Manual Dryout” section on page 9.)

b)Allow the humidifier to run until there is no water left in the cabinet.

c)Unplug humidifier.

WARNING: To reduce the risk of shock, always unplug humidifier before cleaning or servicing. If humidifier is not unplugged, fan could start after housing is removed.

STEP 2:

TO REMOVE PARTS

a)Open the pivoting lid, remove water bottles and set aside. Grasp the fan assembly front center and rear center. Lift off and set aside on a clean flat surface. The wick housings are now accessible.

b)Before removing the internal parts of the humidifier cabinet, we recommend moving the humidifier to an area where floor covers are not susceptible to water damage, i.e.; kitchen or bath area. Lift the wick housing assemblies out of the cabinet and set to the side.

c)Remove the water level float from the humidifier cabinet by pinching the float retainer to release it from the cabinet. The cabinet should be empty before cleaning. The cabinet consists of the lower areas of the humidifier where the wicks are during normal operation time.

d)Empty cabinet of any remaining water.

e)Proceed to “Bi-Weekly Maintenance” or “End of Season Maintenance”.

11

STEP 3:

Install new wicks (use only HDC12 wicks) into the location where the old wicks were positioned. Install the wick housings with the new wicks into the correct location. Reinstall the water level float.

NOTE: Float retainer should be positioned in normal operating mode .Ensure that it is positioned in the bottom of middle slot. (See diagram on page 9.)

Position the fan assembly on top of the cabinet and return the humidifier to the desired location for use. Refill humidifier bottles with fresh cool water and return them to the cabinet. Plug the unit into a 120 volt A.C. power outlet and set the controls per the “CONTROLS AND OPERATION” section (page 7) of this manual.

REMOVING SCALE

1.Fill the reservoir with water and add one 8 oz. (236 ml) cup of white vinegar. Let solution stay for 20 minutes.

2.Clean all interior surfaces with soft brush. Dampen soft cloth with white vinegar and wipe out reservoir to remove scale.

3.Rinse thoroughly with clean warm water to remove scale and cleaning solution before disinfecting the reservoir.

DISINFECTING UNIT

4.Fill humidifier cabinet with 1 teaspoon (5 ml) chlorine bleach in 1 gallon (3.8 litres) of water. Let solution stay for 20 minutes, carefully swishing every few minutes. Wipe down all surfaces normally exposed to water.

5.Empty humidifier cabinet after 20 minutes. Rinse with water until bleach smell is gone. Allow unit to dry out completely.

6.Water bottles should be cleaned in the same manner.

DISINFECTING BASE

7.Pour 1/2 gallon (1.89 litres) of water and 1/2 teaspoon (2.5 millilitres) chlorine bleach into the base. Let solution stay for 20 minutes, carefully swishing every few minutes. Wipe down all surfaces normally exposed to water.

8.Empty base after 20 minutes. Rinse with water until bleach smell is gone. Allow unit to dry out completely

SUMMER STORAGE

1.Clean unit as outlined in Care & Maintenance section.

2.Discard used wick and any water in the reservoir. Do not discard wick retainer. Allow to dry thoroughly before storage. Do not store with water inside reservoir.

3.Do not store unit in an attic or other high-temperature area, as damage is probable.

4.Install new wicks at beginning of season.

12

Troubleshooting

Trouble

Probable Cause

Remedy

Digital display not

• No Power.

• Check 120 volt power source.

illuminated.

• Power button has not

• Press power button once to

 

been depressed.

turn on.

 

 

 

Fan not operating

• ROOM humidity level is

• Raise SET humidity level if desired.

(digital display is

not 3% or more below

• Increase fan speed.

illuminated).

SET humidity level.

• Fill humidifier with water.

 

• Reservoir is empty.

 

Fan running

• Float is obstructed.

• Free float from obstruction.

continuously.

• SET humidity level is

• A SET humidity percentage of 65%

 

65%.

operates fan continuously. Adjust

 

• ROOM humidity level

SET level between 25% - 60%.

 

has not reached 1% or

 

• Continue to run until desired

 

more above SET

 

humidity level is achieved. This

 

humidity level.

 

may take up to 48 hours initially.

 

 

 

• CHECK FILTER (CF)

• Change to a fresh new wick

Poor evaporation of

message flashing.

NOTE: Local water varies from area

• Mineral buildup on filter

to area and under certain conditions

water.

 

(wick).

minerals will build up more rapidly in

 

 

the wick.

 

 

•Unplug unit for 1-2 minutes to reset

 

 

CF function

Neither bottle

• Bottle cap plunger

• Check orientation of bottles.

emptying.

missing post in bottom of

Reposition so that bottle cap

 

cabinet

plunger (stem) is depressed by

 

• Dirty wicks.

post in bottom of cabinet.

 

 

• Change to fresh new wick

 

 

(HDC12).

Bottles continue to

• Air is entering water

•Check “Side-Fill Caps” and valve

empty after complete

bottle.

assembly caps to ensure tightness.

filling.

 

• Check for leak in water bottle.

 

 

• Ensure gaskets for “Side Fill Caps”

 

 

are in place.

 

 

 

Left bottle not

• Right bottle contains

•The humidifier is designed so that

right bottle empties first.

emptying.

water.

• Ensure unit is level - Install caster

 

 

 

 

pads on front casters, as needed.

Bottle empties

• This is normal for initial

• Refill bottle for maximum

completely

filling

run time.

Refill display “F” is on

• Unit is not set up on a

•Position unit level. (Some carpet

before bottles are

level surface. (Float

padding may allow unit to sit lower in

empty.

will not sense water at

the front than rear with the additional

 

the front of the

weight of water bottles in the front of

 

cabinet.)

the cabinet. Install caster pads under

 

 

front casters (see page 5).

“CF” message does

• Loss of power or a

• Replace wick.

not appear even when

power failure has caused

 

wick is dirty or

the CF message to

 

humidifier performance

prematurely reset.

 

is clearly reduced.

 

 

Display blinks or fan

Normal

• No action required.

moves slightly when

 

 

humidifier is plugged in.

 

 

13

Troubleshooting (con’t)

Trouble

Probable Cause

Remedy

Alternating “F” and

• Reservoir is empty.

• Fill humidifier with water.

room humidity

 

 

displayed.

 

 

Alternating “CF” and

• Wick may be dirty.

• Replace wick, if needed

room humidity

 

• Unplug unit for 30 seconds

displayed.

 

and replug in to clear display.

Controls are

CONTROL LOCK (CL)

Hold power button down for 5

unresponsive or CL is

feature has been activated.

seconds. Control Lock function

displayed

 

will deactivate.

CAUTION: Use only EPA Registered Bacteriostat.

Use of other water treatment products may damage the filter elements and void warranty. Under no circumstances should you use water treatment products designed for Rotobelt or Ultrasonic humidifiers.

Parts List for AIRCARE® Evaporative Humidifier

Always order by part number - Not by key number

Key

Part No.

Description

No.

 

 

1

828741-2

Lid, Stationary

2

828855-2

Lid, Pivoting

3

1B72320

Control Assembly

4

824690

“Side-Fill Cap”

5

828192

Bottle, Left (Includes Key Nos. 4 & 7)

6

828726

Bottle, Right (Includes Key Nos. 4 & 7)

7

509229-1

Cap, Valve Assembly

8

828199-10

Cabinet

9

823726

Rail, Kick

10

STD601107

*Screw, Pan Hd. Ty “AB” #10 x 3/4

11

1B5460100

Caster, Twin Wheel

12

830560-1

Support, Filter (Wick) Lower

13

HDC12

†Wick, Evaporative

14

828534

Support, Motor

15

1B72493

Motor

16

822406-1

Fan

17

828197

Support, Upper Filter (Wick)

18

824101

Retainer, Float

19

828200-3

Float, Water Level

20

828920

Pad, Caster (Two)

----

1B72889

Operator’s Manual (not illustrated)

† Stock item - Purchase locally or call 1-800-547- 3888 from the USA or 1-888-744-4687 from Canada to order.

Any attempt to repair the control assembly or motor voids the warranty and may create a hazard unless repair is done by a qualified service technician.

To order parts, call: in USA 1-800-547-3888

in Canada: 1-888-744-4687

14

Repair Parts

Parts List for AIRCARE 14 Gallon Series Model No.s EA1407/ HD1409

15

Loading...
+ 33 hidden pages