Emerson Micro Motion 9739 MVD Configuration And Use Manual

Page 1
CongurationandUseManual
P/NMMI-20016855,Rev .AA
June2010
MicroMotion®9739MVDTransmitters CongurationandUseManual
CongurationOperationMaintenance
Page 2
MicroMotioncustomerservice
LocationTelephoneNumber
CanadaandLatinAmerica+1303-527-5200(U.S.A.)
Japan35769-6803 Asia
Allotherlocations+656777-8211(Singapore)
U.K.08702401978(toll-free) Europe
Allotherlocations+31(0)318495555(TheNetherlands)
CustomersoutsidetheU.S.A.canalsosendanemailtoow.support@emerson.com.
Copyrightsandtrademarks
©2010MicroMotion,Inc.Allrightsreserved.TheMicroMotionandEmersonlogosaretrademarksandservicemarksofEmersonElectricCo.
MicroMotion,ELITE,MVD,ProLink,MVDDirectConnect,andPlantWebaremarksofoneoftheEmersonProcessManagementfamilyof
companies.Allothertrademarksarepropertyoftheirrespectiveowners.
Page 3
Contents
IGettingstarted............................................................................................................1
Chapter1Beforeyoubegin........................................................................................................3
1.1Safetymessages.............................................................................................3
1.2Obtainversioninformation................................................................................3
1.3Availablecommunicationstools........................................................................3
1.4Additionaldocumentationandresources...........................................................4
1.59739MVDtransmittercongurationworksheet..................................................4
Chapter2Quickstartwiththedisplay.......................................................................................19
2.1Applypower..................................................................................................19
2.2Congurationtipsandtricks...........................................................................19
2.3ConguretheprimarymAoutputtoreportmassowrateinauser-selected
measurementunit..........................................................................................20
2.4Performalooptest........................................................................................21
2.5Zerotheowmeter.........................................................................................23
Chapter3QuickstartwithProLinkII.........................................................................................25
3.1Applypower..................................................................................................25
3.2ConnectwithProLinkII..................................................................................25
3.3Congurationtipsandtricks...........................................................................40
3.4ConguretheprimarymAoutputtoreportmassowrateinauser-selected
measurementunit..........................................................................................41
3.5Performalooptest........................................................................................41
3.6TrimmAoutputs............................................................................................43
3.7Zerotheowmeter.........................................................................................43
3.8Testortunethesystemusingsensorsimulation..............................................44
3.9Backuptransmitterconguration....................................................................46
3.10Enable/disableHARTsecurity........................................................................47
Chapter4QuickstartwiththeFieldCommunicator...................................................................49
4.1Applypower..................................................................................................49
4.2ConnectwiththeFieldCommunicator.............................................................49
4.3Congurationtipsandtricks...........................................................................52
4.4ConguretheprimarymAoutputtoreportmassowrateinauser-selected
measurementunit..........................................................................................53
4.5Performalooptest........................................................................................54
4.6TrimmAoutputs............................................................................................55
4.7Zerotheowmeter.........................................................................................55
4.8Testortunethesystemusingsensorsimulation..............................................56
4.9Enable/disableHARTsecurity........................................................................58
IIReferenceinformationforcommissioning.................................................................59
Chapter5Configureprocessmeasurement...............................................................................61
5.1Characterizetheowmeter.............................................................................61
5.2Conguremassowmeasurement.................................................................64
5.3Congurevolumeowmeasurementforliquidapplications..............................69
5.4Conguregasstandardvolumeowmeasurement..........................................74
i
Page 4
5.5CongureFlowDirection....................................................................................79
5.6Conguredensitymeasurement.....................................................................84
5.7Conguretemperaturemeasurement..............................................................88
5.8Congurepressurecompensation..................................................................90
5.9Congurethepetroleummeasurementapplication...........................................91
5.10Conguretheconcentrationmeasurementapplication.....................................93
Chapter6Configuredeviceoptionsandpreferences..................................................................99
6.1Congurethetransmitterdisplay.....................................................................99
6.2Enableordisableoperatoractionsfromthedisplay.......................................103
6.3Conguresecurityforthedisplaymenus......................................................105
6.4Congurethespeedofthetransmitter’sresponsetochangesinprocess
data............................................................................................................107
6.5Congurealarmhandling.............................................................................108
6.6Congureinformationalparameters..............................................................1 12
Chapter7Integratethemeterwiththecontrolsystem............................................................117
7.1CongurethemAoutputs.............................................................................1 17
7.2Congurethefrequencyoutput.....................................................................124
7.3Congurethediscreteoutput........................................................................129
7.4Congurethediscreteinput..........................................................................134
7.5CongurethemAinput................................................................................136
7.6Conguredigitalcommunications.................................................................138
7.7Congureevents.........................................................................................146
7.8Setuppollingforpressure............................................................................148
7.9Setuppollingfortemperature.......................................................................150
IIIOperations,maintenance,andtroubleshooting......................................................153
Chapter8Transmitteroperation.............................................................................................155
8.1Recordtheprocessvariables.......................................................................155
8.2Viewprocessvariables................................................................................155
8.3Viewtransmitterstatus.................................................................................157
8.4Viewandacknowledgestatusalarms............................................................157
8.5Startandstoptotalizersandinventories........................................................161
8.6Resetmassandvolumetotalizers................................................................162
8.7ResetmassandvolumeinventoriesusingProLinkII......................................163
Chapter9Measurementsupport.............................................................................................165
9.1Optionsformeasurementsupport.................................................................165
9.2Validatethemeter........................................................................................165
9.3Performa(standard)D1andD2densitycalibration.......................................168
9.4PerformaD3andD4densitycalibration(T-Seriessensorsonly)....................171
9.5Performtemperaturecalibration....................................................................174
Chapter10Troubleshooting.....................................................................................................177
10.1TransmitterstatusLEDstates.......................................................................177
10.2Statusalarms..............................................................................................178
10.3Flowproblems.............................................................................................184
10.4Densityproblems.........................................................................................187
10.5Temperatureproblems.................................................................................187
10.6Milliampoutputproblems.............................................................................188
ii
Page 5
10.7Frequencyoutputproblems..........................................................................189
10.8Usesensorsimulationfortroubleshooting.....................................................190
10.9Checkpowersupplywiring...........................................................................190
10.10Checksensor-to-transmitterwiring................................................................191
10.11Checkgrounding.........................................................................................191
10.12Checkforradiofrequencyinterference..........................................................191
10.13CheckHARTcommunicationloop.................................................................192
10.14CheckHARTAddressandLoopCurrentMode.......................................................192
10.15CheckHARTburstmode..............................................................................193
10.16CheckmAoutputtrim..................................................................................193
10.17CheckLowerRangeValueandUpperRangeValue.................................................193
10.18CheckmAOutputFaultAction............................................................................193
10.19CheckFrequencyOutputMode...........................................................................193
10.20CheckFrequencyOutputMaximumPulseWidthandFrequencyOutputScaling
Method.........................................................................................................194
10.21CheckFrequencyOutputFaultAction...................................................................194
10.22CheckFlowDirection.......................................................................................194
10.23Checkcutoffs..............................................................................................194
10.24Checkforslugow......................................................................................195
10.25Checkthedrivegain....................................................................................195
10.26Checkthepickoffvoltage.............................................................................197
10.27Checkforelectricalshorts............................................................................198
Appendicesandreference.......................................................................................201
AppendixADefaultvaluesandranges........................................................................................203
A.1Defaultvaluesandranges............................................................................203
AppendixBTransmittercomponentsandinstallationwiring......................................................207
B.1Transmittercomponents...............................................................................207
B.2Transmitter-to-sensorwiring.........................................................................209
B.3Powersupplyterminals................................................................................209
B.4Input/output(I/O)terminals...........................................................................210
AppendixCUsingthetransmitterdisplay...................................................................................213
C.1Componentsofthetransmitterinterface........................................................213
C.2Accessandusethedisplaymenusystem.....................................................215
C.3Displaycodesforprocessvariables..............................................................219
C.4Codesandabbreviationsusedindisplaymenus............................................220
C.5Menumapsforthetransmitterdisplay...........................................................223
AppendixDUsingProLinkIIwiththe9739MVDtransmitter.......................................................231
D.1BasicinformationabouttheProLinkIIsoftwaretool.......................................231
D.2MenumapsforProLinkII.............................................................................232
AppendixEUsingtheFieldCommunicatorwiththe9739MVDtransmitter.................................237
E.1BasicinformationabouttheFieldCommunicator...........................................237
E.2MenumapsfortheFieldCommunicator........................................................238
iii
Page 6
Page 7
I
I I
Gettingstarted
Chapterscoveredinthispart:
♦Beforeyoubegin ♦Quickstartwiththedisplay ♦QuickstartwithProLinkII ♦QuickstartwiththeFieldCommunicator
Page 8
Page 9
Chapter1
Beforeyoubegin
Topicscoveredinthischapter:
♦Safetymessages ♦Obtainversioninformation ♦Availablecommunicationstools ♦Additionaldocumentationandresources
9739MVDtransmittercongurationworksheet

1.1Safetymessages

Safetymessagesareprovidedthroughoutthismanualtoprotectpersonnelandequipment.Readeach safetymessagecarefullybeforeproceedingtothenextstep.

1.2Obtainversioninformation

Tocongure,use,andtroubleshootthetransmitter,youmayneedtoknowtheversioninformationof yourtransmittersoftware,ProLinkIIsoftwareapplication,and/orHARTdevicedescription.
Procedure
SeeTable1-1forinformationonhowtoobtaintheversioninformation.
Table1-1Methodstoobtainversioninformation
ComponentWithdisplayWithProLinkIIWithFieldCommunicator
TransmittersoftwareOFF-LINEMAINT→VER
ProLinkIINotapplicableHelp→AboutProLinkIINotapplicable
HARTdevicedescriptionNotapplicableNotapplicable
View→Installed Options→SoftwareRevisions
Overview→Shortcu­ts→DeviceInforma­tion→Revisions→Xmtr SoftwareRev
Overview→Short­cuts→DeviceInforma­tion→Revisions→DDRe­vision

1.3Availablecommunicationstools

Youcanuseavarietyofcommunicationstoolstointerfacewiththe9739MVDtransmitter.
ConfigurationandUseManual
3
Page 10
Beforeyoubegin
Thefollowingcommunicationstoolsaresupported:
•Transmitterdisplay,ifthetransmitterwasorderedwithadisplay
ProLinkIIsoftware(v2.91orlater)
•FieldCommunicator(DDv2orlater)
Tip
YoumaybeabletouseothertoolsfromEmersonProcessManagement,suchasAMSSuite:Intelligent DeviceManagerortheSmartWirelessTHUM™Adapter.UseofAMSortheSmartWirelessTHUM Adapterisnotdiscussedinthismanual.Foryourreference,theAMSinterfaceissimilartothe interface.FormoreinformationontheSmartWirelessTHUMAdapter,refertothedocumentation availableatwww.micromotion.com.
Thismanualprovidesbasicinformationonusingthetransmitterdisplay,ProLinkII,andthe FieldCommunicator.FormoreinformationonusingProLinkII,seetheProLinkIIusermanualavailable ontheMicroMotionwebsite(www.micromotion.com)orontheMicroMotionuserdocumentationCD.For moreinformationonusingFieldCommunicator,seetheFieldCommunicatordocumentationavailable onthe
MicroMotionwebsite(www.micromotion.com).
ProLinkII

1.4Additionaldocumentationandresources

MicroMotionprovidesadditionaldocumentationtosupporttheinstallationandoperationofthe9739 MVDtransmitter.
SeeTable1-2forthedocumentationresourcesavailabletosupportthe9739MVDtransmitter.All documentationresourcesareavailableontheMicroMotionwebsiteatwww.micromotion.comoronthe MicroMotionuserdocumentationCD.
Table1-2Additionaldocumentationandresources
TopicDocument
SensorSensordocumentation
TransmitterinstallationMicroMotion9739MVDTransmitters:Installation
Manual
Hazardousareainstallation
TransmitterelectronicsmoduleupgradeMicroMotion9739MVDT ransmitterElectronicsModule
Seetheapprovaldocumentationshippedwiththe transmitter,ordownloadtheappropriatedocumentation fromthe
InstallationGuide
MicroMotionwebsiteatwww.micromotion.com.

1.59739MVDtransmitterconfigurationworksheet

Usethe9739MVDtransmittercongurationworksheetforbothplanningandrecordingthetransmitter conguration.
Additionally,thecongurationworksheetshowstheparametersthatareaccessiblefromthedifferent communicationstools.Chooseacommunicationstoolthatprovidesaccesstotheparametersthat youplantocongure.
4
MicroMotion9739MVDTransmitters
Page 11
Table1-39739MVDtransmitterconfigurationsettings
Beforeyoubegin
Configurablewith:
Configuration parameterSettingDisplayProLinkII
SensorType
Flowcalibration factor
D1
D2
Density temperature coefcient(DT)
K1
K2
FD
Temperature calibrationfactor
Massow measurement unit
Ifmassowisa specialunit
Flowdamping
Massowcutoff
Volumetype
StandardGas Density
Volumeow measurement unit
qT-Series(StraightTube)
qOther(CurvedTube)
Basemassunit:
Basetimeunit:
Conversionfactor:
Flowtext:
Totaltext:
qLiquidVolume
qStdGasVolume
üüü
üüü
üüü
üü
üü
üü
üü
üü
üü
üü
üü
üü
üü
üü
üü
üü
üü
üü
üü
üü
FieldCommuni­cator
ConfigurationandUseManual
5
Page 12
Beforeyoubegin
Table1-39739MVDtransmitterconfigurationsettings
Configuration parameterSettingDisplay
Basemassunit:
Basetimeunit:
Ifvolumeowis aspecialunit
Volumeow cutoff
Flowdirection
Conversionfactor:
Flowtext:
Totaltext:
qAbsoluteValue
qBidirectional
qForward
qNegateBidirectional
qNegateForward
qReverse
Density measurement unit
Slugowlow limit
Slugowhigh limit
Slugduration
Densitydamping
Densitycutoff
Temperature measurement unit
q°C
q°F
q°R
qKelvin
Temperature damping
Pressureunits:
Pressure compensation
Flowfactor:
Densityfactor:
Calibrationpressure:
continued
Configurablewith:
ProLinkII
üü
üü
üü
üü
üü
üü
üü
üüü
üü
üü
üü
üü
üü
üüü
üü
üüü
üü
üü
üü
FieldCommuni­cator
6
MicroMotion9739MVDTransmitters
Page 13
Beforeyoubegin
Table1-39739MVDtransmitterconfigurationsettings
Configuration parameterSettingDisplay
APItabletype:
qDegreesAPI,reference
temperature60°F
qRelativeDensity/Specic
Gravity,referencetemperature 60°F
Petroleum measurement application(if available)
qkg/m
APIUnits:
3
atuser-dened referencetemperature (Temperature:____)
qGeneralizedCrudeorJP4
(APIChapter11.1“A”T ables)
qGeneralizedProducts(API
Chapter11.1“B”T ables)
qUserDenedTEC(API
Chapter11.1“C”T ables)
Concentration measurement application(if available)
Weights& Measures application(if available)
Activecurve:
Derivedvariable:
Totalizerresetoptions:
qNotresettablefromdisplayor
digitalcommunications
qResettablefromdigital
communicationsonly
qResettablefrodisplayand
digitalcommunications
qResettablefromdisplayonly
Language:
qEnglish
Transmitter display
qFrench
qGerman
qSpanish
continued
Configurablewith:
üü
ProLinkII
üü
üü
üü
üü
ü
FieldCommuni­cator
ConfigurationandUseManual
7
Page 14
Beforeyoubegin
Table1-39739MVDtransmitterconfigurationsettings
Configuration parameterSettingDisplay
Displayvariables:
•Var1:
•Var2:
•Var3:
•Var4:
•Var5:
•Var6:
•Var7:
•Var8:
•Var9:
•Var10:
•Var11:
•Var12:
•Var13:
•Var14:
•Var15:
Updateperiod(100millisecondsto 10,000millisecondsrange;defaultis 200milliseconds):
Autoscroll:
qEnable
qDisable
Autoscrollrate(defaultis10 seconds):
Backlight:
qOn
qOff
Totalizerstart/stop:
qEnabled
qDisabled
Totalizerreset:
qEnabled
qDisabled
continued
Configurablewith:
ProLinkII
üü
üüü
üü
üü
üü
üü
üü
FieldCommuni­cator
8
MicroMotion9739MVDTransmitters
Page 15
Beforeyoubegin
Table1-39739MVDtransmitterconfigurationsettings
Configuration parameterSettingDisplayProLinkII
Acknowledgeallalarms:
qEnabled
qDisabled
Ofinemenu:
qEnabled
qDisabled
Alarmpassword:
qEnabled
qDisabled
Responsetime:
qNormal
qSpecial
Tag:
Descriptor:
Message:
Date:
Informational parameters
mAOutput1
Sensorserialnumber:
Sensormodel:
Material:
Flange:
Liner:
ProcessVariable:
Measurementunit:
Lowerrangevalue(LRV):
Upperrangevalue(URV):
Massowcutoff:
Addeddamping:
Faultaction:
qUpscale
qDownscale
qInternalzero
qNone
continued
Configurablewith:
üü
üü
üü
ü
üü
üü
ü
üü
üü
üü
üüü
üüü
üüü
üüü
üü
üü
üü
FieldCommuni­cator
ü
ü
ü
ConfigurationandUseManual
9
Page 16
Beforeyoubegin
Table1-39739MVDtransmitterconfigurationsettings
Configuration parameterSettingDisplayProLinkII
FaultLevel:
ProcessVariable:
Measurementunit:
Lowerrangevalue(LRV):
Upperrangevalue(URV):
Massowcutoff:
mAOutput2
Frequency output
Addeddamping:
Faultaction:
qUpscale
qDownscale
qInternalzero
qNone
Faultlevel:
Processvariable:
Scalingmethod:
qFreqency=Flow
qPulses/Unit
qUnits/Pulse
Pulsewidth:
qActiveHigh
qActiveLow
Polarity:
qActiveHigh
qActiveLow
Faultaction:
qUpscale
qDownscale
qInternalzero
qNone
Faultlevel:
Powertype:
qInternal
qExternal
continued
Configurablewith:
üü
üüü
üüü
üüü
üüü
üü
üü
üü
üü
üüü
üüü
üü
üüü
üü
üü
üüü
FieldCommuni­cator
10MicroMotion9739MVDTransmitters
Page 17
Beforeyoubegin
Table1-39739MVDtransmitterconfigurationsettings
Configuration parameterSettingDisplay
Assignment:
qCalibrationinprogress
qDiscreteevent1
qDiscreteevent2
qDiscreteevent3
qDiscreteevent4
qDiscreteevent5
qEvent1
qEvent2
qFaultconditionindication
qFlowswitchindication
Discreteoutput
qForward/Reverseindication
Polarity:
qActiveHigh
qActiveLow
Powertype:
qInternal
qExternal
Faultaction:
qUpscale
qDownscale
qNone
Polarity:
qActiveHigh
qActiveLow
Assignment:
qStartzero:
Discreteinput
qStart/stoptotalizers:
qResetmasstotal:
qResetgasstandardvolume
total:
qResetalltotals:
qResetAPItemperature-
correctedvolumetotal:
continued
Configurablewith:
ProLinkII
üüü
üüü
üüü
üü
üüü
üüü
FieldCommuni­cator
ConfigurationandUseManual
11
Page 18
Beforeyoubegin
Table1-39739MVDtransmitterconfigurationsettings
Configuration parameterSettingDisplay
Processvariable:
qExternalpressure
qInternalpressure
mAinput
HARTAddress orModbus Address
Loopcurrent mode(ProLinkII) ormAoutputac-
FieldCom-
tion( municator)
ModbusASCII
qNone
Lowerrangevalue(LRV):
Upperrangevalue(URV):
qEnable
qDisable
qEnable
qDisable
Burstmode
qEnable
qDisable
Burstmode output
qDynamicvariablesandPV
current
qPrimaryvariable
qPVcurrentandpercentageof
range
qReaddevicevariableswith
status
qTransmittervariables
HARTvariables
•Primaryvariable(PV):
•Secondaryvariable(SV):
•Tertiaryvariable(TV):
•Quaternaryvariable(QV):
continued
Configurablewith:
ProLinkII
üüü
üüü
üüü
üü
üü
ü
üü
üü
üü
FieldCommuni­cator
12MicroMotion9739MVDTransmitters
Page 19
Beforeyoubegin
Table1-39739MVDtransmitterconfigurationsettings
Configuration parameterSettingDisplay
Digital Communications FaultActions Settings
qUpscale
qDownscale
qReportNAN(NotANumber)
qFlowRatesgotozerovalue=
zeroow
qFlowRatesgotozerovalue
=zeroow.Densityand Temperaturegotozero.
qNoAction
Outputassignment:
Processvariable:
Event1
Type:
qHighalarm
qLowalarm
Setpoint:
Outputassignment:
Processvariable:
Event2
Type:
qHighalarm
qLowalarm
Setpoint:
EventType:
qHI
qLO
qIN
DiscreteEvent1
qOUT
ProcessVariable:
SetpointA:
SetpointB:
continued
Configurablewith:
ProLinkII
üü
üü
üü
üü
üü
üü
üü
üü
üü
üü
üü
üü
üü
FieldCommuni­cator
ConfigurationandUseManual
13
Page 20
Beforeyoubegin
Table1-39739MVDtransmitterconfigurationsettings
Configuration parameterSettingDisplay
Action:
qNone(default)
qStartSensorZero
qStart/stopalltotalizers
qResetmasstotal
qResetvolumetotal
qResetgasstandardvolume
total
qResetalltotals
qResettemperature-corrected
volumetotal
qResetCMreferencevolume
total
qResetCMnetmasstotal
qResetCMnetvolumetotal
qIncrementCMmatrix
EventType:
qHI
qLO
qIN
DiscreteEvent2
qOUT
ProcessVariable:
SetpointA:
SetpointB:
continued
Configurablewith:
ProLinkII
üü
üü
üü
üü
üü
FieldCommuni­cator
14
MicroMotion9739MVDTransmitters
Page 21
Beforeyoubegin
Table1-39739MVDtransmitterconfigurationsettings
Configuration parameterSettingDisplay
Action:
qNone(default)
qStartSensorZero
qStart/stopalltotalizers
qResetmasstotal
qResetvolumetotal
qResetgasstandardvolume
total
qResetalltotals
qResettemperature-corrected
volumetotal
qResetCMreferencevolume
total
qResetCMnetmasstotal
qResetCMnetvolumetotal
qIncrementCMmatrix
EventType:
qHI
qLO
qIN
DiscreteEvent3
qOUT
ProcessVariable:
SetpointA:
SetpointB:
continued
Configurablewith:
ProLinkII
üü
üü
üü
üü
üü
FieldCommuni­cator
ConfigurationandUseManual
15
Page 22
Beforeyoubegin
Table1-39739MVDtransmitterconfigurationsettings
Configuration parameterSettingDisplay
Action:
qNone(default)
qStartSensorZero
qStart/stopalltotalizers
qResetmasstotal
qResetvolumetotal
qResetgasstandardvolume
total
qResetalltotals
qResettemperature-corrected
volumetotal
qResetCMreferencevolume
total
qResetCMnetmasstotal
qResetCMnetvolumetotal
qIncrementCMmatrix
EventType:
qHI
qLO
qIN
DiscreteEvent4
qOUT
ProcessVariable:
SetpointA:
SetpointB:
continued
Configurablewith:
ProLinkII
üü
üü
üü
üü
üü
FieldCommuni­cator
16
MicroMotion9739MVDTransmitters
Page 23
Beforeyoubegin
Table1-39739MVDtransmitterconfigurationsettings
Configuration parameterSettingDisplay
Action:
qNone(default)
qStartSensorZero
qStart/stopalltotalizers
qResetmasstotal
qResetvolumetotal
qResetgasstandardvolume
total
qResetalltotals
qResettemperature-corrected
volumetotal
qResetCMreferencevolume
total
qResetCMnetmasstotal
qResetCMnetvolumetotal
qIncrementCMmatrix
EventType:
qHI
qLO
qIN
DiscreteEvent5
qOUT
ProcessVariable:
SetpointA:
SetpointB:
continued
Configurablewith:
ProLinkII
üü
üü
üü
üü
üü
FieldCommuni­cator
ConfigurationandUseManual
17
Page 24
Beforeyoubegin
Table1-39739MVDtransmitterconfigurationsettings
Configuration parameterSettingDisplay
Action:
qNone(default)
qStartSensorZero
qStart/stopalltotalizers
qResetmasstotal
qResetvolumetotal
qResetgasstandardvolume
total
qResetalltotals
qResettemperature-corrected
volumetotal
qResetCMreferencevolume
total
qResetCMnetmasstotal
qResetCMnetvolumetotal
qIncrementCMmatrix
Pollingcontrol:
qNone
qPollAsPrimary
qPollAsSecondary
Polledvariable1
ExternalTag:
Variabletype:
qExternalpressure
qExternaltemperature
qNone
Pollingcontrol:
qNone
qPollAsPrimary
qPollAsSecondary
Polledvariable2
ExternalTag:
Variabletype:
qExternalpressure
qExternaltemperature
qNone
continued
Configurablewith:
ProLinkII
üü
üü
üü
üü
üü
üü
üü
FieldCommuni­cator
18
MicroMotion9739MVDTransmitters
Page 25
Chapter2
Quickstartwiththedisplay
Topicscoveredinthischapter:
♦Applypower ♦Congurationtipsandtricks ♦ConguretheprimarymAoutputtoreportmassowrateinauser-selected
measurementunit ♦Performalooptest ♦Zerotheowmeter

2.1Applypower

Prerequisites
Beforeyouapplypowertotheowmeter,closeandtightenallhousingcovers.
Topreventignitionofammableorcombustibleatmospheres,makesureallcoversare tightlyclosed.Forhazardousareainstallations,applyingpowertotheunitwhilehousing coversareremovedorloosecancauseanexplosion.
Procedure
Turnontheelectricalpoweratthepowersupply.
Theowmeterwillautomaticallyperformdiagnosticroutines.Fortransmitterswithadisplay,thestatus
LEDwillturngreenandbegintoashwhenthestartupdiagnosticsarecomplete.IfthestatusLED
exhibitsdifferentbehavior,analarmconditionispresent.
Postrequisites
Allowtheelectronicstowarmupforapproximately10minutesbeforerelyingonprocessmeasurements.
Althoughthesensorisreadytoreceiveprocessuidshortlyafterpower-up,theelectronicscantakeup
to10minutestowarmupcompletely.

2.2Configurationtipsandtricks

Reviewthesetipsbeforebeginningconguration.
2.2.1AccesstoOFFLINEmenu
AccesstotheOFFLINEmenumaybedisabled.Tocongurethetransmitterusingthedisplay,youmust
enableaccesstotheOFFLINEmenu.
ConfigurationandUseManual
19
Page 26
Quickstartwiththedisplay
2.2.2Defaultvaluesandranges
DefaultvaluesandrangesforthemostcommonlyusedparametersareprovidedinAppendixA.
2.3ConfiguretheprimarymAoutputtoreportmassflow
rateinauser-selectedmeasurementunit
Thisprocedureshowsyouhowtoperformthesetasksusingthedisplay.Forallotherconguration
tasks,includingotheroptionsforthemAoutput,seethecongurationsectionsofthismanual.
Note
Thisprocedureassumesthatyouarestartingfromthefactory-defaultconguration.
Procedure
1.Navigatetothecongurationmenu.
a.Atthetransmitterdisplay,activatetheScrollandSelectopticalswitchessimultaneouslyuntil
SEEALARMappearsonthedisplay .
b.ActivateScrollrepeatedlyuntilOFF-LINEMAINTappearsonthedisplay,thenactivateSelect.
c.ActivateScroll-Select-Scroll.
Thisoperatorsequenceisasafetyprecaution,designedtoprotectthetransmitterfrom accidentalcongurationchangescausedbyunintentionalactivationoftheoff-linemenu.
Important
Ifyouhaveenabledadisplaypassword,theScroll-Select-Scrolloperatorsequenceisdisabled. Youarerequiredtoenterapasswordbeforeyoucancontinue.Thedefaultpasswordis1234.
d.ActivateScrolluntilOFF-LINECONFGappearsonthedisplay,thenactivateSelect.
2.SetMassFlowMeasurementUnitasdesired.
a.WhenCONFIGUNITSappearsonthedisplay,activateSelect.
b.WhenUNITSMASSappearsonthedisplay,activateSelect.
c.ActivateScrolltoviewtheoptionsforMassFlowMeasurementUnit.Whenyouseethe
measurementunityouwanttouse,activateSelect. activateSelect.
d.ActivateScrolluntilUNITSEXITappearsonthedisplay ,thenactivateSelect.
3.SetmAOutputProcessVariabletoMassFlowRate.
a.ActivateScrolluntilCONFIGIOappearsonthedisplay ,thenactivateSelect.
IfSTORE/YES?ashesonthedisplay ,
b.WhenAO1appearsonthedisplay,activateSelect.
c.WhenAO1SRCappearsonthedisplay,activateSelect.
d.ActivateScrolltoviewtheoptionsformAOutputProcessVariableoptions.WhenyouseeMFLOW,
activateSelect.IfSTORE/YES?ashesonthedisplay,activateSelect.
4.SetLowerRangeValue(LRV).
20MicroMotion9739MVDTransmitters
Page 27
Quickstartwiththedisplay
LowerRangeValuespeciesthevalueofMassFlowRatetoberepresentedbyanoutputlevelof 0mAor4mA.
a.ActivateScrolluntilAO14mAorAO10mAappearsonthedisplay,thenactivateSelect.
b.DeneeachcharacterinLowerRangeV alue,includingthesign.
UseSelecttohighlightaspeciccharacter.UseScrolltosetthevalueofthecharacter.
c.Whenyouhavesetallcharactersasdesired,activateScrollandSelectsimultaneouslyuntil
SAVE/YES?ashesonthedisplay,thenactivateSelecttowritethevaluetotransmittermemory .
5.SetUpperRangeValue(URV).
UpperRangeValuespeciesthevalueofMassFlowRatetoberepresentedbyanoutputlevelof20mA.
a.ActivateScrolluntilAO120mAappearsonthedisplay ,thenactivateSelect.
b.DeneeachcharacterinUpperRangeV alue,includingthesign.
UseSelecttohighlightaspeciccharacter.UseScrolltosetthevalueofthecharacter.
c.Whenyouhavesetallcharactersasdesired,activateScrollandSelectsimultaneouslyuntil
SAVE/YES?ashesonthedisplay,thenactivateSelecttowritethevaluetotransmittermemory .
6.ActivateScrolluntilAOEXITappearsonthedisplay,thenactivateSelect.
7.Returnthedisplaytonormaloperation(displayingprocessdata).
a.ActivateScrolluntilIOEXITappearsonthedisplay,thenactivateSelect.
b.ActivateScrolluntilCONFIGEXITappearsonthedisplay ,thenactivateSelect.
c.ActivateScrolluntilOFF-LINEEXITappearsonthedisplay,thenactivateSelect.
d.ActivateScrolluntilEXITappearsonthedisplay,thenactivateSelect.
Tip
Anotherwaytoexittheoff-linemenuistorelyonthetime-outfeature.Ifyouhavenotused theoff-linemenuforapproximately60seconds,thedisplaywillautomaticallyreturntonormal operation.

2.4Performalooptest

Alooptestisawaytoverifythatthetransmitterandthereceivingdevicearecommunicatingproperly.
AlooptestalsohelpsyouknowwhetheryouneedtotrimmAoutputs.Performingalooptestisnot
arequiredprocedure.However,
outputavailableonyourtransmitter.
Procedure
MicroMotionrecommendsperformingalooptestforeveryinputor
1.TestthemAoutput.
a.ChooseOFFLINEMAINT→SIM→AO1SIM,andselectSET4MAoranothermAoutputvalue.
b.ReadthemAcurrentatthereceivingdeviceandcompareittothetransmitteroutput.
ConfigurationandUseManual
Dotstraversethedisplaywhiletheoutputisxed.
Thereadingsdonotneedtomatchexactly .Ifthevaluesareslightlydifferent,youcancorrect thediscrepancybytrimmingtheoutput.
21
Page 28
Quickstartwiththedisplay
c.Atthetransmitter,activateSelect.
d.ScrolltoandselectSET20MA.
Dotstraversethedisplaywhiletheoutputisxed.
e.ReadthemAcurrentatthereceivingdeviceandcompareittothetransmitteroutput.
Thereadingsdonotneedtomatchexactly .Ifthevaluesareslightlydifferent,youcancorrect thediscrepancybytrimmingtheoutput.
f.Atthetransmitter,activateSelect.
2.TestthesecondarymAoutput.
ChooseOFFLINEMAINT→SIM→AO2SIM,andrepeatthelooptestforthesecondarymAoutput.
3.Testthefrequencyoutput.
a.ChooseOFFLINEMAINT→SIM→FOSIM,andselectthefrequencyoutputvalue.
Thefrequencyoutputcanbesetto1,10,or15kHz.
Note
IftheWeights&Measuresapplicationisenabledonthetransmitter,itisnotpossibleto performalooptestofthefrequencyoutput,evenwhenthetransmitterisunsecured.
Dotstraversethedisplaywhiletheoutputisxed.
b.Readthefrequencysignalatthereceivingdeviceandcompareittothetransmitteroutput.
c.Atthetransmitter,activateSelect.
4.Testthediscreteoutput.
a.ChooseOFFLINEMAINT→SIM→DOSIM,andselectSETON.
Dotstraversethedisplaywhiletheoutputisxed.
b.Verifythesignalstateatthereceivingdevice.
c.Atthetransmitter,activateSelect.
d.ScrolltoandselectSETOFF.
e.Verifythesignalstateatthereceivingdevice.
f.Atthetransmitter,activateSelect.
5.Readthediscreteinput.
a.Settheremoteinputdevicesothatthedesiredsignalissenttothetransmitter.
b.Atthetransmitter,chooseOFFLINEMAINT→SIM,andselectREADDI.
c.Verifythesignalstateatthetransmitter.
d.Repeattheprocedurefortheothersignalstate.
6.ReadthemAinput.
a.Settheremoteinputdevicesothatthedesiredcurrentissenttothetransmitter.
b.Atthetransmitter,chooseOFFLINEMAINT→SIM,andselectREADMAI.
22MicroMotion9739MVDTransmitters
Page 29
Quickstartwiththedisplay
c.Verifythecurrentvalue.
Postrequisites
•IfthemAoutputreadingswereslightlyoffatthereceivingdevice,youcancorrectthisdiscrepancy bytrimmingtheoutput.
•IfthemAoutputreadingwassignicantlyoff(±200microamps),orifatanystepthereadingwas faulty,verifythewiringbetweenthetransmitterandtheremotedevice,andtryagain.
•IfthemAinputreadingwasslightlyoffatthetransmitter,trimandcalibratetheinputatthe remoteinputdevice.

2.5Zerotheflowmeter

Zeroingtheowmeterestablishestheowmeter’spointofreferencewhenthereisnoow.
Prerequisites
Toprepareforthezeroprocedure:
1.Allowtheowmetertowarmupforatleast20minutesafterapplyingpower.
2.Runtheprocessuidthroughthesensoruntilthesensortemperaturereachesthenormalprocess operatingtemperature.
3.Stopowthroughthesensorbyshuttingavalvedownstreamfromthesensor .
4.Ensurethatowhascompletelystoppedthroughthesensor,andthatthesensoriscompletely fullofprocessuid.
5.Checktheowrate.Iftheowrateisclosetozero,youshouldnotneedtozerotheowmeter.
Important
Themeterwaszeroedatthefactory,andshouldnotrequireaeldzero.
Note
Donotzerotheowmeterifahigh-severityalarmisactive.Correcttheproblem,thenzerothe owmeter.Youmayzerotheowmeterifalow-severityalarmisactive.
Procedure
1.InitiateowmeterzerobychoosingOFFLINEMAINT→ZERO→CALZERO,andselectCAL/YES?.
Dotstraversethedisplaywhileowmeterzeroisinprogress.
2.Readthezeroresultonthedisplay.
ThedisplaywillreportCALPASSifthezerowassuccessful,orCALFAILifitwasnot.Ifthezerofails, restorethefactoryzero(ifavailable).
2.5.1Restorefactoryzero
Procedure
Restorethefactoryzerowiththedisplay.
ConfigurationandUseManual
23
Page 30
Quickstartwiththedisplay
OFFLINEMAINT→RESTOREZERO→RESTORE/YES?
24MicroMotion9739MVDTransmitters
Page 31
Chapter3
QuickstartwithProLinkII
Topicscoveredinthischapter:
♦Applypower ♦Connectwith ♦Congurationtipsandtricks ♦ConguretheprimarymAoutputtoreportmassowrateinauser-selected
measurementunit ♦Performalooptest ♦TrimmAoutputs ♦Zerotheowmeter ♦T estortunethesystemusingsensorsimulation ♦Backuptransmitterconguration ♦Enable/disableHARTsecurity

3.1Applypower

Prerequisites
Beforeyouapplypowertotheowmeter,closeandtightenallhousingcovers.
ProLinkII
Topreventignitionofammableorcombustibleatmospheres,makesureallcoversare tightlyclosed.Forhazardousareainstallations,applyingpowertotheunitwhilehousing coversareremovedorloosecancauseanexplosion.
Procedure
Turnontheelectricalpoweratthepowersupply.
Theowmeterwillautomaticallyperformdiagnosticroutines.Fortransmitterswithadisplay,thestatus
LEDwillturngreenandbegintoashwhenthestartupdiagnosticsarecomplete.IfthestatusLED
exhibitsdifferentbehavior,analarmconditionispresent.
Postrequisites
Allowtheelectronicstowarmupforapproximately10minutesbeforerelyingonprocessmeasurements.
Althoughthesensorisreadytoreceiveprocessuidshortlyafterpower-up,theelectronicscantakeup
to10minutestowarmupcompletely.

3.2ConnectwithProLinkII

AconnectionfromProLinkIItoyourtransmitterallowsyoutoreadprocessdata,congurethe
transmitter,andperformmaintenanceandtroubleshootingtasks.
ConfigurationandUseManual
25
Page 32
QuickstartwithProLinkII
Prerequisites
YoumusthavethefollowingversionofProLinkIIinstalledonyourcomputer:v2.91orlater.
3.2.1ProLinkIIconnectiontypes
The9739MVDtransmitterhasseveralconnectionsoptionsforcommunicatingviaProLinkII.Y ou
chooseaconnectiontypebasedonwhatyouneedtoaccomplishwiththetransmitterandthedigital
communicationsyouareusing.
The
9739MVDtransmittersupportsthefollowingProLinkIIconnectiontypes:
•Serviceportconnections
•HART/Bell202connections
•HART/RS-485connections
•Modbus/RS-4857-bitconnections(ModbusASCII)
•Modbus/RS-4858-bitconnections(ModbusRTU)
Whenselectingaconnectiontype,considerthefollowing:
•Serviceportconnectionsusestandardconnectionparametersthatarealreadydenedin ProLinkII,andthereforeyoudonothavetocongurethem.
•HART/Bell202connectionsusestandardHARTconnectionparametersthatarealreadydenedin ProLinkII.Theonlyparameteryoumustcongureisthetransmitteraddress.
•Theserviceportterminals(AandB)andtheRS-485terminals(26and27)usethesameinternal wiring.IfyouhavewiredthetransmitterforRS-485digitalcommunications,youcannotmake aserviceportconnection.
•Serviceportconnectionsrequireaccesstotheserviceportterminals,whicharelocatedonthe transmitterdisplayandonlyaccessibleafterremovingthehousingcover.Accordingly,service portconnectionsshouldbeusedonlyfortemporaryconnections,andmayrequireextrasafety precautions.
•Modbusconnections,includingserviceportconnections,aretypicallyfasterthanHART connections.
•WhenyouareusingaHARTconnection, windowatatime.Thisisdonetomanagenetworktrafcandoptimizespeed.
•Youcannotmakeaserviceportconnectionwhileanactiveconnectionexistsbetweenthe transmitterandModbushost.
ProLinkIIwillnotallowyoutoopenmorethanone
26MicroMotion9739MVDTransmitters
Page 33
QuickstartwithProLinkII
3.2.2Makeaserviceportconnection
Prerequisites
Youwillneedthefollowing:
•Aninstalled,licensedcopyofProLinkII
•AnavailableserialorUSBport
•Oneofthefollowingtypesofsignalconverters:
−RS-232toRS-485signalconverter
−USBtoRS-485signalconverter
•Adaptersasrequired(e.g.9-pinto25-pin)
Important
TheSP(ServicePort)clipsonthedisplayofthe theRS-485terminals26and27ofthetransmitter.IfyouhavewiredthetransmitterforRS-485 digitalcommunications,youmustdirectlyconnecttothetransmitterusingtheRS-485terminalblock connectionsordisconnecttheRS-485terminalconnectionstousetheServicePortconnections.
9739MVDtransmitteraredirectlyconnectedto
Procedure
1.AttachthesignalconvertertoyourPC’sserialorUSBport.
2.Atthetransmitter,removethehousingcovertoaccessthetransmitterdisplay.
Ifthetransmitterisinahazardousarea,donotremovethehousingcoverwhile powerisbeingsuppliedtotheunit.Removingthehousingcoverwhilepoweris suppliedtotheunitcouldcauseanexplosion.T oaccesstransmitterinformationina hazardousenvironment,useacommunicationmethodthatdoesnotrequireremoving thetransmitterhousingcover.
3.ConnecttheleadsonyoursignalconvertertotheSP(ServicePort)clips(AandB)ontheface ofthetransmitter.SeeFigure3-1.
Tip
Usually,butnotalways,theblackleadisRS-485/AandtheredleadisRS-485/B.
ConfigurationandUseManual
27
Page 34
QuickstartwithProLinkII
A
B
C
D
Figure3-1ProLinkIIconnectiontoserviceport
APC
B25-to-9pinadapter,ifnecessary;orRS-232toUSBadapter,ifnecessary
CRS-232toRS-485signalconverter
DTransmitter
Notes
•Thisgureshowsaserialportconnection.AUSBportconnectionisalsoavailable.
4.StartProLinkII.
5.ChooseConnection→ConnecttoDevice.
6.SetProtocoltoServicePort.
Tip
Serviceportconnectionsusestandardconnectionparametersandastandardaddress.Y oudonot needtocongurethemhere.
7.SettheCOMPortvaluetothePCCOMportthatyouareusingforthisconnection.
8.ClickConnect.
Iftheconnectionissuccessful:
•Thestatusbarinthemainwindowisupdatedtoshowanactiveconnection.
•TheProcessVariableswindoworCommissioningWizardwindowisdisplayed.
28MicroMotion9739MVDTransmitters
Page 35
Ifanerrormessageappears:
•Switchtheleadsandtryagain.
•EnsurethatyouareusingthecorrectCOMport.
•CheckthephysicalconnectionbetweenthePCandthetransmitter.
3.2.3MakeaHART/Bell202connection
Prerequisites
Youwillneedthefollowing:
•Aninstalled,licensedcopyofProLinkII
•AnavailableserialorUSBport
•Oneofthefollowingtypesofsignalconverters:
−RS-232toBell202signalconverter
−USBtoBell202signalconverter
QuickstartwithProLinkII
•Adaptersasrequired(e.g.9-pinto25-pin)
Important
IftheHARTsecurityswitchissettoON,HARTprotocolcannotbeusedtoperformanyactionthat requireswritingtothetransmitter.Forexample,youcannotchangetheconguration,resettotalizers, orperformcalibrationusingthe HARTsecurityswitchissettoOFF,nofunctionsaredisabled.
YoucanconnectProLinkIItotheHARTclipsonthetransmitter,toanypointinalocalHARTloop,orto anypointinaHARTmultidropnetwork.
Ifthetransmitterisinahazardousarea,donotremovethehousingcoverwhilepoweris beingsuppliedtotheunit.Removingthehousingcoverwhilepowerissuppliedtotheunit couldcauseanexplosion.Toaccesstransmitterinformationinahazardousenvironment, useacommunicationmethodthatdoesnotrequireremovingthetransmitterhousingcover.
FieldCommunicatororProLinkIIwithaHARTconnection.Whenthe
Procedure
1.ToconnecttotheHARTclips:
a.Removethetransmitterhousingcover.
b.AttachtheleadsfromthesignalconvertertotheHARTclipsonthefaceofthetransmitter
andaddresistanceasnecessary .SeeFigure3-2.
ConfigurationandUseManual
ProLinkIImustbeconnectedacrossaresistanceof250–600Ω.
Tip
HARTconnectionsarepolarity-insensitive.Itdoesnotmatterwhichleadyouattachtowhich terminal.
29
Page 36
QuickstartwithProLinkII
A
B
D
C
Figure3-2ProLinkIIconnectiontoHARTclips
APC
BHARTinterface
C250–600Ωresistance
DTransmitter
Notes
•Thisgureshowsaserialportconnection.AUSBportconnectionisalsoavailable.
2.ToconnecttoapointinthelocalHARTloop,attachtheleadsfromthesignalconvertertoany pointintheloopandaddresistanceasnecessary.SeeFigure3-3.
ProLinkIImustbeconnectedacrossaresistanceof250–600Ω.
30MicroMotion9739MVDTransmitters
Page 37
Figure3-3ProLinkIIconnectiontolocalHARTloop
A
C
D
B
R1
R3
R3
E
APC
BHARTinterface
CAnycombinationofresistorsR1,R2,andR3asnecessarytomeetHARTcommunicationresistance
requirements
DDCSorPLC
ETransmitter
QuickstartwithProLinkII
Notes
•Thisgureshowsaserialportconnection.AUSBportconnectionisalsoavailable.
3.ToconnecttoapointintheHARTmultidropnetwork,attachtheleadsfromthesignalconverterto anypointintheloopandaddresistanceasnecessary.SeeFigure3-4.
ProLinkIImustbeconnectedacrossaresistanceof250–600Ω.
ConfigurationandUseManual
31
Page 38
QuickstartwithProLinkII
B
C
D
A
Figure3-4ProLinkIIconnectiontomultidropnetwork
AHARTinterface(toPC)
B250–600Ωresistance
CDevicesonthenetwork
DMasterdevice
Notes
•Thisgureshowsaserialportconnection.AUSBportconnectionisalsoavailable.
4.StartProLinkII.
5.ChooseConnection→ConnecttoDevice.
6.SetProtocoltoHARTBell202.
Tip
HART/Bell202connectionsusestandardconnectionparameters.Youdonotneedtocongure themhere.
7.IfyouareusingaUSBportconnection,enableConverterTogglesRTS.
8.SetAddress/TagtotheHARTpollingaddressconguredinthetransmitter.
32MicroMotion9739MVDTransmitters
Page 39
QuickstartwithProLinkII
Tips
•Ifthisisthersttimeyouareconnectingtothetransmitter,usethedefaultaddress:0.
•IfyouarenotinaHARTmultidropenvironment,theHARTpollingaddressistypicallyleft atthedefaultvalue.
•Ifyouareunsureofthetransmitter’saddress,clickPoll.ProLinkIIwillsearchthenetworkand returnalistofthetransmittersthatitdetects.
9.SettheCOMPortvaluetothePCCOMportthatyouareusingforthisconnection.
10.SetMasterasappropriate.
OptionDescription
SecondaryUsethissettingifanotherHARThostsuchas
aDCSisonthenetwork.
PrimaryUsethissettingifnootherhostisonthe
network.TheFieldCommunicatorisnota host.
11.ClickConnect.
Iftheconnectionissuccessful:
•Thestatusbarinthemainwindowisupdatedtoshowanactiveconnection.
•TheProcessVariableswindoworCommissioningWizardwindowisdisplayed.
Ifanerrormessageappears:
•EnsurethatyouareusingthecorrectCOMport.
•VerifytheHARTpollingaddress.
•CheckthephysicalconnectionbetweenthePCandthetransmitter.
•Increaseordecreaseresistance.
3.2.4MakeaHART/RS-485connection
Prerequisites
Youwillneedthefollowing:
•Aninstalled,licensedcopyofProLinkII
•AnavailableserialorUSBport
•Oneofthefollowingtypesofsignalconverters:
−RS-232toRS-485signalconverter
−USBtoRS-485signalconverter
•Adaptersasrequired(e.g.9-pinto25-pin)
ConfigurationandUseManual
33
Page 40
QuickstartwithProLinkII
A
B
C
D
Important
IftheHARTsecurityswitchissettoON,HARTprotocolcannotbeusedtoperformanyactionthat requireswritingtothetransmitter.Forexample,youcannotchangetheconguration,resettotalizers, orperformcalibrationusingthe
FieldCommunicatororProLinkIIwithaHARTconnection.Whenthe
HARTsecurityswitchissettoOFF,nofunctionsaredisabled.
Procedure
1.AttachthesignalconvertertoyourPC’sserialorUSBport.
2.Atthetransmitter,removethehousingcovertoaccesstheRS-485terminalconnections.
Ifthetransmitterisinahazardousarea,donotremovethehousingcoverwhile powerisbeingsuppliedtotheunit.Removingthehousingcoverwhilepoweris suppliedtotheunitcouldcauseanexplosion.T oaccesstransmitterinformationina hazardousenvironment,useacommunicationmethodthatdoesnotrequireremoving thetransmitterhousingcover.
3.Toconnectdirectlytothetransmitter,connecttheleadsonyoursignalconvertertoterminals26 (RS-485/A)and27(RS-485/B)onyourtransmitter.SeeFigure3-5.
Figure3-5ProLinkIIconnectiontotransmitterterminals
APC
B25-to-9pinadapter,ifnecessary
CRS-485toRS-232signalconverter
DTransmitter
4.ToconnecttoapointintheRS-485network,connecttheleadsonyoursignalconvertertoany pointinthenetworkandaddresistanceasnecessary.SeeFigure3-6.
34
MicroMotion9739MVDTransmitters
Page 41
Figure3-6ProLinkIIconnectiontoanRS-485networkusingHART
A
B
C
D
E
F
APC
B25-to-9pinadapter,ifnecessary
CRS-485toRS-232signalconverter
D120-Ω,1/2-wattterminatingresistorsatbothendsofthesegment,ifnecessary
EDCSorPLC(Auto-detectcommunication)
FTransmitter
QuickstartwithProLinkII
Notes
•Thisgureshowsaserialportconnection.AUSBportconnectionisalsoavailable.
5.Toconnecttoapointinamultidropnetwork,attachtheleadsfromthesignalconvertertoany pointinthewire.SeeFigure3-7.
ConfigurationandUseManual
35
Page 42
QuickstartwithProLinkII
B
C
A
Figure3-7ProLinkIIconnectiontoamultidropnetwork
AHARTinterface(toPC)
BDevicesonthenetwork
CMasterdevice
6.StartProLinkII.
7.ChooseConnection→ConnecttoDevice.
8.Ifnecessary,settheconnectionparameterstomatchtheHART/RS-485parametersconguredin yourtransmitter.
Tominimizecongurationrequirements,the9739MVDtransmitterusesanauto-detectionscheme whenrespondingtoaconnectionrequest.Thetransmitterwillacceptallconnectionrequests withinthelimitsdescribedinTable3-1.
Table3-1Auto-detectionlimits
ParameterOption
Protocol
AddressRespondsto:
Baudrate
Stopbits
ParityEven,odd,none
HART ,ModbusASCII,ModbusRTU
•Serviceportaddress(11 1)
•ConguredHARTaddress(default=0)
•ConguredModbusaddress(default=1)
Standardratesbetween1200and38,400
0,1
9.SettheCOMPortvaluetothePCCOMportthatyouareusingforthisconnection.
36
MicroMotion9739MVDTransmitters
Page 43
10.SetMasterasappropriate:
OptionDescription
SecondaryUsethissettingifanotherhostsuchasaDCS
isonthenetwork.
PrimaryUsethissettingifnootherhostisonthe
network.TheFieldCommunicatorisnota host.
11.ClickConnect.
Iftheconnectionissuccessful:
•Thestatusbarinthemainwindowisupdatedtoshowanactiveconnection.
•TheProcessVariableswindoworCommissioningWizardwindowisdisplayed.
Ifanerrormessageappears:
•EnsurethatyouareusingthecorrectCOMport.
•CheckthephysicalconnectionbetweenthePCandthetransmitter.
QuickstartwithProLinkII
•Forlong-distancecommunication,orifnoisefromanexternalsourceinterfereswiththe signal,install120-Ω½-Wterminatingresistorsinparallelwiththeoutputatbothendsofthe communicationsegment.
3.2.5MakeaModbus/RS-485connection
Prerequisites
Youwillneedthefollowing:
•Aninstalled,licensedcopyofProLinkII
•AnavailableserialorUSBport
•Oneofthefollowingtypesofsignalconverters:
−RS-232toRS-485signalconverter
−USBtoRS-485signalconverter
•Adaptersasrequired(e.g.9-pinto25-pin)
Procedure
1.AttachthesignalconvertertoyourPC’sserialorUSBport.
2.Atthetransmitter,removethehousingcovertoaccesstheRS-485terminalconnections.
Ifthetransmitterisinahazardousarea,donotremovethehousingcoverwhile powerisbeingsuppliedtotheunit.Removingthehousingcoverwhilepoweris suppliedtotheunitcouldcauseanexplosion.T oaccesstransmitterinformationina hazardousenvironment,useacommunicationmethodthatdoesnotrequireremoving thetransmitterhousingcover.
3.Toconnectdirectlytothetransmitter,connecttheleadsonyoursignalconvertertoterminals26 (RS-485/A)and27(RS-485/B)onyourtransmitter.SeeFigure3-8.
ConfigurationandUseManual
37
Page 44
QuickstartwithProLinkII
A
B
C
D
Tip
Usually,butnotalways,theblackleadisRS-485/AandtheredleadisRS-485/B.
Figure3-8ProLinkIIconnectiontotransmitterterminals
APC
B25-to-9pinadapter,ifnecessary
CRS-485toRS-232signalconverter
DTransmitter
4.ToconnecttoapointinRS-485network,connecttheleadsonyoursignalconvertertoanypointin thenetworkandaddresistanceasnecessary.SeeFigure3-9.
Restriction
TheModbushostmustnotbecommunicatingwiththetransmitterwhenyoumaketheProLinkII connection.T omaketheconnection,waituntilthehostcommunicationiscomplete,orterminate thehostconnection.
38
MicroMotion9739MVDTransmitters
Page 45
Figure3-9ProLinkIIconnectiontoRS-485network
A
B
C
D
E
F
APC
B25-to-9pinadapter,ifnecessary
CRS-485toRS-232signalconverter
D120-Ω,1/2-wattterminatingresistorsatbothendsofthesegment,ifnecessary
EDCSorPLC(mustnotbecommunicatingwiththetransmitterduringtheProLinkIIconnection)
FTransmitter
QuickstartwithProLinkII
5.StartProLinkII.
6.ChooseConnection→ConnecttoDevice.
7.Ifnecessary,settheconnectionparameterstomatchtheModbus/RS-485parameterscongured inyourtransmitter.
Tominimizecongurationrequirements,thetransmitterusesanauto-detectionschemewhen respondingtoaconnectionrequest.Thetransmitterwillacceptallconnectionrequestswithinthe auto-detectionlimits(seeT able3-2).
Table3-2Auto-detectionlimits
ParameterOption
Protocol
AddressRespondsto:
Baudrate
Stopbits
Parity
HART,ModbusASCII,ModbusRTU
•Serviceportaddress(11 1)
•ConguredHARTAddress(default=0)
•ConguredModbusAddress(default=1)
Standardratesbetween1200and38,400
0,1
Even,Odd,None
ConfigurationandUseManual
39
Page 46
QuickstartwithProLinkII
Tip
Ifyoudonotknowthetransmitter’sRS-485communicationsettings,youcanconnectthroughthe serviceport,whichalwaysusesdefaultsettings,oruseanothercommunicationstooltoviewor changethesettings.
8.SettheCOMPortvaluetothePCCOMportthatyouareusingforthisconnection.
9.ClickConnect.
Iftheconnectionissuccessful:
•Thestatusbarinthemainwindowisupdatedtoshowanactiveconnection.
•TheProcessVariableswindoworCommissioningWizardwindowisdisplayed.
Ifanerrormessageappears:
•Switchtheleadsandtryagain.
•EnsurethatyouareusingthecorrectCOMport.
•CheckthephysicalconnectionbetweenthePCandthetransmitter.
•Forlong-distancecommunication,orifnoisefromanexternalsourceinterfereswiththe signal,install120-Ω½-Wterminatingresistorsinparallelwiththeoutputatbothendsofthe communicationsegment.

3.3Configurationtipsandtricks

Reviewthesetipsbeforebeginningconguration.
3.3.1HARTsecurity
HARTsecuritymaybeenabledonthe9739MVDtransmitter.T ocongurethetransmitterusingHART protocol,youmustdisableHARTsecurity.
3.3.2Defaultvaluesandranges
DefaultvaluesandrangesforthemostcommonlyusedparametersareprovidedinAppendixA.
3.3.3Restorethefactoryconfiguration
Restoringthefactorycongurationreturnsthetransmittertoaknownoperationalconguration.
Procedure
1.MakeaconnectionfromProLinkIItoyourtransmitter.
2.ChooseProLink→Conguration→Device→RestoreFactoryConguration.
3.IntheCongurationwindow,clicktheDevicetab.
40MicroMotion9739MVDTransmitters
Page 47
QuickstartwithProLinkII
4.ClickRestoreFactoryConguration.
5.ClickOK.
3.4ConfiguretheprimarymAoutputtoreportmassflow rateinauser-selectedmeasurementunit
ThisprocedureshowsyouhowtoperformthesetasksusingProLinkII.Forallothercongurationtasks, includingotheroptionsforthemAoutput,seethecongurationsectionsofthismanual.
Note
Thisprocedureassumesthatyouarestartingfromthefactory-defaultconguration.
Procedure
1.StartProLinkIIandconnecttoyourtransmitter.
2.Setthemeasurementunitformassowrate.
a.ChooseProLink→Conguration.
b.IntheCongurationwindow,clicktheFlowtab.
c.SelectameasurementunitfromtheMassFlowUnitslist,thenclickApply.
3.CongurethemAoutput.
a.IntheCongurationwindow,clicktheAnalogOutputtab.
b.SelectMassFlowRatefromthePrimaryV ariableIslist.
c.EnterappropriatevaluesforLowerRangeValue(LRV)andUpperRangeValue(URV).
LowerRangeValuespeciesthevalueofMassFlowRatetoberepresentedbyanoutputlevelof 0mAor4mA. anoutputlevelof20mA.
4.ClickOKtoapplythechangesandclosetheCongurationwindow.
5.(Optional)ChooseProLink→OutputLevelsandobservethemAoutputreading.
Itshouldvarybetween0mAor4mAand20mAaccordingtothemassowrateofyourprocess.
UpperRangeValuespeciesthevalueofMassFlowRatetoberepresentedby

3.5Performalooptest

9739MVDTransmitterProLink→Test→FixMilliamp1
ProLink→Test→FixMilliamp2 ProLink→Test→FixFreqOut ProLink→Test→FixDiscreteOut ProLink→Test→ReadDiscreteInput ProLink→Test→ReadMilliampInput
Alooptestisawaytoverifythatthetransmitterandthereceivingdevicearecommunicatingproperly.A looptestalsohelpsyouknowwhetheryouneedtotrimmAoutputs.
Alooptestisawaytoverifythatthetransmitterandthereceivingdevicearecommunicatingproperly. AlooptestalsohelpsyouknowwhetheryouneedtotrimmAoutputs.Performingalooptestisnot
ConfigurationandUseManual
41
Page 48
QuickstartwithProLinkII
arequiredprocedure.However,MicroMotionrecommendsperformingalooptestforeveryinputor outputavailableonyourtransmitter.
Procedure
1.TestthemAoutput.
a.ChooseProLink→Test→FixMilliamp1.
b.Enter0mAor4mAinSetOutputTo.ClickFixmA.
c.ReadthemAcurrentatthereceivingdeviceandcompareittothetransmitteroutput.
Thereadingsdonotneedtomatchexactly .Ifthevaluesareslightlydifferent,youcancorrect thediscrepancybytrimmingtheoutput.
d.ClickUnFixmA.
e.Enter20mAinSetOutputT o.ClickFixmA.
f.ReadthemAcurrentatthereceivingdeviceandcompareittothetransmitteroutput.
Thereadingsdonotneedtomatchexactly .Ifthevaluesareslightlydifferent,youcancorrect thediscrepancybytrimmingtheoutput.
g.ClickUnFixmA.
2.TestthesecondarymAoutput.
ChooseProLink→Test→FixMillamp2,andrepeatthelooptestforthesecondarymAoutput.
3.Testthefrequencyoutput.
Note
IftheWeights&Measuresapplicationisenabledonthetransmitter,itisnotpossibletoperforma looptestofthefrequencyoutput,evenwhenthetransmitterisunsecured.
a.ChooseProLink→Test→FixFreqOut.
b.EnterthefrequencyoutputvalueinSetOutputTo.ClickFixFrequency.
c.Readthefrequencysignalatthereceivingdeviceandcompareittothetransmitteroutput.
d.ClickUnFixFreq.
4.Testthediscreteoutput.
a.ChooseProLink→Test→FixDiscreteOutput.
b.SelectOn.
c.Verifythesignalstateatthereceivingdevice.
d.ClickUnFix.
e.SelectOff.
f.Verifythesignalstateatthereceivingdevice.
g.ClickUnFix.
5.Readthediscreteinput.
a.Settheremoteinputdevicesothatthedesiredsignalissenttothetransmitter.
b.ChooseProLink→Test→ReadDiscreteInput.
c.Verifythesignalstateatthetransmitter.
42MicroMotion9739MVDTransmitters
Page 49
QuickstartwithProLinkII
d.Repeattheprocedurefortheothersignalstate.
6.ReadthemAinput.
a.Settheremoteinputdevicesothatthedesiredsignalissenttothetransmitter.
b.ChooseProLink→Test→ReadMAInput.
Postrequisites
•IfthemAoutputreadingswereslightlyoffatthereceivingdevice,youcancorrectthisdiscrepancy bytrimmingtheoutput.
•IfthemAoutputreadingwassignicantlyoff(±200microamps),orifatanystepthereadingwas faulty,verifythewiringbetweenthetransmitterandtheremotedevice,andtryagain.
•IfthemAinputreadingwasslightlyoffatthetransmitter,trimandcalibratetheinputatthe remoteinputdevice.

3.6TrimmAoutputs

TrimmingthemAoutputestablishesacommonmeasurementrangebetweenthetransmitterandthe devicethatreceivesthemAoutput.
TrimmingthemAoutputsisnotarequiredprocedure.However,ifthereisasmalldiscrepancyinthemA readingbetweenthetransmitterandthereceivingdevice,trimmingtheoutputwillcorrectthis.
Note
Anytrimmingperformedontheoutputshouldnotexceed±200microamps.Ifmoretrimmingisrequired, contactMicroMotioncustomersupport.
Procedure
1.ChooseProLink→Calibration→Milliamp1TrimtostartthemAtrimprocedure.
2.FollowtheinstructionsintheguidedmethodtotrimthemAoutput.
3.ChooseProLink→Calibration→Milliamp2TrimtostartthetrimprocedureforthesecondarymAoutput.
4.FollowtheinstructionsintheguidedmethodtotrimthemAoutput.

3.7Zerotheflowmeter

Prerequisites
Toprepareforthezeroprocedure:
1.Allowtheowmetertowarmupforatleast20minutesafterapplyingpower.
2.Runtheprocessuidthroughthesensoruntilthesensortemperaturereachesthenormalprocess operatingtemperature.
3.Stopowthroughthesensorbyshuttingavalvedownstreamfromthesensor .
4.Ensurethatowhascompletelystoppedthroughthesensor,andthatthesensoriscompletely fullofprocessuid.
5.Checktheowrate.Iftheowrateisclosetozero,youshouldnotneedtozerotheowmeter.
ConfigurationandUseManual
43
Page 50
QuickstartwithProLinkII
Important
Themeterwaszeroedatthefactory,andshouldnotrequireaeldzero.
Note
Donotzerotheowmeterifahigh-severityalarmisactive.Correcttheproblem,thenzerothe owmeter.Youmayzerotheowmeterifalow-severityalarmisactive.
Procedure
1.ChooseProLink→Calibration→ZeroCalibration.
2.ModifyZeroTime,ifrequired.
ZeroTimecontrolstheamountoftimethetransmittertakestodetermineitszero-owreferencepoint.
•Alongzerotimemayproduceamoreaccuratezeroreference,butismorelikelytoresultin
azerofailure.Thisisduetotheincreasedpossibilityofnoisyow,whichcausesincorrect calibration.
•Ashortzerotimeislesslikelytoresultinazerofailurebutmayproducealessaccuratezero
reference.
ThedefaultZeroTimeis20seconds.Formostapplications,thedefaultZeroTimeisappropriate.
3.ClickPerformAutoZerotoinitiatethezeroprocedure.
TheCalibrationinProgresslightwillturnredduringthezeroprocedure.Attheendoftheprocedure:
•TheCalibrationinProgresslightwillreturntogreenifthezerowassuccessful.
•TheCalibrationFailurelightwillturnredifthezeroprocedurefailed.
4.Incaseoffailure,youhavetwooptions:
OptionDescription
RestorePriorZeroRestorePriorZerorestorestheowmetertothe
zerovalueithadjustpriortostartingthezero procedure.IfyouclosetheFlowCalibration window,youwillnolongerbeabletorestore thepriorzero.
RestoreFactoryZeroRestoreFactoryZeroisavailableonlyifyou
orderedatransmitterandasensortogether.

3.8Testortunethesystemusingsensorsimulation

Sensorsimulationallowsyoutosetspecicvaluesorvaluerangesformassow,density,and temperature.Thetransmitterwillreportthespeciedvaluesandtakeallappropriateactions,e.g.,apply acutoff,activateanevent,orpostanalarm.Y oucanusethisfeaturetotestthesystem’sresponsetoa varietyofprocessconditions,includingboundaryconditions,problemconditions,oralarmconditions,or totunetheloop.
44
MicroMotion9739MVDTransmitters
Page 51
QuickstartwithProLinkII
Whenyouenablesensorsimulation,thesimulatedvaluesarestoredinthesamememorylocations usedforprocessdatafromthesensor.Therefore,thesimulatedvalueswillbeusedthroughout transmitterfunctioning.Forexample,sensorsimulationwillaffect:
•Allmassow,temperature,ordensityvaluesshownonthedisplayorreportedviaoutputs ordigitalcommunications
•Themasstotalandmassinventoryvalues
•Allvolumecalculationsanddata,includingreportedvalues,volumetotals,andvolumeinventories
•Allmass,temperature,density,orvolumevaluesloggedtoDataLogger
Sensorsimulationdoesnotaffectanydiagnosticvalues.
Unlikeactualmassowanddensityvalues,thesimulatedvaluesarenottemperature-compensated (adjustedfortheeffectoftemperatureonthesensor’sowtubes).
Important
Donotenablesensorsimulationunlessyourprocesscantoleratetheeffectsofthesimulatedprocess values.
Procedure
1.ClickProLink→Conguration→SensorSimulation.
2.Enablesensorsimulation.
3.Formassow,setWaveFormasdesiredandentertherequiredvalues.
OptionRequiredvalues
Fixed
Sawtooth
Sine
4.Fordensity,setWaveFormasdesiredandentertherequiredvalues.
FixedValue
Period
Minimum
Maximum
Period
Minimum
Maximum
ConfigurationandUseManual
45
Page 52
QuickstartwithProLinkII
OptionRequiredvalues
Fixedvalue
Triangularwave
Sinewave
5.Fortemperature,setWaveFormasdesiredandentertherequiredvalues.
OptionRequiredvalues
Fixedvalue
Triangularwave
FixedValue
Period
Minimum
Maximum
Period
Minimum
Maximum
FixedValue
Period
Minimum
Maximum
Sinewave
6.Observethesystemresponsetothesimulatedvaluesandmakeanyappropriatechangesto thetransmittercongurationortothesystem.
7.Modifythesimulatedvaluesandrepeat.
8.Whenyouhavenishedtestingortuning,disablesensorsimulation.
Period
Minimum
Maximum

3.9Backuptransmitterconfiguration

ProLinkIIprovidesacongurationupload/downloadfunctionwhichallowsyoutosaveconguration setstoyourPC.Thisallowsyoutobackupandrestoreyourtransmitterconguration,andisalsoa convenientwaytoreplicateacongurationacrossmultipledevices.
Procedure
ChooseFile→LoadfromXmtrtoFiletosavethetransmittercongurationtoyourPC.
46
MicroMotion9739MVDTransmitters
Page 53
QuickstartwithProLinkII
A
B

3.10Enable/disableHARTsecurity

YouusetheHARTsecurityswitchlocatedonthetransmitterdisplaytodisablecongurationofthe transmitterusingHARTprotocol.WhentheHARTsecurityswitchissettoON,HARTprotocolcannotbe usedtoperformanyactionthatrequireswritingtothetransmitter.Forexample,youcannotchange theconguration,resettotalizers,performcalibration,etc.,usingthe withaHART/Bell202orHART/RS-485connection.WhentheHARTsecurityswitchissettoOFF,no functionsaredisabled.
Important
TheHARTsecurityswitchdoesnotaffectModbuscommunications.
Ifthetransmitterisinahazardousarea,donotremovethehousingcoverwhilepower isbeingsuppliedtotheunit.Removingthehousingcoverwhilepowerissuppliedto theunitcouldcauseanexplosion.ToaccesstheHARTsecurityswitchinahazardous environment,besuretoremovepowerfromthetransmitterbeforeremovingthehousing coverandsettingtheHARTsecurityswitch.
FieldCommunicatororProLinkII
Procedure
1.Removepowerfromthetransmitter.
2.Removethetransmitterhousingcover.
3.MovetheHARTsecurityswitchtothedesiredposition(seeFigure3-10).
Figure3-10HARTsecurityswitch(onblankdisplay)
AHARTsecurityswitch
BUnused
4.Replacethetransmitterhousingcover.
5.Restorepowertothetransmitter.
ConfigurationandUseManual
47
Page 54
Page 55
Chapter4
QuickstartwiththeFieldCommunicator
Topicscoveredinthischapter:
♦Applypower ♦Connectwiththe ♦Congurationtipsandtricks ♦ConguretheprimarymAoutputtoreportmassowrateinauser-selected
measurementunit ♦Performalooptest ♦TrimmAoutputs ♦Zerotheowmeter ♦T estortunethesystemusingsensorsimulation ♦Enable/disableHARTsecurity

4.1Applypower

Prerequisites
Beforeyouapplypowertotheowmeter,closeandtightenallhousingcovers.
FieldCommunicator
Topreventignitionofammableorcombustibleatmospheres,makesureallcoversare tightlyclosed.Forhazardousareainstallations,applyingpowertotheunitwhilehousing coversareremovedorloosecancauseanexplosion.
Procedure
Turnontheelectricalpoweratthepowersupply.
Theowmeterwillautomaticallyperformdiagnosticroutines.Fortransmitterswithadisplay,thestatus
LEDwillturngreenandbegintoashwhenthestartupdiagnosticsarecomplete.IfthestatusLED
exhibitsdifferentbehavior,analarmconditionispresent.
Postrequisites
Allowtheelectronicstowarmupforapproximately10minutesbeforerelyingonprocessmeasurements.
Althoughthesensorisreadytoreceiveprocessuidshortlyafterpower-up,theelectronicscantakeup
to10minutestowarmupcompletely.

4.2ConnectwiththeFieldCommunicator

Prerequisites
ThefollowingHARTdevicedescription(DD)mustbeinstalledontheFieldCommunicator:DDv2.
AconnectionfromtheFieldCommunicatortoyourtransmitterallowsyoutoreadprocessdata,congure
thetransmitter,andperformmaintenanceandtroubleshootingtasks.
ConfigurationandUseManual
49
Page 56
QuickstartwiththeFieldCommunicator
YoucanconnecttheFieldCommunicatortotheHARTclipsonthetransmitter,toanypointinalocal
HARTloop,ortoanypointinaHARTmultidropnetwork.
Ifthetransmitterisinahazardousarea,donotremovethehousingcoverwhilepoweris beingsuppliedtotheunit.Removingthehousingcoverwhilepowerissuppliedtotheunit couldcauseanexplosion.Toaccesstransmitterinformationinahazardousenvironment, useacommunicationmethodthatdoesnotrequireremovingthetransmitterhousingcover.
Important
IftheHARTsecurityswitchissettoON,HARTprotocolcannotbeusedtoperformanyactionthat
requireswritingtothetransmitter.Forexample,youcannotchangetheconguration,resettotalizers,
orperformcalibrationusingthe
HARTsecurityswitchissettoOFF,nofunctionsaredisabled.
FieldCommunicatororProLinkIIwithaHARTconnection.Whenthe
Procedure
1.ToconnecttotheHARTclips:
a.Removethetransmitterhousingcover.
b.AttachtheleadsfromtheFieldCommunicatortotheHARTclipsonthefaceofthetransmitter
andaddresistanceasnecessary.
TheFieldCommunicatormustbeconnectedacrossaresistanceof250–600Ω.
Tip
HARTconnectionsarepolarity-insensitive.Itdoesnotmatterwhichleadyouattachtowhich terminal.
50MicroMotion9739MVDTransmitters
Page 57
Figure4-1FieldCommunicatorconnectiontoHARTclips
C
B
A
A
B
C
AFieldCommunicator
B250–600Ωresistance
CTransmitter
QuickstartwiththeFieldCommunicator
2.ToconnecttoapointinthelocalHARTloop,attachtheleadsfromtheFieldCommunicatortoany pointintheloopandaddresistanceasnecessary.SeeFigure4-2.
The
FieldCommunicatormustbeconnectedacrossaresistanceof250–600Ω.
Figure4-2FieldCommunicatorconnectiontolocalHARTloop
AFieldCommunicator
B250–600Ωresistance
CTransmitterterminals
ConfigurationandUseManual
51
Page 58
QuickstartwiththeFieldCommunicator
A
B
C
D
3.ToconnecttoapointintheHARTmultidropnetwork,attachtheleadsfromtheFieldCommunicator toanypointonthenetwork.SeeFigure4-3.
Figure4-3FieldCommunicatorconnectiontomultidropnetwork
AFieldCommunicator
B250–600Ωresistance
CDevicesonthenetwork
DMasterdevice
4.TurnontheFieldCommunicatorandwaituntilthemainmenuisdisplayed.
5.Ifyouareconnectingacrossamultidropnetwork:
a.SettheFieldCommunicatortopoll.
Thedevicereturnsalladdressesthatarevalid.
b.EntertheappropriateHARTaddress.
ThedefaultHARTaddressis0.However,formultidropoperation,theHARTaddressmustbe uniqueonthenetwork.
6.(Optional)TonavigatetotheOnlinemenu,pressHARTApplication→2Online.
Mostconguration,maintenance,andtroubleshootingtasksareperformedfromtheOnlinemenu.
Tip
YoumayseemessagesrelatedtotheDDoractivealerts.Presstheappropriatebuttonstoignore themessageandcontinue.

4.3Configurationtipsandtricks

Reviewthesetipsbeforebeginningconguration.
52MicroMotion9739MVDTransmitters
Page 59
QuickstartwiththeFieldCommunicator
4.3.1HARTsecurity
HARTsecuritymaybeenabledonthe9739MVDtransmitter.T ocongurethetransmitterusingHART protocol,youmustdisableHARTsecurity.
4.3.2Defaultvaluesandranges
DefaultvaluesandrangesforthemostcommonlyusedparametersareprovidedinAppendixA.
4.4ConfiguretheprimarymAoutputtoreportmassflow rateinauser-selectedmeasurementunit
Note
Thisprocedureassumesthatyouarestartingfromthefactory-defaultconguration.
Procedure
1.MakeaconnectionfromtheFieldCommunicatortoyourtransmitter.
2.NavigatetotheOn-LineMenu.
3.Setthemeasurementunitformassowrate.
a.PressCongure→ManualSetup→Measurements→Flow→MassFlowUnit.
b.Selectthedesiredmeasurementunitfromthelist.
c.PresstheleftarrowuntilyouarereturnedtotheManualSetupmenu.
4.CongurethemAoutput.
a.PressInputs/Outputs→mAOutput1→PrimaryVariable.
b.SelectMassFlowRatefromthelist.
c.PressENTERuntilyouarereturnedtothemAOutput1menu.
d.PressmAOutputSettings.
e.PressPVLRVandenteranappropriatevalueforLowerRangeV alue(LRV).
LowerRangeValuespeciesthevalueofMassFlowRatetoberepresentedbyanoutputlevelof 0mAor4mA.
f.PressENTER.
g.PressPVURVandenteranappropriatevalueforUpperRangeValue(LRV).
UpperRangeValuespeciesthevalueofMassFlowRatetoberepresentedbyanoutputlevel of20mA.
h.PressENTER.
5.(Optional)PressOverview→Shortcuts→Variables→Outputs→Current(mAoutput1)andobservethemA outputreading.
ConfigurationandUseManual
53
Page 60
QuickstartwiththeFieldCommunicator
Itshouldvarybetween0mAor4mAand20mAaccordingtothemassowrateofyourprocess.

4.5Performalooptest

Alooptestisawaytoverifythatthetransmitterandthereceivingdevicearecommunicatingproperly. AlooptestalsohelpsyouknowwhetheryouneedtotrimmAoutputs.Performingalooptestisnot arequiredprocedure.However, outputavailableonyourtransmitter.
Procedure
1.TestthemAoutput.
a.PressServiceTools→Simulate→SimulateOutputs→mAOutputLoopTests,andselect4mA.
b.ReadthemAcurrentatthereceivingdeviceandcompareittothetransmitteroutput.
Thereadingsdonotneedtomatchexactly .Ifthevaluesareslightlydifferent,youcancorrect thediscrepancybytrimmingtheoutput.
c.Select20mA.
MicroMotionrecommendsperformingalooptestforeveryinputor
d.ReadthemAcurrentatthereceivingdeviceandcompareittothetransmitteroutput.
Thereadingsdonotneedtomatchexactly .Ifthevaluesareslightlydifferent,youcancorrect thediscrepancybytrimmingtheoutput.
2.TestthesecondarymAoutput.
PressServiceTools→Maintenance→SimulateOutputs→mAOutput2LoopTest,andrepeattheloop testforthesecondarymAoutput.
3.Testthefrequencyoutput.
Note
IftheWeights&Measuresapplicationisenabledonthetransmitter,itisnotpossibletoperforma looptestofthefrequencyoutput,evenwhenthetransmitterisunsecured.
a.PressServiceTools→Simulate→SimulateOutputs→FrequencyOutputTest,andchoosethe
frequencyoutputlevel.
b.Readthefrequencysignalatthereceivingdeviceandcompareittothetransmitteroutput.
c.ChooseEnd.
4.Testthediscreteoutput.
a.PressServiceTools→Simulate→SimulateOutputs→DiscreteOutputTest.
b.ChooseOff.
c.Verifythesignalstateatthereceivingdevice.
d.ChooseOn.
e.Verifythesignalstateatthereceivingdevice.
54
MicroMotion9739MVDTransmitters
Page 61
QuickstartwiththeFieldCommunicator
f.ChooseEnd.
Postrequisites
•IfthemAoutputreadingswereslightlyoffatthereceivingdevice,youcancorrectthisdiscrepancy bytrimmingtheoutput.
•IfthemAoutputreadingwassignicantlyoff(±200microamps),orifatanystepthereadingwas faulty,verifythewiringbetweenthetransmitterandtheremotedevice,andtryagain.
•IfthemAinputreadingwasslightlyoffatthetransmitter,trimandcalibratetheinputatthe remoteinputdevice.

4.6TrimmAoutputs

TrimmingthemAoutputsisnotarequiredprocedure.However,ifthereisasmalldiscrepancyinthemA readingbetweenthetransmitterandthereceivingdevice,trimmingtheoutputwillcorrectthis.
Note
Anytrimmingperformedontheoutputshouldnotexceed±200microamps.Ifmoretrimmingisrequired, contactMicroMotioncustomersupport.
Procedure
1.PressServiceTools→Maintenance→RoutineMaintenance→TrimmAoutput1tostartthemAtrim procedure.
2.FollowtheinstructionsintheguidedmethodtotrimthemAoutput.
3.ChooseServiceTools→Maintenance→RoutineMaintenance→TrimmAoutput2tostartthetrimprocedure forthesecondarymAoutput.
4.FollowtheinstructionsintheguidedmethodtotrimthemAoutput.

4.7Zerotheflowmeter

Prerequisites
Toprepareforthezeroprocedure:
1.Allowtheowmetertowarmupforatleast20minutesafterapplyingpower.
2.Runtheprocessuidthroughthesensoruntilthesensortemperaturereachesthenormalprocess operatingtemperature.
3.Stopowthroughthesensorbyshuttingavalvedownstreamfromthesensor .
4.Ensurethatowhascompletelystoppedthroughthesensor,andthatthesensoriscompletely fullofprocessuid.
5.Checktheowrate.Iftheowrateisclosetozero,youshouldnotneedtozerotheowmeter.
Important
Themeterwaszeroedatthefactory,andshouldnotrequireaeldzero.
ConfigurationandUseManual
55
Page 62
QuickstartwiththeFieldCommunicator
Note
Donotzerotheowmeterifahigh-severityalarmisactive.Correcttheproblem,thenzerothe owmeter.Youmayzerotheowmeterifalow-severityalarmisactive.
Procedure
Toinitiatetheowmeterzero,pressServiceTools→Maintenance→ZeroCalibration→PerformAutoZero.
ThedisplaywillreportCalibrationinprogress.Whenthecalibrationiscomplete,thedisplayreportsAutozero completeifthezerowassuccessful,orAutozerofailedifitwasnot.

4.8Testortunethesystemusingsensorsimulation

Sensorsimulationallowsyoutosetspecicvaluesorvaluerangesformassow,density,and temperature.Thetransmitterwillreportthespeciedvaluesandtakeallappropriateactions,e.g.,apply acutoff,activateanevent,orpostanalarm.Y oucanusethisfeaturetotestthesystem’sresponsetoa varietyofprocessconditions,includingboundaryconditions,problemconditions,oralarmconditions,or totunetheloop.
Whenyouenablesensorsimulation,thesimulatedvaluesarestoredinthesamememorylocations usedforprocessdatafromthesensor.Therefore,thesimulatedvalueswillbeusedthroughout transmitterfunctioning.Forexample,sensorsimulationwillaffect:
•Allmassow,temperature,ordensityvaluesshownonthedisplayorreportedviaoutputs ordigitalcommunications
•Themasstotalandmassinventoryvalues
•Allvolumecalculationsanddata,includingreportedvalues,volumetotals,andvolumeinventories
•Allmass,temperature,density,orvolumevaluesloggedtoDataLogger
Sensorsimulationdoesnotaffectanydiagnosticvalues.
Unlikeactualmassowanddensityvalues,thesimulatedvaluesarenottemperature-compensated (adjustedfortheeffectoftemperatureonthesensor’sowtubes).
Important
Donotenablesensorsimulationunlessyourprocesscantoleratetheeffectsofthesimulatedprocess values.
Procedure
1.Navigatetothesensorsimulationmenu:ServiceTools→Simulate→SimulateSensor.
2.Enablesensorsimulation.
3.Formassow,setWaveFormasdesiredandentertherequiredvalues.
56
MicroMotion9739MVDTransmitters
Page 63
OptionRequiredvalues
QuickstartwiththeFieldCommunicator
Fixed
Sawtooth
SimFixedValue
SimRampLowPoint
SimRampHighPoint
SimRampPeriod
Sine
SimRampLowPoint
SimRampHighPoint
SimRampPeriod
4.Fordensity,setWaveFormasdesiredandentertherequiredvalues.
OptionRequiredvalues
Fixedvalue
Triangularwave
SimFixedValue
SimRampLowPoint
SimRampHighPoint
SimRampPeriod
Sinewave
SimRampLowPoint
SimRampHighPoint
SimRampPeriod
5.Fortemperature,setWaveFormasdesiredandentertherequiredvalues.
OptionRequiredvalues
Fixedvalue
Triangularwave
SimFixedValue
SimRampLowPoint
SimRampHighPoint
SimRampPeriod
Sinewave
SimRampLowPoint
SimRampHighPoint
SimRampPeriod
6.Observethesystemresponsetothesimulatedvaluesandmakeanyappropriatechangesto thetransmittercongurationortothesystem.
ConfigurationandUseManual
57
Page 64
QuickstartwiththeFieldCommunicator
A
B
7.Modifythesimulatedvaluesandrepeat.
8.Whenyouhavenishedtestingortuning,disablesensorsimulation.

4.9Enable/disableHARTsecurity

YouusetheHARTsecurityswitchlocatedonthetransmitterdisplaytodisablecongurationofthe transmitterusingHARTprotocol.WhentheHARTsecurityswitchissettoON,HARTprotocolcannotbe usedtoperformanyactionthatrequireswritingtothetransmitter.Forexample,youcannotchange theconguration,resettotalizers,performcalibration,etc.,usingthe withaHART/Bell202orHART/RS-485connection.WhentheHARTsecurityswitchissettoOFF,no functionsaredisabled.
Important
TheHARTsecurityswitchdoesnotaffectModbuscommunications.
Ifthetransmitterisinahazardousarea,donotremovethehousingcoverwhilepower isbeingsuppliedtotheunit.Removingthehousingcoverwhilepowerissuppliedto theunitcouldcauseanexplosion.ToaccesstheHARTsecurityswitchinahazardous environment,besuretoremovepowerfromthetransmitterbeforeremovingthehousing coverandsettingtheHARTsecurityswitch.
FieldCommunicatororProLinkII
Procedure
1.Removepowerfromthetransmitter.
2.Removethetransmitterhousingcover.
3.MovetheHARTsecurityswitchtothedesiredposition(seeFigure4-4).
Figure4-4HARTsecurityswitch(onblankdisplay)
AHARTsecurityswitch
BUnused
4.Replacethetransmitterhousingcover.
5.Restorepowertothetransmitter.
58
MicroMotion9739MVDTransmitters
Page 65
II
II II
Referenceinformationfor commissioning
Chapterscoveredinthispart:
♦Congureprocessmeasurement ♦Conguredeviceoptionsandpreferences ♦Integratethemeterwiththecontrolsystem
Page 66
Page 67
Chapter5
Configureprocessmeasurement
Topicscoveredinthischapter:
♦Characterizetheowmeter ♦Conguremassowmeasurement ♦Congurevolumeowmeasurementforliquidapplications ♦Conguregasstandardvolumeowmeasurement ♦CongureFlowDirection ♦Conguredensitymeasurement ♦Conguretemperaturemeasurement ♦Congurepressurecompensation ♦Congurethepetroleummeasurementapplication ♦Conguretheconcentrationmeasurementapplication

5.1Characterizetheflowmeter

DisplayNotavailable
ProLinkII
FieldCommunicatorCongure→ManualSetup→Characterize
ProLink→Conguration→DensityProLink→Conguration→Flow
Characterizingtheowmeteradjuststhetransmitter’smeasurementalgorithmstomatchtheunique traitsofthesensoritispairedwith.Thecharacterizationparameters(alsocalledcalibrationparameters) describethesensor’ssensitivitytoow,density,andtemperature.Dependingonyoursensortype, differentparametersarerequired.Valuesforyoursensorareprovidedby tagorthecalibrationcerticate.
Note
Ifyoursensorandtransmitterwereorderedtogether,thetransmitterhasalreadybeencharacterizedat thefactory.However,youshouldstillverifycharacterizationparameters.
ConfigurationandUseManual
MicroMotiononthesensor
61
Page 68
Configureprocessmeasurement
Procedure
1.SpecifySensorType.
Straight-tube(T-Series)
Curved-tube(allsensorsexceptT-Series)
2.Settheowcharacterizationparameters.Besuretoincludealldecimalpoints.
•Forstraight-tubesensors,setFCF(FlowCalorFlowCalibrationFactor),FTG,andFFQ.
•Forcurved-tubesensors,setFlowCal(FlowCalibrationFactor).
3.Setthedensitycharacterizationparameters.
•Forstraight-tubesensors,setD1,D2,DT,DTG,K1,K2,FD,DFQ1,andDFQ2.
•Forcurved-tubesensors,setD1,D2,TC,K1,K2,andFD.(TCissometimesshownasDT.)
5.1.1Sourcesandformatsforcharacterizationparameters
Differentsensortagsdisplaycharacterizationparametersdifferently,andoldersensorsmaynothaveall therequiredparametersonthetag.
Samplesensortags
Samplesensortagsareshowninthefollowingillustrations:
•Figure5-1:Oldercurved-tubesensors(allsensorsexceptT-Series)
•Figure5-2:Newercurved-tubesensors(allsensorsexceptT-Series)
•Figure5-3:Olderstraight-tubesensors(T-Series)
•Figure5-4:Newerstraight-tubesensors(T-Series)
Figure5-1Tagonoldercurved-tubesensors(allsensorsexceptT-Series)
62MicroMotion9739MVDTransmitters
Page 69
Figure5-2Tagonnewercurved-tubesensors(allsensorsexceptT-Series)
Figure5-3Tagonolderstraight-tubesensor(T-Series)
Configureprocessmeasurement
Figure5-4Tagonnewerstraight-tubesensor(T-Series)
ConfigurationandUseManual
63
Page 70
Configureprocessmeasurement
Densitycalibrationparameters(D1,D2,K1,K2,FD,DT,TC)
IfyoursensortagdoesnotshowaD1orD2value:
•ForD1,entertheDensAorD1valuefromthecalibrationcerticate.Thisvalueistheline-condition densityofthelow-densitycalibrationuid.MicroMotionusesair.IfyoucannotndaDensAor D1value,enter0.001g/cm
•ForD2,entertheDensBorD2valuefromthecalibrationcerticate.Thisvalueistheline-condition densityofthehigh-densitycalibrationuid. orD2value,enter0.998g/cm
IfyoursensortagdoesnotshowaK1orK2value:
•ForK1,entertherst5digitsofthedensitycalibrationfactor.InthesampletaginFigure5-1,this valueisshownas12500.
•ForK2,enterthesecond5digitsofthedensitycalibrationfactor.InthesampletaginFigure5-1, thisvalueisshownas14286.
3
.
MicroMotionuseswater.IfyoucannotndaDensB
3
.
IfyoursensordoesnotshowanFDvalue,contact
ow.support@emerson.com.
IfyoursensortagdoesnotshowaDTorTCvalue,enterthelast3digitsofthedensitycalibration factor.InthesampletaginFigure5-1,thisvalueisshownas4.44.
Flowcalibrationparameters(FCF,FT)
Twoseparatevaluesareusedtodescribeowcalibration:a6-characterFCFvalueanda4-character FTvalue.Bothvaluescontaindecimalpoints.Duringcharacterization,theseareenteredasasingle 10-characterstringthatincludestwodecimalpoints.ThisparameteriscalledeitherFlowcalorFCF.
IfyoursensortagshowstheFCFandtheFTvaluesseparately ,concatenatethetwovaluestoform thesingleparametervalue.
tExample:ConcatenatingFCFandFT
FCF=x.xxxx FT=y.yy Flowcalibrationparameter:x.xxxxy.yy

5.2Configuremassflowmeasurement

Display
ProLinkII
FieldCommunicatorCongure→ManualSetup→Measurements→Flow
OFF-LINEMAINT→OFF-LINECONFG→UNITS→MASS
ProLink→Conguration→Flow
Themassowmeasurementparameterscontrolhowmassowismeasuredandreported.
Themassowmeasurementparametersinclude:
MassFlowMeasurementUnit
FlowDamping
MassFlowCutoff
64
MicroMotion9739MVDTransmitters
Page 71
5.2.1ConfigureMassFlowMeasurementUnit
Configureprocessmeasurement
Display
ProLinkII
FieldCommunicatorCongure→ManualSetup→Measurements→Flow→MassFlowUnit
OFF-LINEMAINT→OFF-LINECONFG→UNITS→MASS
ProLink→Conguration→Flow→MassFlowUnits
MassFlowMeasurementUnitspeciestheunitwillbeusedforthemassowrate.Theunitusedformass totalandmassinventoryisderivedfromthisunit.
Procedure
SetMassFlowMeasurementUnittothedesiredunit.
ThedefaultsettingforMassFlowMeasurementUnitisg/s(gramspersecond).
Tip
Ifthemeasurementunityouwanttouseisnotavailable,youcandeneaspecialmeasurementunit.
OptionsforMassFlowMeasurementUnit
ThetransmitterprovidesastandardsetofmeasurementunitsforMassFlowMeasurementUnit,plusone user-denedspecialmeasurementunit.Differentcommunicationstoolusedifferentlabelsfortheunits.
OptionsforMassFlowMeasurementUnitareshowninTable5-1.
Table5-1OptionsforMassFlowMeasurementUnit
Label
UnitdescriptionDisplayProLinkIIFieldCommunicator
GramspersecondG/Sg/sg/s
GramsperminuteG/MINg/ming/min
GramsperhourG/Hg/hrg/h
Kilogramspersecond
Kilogramsperminute
Kilogramsperhour
Kilogramsperday
Metrictonsperminute
Metrictonsperhour
Metrictonsperday
Poundspersecond
Poundsperminute
Poundsperhour
Poundsperday
Shorttons(2000pounds) perminute
KG/Skg/skg/s
KG/MINkg/minkg/min
KG/Hkg/hrkg/h
KG/Dkg/daykg/d
T/MINmTon/minMetT on/min
T/HmTon/hrMetT on/h
T/DmTon/dayMetT on/d
LB/Slbs/slb/s
LB/MINlbs/minlb/min
LB/Hlbs/hrlb/h
LB/Dlbs/daylb/d
ST/MINsT on/minSTon/min
ConfigurationandUseManual
65
Page 72
Configureprocessmeasurement
Table5-1OptionsforMassFlowMeasurementUnit
UnitdescriptionDisplayProLinkIIFieldCommunicator
Shorttons(2000pounds) perhour
Shorttons(2000pounds) perday
Longtons(2240pounds) perhour
Longtons(2240pounds) perday
SpecialunitSPECL
ST/HsT on/hrSTon/h
ST/DsT on/dayST on/d
LT/HlTon/hrLT on/h
LT/DlTon/dayLT on/d
continued
Label
special
Defineaspecialmeasurementunitformassflow
DisplayNotavailable
ProLinkII
FieldCommunicatorCongure→ManualSetup→Measurements→SpecialUnits→MassSpecialUnits
ProLink→Conguration→SpecialUnits
Spcl
Aspecialmeasurementunitallowsyoutoreportprocessdata,totalizerdata,andinventorydataina unitthatisnothard-codedinthetransmitter.Aspecialmeasurementunitiscalculatedfromanexisting measurementunitusingaconversionfactor.
Restriction
Althoughyoucannotdeneaspecialmeasurementunitusingthedisplay,youcanusethedisplayto selectanexistingspecialmeasurementunitandtoviewprocessdata.
Procedure
1.SpecifyBaseMassUnit.
BaseMassUnitistheexistingmassunitthatthespecialunitwillbebasedon.
2.SpecifyBaseTimeUnit.
BaseTimeUnitistheexistingtimeunitthatthespecialunitwillbebasedon.
3.CalculateMassFlowConversionFactorasfollows:
a.xbaseunits=yspecialunits
b.MassFlowConversionFactor=x/y
4.EnterMassFlowConversionFactor.
5.SetMassFlowLabeltothelabeltobeusedforthemassowunit.
6.SetMassTotalLabeltothelabeltobeusedforthemasstotalandmassinventoryunit.
Thespecialmeasurementunitisstoredinthetransmitter.Youcancongurethetransmittertousethe specialmeasurementunitatanytime.
66
MicroMotion9739MVDTransmitters
Page 73
tExample:Definingaspecialmeasurementunitformassflow
Youwanttomeasuremassowinouncespersecond.
1.SetBaseMassUnittoPounds(lb).
2.SetBaseTimeUnittoSeconds(sec).
3.CalculateMassFlowConversionFactor:
a.1lb/sec=16oz/sec
b.MassFlowConversionFactor=1/16=0.0625
4.SetMassFlowConversionFactorto0.0625.
5.SetMassFlowLabeltooz/sec.
6.SetMassT otalLabeltooz.
5.2.2ConfigureFlowDamping
DisplayNotavailable
ProLinkII
FieldCommunicatorCongure→ManualSetup→Measurements→Flow→FlowDamping
ProLink→Conguration→Flow→FlowDamp
Configureprocessmeasurement
Dampingisusedtosmoothoutsmall,rapiductuationsinprocessmeasurement.TheDampingValue speciesthetimeperiod(inseconds)overwhichthetransmitterwillspreadchangesinthereported processvariable.Attheendoftheinterval,thereportedprocessvariablewillreect63%ofthechange intheactualmeasuredvalue.
Tips
•Ahighdampingvaluemakestheprocessvariableappearsmootherbecausethereportedvalue mustchangeslowly .
•Alowdampingvaluemakestheprocessvariableappearmoreerraticbecausethereported valuechangesmorequickly.
Procedure
SetFlowDampingtothedesiredvalue.
Thedefaultvalueis0.8seconds.Therangeis0to10.24seconds.WhenyouenteravalueforFlow
Damping,thetransmitterautomaticallyroundsitdowntothenearestvalidvalue.ThevalidvaluesforFlow Dampingare:0,0.04,0.08,0.16,...10.24.
Tips
Forgasapplications,MicroMotionrecommendssettingFlowDampingto2.56orhigher.
ConfigurationandUseManual
67
Page 74
Configureprocessmeasurement
EffectofFlowDampingonvolumemeasurement
FlowDampingeffectsvolumemeasurementforbothliquidvolumeandgasstandardvolume.Volumedata
iscalculatedfromthedampedmassowdataratherthanthemeasuredowvalue.
InteractionbetweenFlowDampingandAddedDamping
FlowDampingcontrolstherateofchangeinowprocessvariables.AddedDampingcontrolstherateof changereportedviathemAoutput.IfmAOutputProcessVariableissettoMassFlowRate,andbothFlow DampingandAddedDampingaresettonon-zerovalues,owdampingisappliedrst,andtheadded
dampingcalculationisappliedtotheresultoftherstcalculation.
5.2.3ConfigureMassFlowCutoff
DisplayNotavailable
ProLinkII
FieldCommunicatorCongure→ManualSetup→Measurements→Flow→MassFlowCutoff
ProLink→Conguration→Flow→MassFlowCutoff
MassFlowCutoffspeciesthelowestmassowratethatwillreportedasmeasured.Allmassowrates belowthiscutoffwillbereportedas0.
Procedure
SetMassFlowCutofftothedesiredvalue.
ThedefaultvalueforMassFlowCutoffis0.0g/s.Therecommendedsettingis0.05%ofthesensor's ratedmaximumowrate.
EffectofMassFlowCutoffonvolumemeasurement
MassFlowCutoffdoesnotaffectvolumemeasurement.Volumedataiscalculatedfromtheactualmass
dataratherthanthereportedvalue.
InteractionbetweenMassFlowCutoffandAOCutoff
MassFlowCutoffaffectsallreportedvaluesandvaluesusedinothertransmitterbehavior(e.g.,events
denedonmassow).
AOCutoffaffectsonlymassowvaluesreportedviathemAoutput.
tExample:Cutoffinteraction
Conguration:
mAOutputProcessVariablefortheprimarymAoutput:MassFlowRate
FrequencyOutputProcessVariable:MassFlowRate
AOCutofffortheprimarymAoutput:10grams/second
MassFlowCutoff:15grams/second
Result:Ifthemassowratedropsbelow15grams/second,massowwillbereportedas0,and0 willbeusedinallinternalprocessing.
68
MicroMotion9739MVDTransmitters
Page 75
Configureprocessmeasurement
tExample:Cutoffinteraction
Conguration:
mAOutputProcessVariablefortheprimarymAoutput:MassFlowRate
FrequencyOutputProcessVariable:MassFlowRate
AOCutofffortheprimarymAoutput:15grams/second
MassFlowCutoff:10grams/second
Result:
•Ifthemassowratedropsbelow15grams/secondbutnotbelow10grams/second:
−TheprimarymAoutputwillreportzeroow.
−Thefrequencyoutputwillreporttheactualowrate,andtheactualowratewillbeused inallinternalprocessing.
•Ifthemassowratedropsbelow10grams/second,bothoutputswillreportzeroow,and0will beusedinallinternalprocessing.

5.3Configurevolumeflowmeasurementforliquid applications

Display
ProLinkII
FieldCommunicatorCongure→ManualSetup→Measurements→Flow
Thevolumeowmeasurementparameterscontrolhowliquidvolumeowismeasuredandreported.
Thevolumeowmeasurementparametersinclude:
VolumeFlowType
VolumeFlowMeasurementUnit
VolumeFlowCutoff
Restriction
Youcannotimplementbothliquidvolumeowandgasstandardvolumeow.Y oumustchooseoneor theother.
OFF-LINEMAINT→OFF-LINECONFG→UNITS→VOL
ProLink→Conguration→Flow
5.3.1ConfigureVolumeFlowTypeforliquidapplications
Display
ProLinkII
FieldCommunicatorCongure→ManualSetup→Measurements→GasStandardVolume→VolumeFlowType→Liquid
VolumeFlowTypecontrolswhetherliquidorgasstandardvolumeowmeasurementwillbeimplemented.
ConfigurationandUseManual
OFF-LINEMAINT→OFF-LINECONFG→VOL→VOLTYPELIQUID
ProLink→Conguration→Flow→VolFlowType→LiquidVolume
69
Page 76
Configureprocessmeasurement
Restriction
Ifyouareusingthepetroleummeasurementapplication,youmustsetVolumeFlowTypetoLiquid.Gas standardvolumemeasurementisincompatiblewiththepetroleummeasurementapplication.
Restriction
Ifyouareusingtheconcentrationmeasurementapplication,youmustsetVolumeFlowTypetoLiquid.Gas standardvolumemeasurementisincompatiblewiththeconcentrationmeasurementapplication.
Procedure
SetVolumeFlowTypetoLiquid.
5.3.2ConfigureVolumeFlowMeasurementUnitforliquidapplications
Display
ProLinkII
FieldCommunicatorCongure→ManualSetup→Measurements→Flow→VolumeFlowUnit
VolumeFlowMeasurementUnitspeciestheunitwillbeusedforthevolumeowrate.Theunitusedfor thevolumetotalandthevolumeinventoryisderivedfromthisunit.
OFF-LINEMAINT→OFF-LINECONFG→UNITS→VOL
ProLink→Conguration→Flow→VolFlowUnits
Prerequisites
BeforeyoucongureVolumeFlowMeasurementUnit,besurethatVolumeFlowTypeissettoLiquid.
Procedure
SetVolumeFlowMeasurementUnittothedesiredunit.
ThedefaultsettingforVolumeFlowMeasurementUnitisL/s(literspersecond).
Tip
Ifthemeasurementunityouwanttouseisnotavailable,youcandeneaspecialmeasurementunit.
OptionsforVolumeFlowMeasurementUnitforliquidapplications
ThetransmitterprovidesastandardsetofmeasurementunitsforVolumeFlowMeasurementUnit,plusone user-denedspecialmeasurementunit.Differentcommunicationstoolusedifferentlabelsfortheunits.
OptionsforVolumeFlowMeasurementUnitareshowninT able5-2.
70MicroMotion9739MVDTransmitters
Page 77
Table5-2OptionsforVolumeFlowMeasurementUnitforliquidapplications
Label
UnitdescriptionDisplayProLinkIIFieldCommunicator
CubicfeetpersecondCUFT/Sft3/secCuft/s
CubicfeetperminuteCUF/MNft3/minCuft/min
CubicfeetperhourCUFT/Hft3/hrCuft/h
CubicfeetperdayCUFT/Dft3/dayCuft/d
CubicmeterspersecondM3/Sm3/secCum/s
CubicmetersperminuteM3/MINm3/minCum/min
CubicmetersperhourM3/Hm3/hrCum/h
CubicmetersperdayM3/Dm3/dayCum/d
U.S.gallonspersecondUSGPSUSgal/secgal/s
U.S.gallonsperminuteUSGPMUSgal/mingal/min
U.S.gallonsperhourUSGPHUSgal/hrgal/h
U.S.gallonsperdayUSGPDUSgal/dgal/d
MillionU.S.gallonsperdayMILG/DmilUSgal/dayMMgal/d
Literspersecond
Litersperminute
Litersperhour
Millionlitersperday
Imperialgallonsper second
Imperialgallonsperminute
Imperialgallonsperhour
Imperialgallonsperday
Barrelspersecond
Barrelsperminute
Barrelsperhour
Barrelsperday
Beerbarrelspersecond
Beerbarrelsperminute
Beerbarrelsperhour
Beerbarrelsperday
SpecialunitSPECL
L/Sl/secL/s
L/MINl/minL/min
L/Hl/hrL/h
MILL/Dmill/dayML/d
UKGPSImpgal/secImpgal/s
UKGPMImpgal/minImpgal/min
UKGPHImpgal/hrImpgal/h
UKGPDImpgal/dayImpgal/d
BBL/Sbarrels/secbbl/s
BBL/MNbarrels/minbbl/min
BBL/Hbarrels/hrbbl/h
BBL/Dbarrels/daybbl/d
BBBL/SBeerbarrels/secbbbl/s
BBBL/MNBeerbarrels/minbbbl/min
BBBL/HBeerbarrels/hrbbbl/h
BBBL/DBeerbarrels/daybbbl/d
special
Configureprocessmeasurement
Spcl
ConfigurationandUseManual
71
Page 78
Configureprocessmeasurement
Defineaspecialmeasurementunitforvolumeflow
DisplayNotavailable
ProLinkII
FieldCommunicatorCongure→ManualSetup→Measurements→SpecialUnits→VolumeSpecialUnits
Aspecialmeasurementunitallowsyoutoreportprocessdata,totalizerdata,andinventorydataina unitthatisnothard-codedinthetransmitter.Aspecialmeasurementunitiscalculatedfromanexisting measurementunitusingaconversionfactor.
Restriction
Althoughyoucannotdeneaspecialmeasurementunitusingthedisplay,youcanusethedisplayto selectanexistingspecialmeasurementunitandtoviewprocessdata.
Procedure
1.SpecifyBaseVolumeUnit.
ProLink→Conguration→SpecialUnits
BaseVolumeUnitistheexistingvolumeunitthatthespecialunitwillbebasedon.
2.SpecifyBaseTimeUnit.
BaseTimeUnitistheexistingtimeunitthatthespecialunitwillbebasedon.
3.CalculateVolumeFlowConversionFactorasfollows:
a.xbaseunits=yspecialunits
b.VolumeFlowConversionFactor=x/y
4.EntertheVolumeFlowConversionFactor.
5.SetVolumeFlowLabeltothelabeltobeusedforthevolumeowunit.
6.SetVolumeTotalLabeltothelabeltobeusedforthevolumetotalandvolumeinventoryunit.
Thespecialmeasurementunitisstoredinthetransmitter.Youcancongurethetransmittertousethe specialmeasurementunitatanytime.
tExample:Definingaspecialmeasurementunitforvolumeflow
Youwanttomeasurevolumeowinpintspersecond.
1.SetBaseVolumeUnittoGallons(gal).
2.SetBaseTimeUnittoSeconds(sec).
3.Calculatetheconversionfactor:
a.1gal/sec=8pints/sec
b.VolumeFlowConversionFactor=1/8=0.1250
4.SetVolumeFlowConversionFactorto0.1250.
5.SetVolumeFlowLabeltopints/sec.
6.SetVolumeTotalLabeltopints.
72MicroMotion9739MVDTransmitters
Page 79
Configureprocessmeasurement
5.3.3ConfigureVolumeFlowCutoff
DisplayNotavailable
ProLinkII
FieldCommunicatorCongure→ManualSetup→Measurements→Flow→VolumeFlowCutoff
VolumeFlowCutoffspeciesthelowestvolumeowratethatwillreportedasmeasured.Allvolume owratesbelowthiscutoffwillbereportedas0.
Procedure
SetVolumeFlowCutofftothedesiredvalue.
ThedefaultvalueforVolumeFlowCutoffis0.0L/s.Thelowerlimitis0.Theupperlimitisthesensor’sow calibrationfactor,inunitsofL/s,multipliedby0.2.
InteractionbetweenVolumeFlowCutoffandAOCutoff
VolumeFlowCutoffdenesthelowestliquidvolumeowvaluethatthetransmitterwillreportasmeasured. AOCutoffdenesthelowestowratethatwillbereportedviathemAoutput.IfmAOutputProcessVariableis
settoVolumeFlowRate,thevolumeowratereportedviathemAoutputiscontrolledbythehigherof thetwocutoffvalues.
ProLink→Conguration→Flow→VolFlowCutoff
VolumeFlowCutoffaffectsbothvolumeowvaluesreportedviaoutputsandvolumeowvaluesusedin othertransmitterbehavior(e.g.,eventsdenedonvolumeow).
AOCutoffaffectsonlyowvaluesreportedviathemAoutput.
tExample:CutoffinteractionwithAOCutofflowerthanVolumeFlowCutoff
Conguration:
mAOutputProcessVariablefortheprimarymAoutput:VolumeFlowRate
FrequencyOutputProcessVariable:VolumeFlowRate
AOCutofffortheprimarymAoutput:10liters/second
VolumeFlowCutoff:15liters/second
Result:Ifthemassowratedropsbelow15liters/second,volumeowwillbereportedas0,and0 willbeusedinallinternalprocessing.
tExample:CutoffinteractionwithAOCutoffhigherthanVolumeFlowCutoff
Conguration:
mAOutputProcessVariablefortheprimarymAoutput:VolumeFlowRate
FrequencyOutputProcessVariable:VolumeFlowRate
AOCutofffortheprimarymAoutput:15liters/second
VolumeFlowCutoff:10liters/second
ConfigurationandUseManual
73
Page 80
Configureprocessmeasurement
Result:
•Ifthevolumeowratedropsbelow15liters/secondbutnotbelow10liters/second:
−TheprimarymAoutputwillreportzeroow.
−Thefrequencyoutputwillreporttheactualowrate,andtheactualowratewillbeused inallinternalprocessing.
•Ifthevolumeowratedropsbelow10liters/second,bothoutputswillreportzeroow,and0will beusedinallinternalprocessing.

5.4Configuregasstandardvolumeflowmeasurement

Display
ProLinkII
FieldCommunicatorCongure→ManualSetup→Measurements→GasStandardVolume
OFF-LINEMAINT→OFF-LINECONFG→VOL→VOLTYPEGAS
ProLink→Conguration→Flow→VolFlowType
Thegasstandardvolumeowmeasurementparameterscontrolhowgasstandardvolumeowis measuredandreported.
Thegasstandardvolumeowmeasurementparametersinclude:
VolumeFlowType
StandardGasDensity
GasStandardVolumeFlowMeasurementUnit
GasStandardVolumeFlowCutoff
Restriction
Youcannotimplementbothliquidvolumeowandgasstandardvolumeow.Y oumustchooseoneor theother.
5.4.1ConfigureVolumeFlowTypeforgasapplications
Display
ProLinkII
FieldCommunicatorCongure→ManualSetup→Measurements→GasStandardVolume→V olumeFlowType→GSV
OFF-LINEMAINT→OFF-LINECONFG→VOL→VOLTYPEGAS
ProLink→Conguration→Flow→VolFlowType→StdGasVolume
VolumeFlowTypecontrolswhetherliquidorgasstandardvolumeowmeasurementwillbeimplemented.
Restriction
Ifyouareusingthepetroleummeasurementapplication,youmustsetVolumeFlowTypetoLiquid.Gas standardvolumemeasurementisincompatiblewiththepetroleummeasurementapplication.
74
MicroMotion9739MVDTransmitters
Page 81
Configureprocessmeasurement
Restriction
Ifyouareusingtheconcentrationmeasurementapplication,youmustsetVolumeFlowTypetoLiquid.Gas standardvolumemeasurementisincompatiblewiththeconcentrationmeasurementapplication.
Procedure
SetVolumeFlowTypetoGasStandardVolume.
5.4.2ConfigureStandardGasDensity
DisplayNotavailable
ProLinkII
FieldCommunicatorCongure→ManualSetup→Measurements→GasStandardVolume→GasDensity
StandardGasDensityisusedtoconvertthemeasuredowdatatoreference(standard)values.
Prerequisites
ProLink→Conguration→Flow→StdGasDensity
EnsurethatDensityMeasurementUnitissettotheunitsyouwilluseforStandardGasDensity.
Procedure
EntertheappropriateStandardGasDensityvalueforthegasyouaremeasuring.
Tip
ProLinkIIprovidesaGasWizardthatyoucanusetocalculatethestandarddensityofyourgas,ifyou donotknowit.
5.4.3ConfigureGasStandardVolumeFlowMeasurementUnit
Display
ProLinkII
FieldCommunicatorCongure→ManualSetup→Measurements→GasStandardVolume→GasVolFlowUnit
GasStandardVolumeFlowMeasurementUnitspeciestheunitwillbeusedforthegasstandardvolume owrate.Theunitusedforthegasstandardvolumetotalandthegasstandardvolumeinventoryis derivedfromthisunit.
Prerequisites
OFF-LINEMAINT→OFF-LINECONFG→UNITS→VOL
ProLink→Conguration→Flow→StdGasVolFlowUnits
BeforeyoucongureGasStandardVolumeFlowMeasurementUnit,besurethatVolumeFlowTypeissettoGas StandardVolume.
Procedure
SetGasStandardVolumeFlowMeasurementUnittothedesiredunit.
ThedefaultsettingforGasStandardVolumeFlowMeasurementUnitisSCFM(standardcubicfeetperminute).
ConfigurationandUseManual
75
Page 82
Configureprocessmeasurement
Tip
Ifthemeasurementunityouwanttouseisnotavailable,youcandeneaspecialmeasurementunit.
OptionsforGasStandardVolumeFlowMeasurementUnit
ThetransmitterprovidesastandardsetofmeasurementunitsforGasStandardVolumeFlowMeasurement Unit,plusoneuser-denedspecialmeasurementunit.Differentcommunicationstoolusedifferent
labelsfortheunits.
OptionsforGasStandardVolumeFlowMeasurementUnitareshowninTable5-3.
Table5-3OptionsforGasStandardVolumeMeasurementUnit
Label
UnitdescriptionDisplayProLinkIIFieldCommunicator
Normalcubicmetersper second
Normalcubicmetersper minute
Normalcubicmetersper hour
Normalcubicmetersper day
Normalliterpersecond
NormalliterperminuteNLPMNLPMNotavailable
NormalliterperhourNLPHNLPHNotavailable
NormalliterperdayNLPDNLPDNotavailable
Standardcubicfeetper second
Standardcubicfeetper minute
Standardcubicfeetper hour
Standardcubicfeetper day
Standardcubicmetersper second
Standardcubicmetersper minute
Standardcubicmetersper hour
Standardcubicmetersper day
StandardliterpersecondSLPSSLPS
StandardliterperminuteSLPMSLPM
NM3/SNm3/sec
NM3/MNNm3/min
NM3/HNm3/hr
NM3/DNm3/day
NLPSNLPS
SCFSSCFS
SCFMSCFM
SCFHSCFH
SCFDSCFD
SM3/SSm3/S
SM3/MNSm3/min
SM3/HSm3/hr
SM3/DSm3/day
Notavailable
Notavailable
Notavailable
Notavailable
Notavailable
Notavailable
Notavailable
Notavailable
Notavailable
Notavailable
Notavailable
Notavailable
Notavailable
Notavailable
Notavailable
76
MicroMotion9739MVDTransmitters
Page 83
Configureprocessmeasurement
Table5-3OptionsforGasStandardVolumeMeasurementUnit
UnitdescriptionDisplayProLinkIIFieldCommunicator
StandardliterperhourSLPHSLPH
StandardliterperdaySLPDSLPD
SpecialmeasurementunitSPECL
continued
Label
Notavailable
Notavailable
special
Spcl
DefineaspecialmeasurementunitforGasStandardVolumeflow
DisplayNotavailable
ProLinkII
FieldCommunicatorCongure→ManualSetup→Measurements→SpecialUnits→VolumeSpecialUnits
Aspecialmeasurementunitallowsyoutoreportprocessdata,totalizerdata,andinventorydataina unitthatisnothard-codedinthetransmitter.Aspecialmeasurementunitiscalculatedfromanexisting measurementunitusingaconversionfactor.
Restriction
Althoughyoucannotdeneaspecialmeasurementunitusingthedisplay,youcanusethedisplayto selectanexistingspecialmeasurementunitandtoviewprocessdata.
ProLink→Conguration→SpecialUnits
Procedure
1.SpecifyBaseGasStandardVolumeUnit.
BaseGasStandardVolumeUnitistheexistingGasStandardVolumeunitthatthespecialunitwillbe
basedon.
2.SpecifyBaseTimeUnit.
BaseTimeUnitistheexistingtimeunitthatthespecialunitwillbebasedon.
3.CalculateGasStandardVolumeFlowConversionFactorasfollows:
a.xbaseunits=yspecialunits
b.GasStandardVolumeFlowConversionFactor=x/y
4.EntertheGasStandardVolumeFlowConversionFactor.
5.SetGasStandardVolumeFlowLabeltothelabeltobeusedfortheGasStandardVolumeowunit.
6.SetGasStandardVolumeTotalLabeltothelabeltobeusedfortheGasStandardVolumetotaland GasStandardVolumeinventoryunit.
Thespecialmeasurementunitisstoredinthetransmitter.Youcancongurethetransmittertousethe specialmeasurementunitatanytime.
tExample:DefiningaspecialmeasurementunitforGasStandardVolumeflow
YouwanttomeasureGasStandardVolumeowinthousandsofstandardcubicfeetperminute.
1.SettheBaseGasStandardVolumeUnittoSCFM.
ConfigurationandUseManual
77
Page 84
Configureprocessmeasurement
2.SettheBaseTimeUnittominutes(min).
3.Calculatetheconversionfactor:
a.1thousandsofstandardcubicfeetperminute=1000cubicfeetperminute
b.GasStandardVolumeFlowConversionFactor=1/1000=0.001
4.SetGasStandardVolumeFlowConversionFactorto0.001.
5.SetGasStandardVolumeFlowLabeltoKSCFM.
6.SetGasStandardVolumeTotalLabeltoKSCF.
5.4.4ConfigureGasStandardVolumeFlowCutoff
DisplayNotavailable
ProLinkII
FieldCommunicatorCongure→ManualSetup→Measurements→GasStandardVolume→GSVCutoff
ProLink→Conguration→Flow→StdGasVolFlowCutoff
GasStandardVolumeFlowCutoffspeciesthelowestvolumeowratethatwillreportedasmeasured.All volumeowratesbelowthiscutoffwillbereportedas0.
Procedure
SetVolumeFlowCutofftothedesiredvalue.
ThedefaultvalueforGasStandardVolumeFlowCutoffis0.0.Thelowerlimitis0.0.Thereisnoupperlimit.
InteractionbetweenGasStandardVolumeFlowCutoffandAOCutoff
GasStandardVolumeFlowCutoffdenesthelowestGasStandardVolumeowvaluethatthetransmitterwill reportasmeasured.AOCutoffdenesthelowestowratethatwillbereportedviathemAoutput.If mAOutputProcessV ariableissettoGasStandardVolumeFlowRate,thevolumeowratereportedviathemA
outputiscontrolledbythehigherofthetwocutoffvalues.
GasStandardVolumeFlowCutoffaffectsbothgasstandardvolumeowvaluesreportedviaoutputsand gasstandardvolumeowvaluesusedinothertransmitterbehavior(e.g.,eventsdenedongas standardvolumeow).
AOCutoffaffectsonlyowvaluesreportedviathemAoutput.
tExample:CutoffinteractionwithAOCutofflowerthanGasStandardVolumeFlowCutoff
Conguration:
mAOutputProcessVariablefortheprimarymAoutput:GasStandardVolumeFlowRate
FrequencyOutputProcessVariable:GasStandardVolumeFlowRate
AOCutofffortheprimarymAoutput:10SLPM(standardlitersperminute)
GasStandardVolumeFlowCutoff:15SLPM
Result:IftheGasStandardVolumeowratedropsbelow15SLPM,volumeowwillbereportedas 0,and0willbeusedinallinternalprocessing.
78
MicroMotion9739MVDTransmitters
Page 85
Configureprocessmeasurement
tExample:CutoffinteractionwithAOCutoffhigherthanGasStandardVolumeFlowCutoff
Conguration:
mAOutputProcessVariablefortheprimarymAoutput:GasStandardVolumeFlowRate
FrequencyOutputProcessVariable:GasStandardVolumeFlowRate
AOCutofffortheprimarymAoutput:15SLPM(standardlitersperminute)
GasStandardVolumeFlowCutoff:10SLPM
Result:
•IftheGasStandardVolumeowratedropsbelow15SLPMbutnotbelow10SLPM:
−TheprimarymAoutputwillreportzeroow.
−Thefrequencyoutputwillreporttheactualowrate,andtheactualowratewillbeused inallinternalprocessing.
•IftheGasStandardVolumeowratedropsbelow10SLPM,bothoutputswillreportzeroow, and0willbeusedinallinternalprocessing.

5.5ConfigureFlowDirection

DisplayNotavailable
ProLinkII
FieldCommunicatorCongure→ManualSetup→Measurements→Flow→FlowDirection
FlowDirectioncontrolshowconditionsofforwardowandreverseowaffectowmeasurementand reporting.
FlowDirectionisdenedwithrespecttotheowarrowonthesensor:
•Forwardow(positiveow)movesinthedirectionoftheowarrowonthesensor.
•Reverseow(negativeow)movesinthedirectionoppositetotheowarrowonthesensor.
Procedure
SetFlowDirectionasdesired.
5.5.1OptionsforFlowDirection
FlowDirectioncontrolshowtheoutputsreportowandhowthetotalizersandinventoriesincrementtotals.
ProLink→Conguration→Flow→FlowDirection
ConfigurationandUseManual
79
Page 86
Configureprocessmeasurement
Table5-4OptionsforFlowDirection
FlowDirectionsetting
ProLinkIIFieldCommunicator
ForwardForward
ReverseReverse
AbsoluteValueAbsoluteValue
BidirectionalBidirectional
NegateForward
NegateBidirectional
Negate/ForwardOnly
Negate/Bi-directional
5.5.2EffectofFlowDirectionontransmitteroutputsandtotalizers
FlowDirectionandmAoutputs
mAoutputsareaffectedbyFlowDirectiononlyifmAOutputProcessVariableissettoaowvariable.
TheeffectofFlowDirectiononmAoutputsdependontheLowerRangeValueconguredforthemAoutput:
•IfLowerRangeValueissetto0,seeFigure5-5.
•IfLowerRangeValueissettoanegativevalue,seeFigure5-6.
Figure5-5EffectofFlowDirectiononthemAoutput:LowerRangeValue=0
Notes
LowerRangeValue=0
UpperRangeValue=x
80MicroMotion9739MVDTransmitters
Page 87
Figure5-6EffectofFlowDirectiononthemAoutput:LowerRangeValue<0
Notes
LowerRangeValue=−x
UpperRangeValue=x
Configureprocessmeasurement
tExample:FlowDirection=ForwardandLowerRangeValue=0
Conguration:
FlowDirection=Forward
LowerRangeValue=0g/s
UpperRangeValue=100g/s
Result:
•Underconditionsofreverseoworzeroow,themAoutputis4mA.
•Underconditionsofforwardow,uptoaowrateof100g/s,themAoutputvariesbetween 4mAand20mAinproportiontotheowrate.
•Underconditionsofforwardow,iftheowrateequalsorexceeds100g/s,themAoutputwillbe proportionaltotheowrateupto20.5mA,andwillbelevelat20.5mAathigherowrates.
tExample:FlowDirection=ForwardandLowerRangeValue<0
Conguration:
FlowDirection=Forward
LowerRangeValue=−100g/s
UpperRangeValue=+100g/s
ConfigurationandUseManual
81
Page 88
Configureprocessmeasurement
Result:
•Underconditionsofzeroow,themAoutputis12mA.
•Underconditionsofforwardow,forowratesbetween0and+100g/s,themAoutputvaries between12mAand20mAinproportionto(theabsolutevalueof)theowrate.
•Underconditionsofforwardow,if(theabsolutevalueof)theowrateequalsorexceeds100g/s, themAoutputisproportionaltotheowrateupto20.5mA,andwillbelevelat20.5mAathigher owrates.
•Underconditionsofreverseow,forowratesbetween0and−100g/s,themAoutputvaries between4mAand12mAininverseproportiontotheabsolutevalueoftheowrate.
•Underconditionsofreverseow,iftheabsolutevalueoftheowrateequalsorexceeds100 g/s,themAoutputisinverselyproportionaltotheowratedownto3.8mA,andwillbelevelat
3.8mAathigherabsolutevalues.
tExample:FlowDirection=Reverse
Conguration:
FlowDirection=Reverse
LowerRangeValue=0g/s
UpperRangeValue=100g/s
Result:
•Underconditionsofforwardoworzeroow,themAoutputis4mA.
•Underconditionsofreverseow,forowratesbetween0and−100g/s,themAoutputlevelvaries between4mAand20mAinproportiontotheabsolutevalueoftheowrate.
•Underconditionsofreverseow,iftheabsolutevalueoftheowrateequalsorexceeds100g/s, themAoutputwillbeproportionaltotheabsolutevalueoftheowrateupto20.5mA,andwill belevelat20.5mAathigherabsolutevalues.
FlowDirectionandfrequencyoutputs
FrequencyoutputsareaffectedbyFlowDirectiononlyifFrequencyOutputProcessVariableissettoaow variable.FrequencyoutputlevelsfordifferentcombinationsofFlowDirectionandactualowdirection areshowninT able5-5.
Table5-5EffectoftheFlowDirectionparameterandactualflowdirectiononfrequencyoutputs
Actualflowdirection
FlowDirectionsettingForwardZeroflowReverse
ForwardHz>00Hz0Hz
Reverse0Hz0HzHz>0
BidirectionalHz>00HzHz>0
AbsoluteV alueHz>00HzHz>0
NegateForwardZero
NegateBidirectionalHz>00HzHz>0
(1)
0HzHz>0
(1)Refertothedigitalcommunicationsstatusbitsforanindicationofwhetherowispositiveornegative.
82MicroMotion9739MVDTransmitters
Page 89
Configureprocessmeasurement
FlowDirectionanddiscreteoutputs
DiscreteoutputsareaffectedbyFlowDirectiononlyifDiscreteOutputSourceissettoFlowDirection.Discrete outputstatesfordifferentcombinationsofFlowDirectionandactualowdirectionareshowninT able5-6.
Table5-6EffectoftheFlowDirectionparameterandactualflowdirectionondiscreteoutputs
Actualflowdirection
FlowDirectionsettingForwardZeroflowReverse
Forward
Reverse
Bidirectional
AbsoluteV alue
NegateForward
NegateBidirectional
OFFOFFON
OFFOFFON
OFFOFFON
OFFOFFOFF
ONOFFOFF
ONOFFOFF
FlowDirectionanddigitalcommunications
DigitalcommunicationsvaluesfordifferentcombinationsofFlowDirectionandactualowdirectionare showninT able5-7.
Table5-7EffectoftheFlowDirectionparameterandactualflowdirectiononflowvaluesreported
viadigitalcommunications
Actualflowdirection
FlowDirectionsettingForwardZeroflowReverse
ForwardPositive0Negative
ReversePositive0Negative
BidirectionalPositive0Negative
AbsoluteV aluePositive0Positive
NegateForwardNegative0Positive
NegateBidirectionalNegative0Positive
FlowDirectionandflowtotals
TotalizerandinventorybehaviorsfordifferentcombinationsofFlowDirectionandactualowdirection areshowninT able5-8.
ConfigurationandUseManual
83
Page 90
Configureprocessmeasurement
Table5-8EffectoftheFlowDirectionparameterandactualflowdirectiononflowtotals
Actualflowdirection
FlowDirectionsettingForwardZeroflowReverse
ForwardTotalsincreaseTotalsdonotchangeTotalsdonotchange
ReverseTotalsdonotchangeTotalsdonotchangeT otalsincrease
BidirectionalTotalsincreaseTotalsdonotchangeT otalsdecrease
AbsoluteV alueTotalsincreaseTotalsdonotchangeT otalsincrease
NegateForwardTotalsdonotchangeTotalsdonotchangeT otalsincrease
NegateBidirectionalT otalsdecreaseTotalsdonotchangeT otalsincrease

5.6Configuredensitymeasurement

Display
ProLinkII
FieldCommunicatorCongure→ManualSetup→Measurements→Density
OFF-LINEMAINT→OFF-LINECONFG→UNITS→DENS
ProLink→Conguration→Density
Thedensitymeasurementparameterscontrolhowdensityismeasuredandreported.
Thedensitymeasurementparametersinclude:
DensityMeasurementUnit
•Theslugowparameters:
SlugHighLimit
SlugLowLimit
SlugDuration
5.6.1ConfigureDensityMeasurementUnit
Display
ProLinkII
FieldCommunicatorCongure→ManualSetup→Measurements→Density→DensityUnit
OFF-LINEMAINT→OFF-LINECONFG→UNITS→DENS
ProLink→Conguration→Density→DensityUnits
DensityMeasurementUnitspeciestheunitthatwillbeusedfordensitymeasurement.
Procedure
SetDensityMeasurementUnittothedesiredoption.
ThedefaultsettingforDensityMeasurementUnitisg/cm3(gramspercubiccentimeter).
OptionsforDensityMeasurementUnit
ThetransmitterprovidesastandardsetofunitsforDensityMeasurementUnit.Differentcommunications toolsusedifferentlabels.
OptionsforDensityMeasurementUnitareshowninTable5-9.
84
MicroMotion9739MVDTransmitters
Page 91
Table5-9OptionsforDensityMeasurementUnit
Label
UnitdescriptionDisplayProLinkIIFieldCommunicator
Specicgravityunit(not temperature-corrected)
Gramspercubic centimeter
GramsperliterG/Lg/lg/L
GramspermilliliterG/mLg/mlg/mL
Kilogramsperliter
Kilogramspercubicmeter
PoundsperU.S.gallonLB/GALlbs/Usgallb/gal
PoundspercubicfootLB/CUFlbs/ft3lb/Cuft
Poundspercubicinch
APIgravityDAPIdegAPIdegAPI
ShorttonpercubicyardST/CUYsT/yd3ST on/Cuyd
SGUSGUSGU
G/CM3g/cm3g/Cucm
KG/Lkg/lkg/L
KG/M3kg/m3kg/Cum
LB/CUIlbs/in3lb/CuIn
Configureprocessmeasurement
5.6.2Configureslugflowparameters
DisplayNotavailable
ProLinkII
FieldCommunicatorCongure→ManualSetup→Measurements→Density→SlugLowLimit
Theslugowparameterscontrolhowthetransmitterdetectsandreportstwo-phaseow.
Procedure
1.SetSlugLowLimittothelowestdensityvaluethatisconsiderednormalinyourprocess.
Valuesbelowthiswillcausethetransmittertoperformtheconguredslugowaction.Typically, thisvalueisthelowestdensityvalueinthenormalrangeofyourprocess.
Tip
Gasentrainmentcancauseyourprocessdensitytodroptemporarily.T oreducetheoccurrence ofslugowalarmsthatarenotsignicanttoyourprocess,setSlugLowLimitslightlybelowyour expectedlowestprocessdensity.
ProLink→Conguration→Density→SlugHighLimit ProLink→Conguration→Density→SlugLowLimit ProLink→Conguration→Density→SlugDuration
Congure→ManualSetup→Measurements→Density→SlugHighLimit Congure→ManualSetup→Measurements→Density→SlugDuration
YoumustenterSlugLowLimiting/cm measurement.
ThedefaultvalueforSlugLowLimitis0.0g/cm
ConfigurationandUseManual
3
,evenifanotherunithasbeenconguredfordensity
3
.Therangeis0.0g/cm
3
to10.0g/cm
3
.
85
Page 92
Configureprocessmeasurement
2.SetSlugHighLimittothehighestdensityvaluethatisconsiderednormalinyourprocess.
Tip
Toreducetheoccurrenceofslugowalarmsthatarenotsignicanttoyourprocess,setSlugHigh Limitslightlyaboveyourexpectedhighestprocessdensity.
Valuesabovethiswillcausethetransmittertoperformtheconguredslugowaction.Typically , thisvalueisthehighestdensityvalueinthenormalrangeofyourprocess.
YoumustenterSlugHighLimiting/cm
3
,evenifanotherunithasbeenconguredfordensity
measurement.
ThedefaultvalueforSlugHighLimitis5.0g/cm
3
.Therangeis0.0g/cm
3
to10.0g/cm
3
3.SetSlugDurationtothenumberofsecondsthatthetransmitterwillwaitforaslugowconditionto clearbeforeperformingtheconguredslugowaction.
ThedefaultvalueforSlugDurationis0.0seconds.Therangeis0.0secondsto60.0seconds.
.
Slugflowdetectionandreporting
Slugowistypicallyusedasanindicatoroftwo-phaseow(gasinaliquidprocessorliquidinagas process).Two-phaseowcancauseavarietyofprocesscontrolissues.Byconguringtheslugow parametersappropriatelyforyourapplication,youcandetectprocessconditionsthatrequirecorrection.
Tip
Todecreasetheoccurrenceofslugowalarms,lowerSlugLowLimitorraiseSlugHighLimit.
AconditionofslugowoccurswheneverthemeasureddensitygoesbelowSlugLowLimitorabove SlugHighLimit.Ifthisoccurs:
•Aslugowalarmispostedtotheactivealarmlog.
•Alloutputsthatareconguredtorepresentowrateholdtheirlast“pre-slugow”valueforthe conguredSlugDuration.
IftheslugowconditionclearsbeforeSlugDurationexpires:
•Outputsthatrepresentowratereverttoreportingactualow.
•Theslugowalarmisdeactivated,butremainsintheactivealarmloguntilitisacknowledged.
IftheslugowconditiondoesnotclearbeforeSlugDurationexpires,outputsthatrepresentowrate reportaowrateof0.
IfSlugDurationissetto0.0seconds,outputsthatrepresentowratewillreportaowrateof0assoonas slugowisdetected.
86
MicroMotion9739MVDTransmitters
Page 93
Configureprocessmeasurement
5.6.3ConfigureDensityDamping
DisplayNotavailable
ProLinkII
FieldCommunicatorCongure→ManualSetup→Measurements→Density→DensityDamping
Dampingisusedtosmoothoutsmall,rapiductuationsinprocessmeasurement.TheDampingValue speciesthetimeperiod(inseconds)overwhichthetransmitterwillspreadchangesinthereported processvariable.Attheendoftheinterval,thereportedprocessvariablewillreect63%ofthechange intheactualmeasuredvalue.
Tips
•Ahighdampingvaluemakestheprocessvariableappearsmootherbecausethereportedvalue mustchangeslowly .
•Alowdampingvaluemakestheprocessvariableappearmoreerraticbecausethereported valuechangesmorequickly.
ProLink→Conguration→Density→DensityDamping
Procedure
SetDensityDampingtothedesiredvalue.
Thedefaultvalueis1.6seconds.Therangeis0to10.24seconds.WhenyouenteravalueforFlow
Damping,thetransmitterautomaticallyroundsitdowntothenearestvalidvalue.ThevalidvaluesforFlow Dampingare:0,0.04,0.08,0.16,...10.24.
EffectofDensityDampingonvolumemeasurement
DensityDampingaffectsliquidvolumemeasurement.DensityDampingdoesnotaffectgasstandardvolume
measurement.
InteractionbetweenDensityDampingandAddedDamping
DensityDampingcontrolstherateofchangeinthedensityprocessvariable.AddedDampingcontrolsthe rateofchangereportedviathemAoutput.IfmAOutputProcessVariableissettoDensity,andbothDensity DampingandAddedDampingaresettonon-zerovalues,densitydampingisappliedrst,andtheadded
dampingcalculationisappliedtotheresultoftherstcalculation.
5.6.4ConfigureDensityCutoff
DisplayNotavailable
ProLinkII
FieldCommunicatorCongure→ManualSetup→Measurements→Density→DensityCutoff
ProLink→Conguration→Density→LowDensityCutoff
DensityCutoffspeciesthelowestdensityvaluethatwillreportedasmeasured.Alldensityvaluesbelow thiscutoffwillbereportedas0.
ConfigurationandUseManual
87
Page 94
Configureprocessmeasurement
Procedure
SetDensityCutofftothedesiredvalue.
ThedefaultvalueforDensityCutoffis0.2g/cm
3
.Therangeis0.0g/cm
3
to0.5g/cm
3
EffectofDensityCutoffonvolumemeasurement
DensityCutoffaffectsliquidvolumemeasurement.IfthedensityvaluegoesbelowDensityCutoff,thevolume
owrategoesto0.DensityCutoffdoesnotaffectgasstandardvolumemeasurement.Gasstandard volumevaluesarealwayscalculatedfromthemeasureddensityvalue.

5.7Configuretemperaturemeasurement

.
Display
ProLinkII
FieldCommunicatorCongure→ManualSetup→Measurements→Temperature
OFF-LINEMAINT→OFF-LINECONFG→UNITS→TEMP
ProLink→Conguration→Density→LowDensityCutoff
Thetemperaturemeasurementparameterscontrolhowtemperaturedatafromthesensorisreported. Temperaturedataisusedtocompensateowmeasurementfortheeffectoftemperatureonthesensor tubes.
Thetemperaturemeasurementparametersinclude:
TemperatureMeasurementUnit
TemperatureDamping
5.7.1ConfigureTemperatureMeasurementUnit
Display
ProLinkII
FieldCommunicatorCongure→ManualSetup→Measurements→Temperature→TemperatureUnit
TemperatureMeasurementUnitspeciestheunitthatwillbeusedfortemperaturemeasurement.
OFF-LINEMAINT→OFF-LINECONFG→UNITS→TEMP
ProLink→Conguration→Temperature→T empUnits
Procedure
SetTemperatureMeasurementUnittothedesiredoption.
ThedefaultsettingisDegreesCelsius.
Tip
IfyouareconguringthemAinputtoreceivetemperaturedatafromanexternalmeasurementdevice, youmustsetthemeasurementunittomatchthetemperaturemeasurementunitattheexternal measurementdevice.
OptionsforT emperatureMeasurementUnit
ThetransmitterprovidesastandardsetofunitsforTemperatureMeasurementUnit.Differentcommunications toolsusedifferentlabels.
88
MicroMotion9739MVDTransmitters
Page 95
OptionsforTemperatureMeasurementUnitareshowninT able5-10.
Table5-10OptionsforT emperatureMeasurementUnit
Label
DescriptionDisplayProLinkIIFieldCommunicator
DegreesCelsius°CdegCdegC
DegreesFahrenheit
DegreesRankine
Kelvin
°F
°R
°K
degFdegF
degRdegR
degKKelvin
5.7.2ConfigureTemperatureDamping
DisplayNotavailable
ProLinkII
FieldCommunicatorCongure→ManualSetup→Measurements→Temperature→TempDamping
ProLink→Conguration→Temperature→T empDamping
Configureprocessmeasurement
Dampingisusedtosmoothoutsmall,rapiductuationsinprocessmeasurement.TheDampingValue speciesthetimeperiod(inseconds)overwhichthetransmitterwillspreadchangesinthereported processvariable.Attheendoftheinterval,thereportedprocessvariablewillreect63%ofthechange intheactualmeasuredvalue.
Tips
•Ahighdampingvaluemakestheprocessvariableappearsmootherbecausethereportedvalue mustchangeslowly .
•Alowdampingvaluemakestheprocessvariableappearmoreerraticbecausethereported valuechangesmorequickly.
Procedure
EnterthedesiredvalueforTemperatureDamping.
Thedefaultvalueis4.8seconds.Therangeis0.0secondsto38.4seconds.
WhenyouenteravalueforT emperatureDamping,thetransmitterautomaticallyroundsitdowntothe nearestvalidvalue.ValidvaluesforTemperatureDampingare0,0.6,1.2,2.4,4.8,…38.4.
EffectofTemperatureDamping
TemperatureDampingaffectstheresponsespeedfortemperaturecompensation.Temperature
compensationadjustsprocessmeasurementtocompensatefortheeffectoftemperatureonsensor tubestiffness.
ConfigurationandUseManual
89
Page 96
Configureprocessmeasurement
TemperatureDampingaffectspetroleummeasurementprocessvariablesonlyifthetransmitteriscongured tousetemperaturedatafromthesensor.Ifanexternaltemperaturevalueisusedforpetroleum measurement,TemperatureDampingdoesnotaffectpetroleummeasurementprocessvariables.
TemperatureDampingaffectsconcentrationmeasurementprocessvariablesonlyifthetransmitteris conguredtousetemperaturedatafromthesensor.Ifanexternaltemperaturevalueisusedfor concentrationmeasurement,TemperatureDampingdoesnotaffectconcentrationmeasurementprocess variables.

5.8Configurepressurecompensation

DisplayNotavailable
ProLinkII
FieldCommunicatorCongure→ManualSetup→Measurements→ExternalCompensation
Pressurecompensationadjustsprocessmeasurementtocompensateforthepressureeffectonthe sensor’sowtubes.Pressureeffectisdenedasthechangeinthesensor’ssensitivitytoowand densityassociatedwiththedifferencebetweencalibrationpressureandprocesspressure.
ProLink→Conguration→Pressure→PressureCompensation
Tip
Notallsensorsorapplicationsrequirepressurecompensation.Ifyouareuncertainaboutimplementing pressurecompensation,contact
ow.support@emerson.com.
Procedure
1.Enablepressurecompensation.
2.EnterFlowFactorforyoursensor.
FlowFactoristhepercentchangeintheowrateperPSI.FlowFactorforyoursensorisprovidedon thesensorproductdatasheet.Whenenteringthevalue,reversethesign.
Example:Iftheowfactoris0.000004%perPSI,enter−0.000004%perPSI.
3.EnterDensityFactorforyoursensor.
DensityFactoristhechangeinuiddensity,ing/cm thesensorproductdatasheet.Whenenteringthevalue,reversethesign.
Example:Ifthedensityfactoris0.000006g/cm
4.EnterCalibrationPressureforyoursensor.
3
/PSI.DensityFactorforyoursensorisprovidedon
3
/PSI,enter−0.000006g/cm3/PSI.
90MicroMotion9739MVDTransmitters
Page 97
Configureprocessmeasurement
CalibrationPressureisthepressureatwhichyoursensorwascalibrated,andthereforedenesthe pressureatwhichtherewillbenopressureeffect.CalibrationPressureforyoursensorisprovidedon thesensorcalibrationsheet.Ifthedataisunavailable,enter20PSI.
5.Decidehowpressuredatawillbeprovidedtothetransmitter,andperformtherequiredsetup.
•Ifyouwillpollanexternalpressuredevice,setuppollingforpressure.
•Ifyouwilluseastaticpressurevalue,setPressureUnitstotheunitsyouareusing,enter ExternalPressure,andensurethatpollingforpressureisdisabled.
•Ifyouwillusedigitalcommunicationsoradirectanalogcurrenttowritepressuredatatothe transmitter,setPressureUnitstotheunitstobeusedandensurethatpollingforpressure isdisabled.Thenensurethattheappropriatevalueiswrittentotransmittermemoryat appropriateintervals.
•Ifyouwilluseanexternalmeasurementdevice,congurethemAinputtoExternalPressure. YoumustalsoenableExternalPressureCompensationandsetthePressureUnitstotheunitssetat theexternalmeasurementdevice.

5.9Configurethepetroleummeasurementapplication

DisplayNotavailable
ProLinkII
FieldCommunicatorCongure→ManualSetup→Measurements→PetroleumMeasurement
Thepetroleummeasurementparameterscontrolthevaluesthatwillbeusedinthetransmitter’s petroleummeasurementapplication.
Thepetroleummeasurementparametersinclude:
APITableType
ThermalExpansionCoefcient(TEC)(ifrequiredbyAPITableType)
ReferenceT emperature(ifrequiredbyAPITableType)
Restriction
Thepetroleummeasurementparametersareavailableonlyifthepetroleummeasurementapplicationhas beenpurchasedandisenabledonyourtransmitter.
ProLink→Conguration→APISetup
Procedure
1.SelectAPITableType.
2.IfyousetAPITableTypeto53A,53B,53D,or54C,setReferenceTemperaturetotheappropriatevaluefor yourapplication.Enterthevaluein°C.
3.IfyousetAPIT ableTypeto6C,24C,or54C,setThermalExpansionCoefcienttotheappropriatevalue foryourapplication.
4.(Optional)Setthetemperatureunitconguredonthetransmittertothetemperatureunitusedby yourAPIreferencetable.
ConfigurationandUseManual
91
Page 98
Configureprocessmeasurement
Tip
AlthoughconguringthetemperatureunittomatchthetemperatureunitsusedbyyourAPI referencetableisnotrequired,
MicroMotionrecommendsit.
5.(Optional)Ifyouwanttousetemperaturedatafromanexternaltemperaturesensor:
a.SetTemperatureSourcetoExternal.
b.Dependingonyourexternalsetup,dooneofthefollowing:
•Setuppollingfortemperature
•Usedigitalcommunicationstowritetemperaturedatatothesensoratappropriateintervals
•CongurethemAinputtoreceivetemperaturedatafromanexternalmeasurementdevice
Youcannowcongureyourtransmittertoreportandhandlepetroleummeasurementprocessvariables inthesamewaythatitreportsandhandlesotherprocessvariables.
5.9.1Petroleummeasurementapplication
ThepetroleummeasurementapplicationenablesCorrectionfortheeffectofTemperatureonvolumeof Liquids(CTL),bycalculatingandapplyingaVolumeCorrectionFactor(VCF)tovolumemeasurement. InternalcalculationsareperformedincompliancewithAmericanPetroleumMeasurement(API) standards.
APIreferencetablesareusedtocontrolhowCTLiscalculated.Y ourselectionofAPIT ableTypespecies thetypeofprocessuidthatthecalculationswillassume,andtheCTLsourcedata,thereference temperature,andthedensityunitthatthecalculationswilluse.DependingonyourselectionofAPIT able Type,youmayormaynotneedtospecifyReferenceT emperatureandThermalExpansionCoefcient.See Table5-11foralistingoftheAPIreferencetablesandrelatedinformation.
Table5-11APIreferencetables,associatedprocessfluids,andassociatedcalculationvalues
Reference
TablenameProcessfluidCTLsourcedata
5A
5B
5DLubricatingoils
6C
Generalizedcrude andJP4
Generalized products
Liquidswitha constantdensity baseorknown thermalexpansion coefcient
Observeddensity andobserved temperature
Observeddensity andobserved temperature
Observeddensity andobserved temperature
User-supplied referencedensity(or thermalexpansion coefcient) andobserved temperature
temperatureDensityunit
60°F(non­congurable)
60°F(non­congurable)
60°F(non­congurable)
60°F(non­congurable)
DegreesAPI
Range:0to100
DegreesAPI
Range:0to85
DegreesAPI
Range:−0to+40
DegreesAPI
92MicroMotion9739MVDTransmitters
Page 99
Configureprocessmeasurement
Table5-11APIreferencetables,associatedprocessfluids,andassociatedcalculationvalues
Reference
TablenameProcessfluidCTLsourcedata
23A
23B
23DLubricatingoils
24C
53A
53B
53DLubricatingoils
54C
Generalizedcrude andJP4
Generalized products
Liquidswitha constantdensity baseorknown thermalexpansion coefcient
Generalizedcrude andJP4
Generalized products
Liquidswitha constantdensity baseorknown thermalexpansion coefcient
Observeddensity andobserved temperature
Observeddensity andobserved temperature
Observeddensity andobserved temperature
User-supplied referencedensity(or thermalexpansion coefcient) andobserved temperature
Observeddensity andobserved temperature
Observeddensity andobserved temperature
Observeddensity andobserved temperature
User-supplied referencedensity(or thermalexpansion coefcient) andobserved temperature
temperatureDensityunit
60°F(non­congurable)
60°F(non­congurable)
60°F(non­congurable)
60°F(non­congurable)
15°C(congurable)
15°C(congurable)
15°C(congurable)
15°C(congurable)
continued
Relativedensity
Range:0.6110to
1.0760
Relativedensity
Range:0.6535to
1.0760
Relativedensity
Range:8520to
1.1640
Relativedensity
Basedensity
Range:610to1075
3
kg/m
Basedensity
Range:653to1075
3
kg/m
Basedensity
Range:825to1164
3
kg/m
Basedensityinkg/m
3

5.10Configuretheconcentrationmeasurementapplication

DisplayNotavailable
ProLinkII
FieldCommunicatorCongure→ManualSetup→Measurements→ConcentrationMeasurement
Theconcentrationmeasurementparameterscontrolhowthetransmittercalculatesconcentration fromtemperatureanddensitydata.
ConfigurationandUseManual
ProLink→Conguration→CMSetup
93
Page 100
Configureprocessmeasurement
Theconcentrationmeasurementparametersinclude:
ActiveCurve
DerivedVariable
Restriction
Theconcentrationmeasurementparametersareavailableonlyiftheconcentrationmeasurement applicationhasbeenpurchasedandisenabledonyourtransmitter.
Prerequisites
Beforeyoucancongureconcentrationmeasurement:
•Theconcentrationmeasurementapplicationmustbeenabledonyourtransmitter.
•Thecurveyouwanttousemustbeavailableonyourtransmitter.
Note
Curvescanbemadeavailableonyourtransmittereitherbyloadinganexistingcurveorbyconguring anewcurve.Uptosixcurvescanbeavailableonyourtransmitter,butonlyonecanbeused formeasurementatanygiventime.SeeMicroMotionEnhancedDensityApplication:Theory, Conguration,andUseManualforinformationonloadingorconguringacurve.
Procedure
1.Identifythecurveyouwanttouse.
2.SetDensityMeasurementUnittomatchthedensityunitusedbyyourcurve.
3.SetTemperatureMeasurementUnittomatchthetemperatureunitusedbyyourcurve.
4.SetDerivedVariabletooneofthederivedvariablesavailablewithyourcurve.
Tip
SelectaDerivedVariablethatwillprovidetheconcentrationmeasurementprocessvariablesthatyou wanttouse.Ifyouareusingoneofthestandardcurvesfrom MassConc(Dens).Ifyouareusingacustomcurve,seethereferenceinformationforyourcurve.
5.SetActiveCurvetothecurveyouidentiedinStep1.
6.(Optional)Ifyouwanttheconcentrationmeasurementapplicationtousetemperaturedatafroman externaltemperaturesensor:
a.SetTemperatureSourcetoExternal.
b.Setuppollingfortemperature,orcongurethemAinputtoreceivetemperaturedatafrom
anexternalmeasurementdevice.
Youcannowcongureyourtransmittertoreportandhandleconcentrationprocessvariablesinthe samewaythatitreportsandhandlesotherprocessvariables.
MicroMotion,setDerivedVariableto
94
MicroMotion9739MVDTransmitters
Loading...