• Make sure the document is stored in a safe place such that it is always available when
working on or with the device.
• To avoid danger to individuals or the facility, read the "Basic safety instructions" section
carefully, as well as all other safety instructions in the document that are specific to
working procedures.
• The manufacturer reserves the right to modify technical data without prior notice. Your
Endress+Hauser Sales Center will supply you with current information and updates to
these instructions.
16.12 Certificates and approvals .............. 103
16.13 Application packages .................. 104
16.14 Accessories ......................... 104
16.15 Supplementary documentation ........... 104
Index ................................. 106
4Endress+Hauser
Proline Prosonic Flow E 100 HARTAbout this document
DANGER
WARNING
CAUTION
NOTICE
1 About this document
1.1 Document function
These Operating Instructions contain all the information that is required in various phases
of the life cycle of the device: from product identification, incoming acceptance and
storage, to mounting, connection, operation and commissioning through to
troubleshooting, maintenance and disposal.
1.2 Symbols used
1.2.1 Safety symbols
SymbolMeaning
DANGER!
This symbol alerts you to a dangerous situation. Failure to avoid this situation will
result in serious or fatal injury.
WARNING!
This symbol alerts you to a dangerous situation. Failure to avoid this situation can
result in serious or fatal injury.
CAUTION!
This symbol alerts you to a dangerous situation. Failure to avoid this situation can
result in minor or medium injury.
NOTE!
This symbol contains information on procedures and other facts which do not result in
personal injury.
1.2.2 Electrical symbols
SymbolMeaning
Direct current
Alternating current
Direct current and alternating current
Ground connection
A grounded terminal which, as far as the operator is concerned, is grounded via a
grounding system.
Protective Earth (PE)
A terminal which must be connected to ground prior to establishing any other
connections.
The ground terminals are situated inside and outside the device:
• Inner ground terminal: Connects the protectiv earth to the mains supply.
• Outer ground terminal: Connects the device to the plant grounding system.
1.2.3 Tool symbols
SymbolMeaning
Allen key
Open-ended wrench
Endress+Hauser5
About this documentProline Prosonic Flow E 100 HART
A
1.
1.
-
.
1.2.4 Symbols for certain types of information
SymbolMeaning
Permitted
Procedures, processes or actions that are permitted.
Preferred
Procedures, processes or actions that are preferred.
Forbidden
Procedures, processes or actions that are forbidden.
Tip
Indicates additional information.
Reference to documentation.
Reference to page.
Reference to graphic.
Notice or individual step to be observed.
, 2., 3.…Series of steps.
Result of a step.
Help in the event of a problem.
Visual inspection.
1.2.5 Symbols in graphics
SymbolMeaning
1, 2, 3, ...Item numbers
, 2., 3., …Series of steps
A, B, C, ...Views
A-A, B-B, C-C, ...Sections
Hazardous area
Safe area (non-hazardous area)
Flow direction
1.3 Documentation
For an overview of the scope of the associated Technical Documentation, refer to the
following:
• The W@M Device Viewer : Enter the serial number from the nameplate
(www.endress.com/deviceviewer)
• The Endress+Hauser Operations App: Enter the serial number from the nameplate
or scan the 2-D matrix code (QR code) on the nameplate.
For a detailed list of the individual documents along with the documentation code
→ 104
6Endress+Hauser
Proline Prosonic Flow E 100 HARTAbout this document
1.3.1 Standard documentation
Document typePurpose and content of the document
Technical InformationPlanning aid for your device
The document contains all the technical data on the device and provides
an overview of the accessories and other products that can be ordered for
the device.
Sensor Brief Operating Instructions Guides you quickly to the 1st measured value - Part 1
The Sensor Brief Operating Instructions are aimed at specialists with
responsibility for installing the measuring device.
• Incoming acceptance and product identification
• Storage and transport
• Installation
Transmitter Brief Operating
Instructions
Description of Device ParametersReference for your parameters
Guides you quickly to the 1st measured value - Part 2
The Transmitter Brief Operating Instructions are aimed at specialists with
responsibility for commissioning, configuring and parameterizing the
measuring device (until the first measured value).
• Product description
• Installation
• Electrical connection
• Operation options
• System integration
• Commissioning
• Diagnostic information
The document provides a detailed explanation of each individual
parameter in the Expert operating menu. The description is aimed at
those who work with the device over the entire life cycle and perform
specific configurations.
Additional documents are supplied depending on the device version ordered: Always
comply strictly with the instructions in the supplementary documentation. The
supplementary documentation is an integral part of the device documentation.
1.4 Registered trademarks
HART®
Registered trademark of the FieldComm Group, Austin, Texas, USA
Microsoft®
Registered trademark of the Microsoft Corporation, Redmond, Washington, USA
Endress+Hauser7
Basic safety instructionsProline Prosonic Flow E 100 HART
2 Basic safety instructions
2.1 Requirements for the personnel
The personnel for installation, commissioning, diagnostics and maintenance must fulfill
the following requirements:
Trained, qualified specialists must have a relevant qualification for this specific function
‣
and task.
Are authorized by the plant owner/operator.
‣
Are familiar with federal/national regulations.
‣
Before starting work, read and understand the instructions in the manual and
‣
supplementary documentation as well as the certificates (depending on the
application).
Follow instructions and comply with basic conditions.
‣
The operating personnel must fulfill the following requirements:
Are instructed and authorized according to the requirements of the task by the facility's
‣
owner-operator.
Follow the instructions in this manual.
‣
2.2 Designated use
Application and media
Depending on the version ordered, the measuring device can also measure potentially
explosive, flammable, poisonous and oxidizing media.
Measuring devices for use in hazardous areas, in hygienic applications or where there is an
increased risk due to process pressure, are labeled accordingly on the nameplate.
To ensure that the measuring device remains in proper condition for the operation time:
Keep within the specified pressure and temperature range.
‣
Only use the measuring device in full compliance with the data on the nameplate and
‣
the general conditions listed in the Operating Instructions and supplementary
documentation.
Based on the nameplate, check whether the ordered device is permitted for the
‣
intended use in the hazardous area (e.g. explosion protection, pressure vessel safety).
Use the measuring device only for media to which the process-wetted materials are
‣
sufficiently resistant.
If the measuring device is not operated at atmospheric temperature, compliance with
‣
the relevant basic conditions specified in the associated device documentation is
absolutely essential: "Documentation" section→ 6.
Protect the measuring device permanently against corrosion from environmental
‣
influences.
Incorrect use
Non-designated use can compromise safety. The manufacturer is not liable for damage
caused by improper or non-designated use.
WARNING
L
Danger of breakage due to corrosive or abrasive fluids!
Verify the compatibility of the process fluid with the sensor material.
‣
Ensure the resistance of all fluid-wetted materials in the process.
‣
Keep within the specified pressure and temperature range.
‣
8Endress+Hauser
Proline Prosonic Flow E 100 HARTBasic safety instructions
NOTICE
Verification for borderline cases:
For special fluids and fluids for cleaning, Endress+Hauser is glad to provide assistance
‣
in verifying the corrosion resistance of fluid-wetted materials, but does not accept any
warranty or liability as minute changes in the temperature, concentration or level of
contamination in the process can alter the corrosion resistance properties.
Residual risks
WARNING
L
The electronics and the medium may cause the surfaces to heat up. This presents a
burn hazard!
For elevated fluid temperatures, ensure protection against contact to prevent burns.
‣
2.3 Workplace safety
For work on and with the device:
Wear the required personal protective equipment according to federal/national
‣
regulations.
For welding work on the piping:
Do not ground the welding unit via the measuring device.
‣
If working on and with the device with wet hands:
Due to the increased risk of electric shock, gloves must be worn.
‣
2.4 Operational safety
Risk of injury.
Operate the device in proper technical condition and fail-safe condition only.
‣
The operator is responsible for interference-free operation of the device.
‣
Conversions to the device
Unauthorized modifications to the device are not permitted and can lead to unforeseeable
dangers.
If, despite this, modifications are required, consult with Endress+Hauser.
‣
Repair
To ensure continued operational safety and reliability,
Carry out repairs on the device only if they are expressly permitted.
‣
Observe federal/national regulations pertaining to repair of an electrical device.
‣
Use original spare parts and accessories from Endress+Hauser only.
‣
2.5 Product safety
This measuring device is designed in accordance with good engineering practice to meet
state-of-the-art safety requirements, has been tested, and left the factory in a condition in
which it is safe to operate.
It meets general safety standards and legal requirements. It also complies with the EU
directives listed in the device-specific EU Declaration of Conformity. Endress+Hauser
confirms this by affixing the CE mark to the device.
Endress+Hauser9
Basic safety instructionsProline Prosonic Flow E 100 HART
2.6 IT security
We only provide a warranty if the device is installed and used as described in the
Operating Instructions. The device is equipped with security mechanisms to protect it
against any inadvertent changes to the device settings.
IT security measures in line with operators' security standards and designed to provide
additional protection for the device and device data transfer must be implemented by the
operators themselves.
10Endress+Hauser
Proline Prosonic Flow E 100 HARTProduct description
3
2
1
4
5
6
7
3 Product description
The device consists of a transmitter and a sensor.
The device is available as a compact version:
The transmitter and sensor form a mechanical unit.
3.1 Product design
3.1.1 Device version with HART communication type
A0023153
1Important components of a measuring device
1Sensor
2Transmitter housing
3Main electronics module
4Transmitter housing cover
5Transmitter housing cover (version for optional local display)
6Local display (optional)
7Main electronics module (with bracket for optional local display)
Endress+Hauser11
Incoming acceptance and product identificationProline Prosonic Flow E 100 HART
1
2
1
2
Order code:
Ser. no.:
Ext. ord. cd.:
i
i
Date:
4 Incoming acceptance and product
identification
4.1 Incoming acceptance
Are the order codes on the
A0028673
delivery note (1) and the
product sticker (2) identical?
Are the goods undamaged?
A0028673
Do the nameplate data
A0028673
A0028673
match the ordering
information on the delivery
note?
Is the CD-ROM with the
Technical Documentation
(depends on device version)
and documents present?
• If one of the conditions is not satisfied, contact your Endress+Hauser Sales Center.
• Depending on the device version, the CD-ROM might not be part of the delivery!
The Technical Documentation is available via the Internet or via the Endress+HauserOperations App, see the "Product identification" section → 13.
4.2 Product identification
The following options are available for identification of the measuring device:
• Nameplate specifications
• Order code with breakdown of the device features on the delivery note
• Enter serial numbers from nameplates in W@M Device Viewer
(www.endress.com/deviceviewer): All information about the measuring device is
displayed.
• Enter the serial number from the nameplates into the Endress+Hauser Operations App
or scan the 2-D matrix code (QR code) on the nameplate with the Endress+Hauser
12Endress+Hauser
Operations App: all the information for the measuring device is displayed.
Proline Prosonic Flow E 100 HARTIncoming acceptance and product identification
i
1
2
3
4
5
6
7
8
9
101112
Order code:
Ext. ord. cd.:
Ser. no.:
13
For an overview of the scope of the associated Technical Documentation, refer to the
following:
• The chapters "Additional standard documentation on the device" → 7 and
"Supplementary device-dependent documentation" → 7
• The W@M Device Viewer: Enter the serial number from the nameplate
(www.endress.com/deviceviewer)
• The Endress+Hauser Operations App: Enter the serial number from the nameplate or
scan the 2-D matrix code (QR code) on the nameplate.
4.2.1 Transmitter nameplate
A0030222
2Example of a transmitter nameplate
1Manufacturing location
2Name of the transmitter
3Order code
4Serial number (ser. no.)
5Extended order code (Ext. ord. cd.)
6Electrical connection data, e.g. available inputs and outputs, supply voltage
7Permitted ambient temperature (Ta)
8Degree of protection
92-D matrix code
10 Document number of safety-related supplementary documentation
11 Manufacturing date: year-month
12 CE mark, C-Tick
13 Firmware version (FW)
Endress+Hauser13
Incoming acceptance and product identificationProline Prosonic Flow E 100 HART
Date:
Size:
Material:
Tm:
Patents
i
Ptest =
i
Order code:
Ext. ord. cd.:
Ser. no.:
322541-0000
0044
11
1
2
3 4 5 67
12
8
9
13 1415
16
17
19
18
4.2.2 Sensor nameplate
A0013907
3Example of a sensor nameplate
1Manufacturing location
2Name of the sensor
3Order code
4Serial number (ser. no.)
5Extended order code (Ext. ord. cd.)
6Nominal diameter of sensor
7Test pressure of the sensor
8Flange nominal diameter/nominal pressure
9Material of measuring tube and manifold
10 Medium temperature range
11 CE mark, C-Tick
12 Additional information on version: certificates, approvals
13 Manufacturing date: year-month
14 Flow direction
15 2-D matrix code
16 Degree of protection
17 Approval information for explosion protection and Pressure Equipment Directive
18 Permitted ambient temperature (Ta)
19 Document number of safety-related supplementary documentation
Order code
The measuring device is reordered using the order code.
Extended order code
• The device type (product root) and basic specifications (mandatory features) are
always listed.
• Of the optional specifications (optional features), only the safety and approvalrelated specifications are listed (e.g. LA). If other optional specifications are also
ordered, these are indicated collectively using the # placeholder symbol (e.g. #LA#).
• If the ordered optional specifications do not include any safety and approval-related
specifications, they are indicated by the + placeholder symbol (e.g. XXXXXX-ABCDE
+).
14Endress+Hauser
Proline Prosonic Flow E 100 HARTStorage and transport
5 Storage and transport
5.1 Storage conditions
Observe the following notes for storage:
Store in the original packaging to ensure protection from shock.
‣
Do not remove protective covers or protective caps installed on process connections.
‣
They prevent mechanical damage to the sealing surfaces and contamination in the
measuring tube.
Protect from direct sunlight to avoid unacceptably high surface temperatures.
‣
Store in a dry and dust-free place.
‣
Do not store outdoors.
‣
Storage temperature→ 96
5.2 Transporting the product
Transport the measuring device to the measuring point in the original packaging.
A0029252
Do not remove protective covers or caps installed on process connections. They
prevent mechanical damage to the sealing surfaces and contamination in the
measuring tube.
5.2.1 Measuring devices without lifting lugs
WARNING
L
Center of gravity of the measuring device is higher than the suspension points of the
webbing slings.
Risk of injury if the measuring device slips.
Secure the measuring device against slipping or turning.
‣
Observe the weight specified on the packaging (stick-on label).
‣
A0029214
Endress+Hauser15
InstallationProline Prosonic Flow E 100 HART
5.2.2 Measuring devices with lifting lugs
CAUTION
L
Special transportation instructions for devices with lifting lugs
Only use the lifting lugs fitted on the device or flanges to transport the device.
‣
The device must always be secured at two lifting lugs at least.
‣
5.2.3 Transporting with a fork lift
If transporting in wood crates, the floor structure enables the crates to be lifted lengthwise
or at both sides using a forklift.
5.3 Packaging disposal
All packaging materials are environmentally friendly and 100% recyclable:
• Measuring device secondary packaging: polymer stretch film that conforms to EC
Directive 2002/95/EC (RoHS).
• Packaging:
– Wood crate, treated in accordance with ISPM 15 standard, which is confirmed by the
affixed IPPC logo.
or
– Carton in accordance with European Packaging Directive 94/62EC; recyclability is
confirmed by the affixed RESY symbol.
• Seaworthy packaging (optional): Wood crate, treated in accordance with ISPM 15
standard, which is confirmed by the affixed IPPC logo.
No special measures such as supports etc. are necessary. External forces are absorbed by
the construction of the device.
6.1.1 Mounting position
Mounting location
A0015543
16Endress+Hauser
Proline Prosonic Flow E 100 HARTInstallation
10 × DN3 × DN
2
4
15 × DN3 × DN
15 × DN3 × DN10 × DN3 × DN
Orientation
The direction of the arrow on the nameplate helps you to install the sensor according to
the flow direction (direction of medium flow through the piping).
• Install the measuring device in a parallel plane free of external mechanical stress.
• The internal diameter of the pipe must match the internal diameter of the sensor:
see the "Technical Information" device document, "Design and dimensions" section.
A0015895
OrientationCompact version
AVertical orientation
A0015545
BHorizontal orientation, transmitter
head up
A0015589
CHorizontal orientation, transmitter
head down
A0015590
DHorizontal orientation, transmitter
head at side
A0015592
Inlet and outlet runs
If possible, the sensor should be installed upstream from valves, T-pieces, elbows etc. To
attain the specified level of accuracy of the measuring device, the below mentioned inlet
and outlet runs must be maintained at minimum. If there are several flow disturbances
present, the longest specified inlet run must be maintained.
Endress+Hauser17
A0033877
4Minimum inlet and outlet runs with various flow obstructions
If installing an external device, observe the specified distance.
A0015901
PT Pressure
Installation dimensions
For the dimensions and installation lengths of the device, see the "Technical
Information" document, "Mechanical construction" section.
6.1.2 Requirements from environment and process
Ambient temperature range
Transmitter–25 to +60 °C (–13 to +140 °F)
Local display–20 to +60 °C (–4 to +140 °F), the readability of the display may be
impaired at temperatures outside the temperature range.
Sensor–25 to +60 °C (–13 to +140 °F)
If operating outdoors:
‣
Avoid direct sunlight, particularly in warm climatic regions.
System pressure
It is important that cavitation does not occur, or that gases entrained in the liquids do not
outgas.
Cavitation is caused if the pressure drops below the vapor pressure:
In suction lines
Ensure the system pressure is sufficiently high to prevent cavitation and outgassing.
‣
For this reason, the following mounting locations are recommended:
• At the lowest point in a vertical pipe
• Downstream from pumps (no danger of vacuum)
A0028777
Thermal insulation
In the case of some fluids, it is important to keep the heat radiated from the sensor to the
transmitter to a low level. A wide range of materials can be used for the required
insulation.
18Endress+Hauser
Proline Prosonic Flow E 100 HARTInstallation
t
a
NOTICE
Electronics overheating on account of thermal insulation!
Observe maximum permitted insulation height of the transmitter neck so that the
‣
transmitter head is completely free.
A0034104
tMaximum insulation thickness 2 cm (0.79 in)
aMinimum distance from transmitter to insulation
6.2 Mounting the measuring device
6.2.1 Required tools
For sensor
For flanges and other process connections: Corresponding mounting tools
6.2.2 Preparing the measuring device
1.Remove all remaining transport packaging.
2.Remove any protective covers or protective caps present from the sensor.
3.Remove stick-on label on the electronics compartment cover.
6.2.3 Mounting the measuring device
WARNING
L
Danger due to improper process sealing!
Ensure that the inside diameters of the gaskets are greater than or equal to that of the
‣
process connections and piping.
Ensure that the gaskets are clean and undamaged.
‣
Install the gaskets correctly.
‣
1.Ensure that the direction of the arrow on the sensor matches the flow direction of
the medium.
2.Install the measuring device or turn the transmitter housing so that the cable entries
do not point upwards.
A0029263
Endress+Hauser19
InstallationProline Prosonic Flow E 100 HART
1.
2.
3.
4.
5.
6.
7.
4.
4.
3 mm
6.2.4 Turning the display module
The local display is only available with the following device version:
Order code for "Display; Operation", option B: 4-line; lit, via communication
The display module can be turned to optimize display readability.
Aluminum housing version, AlSi10Mg, coated
A0023192
6.3 Post-mounting check
Is the device undamaged (visual inspection)?
Does the measuring device conform to the measuring point specifications?
For example:
• Process temperature
• Process pressure (refer to the section on "Pressure-temperature ratings" in the "Technical
Information" document)
• Ambient temperature range
• Measuring range
Has the correct orientation for the sensor been selected → 17?
• According to sensor type
• According to medium temperature
• According to medium properties (outgassing, with entrained solids)
Does the arrow on the sensor match the direction of flow of the medium through the piping
→ 17?
Are the measuring point identification and labeling correct (visual inspection)?
Is the device adequately protected from precipitation and direct sunlight?
Are the securing screw and securing clamp tightened securely?
20Endress+Hauser
Proline Prosonic Flow E 100 HARTElectrical connection
7 Electrical connection
NOTICE
The measuring device does not have an internal circuit breaker.
For this reason, assign the measuring device a switch or power-circuit breaker so that
‣
the power supply line can be easily disconnected from the mains.
Although the measuring device is equipped with a fuse, additional overcurrent
‣
protection (maximum 16 A) should be integrated into the system installation.
7.1 Connection conditions
7.1.1 Required tools
• For cable entries: Use corresponding tools
• For securing clamp (on aluminum housing): Allen screw3 mm
• For securing screw (for stainless steel housing): open-ended wrench 8 mm
• Wire stripper
• When using stranded cables: crimper for wire end ferrule
7.1.2 Requirements for connecting cable
The connecting cables provided by the customer must fulfill the following requirements.
Electrical safety
In accordance with applicable federal/national regulations.
Permitted temperature range
• The installation guidelines that apply in the country of installation must be observed.
• The cables must be suitable for the minimum and maximum temperatures to be
expected.
Power supply cable
Standard installation cable is sufficient.
Signal cable
Current output 4 to 20 mA HART
A shielded cable is recommended. Observe grounding concept of the plant.
Pulse/frequency/switch output
Standard installation cable is sufficient.
Cable diameter
• Cable glands supplied:
M20 × 1.5 with cable ⌀ 6 to 12 mm (0.24 to 0.47 in)
• Spring terminals:
Wire cross-sections 0.5 to 2.5 mm2 (20 to 14 AWG)
Endress+Hauser21
Electrical connectionProline Prosonic Flow E 100 HART
L
L
26
27
+
_
24
25
1
2
+
_
+
_
1
2
3
7.1.3 Terminal assignment
Transmitter
Connection version 4-20 mA HART with pulse/frequency/switch output
Order code for "Housing":
Option A: compact, coated aluminum
5Terminal assignment 4-20 mA HART with pulse/frequency/switch output
1Power supply: DC 24 V
2Output 1: 4-20 mA HART (active)
3Output 2: pulse/frequency/switch output (passive)
Connection methods available
Outputs
Power
supply
Possible options for order code
"Electrical connection"
• Option B: thread M20x1
• Option C: thread G ½"
• Option D: thread NPT ½"
A0016888
Terminal number
Order code
"Output"
Option BDC 24 V4-20 mA HART (active)Pulse/frequency/switch
Order code for "Output":
Option B: 4-20 mA HART with pulse/frequency/switch output
Power supplyOutput 1Output 2
2 (L-)1 (L+)27 (–)26 (+)25 (–)24 (+)
output (passive)
7.1.4 Preparing the measuring device
NOTICE
Insufficient sealing of the housing!
Operational reliability of the measuring device could be compromised.
Use suitable cable glands corresponding to the degree of protection.
‣
1.Remove dummy plug if present.
2.If the measuring device is supplied without cable glands:
Provide suitable cable gland for corresponding connecting cable.
3.If the measuring device is supplied with cable glands:
Observe requirements for connecting cables → 21.
22Endress+Hauser
Proline Prosonic Flow E 100 HARTElectrical connection
12
A
7.2 Connecting the measuring device
NOTICE
Limitation of electrical safety due to incorrect connection!
Have electrical connection work carried out by appropriately trained specialists only.
‣
Observe applicable federal/national installation codes and regulations.
‣
Comply with local workplace safety regulations.
‣
The power unit must be tested to ensure it meets safety requirements (e.g. PELV,
‣
SELV).
7.2.1 Connecting the transmitter
ACompact, coated aluminum
1Cable entry for signal transmission
2Cable entry for supply voltage
A0030221
Endress+Hauser23
Electrical connectionProline Prosonic Flow E 100 HART
10 (0.4)
1
3 mm
6Device version with connection example
1Cable
Disconnect the local display from the main electronics module: Operating Instructions
for the device .
Connect the cable in accordance with the terminal assignment .
‣
7.2.2 Ensuring potential equalization
Requirements
No special measures for potential equalization are required.
A0033696
24Endress+Hauser
Proline Prosonic Flow E 100 HARTElectrical connection
4
4...20 mA
5
2
1
3
6
1
2
3
12345
7.3 Special connection instructions
7.3.1 Connection examples
Current output 4 to 20 mA HART
A0029055
7Connection example for 4 to 20 mA HART current output (active)
1Automation system with current input (e.g. PLC)
2Cable shield: the cable shield must be grounded at both ends to comply with EMC requirements; observe cable
specifications
3Connection for HART operating devices → 35
4Resistor for HART communication (≥ 250 Ω): observe maximum load
5Analog display unit: observe maximum load
6Transmitter
Pulse/frequency output
8Connection example for pulse/frequency output (passive)
1Automation system with pulse/frequency input (e.g. PLC)
2Power supply
3Transmitter: Observe input values
A0028761
Endress+Hauser25
Electrical connectionProline Prosonic Flow E 100 HART
1
2
3
Switch output
A0028760
9Connection example for switch output (passive)
1Automation system with switch input (e.g. PLC)
2Power supply
3Transmitter: Observe input values
7.4 Ensuring the degree of protection
The measuring device fulfills all the requirements for the IP66/67 degree of protection,
Type 4X enclosure.
To guarantee IP66/67 degree of protection, Type 4X enclosure, carry out the following
steps after the electrical connection:
1.Check that the housing seals are clean and fitted correctly.
2.Dry, clean or replace the seals if necessary.
3.Tighten all housing screws and screw covers.
4.Firmly tighten the cable glands.
5.To ensure that moisture does not enter the cable entry:
Route the cable so that it loops down before the cable entry ("water trap").
6.Insert dummy plugs into unused cable entries.
7.5 Post-connection check
A0029278
Are cables or the device undamaged (visual inspection)?
Do the cables used meet the requirements→ 21?
Do the cables have adequate strain relief?
Are all the cable glands installed, firmly tightened and leak-tight? Cable run with "water trap"
→ 26?
Does the supply voltage match the specifications on the transmitter nameplate → 94?
If supply voltage is present, is the power LED on the electronics module of the transmitter lit green
→ 11?
Depending on the device version, is the securing clamp or fixing screw firmly tightened?
26Endress+Hauser
Proline Prosonic Flow E 100 HARTOperation options
1234
8 Operation options
8.1 Overview of operating options
A0019598
1Computer with Web browser (e.g. Internet Explorer) or with operating tool (e.g. FieldCare, AMS Device
Manager, SIMATIC PDM)
2Field Xpert SFX350 or SFX370
3Field Communicator 475
4Control system (e.g. PLC)
Endress+Hauser27
Operation optionsProline Prosonic Flow E 100 HART
!
Expert
System
Sensor
Communication
Application
Diagnostics
Access status display
Output
Operating menu for experts
Language
OperatationLanguage
Parameter 1
Setup
Submenu 1
Submenu n
Device tag
Advanced setup
Enter access code
Parameter 1
Parameter n
Submenu 1
Submenu n
Diagnostics
Parameter 1
Parameter n
Submenu 1
Submenu n
Operating menu for operators and maintenances
Parameter n
Operator
Maintenance
Task-oriented
Function-oriented
Expert
Wizard 1 / Parameter 1
Wizard n / Parameter n
Parameter n
Input
8.2 Structure and function of the operating menu
8.2.1 Structure of the operating menu
For an overview of the operating menu for experts: "Description of Device Parameters"
document supplied with the device
28Endress+Hauser
10Schematic structure of the operating menu
A0018237-EN
Proline Prosonic Flow E 100 HARTOperation options
8.2.2 Operating philosophy
The individual parts of the operating menu are assigned to certain user roles (operator,
maintenance etc.). Each user role contains typical tasks within the device lifecycle.
Contains all parameters for error detection and analyzing process and
device errors:
• Diagnostic list
Contains up to 5 currently pending diagnostic messages.
• Event logbook
Contains event messages that have occurred.
• Device information
Contains information for identifying the device.
• Measured values
Contains all current measured values.
• Heartbeat
The functionality of the device is checked on demand and the verification
results are documented.
• Simulation
Is used to simulate measured values or output values.
Expertfunction-oriented Tasks that require detailed
knowledge of the function of the
device:
• Commissioning measurements
under difficult conditions
• Optimal adaptation of the
measurement to difficult
conditions
• Detailed configuration of the
communication interface
• Error diagnostics in difficult cases
Contains all the parameters of the device and makes it possible to access
these parameters directly using an access code. The structure of this menu is
based on the function blocks of the device:
• System
Contains all higher-order device parameters which do not concern the
measurement or the communication interface.
• Sensor
Configuration of the measurement.
• Output
Configuring of the analog current outputs as well as the pulse/frequency
and switch output.
• Communication
Configuration of the digital communication interface and the Web server.
• Application
Configure the functions that go beyond the actual measurement (e.g.
totalizer).
• Diagnostics
Error detection and analysis of process and device errors and for device
simulation and Heartbeat Technology.
Endress+Hauser29
Operation optionsProline Prosonic Flow E 100 HART
8.3 Access to the operating menu via the web browser
8.3.1 Function range
Thanks to the integrated Web server, the device can be operated and configured via a Web
browser and via a service interface (CDI-RJ45) . In addition to the measured values, status
information on the device is also displayed and allows the user to monitor the status of the
device. Furthermore the device data can be managed and the network parameters can be
configured.
For additional information on the Web server, refer to the Special Documentation for
the device
8.3.2 Prerequisites
Computer hardware
InterfaceThe computer must have an RJ45 interface.
ConnectionStandard Ethernet cable with RJ45 connector.
ScreenRecommended size: ≥12" (depends on the screen resolution)
Computer software
Recommended operating
systems
Web browsers supported• Microsoft Internet Explorer 8 or higher
Microsoft Windows 7 or higher.
Microsoft Windows XP is supported.
• Microsoft Edge
• Mozilla Firefox
• Google Chrome
• Safari
Computer settings
User rightsAppropriate user rights (e.g. administrator rights) for TCP/IP and proxy server
settings are necessary (for adjusting the IP address, subnet mask etc.).
Proxy server settings of the
Web browser
JavaScriptJavaScript must be enabled.
Network connectionsOnly the active network connections to the measuring device should be used.
The Web browser setting Use a Proxy Server for Your LAN must be
deselected .
If JavaScript cannot be enabled:
enter http://XXX.XXX.X.XXX/basic.html in the address line of the Web
browser, e.g. http://192.168.1.212/basic.html. A fully functional but
simplified version of the operating menu structure starts in the Web
browser.
Switch off all other network connections such as WLAN.
In the event of connection problems: → 72
30Endress+Hauser
Loading...
+ 80 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.