Endress+Hauser CPA442 Specifications

TI00306C/07/EN/13.13 71202120
Products Solutions Services
Technical Information
Unifit CPA442
Process assembly For safe installation of 12 mm sensors (Pg 13.5) in the food, beverage, pharmaceutical and chemical industry

Application

• Process assembly for 12 mm sensors like pH glass sensors, pH ISFET sensors, oxygen sensors etc.
• Permanent installation in tanks and pipework
•Food
•Pharmaceuticals
•Chemicals
• Water treatment

Your benefits

• Robust assembly
• Complies with 3A standard 74-05
• Complies with the basic criteria of the European Hygienic Equipment Design Group (EHEDG): – In-place cleanability – In-line steam sterilizability
• Flush mounted, leak-tight electrode seal (molded seal)
• Electropolished surface R steel 1.4435 (AISI 316 L) and Alloy C22)
• Hygienic process connections, flanges and further established process connections
• Protection cover for KCl and gel sensors
=0.38 μm or 0.76 μm (stainless
a
Unifit CPA442

Function and system design

Functionality The process assembly CPA442 is used to install 12 mm sensors. You can install the assembly vertically

or angularly in pipes or reactors. The assembly seals the sensor against the process. It is suitable for:
• CIP (cleaning in place)
• SIP (sterilization in place)
The stainless steel version and the Alloy C22 version are autoclavable.

Measuring system A complete measuring system comprises:

• Unifit CPA442 assembly
• 12 mm sensor, e.g. Orbisint CPS11D or Oxymax COS22D
• Transmitter, e.g. Liquiline CM44x
• Measuring cable
a0019355
Example of a measuring system
1 Unifit CPA442 2 Liquiline CM44x 3Measuring cable 4 Orbisint CPS11D A Installation in pipe B Installation in reactor
2 Endress+Hauser
Unifit CPA442

Reliability • The asssemblies are manufactured according to the certified quality criteria of Endress+Hauser.

• The assemblies are made of high-quality materials like stainless steel 1.4435 (AISI 316 L), Alloy C22 or highly resistant plastics like PVDF or PEEK. Only these materials and the respective sealings are in contact with medium. The assemblies are manufactured from solid material (except the 15° versions S and T).
• The assemblies are inline cleanable and inline sterilizable according to the EHEDG criteria.
• Certified versions and certificates are available.
a0019557
a0019558
Endress+Hauser 3
Unifit CPA442
The certificates are subjects to changes. The shown examples might not be the latest versions.
a0019559
4 Endress+Hauser
Unifit CPA442
15°
15°
A B
10°
10°
C

Installation

Installation angle

A
Glass sensor:
B
ISFET pH sensor:
C
Oxygen sensor:
Fig. 1: Permitted installation angles depending on the sensor

Environment

Ambient temperature -15 to 80 °C (+5 to 176 °F)

Storage temperature -15 to 60 °C (+5 to 140 °F)

Installation angle of at least 15° from the horizontal No restrictions, recommended 0 ... 180° Installation angle of at least 10° from the horizontal
a0001792
Endress+Hauser 5
Loading...
+ 11 hidden pages