If necessary, Consequences of noncompliance (if applicable)
Corrective action
‣
WARNING
L
Causes (/consequences)
If necessary, Consequences of noncompliance (if applicable)
Corrective action
‣
CAUTION
L
Causes (/consequences)
If necessary, Consequences of noncompliance (if applicable)
Corrective action
‣
NOTICE
Cause/situation
If necessary, Consequences of noncompliance (if applicable)
Action/note
‣
This symbol alerts you to a dangerous situation.
Failure to avoid the dangerous situation will result in a fatal or serious injury.
This symbol alerts you to a dangerous situation.
Failure to avoid the dangerous situation can result in a fatal or serious injury.
This symbol alerts you to a dangerous situation.
Failure to avoid this situation can result in minor or more serious injuries.
This symbol alerts you to situations which may result in damage to property.
1.2 Symbols used
SymbolMeaning
Additional information, tips
Permitted or recommended
Not permitted or not recommended
Reference to device documentation
Reference to page
Reference to graphic
Result of a step
4Endress+Hauser
Memocheck Sim CYP03DBasic safety instructions
2 Basic safety instructions
2.1 Requirements for personnel
• Installation, commissioning, operation and maintenance of the measuring system may be
carried out only by specially trained technical personnel.
• The technical personnel must be authorized by the plant operator to carry out the specified
activities.
• The electrical connection may be performed only by an electrical technician.
• The technical personnel must have read and understood these Operating Instructions and
must follow the instructions contained therein.
• Faults at the measuring point may only be rectified by authorized and specially trained
personnel.
Repairs not described in the Operating Instructions provided must be carried out only
directly at the manufacturer's site or by the service organization.
2.2 Designated use
Memocheck Sim CYP03D is a test tool for analysis measuring points. It enables the simulation
of user-definable measured values and errors of all sensors incorporating Memosens
Technology.
The main areas of application are:
• Chemicals and process engineering
• Food, pharmaceutical industry and biotechnology
• Water and wastewater treatment
• Hazardous areas
Use of the device for any purpose other than that described, poses a threat to the safety of
people and of the entire measuring system and is therefore not permitted.
The manufacturer is not liable for damage caused by improper or non-designated use.
2.3 Workplace safety
As the user, you are responsible for complying with the following safety conditions:
• Installation guidelines
• Local standards and regulations
• Regulations for explosion protection
Electromagnetic compatibility
• The product has been tested for electromagnetic compatibility in accordance with the
applicable international standards for industrial applications.
• The electromagnetic compatibility indicated applies only to a product that has been
connected in accordance with these Operating Instructions.
Endress+Hauser5
Basic safety instructionsMemocheck Sim CYP03D
2.4 Operational safety
Before commissioning the entire measuring point:
1.Verify that all connections are correct.
2.Ensure that electrical cables and hose connections are undamaged.
3.Do not operate damaged products, and protect them against unintentional operation.
4.Label damaged products as defective.
During operation:
If faults cannot be rectified:
‣
products must be taken out of service and protected against unintentional operation.
2.5 Product safety
2.5.1 State-of-the-art technology
The product is designed to meet state-of-the-art safety requirements, has been tested, and
left the factory in a condition in which it is safe to operate. The relevant regulations and
international standards have been observed.
2.5.2 Safety instructions for electrical equipment in hazardous areas
The Memosens inductive sensor cable connection system consists of:
• Memocheck Sim (Memosens technology) CYP03D
• Measuring cable CYK10
Memosens is approved for measuring applications in explosive atmospheres in accordance
with:
• IECEx design approval, IECEx BVS 12.0007
• ATEX design approval BVS 12 ATEX E 008 with amendments
The EC Declaration of Conformity is an integral part of these Operating Instructions.
The Memocheck Sim CYP03D is powered by three alkaline batteries with a total nominal
voltage of 4.5 V.
Use only the following battery types!
‣
Type
ManufacturerEnergizer
TypeEN91
DesignationLR6 (IEC)
U
battery, nominal
Chemical compositionZinc/manganese dioxide (Zn / MnO2)
Do not open the Memocheck Sim CYP03D in hazardous areas.
‣
6Endress+Hauser
1.5 V
Memocheck Sim CYP03DBasic safety instructions
Before commissioning the Memocheck Sim CYP03D, make sure that the battery
‣
compartment is closed securely with a screw.
The maximum permissible cable length is 100 m (330 ft).
‣
Pay attention to the requirements for electrical installations in explosive atmospheres
‣
(EN/IEC 60079-14).
The conductive protective coating on the device is part of the Ex-related safety concept.
‣
Make sure that there is no damage >4 cm2 to the protective coating.
Hazardous area versions of digital sensors and simulators incorporating Memosens
technology are indicated by a red-orange ring on the plug-in head.
IECEx
Connecting the approved digital Memocheck Sim CYP03D sensor simulator to the IECExcertified, intrinsically safe sensor output circuit of the Liquiline M CM42 (IECEx TUR
11.0007X) transmitter or alternatively to an IECEx-certified, intrinsically safe Memosens
sensor output:
Use only the following IECEx-certified measuring cables:
• CYK10-G*** (IECEx BVS 11.0052X)
• or a Memosens measuring cable that has IECEx certification and is identical in terms of
design, appliance technology and function.
The electrical connection must be made according to the wiring diagram.
‣
The transmitter's Memosens input must support the following maximum values. In
‣
particular, the effective inner inductance and the capacitance of the approved, intrinsically
safe sensor output may not exceed these values:
Connecting the approved digital Memocheck Sim CYP03D sensor simulator to the ATEXcertified, intrinsically safe sensor output circuit of the Liquiline M CM42 transmitter or
alternatively to an ATEX-certified, intrinsically safe Memosens sensor output:
Use only the following ATEX-certified measuring cables:
• CYK10-G*** (BVS 04 ATEX E 121 X incl. amendments)
• or a Memosens measuring cable that has ATEX certification and is identical in terms of
design, appliance technology and function.
The electrical connection must be made according to the wiring diagram.
‣
Endress+Hauser7
Basic safety instructionsMemocheck Sim CYP03D
The transmitter's Memosens input must support the following maximum values. In
‣
particular, the effective inner inductance and the capacitance of the approved, intrinsically
safe sensor output may not exceed these values:
Memocheck SimCYP03D-**+**-20 to +50 ˚C (-4 to 122 ˚F)T4
a
Temperature class
If the ambient temperatures shown above are not exceeded no invalid temperatures for the
particular temperature class will occur at the simulator.
CSA
Pay attention to the documentation and control drawings for the transmitter.
‣
2.5.3 IT security
We only provide a warranty if the device is installed and used as described in the Operating
Instructions. The device is equipped with security mechanisms to protect it against any
inadvertent changes to the device settings.
IT security measures in line with operators' security standards and designed to provide
additional protection for the device and device data transfer must be implemented by the
operators themselves.
8Endress+Hauser
Memocheck Sim CYP03DDevice description
3
3 Device description
3.1 Measuring system
Sensor simulation in a complete measuring system consists of:
• Memocheck Sim CYP03D
• Transmitter with Memosens Technology, e. g. Liquiline M CM42 or Liquiline CM44x
• Memosens data cable CYK10
A0025995
1Measuring system with Memocheck Sim CYP03D
1Transmitter Liquiline CM44x
2Memosens data cable CYK10
3Memocheck Sim CYP03D
Endress+Hauser9
Device descriptionMemocheck Sim CYP03D
3.2 Simulation values
With the Memocheck Sim CYP03D you can simulate the following data:
• Simulation values
• Main values
• Raw values
• Temperature
• Parameter
• pH glass (pH glass)
• pH glass, SIL sensor (pH glass SIL)
• pH ISFET (pH Isfet)
• ORP (ORP)
• pH + ORP combined sensor (pH + ORP)
• Conductivity, conductive (Cond c)
• Conductivity, conductive, 4-pin (Cond c 4-pol)
• Conductivity, inductiveCond i)
• Oxygen, amperometric (Oxygen (amp.))
• Oxygen, optical, Memosens (Oxygen (opt.Memo.))
• Oxygen, optical, fixed cable (Oxy. (opt.fixed))
• Chlorine (Chlorine(CCS142D))
• Free chlorine (Free chlorine)
• Chlorine dioxide (Chlorine dioxide)
• Total chlorine (Total chlorine)
• TurbidityTurbidity)
• Nitrate (Nitrate)
• SAC (SAC)
• The main simulation values can be selected as required within the context of the sensor
specification values
• Repeated ramp with any increment
• Error, e. g. glass breakage, alarm and warning
• Calibration values
You can freely configure all the values so that they match your process. The data listed above
are displayed on the transmitters.
10Endress+Hauser
Memocheck Sim CYP03DIncoming acceptance and product identification
4 Incoming acceptance and product identification
4.1 Incoming acceptance
1.Verify that the packaging is undamaged.
Notify the supplier of any damage to the packaging.
Keep the damaged packaging until the issue has been resolved.
2.Verify that the contents are undamaged.
Notify the supplier of any damage to the delivery contents.
Keep the damaged goods until the issue has been resolved.
3.Check that the delivery is complete and nothing is missing.
Compare the shipping documents with your order.
4.Pack the product for storage and transportation in such a way that it is protected
against impact and moisture.
The original packaging offers the best protection.
Make sure to comply with the permitted ambient conditions.
If you have any questions, please contact your supplier or your local Sales Center.
4.2 Product identification
4.2.1 Nameplate
The nameplate provides you with the following information on your device:
• Manufacturer identification
• Order code
• Extended order code
• Serial number
• Ambient and process conditions
• Input and output values
• Safety information and warnings
Compare the information on the nameplate with the order.
‣
4.2.2 Product page
www.endress.com/cyp03d
4.2.3 Interpreting the order code
The order code and serial number of your product can be found in the following locations:
• On the nameplate
• In the delivery papers
Obtaining information on the product
1.Go to www.endress.com.
2.Call up the site search (magnifying glass).
Endress+Hauser11
Incoming acceptance and product identificationMemocheck Sim CYP03D
3.Enter a valid serial number.
4.Search.
The product structure is displayed in a popup window.
5.Click on the product image in the popup window.
A new window (Device Viewer) opens. All of the information relating to your
device is displayed in this window as well as the product documentation.
4.2.4 Manufacturer address
Endress+Hauser Conducta GmbH+Co. KG
Dieselstraße 24
D-70839 Gerlingen
4.2.5 Scope of delivery
Memocheck Sim CYP03D
• Memocheck Sim CYP03D
• Operating Instructions
• 1 quality certificate as ordered
• Cable as per order (optional)
• Case to store CYP03D and cable (optional)
• Certificate of calibration (optional)
4.3 Certificates and approvals
4.3.1 mark
The product meets the requirements of the harmonized European standards. As such, it
complies with the legal specifications of the EU directives. The manufacturer confirms
successful testing of the product by affixing to it the mark.
4.3.2 Ex approvals
Memocheck Sim CYP03D
• ATEX II 2G Ex ia IIC T4 Gb
• IECEx Ex ia IIC T4 Gb
• CSA IS NI Cl. I, Div. 1&2, Group A-D
12Endress+Hauser
Memocheck Sim CYP03DElectrical connection
II 2G
II 1G
II 3G
Ex ia
Zone 1
Zone 0
Zone 2
CYK10
CYP03D
not permitted
only with ATEXApproval
GN/YE
YE
GN
BN
WH
GND
+
–
Com A
Com B
GY
5 Electrical connection
5.1 Connection in hazardous areas
A0026198-EN
2Operation in hazardous areas
5.2 Connecting the simulator
The electrical connection of the sensor's to the transmitter is established using measuring
cable CYK10.
3Measuring cable CYK10
Endress+Hauser13
A0024019
Operation optionsMemocheck Sim CYP03D
2
3
4
5
6
6 Operation options
6.1 Overview
6.1.1 Display
1
Menu path and/or device designation
2
Battery status
3
Simulated parameter
4
Assignment of soft keys, e.g. menu
5
Assignment of soft keys, e.g. and
6
Sim. main value
A0026099-EN
4Display (e.g. simulation mode)
6.1.2 Key functions
ENTER key
• Switches on the device (press for at least 3 seconds)
• Calls up menu when in simulation mode
• Saves (confirms) data entered
• Selects a menu option
• Switches off the device (press for at least 3 seconds)
or MINUS key or PLUS key
In setup mode, the MINUS and PLUS keys have the following
functions:
• Configuration of parameters and numerical values
• Navigation through menu
In simulation mode, the MINUS and PLUS keys have the following
functions:
"Run through" of sim. main values, with each values being changed by
the amount of the delta value
and Escape function
Press MINUS and PLUS key simultaneously
• Pressing for a short time: takes you up one level in the menu.
• Pressing for a longer time in the main menu: takes you directly to
the simulation mode.
14Endress+Hauser
Memocheck Sim CYP03DOperation options
and Reset
Press MINUS and ENTER key simultaneously for longer
Saved setups remain intact.
and Factory settings
Press PLUS and ENTER key simultaneously for longer
• This resets the device to the factory settings.
• All the setups saved are deleted.
Endress+Hauser15
Operation optionsMemocheck Sim CYP03D
pH glass
pH Isfet
ORP
Cond c
Oxygen (amp.)
Chlorine (CCS142D)
pH glass SIL
Cond i
Oxy. (opt.fixed)
Turbidity
Nitrate
SAC
Language
Display
Setup
Illumination
Contrast
Simulation
Parameter
Memosens
Fixed cable
Sim. main value
Test values
Save setup
Load setup
Advanced
Date
Time
Switch off auto.
Advanc. function
Calibration value
Error simulation
Ramp (continuous)
Diagnostics
Diagnostic list
Order code
Serial number
Software
Hardware
Reset
Factory settings
Switch off
pH + ORP
Cond c 4-pol
Oxygen (opt.Memo.)
Ramp (gradual)
Sensor operation
Cap operation
No. of cap cal
Sensor state
Operating time
*
Charge *
Operating time *
Zero point *
Slope *
Free chlorine
Chlorine dioxide
Total chlorine
6.2 Structure and function of the operating menu
6.2.1 Menu structure
*only Free chlorine, Chlorine dioxide, Total chlorine
16Endress+Hauser
Memocheck Sim CYP03DOperation options
-
-
-
6.2.2 Operating concept
Example of changing values within a menu: defining the start value of a ramp
Maximum and minimum values are displayed in the editor. You can only configure values
within these limits.
1.Use the arrow keys to select the digit of the value that you want to change.
2.Press to change the value.
The digit flashes
3.Press or to increase or decrease the value.
4.Confirm entry with .
The "Escape" function ( and pressed simultaneously) is disabled here in order
to prevent incorrect entries.
5.Select the check mark (arrow key) and press .
The edited value is accepted.
Endress+Hauser17
Loading...
+ 39 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.