A grounded terminal which, as far as the operator is concerned, is grounded via a
grounding system.
Protective Earth (PE)
A terminal which must be connected to ground prior to establishing any other
connections.
The ground terminals are situated inside and outside the device:
• Inner ground terminal: Connects the protectiv earth to the mains supply.
• Outer ground terminal: Connects the device to the plant grounding system.
Communication symbols
SymbolMeaning
Wireless Local Area Network (WLAN)
Communication via a wireless, local network.
LED
Light emitting diode is off.
LED
Light emitting diode is on.
LED
Light emitting diode is flashing.
Symbols for certain types of information
SymbolMeaning
Permitted
Procedures, processes or actions that are permitted.
Preferred
Procedures, processes or actions that are preferred.
Forbidden
Procedures, processes or actions that are forbidden.
Tip
Indicates additional information.
Reference to documentation.
Reference to page.
Reference to graphic.
Visual inspection.
4Endress+Hauser
Page 5
Proline Promass P 300
1.
-
.
2
1
3
Symbols in graphics
SymbolMeaning
1, 2, 3, ...Item numbers
, 2., 3., …Series of steps
A, B, C, ...Views
A-A, B-B, C-C, ...Sections
Hazardous area
Safe area (non-hazardous area)
Flow direction
Function and system design
Measuring principle
The measuring principle is based on the controlled generation of Coriolis forces. These forces are
always present in a system when both translational and rotational movements are superimposed.
Fc = 2 · ∆m (ν · ω)
Fc = Coriolis force
∆m = moving mass
ω = rotational velocity
ν = radial velocity in rotating or oscillating system
The amplitude of the Coriolis force depends on the moving mass ∆m, its velocity ν in the system and
thus on the mass flow. Instead of a constant rotational velocity ω, the sensor uses oscillation.
In the sensor, an oscillation is produced in the measuring tube. The Coriolis forces produced at the
measuring tube cause a phase shift in the tube oscillations (see illustration):
• If there is zero flow (i.e. when the fluid stands still), the oscillation measured at points A and B has
the same phase (no phase difference) (1).
• Mass flow causes deceleration of the oscillation at the inlet of the tubes (2) and acceleration at the
outlet (3).
A0029932
The phase difference (A-B) increases with increasing mass flow. Electrodynamic sensors register the
tube oscillations at the inlet and outlet. System balance is created by exciting an eccentrically
arranged swinging mass to antiphase oscillation. The measuring principle operates independently of
temperature, pressure, viscosity, conductivity and flow profile.
Endress+Hauser5
Page 6
Proline Promass P 300
Density measurement
The measuring tube is continuously excited at its resonance frequency. A change in the mass and
thus the density of the oscillating system (comprising measuring tube and fluid) results in a
corresponding, automatic adjustment in the oscillation frequency. Resonance frequency is thus a
function of medium density. The microprocessor utilizes this relationship to obtain a density signal.
Volume measurement
Together with the measured mass flow, this is used to calculate the volume flow.
Temperature measurement
The temperature of the measuring tube is determined in order to calculate the compensation factor
due to temperature effects. This signal corresponds to the process temperature and is also available
as an output signal.
Measuring system
The device consists of a transmitter and a sensor.
The device is available as a compact version:
The transmitter and sensor form a mechanical unit.
• Material of window in transmitter housing:
– Aluminum, coated: glass
– Stainless, hygienic: polycarbonate
Configuration:
A0026708
• External operation via 4-line, backlit, graphic local display with touch
control and guided menus ("Make-it-run" wizards) for applicationspecific commissioning.
• Via service interface or WLAN interface:
– Operating tools (e.g. FieldCare, DeviceCare, SmartBlue app)
– Web server (access via Web browser, e.g. Microsoft Internet
Explorer, Microsoft Edge)
Sensor
Promass P• For use in sterile applications in heavily regulated industries
– Process connections: stainless steel, 1.4435 BN2 (316L); 1.4404
(316/316L)
– Surface quality: Ra
Ra
0.38 µm (15 µin) (electropolished)
max
– Delta ferrite <1%
0.76 µm (30 µin)
max
6Endress+Hauser
Page 7
Proline Promass P 300
2
1
6
5
7
4
3
3
Equipment architecture
A0027512
1Possibilities for integrating measuring devices into a system
1Control system (e.g. PLC)
2Connecting cable (0/4 to 20 mA HART etc.)
3Fieldbus
4Segment coupler
5Non-hazardous area
6Hazardous area: Zone 2; Class I, Division 2
7Hazardous area: Zone 1; Class I, Division 1
SafetyIT security
Our warranty is valid only if the device is installed and used as described in the Operating
Instructions. The device is equipped with security mechanisms to protect it against any inadvertent
changes to the settings.
IT security measures, which provide additional protection for the device and associated data transfer,
must be implemented by the operators themselves in line with their security standards.
Device-specific IT security
The device offers a range of specific functions to support protective measures on the operator's side.
These functions can be configured by the user and guarantee greater in-operation safety if used
correctly. An overview of the most important functions is provided in the following section.
Function/interfaceFactory settingRecommendation
Write protection via hardware write
protection switch → 8
Access code
(also applies for Web server login or
FieldCare connection) → 8
WLAN
(order option in display module)
WLAN security modeEnabled (WPA2-
WLAN passphrase
(password) → 8
WLAN modeAccess PointOn an individual basis following risk
Web server→ 8Enabled.On an individual basis following risk
CDI-RJ45 service interface → 9–On an individual basis following risk
Not enabled.On an individual basis following risk
assessment.
Not enabled
(0000).
Enabled.On an individual basis following risk
PSK)
Serial numberAssign a customized access code during
Assign a customized access code during
commissioning.
assessment.
Do not change.
commissioning.
assessment.
assessment.
assessment.
Endress+Hauser7
Page 8
Proline Promass P 300
Protecting access via hardware write protection
Write access to the device parameters via the local display, Web browser or operating tool (e.g.
FieldCare, DeviceCare) can be disabled via a write protection switch (DIP switch on the
motherboard). When hardware write protection is enabled, only read access to the parameters is
possible.
Hardware write protection is disabled when the device is delivered.
Protecting access via a password
Different passwords are available to protect write access to the device parameters or access to the
device via the WLAN interface.
• User-specific access code
Protect write access to the device parameters via the local display, Web browser or operating tool
(e.g. FieldCare, DeviceCare). Access authorization is clearly regulated through the use of a userspecific access code.
• WLAN passphrase
The network key protects a connection between an operating unit (e.g. notebook or tablet) and the
device via the WLAN interface which can be ordered as an option.
• Infrastructure mode
When the device is operated in infrastructure mode, the WLAN passphrase corresponds to the
WLAN passphrase configured on the operator side.
User-specific access code
Write access to the device parameters via the local display, Web browser or operating tool (e.g.
FieldCare, DeviceCare) can be protected by the modifiable, user-specific access code.
WLAN passphrase: Operation as WLAN access point
A connection between an operating unit (e.g. notebook or tablet) and the device via the WLAN
interface, which can be ordered as an optional extra, is protected by the network key. The WLAN
authentication of the network key complies with the IEEE 802.11 standard.
When the device is delivered, the network key is pre-defined depending on the device. It can be
changed via the WLAN settings submenu in the WLAN passphrase parameter.
Infrastructure mode
A connection between the device and WLAN access point is protected by means of an SSID and
passphrase on the system side. Please contact the relevant system administrator for access.
General notes on the use of passwords
• The access code and network key supplied with the device should be changed during
commissioning.
• Follow the general rules for generating a secure password when defining and managing the access
code or network key.
• The user is responsible for the management and careful handling of the access code and network
key.
Access via Web server
The device can be operated and configured via a Web browser with the integrated Web server. The
connection is via the service interface (CDI-RJ45) or the WLAN interface. For device versions with
the EtherNet/IP and PROFINET communication protocols, the connection can also be established via
the terminal connection for signal transmission with EtherNet/IP or PROFINET (RJ45 connector).
The Web server is enabled when the device is delivered. The Web server can be disabled if necessary
(e.g. after commissioning) via the Web server functionality parameter.
The device and status information can be hidden on the login page. This prevents unauthorized
access to the information.
For detailed information on device parameters, see:
The "Description of Device Parameters" document → 110
Access via OPC-UA
The "OPC UA Server" application package is available in the device version with the HART
communication protocol → 106.
8Endress+Hauser
Page 9
Proline Promass P 300
The device can communicate with OPC UA clients using the "OPC UA Server" application package.
The OPC UA server integrated in the device can be accessed via the WLAN access point using the
WLAN interface - which can be ordered as an optional extra - or the service interface (CDI- RJ45) via
Ethernet network. Access rights and authorization as per separate configuration.
The following Security Modes are supported as per the OPC UA Specification (IEC 62541):
• None
• Basic128Rsa15 – signed
• Basic128Rsa15 – signed and encrypted
Access via service interface (CDI-RJ45)
The device can be connected to a network via the service interface (CDI-RJ45). Device-specific
functions guarantee the secure operation of the device in a network.
The use of relevant industrial standards and guidelines that have been defined by national and
international safety committees, such as IEC/ISA62443 or the IEEE, is recommended. This includes
organizational security measures such as the assignment of access authorization as well as technical
measures such as network segmentation.
The device can be integrated in a ring topology. The device is integrated via the terminal
connection for signal transmission (output 1) and the connection to the service interface (CDIRJ45) → 95.
Endress+Hauser9
Page 10
Input
Measured variableDirect measured variables
• Mass flow
• Density
• Temperature
Calculated measured variables
• Volume flow
• Corrected volume flow
• Reference density
Measuring rangeMeasuring range for liquids
Proline Promass P 300
DNMeasuring range full scale values
[mm][in][kg/h][lb/min]
8³⁄₈0 to 20000 to 73.50
15½0 to 65000 to 238.9
2510 to 180000 to 661.5
401½0 to 450000 to 1654
5020 to 700000 to 2573
Recommended measuring range
"Flow limit" section → 55
Operable flow range
Over 1000 : 1.
Flow rates above the preset full scale value do not override the electronics unit, with the result that
the totalizer values are registered correctly.
Input signalInput and output versions
→ 12
External measured values
To increase the accuracy of certain measured variables, the automation system can continuously
write various measured values to the measuring device:
• Operating pressure to increase accuracy (Endress+Hauser recommends the use of a pressure
measuring device for absolute pressure, e.g. Cerabar M or Cerabar S)
• Medium temperature to increase accuracy (e.g. iTEMP)
Various pressure transmitters and temperature measuring devices can be ordered from Endress
+Hauser: see "Accessories" section → 109
min(F)
to
max(F)
HART protocol
The measured values are written from the automation system to the measuring device via the HART
protocol. The pressure transmitter must support the following protocol-specific functions:
• HART protocol
• Burst mode
Current input
The measured values are written from the automation system to the measuring device via the
current input → 11.
10Endress+Hauser
Page 11
Proline Promass P 300
Digital communication
The measured values can be written from the automation system to the measuring via:
• FOUNDATION Fieldbus
• PROFIBUS DP
• PROFIBUS PA
• Modbus RS485
• EtherNet/IP
• PROFINET
Current input 0/4 to 20 mA
Current input0/4 to 20 mA (active/passive)
Current span• 4 to 20 mA (active)
• 0/4 to 20 mA (passive)
Resolution1 µA
Voltage dropTypically: 0.6 to 2 V for 3.6 to 22 mA (passive)
Maximum input voltage≤ 30 V (passive)
Open-circuit voltage≤ 28.8 V (active)
Possible input variables• Pressure
• Temperature
• Density
Status input
Maximum input values• DC –3 to 30 V
• If status input is active (ON): Ri >3 kΩ
Response timeAdjustable: 5 to 200 ms
Input signal level• Low signal: DC –3 to +5 V
• High signal: DC 12 to 30 V
Assignable functions• Off
• Reset the individual totalizers separately
• Reset all totalizers
• Flow override
Endress+Hauser11
Page 12
Output
Proline Promass P 300
Output and input variants
Depending on the option selected for output/input 1, different options are available for the other
outputs and inputs. Only one option can be selected for each output/input 1 to 3. The table must be
read vertically (↓).
Example: If the option BA "4–20 mA HART" was selected for output/input 1, one of the options A, B,
D, E, F, H, I or J is available for output 2 and one of the options A, B, D, E, F, H, I or J is available for
output 3.
Order code for "Output; input 1" (020) →Possible options
Current output 4 to 20 mA HARTBA
Current output 4 to 20 mA HART Ex i↓CA
FOUNDATION Fieldbus↓SA
FOUNDATION Fieldbus Ex i↓TA
PROFIBUS DP↓LA
PROFIBUS PA↓GA
PROFIBUS PA Ex i↓HA
Modbus RS485↓MA
EtherNet/IP 2-port switch integrated↓NA
PROFINET 2-port switch integrated↓RA
Order code for "Output; input 2" (021) →↓↓↓↓↓↓↓↓↓↓
Not assignedAAAAAAAAAA
Current output 0/4 to 20 mABBBBBBB
Current output 0/4 to 20 mA (Ex i)CCC
User configurable input/output
Pulse/frequency/switch outputEEEEEEE
Double pulse output
Pulse/frequency/switch output (Ex i)GGG
Relay outputHHHHHHH
Current input 0/4 to 20 mAIIIIIII
Status inputJJJJJJJ
Order code for "Output; input 3" (022) →↓↓↓↓↓↓↓↓↓↓
Not assignedAAAAAAAAAA
Current output 0/4 to 20 mABBBBB
Current output 0/4 to 20 mA (Ex i)C
User configurable input/outputDDDDD
Pulse/frequency/switch outputEEEEE
Double pulse output (slave)
Pulse/frequency/switch output (Ex i)G
Relay outputHHHHH
Current input 0/4 to 20 mAIIIII
Status inputJJJJJ
2)
1)
2)
DDDDDDD
FF
FF
1)A specific input or output can be assigned to a user configurable input/output → 16.
2)If double pulse output (F) is selected for output/input 2 (021), only the double pulse output (F) option is available for selection for output/input 3
(022).
12Endress+Hauser
Page 13
Proline Promass P 300
Output signalHART current output
Current output4 to 20 mA HART
Current spanCan be set to: 4 to 20 mA (active/passive)
Open-circuit voltageDC 28.8 V (active)
Maximum input voltageDC 30 V (passive)
Load250 to 700 Ω
Resolution0.38 µA
DampingConfigurable: 0.07 to 999 s
Assignable measured
variables
Ex-i, passive
• Mass flow
• Volume flow
• Corrected volume flow
• Density
• Reference density
• Temperature
• Electronic temperature
• Oscillation frequency 0
• Oscillation damping 0
• Signal asymmetry
• Exciter current 0
The range of options increases if the measuring device has one or more
application packages.
PROFIBUS PA
PROFIBUS PAIn accordance with EN 50170 Volume 2, IEC 61158-2 (MBP), galvanically
isolated
Data transmission31.25 kbit/s
Current consumption10 mA
Permitted supply voltage9 to 32 V
Bus connectionWith integrated reverse polarity protection
PROFIBUS DP
Signal encodingNRZ code
Data transfer9.6 kBaud…12 MBaud
EtherNet/IP
StandardsIn accordance with IEEE 802.3
PROFINET
StandardsIn accordance with IEEE 802.3
FOUNDATION Fieldbus
FOUNDATION FieldbusH1, IEC 61158-2, galvanically isolated
Data transfer31.25 kbit/s
Endress+Hauser13
Page 14
Current consumption10 mA
Permitted supply voltage9 to 32 V
Bus connectionWith integrated reverse polarity protection
Modbus RS485
Physical interfaceRS485 in accordance with EIA/TIA-485 standard
Terminating resistorIntegrated, can be activated via DIP switches
Current output 0/4 to 20 mA
Current output0/4 to 20 mA
Maximum output values22.5 mA
Current spanCan be set to:
• 4 to 20 mA (active)
• 0/4 to 20 mA (passive)
Ex-i, passive
Proline Promass P 300
Open-circuit voltageDC 28.8 V (active)
Maximum input voltageDC 30 V (passive)
Load0 to 700 Ω
Resolution0.38 µA
DampingAdjustable: 0.07 to 999 s
Assignable measured
variables
• Mass flow
• Volume flow
• Corrected volume flow
• Density
• Reference density
• Temperature
• Electronic temperature
• Oscillation frequency 0
• Oscillation damping 0
• Signal asymmetry
• Exciter current 0
The range of options increases if the measuring device has one or more
application packages.
Pulse/frequency/switch output
FunctionCan be set to pulse, frequency or switch output
VersionOpen collector
Can be set to:
• Active
• Passive
Ex-i, passive
Maximum input valuesDC 30 V, 250 mA (passive)
Open-circuit voltageDC 28.8 V (active)
Voltage dropFor 22.5 mA: ≤ DC 2 V
Pulse output
Maximum input valuesDC 30 V, 250 mA (passive)
Maximum output current22.5 mA (active)
14Endress+Hauser
Page 15
Proline Promass P 300
Open-circuit voltageDC 28.8 V (active)
Pulse widthAdjustable: 0.05 to 2000 ms
Maximum pulse rate10000 Impulse/s
Pulse valueAdjustable
Assignable measured
variables
Frequency output
Maximum input valuesDC 30 V, 250 mA (passive)
Maximum output current22.5 mA (active)
Open-circuit voltageDC 28.8 V (active)
Output frequencyAdjustable: end value frequency 2 to 10000 Hz (f
DampingAdjustable: 0 to 999 s
Pulse/pause ratio1:1
Assignable measured
variables
Switch output
Maximum input valuesDC 30 V, 250 mA (passive)
Open-circuit voltageDC 28.8 V (active)
Switching behaviorBinary, conductive or non-conductive
Switching delayAdjustable: 0 to 100 s
Number of switching
cycles
Assignable functions• Off
• Mass flow
• Volume flow
• Corrected volume flow
• Mass flow
• Volume flow
• Corrected volume flow
• Density
• Reference density
• Temperature
• Electronic temperature
• Oscillation frequency 0
• Oscillation damping 0
• Signal asymmetry
• Exciter current 0
The range of options increases if the measuring device has one or more
application packages.
Unlimited
• On
• Diagnostic behavior
• Limit value
– Mass flow
– Volume flow
– Corrected volume flow
– Density
– Reference density
– Temperature
– Totalizer 1-3
• Flow direction monitoring
• Status
– Partially filled pipe detection
– Low flow
The range of options increases if the measuring device has one or more
application packages.
= 12500 Hz)
max
Endress+Hauser15
Page 16
Double pulse output
FunctionDouble pulse
VersionOpen collector
Can be set to:
• Active
• Passive
Maximum input valuesDC 30 V, 250 mA (passive)
Open-circuit voltageDC 28.8 V (active)
Voltage dropFor 22.5 mA: ≤ DC 2 V
Output frequencyAdjustable: 0 to 1000 Hz
DampingAdjustable: 0 to 999 s
Pulse/pause ratio1:1
Assignable measured
variables
• Mass flow
• Volume flow
• Corrected volume flow
• Density
• Reference density
• Temperature
The range of options increases if the measuring device has one or more
application packages.
Proline Promass P 300
Relay output
FunctionSwitch output
VersionRelay output, galvanically isolated
Switching behaviorCan be set to:
• NO (normally open), factory setting
• NC (normally closed)
Maximum switching
capacity (passive)
Assignable functions• Off
• DC 30 V, 0.1 A
• AC 30 V, 0.5 A
• On
• Diagnostic behavior
• Limit value
– Mass flow
– Volume flow
– Corrected volume flow
– Density
– Reference density
– Temperature
– Totalizer 1-3
• Flow direction monitoring
• Status
– Partially filled pipe detection
– Low flow
The range of options increases if the measuring device has one or more
application packages.
User configurable input/output
One specific input or output is assigned to a user-configurable input/output (configurable I/O)
during device commissioning.
The following inputs and outputs are available for assignment:
• Choice of current output: 4 to 20 mA (active), 0/4 to 20 mA (passive)
• Pulse/frequency/switch output
• Choice of current input: 4 to 20 mA (active), 0/4 to 20 mA (passive)
• Status input
16Endress+Hauser
Page 17
Proline Promass P 300
The technical values correspond to those of the inputs and outputs described in this section.
Signal on alarm
Depending on the interface, failure information is displayed as follows:
HART current output
Device diagnosticsDevice condition can be read out via HART Command 48
PROFIBUS PA
Status and alarm
messages
Failure current FDE (Fault
Disconnection Electronic)
Diagnostics in accordance with PROFIBUS PA Profile 3.02
0 mA
PROFIBUS DP
Status and alarm
messages
Diagnostics in accordance with PROFIBUS PA Profile 3.02
EtherNet/IP
Device diagnosticsDevice condition can be read out in Input Assembly
PROFINET
Device diagnosticsAccording to "Application Layer protocol for decentralized periphery", Version 2.3
FOUNDATION Fieldbus
Status and alarm
messages
Failure current FDE (Fault
Disconnection Electronic)
Diagnostics in accordance with FF-891
0 mA
Modbus RS485
Failure modeChoose from:
• NaN value instead of current value
• Last valid value
Current output 0/4 to 20 mA
4 to 20 mA
Failure modeChoose from:
• 4 to 20 mA in accordance with NAMUR recommendation NE 43
• 4 to 20 mA in accordance with US
• Min. value: 3.59 mA
• Max. value: 22.5 mA
• Freely definable value between: 3.59 to 22.5 mA
• Actual value
• Last valid value
Endress+Hauser17
Page 18
0 to 20 mA
Failure modeChoose from:
• Maximum alarm: 22 mA
• Freely definable value between: 0 to 20.5 mA
Pulse/frequency/switch output
Pulse output
Failure modeChoose from:
• Actual value
• No pulses
Frequency output
Failure modeChoose from:
• Actual value
• 0 Hz
• Defined value (f
Switch output
Failure modeChoose from:
• Current status
• Open
• Closed
2 to 12500 Hz)
max
Proline Promass P 300
Relay output
Failure modeChoose from:
• Current status
• Open
• Closed
Local display
Plain text displayWith information on cause and remedial measures
BacklightRed backlighting indicates a device error.
Status signal as per NAMUR recommendation NE 107
Interface/protocol
• Via digital communication:
– HART protocol
– FOUNDATION Fieldbus
– PROFIBUS PA
– PROFIBUS DP
– Modbus RS485
– EtherNet/IP
– PROFINET
• Via service interface
– CDI-RJ45 service interface
– WLAN interface
Plain text displayWith information on cause and remedial measures
Additional information on remote operation → 91
18Endress+Hauser
Page 19
Proline Promass P 300
Web server
Plain text displayWith information on cause and remedial measures
Light emitting diodes (LED)
Status informationStatus indicated by various light emitting diodes
Ex connection dataSafety-related values
The following information is displayed depending on the device version:
• Supply voltage active
• Data transmission active
• Device alarm/error has occurred
• EtherNet/IP network available
• EtherNet/IP connection established
• PROFINET network available
• PROFINET connection established
• PROFINET blinking feature
Order code for
Output typeSafety-related values
"Output; input 1"
Option BACurrent output
4 to 20 mA HART
UN = 30 V
UM = 250 V
Option GAPROFIBUS PAUN = 30 V
UM = 250 V
Option LAPROFIBUS DPUN = 30 V
UM = 250 V
Option MAModbus RS485UN = 30 V
UM = 250 V
Option SAFOUNDATION FieldbusUN = 30 V
UM = 250 V
Option NAEtherNet/IPUN = 30 V
UM = 250 V
Option RAPROFINETUN = 30 V
UM = 250 V
Order code for
"Output; input 2";
"Output; input 3"
Output typeSafety-related values
Output; input 2Output; input 3
24 (+)25 (–)22 (+)23 (–)
Option BCurrent output
4 to 20 mA
Option DUser configurable input/
output
Option EPulse/frequency/switch
output
UN = 30 V
UM = 250 V
UN = 30 V
UM = 250 V
UN = 30 V
UM = 250 V
Option FDouble pulse outputUN = 30 V
UM = 250 V
Option HRelay outputUN = 30 V
IN =100 mADC/500 mA
UM = 250 V
Option ICurrent input 4 to 20 mA UN = 30 V
UM = 250 V
Option JStatus inputUN = 30 V
UM = 250 V
"Output; input 1"
26 (+)27 (–)
DC
AC
DC
AC
DC
AC
DC
AC
DC
AC
DC
AC
DC
AC
DC
AC
DC
AC
DC
AC
DC
AC
DC
AC
AC
DC
AC
DC
AC
Endress+Hauser19
Page 20
Intrinsically safe values
Proline Promass P 300
Order code for
"Output; input 1"
Option CACurrent output
Option HAPROFIBUS PA Ex iEx ia
Option TAFOUNDATION Fieldbus
1)Only available for the Zone 1; Class I, Division 1 version
2)Only available for the Zone 2; Class I, Division 2 version transmitter
Order code for
"Output; input 2";
"Output; input 3"
Option CCurrent output
Option GPulse/frequency/switch
Output typeIntrinsically safe values
26 (+)27 (–)
Ui = 30 V
4 to 20 mA HART Ex i
Ex i
Output typeIntrinsically safe values or NIFW values
4 to 20 mA Ex i
output Ex i
li = 100 mA
Pi = 1.25 W
Li = 0
Ci = 0
1)
Ui = 30 V
li = 570 mA
Pi = 8.5 W
Li = 10 µH
Ci = 5 nF
1)
Ex ia
Ui = 30 V
li = 570 mA
Pi = 8.5 W
Li = 10 µH
Ci = 5 nF
Output; input 2Output; input 3
24 (+)25 (–)22 (+)23 (–)
Ui = 30 V
li = 100 mA
Pi = 1.25 W
Li = 0
Ci = 0
Ui = 30 V
li = 100 mA
Pi = 1.25 W
Li = 0
Ci = 0
"Output; input 1"
2)
Ex ic
Ui = 32 V
li = 570 mA
Pi = 8.5 W
Li = 10 µH
Ci = 5 nF
2)
Ex ic
Ui = 32 V
li = 570 mA
Pi = 8.5 W
Li = 10 µH
Ci = 5 nF
Low flow cut off
Galvanic isolation
The switch points for low flow cut off are user-selectable.
The outputs are galvanically isolated from one another and from earth (PE).
Protocol-specific dataHART
Manufacturer ID0x11
Device type ID0x3B
HART protocol revision7
Device description files
(DTM, DD)
HART loadMin. 250 Ω
System integrationInformation on system integration: Operating Instructions → 110.
Information and files under:
www.endress.com
• Measured variables via HART protocol
• Burst Mode functionality
20Endress+Hauser
Page 21
Proline Promass P 300
PROFIBUS PA
Manufacturer ID0x11
Ident number0x156D
Profile version3.02
Device description files (GSD,
DTM, DD)
Supported functions• Identification & Maintenance
Configuration of the device
address
Compatibility with
earlier model
System integrationInformation regarding system integration: Operating Instructions → 110.
Information and files under:
• www.endress.com
• www.profibus.org
Simplest device identification on the part of the control system and
nameplate
• PROFIBUS upload/download
Reading and writing parameters is up to ten times faster with PROFIBUS
upload/download
• Condensed status
Simplest and self-explanatory diagnostic information by categorizing
diagnostic messages that occur
• DIP switches on the I/O electronics module
• Local display
• Via operating tools (e.g. FieldCare)
If the device is replaced, the measuring device Promass 300 supports the
compatibility of the cyclic data with previous models. It is not necessary to
adjust the engineering parameters of the PROFIBUS network with the
Promass 300 GSD file.
Earlier models:
• Promass 80 PROFIBUS PA
– ID No.: 1528 (hex)
– Extended GSD file: EH3x1528.gsd
– Standard GSD file: EH3_1528.gsd
• Promass 83 PROFIBUS PA
– ID No.: 152A (hex)
– Extended GSD file: EH3x152A.gsd
– Standard GSD file: EH3_152A.gsd
Description of the function scope of compatibility:
Operating Instructions → 110.
• Cyclic data transmission
• Block model
• Description of the modules
PROFIBUS DP
Manufacturer ID0x11
Ident number0x156F
Profile version3.02
Device description files (GSD,
DTM, DD)
Supported functions• Identification & Maintenance
Configuration of the device
address
Information and files under:
• www.endress.com
On the product page for the device: Documents/Software → Device drivers
• www.profibus.org
Simplest device identification on the part of the control system and
nameplate
• PROFIBUS upload/download
Reading and writing parameters is up to ten times faster with PROFIBUS
upload/download
• Condensed status
Simplest and self-explanatory diagnostic information by categorizing
diagnostic messages that occur
• DIP switches on the I/O electronics module
• Via operating tools (e.g. FieldCare)
Endress+Hauser21
Page 22
Proline Promass P 300
Compatibility with
earlier model
System integrationInformation regarding system integration: Operating Instructions → 110.
If the device is replaced, the measuring device Promass 300 supports the
compatibility of the cyclic data with previous models. It is not necessary to
adjust the engineering parameters of the PROFIBUS network with the
Promass 300 GSD file.
Response times• Direct data access: typically 25 to 50 ms
• Auto-scan buffer (data range): typically 3 to 5 ms
Device typeSlave
Slave address range1 to 247
Broadcast address range0
Function codes• 03: Read holding register
• 04: Read input register
• 06: Write single registers
• 08: Diagnostics
• 16: Write multiple registers
• 23: Read/write multiple registers
Broadcast messagesSupported by the following function codes:
• 06: Write single registers
• 16: Write multiple registers
• 23: Read/write multiple registers
24Endress+Hauser
Page 25
Proline Promass P 300
Supported baud rate• 1200 BAUD
• 2400 BAUD
• 4800 BAUD
• 9600 BAUD
• 19200 BAUD
• 38400 BAUD
• 57600 BAUD
• 115200 BAUD
Data transfer mode• ASCII
• RTU
Data accessEach device parameter can be accessed via Modbus RS485.
For Modbus register information
Compatibility with
earlier model
System integrationInformation on system integration: Operating Instructions → 110.
If the device is replaced, the measuring device Promass 300 supports the
compatibility of the Modbus registers for the process variables and the
diagnostic information with the previous model Promass 83. It is not
necessary to change the engineering parameters in the automation system.
Description of the function scope of compatibility:
Terminal assignment of the remote display and operating module → 31.
Device plugs may not be used in hazardous areas!
Device plugs for fieldbus systems:
Order code for "Input; output 1"
• Option SA "FOUNDATION Fieldbus" → 26
• Option GA "PROFIBUS PA" → 26
• Option RA "PROFINET" → 27
• Option NA "EtherNet/IP" → 27
Device plug for connecting to the service interface:
Order code for "Accessory mounted"
option NB, adapter RJ45 M12 (service interface) → 28
Order code for "Input; output 1", option SA "FOUNDATION Fieldbus"
Order code forCable entry/connection → 30
"Electrical connection"23
M, 3, 4, 57/8" connector–
Order code for "Input; output 1", option GA "PROFIBUS PA"
Order code forCable entry/connection → 30
"Electrical connection"23
L, N, P, UConnector M12 × 1–
26Endress+Hauser
Page 27
Proline Promass P 300
1
2
4
3
1
2
4
3
Order code for "Input; output 1", option RA "PROFINET"
Order code forCable entry/connection → 30
"Electrical connection"23
L, N, P, UConnector M12 × 1–
1) 2)
1) 2)
1) 2)
R
, S
, T
, V
1) 2)
Connector M12 × 1Connector M12 × 1
1)Cannot be combined with an external WLAN antenna (order code for "Enclosed accessories", option P8) of
an RJ45 M12 adapter for the service interface (order code for "Accessories mounted", option NB) or of the
remote display and operating module DKX001.
2)Suitable for integrating the device in a ring topology.
Order code for "Input; output 1", option NA "EtherNet/IP"
Order code forCable entry/connection → 30
"Electrical connection"23
L, N, P, UConnector M12 × 1–
1) 2)
1) 2)
1) 2)
R
, S
, T
, V
1) 2)
Connector M12 × 1Connector M12 × 1
1)Cannot be combined with an external WLAN antenna (order code for "Enclosed accessories", option P8) of
an RJ45 M12 adapter for the service interface (order code for "Accessories mounted", option NB) or of the
remote display and operating module DKX001
2)Suitable for integrating the device in a ring topology.
Order code for "Accessory mounted", option NB "Adapter RJ45 M12 (service interface)"
Order codeCable entry/coupling → 30
"Accessory mounted"Cable entry
NBPlug M12 × 1–
Pin assignment, device plugFOUNDATION Fieldbus
PROFIBUS PA
Cable entry
2
3
PinAssignmentCodingPlug/socket
1+Signal +APlug
2-Signal –
3Grounding
4Not assigned
PinAssignmentCodingPlug/socket
1+PROFIBUS PA +APlug
2Grounding
3-PROFIBUS PA –
4Not assigned
Endress+Hauser27
Page 28
PROFINET
3
2
4
3
2
4
3
2
4
PinAssignment
1+TD +
2+RD +
3-TD –
4-RD –
A0032047
Recommended plug:
• Binder, series 763, part no. 99 3729 810 04
• Phoenix, part no. 1543223 SACC-M12MSD-4Q
• When using the device in a hazardous location, use a suitably certified plug.
CodingPlug/socket
DSocket
EtherNet/IP
PinAssignment
1+Tx
2+Rx
3-Tx
4-Rx
A0032047
CodingPlug/socket
DSocket
Proline Promass P 300
Supply voltage
Recommended plug:
• Binder, series 763, part no. 99 3729 810 04
• Phoenix, part no. 1543223 SACC-M12MSD-4Q
• When using the device in a hazardous location, use a suitably certified plug.
Service interface
Order code for "Accessories mounted", option NB: Adapter RJ45 M12 (service interface)
PinAssignment
1+Tx
2+Rx
3-Tx
4-Rx
A0032047
Recommended plug:
• Binder, series 763, part no. 99 3729 810 04
• Phoenix, part no. 1543223 SACC-M12MSD-4Q
• When using the device in a hazardous location, use a suitably certified plug.
Order code for
"Power supply"
Option DDC24 V±20%–
Option EAC100 to 240 V –15…+10%50/60 Hz
Option I
CodingPlug/socket
DSocket
terminal voltageFrequency range
DC24 V±20%–
AC100 to 240 V –15…+10%50/60 Hz
28Endress+Hauser
Page 29
Proline Promass P 300
Power consumptionTransmitter
Max. 10 W (active power)
switch-on currentMax. 36 A (as per NAMUR Recommendation NE21)
Current consumptionTransmitter
• Max. 400 mA (24 V)
• Max. 200 mA (110 V, 50/60 Hz; 230 V, 50/60 Hz)
Power supply failure
• Totalizers stop at the last value measured.
• Depending on the device version, the configuration is retained in the device memoryor in the
pluggable data memory (HistoROM DAT).
• Error messages (incl. total operated hours) are stored.
Endress+Hauser29
Page 30
Electrical connectionConnecting the transmitter
1
2
3
4
1
2
3
4
• Terminal assignment→ 25
• Device plugs available→ 26
1Terminal connection for supply voltage
2Terminal connection for signal transmission, input/output
3Terminal connection for signal transmission, input/output or terminal for network connection via service
interface (CDI-RJ45); Optional: terminal connection for external WLAN antenna or connection for remote
display and operating module DKX001
4Protective ground (PE)
An adapter for RJ45 and the M12 connector is optionally available:
Order code for "Accessories", option NB: "Adapter RJ45 M12 (service interface)"
The adapter connects the service interface (CDI-RJ45) to an M12 connector mounted in the
cable entry. Therefore the connection to the service interface can be established via an M12
connector without opening the device.
Network connection via service interface (CDI-RJ45) → 97
Proline Promass P 300
A0026781
Connecting in a ring topology
Device versions with EtherNet/IP and PROFINET communication protocols can be integrated into a
ring topology. The device is integrated via the terminal connection for signal transmission (output 1)
and the connection to the service interface (CDI-RJ45).
Integrate the transmitter into a ring topology:
• EtherNet/IP → 95
• PROFINET → 96
A0026781
1Terminal connection for supply voltage
2Terminal connection for signal transmission: PROFINET or EtherNet/IP (RJ45 connector)
3Connection to service interface (CDI-RJ45)
4Protective earth (PE)
If the device has additional input/outputs, these are routed via the cable entry for the
connection to the service interface (CDI-RJ45).
30Endress+Hauser
Page 31
Proline Promass P 300
1
3
81
Vcc
82
Gnd
83A84
B
81
Vcc
82
Gnd
83A84
B
2
4
5
4
4...20 mA
5
2
1
3
6
Connecting the remote display and operating module DKX001
The remote display and operating module DKX001 is available as an optional extra → 107.
• The remote display and operating module DKX001 is only available for the following housing
version: order code for "Housing": option A "Aluminum, coated"
• The measuring device is always supplied with a dummy cover when the remote display and
operating module DKX001 is ordered directly with the measuring device. Display or operation
at the transmitter is not possible in this case.
• If ordered subsequently, the remote display and operating module DKX001 may not be
connected at the same time as the existing measuring device display module. Only one
display or operation unit may be connected to the transmitter at any one time.
2Connection example for 4 to 20 mA HART current output (active)
1Automation system with current input (e.g. PLC)
2Cable shield: the cable shield must be grounded at both ends to comply with EMC requirements; observe cable
specifications → 39
3Connection for HART operating devices → 91
4Resistor for HART communication (≥ 250 Ω): observe maximum load → 13
5Analog display unit: observe maximum load → 13
6Transmitter
Endress+Hauser31
Page 32
Proline Promass P 300
2
3
4...20 mA
41
5
2
4...20 mA
4
1
2
3
3
6
5
A0028762
3Connection example for 4 to 20 mA HART current output (passive)
1Automation system with current input (e.g. PLC)
2Power supply
3Cable shield: the cable shield must be grounded at both ends to comply with EMC requirements; observe cable
4Connection example for HART input with a common negative (passive)
1Automation system with HART output (e.g. PLC)
2Active barrier for power supply (e.g. RN221N)
3Cable shield: the cable shield must be grounded at both ends to comply with EMC requirements; observe cable
specifications
4Analog display unit: observe maximum load
5Pressure transmitter (e.g. Cerabar M, Cerabar S): see requirements
6Transmitter
32Endress+Hauser
Page 33
Proline Promass P 300
2134
78
66
6
6
5
6
6
5
2
1
A
B
3
4
4
A
B
A
B
PROFIBUS PA
A0028768
5Connection example for PROFIBUS PA
1Control system (e.g. PLC)
2PROFIBUS PA segment coupler
3Cable shield: the cable shield must be grounded at both ends to comply with EMC requirements; observe cable
1Control system (e.g. PLC)
2Ethernet switch
3Observe cable specifications → 39
4Connecting cable between the two transmitters
5Transmitter
FOUNDATION Fieldbus
9Connection example for FOUNDATION Fieldbus
1Control system (e.g. PLC)
2Power Conditioner (FOUNDATION Fieldbus)
3Cable shield: the cable shield must be grounded at both ends to comply with EMC requirements; observe cable
10Connection example for Modbus RS485, non-hazardous area and Zone 2; Class I, Division 2
1Control system (e.g. PLC)
2Cable shield: the cable shield must be grounded at both ends to comply with EMC requirements; observe cable
specifications
3Distribution box
4Transmitter
Current output 4-20 mA
A0028758
11Connection example for 4-20 mA current output (active)
1Automation system with current input (e.g. PLC)
2Analog display unit: observe maximum load
3Transmitter
A0028759
12Connection example for 4-20 mA current output (passive)
1Automation system with current input (e.g. PLC)
2Active barrier for power supply (e.g. RN221N)
3Analog display unit: observe maximum load
4Transmitter
36Endress+Hauser
Page 37
Proline Promass P 300
1
2
3
12345
1
2
3
1
2
3
4
Pulse/frequency output
A0028761
13Connection example for pulse/frequency output (passive)
1Automation system with pulse/frequency input (e.g. PLC)
2Power supply
3Transmitter: Observe input values → 14
Switch output
14Connection example for switch output (passive)
1Automation system with switch input (e.g. PLC)
2Power supply
3Transmitter: Observe input values → 14
Double pulse output
15Connection example for double pulse output (active)
1Automation system with relay input (e.g. PLC)
2Power supply
3Transmitter: Observe input values → 16
Current input
A0028760
A0028915
18Connection example for 4 to 20 mA current input
1Power supply
2Terminal box
3External measuring device (for reading in pressure or temperature, for instance)
4Transmitter
38Endress+Hauser
Page 39
Proline Promass P 300
1
2
3
Status input
19Connection example for status input
1Automation system with status output (e.g. PLC)
2Power supply
3Transmitter
Potential equalizationRequirements
No special measures for potential equalization are required.
Please consider the following to ensure correct measurement:
• Same electrical potential for the fluid and sensor
• Company-internal grounding concepts
A0028764
terminals
Spring-loaded terminals: Suitable for strands and strands with ferrules.
Conductor cross-section 0.2 to 2.5 mm2 (24 to 12 AWG).
Cable entries
• Cable gland: M20 × 1.5 with cable ⌀ 6 to 12 mm (0.24 to 0.47 in)
• Thread for cable entry:
– NPT ½"
– G ½"
– M20
• Device plug for digital communication: M12
Only available for certain device versions → 26.
Cable specificationPermitted temperature range
• The installation guidelines that apply in the country of installation must be observed.
• The cables must be suitable for the minimum and maximum temperatures to be expected.
Power supply cable
Standard installation cable is sufficient.
Signal cable
Current output 4 to 20 mA HART
A shielded cable is recommended. Observe grounding concept of the plant.
PROFIBUS PA
Twisted, shielded two-wire cable. Cable type A is recommended .
For further information on planning and installing PROFIBUS networks see:
• Operating Instructions "PROFIBUS DP/PA: Guidelines for planning and commissioning"
(BA00034S)
• PNO Directive 2.092 "PROFIBUS PA User and Installation Guideline"
• IEC 61158-2 (MBP)
PROFIBUS DP
The IEC 61158 standard specifies two types of cable (A and B) for the bus line which can be used for
every transmission rate. Cable type A is recommended.
Endress+Hauser39
Page 40
Proline Promass P 300
Cable typeA
Characteristic impedance135 to 165 Ω at a measuring frequency of 3 to 20 MHz
Cable capacitance< 30 pF/m
Wire cross-section
Cable typeTwisted pairs
Loop resistance≤110 Ω/km
Signal dampingMax. 9 dB over the entire length of the cable cross-section
ShieldCopper braided shielding or braided shielding with foil shield. When grounding
For further information on planning and installing PROFIBUS networks see:
• Operating Instructions "PROFIBUS DP/PA: Guidelines for planning and commissioning"
(BA00034S)
• PNO Directive 2.092 "PROFIBUS PA User and Installation Guideline"
• IEC 61158-2 (MBP)
EtherNet/IP
The standard ANSI/TIA/EIA-568-B.2 Annex specifies CAT 5 as the minimum category for a cable
used for EtherNet/IP. CAT 5e and CAT 6 are recommended.
For more information on planning and installing EtherNet/IP networks, please refer to the
"Media Planning and Installation Manual. EtherNet/IP" of ODVA Organization
> 0.34 mm2 (22 AWG)
the cable shield, observe the grounding concept of the plant.
PROFINET
Standard IEC 61156-6 specifies CAT 5 as the minimum category for a cable used for PROFINET. CAT
5e and CAT 6 are recommended.
For more information on planning and installing PROFINET networks, see: "PROFINET Cabling
and Interconnection Technology", Guideline for PROFINET
FOUNDATION Fieldbus
Twisted, shielded two-wire cable.
For further information on planning and installing FOUNDATION Fieldbus networks see:
• Operating Instructions for "FOUNDATION Fieldbus Overview" (BA00013S)
• FOUNDATION Fieldbus Guideline
• IEC 61158-2 (MBP)
Modbus RS485
The EIA/TIA-485 standard specifies two types of cable (A and B) for the bus line which can be used
for every transmission rate. Cable type A is recommended.
Cable typeA
Characteristic impedance135 to 165 Ω at a measuring frequency of 3 to 20 MHz
Cable capacitance< 30 pF/m
Wire cross-section
Cable typeTwisted pairs
Loop resistance≤110 Ω/km
Signal dampingMax. 9 dB over the entire length of the cable cross-section
ShieldCopper braided shielding or braided shielding with foil shield. When grounding
> 0.34 mm2 (22 AWG)
the cable shield, observe the grounding concept of the plant.
Current output 0/4 to 20 mA
Standard installation cable is sufficient.
40Endress+Hauser
Page 41
Proline Promass P 300
Pulse/frequency/switch output
Standard installation cable is sufficient.
Double pulse output
Standard installation cable is sufficient.
Relay output
Standard installation cable is sufficient.
Current input 0/4 to 20 mA
Standard installation cable is sufficient.
Status input
Standard installation cable is sufficient.
Connecting cable for transmitter - remote display and operating module DKX001
Standard cable
A standard cable can be used as the connecting cable.
Standard cable4 cores (2 pairs); pair-stranded with common shield
The outputs have the following base accuracy specifications.
Repeatability
Current output
Accuracy±5 µA
Pulse/frequency output
o.r. = of reading
AccuracyMax. ±50 ppm o.r. (over the entire ambient temperature range)
o.r. = of reading; 1 g/cm3 = 1 kg/l; T = medium temperature
Base repeatability
Design fundamentals → 45
Mass flow and volume flow (liquids)
±0.05 % o.r.
Density (liquids)
±0.00025 g/cm
3
Temperature
±0.25 °C ± 0.0025 · T °C (±0.45 °F ± 0.0015 · (T–32) °F)
Response time
Influence of ambient
The response time depends on the configuration (damping).
Current output
temperature
Temperature coefficientMax. 1 μA/°C
Pulse/frequency output
Temperature coefficientNo additional effect. Included in accuracy.
Endress+Hauser43
Page 44
Proline Promass P 300
3
[°C]
-40
0
-80
40 80 120 160 200 240 280 320 360 400
[°F]
-50
0
50
100150
200
0
2
4
6
8
10
14
12
16
1
2
18
20
Influence of medium
temperature
Mass flow and volume flow
o.f.s. = of full scale value
When there is a difference between the temperature for zero point adjustment and the process
temperature, the additional measured error of the sensor is typically ±0.0002 % o.f.s./°C (±0.0001 %
o. f.s./°F).
The effect is reduced if zero point adjustment is performed at process temperature.
Density
When there is a difference between the density calibration temperature and the process
temperature, the typical measured error of the sensor is ±0.0001 g/cm3 /°C (±0.00005 g/cm3 /°F).
Field density calibration is possible.
Wide-range density specification (special density calibration)
If the process temperature is outside the valid range (→ 42) the measured error is
±0.0001 g/cm3 /°C (±0.00005 g/cm3 /°F)
Influence of medium
pressure
A0016610
1Field density calibration, for example at +20 °C (+68 °F)
2Special density calibration
Temperature
±0.005 · T °C (± 0.005 · (T – 32) °F)
The table below shows the effect on accuracy of mass flow due to a difference between calibration
pressure and process pressure.
o.r. = of reading
It is possible to compensate for the effect by:
• Reading in the current pressure measured value via the current input.
• Specifying a fixed value for the pressure in the device parameters.
Operating Instructions→ 110.
DN[% o.r./bar][% o.r./psi]
[mm][in]
8³⁄₈–0.002–0.0001
15½–0.006–0.0004
251–0.005–0.0003
44Endress+Hauser
Page 45
Proline Promass P 300
ZeroPoint
BaseAccu
⋅ 100
³
± BaseAccu
ZeroPoint
BaseAccu
⋅ 100
<
ZeroPoint
MeasValue
⋅ 100
±
½ ⋅ ZeroPoint
BaseRepeat
⋅ 100
³
½ ⋅ ZeroPoint
BaseRepeat
⋅ 100
<
ZeroPoint
MeasValue
⋅ 100
½ ⋅±
E [%]
0102030405060
0
0.5
1.0
1.5
2.0
2.5
708090100
Q [%]
DN[% o.r./bar][% o.r./psi]
[mm][in]
401½–0.007–0.0005
502–0.006–0.0004
Design fundamentals
o.r. = of reading, o.f.s. = of full scale value
BaseAccu = base accuracy in % o.r., BaseRepeat = base repeatability in % o.r.
MeasValue = measured value; ZeroPoint = zero point stability
Calculation of the maximum measured error as a function of the flow rate
Flow rateMaximum measured error in % o.r.
A0021332
A0021333
Calculation of the maximum repeatability as a function of the flow rate
Flow rateMaximum repeatability in % o.r.
A0021335
A0021336
A0021339
A0021334
A0021340
A0021337
Endress+Hauser45
Example for maximum measured error
EMaximum measured error in % o.r. (example)
QFlow rate in % of maximum full scale value
Installation
No special measures such as supports etc. are necessary. External forces are absorbed by the
construction of the device.
A0030317
Page 46
Mounting location
1
2
3
4
5
Proline Promass P 300
A0028772
To prevent measuring errors arising from accumulation of gas bubbles in the measuring tube, avoid
the following mounting locations in the pipe:
• Highest point of a pipeline.
• Directly upstream of a free pipe outlet in a down pipe.
Installation in down pipes
However, the following installation suggestion allows for installation in an open vertical pipeline.
Pipe restrictions or the use of an orifice with a smaller cross-section than the nominal diameter
prevent the sensor running empty while measurement is in progress.
20Installation in a down pipe (e.g. for batching applications)
1Supply tank
2Sensor
3Orifice plate, pipe restriction
4Valve
5Batching tank
DNØ orifice plate, pipe restriction
[mm][in][mm][in]
8³⁄₈60.24
15½100.40
251140.55
Orientation
The direction of the arrow on the sensor nameplate helps you to install the sensor according to the
401½220.87
502281.10
flow direction (direction of medium flow through the piping).
46Endress+Hauser
A0028773
Page 47
Proline Promass P 300
12
OrientationRecommendation
AVertical orientation
A0015591
BHorizontal orientation, transmitter at
top
1)
2)
Exceptions:
→ 21, 47
A0015589
CHorizontal orientation, transmitter at
bottom
3)
Exceptions:
→ 21, 47
A0015590
DHorizontal orientation, transmitter at
side
A0015592
1)This orientation is recommended to ensure self-draining.
2)Applications with low process temperatures may decrease the ambient temperature. To maintain the
minimum ambient temperature for the transmitter, this orientation is recommended.
3)Applications with high process temperatures may increase the ambient temperature. To maintain the
maximum ambient temperature for the transmitter, this orientation is recommended.
Inlet and outlet runs
Special mounting
instructions
If a sensor is installed horizontally with a curved measuring tube, match the position of the sensor to
the fluid properties.
A0028774
21Orientation of sensor with curved measuring tube
1Avoid this orientation for fluids with entrained solids: Risk of solids accumulating.
2Avoid this orientation for outgassing fluids: Risk of gas accumulating.
No special precautions need to be taken for fittings which create turbulence, such as valves, elbows
or T-pieces, as long as no cavitation occurs → 55.
Drainability
When installed vertically, the measuring tube can be drained completely and protected against
buildup.
When the sensor is installed in a horizontal line, eccentric clamps can be used to ensure complete
drainability. When the system is pitched in a specific direction and at a specific slope, gravity can be
used to achieve complete drainability. The sensor must be mounted in the correct position to ensure
full drainability in the horizontal position. Markings on the sensor show the correct mounting
position to optimize drainability.
Endress+Hauser47
Page 48
Proline Promass P 300
1
2
4
3
A
B
C
A0016583
1Eccentric clamp connection
2"This side up" label indicates which side is up
3Slope the device in accordance with the hygiene guidelines. Slope: approx. 2 ° or 35 mm/m (0.42 in/feet)
4Transmitter
5Line on the underside indicates the lowest point of the eccentric process connection.
Securing with mounting clamp in the case of hygiene connections
It is not necessary to provide additional support for the sensor for operational performance purposes.
If, however, additional support is required for installation purposes, the following dimensions must
be observed.
Use mounting clamp with lining between clamp and measuring instrument.
A0030298
DNABC
[mm][in][mm][in][mm][in][mm][in]
8³⁄₈29811.73331.3281.1
15½40215.83331.3281.1
25154221.34331.3381.5
401 ½65825.9136.51.44562.2
50277230.3944.11.74752.95
Zero point adjustment
All measuring devices are calibrated in accordance with state-of-the-art technology. Calibration
takes place under reference conditions→ 42. Therefore, a zero point adjustment in the field is
generally not required.
Experience shows that zero point adjustment is advisable only in special cases:
• To achieve maximum measuring accuracy even with low flow rates.
• Under extreme process or operating conditions (e.g. very high process temperatures or very highviscosity fluids).
48Endress+Hauser
Page 49
Proline Promass P 300
146 (5.75)
48 (1.9)
12 (0.47)
280 (11.0)255 (10.0)
134 (5.3)
30 (1.18)
Protective cover
A0029553
Environment
Ambient temperature range
Storage temperature
Climate class
Degree of protection
Measuring device• –40 to +60 °C (–40 to +140 °F)
• Order code for "Test, certificate", option JP:
–50 to +60 °C (–58 to +140 °F)
Readability of the local
display
–20 to +60 °C (–4 to +140 °F)
The readability of the display may be impaired at temperatures outside the
temperature range.
Dependency of ambient temperature on medium temperature→ 50
If operating outdoors:
‣
Avoid direct sunlight, particularly in warm climatic regions.
You can order a weather protection cover from Endress+Hauser. → 107.
–50 to +80 °C (–58 to +176 °F)
DIN EN 60068-2-38 (test Z/AD)
Measuring device
• As standard: IP66/67, type 4X enclosure
• When housing is open: IP20, type 1 enclosure
• Display module: IP20, type 1 enclosure
• With the order code for "Sensor options", option CM: IP69 can also be ordered
External WLAN antenna
IP67
Vibration resistance
• Oscillation, sinusoidal, following IEC 60068-2-6
– 2 to 8.4 Hz, 3.5 mm peak
– 8.4 to 2000 Hz, 1 g peak
• Oscillation, broadband noise following IEC 60068-2-64
– 10 to 200 Hz, 0.003 g2/Hz
– 200 to 2000 Hz, 0.001 g2/Hz
– Total: 1.54 g rms
Shock resistance
Shock, half-sine according to IEC 60068-2-27
6 ms 50 g
Endress+Hauser49
Page 50
Proline Promass P 300
T
a
T
m
B
A
Shock resistance
Interior cleaning
Electromagnetic
compatibility (EMC)
Medium temperature range
Shock due to rough handling following IEC 60068-2-31
• Cleaning in place (CIP)
• Sterilization in place (SIP)
• Cleaning with pigs
Options
Oil- and grease-free version for wetted parts, without declaration
Order code for "Service", option HA
• As per IEC/EN 61326 and NAMUR Recommendation 21 (NE 21)
• Device version with PROFIBUS DP: Complies with emission limits for industry as per EN 50170
Volume 2, IEC 61784
The following applies for PROFIBUS DP: If baud rates > 1.5 MBaud, an EMC cable entry must be
used and the cable shield must continue as far as the terminal wherever possible.
Details are provided in the Declaration of Conformity.
Process
Standard version–50 to +150 °C (–58 to +302 °F)Order code for "Measuring tube
mat., wetted surface", option BB,
BC, BD
Extended temperature version–50 to +205 °C (–58 to +401 °F)Order code for "Measuring tube
mat., wetted surface", option TD, TG
Dependency of ambient temperature on medium temperature
A0031121
22Exemplary representation, values in the table below.
TaAmbient temperature
TmMedium temperature
AMaximum permitted medium temperature Tm at T
require a reduced ambient temperature T
BMaximum permitted ambient temperature Ta for the maximum specified medium temperature Tm of the
sensor
a
= 60 °C (140 °F); higher medium temperatures T
a max
m
Values for devices used in the hazardous area:
Separate Ex documentation (XA) for the device → 110.
50Endress+Hauser
Page 51
Proline Promass P 300
PN40
300[° ]F350250200150100500- 05
[°C]200150100500- 05
400
0
10
20
30
40
50
200
400
600
0
800
[bar][psi]
Class 150
300[°F]350250200150100500-50
[°C]200150100500-50
400
0
10
20
30
[bar][psi]
200
400
0
Not insulatedInsulated
ABAB
VersionT
Standard version60 °C (140 °F)150 °C (302 °F)––60 °C (140 °F)110 °C (230 °F)55 °C (131 °F)150 °C (302 °F)
Extended temperature version60 °C (140 °F)205 °C (401 °F)––60 °C (140 °F)110 °C (230 °F)50 °C (122 °F)205 °C (401 °F)
a
T
m
TaT
m
T
a
T
m
T
a
T
m
Density
Pressure-temperature
ratings
0 to 5000 kg/m3 (0 to 312 lb/cf)
The following pressure/temperature diagrams apply to all pressure-bearing parts of the device and
not just the process connection. The diagrams show the maximum permissible medium pressure
depending on the specific medium temperature.
Pressure-temperature curves with temperature range +151 to +205 °C (+304 to +401 °F)
exclusively for extended temperature version of measuring devices.
Flange according to EN 1092-1 (DIN 2501)
A0027769-EN
23With flange material 1.4404 (F316/F316L).
Flange according to ASME B16.5
A0026540-EN
24With flange material 1.4404 (F316/F316L)
Endress+Hauser51
Page 52
Flange JIS B2220
20K
300[°F]350250200150100500-50
[°C]200150100500-50
400
0
10
20
30
[bar][psi]
200
400
0
DN8…40
≥ DN50
-80
080160
240
320400
0
10
20
30
40
0
[°C]
[°F]
[bar][psi]
20015010050
0
-50
50
300[°F]350250200150100500-50
[°C]200150100500-50
400
0
10
20
30
[bar][psi]
200
400
0
25With flange material 1.4404 (F316/F316L)
Flange DIN 11864-2 Form A
Proline Promass P 300
A0026541-EN
26With flange material 1.4435 (316L)
Neumo BioConnect, BBS flange
27With flange material 1.4435 (316L)
A0027781-EN
A0027782-EN
52Endress+Hauser
Page 53
Proline Promass P 300
DN8…40
300[°F]350250200150100500-50
[°C]200150100500-50
400
0
10
20
30
40
50
[bar][psi]
200
400
600
0
≥DN50
DN8…40
≥ DN50
-80
080160
240
320400
0
10
20
30
40
0
[°C]
[°F]
[bar][psi]
20015010050
0
-50
50
-80
080160
240
320400
0
10
20
0
[°C]
[°F]
[bar][psi]
20015010050
0
-50
Thread DIN 11851
A0027783-EN
28With flange material 1.4435 (316L)
DIN 11851 allows for applications up to +140 °C (+284 °F) if suitable sealing materials are used.
Please take this into account when selecting seals and counterparts, as these components can limit
the pressure and temperature range.
Thread DIN 11864-1 Form A
29With connection material 1.4435 (316L)
Thread ISO 2853
A0027784-EN
A0027785-EN
30With connection material 1.4435 (316L)
Endress+Hauser53
Page 54
Proline Promass P 300
-80
080160
240
320400
0
2
6
0
[°C]
[°F]
[bar][psi]
20015010050
0
-50
8
DN8…40
≥ DN50
-80
080160
240
320400
0
10
20
30
40
0
[°C]
[°F]
[bar][psi]
20015010050
0
-50
50
Thread SMS 1145
A0027786-EN
31With connection material 1.4435 (316L)
SMS 1145 allows for applications up to 6 bar (87 psi) if suitable sealing materials are used. Please
take this into account when selecting seals and counterparts, as these components can limit the
pressure and temperature range.
Clamp connections DIN 11864-3 Form A
32With connection material 1.4435 (316L)
Tri-Clamp; clamp connection ISO 2852, DIN 32676, BBS, Neumo BioConnect
The clamp connections are suitable up to a maximum pressure of 16 bar (232 psi). Please observe
the operating limits of the clamp and seal used as they can be over 16 bar (232 psi). The clamp and
seal are not included in the scope of supply.
Sensor housing
The sensor housing is filled with dry nitrogen gas and protects the electronics and mechanics inside.
If a measuring tube fails (e.g. due to process characteristics like corrosive or abrasive fluids), the
fluid will initially be contained by the sensor housing.
If the sensor is to be purged with gas (gas detection), it should be equipped with purge connections.
Do not open the purge connections unless the containment can be filled immediately with a
dry, inert gas. Use only low pressure to purge. Maximum pressure: 5 bar (72.5 psi).
Sensor housing nominal pressure rating and burst pressure
The following sensor housing nominal pressure ratings/burst pressures are only valid for standard
devices and/or devices equipped with closed purge connections (not opened/as delivered).
If a device fitted with purge connections (order code for "Sensor option", option CH "Purge
54Endress+Hauser
connection") is connected to the purge system, the maximum nominal pressure is determined by the
purge system itself or by the device, depending on which component has the lower nominal pressure
classification.
The sensor housing burst pressure refers to a typical internal pressure which is reached prior to
mechanical failure of the sensor housing and which was determined during type testing. The
A0027784-EN
Page 55
Proline Promass P 300
corresponding type test declaration can be ordered with the device (order code for "Additional
approval", option LN "Sensor housing burst pressure, type test").
Flow limit
DNSensor housing nominal
pressure
(designed with a safety factor
≥ 4)
[mm][in][bar][psi][bar][psi]
8³⁄₈253621902755
15½253621752538
251253621652392
401½253621522204
502253621031494
Sensor housing burst pressure
For information on the dimensions: see the "Mechanical construction" section→ 57
Select the nominal diameter by optimizing between the required flow range and permissible pressure
loss.
For an overview of the full scale values for the measuring range, see the "Measuring range"
section → 10
• The minimum recommended full scale value is approx. 1/20 of the maximum full scale value
• In most applications, 20 to 50 % of the maximum full scale value can be considered ideal
• A low full scale value must be selected for abrasive media (such as liquids with entrained solids):
flow velocity < 1 m/s (< 3 ft/s).
To calculate the flow limit, use the Applicator sizing tool → 109
Pressure loss
System pressure
Thermal insulation
To calculate the pressure loss, use the Applicator sizing tool → 109
It is important that cavitation does not occur, or that gases entrained in the liquids do not outgas.
This is prevented by means of a sufficiently high system pressure.
For this reason, the following mounting locations are recommended:
• At the lowest point in a vertical pipe
• Downstream from pumps (no danger of vacuum)
A0028777
In the case of some fluids, it is important to keep the heat radiated from the sensor to the
transmitter to a low level. A wide range of materials can be used for the required insulation.
The following device versions are recommended for versions with thermal insulation:
• Version with extended neck for insulation:
Order code for "Sensor option", option CG with an extended neck length of 105 mm (4.13 in).
• Extended temperature version:
Order code for "Measuring tube material", option TD or TG with an extended neck length of
105 mm (4.13 in).
Endress+Hauser55
Page 56
Proline Promass P 300
NOTICE
Electronics overheating on account of thermal insulation!
Maximum permissible temperature at the lower end of the transmitter housing: 80 °C (176 °F)
‣
Thermal insulation with extended neck free: We recommend that you do not insulate the
‣
extended neck in order to ensure optimum dissipation of heat.
33Thermal insulation with extended neck free
A0034391
Heating
Vibrations
Some fluids require suitable measures to avoid loss of heat at the sensor.
Heating options
• Electrical heating, e.g. with electric band heaters
• Via pipes carrying hot water or steam
• Via heating jackets
Heating jackets for the sensors can be ordered as accessories from Endress+Hauser .→ 108
NOTICE
Danger of overheating when heating
Ensure that the temperature at the lower end of the transmitter housing does not exceed
‣
80 °C (176 °F).
Ensure that sufficient convection takes place at the transmitter neck.
‣
Ensure that a sufficiently large area of the transmitted neck remains exposed. The uncovered
‣
part serves as a radiator and protects the electronics from overheating and excessive cooling.
If using in potentially explosive atmospheres, observe the information in the device-specific Ex
‣
documentation. For detailed information on the temperature tables, see the separate document
entitled "Safety Instructions" (XA) for the device.
The high oscillation frequency of the measuring tubes ensures that the correct operation of the
measuring system is not influenced by plant vibrations.
56Endress+Hauser
Page 57
Proline Promass P 300
D
L
E
F
A
CB
K
G
H
I
M
Mechanical construction
Dimensions in SI unitsCompact version
A0033783
Order code for "Housing", option A "Aluminum, coated"
1)Depending on the cable gland used: values up to + 30 mm
2)With order code for "Sensor option", option CG or order code for "Measuring tube material", option TD, TG:
values +70 mm
3)Depends on the process connection in question
2)
2)
F
GHIKLM
3)
92
3)
92
3)
92
3)
142
3)
169
58Endress+Hauser
Page 59
Proline Promass P 300
C
D
A
B
E
L
Flange connections
Fixed flange EN 1092-1, ASME B16.5, JIS B2220
A0015621
Length tolerance for dimension L in mm:
+1.5 / –2.0
Flange according to EN 1092-1 (DIN 2501): PN 40
1.4404 (316/316L)
Order code for "Process connection", option D2W
DN
[mm]
1)
8
1595654 × Ø1420.017.3440
25115854 × Ø1419.028.5580
401501104 × Ø1821.043.1707
501651254 × Ø1825.054.5828
Surface roughness (flange): EN 1092-1 Form B1 (DIN 2526 Form C), Ra 3.2 to 12.5 µm
1)DN 8 with DN 15 flanges as standard
Flange according to ASME B16.5: Cl 150
1.4404 (316/316L)
Order code for "Process connection", option AAW
DN
[mm]
1)
8
159060.34 × Ø15.711.215.7440
2511079.44 × Ø15.714.226.7580
4012598.44 × Ø15.717.540.9707
50150120.78 × Ø19.119.152.6828
Surface roughness (flange): Ra 3.2 to 6.3 µm
A
[mm]
95654 × Ø1417.017.3336
A
[mm]
9060.34 × Ø15.711.215.7336
[mm]
B
[mm]
B
C
[mm]
C
[mm]
D
[mm]
[mm]
E
[mm]
D
E
[mm]
L
[mm]
L
[mm]
1)DN 8 with DN 15 flanges as standard
Endress+Hauser59
Page 60
Proline Promass P 300
Flange JIS B2220: 20K
1.4404 (316/316L)
Order code for "Process connection", option NEW
DN
[mm]
1)
8
1595704 × Ø1516.015440
25125904 × Ø1917.525580
401401054 × Ø1920.040707
501551208 × Ø1927.550828
Surface roughness (flange): Ra 3.2 to 6.3 µm
1)DN 8 with DN 15 flanges as standard
A
[mm]
95704 × Ø1516.015336
B
[mm]
C
[mm]
D
[mm]
E
[mm]
L
[mm]
60Endress+Hauser
Page 61
Proline Promass P 300
C
D
A
B
E
XX
L
Fixed flange DIN 11864-2
A0015627
34Detail X: Asymmetrical process connection; the part shown in blue is provided by the supplier.
Length tolerance for dimension L in mm:
+1.5 / –2.0
Flange DIN 11864-2 Form A, for pipe according to DIN 11866 series A, flange with notch
1.4435 (316L)
Order code for "Process connection", option KJW
DN
[mm]
859.0424 × Ø91016.00384
1559.0424 × Ø91016.00488
2570.0534 × Ø91026.00626
4082.0654 × Ø91038.00753
5094.0774 × Ø91050.00877
Ra
0.76 µm: order code for "Measuring tube material", option BB, TD
max
Ra
0.38 µm electropolished: order code for "Measuring tube material", option BC, TG
max
A
[mm]
B
[mm]
C
[mm]
D
[mm]
E
[mm]
L
[mm]
Endress+Hauser61
Page 62
Proline Promass P 300
L
C
D
A
B
E
XX
BBS fixed flange
35Detail X: Asymmetrical process connection; the part shown in blue is provided by the supplier.
Length tolerance for dimension L in mm:
+1.5 / –2.0
A0016910
BBS flange small (sterile orbital), for pipe according to DIN 11866 series A, female
1.4435 (316L)
Order code for "Process connection", option BSK
DN
[mm]
859424 × Ø91010.00384
1559424 × Ø91016.00488
2570534 × Ø91026.00626
4082654 × Ø91038.00753
5094774 × Ø91050.00877
Ra
0.76 µm: order code for "Measuring tube material", option BB, TD
max
Ra
0.38 µm electropolished: order code for "Measuring tube material", option BC, TG
max
BBS flange small (sterile orbital), for pipe according to DIN 11866 series B, female
1.4435 (316L)
Order code for "Process connection", option BSL
DN
[mm]
859424 × Ø91014.00384
1562454 × Ø91018.10488
2574574 × Ø91029.70626
4088714 × Ø91044.30753
50103854 × Ø91056.30877
Ra
0.76 µm: order code for "Measuring tube material", option BB, TD
max
Ra
0.38 µm electropolished: order code for "Measuring tube material", option BC, TG
max
A
[mm]
A
[mm]
B
[mm]
B
[mm]
C
[mm]
C
[mm]
D
[mm]
D
[mm]
E
[mm]
E
[mm]
L
[mm]
L
[mm]
62Endress+Hauser
Page 63
Proline Promass P 300
L
C
D
A
B
E
XX
Neumo BioConnect fixed flange
A0016907
36Detail X: Asymmetrical process connection; the part shown in blue is provided by the supplier.
Length tolerance for dimension L in mm:
+1.5 / –2.0
Neumo BioConnect flange (sterile orbital), for pipe according to DIN 11866 series A, flange form R
1.4435 (316L)
Order code for "Process connection", option BSB
DN
[mm]
865454 × Ø91010.00384
1575554 × Ø91016.00488
2585654 × Ø91226.00626
40100804 × Ø91238.00753
50110904 × Ø91450.00877
Ra
0.76 µm: order code for "Measuring tube material", option BB, TD
max
Ra
0.38 µm electropolished: order code for "Measuring tube material", option BC, TG
max
A
[mm]
B
[mm]
C
[mm]
D
[mm]
E
[mm]
L
[mm]
Endress+Hauser63
Page 64
Clamp connections
L
A
B
≥(1")DN 25DN 8, 15 (¾", ½")
A
B
Tri-Clamp
37Engineering unit mm (in)
Length tolerance for dimension L in mm:
+1.5 / –2.0
Proline Promass P 300
A0023342
Tri-Clamp, for pipe according to DIN 11866 series C
1.4435 (316L)
Order code for "Process connection", option FHW
DN
[mm]
Clamp
[in]
A
[mm]
B
[mm]
8½25.09.40362
15¾25.015.75466
251
401½
502
Ra
0.76 µm: order code for "Measuring tube material", option BB, TD
max
Ra
0.38 µm electropolished: order code for "Measuring tube material", option BC, TG
max
1)
1)
1)
50.422.10606
50.434.80731
63.947.50853
1)The connection complies with the hygienic clamp dimensions as per ASME BPE.
½" Tri-Clamp, for pipe according to DIN 11866 series C
1.4435 (316L)
Order code for "Process connection", option FCW
DN
[mm]
Clamp
[in]
A
[mm]
B
[mm]
15½25.09.40466
Ra
0.76 µm: order code for "Measuring tube material", option BB, TD
max
Ra
0.38 µm electropolished: order code for "Measuring tube material", option BC, TG
max
L
[mm]
L
[mm]
¾" Tri-Clamp, for pipe according to DIN 11866 series C
1.4435 (316L)
Order code for "Process connection", option FFW
DN
[mm]
Clamp
[in]
A
[mm]
B
[mm]
L
[mm]
8¾25.015.75362
Ra
0.76 µm: order code for "Measuring tube material", option BB, TD
max
Ra
0.38 µm electropolished: order code for "Measuring tube material", option BC, TG
max
64Endress+Hauser
Page 65
Proline Promass P 300
L
A
B
1" Tri-Clamp, for pipe according to DIN 11866 series C
1.4435 (316L)
Order code for "Process connection", option FPW
DN
[mm]
81
151
Ra
0.76 µm: order code for "Measuring tube material", option BB, TD
max
Ra
0.38 µm electropolished: order code for "Measuring tube material", option BC, TG
max
Clamp
[in]
1)
1)
A
[mm]
B
[mm]
50.422.10362
50.422.10466
1)The connection complies with the hygienic clamp dimensions as per ASME BPE.
Tri-Clamp eccentric, for pipe according to DIN 11866 series C
1.4435 (316L)
DN
[mm]
Order code for
"Process connection",
Clamp
[in]
A
[mm]
B
[mm]
option
8FEB½25.09.40362
15FED¾25.015.75466
25FEF1
40FEH1½
50FEK2
Ra
0.76 µm: order code for "Measuring tube material", option BB, TD
max
Ra
0.38 µm electropolished: order code for "Measuring tube material", option BC, TG
max
1)
1)
1)
50.422.10606
50.434.80738
63.947.50860
Additional information on "Eccentric clamps"
L
[mm]
L
[mm]
1)The connection complies with the hygienic clamp dimensions as per ASME BPE.
Clamp connection DIN 32676, ISO 2852
A0015625
Length tolerance for dimension L in mm:
+1.5 / –2.0
Clamp DIN 32676, for pipe according to DIN 11866 series A
1.4435 (316L)
Order code for "Process connection", option KPW
DN
[mm]
834.016.00362
1534.016.00466
2550.526.00606
A
[mm]
B
[mm]
L
[mm]
Endress+Hauser65
Page 66
Clamp DIN 32676, for pipe according to DIN 11866 series A
1.4435 (316L)
Order code for "Process connection", option KPW
DN
[mm]
A
[mm]
B
[mm]
4050.538.00732
5064.050.00854
Ra
0.76 µm: order code for "Measuring tube material", option BB, TD
max
Ra
0.38 µm electropolished: order code for "Measuring tube material", option BC, TG
max
Clamp ISO 2852 , for pipe according to ISO 2037
1.4435 (316L)
Order code for "Process connection", option JSA
DN
[mm]
A
[mm]
B
[mm]
850.522.6362
1550.522.6466
2550.522.6606
4050.535.6731
5064.048.6853
Ra
0.76 µm: order code for "Measuring tube material", option BB, TD
max
Ra
0.38 µm electropolished: order code for "Measuring tube material", option BC, TG
max
Proline Promass P 300
L
[mm]
L
[mm]
Clamp ISO 2852, for pipe according to DIN11866 series B
1.4435 (316L)
Order code for "Process connection", option JSC
DN
[mm]
A
[mm]
B
[mm]
834.014.00362
1534.018.10466
2550.529.70606
4064.044.30731
5077.556.30853
Ra
0.76 µm: order code for "Measuring tube material", option BB, TD
max
Ra
0.38 µm electropolished: order code for "Measuring tube material", option BC, TG
max
L
[mm]
66Endress+Hauser
Page 67
Proline Promass P 300
A
B
X
L
X
L
X
A
B
X
Clamp connection DIN 11864-3
A0016903
38Detail X: Asymmetrical process connection; the part shown in blue is provided by the supplier.
Length tolerance for dimension L in mm:
+1.5 / –2.0
Clamp DIN 11864-3 Form A, with notch, for pipe according to DIN 11866 series A
1.4435 (316L)
Order code for "Process connection", option KLW
DN
[mm]
834.016.05370
1534.016.05474
2550.526.05614
4064.038.05738
5077.550.05853
Ra
0.76 µm: order code for "Measuring tube material", option BB, TD
max
Ra
0.38 µm electropolished: order code for "Measuring tube material", option BC, TG
max
A
[mm]
B
[mm]
BBS clamp connection
L
[mm]
A0016908
39Detail X: Asymmetrical process connection; the part shown in blue is provided by the supplier.
Length tolerance for dimension L in mm:
+1.5 / –2.0
Endress+Hauser67
Page 68
Proline Promass P 300
L
X
A
B
X
BBS Quick Connect (sterile orbital), for pipe according to DIN 11866 series A, female
1.4435 (316L)
Order code for "Process connection", option BSE
DN
[mm]
825.010.00362
1550.516.00466
2550.526.00606
4064.038.00732
5077.550.00854
Ra
0.76 µm: order code for "Measuring tube material", option BB, TD
max
Ra
0.38 µm electropolished: order code for "Measuring tube material", option BC, TG
max
BBS Quick Connect (sterile orbital) for pipe according to DIN11866 series B, female
1.4435 (316L)
Order code for "Process connection", option BSJ
DN
[mm]
825.014.00362
1550.518.10466
2550.529.70606
4064.044.30738
5077.556.30860
Ra
0.76 µm: order code for "Measuring tube material", option BB, TD
max
Ra
0.38 µm electropolished: order code for "Measuring tube material", option BC, TG
max
A
[mm]
A
[mm]
B
[mm]
B
[mm]
L
[mm]
L
[mm]
Neumo BioConnect clamp connection
A0016905
40Detail X: Asymmetrical process connection; the part shown in blue is provided by the supplier.
Length tolerance for dimension L in mm:
+1.5 / –2.0
68Endress+Hauser
Page 69
Proline Promass P 300
L
A
B
Clamp, Neumo BioConnect (sterile orbital), for pipe according to DIN 11866 series A, clamp form R
1.4435 (316L)
Order code for "Process connection", option BSA
DN
[mm]
A
[mm]
B
[mm]
L
[mm]
825.010.00362
1525.016.00466
2550.426.00606
4064.038.00732
5077.450.00854
Ra
0.76 µm: order code for "Measuring tube material", option BB, TD
max
Ra
0.38 µm electropolished: order code for "Measuring tube material", option BC, TG
max
Eccentric clamp connection DIN 32676, ISO 2852
Length tolerance for dimension L in mm:
+1.5 / –2.0
Eccentric clamp DIN 32676, for pipe according to DIN 11866 series A
1.4435 (316L)
Order code for "Process connection", option KRW
DN
[mm]
834.010.00362
1534.016.00466
2550.526.00606
5064.050.00860
Ra
0.76 µm: order code for "Measuring tube material", option BB, TD
max
Ra
0.38 µm electropolished: order code for "Measuring tube material", option BC, TG
max
Eccentric clamp ISO 2852, for pipe according to DIN11866 series B
1.4435 (316L)
Order code for "Process connection", option JEC
DN
[mm]
834.010.30362
1534.014.00466
2534.018.10606
4050.529.70738
A
[mm]
A
[mm]
B
[mm]
B
[mm]
A0016543
L
[mm]
L
[mm]
Endress+Hauser69
Page 70
Proline Promass P 300
X
L
A
B
X
Eccentric clamp ISO 2852, for pipe according to DIN11866 series B
1.4435 (316L)
Order code for "Process connection", option JEC
DN
[mm]
5064.044.30853
Ra
0.76 µm: order code for "Measuring tube material", option BB, TD
max
Ra
0.38 µm electropolished: order code for "Measuring tube material", option BC, TG
max
Additional information on "Eccentric clamps"
Eccentric clamp ISO 2852, for pipe according to DIN11866 series B, for connection to DN15 pipes
1.4435 (316L)
Order code for "Process connection", option JED
DN
[mm]
2550.518.10606
Ra
0.76 µm: order code for "Measuring tube material", option BB, TD
max
Ra
0.38 µm electropolished: order code for "Measuring tube material", option BC, TG
max
Additional information on "Eccentric clamps"
A
[mm]
A
[mm]
B
[mm]
B
[mm]
L
[mm]
L
[mm]
Eccentric clamp connection DIN 11864-3
A0016904
41Detail X: Asymmetrical process connection; the part shown in blue is provided by the supplier.
Length tolerance for dimension L in mm:
+1.5 / –2.0
Eccentric clamp DIN 11864-3 Form A, with notch, for pipe according to DIN 11866 series A
1.4435 (316L)
Order code for "Process connection", option KNW
DN
[mm]
834.010.00370
1534.016.00474
2550.526.00624
5077.550.00869
Ra
0.76 µm: order code for "Measuring tube material", option BB, TD
max
Ra
0.38 µm electropolished: order code for "Measuring tube material", option BC, TG
max
Additional information on "Eccentric clamps"
A
[mm]
B
[mm]
L
[mm]
70Endress+Hauser
Page 71
Proline Promass P 300
X
L
A
B
X
BBS eccentric clamp connection
42Detail X: Asymmetrical process connection; the part shown in blue is provided by the supplier.
Length tolerance for dimension L in mm:
+1.5 / –2.0
BBS Quick Connect (sterile orbital), eccentric, for pipe according to DIN11866 series A, female
1.4435 (316L)
Order code for "Process connection", option BEJ
DN
[mm]
825.010.00362
1550.516.00466
2550.526.00606
5077.550.00860
Ra
0.76 µm: order code for "Measuring tube material", option BB, TD
max
Ra
0.38 µm electropolished: order code for "Measuring tube material", option BC, TG
max
Additional information on "Eccentric clamps"
A
[mm]
B
[mm]
[mm]
A0016909
L
BBS Quick Connect (sterile orbital), eccentric, for pipe according to DIN11866 series B, female
1.4435 (316L)
Order code for "Process connection", option BEK
DN
[mm]
A
[mm]
B
[mm]
L
[mm]
825.010.30362
1550.514.00466
2550.518.10606
4050.529.70738
5064.044.30860
Ra
0.76 µm: order code for "Measuring tube material", option BB, TD
max
Ra
0.38 µm electropolished: order code for "Measuring tube material", option BC, TG
max
Additional information on "Eccentric clamps"
Endress+Hauser71
Page 72
Proline Promass P 300
X
L
A
B
X
L
A
B
Neumo BioConnect eccentric clamp connection
43Detail X: Asymmetrical process connection; the part shown in blue is provided by the supplier.
Length tolerance for dimension L in mm:
+1.5 / –2.0
A0016906
Neumo BioConnect eccentric clamp connection, for pipe according to DIN 11866 series C, clamp form R
1.4435 (316L)
Order code for "Process connection", option BEA
DN
[mm]
82510.00362
152516.00466
252526.00610
502550.00859
Ra
0.76 µm: order code for "Measuring tube material", option BB, TD
max
Ra
0.38 µm electropolished: order code for "Measuring tube material", option BC, TG
max
Additional information on "Eccentric clamps"
A
[mm]
B
[mm]
L
[mm]
Cable glands
Thread DIN 11851, DIN 11864-1, SMS 1145, BBS
Length tolerance for dimension L in mm:
+1.5 / –2.0
72Endress+Hauser
A0015628
Page 73
Proline Promass P 300
Thread DIN 11851, for pipe according to DIN 11866 series A
1.4435 (316L)
Order code for "Process connection", option KCW
DN
[mm]
A
[mm]
B
[mm]
8Rd 34 × ¹⁄₈16362
15Rd 34 × ¹⁄₈16466
25Rd 52 × ¹⁄₆26606
40Rd 65 × ¹⁄₆38738
50Rd 78 × ¹⁄₆50864
Ra
0.76 µm: order code for "Measuring tube material", option BB, TD
max
Ra
0.38 µm electropolished: order code for "Measuring tube material", option BC, TG
max
Thread DIN 11851, Rd 28 × ¹⁄₈", for pipe according to DIN 11866 series A
1.4435 (316L)
Order code for "Process connection", option KAW
DN
[mm]
A
[mm]
B
[mm]
8Rd 28 × ¹⁄₈"10.00362
15Rd 28 × ¹⁄₈"10.00466
Ra
0.76 µm: order code for "Measuring tube material", option BB, TD
max
Ra
0.38 µm electropolished: order code for "Measuring tube material", option BC, TG
max
L
[mm]
L
[mm]
Thread DIN 11864-1 Form A, for pipe according to DIN 11866 series A
1.4435 (316L)
Order code for "Process connection", option KGW
DN
[mm]
A
[mm]
B
[mm]
8Rd 34 × ¹⁄₈"16362
15Rd 34 × ¹⁄₈"16466
25Rd 52 × ¹⁄₆"26620
40Rd 65 × ¹⁄₆"38738
50Rd 78 × ¹⁄₆"50864
Ra
0.76 µm: order code for "Measuring tube material", option BB, TD
max
Ra
0.38 µm electropolished: order code for "Measuring tube material", option BC, TG
max
Threaded hygienic connection SMS 1145
1.4435 (316L)
Order code for "Process connection", option SAW
DN
[mm]
A
[mm]
B
[mm]
8Rd 40 × ¹⁄₆"22.50362
15Rd 40 × ¹⁄₆"22.50466
25Rd 40 × ¹⁄₆"22.50606
40Rd 60 × ¹⁄₆"35.50742
50Rd 70 × ¹⁄₆"48.50864
Ra
0.76 µm: order code for "Measuring tube material", option BB, TD
max
Ra
0.38 µm electropolished: order code for "Measuring tube material", option BC, TG
max
L
[mm]
L
[mm]
Endress+Hauser73
Page 74
BBS thread (sterile orbital), for pipe according to DIN 11866 series A
L
A
B
1.4435 (316L)
Order code for "Process connection", option BSC
DN
[mm]
A
[mm]
B
[mm]
8M22 ×1.510.00362
15M30 × 216.00466
25M42 × 226.00606
40M52 × 238.00732
50M68 × 250.00854
Ra
0.76 µm: order code for "Measuring tube material", option BB, TD
max
Ra
0.38 µm electropolished: order code for "Measuring tube material", option BC, TG
max
BBS thread (sterile orbital), for pipe according to DIN 11866 series B
1.4435 (316L)
Order code for "Process connection", option BSD
DN
[mm]
A
[mm]
B
[mm]
8M26 × 1.510.00362
15M30 × 216.00466
25M42 × 226.00606
40M56 × 238.00738
50M68 × 250.00860
Ra
0.76 µm: order code for "Measuring tube material", option BB, TD
max
Ra
0.38 µm electropolished: order code for "Measuring tube material", option BC, TG
max
Proline Promass P 300
L
[mm]
L
[mm]
Threaded hygienic connection ISO 2853
Length tolerance for dimension L in mm:
+1.5 / –2.0
Threaded hygienic connection ISO 2853, for pipe according to ISO 2037
1.4435 (316L)
Order code for "Process connection", option JSD
DN
[mm]
837.1322.60370
1537.1322.60474
2537.1322.60614
A
[mm]
B
[mm]
A0015623
L
[mm]
74Endress+Hauser
Page 75
Proline Promass P 300
L
45°
A
1
½" NPT
35 (1.38)
1
1
L
A
11
DN 8 ( ")...25 (1")!⁄#
DN 40(1 ")…50 (2")
½
Threaded hygienic connection ISO 2853, for pipe according to ISO 2037
1.4435 (316L)
Order code for "Process connection", option JSD
DN
[mm]
4050.6535.60742
5064.1048.60864
Ra
0.76 µm: order code for "Measuring tube material", option BB, TD
max
Ra
0.38 µm electropolished: order code for "Measuring tube material", option BC, TG
max
A
[mm]
B
[mm]
Accessories
Rinse connections
L
[mm]
44
1Connection nipple for purge connections: order code for "Sensor options", option CH "Purge connection"
A0029971
DNAL
[mm][mm][mm]
847110
1547204
2547348
4068.15418
5081.65473
Endress+Hauser75
Page 76
Protective cover
146 (5.75)
48 (1.9)
12 (0.47)
280 (11.0)255 (10.0)
134 (5.3)
30 (1.18)
136 (5.35)
114 (4.49)
86 (3.39)
78 (3.07)
!103 (4.06)
!12 (0.47)
7.0 (0.27)
68 (2.7)105 (4.1)
173 (6.8)
Remote display and operating module DKX001
Proline Promass P 300
A0029553
45Engineering unit mm (in)
External WLAN antenna
The external WLAN antenna is not suitable for use in hygienic applications.
External WLAN antenna mounted on device
46Engineering unit mm (in)
A0028921
A0028923
76Endress+Hauser
Page 77
Proline Promass P 300
70 (2.8)
1500 (59.1)
72 (2.8)
External WLAN antenna mounted with cable
The external WLAN antenna can be mounted separately from the transmitter if the transmission/
reception conditions at the transmitter mounting location are poor.
A0033597
47Engineering unit mm (in)
Endress+Hauser77
Page 78
Dimensions in US unitsCompact version
D
L
E
F
A
CB
K
G
H
I
M
Order code for "Housing", option A "Aluminum, coated"
1)Depending on the cable gland used: values up to + 1.18 in
2)With order code for "Sensor option", option CG or order code for "Measuring tube material", option TD, TG:
values +2.76 in
3)Depends on the process connection in question
Flange connections
Fixed flange ASME B16.5
A0015621
Length tolerance for dimension L in inch:
+0.06 / –0.08
Flange according to ASME B16.5: Cl 150
1.4404 (316/316L)
Order code for "Process connection", option AAW
DN
[in]
³⁄₈
1)
A
[in]
B
[in]
C
[in]
D
[in]
[in]
E
L
[in]
3.542.374 × Ø0.620.440.6213.23
½3.542.374 × Ø0.620.440.6217.32
14.333.134 × Ø0.620.561.0522.83
1½4.923.874 × Ø0.620.691.6127.83
25.914.754 × Ø0.750.752.0732.6
Surface roughness (flange): Ra 3.2 to 6.3 µm
1)DN ³⁄₈" with DN ½" flanges as standard
Endress+Hauser79
Page 80
Proline Promass P 300
L
C
D
A
B
E
XX
L
A
B
≥(1")DN 25DN 8, 15 (¾", ½")
A
B
Neumo BioConnect fixed flange
48Detail X: Asymmetrical process connection; the part shown in blue is provided by the supplier.
Length tolerance for dimension L in inch:
+0.06 / –0.08
A0016907
Neumo BioConnect flange (sterile orbital), for pipe according to DIN 11866 series A, flange form R
1.4435 (316L)
Order code for "Process connection", option BSB
DN
[in]
³⁄₈2.561.774 × ⌀0.350.390.3915.12
½2.952.174 × ⌀0.350.390.6319.21
13.352.564 × ⌀0.350.471.0224.65
1½3.943.154 × ⌀0.350.471.529.65
24.333.544 × ⌀0.350.551.9734.53
Ra
30 µin: order code for "Measuring tube material", option BB, TD
max
Ra
15 µin electropolished: order code for "Measuring tube material", option BC, TG
max
A
[in]
B
[in]
C
[in]
D
[in]
E
[in]
Clamp connections
Tri-Clamp
L
[in]
49Engineering unit mm (in)
Length tolerance for dimension L in inch:
+0.06 / –0.08
80Endress+Hauser
A0023342
Page 81
Proline Promass P 300
Tri-Clamp; for pipe according to DIN 11866 series C
1.4435 (316L)
Order code for "Process connection", option FHW
DN
[in]
Clamp
[in]
A
[in]
B
[in]
³⁄₈½0.980.3714.25
½¾0.980.6218.35
111.980.8723.86
1½1 ½1.981.3728.78
222.521.8733.58
Ra
30 µin: order code for "Measuring tube material", option BB, TD
max
Ra
15 µin electropolished: order code for "Measuring tube material", option BC, TG
max
½" Tri-Clamp; for pipe according to DIN 11866 series C
1.4435 (316L)
Order code for "Process connection", option FCW
DN
[in]
Clamp
[in]
A
[in]
B
[in]
½½0.980.3718.35
Ra
30 µin: order code for "Measuring tube material", option BB, TD
max
Ra
15 µin electropolished: order code for "Measuring tube material", option BC, TG
max
L
[in]
L
[in]
¾" Tri-Clamp; for pipe according to DIN 11866 series C
1.4435 (316L)
Order code for "Process connection", option FFW
DN
[in]
Clamp
[in]
A
[in]
B
[in]
³⁄₈¾0.980.6214.25
Ra
30 µin: order code for "Measuring tube material", option BB, TD
max
Ra
15 µin electropolished: order code for "Measuring tube material", option BC, TG
max
1" Tri-Clamp; for pipe according to DIN 11866 series C
1.4435 (316L)
Order code for "Process connection", option FPW
DN
[in]
Clamp
[in]
A
[in]
B
[in]
³⁄₈11.980.8714.25
½11.980.8718.35
Ra
30 µin: order code for "Measuring tube material", option BB, TD
max
Ra
15 µin electropolished: order code for "Measuring tube material", option BC, TG
max
Tri-Clamp eccentric; for pipe according to DIN 11866 series C
1.4435 (316L)
DN
[in]
Order code for
"Process connection",
Clamp
[in]
A
[in]
B
[in]
option
³⁄₈FEB½0.980.3714.25
½FED¾0.980.6218.35
1FEF11.980.8723.86
1½FEH1½1.981.3729.06
L
[in]
L
[in]
L
[in]
Endress+Hauser81
Page 82
Proline Promass P 300
L
X
A
B
X
Tri-Clamp eccentric; for pipe according to DIN 11866 series C
1.4435 (316L)
DN
[in]
2FEK22.521.8733.86
Ra
30 µin: order code for "Measuring tube material", option BB, TD
max
Ra
15 µin electropolished: order code for "Measuring tube material", option BC, TG
max
Additional information on "Eccentric clamps"
Neumo BioConnect clamp connection
Order code for
"Process connection",
option
Clamp
[in]
A
[in]
B
[in]
L
[in]
50Detail X: Asymmetrical process connection; the part shown in blue is provided by the supplier.
Length tolerance for dimension L in inch:
+0.06 / –0.08
Clamp; Neumo BioConnect (sterile orbital), for pipe according to DIN 11866 series A, clamp form R
1.4435 (316L)
Order code for "Process connection", option BSA
DN
[in]
³⁄₈0.980.3914.25
½0.980.6318.35
11.981.0223.86
1½2.521.528.82
23.051.9733.62
Ra
30 µin: order code for "Measuring tube material", option BB, TD
max
Ra
15 µin electropolished: order code for "Measuring tube material", option BC, TG
max
A
[in]
B
[in]
L
[in]
A0016905
82Endress+Hauser
Page 83
Proline Promass P 300
X
L
A
B
X
L
A
B
Neumo BioConnect eccentric clamp connection
A0016906
51Detail X: Asymmetrical process connection; the part shown in blue is provided by the supplier.
Length tolerance for dimension L in inch:
+0.06 / –0.08
Neumo BioConnect eccentric clamp connection, for pipe according to DIN 11866 series C, clamp form R
1.4435 (316L)
Order code for "Process connection", option BEA
DN
[in]
³⁄₈0.980.3914.25
½0.980.6318.35
10.981.0224.02
20.981.9743.39
Ra
30 µin: order code for "Measuring tube material", option BB, TD
max
Ra
15 µin electropolished: order code for "Measuring tube material", option BC, TG
max
Additional information on "Eccentric clamps"
A
[in]
B
[in]
L
[in]
Cable glands
Threaded hygienic connection SMS 1145
A0015628
Length tolerance for dimension L in inch:
+0.06 / –0.08
Endress+Hauser83
Page 84
Threaded hygienic connection SMS 1145
L
45°
A
1
½" NPT
35 (1.38)
1
1
L
A
11
DN 8 ( ")...25 (1")!⁄#
DN 40(1 ")…50 (2")
½
1.4435 (316L)
Order code for "Process connection", option SAW
DN
[in]
³⁄₈Rd 40 × ¹⁄₆0.8914.25
½Rd 40 × ¹⁄₆0.8918.35
1Rd 40 × ¹⁄₆0.8923.86
1½Rd 60 × ¹⁄₆1.429.21
2Rd 70 × ¹⁄₆1.9134.02
Ra
30 µin: order code for "Measuring tube material", option BB, TD
max
Ra
15 µin electropolished: order code for "Measuring tube material", option BC, TG
max
A
[in]
Accessories
Rinse connections
B
[in]
Proline Promass P 300
L
[in]
52
1Connection nipple for purge connections: order code for "Sensor options", option CH "Purge connection"
DNAL
[in][in][in]
³⁄₈1.854.33
½1.858.03
11.8513.7
1½2.68316.46
23.21518.62
84Endress+Hauser
A0029971
Page 85
Proline Promass P 300
146 (5.75)
48 (1.9)
12 (0.47)
280 (11.0)255 (10.0)
134 (5.3)
30 (1.18)
136 (5.35)
114 (4.49)
86 (3.39)
78 (3.07)
!103 (4.06)
!12 (0.47)
7.0 (0.27)
68 (2.7)105 (4.1)
173 (6.8)
Protective cover
A0029553
Remote display and operating module DKX001
53Engineering unit mm (in)
External WLAN antenna
External WLAN antenna mounted on device
54Engineering unit mm (in)
A0028921
The external WLAN antenna is not suitable for use in hygienic applications.
A0028923
Endress+Hauser85
Page 86
Proline Promass P 300
70 (2.8)
1500 (59.1)
72 (2.8)
External WLAN antenna mounted with cable
The external WLAN antenna can be mounted separately from the transmitter if the transmission/
reception conditions at the transmitter mounting location are poor.
A0033597
55Engineering unit mm (in)
Weight
All values (weight exclusive of packaging material) refer to devices with EN/DIN PN 40 flanges.
Weight specifications including transmitter as per order code for "Housing", option A "Aluminum,
coated".
Different values due to different transmitter versions:
• Transmitter version for the hazardous area
(Order code for "Housing", option A "Aluminum, coated"; Ex d): +2 kg (+4.4 lbs)
• Transmitter version for hygienic area
(Order code for "Housing", option B "Stainless, hygienic"): +0.2 kg (+0.44 lbs)
Weight in SI units
DN
[mm]
812
1514
2520
4036
5059
Weight [kg]
Weight in US units
DN
[in]
3/826
½31
144
1½79
2130
Weight [lbs]
MaterialsTransmitter housing
Order code for "Housing":
• Option A "Aluminum, coated": aluminum, AlSi10Mg, coated
• Option B "Stainless, hygienic": stainless steel, 1.4404 (316L)
86Endress+Hauser
Page 87
Proline Promass P 300
1
2
3
4
Window material
Order code for "Housing":
• Option A "Aluminum, coated": glass
• Option B "Stainless, hygienic": polycarbonate
Seals
Order code for "Housing":
Option B "Stainless, hygienic": EPDM and silicone
Cable entries/cable glands
56Possible cable entries/cable glands
1Female thread M20 × 1.5
2Cable gland M20 × 1.5
3Adapter for cable entry with internal thread G ½" or NPT ½"
4Device plugs
Order code for "Housing", option A "Aluminum, coated"
The various cable entries are suitable for hazardous and non-hazardous areas.
Cable entry/cable glandMaterial
Cable gland M20 × 1.5Plastic/nickel-plated brass
Adapter for cable entry with internal thread G ½"Nickel-plated brass
Adapter for cable entry with internal thread NPT ½"
Order code for "Housing", option B "Stainless, hygienic"
The various cable entries are suitable for hazardous and non-hazardous areas.
Cable entry/cable glandMaterial
Cable gland M20 × 1.5Plastic
Adapter for cable entry with internal thread G ½"Nickel-plated brass
Adapter for cable entry with internal thread NPT ½"
• Flanges according to EN 1092-1 (DIN 2501) / according to ASME B16.5 / according to JIS B2220:
Stainless steel, 1.4404 (316/316L)
• All other process connections:
Stainless steel, 1.4435 BN2 (316L)
Available process connections→ 89
Seals
Welded process connections without internal seals
Accessories
Protective cover
Stainless steel, 1.4404 (316L)
External WLAN antenna
• Antenna: ASA plastic (acrylic ester-styrene-acrylonitrile) and nickel-plated brass
• Adapter: Stainless steel and nickel-plated brass
• Cable: Polyethylene
• Plug: Nickel-plated brass
• Angle bracket: Stainless steel
88Endress+Hauser
Page 89
Proline Promass P 300
Process connections
• Fixed flange connections:
– EN 1092-1 (DIN 2501) flange
– EN 1092-1 (DIN 2512N) flange
– ASME B16.5 flange
– JIS B2220 flange
– DIN 11864-2 Form A flange, DIN 11866 series A, flange with notch
– BBS flange small (sterile orbital), DIN 11866 series A, female
– BBS flange small (sterile orbital), DIN 11866 series B, female
• Clamp connections:
– Tri-Clamp (OD tubes), DIN 11866 series C
– DIN 11864-3 Form A clamp, DIN 11866 series A, with notch
– DIN 32676 clamp, DIN 11866 series A
– ISO 2852 clamp, ISO 2037
– ISO 2852 clamp, DIN 11866 series B
– BBS Quick-Connect (sterile orbital), DIN 11866 series A, female
– BBS Quick-Connect (sterile orbital), DIN 11866 series B, female
– Neumo BioConnect clamp, DIN 11866 series A, clamp form R
• Eccentric clamp connection:
– Eccen. Tri-Clamp, DIN 11866 series C
– DIN 11864-3 Form A clamp, DIN 11866 series A, with notch
– DIN 32676 clamp, DIN 11866 series A
– ISO 2852 clamp, DIN 11866 series B
– BBS Quick-Connect (sterile orbital), DIN 11866 series A, female
– BBS Quick-Connect (sterile orbital), DIN 11866 series B, female
– Neumo BioConnect clamp, DIN 11866 series A, clamp form R
• Thread:
– DIN 11851 thread, DIN 11866 series A
– SMS 1145 thread
– ISO 2853 thread, ISO 2037
– DIN 11864-1 Form A thread, DIN 11866 series A
– BBS thread (sterile orbital), DIN 11866 series A
– BBS thread (sterile orbital), DIN 11866 series B
Process connection materials → 88
Surface roughness
Operating concept
All data relate to parts in contact with fluid. The following surface roughness quality can be ordered.
• Ra
= 0.76 µm (30 µin)
max
• Ra
= 0.38 µm (15 µin) electropolished
max
• Delta ferrite <1%
Operability
Operator-oriented menu structure for user-specific tasks
• Commissioning
• Operation
• Diagnostics
• Expert level
Fast and safe commissioning
• Guided menus ("Make-it-run" wizards) for applications
• Menu guidance with brief descriptions of the individual parameter functions
• Device access via Web server or SmartBlue app → 109
• WLAN access to the device via mobile handheld terminal, tablet or smart phone
Reliable operation
• Operation in local language → 90
• Uniform operating philosophy applied to device and operating tools
• If replacing electronic modules, transfer the device configuration via the integrated memory
(HistoROM backup) which contains the process and measuring device data and the event logbook.
No need to reconfigure.
• Troubleshooting measures can be called up via the device and in the operating tools
• Diverse simulation options, logbook for events that occur and optional line recorder functions
Languages
Can be operated in the following languages:
• Via local operation
English, German, French, Spanish, Italian, Dutch, Portuguese, Polish, Russian, Turkish, Chinese,
Japanese, Korean, Bahasa (Indonesian), Vietnamese, Czech, Swedish
• Via Web browser
English, German, French, Spanish, Italian, Dutch, Portuguese, Polish, Russian, Turkish, Chinese,
Japanese, Korean, Bahasa (Indonesian), Vietnamese, Czech, Swedish
• Via "FieldCare", "DeviceCare" operating tool: English, German, French, Spanish, Italian, Chinese,
Japanese
Local operationVia display module
Two display modules are available:
• Order code for "Display; operation", option F "4-line, illuminated, graphic display; touch control"
• Order code for "Display; operation", option G "4-line, illuminated, graphic display; touch control +
WLAN"
Information about WLAN interface → 97
57Operation with touch control
Display elements
• 4-line, illuminated, graphic display
• White background lighting; switches to red in event of device errors
• Format for displaying measured variables and status variables can be individually configured
• Permitted ambient temperature for the display: –20 to +60 °C (–4 to +140 °F)
The readability of the display may be impaired at temperatures outside the temperature range.
Operating elements
• External operation via touch control (3 optical keys) without opening the housing: , ,
• Operating elements also accessible in the various zones of the hazardous area
Via remote display and operating module DKX001
The remote display and operating module DKX001 is available as an optional extra → 107.
• The remote display and operating module DKX001 is only available for the following housing
version: order code for "Housing": option A "Aluminum, coated"
• The measuring device is always supplied with a dummy cover when the remote display and
operating module DKX001 is ordered directly with the measuring device. Display or operation
at the transmitter is not possible in this case.
• If ordered subsequently, the remote display and operating module DKX001 may not be
connected at the same time as the existing measuring device display module. Only one
display or operation unit may be connected to the transmitter at any one time.
A0026785
90Endress+Hauser
Page 91
Proline Promass P 300
A0026786
58Operation via remote display and operating module DKX001
Display and operating elements
The display and operating elements correspond to those of the display module → 90.
Material
The housing material of the display and operating module DKX001 depends on the choice of
transmitter housing material.
Transmitter housingRemote display and operating module
Order code for "Housing"MaterialMaterial
Option A "Aluminum, coated"AlSi10Mg, coatedAlSi10Mg, coated
Cable entry
Corresponds to the choice of transmitter housing, order code for "Electrical connection".
Connecting cable
→ 41
Dimensions
→ 76
Remote operationVia HART protocol
This communication interface is available in device versions with a HART output.
Endress+Hauser91
Page 92
Proline Promass P 300
1235
8
47
6
1
10
2
3
457
69
8
59Options for remote operation via HART protocol (active)
1Control system (e.g. PLC)
2Field Communicator 475
3Computer with Web browser (e.g. Internet Explorer) for access to the integrated device Web server or
computer with an operating tool (e.g. FieldCare, DeviceCare, AMS Device Manager, SIMATIC PDM) with COM
DTM "CDI Communication TCP/IP"
4Commubox FXA195 (USB)
5Field Xpert SFX350 or SFX370
6Field Xpert SMT70
7VIATOR Bluetooth modem with connecting cable
8Transmitter
A0028747
60Options for remote operation via HART protocol (passive)
1Control system (e.g. PLC)
2Transmitter power supply unit, e.g. RN221N (with communication resistor)
3Connection for Commubox FXA195 and Field Communicator 475
4Field Communicator 475
5Computer with Web browser (e.g. Internet Explorer) for access to the integrated device Web server or
computer with an operating tool (e.g. FieldCare, DeviceCare, AMS Device Manager, SIMATIC PDM) with COM
DTM "CDI Communication TCP/IP"
6Commubox FXA195 (USB)
7Field Xpert SFX350 or SFX370
8Field Xpert SMT70
9VIATOR Bluetooth modem with connecting cable
10 Transmitter
Via FOUNDATION Fieldbus network
This communication interface is available in device versions with FOUNDATION Fieldbus.
92Endress+Hauser
A0028746
Page 93
Proline Promass P 300
12
3
1
9898
4
66
5
12
3
444
A0028837
61Options for remote operation via FOUNDATION Fieldbus network
1Automation system
2Computer with FOUNDATION Fieldbus network card
3Industry network
4High Speed Ethernet FF-HSE network
5Segment coupler FF-HSE/FF-H1
6FOUNDATION Fieldbus FF-H1 network
7Power supply FF-H1 network
8T-box
9Measuring device
Via PROFIBUS DP network
This communication interface is available in device versions with PROFIBUS DP.
62Options for remote operation via PROFIBUS DP network
1Automation system
2Computer with PROFIBUS network card
3PROFIBUS DP network
4Measuring device
Via PROFIBUS PA network
This communication interface is available in device versions with PROFIBUS PA.
A0020903
Endress+Hauser93
Page 94
2
3
1
76
5
4
63Options for remote operation via PROFIBUS PA network
12
3
1Automation system
2Computer with PROFIBUS network card
3PROFIBUS DP network
4Segment coupler PROFIBUS DP/PA
5PROFIBUS PA network
6T-box
7Measuring device
Proline Promass P 300
A0028838
Via Modbus RS485 protocol
This communication interface is available in device versions with a Modbus-RS485 output.
A0029437
64Options for remote operation via Modbus-RS485 protocol (active)
1Control system (e.g. PLC)
2Computer with Web browser (e.g. Internet Explorer) for accessing the integrated device Web server or with
operating tool (e.g. FieldCare, DeviceCare) with COM DTM "CDI Communication TCP/IP" or Modbus DTM
3Transmitter
Via EtherNet/IP network
This communication interface is available in device versions with EtherNet/IP.
94Endress+Hauser
Page 95
Proline Promass P 300
4
213
555
4
213
5
55
Star topology
A0032078
65Options for remote operation via EtherNet/IP network: star topology
1Automation system, e.g. "RSLogix" (Rockwell Automation)
2Workstation for measuring device operation: with Custom Add-On Profile for "RSLogix 5000" (Rockwell
Automation) or with Electronic Data Sheet (EDS)
3Computer with Web browser (e.g. Internet Explorer) for accessing the integrated device Web server or
computer with operating tool (e.g. FieldCare, DeviceCare) with COM DTM "CDI Communication TCP/IP"
4Ethernet switch
5Measuring device
Ring topology
The device is integrated via the terminal connection for signal transmission (output 1) and the
service interface (CDI-RJ45).
66Options for remote operation via EtherNet/IP network: ring topology
1Automation system, e.g. "RSLogix" (Rockwell Automation)
2Workstation for measuring device operation: with Custom Add-On Profile for "RSLogix 5000" (Rockwell
Automation) or with Electronic Data Sheet (EDS)
3Computer with Web browser (e.g. Internet Explorer) for accessing the integrated device Web server or
computer with operating tool (e.g. FieldCare, DeviceCare) with COM DTM "CDI Communication TCP/IP"
4Ethernet switch
5Measuring device
A0033725
Via PROFINET network
This communication interface is available in device versions with PROFINET.
Endress+Hauser95
Page 96
Proline Promass P 300
312
444
3
12
444
Star topology
67Options for remote operation via PROFINET network: star topology
1Automation system, e.g. Simatic S7 (Siemens)
2Computer with Web browser (e.g. Internet Explorer) for accessing the integrated device Web server or
computer with operating tool (e.g. FieldCare, DeviceCare, SIMATIC PDM) with COM DTM "CDI Communication
TCP/IP"
3Switch, e.g. Scalance X204 (Siemens)
4Measuring device
A0026545
Ring topology
This communication interface is available in device versions with PROFINET.
68Options for remote operation via PROFINET network: ring topology
1Automation system, e.g. Simatic S7 (Siemens)
2Computer with Web browser (e.g. Internet Explorer) for accessing the integrated device Web server or
computer with operating tool (e.g. FieldCare, DeviceCare, SIMATIC PDM) with COM DTM "CDI Communication
TCP/IP"
3Switch, e.g. Scalance X204 (Siemens)
4Measuring device
A0033719
96Endress+Hauser
Page 97
Proline Promass P 300
2
3
213
4
3
4
567
Service interfaceVia service interface (CDI-RJ45)
A point-to-point connection can be established to configure the device onsite. With the housing
open, the connection is established directly via the service interface (CDI-RJ45) of the device.
An adapter for RJ45 and the M12 connector is optionally available:
Order code for "Accessories", option NB: "Adapter RJ45 M12 (service interface)"
The adapter connects the service interface (CDI-RJ45) to an M12 connector mounted in the
cable entry. Therefore the connection to the service interface can be established via an M12
connector without opening the device.
A0027563
69Connection via service interface (CDI-RJ45)
1Computer with Web browser (e.g. Microsoft Internet Explorer, Microsoft Edge) for accessing the integrated
device Web server or with "FieldCare", "DeviceCare" operating tool with COM DTM "CDI Communication TCP/IP"
or Modbus DTM
2Standard Ethernet connecting cable with RJ45 connector
3Service interface (CDI-RJ45) of the measuring device with access to the integrated Web server
Via WLAN interface
The optional WLAN interface is available on the following device version:
Order code for "Display; operation", option G "4-line, illuminated, graphic display; touch control +
WLAN"
1Transmitter with integrated WLAN antenna
2Transmitter with external WLAN antenna
3LED lit constantly: WLAN reception is enabled on measuring device
4LED flashing: WLAN connection established between operating unit and measuring device
5Computer with WLAN interface and Web browser (e.g. Microsoft Internet Explorer, Microsoft Edge) for
6Mobile handheld terminal with WLAN interface and Web browser (e.g. Microsoft Internet Explorer, Microsoft
7Smart phone or tablet (e.g. Field Xpert SMT70)
Endress+Hauser97
accessing the integrated device Web server or with operating tool (e.g. FieldCare, DeviceCare)
Edge) for accessing the integrated device Web server or operating tool (e.g. FieldCare, DeviceCare)
A0034570
Page 98
Proline Promass P 300
213
555
4
6
7
8
FunctionWLAN: IEEE 802.11 b/g (2.4 GHz)
• Access point with DHCP server (default setting)
• Network
EncryptionWPA2-PSK AES-128 (in accordance with IEEE 802.11i)
Configurable WLAN channels1 to 11
Degree of protectionIP67
Available antennas• Internal antenna
• External antenna (optional)
In the event of poor transmission/reception conditions at the place of
installation.
Available as an accessory → 107.
Only one antenna active in each case!
Max. range50 m (164 ft)
Materials:
External WLAN antenna
• Antenna: ASA plastic (acrylic ester-styrene-acrylonitrile) and nickelplated brass
• Adapter: Stainless steel and nickel-plated brass
• Cable: Polyethylene
• Connector: Nickel-plated brass
• Angle bracket: Stainless steel
Network integration
With the optional OPC-UA-Server application package, the device can be integrated into an Ethernet
network via the service interface (CDI-RJ45 and WLAN) and communicate with OPC-UA clients. If
the device is used in this way, IT security must be considered.
For permanent access to device data and for device configuration via the Web server, the device is
incorporated directly in a network via the service interface (CDI-RJ45). In this way, the device can be
accessed any time from the control station. The measured values are processed separately via the
inputs and outputs through the automation system.
1Automation system, e.g. Simatic S7 (Siemens)
2Ethernet switch
3Edge Gateway
4Cloud
5Measuring device
6Ethernet network
7Measured values via inputs and outputs
8Optional WLAN interface
A0033618
The optional WLAN interface is available on the following device version:
Order code for "Display; operation", option G "4-line, illuminated, graphic display; touch control +
WLAN"
Special Documentation for the OPC-UA-Server application package → 111.
98Endress+Hauser
Page 99
Proline Promass P 300
Supported operating tools
Different operating tools can be used for local or remote access to the measuring device. Depending
on the operating tool used, access is possible with different operating units and via a variety of
interfaces.
Supported operating
tools
Web browserNotebook, PC or tablet
DeviceCare SFE100Notebook, PC or tablet
FieldCare SFE500Notebook, PC or tablet
Device XpertField Xpert SFX
Operating unitInterfaceAdditional information
with Web browser
with Microsoft Windows
system
with Microsoft Windows
system
100/350/370
• CDI-RJ45 service
interface
• WLAN interface
• Ethernet-based
fieldbus (EtherNet/IP,
PROFINET)
• CDI-RJ45 service
interface
• WLAN interface
• Fieldbus protocol
• CDI-RJ45 service
interface
• WLAN interface
• Fieldbus protocol
HART and
FOUNDATION Fieldbus
fieldbus protocol
Special Documentation for
device → 111
→ 109
→ 109
Operating Instructions
BA01202S
Device description files:
Use update function of
handheld terminal
Other operating tools based on FDT technology with a device driver such as DTM/iDTM or
DD/EDD can be used for device operation. These operating tools are available from the
individual manufacturers. Integration into the following operating tools, among others, is
supported:
• FactoryTalk AssetCentre (FTAC) by Rockwell Automation → www.rockwellautomation.com
• Process Device Manager (PDM) by Siemens → www.siemens.com
• Asset Management Solutions (AMS) by Emerson → www.emersonprocess.com
• FieldCommunicator 375/475 by Emerson → www.emersonprocess.com
• Field Device Manager (FDM) by Honeywell → www.honeywellprocess.com
• FieldMate by Yokogawa → www.yokogawa.com
• PACTWare → www.pactware.com
The associated device description files are available at: www.endress.com → Downloads
Web server
Thanks to the integrated Web server, the device can be operated and configured via a Web browser
and via a service interface (CDI-RJ45) or via a WLAN interface. The structure of the operating menu
is the same as for the local display. In addition to the measured values, status information on the
device is also displayed and allows the user to monitor the status of the device. Furthermore the
device data can be managed and the network parameters can be configured.
A device that has a WLAN interface (can be ordered as an option) is required for the WLAN
connection: order code for "Display; operation", option G "4-line, illuminated; touch control + WLAN".
The device acts as an Access Point and enables communication by computer or a mobile handheld
terminal.
Supported functions
Data exchange between the operating unit (such as a notebook for example) and the measuring
device:
– Upload the configuration from the measuring device (XML format, configuration backup)
– Save the configuration to the measuring device (XML format, restore configuration)
– Export event list (.csv file)
– Export parameter settings (.csv file or PDF file, document the measuring point configuration)
– Export the Heartbeat verification log (PDF file, only available with the "Heartbeat Verification"
application package)
Endress+Hauser99
Page 100
Proline Promass P 300
– Flash firmware version for device firmware upgrade, for instance
– Download driver for system integration
– Visualize up to 1000 saved measured values (only available with the Extended HistoROM
application package → 105)
Web server special documentation → 111
HistoROM data management
The measuring device features HistoROM data management. HistoROM data management
comprises both the storage and import/export of key device and process data, making operation and
servicing far more reliable, secure and efficient.
When the device is delivered, the factory settings of the configuration data are stored as a
backup in the device memory. This memory can be overwritten with an updated data record, for
example after commissioning.
Additional information on the data storage concept
There are different types of data storage units in which device data are stored and used by the device:
Device memoryT-DATS-DAT
Available data• Event logbook such as diagnostic events for example
• Parameter data record backup
• Device firmware package
• Driver for system integration for exporting via Web
server, e.g:
– GSD for PROFIBUS DP
– GSD for PROFIBUS PA
– GSDML for PROFINET
– EDS for EtherNet/IP
– DD for FOUNDATION Fieldbus
Storage location Fixed on the user interface board in the connection
compartment
• Measured value logging
("Extended HistoROM" order
option)
• Current parameter data record
(used by firmware at run time)
• Peakhold indicator (min/max
values)
• Totalizer values
Attachable to the user interface
board in the connection
compartment
• Sensor data: nominal diameter
etc.
• Serial number
• Calibration data
• Device configuration (e.g. SW
options, fixed I/O or multi I/O)
In the sensor plug in the transmitter
neck part
Data backup
Automatic
• The most important device data (sensor and transmitter) are automatically saved in the DAT
modules
• If the transmitter or measuring device is replaced: once the T-DAT containing the previous device
data has been exchanged, the new measuring device is ready for operation again immediately
without any errors
• If the sensor is replaced: once the sensor has been replaced, new sensor data are transferred from
the S-DAT in the measuring device and the measuring device is ready for operation again
immediately without any errors
• If exchanging the electronics module (e.g. I/O electronics module): Once the electronics module
has been replaced, the software of the module is compared against the current device firmware.
The module software is upgraded or downgraded where necessary. The electronics module is
available for use immediately afterwards and no compatibility problems occur.
Manual
Additional parameter data record (complete parameter settings) in the integrated device memory
HistoROM backup for:
• Data backup function
Backup and subsequent restoration of a device configuration in the device memory HistoROM
backup
• Data comparison function
Comparison of the current device configuration with the device configuration saved in the device
memory HistoROM backup
100Endress+Hauser
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.