A terminal to which DC voltage is applied or through which direct current flows.
A0011197
Alternating current
A0011198
A terminal to which alternating voltage (sine-wave) is applied or through which alternating current flows.
Ground connection
A grounded terminal which, as far as the operator is concerned, is grounded via a grounding system.
A0011200
Protective ground connection
A terminal which must be connected to ground prior to establishing any other connections.
A0011199
Equipotential connection
A connection that has to be connected to the plant grounding system: This may be a potential equalization
A0011201
line or a star grounding system depending on national or company codes of practice.
Tool symbols
Symbol
Meaning
Torx screwdriver
A0013442
Flat blade screwdriver
A0011220
Phillips head screwdriver
A0011219
Allen key
A0011221
Hexagon wrench
A0011222
Symbols for certain types of information
Symbol
Meaning
Allowed
Indicates procedures, processes or actions that are allowed.
A0011182
Preferred
Indicates procedures, processes or actions that are preferred.
A0011183
Forbidden
Indicates procedures, processes or actions that are forbidden.
A0011184
Tip
Indicates additional information.
A0011193
Reference to documentation
Refers to the corresponding device documentation.
A0011194
Reference to page
Refers to the corresponding page number.
A0011195
Reference to graphic
Refers to the corresponding graphic number and page number.
A0011196
Endress+Hauser3
Symbols in graphics
,…,
-
.
SymbolMeaning
1, 2, 3,...
A, B, C, ...
A-A, B-B, C-C, ...
Item numbers
Series of steps
Views
Sections
Flow direction
A0013441
Hazardous area
Indicates a hazardous area.
A0011187
Safe area (non-hazardous area)
Indicates a non-hazardous location.
A0011188
Function and system design
Proline t-mass A 150
Measuring principle
Measuring system
The thermal measuring principle is based on the cooling of a heated resistance thermometer (PT100), from
which heat is extracted by the passing gas. The gas passes two PT100 resistance thermometers in the
measurement section. One of these is used in the conventional way as a temperature probe, while the other
serves as a heating element. The temperature probe monitors and records the effective process temperature
while the heated resistance thermometer is kept at a constant differential temperature (compared to the
measured gas temperature) by controlling the electrical current used by the heating element. The greater the
mass current passing over the heated resistance thermometer, the greater the extent to which cooling takes
place and therefore the stronger the current required to maintain a constant differential temperature. This means
that the heat current measured is an indicator of the mass flow rate of the gas.
A0016823
The device consists of a transmitter and a sensor.
One device version is available: compact version comprising transmitter and sensor.
Transmitter
t-mass 150
Materials:
Aluminum coating AlSi10Mg
Configuration:
• Four-line local display with key operation and guided menu ("Setup") for
applications
• Operating tools (e.g. FieldCare)
A0015480
Other special features:
May also be ordered without local display
4Endress+Hauser
Proline t-mass A 150
Sensor
t-mass A
Characteristic values
Measured variableDirect measured variables
• Mass flow
• Gas temperature
Calculated measured variables
• Corrected volume flow
• FAD (free air delivery) volume flow
Measuring range
The available measuring range depends on the choice of gas, the size of the pipe and the use of a flow conditioner.
The measuring device is calibrated with air (under ambient conditions) and the value is converted in order to
adapt it to the customer's gas if necessary.
To obtain information on other gases and process conditions, please contact your Endress+Hauser sales
office.
To calculate the measuring range with and without a flow conditioner (option L (®ä 12)
(®ä 15)), use the Applicator (®ä 31) product selection tool.
The following tables list the ranges available for air (without flow conditioner).
Measuring range "Calibration flow", option G and H (®ä 12)
SI units for EN (DIN) flange versions
DN
[mm]min.Max.min.Max.min.Max.
150.5530.38410.443
2522001.51551.6164
4065554.64294.9453
50109107.77048.2744
[kg/h][Nm3/h] at 0 °C (1.013 bar a)[Nm3/h at 15 °C (1.013 bar a)
US units for ASME flange versions
DN
[in]min.Max.min.Max.min.Max.
½1.11160.23240.2425
14.44400.9911.096
1½13.212202.72522.9266
222.020024.54134.8436
[lb/h][Scf/min] at 32 °F (14.7 psi a)[Scf/min] at 59 °F (14.7 psi a)
Endress+Hauser5
Proline t-mass A 150
Measuring range "Calibration flow" option K (®ä 12)
SI units for EN (DIN) flange versions
DN[kg/h][Nm3/h] at 0 °C (1.013 bar a)[Nm3/h at 15 °C (1.013 bar a)
[mm]min.Max.min.Max.min.Max.
150.5800.38620.2465
2523001.52321.0245
4068334.66442.3681
501013657.710564.81116
US units for ASME flange versions
DN
[in]min.Max.min.Max.min.Max.
½1.11740.23360.2438
14.46600.91361.0144
1½13.218302.73782.9399
222.030034.56204.8656
[lb/h][Scf/min] at 32 °F (14.7 psi a)[Scf/min] at 59 °F (14.7 psi a)
Operable flow range
Over 100:1 (over 150:1 for calibration option code K).
Even in the extended measuring range (above the calibrated end value), the flow rate is captured and provided
as an output signal. However, the extended range is not subject to the specified measuring uncertainty.
Output
Output signalCurrent output
Current output
Maximum output values
Load
Resolution
Damping
Assignable measured
variables
4-20 mA HART, active
• DC 24 V (no flow)
• 22 mA
If the option Defined value is selected in the Failure mode parameter :
22.5 mA
0 to 750 Ω
16 Bit or 0.38 µA
Adjustable: 0 to 999 s
• Mass flow
• Corrected volume flow
• FAD volume flow
• Temperature
Pulse/frequency/switch output
Function
Version
Maximum input values
Voltage drop
Pulse output
Can be set to pulse, frequency or switching output
Passive, open collector
• DC 30 V
• 25 mA
For 25 mA: £ DC 2 V
6Endress+Hauser
Proline t-mass A 150
Pulse width
Pulse value
Assignable measured
variables
Frequency output
Maximum frequency
Damping
Pulse/pause ratio
Assignable measured
variables
Switching output
Switching behavior
Switching delay
Number of switching cycles
Assignable functions
Adjustable: 0.5 to 2000 ms ® pulse rate: 0 to 1000 Pulse/s
Adjustable
• Mass flow
• Corrected volume flow
• FAD volume flow
Adjustable: 0 to 1000 Hz
Adjustable: 0 to 999 s
1:1
• Mass flow
• Corrected volume flow
• FAD volume flow
• Temperature
Binary, conductive or non-conductive
Adjustable: 0 to 100 s
Unlimited
• Off
• On
• Diagnostic behavior
• Limit value
• Status
Signal on alarm
Depending on the interface, failure information is displayed as follows:
Current output
Failure mode
Minimum alarm
Maximum alarm
Adjustable value
Can be selected (as per NAMUR recommendation NE 43)
3.6 mA
22 mA
3.59 to 22.5 mA
Pulse/frequency/switch output
Pulse output
Failure mode
Frequency output
Failure mode
Switching output
Failure mode
Choose from:
• Actual value
• No pulses
Choose from:
• Actual value
• Defined value: 0 to 1250 Hz
• 0 Hz
Choose from:
• Current status
• Open
• Closed
Local display
Plain text display
With information on cause and corrective action
Endress+Hauser7
Status signal as per NAMUR recommendation NE 107
Operating tool
• Via digital communication: HART protocol
• Via service interface
Proline t-mass A 150
Plain text display
Low flow cut off
Galvanic isolation
The switch point for low flow cut off is programmable.
The following connections are galvanically isolated from each other:
• Outputs
• Voltage supply
Protocol-specific dataHART
Manufacturer ID
Device type ID0x66
HART protocol revision6.0
Device description files (DTM,
DD)
HART loadMin. 250 Ω
Dynamic variablesThe measured variables can be freely assigned to the dynamic variables.
With information on cause and corrective action
Additional information on remote operation(®ä 28)
0x11
Information and files under:
www.endress.com
Measured variables for PV (primary dynamic variable)
• Mass flow
• Corrected volume flow
• FAD volume flow
• Temperature
Measured variables for SV, TV, QV (secondary, tertiary and quaternary
dynamic variable)
• Mass flow
• Corrected volume flow
• FAD volume flow
• Temperature
• Totalizer
8Endress+Hauser
Proline t-mass A 150
4
1
2
3
12
23
2224
2526
27
Power supply
Terminal assignmentTransmitter
Connection version 4-20 mA HART, pulse/frequency/switching output
1Supply voltage
2Signal transmission: Pulse/frequency/switching output
3Signal transmission: 4-20 mA HART
4Ground terminal for cable shield
A0017178
Power consumption
Supply voltage
Order characteristic for
"Power supply"
Option DDC 24 V (18 to 30 V)
1 (L+)2 (L-)
Terminal numbers
Signal transmission
Order characteristic for
"Output"
Option A4-20 mA HART active-
Option B4-20 mA HART activePulse/frequency/switch output
Option K-Pulse/frequency/switch output
Output 1Output 2
26 (+)27 (-)24 (+)25 (-)
Terminal numbers
Supply voltage
DC 24 V (18 to 30 V)
The power supply circuit must comply with SELV/PELV requirements.