Endress+Hauser CYM291 Operating Manual

Page 1
Products ServicesSolutions
Operating Instructions
Liquiline To Go Ex CYM291
Page 2
Page 3
Basics
Warranty
Defects occurring within 3 years from delivery date shall be remedied free of charge at our plant (carriage and insurance paid by sender). Subject to change.
Return of Products Under Warranty
Please contact our Service Team before returning a defective device. Ship the cleaned device to the address you have been given. If the device has been in contact with process uids, it must be decontaminated/ disinfected before shipment. In that case, please attach a corresponding certicate, for the health and safety of our service personnel.
Disposal
Please observe the applicable local or national regulations concerning the disposal of “waste electrical and electronic equipment”.
3
Registered Trademarks
The following names are registered trademarks. For practical reasons they are shown without trademark symbol in this manual.
• Memosens®
• Liquiline®
• Sensocheck®
• Sensoface®
Page 4
4
Package Contents ............................................................................... 6
Documentation ................................................................................... 7
Overview ............................................................................................. 8
Value-Added Features ..............................................................................................9
Protective Cover ....................................................................................................... 10
Hook ............................................................................................................................10
Display ........................................................................................................................ 11
Keypad ........................................................................................................................12
Start-Up ............................................................................................. 13
Inserting the Batteries ........................................................................................... 13
Batteries for Application in Hazardous Locations........................................ 14
Connecting a Sensor .............................................................................................. 15
Switching On the Meter ........................................................................................ 16
Icons ............................................................................................................................ 16
Conguring .......................................................................................17
Conguration (pH) .................................................................................................. 17
Conguration (Cond) ............................................................................................. 18
Conguration (Oxy) ................................................................................................ 19
pH Calibration ................................................................................... 20
Conductivity Calibration .................................................................. 24
Oxygen Calibration ..........................................................................27
Measuring .........................................................................................31
Toggling the Measured Value Display .............................................................31
Adjusting the Temperature .................................................................................. 31
Data Logger ......................................................................................32
Operating Modes of the Data Logger (Logger Type) .................................33
Data Logger Menu .................................................................................................. 35
Conguring the Data Logger .............................................................................. 36
Starting the Data Logger using CONT ............................................................. 37
Starting the Data Logger using START ............................................................ 37
Displaying the Logger Data ................................................................................. 38
Stopping the Data Logger ...................................................................................39
Clearing the Data Logger ..................................................................................... 39
Table of Contents
Page 5
Table of Contents
Clock ..................................................................................................40
Error and Status Messages ...............................................................41
”Sensoface” Messages ............................................................................................ 42
Error Messages ......................................................................................................... 43
pH Product Line ................................................................................ 44
pH Sensors ................................................................................................................. 44
Endress+Hauser Buer Solutions (pH) ...........................................................45
Ready-to-use quality pH buer solutions.......................................................45
Accessories for pH ................................................................................................... 45
Conductivity Product Line ............................................................... 46
Conductivity Sensors ............................................................................................. 46
CLY11 Conductivity Calibration Solutions......................................................47
Accessories for Conductivity ............................................................................... 47
Oxygen Product Line ........................................................................ 48
Oxygen Sensors ....................................................................................................... 48
Accessories for Oxygen ......................................................................................... 48
Specications .................................................................................... 49
Index .................................................................................................. 54
5
Page 6
6
Check the shipment for transport damage and completeness. The package of the Liquiline To Go Ex CYM291 includes:
Meter incl. 4 AA batteries and premounted quiver
Carrying strap
Data carrier with detailed user manuals
USB cable, 1.5 m
Safety instructions
Quickstart instructions in various languages
Certicates

Package Contents

Liquiline To Go Ex CYM291
Page 7

Documentation

Specic Test Report
CD-ROM
Complete documentation:
• User manuals in dierent languages
• Safety instructions
• Certicates
• Quickstart guides
Safety Instructions
In ocial EU languages and others.
• EU Declarations of Conformity
7
CAUTION!
These safety instructions are part of the product documentation and must be observed.
Quickstart Guides
Installation and rst steps:
• Operation
• Menu structure
• Calibration
• Error messages and recommended actions
Various languages on CD-ROM:
• German
• English
• French
• Italian
• Spanish
• Portuguese (Brazil)
Certicates
Page 8
8

Overview

The Liquiline To Go Ex CYM291 is
a portable multiparameter meter for measuring pH, ORP, conductivity or oxygen. With a plain text line on a high-contrast LCD, operation is largely intuitive.
Quiver
The meter stands out by the following features:
• Application in hazardous locations up to Zone 0
• Use of digital Memosens sensors
• A detachable quiver protects the sensor and prevents it from drying out. Furthermore, it can be used for calibration.
• The rugged housing is made of a high-performance polymer. It provides high impact resistance and dimensional stability even when exposed to extreme moisture.
• Scratch-proof clear glass display, perfectly readable even after years
• Very long operating time with one set of batteries (4 x AA) for reliable operation even at high or very low operating temperatures
• Data logger with 5000 values
• Micro USB port
• Sensoface icons provide single-glance information on the sensor condition (page9)
• Real-time clock and indication of battery charging level
Page 9
Overview

Value-Added Features

Memosens
The Liquiline To Go Ex CYM291 can communicate with Memosens sensors. These digital sensors are automatically identied and the meter switches to the appropriate measurement method. When a Memosens sensor is connected to the meter, it is indicated by the logo shown on the right. Furthermore, Memosens allows the storage of calibration data, which will be available and can still be used when the sensor is connected to another Memosens-capable device.
Sensoface
Sensoface provides quick information on the sensor condition. The three “smiley” faces as shown on the right represent the sensor condition during measurement and after a calibration. When the condition deteriorates, an INFO …” message gives a hint to the cause.
9
Programmed buers
“Programmed buers” is a very convenient method for pH calibration with automatic buer recogni­tion. You only have to select the buer set with the buers used. The buers can then be used in any order.
Page 10
10
Overview

Protective Cover

The front of the meter is protected by a cover, which can be completely ipped over and secured to the back for operation.

Hook

A fold-out hook on the back allows suspending the meter. This leaves your hands free for the actual measurement. The rating plate is located beneath the hook.
Protective Cover and Hook Combined
Cover and hook can be joined together to form a benchtop stand allowing com­fortable and fatigue-free working at a lab bench or desk.
Page 11
Overview

Display

The meter has a three-line display for representing alphanumeric informa­tion such as measurement and calibra­tion data, temperatures and date/time. Additional information is provided by means of icons (Sensoface, battery icon, etc.). Some typical displays are shown here.
11
pH measurement (display of measured value, tempera­ture, date and time)
Oxygen calibration – step 1 (calibration in air)
Logger data for conductivity (display of measured value, memory location, temperature, date and time)
Oxygen calibration – step 2 (adjusting the relative humidity)
Clock (display of hours and minutes, seconds and date)
Page 12
12
Overview

Keypad

The keys of the membrane keypad have a noticeable pressure point.
They have the following functions:
Switches the meter on and displays the device and calibration data
(see Start-up)
Switches the meter on / Activates measuring mode / Stops the data logger
Starts calibration
Activates conguration / Conrms entries
Displays time and date, allows setting the clock using set
View stored values
Holds and saves a measured value, allows setting and starting the logger by press­ing (page 32)
When this icon is displayed,
▲ ▼
you can use the arrow keys for navigation.
Page 13

Start-Up

Check the shipment for transport damage and completeness (see Package Contents).
NOTICE!
Do not operate the device when one of the following conditions applies:
• the device shows visible damage
• the device fails to perform the intended function
• prolonged storage at temperatures above 70 °C
• severe transport stresses In this case, a professional routine test must be performed. This test should be carried out at our factory.
Precautions for application in hazardous locations
WARNING!
• Only open the battery compartment of the Liquiline To Go Ex CYM291 outside the hazardous location.
• Never try to open the device. If a repair should be required, return the device to our factory.
• Never use the USB port within the hazardous location.
13

Inserting the Batteries

With four AA batteries, the Liquiline ToGo Ex CYM291 has an operating time of approx. 500h. Open the battery compartment on the rear of the device. Be sure to observe the correct polarity when inserting the batter‑ ies (see markings in the battery chamber). Close the battery compartment cover and screw it handtight.
Page 14
14
A battery icon in the display indicates the battery power level:
Icon fully lled Batteries at full capacity
Icon partially lled Battery capacity is sucient
Icon empty Battery capacity not sucient;
calibration is possible, no logging
Icon blinks Only a few operating hours remaining,
measurement is still possible NOTICE! It is absolutely necessary to replace the batteries.
WARNING!
When operating the Liquiline To Go Ex CYM291 in a hazardous loca‑ tion, only the battery types listed below may be used. The batteries must be from the same manufacturer and of identical type and capacity. Never use new and used batteries together (“Certicates”, Control Drawing 209.009‑150).
Start-Up

Batteries for Application in Hazardous Locations

Batteries (4x each) Temp. class Ambient temperature range
Duracell MN1500 T4 –10 °C ≤ Ta ≤ +40 °C
Energizer E91 T3 –10 °C ≤ Ta ≤ +50 °C
Power One 4106 T3 –10 °C ≤ Ta ≤ +50 °C
Panasonic Pro Power LR6 T3 –10 °C ≤ Ta ≤ +50 °C
Page 15
Start-Up
15

Connecting a Sensor

The Liquiline To Go Ex CYM291 provides several connections so that many types of sensors can be used for measurement (see illustration below). Note that only one sensor may be connected to the meter at a time. The meter recognizes the connect­ed Memosens sensor and displays the Memosens logo.
Separate temperature probe
After power-on, a separate temperature probe is automatically recognized. When you want to replace the temperature probe, you must switch o the meter and then switch it on again.
a b c d
Memosens sensors have a cable coupling, which allows convenient replacement of sensors while the cable remains connected to the meter. The connecting cable is connected to socket b (M8, 4 pins) or e (M12, 8 pins).
e
Connections
a - Micro USB port b - M8, 4 pins, for Memosens lab cable c - Temperature probe – GND d - Temperature probe e - M12, 8 pins, for Memosens sensors
Page 16
16

Switching On the Meter

When you have connected the sensor, you can switch the meter on by pressing the or key. When the meter is switched on with the key, rst a self test is performed and then the calibration data and settings are displayed before the meter switches to measuring mode. When the meter is switched on with the key, it immediately switches to measuring mode. Depending on the connected sensor and the specic measur­ing task, several steps for conguration and calibration must be performed as described on the following pages.

Icons

Important information about the state of the device:
Start-Up
Sensor condition
Error message
Memosens sensor
Date
Battery
charging level
Measured value
Measured
variable
Temperature
detection
(sensor,
separate or
manual)
Time
Page 17
Conguring
Conguration (pH)
Conguration is required to match the connected sensor and the desired measurement performance. Furthermore, you can select the suitable calibration method. The following table gives you an over­view. Factory settings are shown in bold print.
Measurement
17
pH
“Setup” display
Display 1 pH x.xx | pH x.xxx | mV Display 2 OFF | date + time | date | time CAL Timer OFF | 1 ... 99 days
CAL
CAL POINTS 1 | 2 | 3 | 1-2-3 (for PROG.BUFFERS, Manual, FREE CAL)
BUFFER SET (PROG.BUF­FERS, FREE CAL)
Auto OFF OFF| 0.1h | 1h | 6h | 12h Temp Unit °C | °F Time Format 24h | 12h Date Format dd.mm.yy | mm.dd.yy
Default
Select using arrow keys, conrm by pressing .
PROG.BUFFERS|Manual|DATA INPUT|(ISFET-Zero)| FREE CAL
-01- Endress+Hauser 2.00|4.01|6.98|9.95|11.87
-02- Mettler-Toledo 2.00|4.01|7.00|9.21
-03- Knick CaliMat 2.00|4.00|7.00|9.00|12.00
-04- Ciba (94) 2.06|4.00|7.00|10.00
-05- NIST technical 1.68|4.00|7.00|10.01|12.46
-06- NIST standard 1.679|4.006|6.865|9.180
-07- HACH 4.01|7.00|10.01|12.00
-08- WTW techn. buers 2.00|4.01|7.00|10.00
-09- Hamilton 2.00|4.01|7.00|10.01|12.00
-10- Reagecon 2.00|4.00|7.00|9.00|12.00
-11- DIN 19267 1.09|4.65|6.79|9.23|12.75
NO | YES (reset to factory settings) Note: All data logger entries will be deleted.
This icon prompts you to select a menu item using the arrow keys – the selection is conrmed by pressing .
Page 18
18
Cond
Measurement
Conguration (Cond)
Conguration is required to match the connected sensor and the desired measurement performance. Furthermore, you can select the suitable calibration method. The following table gives you an over­view. Factory settings are shown in bold print.
“Setup” display
Display Cond | SAL g/kg | TDS mg/l
MOHM cm OFF | On
Cond Unit mS/cm | S/m
TDS Factor 0.0 … 1.0 (if display = TDS)
TC * OFF | LINEAR | NLF | NACL | HCL | NH3 | NAOH
TC LINEAR 0.0 … 20.0 %/K | 2.0 %/K (if TC = LINEAR)
REF. Temp. 0 … 100 °C | 25 °C ( 32 … 212 °F | 77 °F)
s t
CAL CELL CONST. | COND | <25°C> 74.0µS/cm |
Auto OFF OFF | 0.1h | 1h | 6h | 12h
Temp. Unit °C | °F
Time Format 24h | 12h
Date Format dd.mm.yy | mm.dd.yy
Default NO | YES (reset to factory settings)
Select using arrow keys, conrm by pressing .
(if display = Cond)
(if TC = LINEAR)
<25°C> 149.6µS/cm | <25°C> 1.406mS/cm | <25°C> 12.64mS/cm | <25°C> 107.00mS/cm |FREE CAL
Note: All data logger entries will be deleted.
s
This icon prompts you to select a menu item using the arrow keys –
t
the selection is conrmed by pressing
* Temperature compensation
.
Page 19
Conguring
Conguration (Oxy)
Conguration is required to match the connected sensor and the desired measurement performance. Furthermore, you can select the suitable calibration method. The following table gives you an over­view. Factory settings are shown in bold print.
Measurement
19
Oxy
“Setup” display
Display 1 Saturation in % air | Concentration in mg/l
Display 2 OFF | date + time | date | time Altitude 0 … 4000 m Salt Correct 0.0 … 45.0 g/kg CAL AIR CAL | ZERO CAL | DATA INPUT | FREE CAL CAL Timer OFF | 1 … 99 days Auto OFF OFF | 0.1h | 1h | 6h | 12h Temp Unit °C | °F Time Format 24h | 12h Date Format dd.mm.yy | mm.dd.yy Default NO | YES (reset to factory settings)
This icon prompts you to select a menu item using the arrow keys – the selection is conrmed by pressing .
Select using arrow keys, conrm by pressing .
Note: All data logger entries will be deleted.
Page 20
20
pH
Measurement

pH Calibration

“Programmed Buers” Calibration
(Calibration with automatic buer recognition)
The calibration method is selected in the conguration menu.
Calibration is required to adjust the sensor to the meter. It is indis‑
pensable for achieving comparable and reproducible measurement
results.
CAL PROG. BUFFERS
CAL 1/2/3 PRESS CAL
pH, mV value,
Calibration method, the number of calibration points and the buer set have been selected in the conguration menu.
Dip sensor in 1st/2nd/3rd buer solution. It does not matter which buer solution is taken rst.
temperature
Depending on the number of calibration points, the procedure described above for CAL 1/2/3 is repeated.
mV value blinks until calibration is completed, then successive display of: CAL DATA 1/2/3 CAL POINTS ZERO POINT SLOPE Then the meter switches to measuring mode.
Please note: To abort calibration, you can press at any time. This will be conrmed by the display message “CAL ABORTED”. Exception: When you have selected “CAL POINTS 1‑2‑3” and the rst calibration step has been completed, the calibration process cannot be stopped any more.
Page 21
pH Calibration
DATA INPUT Calibration
(Calibration by entering known sensor values)
The calibration method is selected in the conguration menu.
Measurement
CAL DATA INPUT
21
pH
ZERO POINT
SLOPE
The calibration data will be displayed successively: Date and time ZERO POINT SLOPE Then the meter switches to measuring mode.
Please note: To abort calibration, you can press at any time.
st to select the value for the zero point.
Use
st to select the value for the slope.
Use
Page 22
22
pH
Measurement
pH Calibration
MANUAL Calibration
(Manual calibration)
The calibration method is selected in the conguration menu.
CAL MANUAL
CAL 1/2/3 PRESS CAL
pH display blinks PRESS CAL
mV display blinks
mV value blinks until calibration is completed, then successive display of: CAL DATA 1/2/3 CAL POINTS ZERO POINT SLOPE Then the meter switches to measuring mode.
The number of calibration points has been selected in the conguration menu.
Take the temperature‑corrected pH value from the buer description and enter it using
Depending on the number of calibration points, the procedure described above for CAL 1/2/3 is repeated.
st.
Please note: To abort calibration, you can press This will be conrmed by the display message “CAL ABORTED”. Exception: When you have selected “CAL POINTS 1‑2‑3” and the rst calibration step has been completed, the calibration process cannot be stopped any more.
at any time.
Page 23
Measurement
23
pH
FREE CAL Calibration
(Free selection of calibration method)
FREE CAL calibration is selected in the conguration menu.
CAL PROG. BUFFERS blinks
Perform the selected calibration (see PROG. BUFFERS, DATA INPUT or MANUAL calibration).
st to select the required calibration
Use method (PROG. BUFFERS, DATA INPUT or MANUAL).
Page 24
24
Cond
Measurement

Conductivity Calibration

CELL CONST. Calibration
(Calibration by entry of cell constant)
The calibration method is selected in the conguration menu.
CAL CELL CONST.
Value blinks
Calibration will be performed. Automatic return to measuring mode.
The conductivity will be shown in the display and can be compared with a reference solution (temperature‑corrected).
st to select the value for the cell constant.
Use
COND Calibration
(Calibration by entry of conductivity)
The calibration method is selected in the conguration menu.
Measurement
CAL COND
Dip sensor in solution.
Value blinks
Calibration will be performed. Automatic return to measuring mode.
Use st to adjust the temperature‑corrected conductivity value. NOTICE: Here, the meter does not perform a temperature compensation!
Page 25
Conductivity Calibration
Calibration with Calibration Solution
(Automatic calibration with preselected calibration solution)
The calibration method is selected in the conguration menu.
NOTICE!
• Make sure that the values of the calibration solutions used correspond exactly to those specied in this manual. If not, the resulting cell constant will be incorrect.
• When calibrating in a liquid, make sure that the sensor, the separate temperature probe (if present) and the calibration solution have the same temperature. Only this ensures that the cell constant is determined correctly.
Measurement
25
Cond
CAL <25°C> 107.00mS/cm PRESS CAL
Measured value Temperature Cal solution conductivity Hourglass blinks
Calibration will be performed. Automatic return to measuring mode.
Dip sensor in calibration solution. The meter automatically compensates for the temperature deviation!
Page 26
26
Cond
Measurement
Conductivity Calibration
FREE CAL Calibration
(Free selection of calibration method)
FREE CAL calibration is selected in the conguration menu.
CAL CELL CONST. blinks
Perform the selected calibration (see CELL CONST., COND or calibration solution).
st to select the desired calibration
Use method (CELL CONST., COND, <25°C>
74.0µS/cm, <25°C> 149.6µS/cm, <25°C>
1.406mS/cm, <25°C> 12.64mS/cm, <25°C>
107.00mS/cm).
Page 27

Oxygen Calibration

AIR CAL Calibration
(Calibrating the slope in air)
The calibration method is selected in the conguration menu.
Measurement
27
Oxy
CAL AIR CAL
Measured value Temperature xxx rH blinks
xxx nA blinks Hourglass blinks
CAL DATA
Date SLOPE xxx nA
Automatic return to measuring mode.
Place sensor in air and wait until the measured values have stabilized.
Use humidity.
Calibration will be performed.
Slope calibration data is displayed.
st to adjust the correct value for relative
Please note: To abort calibration, you can press
at any time.
Page 28
28
Oxy
Measurement
CAL ZERO CAL
Oxygen Calibration
ZERO CAL Calibration
(Zero calibration with oxygen-free medium)
The calibration method is selected in the conguration menu.
xxx nA PRESS CAL blinks
CAL DATA Calibration will be performed.
Date ZERO P. xxx nA
Date SLOPE xxx nA
Automatic return to measuring mode.
Please note: To abort calibration, you can press
Place sensor in oxygen‑free medium (e.g., nitrogen 5.0) and wait until the measured values have stabilized.
Zero calibration data is displayed.
Slope calibration data is displayed.
at any time.
Page 29
Oxygen Calibration
DATA INPUT Calibration
(Calibration by entering known sensor values)
The calibration method is selected in the conguration menu.
Measurement
CAL DATA INPUT
29
Oxy
xx blinks nA ZERO POINT
xxx blinks nA SLOPE
Calibration will be performed. Automatic return to measuring mode.
Please note: To abort calibration, you can press
st to adjust the known value for the
Use sensor zero point.
st to adjust the known value for the
Use sensor slope.
at any time.
Page 30
30
Oxy
Measurement
Oxygen Calibration
FREE CAL Calibration
(Free selection of calibration method)
FREE CAL calibration is selected in the conguration menu.
CAL AIR CAL blinks
Perform the selected calibration (see AIR CAL, ZERO CAL or DATA INPUT calibration).
st to select the desired calibration
Use method (AIR CAL, ZERO CAL, DATA INPUT).
Page 31

Measuring

31
pH Oxy Cond
Once you have completed all preparations, you can start with the actual measurement.
1) Connect the desired sensor to the meter. Some sensors require a special preparation. Please proceed according to the operating instructions for the sensor.
2) Switch the device on using the
3) Depending on the measurement method and the sensor used, immerse the sensing part of the sensor in the medium to be measured.
4) Watch the display and wait for the reading to stabilize.
5) By pressing the value (see data logger, page 32).
key, you can hold and save a measured
or key.

Toggling the Measured Value Display

During measurement, you can toggle the measured value display by pressing :
• pH: between pH and mV
• Cond: between compensated and uncompensated measured value (when temperature compensation, SAL or TDS are activated)
• Oxy: not applicable
Keys for
measurement

Adjusting the Temperature

When you connect a sensor without temperature detector, you can manually adjust the temperature for measurement or calibration:
1) Press to access measuring mode. The adjusted temperature will be displayed.
2) Set the desired temperature value using the or arrow. Holding the key depressed changes the temperature value at high speed.
Page 32
32

Data Logger

pH Oxy Cond
Data Logger
The meter provides a data logger. Prior to use, it must be congured and then activated. You can choose from the following logger types:
• DIFF (signal-controlled logging of measured variable and temperature)
• INT (time-controlled logging at a xed interval)
• DIFF+INT (combined time- and signal-controlled logging)
• SHOT (manual logging by pressing the key)
The data logger records up to 5000 entries and saves them in a circular buer. Already existing entries will be overwritten. The following data are recorded: primary value, temperature, time stamp and device status.
Display: Icons related to the data logger
Memory address (0026)
Measured value is being saved:
: Measured values are displayed
Next value
Primary process
variable
Temperature
DateData logger activated
Time
Logger type
Page 33
Data Logger
pH Oxy Cond

Operating Modes of the Data Logger (Logger Type)

Manual logging when logger is activated (SHOT)
In this mode, a measured value is recorded when the key is pressed.
Measurement Logger activated
The measured value is saved to the address of the last recorded value + 1
Manual logging when logger is deactivated
Measurement Logger deactivated
33
Measured value is maintained Proposed address blinks (address of the last recorded value + 1)
Measured value is saved to the desired address (e.g., for overwriting an incorrect measurement).
Interval (INT)
In this mode, the measured values are cyclically recorded.
Measured parameter
= entry
If desired: Select start address using st.
Example: Logger type “INT”
Time
Page 34
34
Data Logger
pH Oxy Cond
Dierence (DIFF)
When the delta range (process variable and/or temperature) related to the last entry is exceeded, a new entry is created and the delta range is displaced upwards or downwards by the delta value. The rst entry is automatically created when the data logger is started.
Example: First entry = 2, Delta = 1
Measured parameter
4
Delta
3
Delta
2
Delta
1
Delta
= entry
Dierence + Interval combined (DIFF+INT)
When the delta range related to the last DIFF entry is exceeded, a new entry is creat­ed (example: entry A) and the delta range is displaced upwards or downwards by the delta value. As long as the measured value remains within the delta range, logging is performed at the preset interval. The rst DIFF entry is automatically created when the data logger is started.
Measured parameter
4
3
Delta
2
Delta
1
Start logger
Delta
A
Example: Logger type “Interval + Dierence”
Interval
= entry
Time
Time
Page 35
Data Logger

Data Logger Menu

Logger display
CONT Select start address and start the data logger
START Deletes all entries and starts the data logger at start
s t
DEL Deletes all entries
SET Select logger type and congure:
Overview of data logger menu (default in bold print)
Select using arrow keys, conrm by pressing
address 0001
DIFF, INT, DIFF+INT, SHOT (see table below)
35
pH Oxy Cond
.
Logger type
DIFF Delta % air OFF | 0.1 … 100 % air | 1.0 % air
OFF | 0.01 … 20 mg/l | 1.00 mg/l
Delta pH OFF | pH 0.01…14.00 | pH 1.00
OFF | 1… 1000 mV | 1 mV
Delta Cond OFF | 1 … 1000 mS/cm
OFF | 0.1 … 100 S/m | 1 S/m
Delta SAL OFF | 0.1 … 45.0 g/kg Delta TDS OFF | 1 … 1999 mg/l Delta °C / °F OFF | 0.1 … 50.0 °C | 1.0 °C
OFF | 0.1 …90 °F | 1.0 °F
INT Interval h:mm:ss
0:00:01 … 9:59:59 | 12:02:00 AM
DIFF+INT DIFF See logger type DIFF
INT See logger type INT
SHOT Currently selected process variable is recorded
Page 36
36
pH Oxy Cond
Conguring the Data Logger
Prerequisite: The data logger is stopped (press ).
Measurement
Measured value is maintained
Logger: CONT blinks
t
Logger: START blinks
t
Logger: DEL blinks
t
Logger: SET blinks
Data Logger
Logger: Current logger type blinks
Select the appropriate parameters using st and conrm each selection by press­ing . When conguration is nished, CONT blinks. You can start the data logger by selecting START or CONT (see page 37).
Select desired logger type using st: DIFF, INT, DIFF+INT or SHOT.
Page 37
Data Logger
37
pH Oxy Cond

Starting the Data Logger using CONT

Prerequisite: Data logger is congured. Every time the meter has been switched o, the data logger must be restarted (exception: SHOT).
Measurement
Measured value is maintained
Logger: CONT blinks
Address of the last recorded value +1 blinks (proposed start address)
The measured value is saved to the selected start address (exception: SHOT). “… FREE MEMORY” is displayed. “LOGGER” and “active logger type” icons are displayed.
If desired: Select start address using st.

Starting the Data Logger using START

Prerequisite: Data logger is congured. All existing entries are deleted. The start address for saving the values is 0001. Every time the meter has been switched o, the data logger must be restarted (exception: SHOT).
Measurement
Measured value is maintained
Logger: CONT blinks
t
Logger: START blinks
All entries will be deleted. “5000 FREE MEMORY” is displayed. “LOGGER” and “active logger type” icons are displayed.
Page 38
38
pH Oxy Cond

Displaying the Logger Data

Pressing the key displays all stored values.
Measurement
Data Logger
The “RCL” icon and the last record­ed value is displayed.
or
Return to measurement
Use st to select the desired address. Empty memory locations will also be displayed.
Example: Measured value stored at location 0026
Example: Empty memory location 0004
Page 39
Data Logger
39
pH Oxy Cond

Stopping the Data Logger

You can stop the data logger at any time by pressing the key.
Measurement, logger activated
Data logger is stopped. “LOGGER” and “active logger type” icons are no longer displayed. It is still possible to hold a measured value by pressing and send it to any desired address.

Clearing the Data Logger

Selecting “DEL” deletes all data records.
Measurement
Measured value is maintained
Logger: CONT blinks
t
Logger: START blinks
t
Logger: DEL blinks PRESS SET
All stored data are deleted. “0000 DELETED” is displayed.
Page 40
40
pH Oxy Cond
Press the key to access the clock mode. Date and time will be displayed in the format as set in the conguration menu. To set the clock, proceed as follows:
Display of time+date

Clock

Hour display blinks SET HOUR
Minute display blinks SET MINUTE
Seconds display blinks and shows 00
Year display blinks SET YEAR
Month display blinks SET MONTH
Day display blinks SET DAY
Set value.
Set value.
Clock is started, the seconds count up.
Set value.
Set value.
Set value.
Display of corrected time+date
Page 41

Error and Status Messages

41
pH Oxy Cond
Error messages are indicated as “ERROR …” on the display. Information on the sensor condition is indicated by the “Sensoface” icon (friendly, neutral, sad) possibly accom­panied by an info message (“INFO …”).
Example of an error message: ERROR 1 (value out of range)
Sensoface (the “smiley” icon) provides information on the sensor condition (main­tenance request). Measurement can still be performed. After a calibration, the corre­sponding Sensoface icon (friendly, neutral, sad) is shown together with the calibration data. Otherwise, Sensoface is only visible in measuring mode.
The most important error messages and “Sensoface” info messages are shown on the inside of the protective cover. A complete list of messages and their meanings is provided in the following tables.
Example of a “Sensoface” message: INFO 1 (cal timer expired)
Page 42
42
pH Oxy Cond

”Sensoface” Messages

The “Sensoface” icon provides information on the sensor condition:
Sensoface Meaning
Sensor is okay
Calibrate the sensor soon
Calibrate or replace the sensor
The “neutral” and “sad” Sensoface icons are accompanied by an “INFO …” message to give a hint to the cause of deterioration.
Sensoface Message Cause
INFO 1 Calibration timer
INFO 3 Sensocheck
INFO 5 Zero/Slope
INFO 6 Response time
INFO 7 Operating point (asymmetry potential)
INFO 8 Leakage current
INFO 9 ORP oset
INFO 10 Polarization
Page 43
Error and Status MessagesError and Status Messages
pH Oxy Cond

Error Messages

The following error messages can be shown in the display.
Message Cause Remedy
43
blinks ERROR 1 Value out of range ERROR 2 ORP value out of range ERROR 3 Temperature value out of
ERROR 4 Zero point too high/low Thoroughly rinse the sensor and recal-
ERROR 5 Slope too high/low
ERROR 6 Cell constant too high/low Enter nominal cell constant or calibrate
ERROR 8 Calibration error:
ERROR 9 Calibration error:
ERROR 10 Calibration media
ERROR 11 Measured value unstable
ERROR 14 Time and date invalid Set time and date ERROR 18 Conguration invalid Restart, reset to factory settings,
ERROR 19 Factory settings error Device defective, send it in. ERROR 21 Sensor error
ERROR 22 Sensor conict Connect only one sensor.
Battery empty Replace batteries
Check whether the measurement conditions correspond to the adjusted
range
Identical buers
Buer unknown
interchanged
Stability criterion not met
(Memosens)
measuring range.
ibrate. If this does not help, replace the sensor.
the sensor using a known solution. Use a buer solution with a dierent
nominal value before starting the next calibration step.
Make sure that you use the same buer set as congured.
Repeat calibration.
Leave the sensor in the liquid until the temperature is stable. If this does not help, replace the sensor.
congure and calibrate. If this does not help, send in the device for repair.
Connect operational Memosens sensor.
Page 44
44
pH

pH Sensors

pH Product Line

Product Name
(Link to Product Congurator)
Orbisint CPS11D Digital pH sensor
Orbisint CPS12D Digital ORP sensor
Memosens CPS16D Combined digital pH/ORP sensor
Memosens CPS31D Digital pH sensor
Ceraliquid CPS41D Digital pH sensor
Ceraliquid CPS42D Digital ORP sensor
Ceragel CPS71D Digital ORP sensor
Ceragel CPS72D Digital ORP sensor
Memosens CPS76D Combined digital pH/ORP sensor
Orbipore CPS91D Digital pH sensor
Orbipore CPS92D Digital ORP sensor
Memosens CPS96D Combined digital pH/ORP sensor
Ceramax CPS341D Digital non-glass pH sensor
Tophit CPS441D Digital non-glass pH sensor
Tophit CPS471D Digital non-glass pH sensor
Tophit CPS491D Digital non-glass pH sensor
Orbipac CPF81D Digital pH sensor
Orbipac CPF82D Digital ORP sensor
The Product Congurator can be accessed at: www.endress.com/<product name>
Sensor Type
Memosens sensors have a cable coupling, which allows convenient replacement of sensors while the cable remains connected to the meter.
Page 45
45
pH
Endress+Hauser Buer Solutions (pH)
Ready-to-use quality pH buer solutions
Quality buers from Endress+Hauser - CPY20
Solutions which are traced by a DAkkS-accredited Endress+Hauser buer laboratory (DkkS = German Accreditation Body) to a primary reference material of the PTB and to standard reference material of the National Institute of Standards and Technol­ogy (NIST) in accordance with DIN 19266 are used as secondary reference buer solutions.
CPY20 Buer Sets Quantity
pH 2.00 5000 ml / 100 ml / 250 ml pH 4.01 5000 ml / 100 ml / 250 ml / 18 ml pH 6.98 5000 ml / 100 ml / 250 ml / 18 ml pH 9.95 5000 ml / 100 ml / 250 ml pH 11.87 5000 ml / 100 ml / 250 ml

Accessories for pH

Item
1.5 m sensor cable ATEX CYK20 BAB1C2
CPY7 electrolyte vessel, reservoir for KCI electrolyte, 150ml
The Product Congurator can be accessed at:
www.endress.com
Page 46
46
Cond

Conductivity Sensors

Conductivity Product Line

Product Name
(Link to Product Congurator)
Condumax CLS15D Conductivity sensor
Condumax CLS16D Conductivity sensor
Condumax CLS21D Conductivity sensor
Memosens sensors have a cable coupling, which allows convenient replacement of sensors while the cable remains connected to the meter.
Sensor Type
Page 47
47
Cond

CLY11 Conductivity Calibration Solutions

CLY11-A, 74 μS/cm (reference temp. 25°C (77 °F)), 500ml (16.9 fl.oz)
CLY11-B, 149.6 μS/cm (reference temp. 25°C (77 °F)), 500ml (16.9 fl.oz)
CLY11-C, 1.406 mS/cm (reference temp. 25°C (77 °F)), 500ml (16.9 fl.oz)
CLY11-D, 12.64 mS/cm (reference temp. 25°C (77 °F)), 500ml (16.9 fl.oz)
CLY11-E, 107.00 mS/cm (reference temp. 25°C (77 °F)), 500ml (16.9 fl.oz)

Accessories for Conductivity

Item (Link to Product Congurator)
1.5 m sensor cable ATEX CYK20 BAB1C2
Conducal CLY421 calibration set
• Conductivity calibration set (case) for ultrapure water applications
• Complete, factory-calibrated measuring set with certicate, traceable to SRM of NIST and DKD, for comparative measurement in ultrapure water up to max.20μS/cm
• Product Congurator on the product page: www.endress.com/cly421
The Product Congurator can be accessed at:
www.endress.com
Page 48
48

Oxygen Product Line

Oxy

Oxygen Sensors

Product Name
(Link to Product Congurator)
Oxymax COS22D digital oxygen sensor

Accessories for Oxygen

Item
COS22Z maintenance kit
1.5 m sensor cable ATEX CYK20 BAB1C2
The Product Congurator can be accessed at:
www.endress.com
Page 49
Specications
49
pH Oxy Cond
Connections 1x M8 socket, 4 pins, for Memosens lab cable
Display LCD STN 7-segment display with 3 lines and icons
Sensoface Status indication (friendly, neutral, sad) Status indicators For battery power level, logger Notices Hourglass Keypad
Data logger With up to 5000 memory locations Recording Manual, interval- or event-controlled
Communication USB 2.0 Profile HID, driverless installation Usage Data exchange
Diagnostics
Sensor data (Memosens only)
Calibration data Calibration date; zero and slope, or cell constant, resp. Device self-test Automatic memory test (FLASH, EEPROM, RAM) Device data Device type, software version, hardware version
Data retention Parameters, calibration data > 10 years
EMC EN 61326-1 (General Requirements)
Emitted interference Class B (residential area) Immunity to interference Industry
Explosion protection CYM291
1 x M12 socket, 8 pins, for Memosens sensors 2 x 4-mm socket for separate temperature detector 1 x micro USB-B
Be sure to observe the safety instructions when using the USB port.
], [ ], [ ], [ ], [], [], [ ], [ ], [ ]
[
Manufacturer, sensor type, serial number, operating time
EN 61326-2-3 (Particular Requirements for Transmitters)
Global IECEx Ex ia IIC T4/T3 Ga Europe ATEX II 1 G Ex ia IIC T4/T3 Ga
Page 50
50
pH Oxy Cond
RoHS conformity According to directive 2011/65/EU
Power supply 4 x AA batteries
Operating time Approx. 500 h (alkaline)
Nominal operating conditions
Ambient temperature -10 °C ≤ Ta ≤ +40 °C (+14 ... +104 °F) T4 Duracell MN1500
Transport/Storage temp. -25 … +70 °C (-13 ... +158 °F) Relative humidity 0 … 95 %, short-term condensing allowed
Housing
Material PA12 GF30 (silver gray RAL 7001) + TPE (black) Protection IP 66/67 with pressure compensation Dimensions Approx. (132 x 156 x 30) mm Weight Approx. 500 g
For battery types, see Control Drawing No. 209.009-150
-10 °C ≤ Ta ≤ +50 °C (+14 ... +122 °F) T3 Energizer E91, Power One 4106 and Panasonic Pro Power LR6
Page 51
Specications
51
pH
Memosens pH input (also ISFET)
Display ranges
Memosens input ORP
Display ranges
Sensor standardization * ORP calibration (zero adjustment) Permissible calibration range ΔmV (offset) -700 … +700 mV
Sensor standardization * pH calibration
Operating modes * PROG.BUFFERS Calibration with automatic buffer recognition
“Programmed buffers” buffer sets *
Permissible calibration range Zero point 6 ... 8 pH
Calibration timer * Interval 1 … 99 days, can be switched off
Sensoface Provides information on the sensor condition
Evaluation of zero/slope, response, calibration interval
1)
1)
M8 socket, 4 pins, for Memosens lab cable or M12 socket, 8 pins, for Memosens sensors
pH -2.00 … +16.00 mV -1999 … +1999 mV
Temperature -50 … +250 °C (-58 ... +482 °F)
M8 socket, 4 pins, for Memosens lab cable or M12 socket, 8 pins, for Memosens sensors
mV -1999 … +1999 mV Temperature -50 … +250 °C (-58 ... +482 °F)
MANUAL Manual calibration with entry of individual
DATA INPUT Data entry of zero and slope
-01- Endress+Hauser 2.00/4.01/6.98/9.95/11.87
-02- Mettler-Toledo 2.00/4.01/7.00/9.21
-03- Knick CaliMat 2.00/4.00/7.00/9.00/12.00
-04- Ciba (94) 2.06/4.00/7.00/10.00
-05- NIST technical 1.68/4.00/7.00/10.01/12.46
-06- NIST standard 1.679/4.006/6.865/9.180
-07- HACH 4.01/7.00/10.01/12.00
-08- WTW techn. buffers 2.00/4.01/7.00/10.00
-09- Hamilton 2.00/4.01/7.00/10.01/12.00
-10- Reagecon 2.00/4.00/7.00/9.00/12.00
-11- DIN 19267 1.09/4.65/6.79/9.23/12.75
With ISFET: Operating point (asymmetry)
Slope Approx. 74 … 104 % (possibly restricting notes from Sensoface)
buffer values
-750 … +750 mV
* User-defined
1) Ranges depending on Memosens sensor
Page 52
52
Cond
Conductivity input M8 socket, 4 pins, for Memosens lab cable
Measuring ranges CLS15D k = 0.01 : 0-20 µS/cm
CLS16D k = 0.1 : 0.04 µS/cm - 500 µS/cm
CLS21D k = 1 : 10.0 µS/cm - 20.0 mS/cm
Permissible cell constant 0.005 … 200.0 cm
Measuring cycle Approx. 1 s
Temperature compensation Linear 0 … 20 %/K, reference temperature adjustable
Display resolution (autoranging)
Sensor standardization Cell constant Input of cell constant with simultaneous dis-
Measurement error
1,2,3)
nLF: 0 … 120 °C NaCl HCl (ultrapure water with traces) NH3 (ultrapure water with traces) NaOH (ultrapure water with traces)
Conductivity 0.001 μS/cm (c < 0.05 cm
Resistivity 00.00 … 99.99 MΩ cm Salinity 0.0 … 45.0 g/kg (0 … 30 °C) TDS 0 … 1999 mg/l (10 … 40 °C)
Input of solution Input of conductivity of the calibration
Auto Automatic determination of the cell constant
< 0.5 % meas.val. + 0.4 µS * c
k = 0.1 : 0-200 µS/cm
-1
(adjustable)
0.01 μS/cm (c = 0.05 … 0.2 cm–1)
0.1 μS/cm (c > 0.2 cm–1)
play of conductivity value and temperature
solution with simultaneous display of cell constant and temperature
with calibration solution
4)
–1
)
1) according to EN 60746-1, at nominal operating conditions
2) ± 1 count
3) plus sensor error
4) c = cell constant
Page 53
Specications
53
Oxy
Memosens input, oxygen M8 socket, 4 pins, for Memosens lab cable or
Display ranges
Temperature meas. range 1)-20 … +150 °C
Sensor standardization Automatic calibration in air (100 % RH)
1) Ranges depending on Memosens sensor
1)
M12 socket, 8 pins, for Memosens sensors Saturation 0.000 ... 200.0 % Concentration 000 μg/l … 20.00 mg/l
Zero calibration
Page 54
54
0000 DELETED (“data deleted” display) 39
A
AA batteries 13 Accessories for conductivity 47 Accessories for oxygen 48 Accessories for pH 45 AIR CAL (Oxy calibration) 27 Altitude (Oxy conguration) 19 Arrow keys 12 Automatic calibration (Cond) 25 Automatic pH calibration (Programmed buers) 20
B
Batteries 14 Batteries for application in hazardous locations 14 Battery capacity 14 Battery charge indicator 14 Battery compartment 13 Battery icon 14 Battery replacement 13 Benchtop stand 10 Buer sets CPY20 45 Buer set selection 17 Buer solutions 45
C
Calibration, conductivity 24 Calibration set Conducal CLY421 47 CD-ROM 7 Cell constant (Cond calibration) 24 Certicates 7 Charge level of batteries 14 Clearing the datalogger 39 Clock 40 Cond conguration 18 Conductivity calibration, CELL CONST 24 Conductivity calibration, FREE CAL 26 Conductivity calibration solutions CLY11 47

Index

Page 55
Conductivity calibration, with cal solution 25 Conductivity sensors 46 Conductivity sensors, product line 46 Conguration (Cond) 18 Conguration (Oxy) 19 Conguration (pH) 17 Conguring the data logger 36 Connecting a sensor 15 Connecting cable for Memosens 15 Connections 15 Continuous recording of measured values 33 Control functions 12 CONT, starting the data logger 37 Cyclic recording of measured values 33
D
DATA INPUT (Oxy calibration) 29 DATA INPUT (pH calibration) 21 Data logger 32 Data logger conguration 36 Data logger icons 32 Data logger menu 35 Data memory 32 Data of the meter 49 Date 40 Default (conguration) 17 Deleting data logger entries 39 Delta range (data logger) 34 Device conguration (Cond) 18 Device conguration (Oxy) 19 Device conguration (pH) 17 Device messages 41 Dierence (data ogger mode) 34 Dierence+Interval (data logger mode) 34 Digital sensors, conductivity 46 Digital sensors, oxygen 48 Digital sensors, pH 44
55
Page 56
56
Display 11 Display icons 16 Displaying the data logger 32 Displaying the time and date 40 Disposal 3 Documentation 7 Duracell MN1500 battery 14
E
Electrolyte vessel, CPY7 45 Energizer E91 battery 14 ERROR (error codes) 43 Error messages 41 Error messages, overview 43 EU Declarations of Conformity 7
F
FREE CAL, free selection of calibration method (Cond) 26 FREE CAL, free selection of calibration method (Oxy) 30 FREE CAL, free selection of calibration method (pH) 23
H
Hazardous location, batteries 14 Holding the measured value 33 Hook 10 Hours, display 40
I
Icons 16 Icons for data logger 32 INFO messages 42 Inserting the batteries 13 Interrupting the data logger 39 Interval (data logger mode) 33 Introduction 8
Index Index
Page 57
Index
K
Keypad 12
L
Logger 32 Logger display 35 Logger type (data logger modes) 33
M
Manual calibration (Cond) 24 Manual calibration (pH) 22 Measured-value recording 33 Measuring 31 Memory for measured values 32 Memosens cable, description 9 Memosens connecting cable 15 Memosens sensors, conductivity 46 Memosens sensors, pH 44 Menu of data logger 35 Menu structure of Cond conguration 18 Menu structure of data logger 35 Menu structure of Oxy conguration 19 Menu structure of pH conguration 17 Messages 41 Messages, “Sensoface” 42 Micro USB port 15 Minutes, display 40
N
Nitrogen 5.0 28
O
on/o key 12 Operating modes of the data logger 33 ORP sensors 44 Overview of Cond conguration 18 Overview of error messages 43 Overview of Oxy conguration 19 Overview of pH conguration 17
57
Page 58
58
Oxy conguration 19 Oxygen calibration, AIR CAL 27 Oxygen calibration, DATA INPUT 29 Oxygen calibration, FREE CAL 30 Oxygen calibration, ZERO CAL 28 Oxygen-free medium 28 Oxygen sensors 48
P
Package contents 6 Panasonic Pro Power LR6 battery 14 Parameter setting, data logger 36 Parameter settings (Cond conguration) 18 Parameter settings (Oxy conguration) 19 Parameter settings (pH conguration) 17 pH buer solutions 45 pH calibration, DATA INPUT 21 pH calibration, FREE CAL 23 pH calibration, MANUAL 22 pH calibration, PROG. BUFFERS 20 pH conguration 17 pH sensors 44 Ports 15 Power-on 16 Power One 4106 battery 14 Product range 44 Programmed buers, description 9 Programmed buers (pH calibration) 20 Protective cover 10
Q
Quickstart guides 7
Index Index
Page 59
Index
R
Rating plate 10 Real-time clock 40 Recorded data, display 38 Redox sensors 44 Registered trademarks 3 Replacing the batteries 13 Reservoir for KCI electrolyte 45 Reset to factory settings (Default) 17 Return of products under warranty 3
S
Safety instructions 7 Saving the currently measured value 33 Seconds, display 40 Sensoface, description 9 Sensoface messages 42 Sensor connection 15 Sensors for oxygen, product line 48 Sensors for pH, product line 44 Sensor without temperature detector 31 Settings for conductivity measurement 18 Settings for oxygen measurement 19 Settings for pH measurement 17 Setting the data logger 36 Setup (Cond conguration) 18 Setup (Oxy conguration) 19 Setup (pH conguration) 17 SHOT (data logger mode) 33 Smiley 42 Specications 49 Specic test report 7 Start address (data logger) 33 Starting the data logger using CONT 37 Starting the data logger using START 37 START, starting the data logger 37 Start-up 13
59
Page 60
60
Status messages 41 Stopping the data logger 39 Structure of data logger 35 Suspending the meter 10 Switching on the meter 12, 16 Switching the measured value display 31 Symbols in display 16
T
T3, temperature class 14 T4, temperature class 14 Table of Cond conguration 18 Table of error messages 43 Table of Oxy conguration 19 Table of pH conguration 17 Technical data 49 Temperature class 14 Temperature, manual adjustment 31 Temperature probe 15 Toggling the measured value display 31 Trademarks 3 Triangle icons 12
V
Value-added features 9 Viewing recorded data 38 Viewing the logger data 38
W
Warranty 3
Z
ZERO CAL (Oxy calibration) 28
Index
Page 61
Page 62
Page 63
Page 64
088989
www.addresses.endress.com
Endress+Hauser Conducta GmbH+Co. KG Dieselstraße 24 70839 Gerlingen - Germany Phone: +49 7156 209 790 Fax: +49 7156 28 158
TA-209.4MU-EHE01 20150413 Software version: 1.x 71265352
Loading...