• Increased plant safety thanks to complete sensor and error simulation
• Easy, quick and reliable sensor simulation with user-definable measured values for
CYP03D and fixed values for CYP02D
• Error simulation
• Verification of all parameters: pH (glass and ISFET), ORP, conductivity (conductive
and toroidal), dissolved oxygen (amperometric and optical), chlorine, turbidity and
nitrate
• High degree of flexibility when commissioning the measuring point
• Utmost reliability owing to Memosens Technology:
– No contact corrosion
– Completely watertight
– No interference from external potential thanks to galvanic isolation of simulator
and transmitter
Page 2
Function and system design
3
SIMU
1
2
3
OK
Memocheck CYP02D / Memocheck Sim CYP03D
Simulation setup in nonhazardous areas
Sensor simulation in a complete measuring system consists of:
• Memocheck CYP02D or Memocheck Sim CYP03D
• Transmitter with Memosens Technology, e. g. Liquiline M CM42 or Liquiline CM44x
• Memosens data cable CYK10
A0025995
1Measuring system with Memocheck Sim CYP03D
1Transmitter Liquiline CM44x
2Memosens data cable CYK10
3Memocheck Sim CYP03D
2Measuring system with Memocheck CYP02D
1Liquiline M CM42 transmitter
2Memosens data cable CYK10
3Memocheck CYP02D
A0025994
2Endress+Hauser
Page 3
Memocheck CYP02D / Memocheck Sim CYP03D
II 2G
II 1G
II 3G
Ex ia
Zone 1
Zone 0
Zone 2
CYK10
CYP03D
not permitted
only with ATEXApproval
Simulation setup in
hazardous areas
Sensor simulation in a complete measuring system consists of:
• Memocheck Sim CYP03D
• Transmitter with Memosens Technology and Ex approval, e. g. Liquiline M CM42
• Memosens data cable CYK10
A0026198-EN
3Operation in hazardous areas
Endress+Hauser3
Page 4
Performance characteristics
Memocheck CYP02D / Memocheck Sim CYP03D
Memocheck CYP02D
simulation values
Memocheck CYP02D consists of two interconnected sensor plug-in heads which are independent of
one another, and which each supply one specific parameter. You can simulate the following values
depending on your order:
Simulation values output for pH glass and pH ISFET
Fixed simulation
value
pH glass425 ˚C (77 ˚F)
pH ISFET760 ˚C (140 ˚F)
Temperature
Simulation values output for pH glass and pH glass
Fixed simulation
value
pH glass425 ˚C (77 ˚F)
pH glass760 ˚C (140 ˚F)
Temperature
Simulation values output for pH glass and ORP
Fixed simulation
value
pH glass425 ˚C (77 ˚F)
ORP potential750 mV90 ˚C (194 ˚F)
Temperature
Simulation values output for pH glass and oxygen
Fixed simulation
value
pH glass425 ˚C (77 ˚F)
Oxygen60 nA
1)Sensor current
1)
Temperature
60 ˚C (140 ˚F)
Simulation values output for pH glass and conductive measurement of conductivity
Fixed simulation
value
pH glass425 ˚C (77 ˚F)
Conductivity20 µS/cm60 ˚C (140 ˚F)
Temperature
Simulation values output for pH glass and chlorine
Fixed simulation
value
pH glass425 ˚C (77 ˚F)
Chlorine60 nA
1)Sensor current
1)
Temperature
25 ˚C (77 ˚F)
The data listed above are displayed on the transmitters.
Memocheck CYP02D is maintenance-free.
4Endress+Hauser
Page 5
Memocheck CYP02D / Memocheck Sim CYP03D
Memocheck Sim CYP03D
simulation values
With the Memocheck Sim CYP03D you can simulate the following data:
• Simulation values
– Main values
– Raw values
– Temperature
• The main simulation values can be selected as required within the context of the sensor
specification values
• Repeated ramp with any increment
• Error, e. g. glass breakage, alarm and warning
• Calibration values
You can freely configure all the values so that they match your process. The data listed above are
displayed on the transmitters.
The Memocheck Sim CYP03D can, with the quality or calibration certificate, also be used as a
qualification tool for your measuring point.
The quality and calibration certificates can be renewed:
For such services, you will need to return Memocheck Sim CYP03D to Endress+Hauser.
In the case of requalification, the device is tested fully and a new quality certificate is issued.
In the case of recalibration, in addition to requalification the device is also incorporated into a
calibration procedure. A quality certificate and a certificate of calibration are issued. The
recommended testing interval is 1 year.
Environment
Ambient temperature rangeMemocheck CYP02D
–15 to 70 °C (5 to 158 °F)
Memocheck Sim CYP03D
–20 to 50 °C (–4 to 120 °F)
Storage temperatureMemocheck CYP02D
–15 to 70 °C (5 to 158 °F)
Memocheck Sim CYP03D
-20 to 55 °C (-4 to 130 °F)
Degree of protectionMemocheck CYP02D
IP65
Memocheck Sim CYP03D
IP55
Endress+Hauser5
Page 6
Memocheck CYP02D / Memocheck Sim CYP03D
89 (3.5)
25.8 (1.02)
25.8 (1.02)
37.4 (1.47)
Ø 20 (0.79)
40.0 (1.57)
25.0 (0.98)
58.0 (2.28)
94.0 (3.7)
36.0 (1.42)
200.0 (7.87)
Electromagnetic
compatibility
Dimensions
Interference emission and interference immunity as per EN 61326-1:2013, Class A for Industry
Mechanical construction
A0026004
4Memocheck CYP02D
5Memocheck Sim CYP03D
All dimensions in mm (in)
6Endress+Hauser
A0026005
Page 7
Memocheck CYP02D / Memocheck Sim CYP03D
Weight (CYP03D only, incl.
batteries)
Materials
Batteries
mark
0.3 kg (0.7 lbs)
Memocheck CYP02D:
White plastic:PET
Blue plastic:PPS GF 40
Memocheck Sim CYP03D:
Housing:ABS (UL 94 HB)
No batteries are required for the Memocheck CYP02D.
For the Memocheck Sim CYP03D, use only the following battery types, as only these are covered by
the Ex approval:
• Energizer, EN91 (AA, 1.5 V, LR6 as per IEC), x 3
• Battery storage temperature: -20 to 35 ˚C (-4 to 95 ˚F)
Certificates and approvals
The product meets the requirements of the harmonized European standards. As such, it complies
with the legal specifications of the EU directives. The manufacturer confirms successful testing of the
product by affixing to it the mark.
Ex approvals
Memocheck Plus CYP02D
• ATEX/IECEx II 2G Ex ia IIC T6 Gb
• FM/CSA IS NI Cl.I Div.1&2, Group A-D
Memocheck Sim CYP03D
• ATEX II 2G Ex ia IIC T4 Gb
• IECEx Ex ia IIC T4 Gb
• CSA IS NI Cl. I, Div. 1&2, Group A-D
Endress+Hauser7
Page 8
Ordering information
Memocheck CYP02D / Memocheck Sim CYP03D
Product page
Product Configurator
Scope of delivery
www.endress.com/cyp02d
www.endress.com/cyp03d
On the product page there is a Configure button to the right of the product image.
1.Click this button.
The Configurator opens in a separate window.
2.Select all the options to configure the device in line with your requirements.
In this way, you receive a valid and complete order code for the device.
3.Export the order code as a PDF or Excel file. To do so, click the appropriate button on the right
above the selection window.
For many products you also have the option of downloading CAD or 2D drawings of the
selected product version. Click the CAD tab for this and select the desired file type using
picklists.
Memocheck Plus CYP02D
• 1 Memocheck CYP02D with 2 plug-in heads
• 1 Operating Instructions (de + en)
Memocheck Sim CYP03D
• Memocheck Sim CYP03D
• Operating Instructions
• 1 quality certificate as ordered
• Cable as per order (optional)
• Case to store CYP03D and cable (optional)
• Certificate of calibration (optional)
8Endress+Hauser
Page 9
Memocheck CYP02D / Memocheck Sim CYP03D
Accessories
The following are the most important accessories available at the time this documentation was
issued.
For accessories not listed here, please contact your Service or Sales Center.
‣
Memosens data cable
Storage case
Order No.Memosens data cable CYK10 (optional)
71128718CYK10-A032 + adapter, cable ends; Non-Ex
71128721CYK10-G032 + adapter; only for CYP03D, Ex
For the Memocheck CYP02D, you can use the Memosens data cable CYK10 belonging to the
measuring point.
To connect the Memocheck Sim CYP03D to transmitters with M12 sockets and Pg couplings, you
require the Memosens data cable CYK10 provided. The cable is always supplied with an adapter
piece so that it fits both M12 sockets and Pg couplings. If you wish to simulate fixed cable sensors
(turbidity, nitrate, toroidal conductivity, oxygen optical) with the Memocheck Sim CYP03D, you need
this cable. When using sensors with an inductive Memosens plug-in head (pH/ORP, oxygen,
conductive conductivity, chlorine), the appropriate cable is already included in the measuring point.
Order No.
71183327Ex
In the hazardous area, the Memocheck simulator case should only be opened to remove or put back
the Memocheck simulator. When opened, the case should never be exposed to process-related
intensive electrostatic charges.
Case for Memocheck Sim CYP03D
Endress+Hauser9
Page 10
Page 11
Page 12
www.addresses.endress.com
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.