• No flow needed - measurement possible in still water
• COS61D - the Liquiline sensor
– Plug&Play: Safe communication based on Memosens protocol
– Optionally with M12 plug for fast connection to the transmitter
• COS61 - the Liquisys sensor
– Compatible with tried-and-tested COS31 with COM2x3W: Easy measuring point
changeover to optical technology
– Compatible with COS41 with COM2x3D with conversion kit
Function and system design
Oxymax COS61D/COS61
Measuring principle
Sensor structure
Oxygen-sensitive molecules (markers) are integrated into an optically active layer (fluorescence
layer).
The fluorescence layer, an optical insulating layer and a cover layer are applied on top of one another
on a carrier. The cover layer is in direct contact with the medium.
The sensor optics are directed at the rear of the carrier and therefore at the fluorescence layer.
Measurement process (principle of quenching)
If the sensor is immersed in the medium, an equilibrium is very quickly established between the
oxygen partial pressure in both the medium and the fluorescence layer.
1.The sensor optics send green light pulses to the fluorescence layer.
2.The markers "respond" (fluoresce) with red light pulses.
The duration and intensity of the response signals are directly dependent on the oxygen
contents and oxygen partial pressure.
If the medium is free from oxygen, the response signals are long and very intense.
Oxygen molecules mask the marker molecules. As a result, the response signals are shorter and less
intense.
Measurement result
The sensor returns a signal that is proportional to the oxygen concentration in the medium.
‣
The medium temperature and air pressure are already taken into account to calculate the oxygen
concentration in the sensor.
The sensor provides measured values for temperature and partial pressure as well as a raw measured
value. This value corresponds to the fluorescence decay time and is approx. 20 μs in air and
approx. 60 μs in oxygen-free media.
For optimum measurement results
1.During calibration, enter the current air pressure at the transmitter.
2.In the case of saline media:
Enter the salinity.
3.For measurements in the units %Vol or %SAT:
Also enter the current operating pressure in the measuring mode.
• Operating Instructions for Memosens, BA01245C
For all transmitters, analyzers and samplers in the Liquiline CM44x/P/R, Liquiline System
CA80XX and Liquistation CSFxx product families
• Operating Instructions for Liquisys COM2x3, BA00199C
2Endress+Hauser
Oxymax COS61D/COS61
12
3
4
5
6
Measuring system
COS61D
A complete measuring system consists of the following components at least:
• Oxymax COS61D oxygen sensor
• Liquiline CM44x multi-channel transmitter
• Sensor cable, optionally available with M12 connector
• Assembly, e.g. flow assembly COA250, immersion assembly CYA112 or
retractable assembly COA451
Optionally:
• Assembly holder Flexdip CYH112 for immersion operation
• Extension cable CYK11
• Cleaning system
1Example of a measuring system with COS61D
1
Sensor cable
2
Liquiline CM44x
3
Flexdip CYH112
A0012882
4
Basin rim with railing
5
Oxymax COS61D
6
FlexdipCYA112
Endress+Hauser3
COS61
12
3
4
5
6
A complete measuring system comprises:
• Oxymax COS61 oxygen sensor
• Transmitter, e.g. Liquisys COM2x3-W
• Sensor cable
• Assembly, e.g. flow assembly COA250, immersion assembly CYA112 or
retractable assembly COA451
Optionally:
• Flexdip CYH112 assembly holder for immersion operation
• VS junction box (for cable extension)
• Cleaning system
Oxymax COS61D/COS61
2Example of a measuring system with COS61
1
Liquisys COM253
2
Oxymax COS61
3
Pipe (ascending pipe)
Input
Measured values
Measuring ranges
Dissolved oxygen [mg/l, µg/l, ppm, ppb or %SAT or hPa]
Measuring ranges apply for 25 ˚C (77 ˚F) and 1013 hPa (15 psi)
With Liquiline CM44x, CM44xR, CM44P or with Liquisys COM2x3-W:
• 0 to 20 mg/l
• 0 to 400 hPa
• 0 to 200 % SAT
A0012885
4
Direction of medium flow
5
Cleanfit COA451
6
Sensor cable
4Endress+Hauser
Oxymax COS61D/COS61
GN
PK
GY
YE
COM A
COM B
U
+
U
–
Power supply
Electrical connection
COS61D
Connection data
• Sensor cable connected directly to the terminal connector of the basic module
• Optional: sensor cable plug connected to the M12 sensor socket of the transmitter
With this type of connection, the transmitter is already wired at the factory.
3Sensor fixed cable with terminated cable cores
COS61 to field device
You connect the sensor directly to the transmitter via the special measuring cable with SXP
connector.
Sensor with fixed cable (OMK)Sensor with TOP68 plug connection (CYK71)
CoreAssignmentCoreAssignment
B
B
YE+U
Coax, inside-U
B
B
1.Remove the SXP connector (transmitter side!) from the cable.
2.Refer to the table for the cable assignment and terminals of Liquisys COM223-WX/WS.
Please note that the cable assignment varies depending on the sensor version (fixed cable
or TOP68 plug connection).
Performance characteristics
From air to nitrogen at reference operating conditions:
t90 : 60 s
Endress+Hauser5
Loading...
+ 11 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.